comparison of three nirs devices for the measurement of ... · the velocity and degree of flow...
TRANSCRIPT
![Page 1: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/1.jpg)
Academiejaar 2015 – 2016
Comparison of three NIRS devices for the measurement of microvascular reactivity
Kevin Steenhaut
Promotor 1: Prof. dr. Anneliese Moerman Promotor 2: Prof. dr. Stefan De Hert
Masterproef voorgedragen in de master in de specialistische geneeskunde Anesthesie en reanimatie
![Page 2: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/2.jpg)
![Page 3: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/3.jpg)
Academiejaar 2015 – 2016
Comparison of three NIRS devices for the measurement of microvascular reactivity
Kevin Steenhaut
Promotor 1: Prof. dr. Anneliese Moerman Promotor 2: Prof. dr. Stefan De Hert
Masterproef voorgedragen in de master in de specialistische geneeskunde Anesthesie en reanimatie
![Page 4: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/4.jpg)
Toelating tot bruikleen
“De auteur geeft de toelating deze masterproef voor consultatie beschikbaar te stellen en delen van de masterproef te kopiëren voor persoonlijk gebruik. Elk ander gebruik valt onder de beperkingen van het auteursrecht, in het bijzonder met betrekking tot de verplichting de
bron uitdrukkelijk te vermelden bij het aanhalen van resultaten uit deze masterproef.”
“The author gives permission to make this master thesis available for consultation and to copy parts of this master thesis for personal use. In the case of any other use, the limitations
of the copyright have to be respected, in particular with regard to the obligation to state expressly the source when quoting results from this master thesis.”
Gent, 27 april 2016
XKevin Steenhaut
![Page 5: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/5.jpg)
Table of Content
List of abbreviations
Abstract
1. Introduction
1.1 Background
1.2 NIRS: principles of operation
1.3 Assumptions and limitations
1.4 Tissue oxygenation changes during vascular occlusion test
2. Aim
3. Materials and methods
3.1 Subject preparation
3.2 Different NIRS devices
3.3 Interventions
3.4 Missing values
3.5 Selected parameters of microvascular reactivity
3.6 Statistics
4. Results
4.1 Comparison of static parameters between the 3 devices
4.2 Comparison of dynamic parameters between the 3 devices
5. Discussion
5.1 Important findings
5.2 Comparison with previous studies
5.3 Clinical relevance
6. Conclusion
7. References
![Page 6: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/6.jpg)
List of abbreviations
AUC= area under the curve
A/V ratio= arterial/venous ratio
BIS= bispectral index
BL/BSLN= baseline
CABG= coronary artery bypass grafting
CPB= cardiopulmonary bypass
Hb= hemoglobin, HHb= deoxygenated Hb, O2Hb= oxygenated Hb, THb= total Hb
ICU= intensive care unit
IQR= interquartile range
NIRS= near-infrared spectroscopy
PORH= post-occlusive reactive hyperemia
SD= standard deviation
SRS= Spatially Resolved Spectroscopy
StO2= peripheral tissue saturation
M1= first minute
TOI= tissue oxygenation index
VOT= vascular occlusion test
![Page 7: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/7.jpg)
1
Abstract
Comparison of three NIRS devices for the measurement of microvascular reactivity
Kevin Steenhaut MD, Stefan De Hert MD, PhD, Anneliese Moerman MD, PhD
Ghent University Hospital, Dept of Anaesthesiology, Ghent, Belgium
Background and Goal
An increasing number of NIRS devices are used to provide measurements of microvascular
reactivity. The interchangeability of the different devices is however unclear. The aim of the
present study is to analyse tissue oxygenation measurements by three different NIRS devices
during a vascular occlusion test (VOT). The hypothesis is that measurements from the different
devices are similar.
Materials and Methods
Forty consenting adults scheduled for elective CABG surgery were recruited. Three disposable
NIRS sensors (INVOS 5100C; Foresight Elite and NIRO-200NX) were applied to the left
forearm over the brachioradial muscle. A standard blood pressure cuff at the upper arm was
inflated to a pressure of 50 mmHg above the individual systolic pressure. After 3 minutes of
ischaemia, cuff pressure was rapidly released. Tissue oxygenation (StO2) measurements
included baseline StO2 (BL), downsloping rate in first minute (DS-M1) and over 3 minutes
(DS), minimum value (Min), upsloping rate (US), rise time (Rt), maximum value (Max) and
settling time (St).
Comparisons between devices were performed with the Kruskal-Wallis test. Pairwise
differences among devices were examined by the Mann-Whitney U test.
Results and Discussion
There were no significant differences at baseline. Pairwise comparisons between devices
showed that INVOS has significantly higher downsloping rates, lower minimum values and
higher upsloping rates, while NIRO has lower maximum values and Foresight has longer rise
and settling times compared to the two other devices.
Table 1. Comparison of tissue oxygenation measurements during VOT
INVOS Foresight NIRO
BL (%) 66 [61-73] 70 [65-73] 69 [65-73]
DS-M1 (%/min) 17 [13-24]* 11 [6-15] 12 [7-16]
DS (%/min) 15 [11-21]* 11 [8-13] 12 [9-15]
Min (%) 36 [21-48]* 45 [40-51] 46 [36-51]
US (%/min) 311 [92-523]* 114 [65-199]* 202 [88-269]*
Rt (sec) 25 [23-35] 40 [28-50]* 27 [21-31]
Max (%) 82 [77-86] 81 [78-87] 79 [75-82]*
St (sec) 181 [146-223] 226 [181-266]* 187 [127-248]
* p<0.05 vs. the two other devices. Data are presented as median [IQR].
Conclusion
Although no significant differences were found at baseline, analysis of different parameters of
microvascular reactivity shows that different information is retrieved depending on the NIRS
device used. This phenomenon should be kept in mind when using NIRS as monitoring
technique for tissue oxygenation, especially during VOT.
