Pacemaker Troubleshooting

• Implant related Troubleshooting

• Lead related Troubleshooting

• Pacemaker malfunction

Page 3: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Early complications of pacemaker

implantation

– Pneumothorax/Hemothorax/vascular

hemorrhage/AIR Emboli

– SVT, VT/ Cardiac arrest

– Lead dislodgement/Lead perforation

Page 4: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pneumothorax

• Absence of lung markings over the lung

field ipsilateral to the pacemaker pocket

assessed from the fluoroscopy or pre-

discharge x-ray.

• Non-puncture related, might occur at

contralateral side

Page 5: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pneumothorax Sign

• Dyspnea(80~100%)

• Chest pain(75~90%)

• Dry cough(25~35%)

• Hypotension、Tachycardia、SaO2 ↓

Page 6: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pneumothorax

• 0.66% (190/28,860 patients) in Danish Pacemaker

Register

– more often in women [OR 1.9],

– age >80 years [OR 1.4],

– prior history of chronic obstructive pulmonary disease

[OR 3.9]

– implantation of a dual-chamber PM [OR 1.5]

– venous access with subclavian vein puncture [OR 7.8]

– venous access with both subclavian vein puncture and

cephalic vein cut-down [OR 5.7]

– implantation in a non-university center [OR 2.1].

Page 7: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Old lady, kyphoscoliosis,

chronic obstructive pulmonary disease

Page 8: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

venous access with

subclavian vein puncture

Page 9: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

How to avoid pneumothorax

• The cephalic vein cut-down technique

should be applied whenever possible to

avoid this complication.

Page 10: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pneumothorax nursing care

• Administer oxygen as prescribed.

• Position the client in high fowler’s position.

• Prepare for chest tube placement until the lung

has expanded fully.

• Monitor chest tube drainage system.

Page 11: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Air Emboli

Page 12: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Air Emboli

Page 13: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Air Emboli

• More occurs in

– Un-cooperated patients

– Under respiratory distress

– Old age

– Snoring patients

• Management

– IV resuscitation

– Raise patients’ legs

– Increase FiO2

Page 14: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

SVT / VT during implantation

• Check Vital sign

• Stable Medication

RV lead pacing

• Unstable Cardioversion

Page 15: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead perforation / cardiac

tamponade

• Rising stimulus threshold, RBBB

morphology

• Intercostal diaphragmatic pacing

• Hypotension

Page 16: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead perforation / cardiac

tamponade

• Echo

• Cardiocentesis

Page 17: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Rare but it happened sometimes

Page 18: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Hematoma formation at pulse

generator / due to

anticoagulants

Page 19: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Anticoagulation therapy

• Warfarin was temporarily discontinued before device

implantation when possible to achieve an INR value

of < 1.7

• Administration of LMWH was stopped 24 h before the

procedure

• Antiplatelet therapy with ASA or clopidogrel was

allowed to continue

• Treatment with warfarin was resumed after 24 h and

with LMWH after 12–24 h

Page 20: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Cardiac arrest

• TPM pacing

• TCP pacing

• RV lead pacing

Page 21: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead related complications

1. Lead dislodgement

Atrial > Ventricular

2. Lead fracture

3. Loss of integrity of insulation

Page 22: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead failure

• Development of high pacing thresholds

or sensing problems resulting in the

need to program the device to a

different pacing mode or the need for

reoperation.

Page 23: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區
Page 24: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead dislocation

Page 25: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Atrial lead dislodge more

frequently

Page 26: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead Dislodgement

Page 27: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead Dislodgement

Diagnostic features

– changes in the morphology of capture beats

– changes in dipole of the pacing stimulus

– changes in the lead position identified on a chest radiograph

Page 28: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Lead Dislodgment

Treatment – surgical intervention to reposition the lead

• an adequate heel on the intracardiac portion of the lead

• look for a 2 to 3mV current of injury pattern

• electrical and mechanical stability of the lead may be assessed

– Twiddler’s syndrome • the portion of the lead within the pocket should be

carefully inspected.

• If damage to the conductor coil or insulation is noted, the lead should not be reused.

Page 29: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Order a Chest X-ray The chest x-ray revealed a dislodged lead

Page 30: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pacemaker system malfunction-

Troubleshooting

Page 31: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pacing Stimuli Present with Failure to

Capture - causes

Lead dislodgment • Early: unstable position

• Late: Twiddler’s syndrome

Lead maturation • Early: inflammatory response

• Late: progressive fibrosis

Late high thresholds • Progressive fibrosis

• Myocardial infarction

• Cardiomyopathy

• Metabolic/drugs

• Damaged lead or tissue interface

Insulation failure Conductor failure

• Lead fracture • Loose set-screw

Battery depletion Functional non capture

• Pseudomalfunction

Page 32: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Electrical stimuli delivered by

the pacemaker do not initiate

depolarization of the atria or

ventricle

Loss of Capture

Page 33: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Loss of Capture

Possible Causes Corrective Measures •Threshold rise •Increase output (mA)/check thresholds

•Fractured/dislodged lead •Replace/reposition lead

•Battery depletion •Replace battery

•QRS not visible •Adjust ECG

•Tissue is refractory •Assess mode selection

•Faulty cable connections •Check connections

•Switch polarity (epicardial system)

Page 34: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Capture

Loss of Ventricular Capture

Atrial/Ventricular Stimulation Thresholds

Page 35: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Pacemaker fails to emit stimuli

at the programmed intervals

No Output

Page 36: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

No Output

Possible Causes Corrective Measures •Battery depletion •Replace battery

•Pacemaker OFF •Verify pacemaker settings

•Faulty cable connections •Check cable connections

•Fractured/dislodged lead •Replace/reposition lead

•Oversensing •Verify/adjust sensitivity

Page 37: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Failure of the pacemaker to sense

intrinsic R-waves or intrinsic

P-waves

Undersensing

Page 38: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Undersensing

Possible Causes Corrective Measures •Decreased QRS voltage •Increase sensitivity

•Fractured/dislodged lead •Replace/reposition Lead

•Battery depletion •Replace battery

•Inappropriate sensitivity setting •Sensing test/increase sensitivity

•Fusion beat

Page 39: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Sensing

Atrial Undersensing

Atrial/Ventricular Sensing Thresholds

Page 40: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Undersensing . . .Overpacing • Pacemaker does not “see” the intrinsic

beat, and therefore does not respond

appropriately

Intrinsic beat

not sensed

Scheduled pace

delivered

VVI / 60

Page 41: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Inhibition of the pacemaker by events

pacemaker should ignore, e.g. EMI,

T-waves and myopotentials

Oversensing

Page 42: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Oversensing

Possible Causes Corrective Measures •Fractured/dislodged lead •Replace/reposition lead

•Environmental interference •Eliminate interference

•T-wave oversensing •Sensing test/decrease sensitivity

•Faulty cable connections •Check connections

Page 43: 心臟植入性電子儀器(CIED )護理照護指引- Cathroom Troubleshooting_20130914中區

Oversensing …Underpacing

• An electrical signal other than the

intended P or R wave is detected Marker channel

shows intrinsic

activity...

...though no

activity is present

VVI / 60