data independent analysis on thermo scientific orbitrap ms systems

30
1 The world leader in serving science Data-Independent Analysis on Thermo Scientific Orbitrap MS Systems

Upload: thermo-fisher-scientific

Post on 27-May-2015

985 views

Category:

Science


7 download

DESCRIPTION

Data-independent acquisition (DIA) is becoming increasingly important for quantitative proteomics, especially for large-scale targeted protein quantification. DIA provides better reproducibility, acquires fragment ions of all precursors, and breaks through the limit of quantification throughput. DIA with the Thermo Scientific™ Orbitrap™ LC-MS enables ultimate flexibility. The Thermo Scientific™ Orbitrap™ LC-MS pushes the boundaries of large scale quantitation, with highest mass accuracy, Orbitrap high resolution, short cycle times, and narrow isolation windows. Learn more at www.thermoscientific.com/DIAwebinar

TRANSCRIPT

Page 1: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

1

The world leader in serving science

Data-Independent Analysis on Thermo Scientific Orbitrap MS Systems

Page 2: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

2

Main Topics

Introduction: Why Do DIA?

• Advantages of data independent acquisition, where, when, and why its useful

Where does DIA fit in?

• Positioning of DIA methods with other workflows like DDA and PRM

Data Independent Analysis

• Prerequisites, hardware and software

Challenges in DIA

• The three characteristics of the ideal DIA experiment and the compromises among them

Flexibility in DIA Methods

• Multiple varieties of MS2 and MS1+MS2 methods to suit experimental needs

Conclusion: If you Are Going to Make an Archive…

• The new Thermo ScientificTM Q ExactiveTM HF MS as a DIA workhorse

Page 3: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

3

Introduction: Traditional Quantitation Tools

Relative Quantification (lower accuracy and precision)

Absolute Quantification (highest accuracy and precision)

Discovery Routine

MS1 level quan., DDA methods, 1000’s analytes DDA or SRM

up to 100 analytes

SRM, MS2 level quan., 1-10 analytes

Page 4: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

4

Introduction: Quantitation Transformed

Relative Quantification (lower accuracy and precision)

Absolute Quantification (highest accuracy and precision)

Discovery Routine

MS1 level quan., DDA methods, 1000’s analytes DIA or PRM

several 100 analytes

SRM, MS2 level quan., 1-10 analytes

Page 5: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

5

Introduction: A Time and Place for DIA

Typical Samples Published On:

Best Suited When

Protein Complexes

Enriched Samples (e.g. phosphopeptides)

Cellular Fractions

IP Pulldowns

GLOBAL

QUANTITA

TION

PHASE Data

Dependent Acquisition

Data Independe

nt Acquisition

wiSIM

Methods for a variety of sample conditions:

pSMART

basicDIA

msxDIA

Page 6: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

6

Introduction: Advantages of DIA

No targeting or timing

Retrospectively find new targets of interest

Run unmodified sample

Why perform Data Independent Analysis?

Page 7: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

7

HIGH QUALITY SPECTRAL LIBRARIES are critical to a successful DIA experiment

Generating Spectral Libraries

Validated High Quality Library

SEQUEST Search

y₁₆⁺

1815.73828

y₁₅⁺

1684.69775

y₂⁺

248.16026

b₁₄⁺

1497.80652

b₁₀²⁺

564.29541

b₁₃²⁺-H₂O

683.85907

y₈⁺

907.41931

b₁₁⁺

1226.65344

y₁₀⁺

1035.47791

b₇⁺

754.38391

y₁₁⁺

1195.50806

500 1000 1500 2000

m/z

0.0

0.5

1.0

1.5

Inte

nsity

[co

un

ts] (1

0^6

)

Extracted from: D:\Yeast_Smart\W303_Yeast_50cm_140min_120k30k_2.raw #67524 RT: 120.83 FTMS, [email protected], z=+3, Mono m/z=1104.85144 Da, MH+=3312.53977 Da, Match Tol.=0.02 Da

Matches

VALIDATED at

1% FDR

High resolution accurate mass archive of all identified peptides

• Peptide Intact Mass

• Fragment Masses

• Fragment Intensities

• Retention Time

• Isotope Ratios

under actual experimental conditions

1 µg

HeLa

60 min

The Q Exactive HF MS generates a more comprehensive spectral library

Page 8: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

8

Using Libraries to Validate Results

DIA Experimental Data Wide Window MS2 Fragment Ion Spectra

Validated Spectral Library

y₁₆⁺

1815.73828

y₁₅⁺

1684.69775

y₂⁺

248.16026

b₁₄⁺

1497.80652

b₁₀²⁺

564.29541

b₁₃²⁺-H₂O

683.85907

y₈⁺

907.41931

b₁₁⁺

1226.65344

y₁₀⁺

1035.47791

b₇⁺

754.38391

y₁₁⁺

1195.50806

500 1000 1500 2000

m/z

0.0

0.5

1.0

1.5

Inte

nsity

[co

un

ts] (1

0^6

)