![Page 8: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/8.jpg)
2
1. INTRODUCTION
1.1 Background
Alterations in microvascular perfusion are associated with impaired tissue oxygenation and
organ dysfunction.1 Therefore, assessment of microcirculation could provide the clinician
with information about the tissue oxygenation status, the severity of the disease and the
results of the applied therapies. Parameters derived from the functional evaluation of the
microcirculation by a vascular occlusion test (VOT) have been shown to predict outcome in
septic and trauma patients independently from the macrocirculation.2 A minimum tissue
oxygenation value precedes peak lactate levels by more than 90 min in patients with
hypovolemic shock, indicating that regional tissue perfusion might be an earlier indicator of
perfusion deficits.3 Whereas resting tissue oxygen saturation is largely insensitive to
hypoperfusion, dynamic parameters can provide information regarding tolerance to ischemia
and its recovery potential after ischemia.
Near-infrared spectroscopy (NIRS) is increasingly recognized as a method for non-invasive
assessment of microvascular reactivity. It enables to quantify endothelium-mediated changes
in vascular tone, elicited by creating post-occlusive reactive hyperemia (PORH).4 PORH
refers to the reproducible transient increase in blood flow after release of an arterial occlusion.
The velocity and degree of flow restoration depend on the capacity of the microvasculature to
recruit arterioles and capillaries, thereby reflecting the integrity of the microcirculation.4
An increasing number of NIRS devices are being used in research investigations to provide
measurements of vascular reactivity. However, comparability between the different devices is
unclear as there are several technical differences and every technology applies a different
computational algorithm to generate NIRS values.5 Currently, it is not known if and how well
measures from different devices are related.
![Page 9: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/9.jpg)
3
1.2 NIRS: Principles of operation
The physical and mathematical basis for NIRS is provided by the Beer-Lambert law, which
states that the quantity of light absorbed by a substance (A) is directly proportional to the
specific absorption coefficient of the substance at a particular wavelength (ɛ), the
concentration of the substance (c) and the path length of the light through the solution (l) (A =
ɛ . c . l).
The relative transparency of biological tissues to light in the near-infrared part of the spectrum
(700-1000 nm) enables light photons to pass through the tissues, where they are attenuated
due to a combination of absorption and scattering. Because of scattering by the tissue
components, the light does not travel in a straight line. Therefore, the Modified Beer-Lambert
law is applied: (A = ɛ . c . l . B + k), where B is the differential path length factor and k is an
additive geometry-dependent term, reflecting scatter loss. The geometrical path length l has to
be multiplied by B to find the true optical distance, because light that reaches the detector will
have been scattered multiple times and therefore has travelled a much greater distance than
the actual light emitter-detector distance. K corrects for the fact that not all emitted light
reaches the detector, because some of it is scattered away from the detector, giving scattering
losses. Scattering is a function of the tissue composition and the number of various tissue
interfaces. Because B and k are unknown factors, no absolute values can be measured with
the Modified Beer-Lambert law. NIRS technology is based on the assumption that the
quantity of scattering remains constant and that changes in attenuation result solely from
changes in absorption.
Several biological molecules, termed chromophores, absorb light in the near-infrared
spectrum. However, only hemoglobin and cytochrome oxidase are present in variable
concentrations, reflecting blood and intracellular oxygenation, respectively. Other
chromophores are assumed to be constant over the period of monitoring.
The wavelengths of near-infrared light used in commercial devices are selected to be sensitive
to hemoglobin. Cytochroom oxidase has a crucial role in mitochondrial oxidative energy
metabolism, and therefore provides a potential biomarker of the cellular oxygenation state,
with substantial physiological and clinical importance. However, it is present in much lower
concentrations in the tissue than oxygenated and deoxygenated hemoglobin and its absorption
spectrum overlaps that of these chromophores, and therefore, the validity of cytochroom
![Page 10: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/10.jpg)
4
oxidase measurements is debated, and the signal is not incorporated into any clinical monitors
yet.
Oximetry relies in the fact that absorption of near-infrared light at specific wavelengths is
different in deoxygenated hemoglobin (HHb) when compared with oxygenated hemoglobin
(O₂Hb).
Commercial devices generally use wavelengths between 690 and 880 nm where the
absorption spectra of O₂Hb and HHb are maximally separated and there is minimal overlap
with that of water absorption (980 nm). Optical absorption at 1 wavelength for each
chromophore of interest must be known. NIRS devices use near-infrared light at two or more
specific wavelengths to differentiate between O2Hb and HHb. No attempt is made to measure
optical scattering. The scale of measured changes is dependent on the application of
assumptions of the scattering properties at different wavelengths, and is incorporated into the
algorithm of the respective devices. Algorithmic formulae are complex and their validity is
contingent on the assumptions made. The variability in algorithms between NIRS devices
implicates that there are differences in the chromophore concentrations derived, making
comparisons between oximeters produced by different manufacturers problematic.
NIRS uses reflected light rather than transmitted light to study the absorption of light in tissue
samples. Reflectance probes locate the light emitter and detector adjacent to one another. The
light takes a ‘banana-shaped’ pathway through the tissues, with the depth of photon
penetration proportional to the source-detector separation (principle of spatial resolution). In
order to compensate for superficial tissue, which is not the tissue of interest, differentially
spaced light detectors are used. Owing to the principle of spatial resolution, the closer receiver
will measure more superficial tissue while the distal optode measures both superficial and
deeper tissue. After subtraction of the interference from superficial tissues- the mathematical
details of which are not provided by the manufacturers- oxygenation in the deeper tissues is
derived.
