To improve productivity and lower the capital cost, a single gas chromatograph has been configured to perform two standard methods: CAN/CGSB-3.0 and ASTM D-5580. Detailed hydrocarbon (DHA) analysis of automotive gasoline has become a routine gas chromatographic method. The Canadian standard CAN/CGSB-3.0 requires cryogenic temperature programing to provide better separation of early-eluting components than conventional non-cryogenic DHA measurements. Determination of aromatics in the presence of methanol and ethanol is also an important requirement for bio-ethanol; method ASTM D-5580 is the standard method for this analysis.

TRANSCRIPT

Page 1: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Zhuangzhi ‘Max’ Wang, PhD, Richard Whitney, PhD, Clifford M. Taylor, Nicole Lock, Laura Chambers, GCMS Product Manager, Shimadzu Scientific Instruments, Columbia, MD, USA, 800-477-1227, www.ssi.shimadzu.com

Page 2: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

To improve productivity and lower the capital cost, a single gas chromatograph has been configured to perform two standard methods: CAN/CGSB-3.0 and ASTM D-5580.

Detailed hydrocarbon (DHA) analysis of automotive gasoline has become a routine gas chromatographic method. The Canadian standard CAN/CGSB-3.0 requires cryogenic temperature programing to provide better separation of early-eluting components than conventional non-cryogenic DHA measurements. Determination of aromatics in the presence of methanol and ethanol is also an important requirement for bio-ethanol; method ASTM D-5580 is the standard method for this analysis.

In this work, the two applications have been successfully integrated into one GC. The methods of three tests – one for CAN/CGSB-3.0 and two for ASTM D-5580, can be switched smoothly with minimum effort. Automated software control provides separation of hydrocarbons in groups and reporting of the results for method CAN/CGSB-3.0. Repeatability (%RSD) of ASTM D-5580 was less than 1.2% for analysis No. 1, and less than 1.8% for analysis No. 2.

This configuration is not limited to motor gasoline analysis though; it can be applied to other finished products such as blend alcohol and other new green energy products.

Introduction

Page 3: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Instrumentation

Columns: 20% TCEP, 0.558m X

0.75mm Rtx-1 30m X 0.53mm X

5µm

LN2 Cryo-cooling

Injection Volume: 1 µL Valve box Temp: 80C APC1 Pressure: 210.0KPa GCSolution ver. 2.41 PONASolution ver. 2.60

Rtx-1

TCEP

PONA

GC-2010Plus

Page 4: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

ObjectiveSeparation of hydrocarbons into types

Paraffins, Olefins, Naphthene, Aromatics

Separation of each type according to carbon number

MethodSingle capillary column separation of all components, followed by grouping with a post-run program - PONA Solution

Canadian PONA Method CAN/CGSB-3.0

Page 5: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Column: Rtx-1 PONA 100mX0.25mmX0.5µmSample: Premium GasCarrier Gas: He, Linear velocity: 25.6cm/secSplit ratio: 270, Gas Saver: On (split ratio 270 to 5

after 5min) Injector temperature: SPL 275˚COven Temperature: 0˚C hold 15min, 1˚C/min to

50˚C, 2˚C/min to 130˚C, 4˚C/min to 270˚CDetector: FID 300 ˚C Injection volume: 0.5 µL

Canadian PONA Experimental

Page 6: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Chromatogram

5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 60.0 65.0 70.0 75.0 80.0 85.0 90.0 95.0 100.0 105.0 110.0 115.0 120.0 125.0 130.0 135.0 min-0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0uV(x10,000) Chromatogram

0.5L injection of premium gas

Page 7: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Manual Chromatogram Quantification

All components are identified and classified

Page 8: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Canadian PONA ReportSample name Premium GasSample ID Premium GasOperater AdminID table CanadianPONAComments Premium GasSampling date 2/21/2012Analysis date 4/6/2011Identified peaks [323/367]Total area 19776176

Hydrocarbon Type % by Mass % by Mole % by VolumeParaffins 3.81 3.80 4.46Isoparaffins 39.31 39.49 42.53Olefins 11.37 11.66 12.14Naphthenes 4.97 5.09 4.79Aromatics 29.53 31.95 25.11Oxygenates 11.02 8.01 10.99

Density(15deg) 0.75 g/cm3Average molecular weight 99.34Reseach Octane Number 89.79Vapor pressure(37.78deg) 41.56 kPa