Extracted from: D:\Yeast_Smart\W303_Yeast_50cm_140min_120k30k_2.raw #67524 RT: 120.83 FTMS, [email protected], z=+3, Mono m/z=1104.85144 Da, MH+=3312.53977 Da, Match Tol.=0.02 Da

y11

y10

y9

Y8

y7

• Number of Coeluting Library Transitions

• Retention Time Correlation

• Dot Product with Library Spectra

• Mass Accuracy

How many library

fragments are in

the integrated peak

How close is the

RT of the integrated

peak to the library

peptide (blue line)

What is the relative

intensity of

integrated peak

relative to the real

fragment masses

What is the mass

error of the

integrated peak

relative to the real

fragment masses

Page 9: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems
Page 10: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

10

basicDIA: 25 Da Segments

Quantify and confirm on the MS2 level using wide isolation windows

400 – 1000 m/z

MS2 No. 1

MS2 No.30

120 Windows of 6 Da Each with 1 Da Overlap

MS1 No.1

m/z 400

400 – 1000 m/z MS2 No.2

m/z 1000

60

k

12

0k

4 s

ec

12

0k

MS2 No. 1

MS2 No.30

60

k

DIA • 26 amu isolation

• Variable first mass

• 60 000 resolution

• AGC: 2e5

• 105ms Injection

• NCE 28%

• Loop Count 32

Windows 412.5 612.5 812.5 1012.5

437.5 637.5 837.5 1037.5

462.5 662.5 862.5 1062.5

487.5 687.5 887.5 1087.5

512.5 712.5 912.5 1112.5

537.5 737.5 937.5 1137.5

562.5 762.5 962.5 1162.5

587.5 787.5 987.5 1187.5

412.5 612.5 812.5

437.5 637.5 837.5

462.5 662.5 862.5

487.5 687.5 887.5

Page 11: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

11

60k

basicDIA at 60K FWHM

25 Da windows covering the mass range from 400 to 1200 (optional MS1)

MS2 No. 1

MS2 No.32

32 Windows of 26 Da Each with 1 Da Overlap m/z 400 m/z 1200

60k

4 s

ec

MS2 No. 1

MS2 No.32

basicDIA provides good reproducibility due to short cycle times

Page 12: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

12

msxDIA

Speed / Cycle Time of using 20 Da windows with the selectivity of 10 Da windows

400 – 1000 m/z

MS2 No. 1

MS2 No.30

120 Windows of 6 Da Each with 1 Da Overlap

MS1 No.1

m/z 400

400 – 1000 m/z MS2 No.2

m/z 1000

60

k

12

0k

20 s

ec

4 s

ec

12

0k

MS2 No. 1

MS2 No.30

60

k

Full MS • 400-1000 m/z

• Resolution: 120 000

• AGC: 3e6

• 60 ms Injection Profile

DIA • 10 amu isolation

• Variable first mass

• 60 000 resolution

• AGC: 1e5

• 54 ms Injection

• NCE 28%

• MSX Count 2

• Loop Count 16

Windows

835.6297 985.6979 635.5388 615.5297 775.6024

765.5979 475.466 925.6706 715.5751 605.5251

915.6661 745.5888 785.607 445.4524 575.5115

685.5615 655.5479 855.6388 555.5024 845.6343

515.4842 795.6115 995.7025 425.4433 885.6525

465.4615 875.6479 625.5342 905.6615 565.5069

705.5706 665.5524 585.516 595.5206 945.6797

505.4796 495.4751 645.5433 805.6161 675.557

955.6843 455.4569 725.5797 405.4342 695.5661

525.4887 825.6252 935.6752 535.4933 435.4478

815.6206 415.4387 965.6888 485.4706 865.6434

545.4978 975.6934 755.5933 895.657 735.5842

Randomized

List

Generated

by Skyline

………... continues to 5000 targets

Page 13: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

13

msxDIA

Speed / Cycle Time of using 20 Da windows with the selectivity of 10 Da windows

400 – 1000 m/z

MS2 No. 1

MS2 No.30

120 Windows of 6 Da Each with 1 Da Overlap

MS1 No.1

m/z 400

400 – 1000 m/z MS2 No.2

m/z 1000

60

k

12

0k

20 s

ec

4 s

ec

12

0k

MS2 No. 1

MS2 No.30

60

k

Full MS • 400-1000 m/z

