Dissolved Dissolved Oxygen and Oxygen and Biochemical Biochemical Oxygen Demand Oxygen Demand Analyses Analyses Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit Note: These Procedures are available in the OTCU Laboratory manual which is available on the OTCU website.

Upload: vuongkhue

Post on 20-Jun-2018

226 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Dissolved Oxygen Dissolved Oxygen and Biochemical and Biochemical Oxygen Demand Oxygen Demand

AnalysesAnalyses

Prepared ByMichigan Department of Environmental Quality

Operator Training and Certification UnitNote: These Procedures are available in the OTCU Laboratory manual which is

available on the OTCU website.

Page 2: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Dissolved OxygenDissolved Oxygen

D.O.Amount of “FREE” Oxygen

O2In the Water

Required By :Required By :FISH

MICROORGANISMS( Bacteria)

Page 3: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

RESPIRATIONRESPIRATIONOrganic Matter Converted to:

Carbon Dioxide and

Water

CH2O + O2 CO2 + H2O

Page 4: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

RESPIRATIONRESPIRATIONCH2O + O2 CO2 + H2O

Reason for TreatmentBasis of

Secondary Treatment

Basis of BOD Analysis

Page 5: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

NitrificationNitrificationBIOLOGICAL Oxidation of Nitrogen

FromAMMONIA

(NH3)

toNITRITE

( NO2)

to NITRATE

(NO3)

NH3 + O2 NO2 + O2 NO3

Page 6: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

NITRIFICATIONNITRIFICATIONNH3 + O2 NO2 + O2 NO3

May be Required:

Oxygen Demand Reduction

Ammonia Removal (toxic)

Page 7: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Reason for D.O.Reason for D.O.AnalysisAnalysis

Used in BOD AnalysisTreatment Processes

Treatment Plant EffluentsReceiving Stream

Page 8: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Maximum D.O. Concentrations in Water (Saturation)

Temperature0C

Max. Concentrationmg/L

4 13.18 11.912 10.816 10.020 9.224 8.528 7.930 7.6

0 14.6

Page 9: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

SAMPLE COLLECTIONSAMPLE COLLECTION

21 %Atmosphere

210,000 mg/L

Sample

0 to 10 mg/L

Page 10: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

SAMPLE COLLECTIONSAMPLE COLLECTION

Page 11: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

SAMPLE COLLECTIONSAMPLE COLLECTION

Page 12: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Do Not Do Not AERATE AERATE SampleSample

Do Not Splash Sample in AirDo Not Splash Sample in Air or or

Let Air Circulate in SampleLet Air Circulate in Sample

NONO YESYES

Page 13: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D.O. D.O. Procedure

Page 14: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

WINKLER

DISSOLVED OXYGEN

DETERMINATION

Page 15: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

WINKLER Dissolved Oxygen PROCEDUREwith the

ALSTERBERGAZIDE

MODIFICATIONAZIDE Destroys NITRITE

Page 16: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent
Page 17: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Winkler MethodIodometric Method

Mn SO4 + 2 KOH Mn(OH)2 + K2SO4

MnO(OH)2 + 2 H2SO4 Mn(SO4)2 + 3 H2O

2 Mn(OH)2 O22 MnO(OH)2+

+Mn(SO4)2 2 KI Mn SO4 + K2SO4 + I2

2. Takes Time

1. Takes Mixing

3. Free IODINE released in relation to D.O. in Sample

Page 18: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Winkler Procedure

1 mL 1 mL ManganousManganous

SulfateSulfate

1 mL1 mLAlkalineAlkaline

Azide SolutionAzide Solution

Page 19: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

StopperStopper Mix WellMix Well

AllowAllowFlocFloc

to Settleto Settle

Page 20: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Repeat MixingRepeat Mixing SettleSettleAgainAgain