1.3 Assumptions and limitations
First of all, NIRS does not quantify oxygen molecules but calculates the ratio of light
absorbencies at predefined wavelengths. External light sources may cause significant
artefacts. Secondly, the algorithms used to calculate oxygen saturation assume a fixed
distance for light to travel through the sampled area (the optical path length). However,
![Page 11: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/11.jpg)
5
different tissue components produce very different amounts of photon scattering and
absorption. As a result, variations in probe positioning as well as inter-individual variations in
the composition of tissue may result in 10 to 15% variability of the true optical path length
measurement. The significant inter-individual biological variability in tissue composition
causes a wide variation in ‘normal’ baseline values of volunteers. Therefore, NIRS devices
are best used as trend monitors. Rather than to base therapeutic decisions on absolute
numbers, it is safer to rely on proportional changes of an individual’s baseline value as a basis
for clinical decision making. Thirdly, the physiological correlate to which tissue saturation
measurements obtained with NIRS relate, remains a matter of debate because the interrogated
tissue sample contains all the different vascular components and represents a mixture of
arterial, capillary and venous oxygen saturations.
1.4 Tissue oxygenation changes during vascular occlusion test
Fig 1. StO2 derived parameters during a vascular occlusion test. 1 ischemic downslope, 2 reperfusion
upslope, 3 hyperemic response.7
At resting state, a baseline value is given. After cuff-inflation, a rapid desaturation (=ischemic
downslope) is seen and after 3 minutes, when the cuff is released, the surge of blood flow
causes a rapid resaturation (=reperfusion upslope) and an endothelium-dependent
vasodilatation that creates a transient increase in blood flow to a level higher than at resting
state, this is called the hyperemic response or post-occlusive reactive hyperemia (PORH).
2. AIM
The aim of the present investigation is to determine the interrelationship among different
measurements of microvascular reactivity, obtained with 3 different NIRS devices. The
hypothesis is that measures of microvascular reactivity obtained with the different devices
would be significantly correlated to each other.
![Page 12: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/12.jpg)
6
3. MATERIALS AND METHODS
This is a prospective, observational and descriptive study. The data were acquired as part of a
research project investigating the effects of the priming solution of the cardiopulmonary
bypass (CPB) on microvascular reactivity. After approval by the local research ethics
committee and after obtaining their written informed consent, 40 adult patients (33 males/7
females, mean age 66 +/- 9) scheduled for elective coronary artery bypass grafting surgery
were recruited. Exclusion criteria were an ejection fraction < 25%, diabetes, renal
insufficiency (creatinine > 2.0 mg/dl), significant hepatic disease (liver function tests > 3x
upper limit of normal), history of cerebrovascular disease, significant carotid artery stenosis
(> 60%), perioperative use of corticosteroids, and need for vasopressor or inotropic therapy
before surgery.
3.1 Subject preparation
All subjects needed to fasten at least 6 hours prior to anesthesia and were asked to refrain
from nicotine. On the morning of surgery, patients were allowed to take their routine
medication, except for angiotensin-converting enzyme inhibitors and angiotensin II
antagonists. Patients were premedicated with oral diazepam (5-10 mg). Standard monitoring
was used throughout the procedure, including elektrocardiogram, pulse oximetry and
bispectral index (BIS). Arterial blood pressure was recorded continuously via the right radial
artery catheter. Three disposable NIRS sensors (INVOS 5100C, Covidien, Mansfield, MA;
NIRO-200NX, Hamamatsu Photonics, Tokyo, Japan and Foresight Elite, CAS Medical
Systems, Branford, CT, USA) were applied to the left forearm in a circumferential orientation
(over the brachioradialis muscle, ~5 to 10 cm distal from the proximal head of the radius) for
measurement of microvascular reactivity.
![Page 13: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/13.jpg)
7
3.2 Different NIRS devices
INVOS 5100 Foresight Elite
Equanox 7600 NIRO 200-NX
Several NIRS devices for measuring tissue oxygen saturation are commercially available,
three of which are FDA-approved: INVOS 5100, Foresight and Equanox 7600 (Nonin
Medical Inc., Minneapolis, MN, USA). NIRO 200-NX is not FDA-approved. Despite the
identical basic technology using near-infrared wavelengths to detect changes in the
concentration of O2Hb and HHb, there are several technical differences. NIRO employs the
technique of Spatially Resolved Spectroscopy (SRS, multiple closely spaced detectors to
measure light attenuation as a function of source-detector separation) to measure the tissue
oxygenation index (TOI) and change in hemoglobin. Independently of the SRS method,
NIRO measures changes in concentration of O2Hb, HHb and THb using the modified Beer-
Lambert method. INVOS, Foresight and Equanox use the modified Beer-Lambert law to
measure tissue oxygen saturation, and eliminate the contribution of superficial tissue by using
the principle of Spatial Resolution (depth of photon penetration proportional to the source-
![Page 14: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/14.jpg)
8
detector separation). INVOS 5100 features 2, NIRO 200-NX features 3 and Foresight Elite
features 5 wavelengths of near-infrared light. Theoretically speaking, more wavelengths
should lead to greater accuracy and enhanced tissue recognition.
3.3 Interventions
Sensitivity to changes in oxygenation was evaluated with a vascular occlusion test.
Measurements were performed on the awake patient before induction of anesthesia. A
sphygmomanometer cuff was wrapped around the arm over the left brachial artery. Arterial
occlusion was achieved by inflating a standard blood pressure cuff (EH50U, Siemens) at the
upper arm to a pressure of 50 mmHg above the individual systolic pressure of each subject.
The cuff was automatically inflated in less than 2 seconds to the pressure needed for the
arterial occlusion. After 3 minutes of ischemia, cuff pressure was rapidly released and StO2
response was recorded until it stabilized at the baseline value. Regarding the different values
given by NIRO 200-NX, only the TOI value was used in this study. This allows a comparison
between the tissue oxygen saturation values obtained by spatial resolution (INVOS and
Foresight) and tissue oxygen saturation values obtained by spatially resolved spectroscopy
(NIRO).
