Page 1: Development of neutron  dosimetry  & spectrometry using track detectors

Alvin Sashala Naik

Under the coordination of:Dr. A. Parravicini & Prof. M. Caresana

Page 2: Development of neutron  dosimetry  & spectrometry using track detectors

Annual ARDENT meeting Vienna 2012 2

Mauritius island!!

21/04/23

Page 3: Development of neutron  dosimetry  & spectrometry using track detectors

21/04/23 Annual ARDENT meeting Vienna 2012 3

Page 4: Development of neutron  dosimetry  & spectrometry using track detectors

Mauritius

• Developing country with a

diversified economy

• Independent since 1968

• Multicultural population of 1.2 M

inhabitants

• Volcanic island of lagoons and

white sand beaches

• Unfortunately, lack of

infrastructures at the UoM to study

advanced experimental physics

Annual ARDENT meeting Vienna 2012 421/04/23

Page 5: Development of neutron  dosimetry  & spectrometry using track detectors

Education BSc degree in Applied Physics

University of Paris Sud 11

MSc degree in Fusion Science and Technologies

• Commissariat à l’Energie Atomique (CEA), Ecole Polytechnique de

Palaiseau & Université Paris Sud 11

• Diagnostics for Inertial and Magnetic confinement fusion devices

• Simulation of the wall surface damage from high energy particles

in the International Thermonuclear Experimental Reactor

(ITER Organization)

Annual ARDENT meeting Vienna 2012 521/04/23

Page 6: Development of neutron  dosimetry  & spectrometry using track detectors

ARDENT

• PhD on Neutron dosimetry and spectrometry with track detectors

• Early Stage Researcher (ESR) n° 13• Starting PhD on January 2013 due to delay on visa

and residence permit process...• Further partnerships with AIT, CERN• Possible research collaboration with CNAO

Annual ARDENT meeting Vienna 2012 621/04/23

Page 7: Development of neutron  dosimetry  & spectrometry using track detectors

Host organisations

• MI.AM

– Development of instruments for measuring ambient radiation

– Radon detection– POLITRACK system for

automatic analysis of track detectors

• POLIMI

– Hosting university for the PhD project

– Development of adapted Monte Carlo codes

– Statistical physics– Further investigation

into personal and environmental neutron dosimetry

Annual ARDENT meeting Vienna 2012 721/04/23

Page 8: Development of neutron  dosimetry  & spectrometry using track detectors

Development of neutron dosimetry and spectrometry using track

detectors

1. Use of CR-39 track detector coupled to a boron converter, as thermal neutron detector (exploiting the n,α reaction on 10B), inside Bonner spheres and REM counters.– The neutron is detected by the 1.47 MeV α

particle.

8Annual ARDENT meeting Vienna 201221/04/23

Page 9: Development of neutron  dosimetry  & spectrometry using track detectors

Use of CR39 as thermal neutron detector

9Annual ARDENT meeting Vienna 201221/04/23

Page 10: Development of neutron  dosimetry  & spectrometry using track detectors

Development of neutron dosimetry and spectrometry using track

detectors• Use the recoil protons (radiator-degrader

technique). – The number of tracks is proportional to the fast neutron

fluence. – The detection efficiency depends on the neutron energy.

• Calculation of particle LET and impinging angle. – Direct estimate of the equivalent dose by calculating, for

any particle the dose and the quality factor Q(LET) .

21/04/23 Annual ARDENT meeting Vienna 2012 10

Page 11: Development of neutron  dosimetry  & spectrometry using track detectors

11

Cone shape track left after chemical treatment (Etching)

Page 12: Development of neutron  dosimetry  & spectrometry using track detectors

21/04/23 Annual ARDENT meeting Vienna 2012 12

Page 13: Development of neutron  dosimetry  & spectrometry using track detectors

Image processing with POLITRACK

21/04/23 Annual ARDENT meeting Vienna 2012 13

Page 14: Development of neutron  dosimetry  & spectrometry using track detectors

• Vt measured from geometrical parameters of tracks formed

• LET deduced from V

21/04/23 Annual ARDENT meeting Vienna 2012 14

Page 15: Development of neutron  dosimetry  & spectrometry using track detectors

Work in perspective:

- A full Monte Carlo characterization of the dosimeters- Angular response still to be understood

- Metrological characterization for personal and environmental dosimetry

- Medical physics application (?) : secondary dose evaluation due to fragments, with LET spectrometry

15Annual ARDENT meeting Vienna 201221/04/23

Page 16: Development of neutron  dosimetry  & spectrometry using track detectors

Thank you for your attention!