Contact and TimeContact and Time

Page 21: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

WhiteWhite - No Dissolved Oxygen- No Dissolved Oxygen

Page 22: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

SulfamicSulfamicAcidAcid

1 mL 1 mL Sulfuric Sulfuric

AcidAcid

MixMix

Page 23: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

IodineIodineSolutionSolution

MeasureMeasure100 mL100 mL

Page 24: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

TitrationSodium ThiosulfateSodium Thiosulfate

(Thio)(Thio)Phenylarsene OxidePhenylarsene Oxide

(PAO)(PAO)OR

Page 25: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

End Point Indicator

StarchStarch

Page 26: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BlueBlue

ClearClear

Page 27: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent
Page 28: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Mg/L D.O. =

mL of Titrant X Normality of Titrant X 8 X 1000

mL of Sample

0.0125 N X 8 X 1000 = 1100

Page 29: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Mg/L D.O. =

mL of Titrant X 1When 100 ml of Sample is Titrated with 0.0125 Normal Titrant

Each mL used Equals 1 mg/L of D.O.

If4.6 mLis Used

Then the D.O. in the Sample is4.6 mg/L

Page 30: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

1 mL of Titrant = 1 mg/L D.O.

With Any of the Following Combinations ofSample Volumes and Normalities

Sample Volume Normality

100 mL 0.0125 N

200 mL 0.0250 N

300 mL 0.0375 N

Page 31: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Outline Of Winkler Dissolved Oxygen Procedure

CarefullyCollectSample

In 300 mLBOD Bottle

3.1Add1 mL

MnSO4 Soln.and

1 mLAlkali-iodide-azide

Reagent

3.11Yellow

ToBrownFloc,

D.O. Present

WhiteFloc,

No D.O.

3.12RepeatMixing

andSettling

3.13Add1 mLH2SO4

andMix

Mix ByInverting

andAllow To

Settle

Page 32: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Titration of Iodine Solution

3.21

Pour100 mL

Into Flask

3.22

TitrateWithTHIO

Reddish-Brown

3.23Add

StarchIndicator

PaleYellow

Blue

3.24

Titrateto

Clear

Clear

Page 33: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Winkler MethodWinkler MethodIodometric MethodIodometric Method

Mn SOMn SO44 ++ 2 KOH2 KOH Mn(OH)Mn(OH)22 ++ KK22SOSO44

MnO(OH)MnO(OH)22 ++ 2 H2 H22SOSO44 Mn(SOMn(SO44))22 ++ 3 H3 H22OO

2 Mn(OH)2 Mn(OH)22 OO222 MnO(OH)2 MnO(OH)22++

++Mn(SOMn(SO44))22 2 KI2 KI Mn SOMn SO44 ++ KK22SOSO44 ++ II22

2. Takes Time2. Takes Time

1. Takes Mixing1. Takes Mixing

3. Free 3. Free IODINEIODINE released in relation to D.O. in Sample released in relation to D.O. in Sample

Page 34: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D.O. D.O. Procedure

Electrode Methods

Page 35: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

The membrane electrode is composed of two solid metal electrodes in contact with

supporting electrolyte separated from the test solution by a gas permeable membrane.

Oxygen dissolved in the sample diffuses through the membrane on the DO probe and is chemically reduced (accepts electrons),

producing an electrical current between the anode and cathode in the probe. The amount

of current is proportional to the concentration of DO. Following proper

calibration, the meter relates this current to the concentration of DO.

MEMBRANE ELECTRODE METHODMEMBRANE ELECTRODE METHOD

Page 36: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

CLARK ELECTRODECLARK ELECTRODE

Sample

GasPermeableMembrane

InternalSolution

CathodeAnode

O2

O2

O2O2

O2

O2 O2

O2O2

O2

O2

O2

O2

O2 + 4H+ + 4e- 2H2O

e-

Meter

Page 37: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

1. Fill two BOD bottles completely full of BOD dilution water, being very careful not to introduce air into either bottle.