We chose to perform the measurements on the forearm because physiologically the forearm is
a predominant place for vasoconstriction in case of circulatory distress. So the vascular
response will be altered sooner and more intensely.8
In our study we used a VOT with a fixed time occlusion of 3 minutes. Several studies have
claimed it is better to occlude until a certain threshold is reached (mostly 40%), while others
claim a 3 min fixed time is better because then the induced stimulus is the same for
everybody. Still no consensus has been reached regarding this issue.
3.4 Missing values
Data of 7 people was missing for Foresight, of 2 people for INVOS and of 1 person for NIRO.
![Page 15: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/15.jpg)
9
3.5 Selected parameters of microvascular reactivity (see Fig 2)
Several parameters of microvascular reactivity can be determined. Following parameters
where selected for this study:
(1) Baseline StO2 (%)
(2) Downslope (M1), desaturation rate (%/min) during first 60 seconds
(3) Downslope, desaturation rate (%/min) from baseline until nadir
(4) Minimum StO2 (%)
(5) Rise time (sec) = time from cuff release to maximum value
(6) Maximum StO2 (%)
(7) Upslope, resaturation rate (%/min) from minimum until maximum value
(8) Settling time (sec) = time from cuff release to second time baseline
Fig 2. Graphic presentation of parameters of microvascular reactivity
These 8 parameters give an insight in the microvascular reactivity of the peripheral tissues at
the time of the vascular occlusion test. The baseline, minimum and maximum values can be
considered as static parameters, the five other parameters as dynamic parameters because they
are time-related. The downslope in the ischemic phase has a biphasic pattern. The initial
desaturation rate is lineair, but after the tissue oxygen level decreases to about 50 %, the slope
often changes and becomes non-lineair. Measuring the first fraction of change of desaturation
![Page 16: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/16.jpg)
10
rate reduces bias by the non-lineair part of the slope.9 For this reason the downslope during
the first minute was included in the analysis. The desaturation rate is a measure for the
oxygen consumption rate and local metabolic demand, while the upslope and rise time are
measures of the endothelial function (arteriolar and capillary recruitment) and the hyperemic
response is a measure for the vascular reserve.4 The hyperemic response is dependent of the
capillary integrity, the local blood volume, the local vasomotor tone, the perfusion pressure,
StO2 and total Hb. A comparison of the PORH parameters revealed that the time parameters
of reactive hyperemia most clearly distinguish between the group of patients with peripheral
vascular disease and the group of healthy volunteers and correlate best with the values of the
ankle-brachial index and transcutaneous tissue oxygenation.6,10 For this reason, and because
of the high intra-individual variability as shown in a study by Gomez et al11, the area under
the curve (AUC) was not included in the analysis.
3.7 Statistics
Statistical analysis was performed using the statistical software SPSS Statistics 22 (SPSS Inc.,
Chicago, IL). The raw data were tested for normality using the Shapiro-Wilk test and were
considered normally distributed if p > 0.05. The non-parametric data are presented as median
[IQR]. Comparisons between devices were performed with the Kruskal-Wallis test. Pairwise
differences among devices were examined for significance by using the Mann-Whitney U
test. The level of statistical significance was set at corrected 2-sided p-value <0.05.
4. RESULTS
4.1 Comparison of the static parameters between the 3 devices
Table 1. Results of comparison of the static parameters (values presented as median [IQR])
Foresight Elite INVOS 5100C NIRO 200-NX
Baseline StO2 (%) 70 [65-73] 66 [61-73] 69 [65-73]
Minimum StO2 (%) 45 [40-51] 36 [21-48]* 46 [36-51]
Maximum StO2 (%) 81 [78-87] 82 [77-86] 79 [75-82]*
* p<0.05 vs. the two other devices
![Page 17: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/17.jpg)
11
Figure 3. Results of comparison of the static parameters (values presented in %)
Values by Foresight are in blue, by INVOS in green and by NIRO in grey (presented in that order from left to
right). The boxes represent the interquartile ranges, the whiskers the range and the horizontal bar the median.
Outliers are marked by colored dots.
The Kruskal-Wallis test demonstrated no difference in the baseline StO2 values between the 3
devices. During the vascular occlusion test a significant difference was observed between
INVOS and NIRO for both minimum and maximum values (p=0.012 and p=0.034,
respectively), and between INVOS and Foresight for the minimum value (p=0.001). Foresight
and NIRO differed for the maximum value (p=0.022).
4.2 Comparison of dynamic parameters between the 3 devices
Table 2. Results of comparison of dynamic parameters (values presented as median [IQR])
Foresight Elite INVOS 5100C NIRO 200-NX
Downslope(M1) (%/min) 11 [6-15] 17 [13-24]* 12 [7-16]
Downslope (%/min) 11 [8-13] 15 [11-21]* 12 [9-15]
Rise time (sec) 40 [28-50]* 25 [23-35] 27 [21-31]
Upslope (%/min) 114 [65-199]* 311 [92-523]* 202 [88-269]*
Settling time (sec) 226 [181-266]* 181 [146-223] 187 [127-248]
* p<0.05 vs. the two other devices (results obtained with Kruskal-Wallis test)
![Page 18: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/18.jpg)
12
Fig. 4 Results of comparison of the desaturation rates (values given in %/min)
Fig. 5 Results of comparison of the time parameters during reperfusion (values given in sec)
Values by Foresight are in blue, by INVOS in green and by NIRO in grey (presented in that order from left to
right). The boxes represent the interquartile ranges, the whiskers the range and the horizontal bar the median.
Outliers are marked by colored dots and *.