• Resolution: 120 000

• AGC: 3e6

• 60 ms Injection Profile

DIA • 10 amu isolation

• Variable first mass

• 60 000 resolution

• AGC: 1e5

• 54 ms Injection

• NCE 28%

• MSX Count 2

• Loop Count 16

………... continues to 5000 targets

Page 14: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

14

DIA on the Q Exactive HF MS: msxDIA

Speed / Cycle Time of using 20 Da windows with the selectivity of 10 Da windows

400 – 1000 m/z

MS2 No. 3

MS2 No.30

60 Windows of 10 Da Each

Full Coverage of

400–1000 m/z

Based upon: Egertson et al. (2013) Nature Methods 10, 744-766

Multiplex 2 x 10 Da Windows Simultaneously

MS1 No.1

m/z 400

400 – 1000 m/z MS1 No.2

MS2 No. 1

MS2 No. 2

m/z 1000

MS2 No. 1

Random

ly D

istr

ibute

d W

indow

s

60

k

12

0k

12

0k

60

k

5 s

ec

2.5

se

c

60k

Page 15: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

15

DIA on the Q Exactive HF MS: msxDIA

Speed / Cycle Time of using 20 Da windows with the selectivity of 10 Da windows

400 – 1000 m/z

MS2 No. 3

MS2 No.30

60 Windows of 10 Da Each

Full Coverage of

400–1000 m/z

Based upon: Egertson et al. (2013) Nature Methods 10, 744-766

Multiplex 2 x 10 Da Windows Simultaneously

MS1 No.1

m/z 400

400 – 1000 m/z MS1 No.2

MS2 No. 1

MS2 No. 2

m/z 1000

MS2 No. 1

Random

ly D

istr

ibute

d W

indow

s

60

k

12

0k

12

0k

60

k

5 s

ec

2.5

se

c

An approach providing a good balance of selectivity, speed, and sensitivity suited for complex mixtures

60k

Page 16: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

16

DIA on the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ MS: WiSIM

The power of two detectors working in parallel: unsurpassed speed and sensitivity

400 – 1000 m/z

MS2 No. 1

MS2 No.30

120 Windows of 6 Da Each with 1 Da Overlap

MS1 No.1

m/z 400

400 – 1000 m/z MS2 No.2

m/z 1000

60

k

12

0k

20 s

ec

4 s

ec

12

0k

MS2 No. 1

MS2 No.30

60

k

SIM • 400-600, 600-800, 800-

1000 m/z

• 200 amu isolation

• Resolution: 240 000

• AGC: 3e4

• 50 ms injection

• Profile

tMS2 • 12 amu isolation

• 150-1850 m/z

• Ion Trap

• AGC: 5e4

• 47 ms Injection

• CID 30

• Rapid Scan

Windows

406 514 606 714 806 914

418 526 618 726 818 926

430 538 630 738 830 938

442 550 642 750 842 950

454 562 654 762 854 962

466 574 666 774 866 974

478 586 678 786 878 986

490 598 690 798 890 998

502 702 902

Large-Scale Targeted Protein Quantification Using WiSIM-DIA on

an Orbitrap Fusion Tribrid Mass Spectrometer

Kiyonami R, Senko M, Zabrouskov V, Huhmer A, Egertson J, Ting S, and MacCoss M.

Page 17: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

17

DIA on the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ MS: WiSIM

The power of two detectors working in parallel: unsurpassed speed and sensitivity

400 – 1000 m/z

MS2 No. 1

MS2 No.30

120 Windows of 6 Da Each with 1 Da Overlap

MS1 No.1

m/z 400

400 – 1000 m/z MS2 No.2

m/z 1000

60

k

12

0k

20 s

ec

4 s

ec

12

0k

MS2 No. 1

MS2 No.30

60

k

SIM • 400-600, 600-800, 800-

1000 m/z

• 200 amu isolation

• Resolution: 240 000

• AGC: 3e4

• 50 ms injection

• Profile

tMS2 • 12 amu isolation

• 150-1850 m/z

• Ion Trap

• AGC: 5e4

• 47 ms Injection

• CID 30

• Rapid Scan

Large-Scale Targeted Protein Quantification Using WiSIM-DIA on

an Orbitrap Fusion Tribrid Mass Spectrometer

Kiyonami R, Senko M, Zabrouskov V, Huhmer A, Egertson J, Ting S, and MacCoss M.