2. Analyze one bottle for D.O. using the Winkler titration.

3. Insert the electrode into the second bottle, turn on the stirring mechanism, and wait for the reading to stabilize.

4. Calibrate the meter to the D.O. value obtained in the titration.

5. The meter is now ready for sample analysis

MEMBRANE ELECTRODE METHODMEMBRANE ELECTRODE METHOD

Comparison with Winkler Titration

Calibration

Page 38: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

This procedure varies considerably among the various instrument models available. Therefore, the procedure must be obtained from the instrument manual, but the following points should be noted.

1. Where possible with the specific equipment being used, compensation should be made during calibration for both ambient temperature and local atmospheric pressure. This pressure should be determined using a reliable onsite barometer. The oxygen solubility table following this procedure may be used.

2. Carefully blot any water droplets from the membrane using a soft tissue.

3. During calibration, be sure the membrane is exposed to fresh air. Laying the electrode on the bench for calibration is usually adequate.

4. Complete the calibration as soon as possible before the electrode membrane begins to dry.

5. The temperature registered on the meter should be checked against a trusted thermometer often.

6. Daily calibration of the D.O. meter is required. Calibration should also be verified after every five or six sample measurements.

7. Assure sufficient sample flow across membrane surface during analysis to overcome erratic response.

MEMBRANE ELECTRODE METHODMEMBRANE ELECTRODE METHOD“AIR” Calibration

Page 39: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent
Page 40: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

CLARK ELECTRODECLARK ELECTRODECalibration

TemperatureTemperatureAtmospheric PressureAtmospheric Pressure

(Use a Reliable Barometer)(Use a Reliable Barometer)

NONO

NOTNOTREQUIREDREQUIRED

Page 41: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Luminescence D.O. Probe

Page 42: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Luminescence D.O. Probe

Page 43: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

How Does LDO Work?

• A sensor is coated with a luminescent material.

• Blue light from an LED strikes the luminescent chemical on the sensor.

• The luminescent chemical instantly becomes excited.

LEDPhoto Diode

Probe

Sensor

LED

Page 44: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• As the excited chemical relaxes, it releases red light.

LEDPhoto Diode

Probe

Sensor

LED• The red light is

detected by a photo diode.

• The time it takes for the chemical to return to a relaxed state is measured

How Does LDO Work?

Page 45: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• When oxygen contacts the luminescent chemical, the intensity of the red light decreases

• The amount of time it takes for the material to relax is reduced

LEDPhoto Diode

Probe

Sensor

LED

Oxygen

LuminescentIndicator

Molecules Gas PermeablePolymer Matrix

Clear, GasImpermeable

Substrate

How Does LDO Work?

Page 46: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• The intensity of the red light is not what’s being measured.

• What’s being measured is the time it takes after excitation for red light to be given off. – Lifetime of

luminescence

LEDPhoto Diode

Probe

Sensor

LED

How Does LDO Work?

Page 47: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• A red LED is also present in the probe.

• Between flashes of the

blue LED, a red LED of known intensity, is flashed on the sensor.

LEDPhoto Diode

Probe

Sensor

LED

• The red LED acts as an internal standard (or reference) for a comparison to the red light given off by the luminescent chemical.

How Does LDO Work?

Page 48: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Why is this a Big Deal?

No membrane to replaceNo more stretching of Teflon and worrying about air bubbles No more punctured membranes

No electrolyte to foul or poisonNo H2S poisoning of the electrolyte

No anode or cathodeNo cleaning of anodesNo more coating of electrodes

Reduced Maintenance

Page 49: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Why is this a Big Deal?

No membrane to replaceNo more stretching of Teflon and worrying about air bubbles No more punctured membranes

No electrolyte to foul or poisonNo H2S poisoning of the electrolyte

No anode or cathodeNo cleaning of anodesNo more coating of electrodes

Reduced Maintenance

Page 50: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Why is this a Big Deal?

No membrane to replaceNo more stretching of Teflon and worrying about air bubbles No more punctured membranes

No electrolyte to foul or poisonNo H2S poisoning of the electrolyte

No anode or cathodeNo cleaning of anodesNo more coating of electrodes

Reduced Maintenance

Page 51: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• Frequent Calibration Not Required– No anode to consume and no

electrolyte to deplete means extremely stable measurements

– Internal standard with Red LED– No interference from pH swings,

wastewater chemicals, H2S, or heavy metals

Why is this a Big Deal?