![Page 19: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/19.jpg)
13
The results from pairwise comparisons among devices with the Mann-Whitney-U test are
shown in table 3:
Table 3. Pairwise comparison of the dynamic parameters (values given are p-values)
ISCHEMIA REPERFUSION
Downslope(M1) Downslope Rise time Upslope Settling time
Foresight vs
NIRO 0,361 0,208
0,001* 0,040* 0,020*
Foresight vs
INVOS 0,001* 0,001*
0,003* 0,001* 0,002*
INVOS vs
NIRO 0,002* 0,006*
0,513 0,024* 0,668
* p< 0,05
The Kruskal-Wallis test showed that there are significant differences in all 5 dynamic
parameters obtained by the 3 different NIRS devices. After pairwise comparison, it became
clear that INVOS had significantly higher desaturation rates in general and in the first minute
compared to Foresight and NIRO. Also the resaturation rate differed significantly between all
three devices, the fastest resaturation rate was seen with INVOS, the slowest with Foresight.
Foresight had significantly slower rise and settling times than NIRO and INVOS.
Interestingly, the upslope differed significantly between INVOS and NIRO, but the time-
related (rise time and settling time) parameters were similar.
Comparing Foresight with NIRO showed similar parameters during ischemia but significantly
different parameters during reperfusion. Foresight and INVOS differed significantly for all
five dynamic parameters.
No clear-cut difference between the values given by NIRO, using the SRS technique, and the
two other devices, using the spatial resolution technique, can be made. NIRO values differ
significantly from INVOS during ischemia, but not from Foresight, while vice versa is true
during reperfusion. Since no real reference value exists for StO2, it is not possible to state if
one technique is more valid than the other.
![Page 20: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/20.jpg)
14
5. DISCUSSION
5.1 Important findings
This study is the first study directly comparing these three NIRS devices for the measurement
of microvascular reactivity in peripheral tissues during a vascular occlusion test. When
comparing different NIRS devices, three parameters are of importance: the absolute StO2
values during the resting state, the changes in StO2 when oxygenation is altered and the
repeatability, i.e. the similarity of repeated measurements. In our study no significant
differences were found at baseline, but analysis of different parameters of microvascular
reactivity shows that different information is retrieved depending on the NIRS device used.
Repeatability was not tested in this study. So, measures of microvascular reactivity obtained
with different NIRS devices don’t seem to correlate significantly to each other.
As stated in the introduction, there are multiple possible reasons for this observation. The
NIRS devices differ significantly in applied computational algorithms to derive the oxygen
saturation values. Also, penetration depth differs depending on the wavelength and intensity
of the emitted light, the sensitivity of the light detector and the spacing between the light
emitter and light detectors. For example, the fact that INVOS has significantly lower
minimum values and significantly higher maximum values than NIRO, while both devices
have similar time parameters during reperfusion, is most likely a consequence of a difference
in sensitivity between both devices.12 The actual sources of device differences remain to be
elucidated and would probably require access to the raw optical data and exact algorithms and
calibrations which are now kept secret by the different companies. Indeed, every company
makes assumptions regarding the wavelength dependence of scattering that is necessary to
derive the spectral shape of absorption coefficient and thereby deduce StO2. These
assumptions in combination with assumptions about the water content and the different light
emitter and detector geometry probably account for the differences between devices.
5.2 Comparison with previous studies
Most previous studies have used NIRS devices for measuring cerebral tissue oxygenation,
while only a few have used them for measuring peripheral tissue oxygenation. Numerous
other studies have used NIRS devices for measurement of cerebral and peripheral tissue
oxygenation in neonates during the transition period.
![Page 21: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/21.jpg)
15
It is uncertain how results from cerebral measurements can be applied for interpreting
peripheral measurements, because in cerebral oximetry a large variation in reading errors
between subjects is seen, with the mean bias possibly related to variations in the ratio of
arterial and venous blood in the sampling area of the brain. This ratio is probably not fixed, as
assumed by the manufacturers, but dynamically changes with hypoxia. In peripheral
measurements, the A/V ratio is more or less fixed.13-16
Another issue is the influence of extracranial contamination on the cerebral oxygenation
values measured, which has been the subject of multiple studies .8,13,14,17 This also made
possible interference of skin oxygenation on the measurement of peripheral tissue
oxygenation a paradigm on which a couple of studies have laid their focus.19,20 The results of
these studies remain conflicting. Most studies do seem to show that the SRS technique, used
by NIRO to derive the TOI value, more effectively rejects artifacts from superficial
hemodynamic changes in cutaneous microcirculation than the spatial resolution technique.19,21
The results from studies on neonates also cannot simply be extrapolated to be compared with
the results of peripheral tissue oxygenation measurements in adults because the use of
neonatal sensors and a different calibration algorithm influence the results.
Finally, the lack of standardization of the VOT leads to great difficulties when trying to
compare results from different studies.4 The length of occlusion time (fixed time or until
preset threshold), the place of vascular occlusion, the type of sensor used, the type of study
population (healthy volunteers vs patients with significant co-morbidities) and site of
measurements differ greatly among the different studies. It has been proven that
measurements depend strongly on the site, sensors and probes used. For example, a study by
Bezemer et al shows that the reperfusion upslope is greater when 25 mm probes are used vs
15 mm probes and is also greater when measured at the thenar muscle compared to the
forearm.10 INVOS values measured with pediatric or neonatal sensors are about 10% higher
than when adult sensors are used due to different calibration and different algorithm.22 The
results after a VOT also show a great intra-individual variability of the values in stable
conditions. The intra-subject variability can vary up to 14% points and responses to VOT
differ over time in the same patient.9 Needless to say, using a standardized VOT and using
agreed upon sensors is paramount for better comparability of future studies.