Page 18: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

18

DIA on the Orbitrap Fusion MS: WiSIM

Quantify using ultra-high resolution MS1 with MS2 confirmation and IT sensitivity

SIM 400 – 600 m/z

MS2 No. 1

MS2 No.17

17 Windows of 12 Da Each MS1 No.1

SIM 600 – 800 m/z

Ion

Tra

p

24

0k

3.6

se

c

SIM 800 – 1000 m/z

MS1 No.2

24

0k

MS2 No. 1

MS2 No.17

Ion

Tra

p

MS1 No.3

24

0k

MS2 No. 1

MS2 No.17

Ion

Tra

p

Parallelized

Parallelized

Parallelized

Page 19: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

19

DIA on the Orbitrap Fusion MS: WiSIM

Quantify using ultra-high resolution MS1 with MS2 confirmation and IT sensitivity

SIM 400 – 600 m/z

MS2 No. 1

MS2 No.17

17 Windows of 12 Da Each MS1 No.1

SIM 600 – 800 m/z

Ion

Tra

p

24

0k

3.6

se

c

SIM 800 – 1000 m/z

MS1 No.2

24

0k

MS2 No. 1

MS2 No.17

Ion

Tra

p

MS1 No.3

24

0k

MS2 No. 1

MS2 No.17

Ion

Tra

p

Parallelized

Parallelized

Parallelized

wiSIM provides fully parallelized speed, IT sensitivity, and ultimate resolution

Page 20: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

20

Achieving the Utmost DIA Performance

Short cycle times, good reproducibility

Excellent selectivity and sensitivity

Balance of speed, sensitivity, selectivity

Fully parallelized speed, IT sensitivity, ultimate resolution

Page 21: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

21

DIA: In Summary

y₁₆⁺

1815.73828

y₁₅⁺

1684.69775

y₂⁺

248.16026

b₁₄⁺

1497.80652

b₁₀²⁺

564.29541

b₁₃²⁺-H₂O

683.85907

y₈⁺

907.41931

b₁₁⁺

1226.65344

y₁₀⁺

1035.47791

b₇⁺

754.38391

y₁₁⁺

1195.50806

500 1000 1500 2000

m/z

0.0

0.5

1.0

1.5

Inte

nsity

[co

un

ts] (1

0^6

)

Extracted from: D:\Yeast_Smart\W303_Yeast_50cm_140min_120k30k_2.raw #67524 RT: 120.83 FTMS, [email protected], z=+3, Mono m/z=1104.85144 Da, MH+=3312.53977 Da, Match Tol.=0.02 Da

Obtain super deep proteomic profiling using high resolution / accurate mass in under an hour.

Page 22: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

22

DIA: In Summary

Access the highest resolutions and mass accuracy.

Page 23: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

23

Access the highest resolutions and mass accuracy.

DIA: In Summary

Page 24: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

24

DIA: In Summary

Have the flexibility to quantify on either the MS1 or MS2 level at high resolution

Page 25: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

25

DIA: In Summary

Have the selectivity of narrow windows AND the speed of wide windows thanks to precursor multiplexing

Page 26: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

26

DIA: In Summary

Ensure reproducible quantitation with scan speeds up to 18 Hz !

10 points 3 points

RSD 8-15% RSD >25%

Page 27: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

27

DIA: In Summary

Next generation analyzers provide twice the resolution at the same speed

0

20000

40000

60000

80000

100000

120000

400 600 800 1000 1200 1400 1600 1800 2000

Re

so

luti

on

m/z

Resolution at 7 Hz Analyzer Setting Over Typical Fragment Range

35k 27k

19k

Page 28: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

28

DIA: In Summary

dotp

0.99

The capabilities of the Orbitrap platform make it the ideal DIA instrument

Next generation analyzers provide twice the resolution at the same speed

0

20000

40000

60000

80000

100000

120000

400 600 800 1000 1200 1400 1600 1800 2000

Re

so

luti

on

m/z

Resolution at 7 Hz Analyzer Setting Over Typical Fragment Range

35k 27k

19k

Page 29: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems

29

Concluding Remarks

Transform your DIA experiment with the latest Orbitrap instruments

Page 30: Data Independent Analysis on Thermo Scientific Orbitrap MS Systems