Page 52: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• Accurate and Stable Readings –With nothing to interfere with the

readings, LDO produces more stable measurements for a longer time

• Speed!– Turn it on and it’s running!– Response time of less than 30 seconds

to 90%!

Why is this a Big Deal?

Page 53: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

• Simple Operation and Maintenance– Only one replacement part– Inexpensive sensor cap is simple to

replace quickly

Why is this a Big Deal?

Page 54: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

NOW EPA APPROVEDNOW EPA APPROVEDListed In

Federal RegisterVol. 77,No. 97

Friday, May 18, 2012

Listed as ASTM Method D888-09 (C)Footnote 63 – Hach Method 10360

(BOD and cBOD)Footnote 64 In-Situ Method 1002-8-2009

(Dissolved Oxygen Measurement by Optical Probe)

URL - http://www.gpo.gov/fdsys/browse/collection.action?collectionCode=FR&browsePath=2012%2F05%2F05-18%5C%2F6%2FEnvironmental+Protection+Agency&isCollapsed=false&leafLevelBrowse=false&isDocumentResults=true&ycord=546

Page 55: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D.0. Meter

D.0. Meter

Page 56: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D.O. METERD.O. METER

AdvantagesAdvantagesSaves TimeSaves Time

Continuous MonitoringContinuous MonitoringLess Chemical InterferenceLess Chemical Interference

PortablePortable

Page 57: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D.O. METERD.O. METERLimitationsLimitations

(Membrane Electrode)(Membrane Electrode)

Daily CalibrationDaily CalibrationFlow Past MembraneFlow Past MembraneMembrane May FoulMembrane May Foul

Requires TrainingRequires Training

Page 58: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD5

Page 59: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

B.O.D.Biochemical Oxygen Demand

The Quantity of Oxygen Used in the Biochemical Oxidation of Organic Material.

Under:Specified TimeSpecified TemperatureSpecified Conditions

5 Days200 C

In the DarkIn the Presence

of Bacteria

Page 60: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Cx H

Y OzC xH

YO z

CxHYOz

C xH YO zCx H

Y Oz

NH3

NH3

Bacteria

Measured Volume of Wastewater is Measured Volume of Wastewater is Added to BOD Bottle.Added to BOD Bottle.

Contains:Contains:OrganicsOrganics

AmmoniaAmmonia

BacteriaBacteria

Page 61: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Cx H

Y Oz

C xH YO z

CxHYOz

C xH YO z

Cx H

Y Oz

NH3

NH3

Bacteria

O2

O2

O2

O2

O2

O2

Dilution Water AddedDilution Water Added

Dilution Water Contains:Dilution Water Contains:

NutrientsNutrients

OxygenOxygen

MeasureMeasureD.O. ConcentrationD.O. Concentration

Page 62: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

O2

O2

O2

Incubate 5 DaysIncubate 5 Days

Some Oxygen Used:Some Oxygen Used:

RespirationRespiration

NitrificationNitrificationCx H

Y Oz + O

2 H2 O + CO

2

NH3 + O

2 NO3

MeasureMeasureD.O. ConcentrationD.O. Concentration

Page 63: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

O2

O2

O2

Cx H

Y Oz + O

2 H2 O + CO

2

NH3 + O

2 NO3

Measure Oxygen LossMeasure Oxygen Loss(Demand)(Demand)

D.O. Day 1 D.O. Day 1 -- D.O. Day 5 D.O. Day 5

= Oxygen Demand= Oxygen Demand

(Of What Is In The Bottle)