Peak NameHydrocarbon

TypeRetention

Time % by Mass % by Mole % by VolumeIsobutane P 8.66 0.09 0.091 0.1211-Butene X 9.44 0.019 0.018 0.023Isobutene O 9.48 0.013 0.013 0.016n-Butane A 9.83 0.88 0.889 1.132trans-2-Butene X 10.3 0.054 0.053 0.064cis-2-Butene O 11.02 0.062 0.06 0.0723-Methyl-1-butene O 13.15 0.087 0.085 0.1Isopentane O 14.71 5.616 5.636 6.721-Pentene P 16.34 0.262 0.256 0.2952-Methyl-1-butene O 17.17 0.451 0.44 0.5n-Pentane P 17.68 1.337 1.342 1.585trans-2-Pentene P 18.78 1.042 1.017 1.161cis-2-Pentene O 19.79 0.495 0.483 0.5442-Methyl-2-butene N 20.44 1.267 1.236 1.379"2,2-Dimethylbutane" O 22.51 0.406 0.405 0.461Cyclopentene O 25.07 0.205 0.195 0.1864-Methyl-1-pentene O 26.24 0.038 0.038 0.0423-Methyl-1-pentene O 26.32 0.069 0.067 0.074Cyclopentane P 26.85 0.254 0.248 0.247"2,3-Dimethylbutane" O 27.58 1.397 1.395 1.561Methyl-tert-butylether O 27.98 7.979 10.963 10.937"2,3-Dimethyl-1-butene" P 28.45 0.031 0.03 0.0332-Methylpentane P 28.63 2.516 2.513 2.846trans-4-Methyl-2-pentene P 29.08 0.133 0.13 0.1443-Methylpentane X 31.17 1.546 1.543 1.7192-Methyl-1-pentene A 32.33 0.194 0.19 0.2061-Hexene X 32.55 0.087 0.085 0.094

Page 9: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

ASTM D5580-02

Scope:Determination of benzene, toluene, ethylbenzene, xylenes, C9 and heavier aromatics, and total aromatics

Sample:Finished motor gasoline

Page 10: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Two Step Analysis

No.1Determination of benzene and toluene, with 2-Hexanone as internal standard; backflush C8+ aromatics and hydrocarbons

No.2Determination of ethylbenzene, xylenes, and C9+ aromatics, with 2-Hexanone as internal standard

Page 11: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

ASTM D-5580 Experimental

INJ Temp: 200 C, Carrier Gas: He, Flow Control Mode:

Pressure, INJ Pressure: 160KPa, TCEP Column Flow: 10

mL/min, Rtx-1 column flow: 10 mL/min, Split vent flow: 100

mL/min, Purge Flow: 1mL/min, Split Ratio: 11:1

Valve Box Temp: 80.0C

Oven1 Temp: 40.0 C hold 6 min, 2 C /min to 80 C, 3

C/min to 115 C hold 2.33 min

FID Temp: 250 C, Makeup Gas: He, Makeup Flow:

20.0mL/min, H2 Flow: 40.0mL/min, Air Flow: 400.0mL/min.

TCD Temp: 200 C, Makeup Gas: He, Makeup Flow: 8

mL/min, Current: 60 mA

T1: 1.8 min, T2: 6.2 min, T3: 21.0 min, T4: 30.0 min

Page 12: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 min

-0.1

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

1.1

1.2

uV(x10,000) Chromatogram

Ben

zene

Tolu

ene

Eth

ylbe

nzen

e

o-X

ylen

e

2-H

exan

one

TCD Chromatogram

T1 T2

Page 13: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Analysis No.1

2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0 32.5 35.0 37.5 min

-0.25

0.00

0.25

0.50

0.75

1.00

1.25

1.50

1.75

2.00

2.25

2.50

2.75

3.00

uV(x1,000,000) Chromatogram

Be

nz

en

e

To

lue

ne

2-H

ex

an

on

e

Backflush PeakC8+ aromatics and hydrocarbons

Page 14: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Calibration Curves: Benzene & Toluene

Page 15: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Premium Gas Analysis No.1

0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0 32.5 35.0 37.5 min

-0.1

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

1.1

1.2

uV(x10,000,000)

%RSD < 1.2%

(n=8)

Page 16: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Analysis No.2

2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0 32.5 35.0 37.5 min

-0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

uV(x1,000,000) Chromatogram

2-H

ex

an

on

e

Eth

ylb

en

ze

ne

o-X

yle

ne

1,2

,4-T

rim

eth

ylb

enz

en

e

Page 17: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Calibration Curves Ethylbenzene, o-Xylene, and 1,2,4-Trimethylbenzene

Page 18: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0 32.5 35.0 37.5 min

-0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

uV(x1,000,000)

Chromatogram

Premium Gas Analysis No.2

%RSD < 1.8%

(n=8)

Page 19: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Conclusions

Mechanical valve-based multidimensional GC-2010 Plus has

successfully achieved ASTM D5580 requirements.

Excellent retention time stability allows for the ease of

determining backflush times.

Excellent qualitative and quantitative results.

In addition to ASTM D-5580 method, a separate analytical

line has been incorporated to be able to run Canadian PONA

analysis on the same GC, which has lowered instrumentation

cost and improved productivity.

Page 20: Combining Methods CAN/CGSB-3.0 and ASTM D-5580 in a Single GC Platform

Acknowledgements

Mr. Mohamed Salem and Mr. Patrick Armstrong (Shimadzu R&D group) developed Canadian PONA software.

Mr. Inoye Tataaki and Mr. Ryosuke Kamae (Shimadzu Japan) developed and supported PONASolution software.