![Page 22: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/22.jpg)
16
As stated above, this study is the first study directly comparing these three devices for the
measurement of microvascular reactivity in peripheral tissues during a vascular occlusion test.
A few other studies have compared two of these three devices or have compared one of these
three devices to other devices.
A study by Hyttel et al compared INVOS, Foresight and Equanox in a similar fashion to our
study. His results regarding the comparison between INVOS and Foresight were dissimilar to
ours showing a significant difference in baseline values between the two devices, no
significant difference in desaturation downslope and a significant difference in post-cuff
release maximum values.23 A possible explanation for this discrepancy is that the
measurements were done on young healthy adults compared to adults scheduled for CABG in
our study. Or maybe, the difference is due to the difference between values given by
Foresight and the values given by Foresight Elite. His conclusion, on the other hand, was the
same as ours, values of peripheral tissue oxygenation on the forearm from the three different
devices cannot be used interchangeably.
Another study by Hyttel et al compared the mean values of regional tissue oxygenation, the
reproducibility and dynamic range of four NIRS-instruments (INVOS 5100, NIRO 200NX,
NIRO 300 and Oxyprem) on the human forearm.24 The baseline values between INVOS and
NIRO 200 NX were similar (like in our study), as were the reproducibility (not tested in our
study) and the dynamic range (mixed results in our study). The VOT though was combined
with exercise, so that differs from our study.
A study by Lee et al compared INVOS and Inspectra for the measurement of tissue
oxygenation during a VOT in healthy volunteers.22 The Inspectra model 325 is a NIRS device
specifically designed for peripheral measurements only. The study showed that both devices
give significantly different baseline values, deoxygenation rates, reoxygenation rates and
hyperemia values. Compared the our study, baseline values by INVOS were higher as were
the maximum values post-cuff release. The desaturation and resaturation rates on the other
hand were similar to our study.
A study by Fellahi et al. compared INVOS and Equanox for peripheral oximetry in healthy
volunteers and came to the conclusion that both devices are not comparable for measuring
both absolute and dynamic changes of peripheral StO2 and NIRS-derived parameters during
![Page 23: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/23.jpg)
17
VOT’s.25 The measurements by INVOS differed from our study showing higher baseline
values. Of course, an obvious explanation could be the study population: healthy volunteers
vs pre-CABG patients.
To date, no studies have compared Foresight and NIRO with regard to peripheral tissue
oxygenation measuring.
The observation that INVOS has significantly higher desaturation rates and significantly
lower values at lower tissue oxygenation levels than Foresight and NIRO has been reported
before.22 These results seem to imply that INVOS has the highest sensitivity for changes in
oxygenation of all 3 devices.
Although we did not check the repeatability of the measurements in this study, several other
studies have. They tend to show better repeatability for Foresight and NIRO compared to
INVOS.12,23 So the high sensitivity of INVOS seems to come at the expense of good
repeatability.
Either INVOS data show a greater variability due to less accurate measurement technology, or
alternatively, Foresight and NIRO data show less variability because of a more pronounced
signal attenuation technology, providing tissue oxygenation values that do not as readily
reflect true physiological changes. Clearly, more studies are needed to clarify this issue.
5.3 Clinical relevance
With NIRS technology, the standardization process still has to be initiated, and we are
currently confronted with a wide variety of devices measuring regional tissue oxygen
saturation by using (two to five) different wavelengths, and they come with multiple shaped
optodes or sensors intended to be used at different anatomic regions (forehead, thenar
eminence, somatic organs, muscle etc.). There are several clinical implications and
consequences when it comes to choosing a particular NIRS device: when NIRS oximetry is
intended to be used for trend monitoring, the repeatability of measurement is less important. If
the sensitivity of a NIRS device to changes in hemoglobin oxygen saturation is low, the risk
of undetected true hypoxia will be high. However, if NIRS is to be used as a spot
measurement (e.g. in the emergency room) or if the monitoring is started when the patient
status is uncertain (e.g. in the ICU in a deteriorating patient without having a baseline value),
![Page 24: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/24.jpg)
18
a good repeatability of measurement or a close proximity to absolute (true) values becomes
more important. Since high sensitivity comes at the expense of good repeatability, a
pragmatic approach would be to care less about the closeness to the ‘true’ values, but to settle
for the device with the best combination of repeatability and sensitivity to changing
oxygenation.26
An important paradigm is that no monitoring device, however insightful its data, can improve
patient-centered outcomes, unless it is coupled to a treatment which itself improves
outcome.27 In this regard, a few studies have examined whether tissue oxygenation values can
be used as a target for goal-directed therapy in high-risk surgery. A pilot study by Van Beest
et al. is an example of one such studies and they hypothesized that intraoperative optimization
of StO2 by a perioperative treatment protocol can improve tissue perfusion and thus reduce
postoperative complications.28 The treatment protocol involved starting a dobutamine infusion
when the tissue saturation dropped below 80 %. The conclusion of the study was that no
statistically significant difference in outcome was realized through intraoperative optimization
of StO2 values. Further research is obligatory to define both the optimal StO2 threshold and
intervention to treat tissue hypoperfusion. And as has become clear from this study, each
NIRS device will need a customized treatment protocol.
6. CONCLUSION
Using NIRS for peripheral tissue oxygenation measurements leaves us with 4 main problems:
1) Since no real reference value exists for StO2, it is not possible to state if one monitor is
more valid than another.
2) The accuracy of quantitative data by NIRS is limited by inaccuracies in the estimation
of optical path length for light transmitted through tissue. Until real time path length
measurements are incorporated, relative changes will form the basis of NIRS.
3) There are significant differences in absolute values and dynamic measurements
between different devices (as shown in this and other studies)
4) There is a large intra- and inter-individual variability and a variable repeatability.