Page 64: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD

REAGENTS Distilled Water

High Quality

Free of Toxic Material

Free of Oxygen Demanding Substances

Page 65: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD

REAGENTS

Phosphate Buffer Magnesium Sulfate Calcium Chloride Ferric Chloride

Distilled Water

Provide Essential NutrientsBuffer pH

Page 66: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

OTHER REAGENTSOTHER REAGENTS

DECHLORINATING AGENTSodium Sulfite - Na2SO3

NITRIFICATION INHIBITORCBOD

QUALITY CONTROL CHECKAccuracy

Glucose - Glutamic Acid Solution

Page 67: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

SAMPLE PRETREATMENTTemperature

Near 200C

pHBetween 6.5 and 7.5

(Adjust if > 8.5 or < 6.0 and seed)

Supersaturated D.O.Agitate

DechlorinateProper Amount of Sodium Sulfite

Page 68: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

DECHLORINATION100 mL of Sample

+ Potassium Iodide + Sulfuric Acid + Starch

Titrate with Sodium Sulfite

to Starch Iodide

Endpoint

Page 69: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

DILUTION WATERDILUTION WATERDistilled Water

High QualityNo Toxics

No Organicsplus

BUFFER pH 7.2

plus

NUTRIENTS(Phosphorus and Ammonia)MagnesiumCalciumIron

Page 70: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

DILUTION WATER PREPARATION

NEEDED VOLUME for Each Day’s UseADD NUTRIENTS (1 mL per Liter)

SATURATE WITH OXYGEN

VacuumShake(small volume)

orDraw Vacuum

Page 71: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

DILUTION WATER PREPARATIONNEEDED VOLUME

ADD NUTRIENTSSATURATE WITH OXYGEN

STOREIn Incubator

ADD BUFFERDay of Use

Page 72: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDUREBOD PROCEDUREDILUTE SAMPLE

Minimum Residual, 1.0 mg/LMinimum Depletion, 2.0 mg/LAt Least Two DilutionsThoroughly Mix Sample

ADD NITRIFICATION INHIBITORIf Required for CBOD

TCMP0.10 gram/bottle

Two “shots”

Page 73: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDUREBOD PROCEDUREDILUTE SAMPLE

Minimum Residual, 1.0 mg/LMinimum Depletion, 2.0 mg/LAt Least Two DilutionsThoroughly Mix Sample

ADD NITRIFICATION INHIBITORIf Required for CBOD

Add SEED (Bacteria)If Required

Disinfected SamplesIndustrial SamplesReference Samples

Page 74: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDUREBOD PROCEDURESource of Seed (bacteria)

Settled SewagePrimary Effluent

CommerciallyAvailable

Adapted Seed

Page 75: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDURE, (cont.)BOD PROCEDURE, (cont.)

FILL BOTTLE (with dilution water)

MEASURE INITIAL D.O.

STOPPER(No Air Bubbles)

SEAL(Water and Cover)

Page 76: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDURE, (cont.)BOD PROCEDURE, (cont.)

INCUBATE

MEASURE FINAL D.O.

(wash bottles)

20 ± 1°C5 Days ± 6 hour

Page 77: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

DILUTION WATER DILUTION WATER BLANKBLANK

CHECK ON QUALITY

MAXIMUM DEPLETIONMAXIMUM DEPLETION0.2 mg/L0.2 mg/L

NOT USED IN CALCULATIONS

Page 78: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D.O. In - D.O. Out

= Oxygen Demand of Diluted Sample

C1 X V1 = C2 X V2

SampleBOD X Sample

Volume = BODDiluted X Diluted

Volume

BOD Sample = BOD DilutedSample Volume X Volume

Diluted

BOD Sample =Depletion X 300 mL

= DEPLETION

B.O.D. Calculation

Sample Volume

Page 79: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

B.O.D. mg/L =B.O.D. mg/L =D.0. DEPLETION (mg/L)D.0. DEPLETION (mg/L)SAMPLE VOLUME (mL)SAMPLE VOLUME (mL) X 300 mLX 300 mL

D.O. DEPLETION = D.O. Initial - D.O. 5-Day

Minimum Depletion - 2.0 mg/LMinimum Residual - 1.0 mg/L

Page 80: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

B.O.D. Example ProblemB.O.D. Example ProblemCalculate the B.O.D.:

B.O.D., mg/L = D.O. Depletion, mg/LVolume Sample, mL

X 300 mL

Initial D.O. - Residual D.O. Volume Sample, mLs

X 300 mL=

= 8.1 mg/L - 2.1 mg/L 60 mL

X 300 mL

= 6.0 mg/L 60 mL

X 300 mL

= 0.1 mg/L X 300 = 30 mg/L

Initial Sample D.O. = 8.1 mg/L5-Day Sample D.O. = 2.1 mg/LVol. Of Sample in 300 mL Bottle = 60 mL

Page 81: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

B.O.D. PRACTICE PROBLEMCalculate the B.O.D. value to be reported for each of the samples below. Be sure to consider the minimum depletion and residual requirements.

Sample 1 Dilution A Dilution BmL Sample 15 30Initial D.O., mg/L 8.1 8.25-Day D.O., mg/L 6.6 4.2Depletion 1.5 4.0

B.O.D., mg/L = D.O. Depletion, mg/LVolume Sample, mL

X 300 mL

= 4.0 mg/L 30 mL

X 300 mL

= 40 mg/L

Page 82: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

B.O.D. PRACTICE PROBLEMCalculate the B.O.D. value to be reported for each of the samples below. Be sure to consider the minimum depletion and residual requirements.

Sample 2 Dilution A Dilution BmL Sample 15 30Initial D.O., mg/L 8.0 8.25-Day D.O., mg/L 4.3 0.7Depletion 3.7

B.O.D., mg/L = D.O. Depletion, mg/LVolume Sample, mL

X 300 mL

= 3.7 mg/L 15 mL

X 300 mL

= 74 mg/L

Page 83: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

B.O.D. PRACTICE PROBLEMSample 3 Dilution A Dilution B

mL Sample 15 30Initial D.O., mg/L 8.1 8.15-Day D.O., mg/L 5.6 3.3Depletion 2.5 4.8

BODB = 4.8 mg/L 30 mL

X 300 mL = 48 mg/L

= 49 mg/L

BODA = 2.5 mg/L 15 mL

X 300 mL = 50 mg/L

50 + 48 2

Page 84: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDUREBOD PROCEDUREDILUTE SAMPLE

Minimum Residual, 1.0 mg/LMinimum Depletion, 2.0 mg/LAt Least Two DilutionsThoroughly Mix Sample

ADD NITRIFICATION INHIBITORIf Required for CBOD

Add SEED (Bacteria)If Required

De-chlorinated SamplesIndustrial SamplesReference Samples

Page 85: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD PROCEDUREBOD PROCEDURESource of Seed (bacteria)

Settled SewagePrimary Effluent

CommerciallyAvailable

Adapted Seed

Page 86: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

SEEDED BOD SEEDED BOD PROCEDUREPROCEDURE

Page 87: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

1

2

Sample

Organics

Seed

AndOrganics

Bacteria

Depletion3.7 mg/L

Depletion5.0 mg/L

10 mL of seed caused 5.0 mg/L Depletion1 mL would have caused 5.0 mg/L

10 mL = 0.5 mg/L

100 mL

10 mL

1 mL

So: 0.5 mg/L would be used by 1 mL of seedSo: 0.5 mg/L would be used by 1 mL of seedin seeded samplein seeded sample

Page 88: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD(seeded)=D1 - D2

Sample Volume X 300 mL

SEEDED BOD SEEDED BOD CALCULATIONCALCULATION

BOD = Depletion Caused by SampleVolume of Sample Used X 300 mL

Page 89: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD(seeded)=D1 - D2

Sample Volume X 300 mL

D1 = Depletion Due To Sample and Seed

D2 = D.O. Depletion Due to Just Seed

SEEDED BOD CALCULATIONSEEDED BOD CALCULATION

(Total Depletion in Bottle with Seed and Sample)

(Seed Depletion in Bottle with Seed and Sample)

??

Page 90: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD(seeded)=D1 - D2

Sample Volume X 300 mL

D2 = D.O. Depletion Due to Just Seed

D2 =D.O. Depletion of a Sample of Seed

Volume of Seed Used

(Seed Depletion in Bottle with Seed and Sample)

D2 Must Be CalculatedBased on Bottle with Just Seed

In The Bottle with Just Seed

Page 91: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD(seeded)=D1 - D2

Sample Volume X 300 mL

D2 = D.O. Depletion Due to Just Seed

D2 =D.O. Depletion of a Sample of Seed

Volume of Seed Used

(Seed Depletion in Bottle with Seed and Sample)

In the Example

D2 =5.0 mg/L

10mL = 0.5 mg/L

Page 92: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD(seeded) =D1 - D2

Sample Volume X 300 mL

BOD(seeded) =

3.7 mg/L- 0.5 mg/L100 mL

X 300 mL

= 3.2 mg/L100 mL X 300 mL

= 9.6 mg/L

Page 93: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D2 = D.O. Depletion Due to Just Seed

D2 =D.O. Depletion of a Sample of Seed

Volume of Seed Used

(Seed Depletion in Bottle with Seed and Sample)

What If More Than 1mLWas Used for Seeding?

D2 = = 0.5 mg/L5.0 mg/L10 mL

Multiply by mL Used for Seeding

/mL

Page 94: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

1

2

ReferenceSolution

Organics

Seed

AndOrganics

Bacteria

5 mL

5 mL

1 mL

Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample. (300 mL BOD bottles were used)

Example Seeded BOD Calculation.

Seed Material Reference SampleVolume Used 5 mL 5 mLInitial D.O. 8.6 mg/L 8.7 mg/LFinal D.O. 5.8 mg/L 4.8 mg/L

Page 95: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D1 - D2BODSample Volume, mL X 300 mL

D1 = Depletion of Sample & Seed

seeded =

D1 = 8.7 mg/L - 4.8 mg/L

D1 = 3.9 mg/L

Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample. (300 mL BOD bottles were used)

Example Seeded BOD Calculation.

Seed Material Reference SampleVolume Used 5 mL 5 mLInitial D.O. 8.6 mg/L 8.7 mg/LFinal D.O. 5.8 mg/L 4.8 mg/L

Page 96: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D1 - D2BODSample Volume, mL X 300 mLseeded =

Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample. (300 mL BOD bottles were used)

Example Seeded BOD Calculation.

Seed Material Reference SampleVolume Used 5 mL 5 mLInitial D.O. 8.6 mg/L 8.7 mg/LFinal D.O. 5.8 mg/L 4.8 mg/L

D2 = Depletion of Seed (In Seeded Sample)

D2 =8.6 mg/L – 5.8 mg/L

5 mL

D2 = 2.8 mg/L 5 mL

= 0.56

Page 97: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

BOD = 3.34 mg/L5 mL X 300 mL

BOD = 200.4 mg/L

BOD = 3.9 mg/L5 mL X 300 mL- 0.56 mg/L

Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample. (300 mL BOD bottles were used)

Example Seeded BOD Calculation.

Seed Material Reference SampleVolume Used 5 mL 5 mLInitial D.O. 8.6 mg/L 8.7 mg/LFinal D.O. 5.8 mg/L 4.8 mg/L

D1 - D2BODSample Volume, mL X 300 mLseeded =

Page 98: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Calculate the B.O.D. of an de-chlorinated effluent sample given that 1mL of seed material was used in the sample. (300 mL BOD bottles were used)

Example #2 Seeded BOD Calculation.

Seed Material SampleVolume Used 10 mL 200 mLInitial D.O. 8.3 mg/L 8.5 mg/LFinal D.O. 5.6 mg/L 3.7 mg/L

D1 - D2BODSample Volume, mL X 300 mLseeded =

D1 = Depletion of Sample & Seed

D1 = 8.5 mg/L - 3.7 mg/L

D1 = 4.8 mg/L

Page 99: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Calculate the B.O.D. of an de-chlorinated effluent sample given that 1mL of seed material was used in the sample. (300 mL BOD bottles were used)

Example #2 Seeded BOD Calculation.