Further development of the technology to improve the precision and reproducibility, while
maintaining good sensitivity seems paramount. In the future, using a standardized VOT and
using agreed upon sensors is indicated. Further research is obligatory to define both the
optimal StO2 threshold and interventions to treat tissue hypoperfusion.
![Page 25: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/25.jpg)
19
7. REFERENCES
1. Vallet B. Endothelial cell dysfunction and abnormal tissue perfusion. Crit Care Med 2002;
30 [Suppl.]: S229 –S234
2. De Backer D, Ortiz JA, Salgado D. Coupling microcirculation to systemic hemodynamics.
Curr Opin Crit Care. 16(3): 250-254
3. Putnam B, Bricker S, Fedorka P, et al. The correlation of near-infrared spectroscopy with
changes in oxygen delivery in a controlled model of altered perfusion. Am Surg
2007;73:1017e22.
4. Gerovasili V, Dimopoulos S, Tzanis G, Anastasiou-Nana M, Nanas S. Utilizing the
vascular occlusion technique with NIRS technology. Int J Ind Ergonom 2010; 40: 218-22
5. Moerman A, Wouters P. Near-infrared spectroscopy (NIRS) monitoring in contemporary
anesthesia and critical care. Acta Anaesth Belg 2010; 61: 185-94
6. Kragelj R, Jarm T, Erjavec T, Presern-Strukelj M, Miklavcic D. Parameters of
postocclusive reactive hyperemia measured by near infrared spectroscopy in patients with
peripheral vascular disease and in healthy volunteers. Ann Biomed Eng. 2001 Apr;29(4):311-
20.
7. Bernet C, Desebbe O, Bordon S, Lacroix C, Rosamel P, Farhat F, Lehot JJ, Cannesson M.
The impact of induction of general anesthesia and a vascular occlusion test on tissue oxygen
saturation derived parameters in high-risk surgical patients. J Clin Monit Comput. 2011
Aug;25(4):237-44.
8. Scheeren TWL, Schrober P, Schwarte LA. Monitoring tissue oxygenation by near infrared
spectroscopy (NIRS): background and current applications. J Clin Monit Comput. 2012;
26:279-87
9. Gómez H, Torres A, Polanco P, Kim HK, Zenker S, Puyana JC, Pinsky MR. Use of non-
invasive NIRS during a vascular occlusion test to assess dynamic tissue O(2) saturation
response. Intensive Care Med. 2008 Sep;34(9):1600-7.
10. Bezemer R, Lima A, Myers D, Klijn E, Heger M, Goedhart PT, Bakker J, Ince C.
Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared
spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers.
Crit Care. 2009;13 Suppl 5:S4.
11. Gómez H, Mesquida J, Simon P, Kim HK, Puyana JC, Ince C, Pinsky MR.
Characterization of tissue oxygen saturation and the vascular occlusion test: influence of
measurement sites, probe sizes and deflation thresholds. Crit Care. 2009;13 Suppl 5:S3
![Page 26: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/26.jpg)
20
12. Pocivalnik M, Pichler G, Zotter H, Tax N, Müller W, Urlesberger B; Regional tissue
oxygen saturation: comparability and reproducibility of different devices. J. Biomed. Opt.
0001;16(5):057004-057004-5.
13. Bickler PE, Feiner JR, Rollins MD. Factors affecting the performance of 5 cerebral
oximeters during hypoxia in healthy volunteers. Anesth Analg. 2013 Oct;117(4):813-23.
14. Sørensen H, Rasmussen P, Siebenmann C, Zaar M, Hvidtfeldt M, Ogoh S, Sato K, Kohl-
Bareis M, Secher NH, Lundby C. Extra-cerebral oxygenation influence on near-infrared-
spectroscopy-determined frontal lobe oxygenation in healthy volunteers: a comparison
between INVOS-4100 and NIRO-200NX. Clin Physiol Funct Imaging. 2015 May;35(3):177-
84.
15. Henrik Sørensen, Niels H. Secher, and Peter Rasmussen. A note on arterial to venous
oxygen saturation as reference for NIRS-determined frontal lobe oxygen saturation in healthy
humans. Front Physiol. 2013; 4: 403.
16. Watzman HM, Kurth CD, Montenegro LM, Rome J, Steven JM, Nicolson SC.
Arterial and venous contributions to near-infrared cerebral oximetry. Anesthesiology. 2000
Oct;93(4):947-53.
17. Sørensen H, Secher NH, Siebenmann C, Nielsen HB, Kohl-Bareis M, Lundby C,
Rasmussen P. Cutaneous vasoconstriction affects near-infrared spectroscopy determined
cerebral oxygen saturation during administration of norepinephrine. Anesthesiology. 2012
Aug;117(2):263-70.
18. Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen
saturation: a comparison of three cerebral oximetry technologies. Anesthesiology. 2012
Apr;116(4):834-40.
19. Messere A, Roatta S. Influence of cutaneous and muscular circulation on spatially
resolved versus standard Beer-Lambert near-infrared spectroscopy. Physiol Rep. 2013 Dec
5;1(7).
20. Buono M. J., Miller P. W., Hom C., Pozos R. S., Kolkhorst F. W. Skin blood flow affects
in vivo near‐infrared spectroscopy measurements in human skeletal muscle. Jpn. J.
Physiol.2005; 55:241-244
21. Messere A, Roatta S. Local and remote thermoregulatory changes affect NIRS
measurement in forearm muscles. Eur J Appl Physiol. 2015 Nov;115(11):2281-91
22. Lee JH, Park YH, Kim HS, Kim JT. Comparison of two devices using near-infrared
spectroscopy for the measurement of tissue oxygenation during a vascular occlusion test in
healthy volunteers (INVOS® vs. InSpectra™). J Clin Monit Comput. 2015 Apr;29(2):271-8.