Seed Material SampleVolume Used 10 mL 200 mLInitial D.O. 8.3 mg/L 8.5 mg/LFinal D.O. 5.6 mg/L 3.7 mg/L

D1 - D2BODSample Volume, mL X 300 mLseeded =

D2 = Depletion of Seed (In Seeded Sample)

D2 =8.3 mg/L – 5.6 mg/L

10 mL

D2 = 2.7 mg/L 10 mL

= 0.27

Page 100: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Calculate the B.O.D. of an de-chlorinated effluent sample given that 1mL of seed material was used in the sample. (300 mL BOD bottles were used)

Example #2 Seeded BOD Calculation.

Seed Material SampleVolume Used 10 mL 200 mLInitial D.O. 8.3 mg/L 8.5 mg/LFinal D.O. 5.6 mg/L 3.7 mg/L

D1 - D2BODSample Volume, mL X 300 mLseeded =

BOD = 4.53 mg/L200 mL X 300 mL

BOD = 6.8 mg/L

BOD = 4.8 mg/L200 mL X 300 mL- 0.27 mg/L

Page 101: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Work Calculations on Separate PaperAnswers Given on Next Slides

BOD Practice Problems

1. Calculate the B.O.D. given:D.O. IN = 7.0 mg/LD.O. OUT (5-day) = 3.5 mg/L

Vol. Sample in B.O.D. bottle = 15 mL

2. Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample.

(300 mL BOD bottles were used)

Seed Material Reference SampleVolume Used 9 mL 5 mL

Initial D.O. 8.7 mg/L 8.6 mg/LFinal D.O. 5.1 mg/L 5.0 mg/L

Page 102: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

1. Calculate the B.O.D. given:D.O. IN = 7.0 mg/LD.O. OUT (5-day) = 3.5 mg/L

Vol. Sample in B.O.D. bottle = 15 mL

BOD = Depletion, mg/LSample Volume, mL X 300 mL

BOD = 3.5 mg/L15mL X 300 mL

BOD = 70 mg/L

BOD = 7.0 mg/L - 3.5 mg/L15 mL X 300 mL

Page 103: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

2. Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample.

(300 mL BOD bottles were used)

D1 - D2BOD Sample Volume, mL X 300 mL

D1 = Depletion of Sample & Seed

D2 = Depletion of Seed (In Seeded Sample)

seeded =

Seed Material Reference SampleVolume Used 9 mL 5 mL

Initial D.O. 8.7 mg/L 8.6 mg/LFinal D.O. 5.1 mg/L 5.0 mg/L

D1 = 8.6 mg/L - 5.0 mg/LD1 = 3.6 mg/L

D2 = 8.7 mg/L – 5.1 mg/L9 mL

D2 =3.6 mg/L

9 mL = 0.40

Page 104: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

D1 - D2BODSample Volume, mLX 300 mL

BOD = 3.2 mg/L5 mL X 300 mL

BOD = 192 mg/L

BOD = 3.6 mg/L - 0.40 mg/L5 mL X 300 mL

D1 = Depletion of Sample & SeedD2 = Depletion of Seed

seeded =

2. Calculate the B.O.D. of a reference sample given that 1mL of seed material was used in the reference sample.

(300 mL BOD bottles were used)Seed Material Reference Sample

Volume Used 9 mL 5 mLInitial D.O. 8.7 mg/L 8.6 mg/LFinal D.O. 5.1 mg/L 5.0 mg/L

Page 105: Dissolved Oxygen and Biochemical Oxygen Demand …€¦ · PPT file · Web view2016-02-26 · Blue light from an LED strikes the luminescent chemical on the sensor. The luminescent

Dissolved Oxygen Dissolved Oxygen and Biochemical and Biochemical Oxygen Demand Oxygen Demand

AnalysesAnalyses

Prepared ByMichigan Department of Environmental Quality

Operator Training and Certification Unit