![Page 27: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/27.jpg)
21
23. Simon Hyttel-Sorensen, Trine Witzner Hessel, and Gorm Greisen. Peripheral tissue
oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearm. J
Clin Monit Comput. 2014; 28(2): 149–155.
24. Hyttel-Sorensen S, Sorensen LC, Riera J, Greisen G. Tissue oximetry: a comparison of
mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-
instruments on the human forearm. Biomed Opt Express. 2011 Nov 1;2(11):3047-57.
25. Fellahi JL, Butin G, Fischer MO, Zamparini G, Gérard JL, Hanouz JL. Dynamic
evaluation of near-infrared peripheral oximetry in healthy volunteers: a comparison between
INVOS and EQUANOX. J Crit Care. 2013 Oct;28(5):881
26. Scheeren TWL, Bendjelid K. Journal of clinical monitoring and computing 2014 end of
year summary: near infrared spectroscopy (NIRS). Journal of Clinical Monitoring and
Computing. 2015;29(2):217-220.
27. Michael R Pinsky and Didier Payen. Functional hemodynamic monitoring. J Crit Care.
2005 Dec; 9(6): 566-572
28. Van Beest PA, Vos JJ, Poterman M, Kalmar AF, Scheeren TW. Tissue oxygenation as a
target for goal-directed therapy in high-risk surgery: a pilot study. BMC Anesthesiol. 2014
Dec 16;14:122.
![Page 28: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/28.jpg)
22
Vergelijking tussen drie NIRS toestellen voor het meten van microvasculaire reactiviteit
Dr Kevin Steenhaut, Prof dr. Stefan De Hert, Prof. dr. Anneliese Moerman
Universiteit Gent, Afdeling anesthesie en reanimatie, Gent, België
Achtergrond en doelstelling
Nabije-infrarood spectroscopie (NIRS) wordt in toenemende mate erkent als een methode om
niet-invasief de microvasculaire reactiviteit in weefsels te bepalen. NIRS laat toe om
endotheel-gemedieerde veranderingen in vasculaire tonus, die optreden na een periode van
ischemie, te kwantificeren. Veranderingen in microvasculaire reactiviteit zijn geassocieerd
met gestoorde weefseloxygenatie en orgaandysfunctie. Bepaling van de microvasculaire
reactiviteit kan dus belangrijke informatie opleveren over de status van de weefseloxygenatie,
de ernst van de aandoening en de effecten van de toegepaste behandelingen. Het meten van de
weefseloxygenatie in rust laat niet toe hypoperfusie op te sporen, daarvoor is een vasculaire
occlusie test nodig.
Er komen steeds meer en meer NIRS toestellen op de markt en de mate waarin deze
verschillende toestellen gelijkaardige informatie opleveren, blijft onduidelijk. De doelstelling
van deze studie is om metingen van veranderingen in weefseloxygenatie tijdens een
vasculaire occlusie test door drie verschillende NIRS toestellen met elkaar te vergelijken.
Onze hypothese is dat deze metingen gelijkaardig zijn.
Methodiek
Veertig volwassenen die gepland waren om een electieve CABG te ondergaan, werden
geïncludeerd. Alle patiënten gaven hun toestemming tot deelname aan de studie door middel
van informed consent. Een sensor van elk NIRS toestel (INVOS 5100C; Foresight Elite en
NIRO-200NX) werd op de linker voorarm over de musculus brachioradialis aangebracht. Een
standaard bloeddrukmeter werd op de linker bovenarm aangebracht en werd opgeblazen tot
een druk van 50 mmHg boven de systolische bloeddruk van de patiënt. Na een ischemieduur
van 3 minuten werd de druk snel gelost. Volgende parameters van microvasculaire reactiviteit
werden bepaald voor verdere analyse: de waarde in rust, de snelheid van desaturatie in de
eerste minuut en over 3 minuten, de minimumwaarde, de snelheid van resaturatie, de bereikte
maximumwaarde, de tijd tussen lossen van de druk en bereiken van de maximum waarde en
de tijd tussen het lossen van de druk en het opnieuw stabiliseren rond de rustwaarde.
![Page 29: Comparison of three NIRS devices for the measurement of ... · The velocity and degree of flow restoration depend on the capacity of the microvasculature to recruit arterioles and](https://reader030.vdocuments.net/reader030/viewer/2022040400/5e6f0e14fa3f6c0ca817e6a6/html5/thumbnails/29.jpg)
23
De verschillende metingen werden met elkaar vergeleken door middel van de Kruskal-Wallis
test. Het bestaan van significante verschillen tussen twee van de drie toestellen werd
onderzocht met de Mann-Whitney U test.
Resultaten en discussie
Er waren geen significante verschillen in rust. Analyse van de veranderingen in
weefseloxygenatie tijdens de vasculaire occlusie test toonde aan dat er voor alle parameters
significante verschillen zijn tussen ten minste twee van de drie NIRS toestellen. Mogelijke
verklaringen hiervoor zijn het feit dat elk NIRS toestel een ander algoritme gebruikt voor zijn
metingen. Ook de sensoren van de verschillende NIRS toestellen hebben andere kenmerken.
Aangezien er op dit moment geen gouden standaard bestaat voor het bepalen van de
oxygenatie in weefsels, is het onmogelijk om te achterhalen welk NIRS toestel de meest
correcte metingen uitvoert.
Conclusie
Ondanks het feit dat er geen significante verschillen gevonden werden in rusttoestand, toont
de analyse aan dat er wel significante verschillen zijn tussen de drie NIRS toestellen wanneer
de weefseloxygenatie verandert. Bij het gebruik van NIRS toestellen moet hier steeds
rekening mee gehouden worden. Technische progressie in de NIRS technologie is nodig om
meer betrouwbare en meer correcte resultaten te bekomen.