Presentation Overview
Background of MALDI imaging Project motivation Problem statement Design criteria Design matrices Final design Budget evaluation Future work
![Page 3: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/3.jpg)
MALDI-MSI
www.mc.vanderbilt.edu
![Page 4: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/4.jpg)
Data Acquisition from MALDI-MSI
Alanine, neuropeptide in crab brain m/z = 1474.6
Valine, m/z = 1502.7
Alanine
Valine
Greg Barrett-Wilt, PhD., UW Biotechnology Center
![Page 5: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/5.jpg)
Benefits of MALDI-MSI Analysis of entire
sample in one reading
Previous knowledge of molecular composition is not necessary
Allows for investigation of disease formation, progression, and treatment
www.maldi-msi.org
![Page 6: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/6.jpg)
Matrix Application
Matrix application is vital for quality image resolution
Must contact sample as fine, liquid mist
Current procedure involves manual application with airbrush
100µm raster step
Greg Barrett-Wilt, PhD., UW Biotechnology Center
![Page 7: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/7.jpg)
Problem Statement
Sample preparation methods MALDI-MSI are difficult to control
Requires accuracy and precision
A device to apply a fine, uniform coating of light-absorbing compounds
![Page 8: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/8.jpg)
Project Motivation Objective 1:
Simplify matrix application, facilitate MALDI
Objective 2: Centralize a reliable
tool at UW Biotechnology Center
Implications for campus-wide, even global, research
www.buswire.ocr.wisc.edu
![Page 9: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/9.jpg)
Design Criteria
Spray an even coating of matrix over an 81cm x 123 cm tissue sample
Adjustable spray aperture, air pressure and positioning of the plate and sprayer
Enclosed in a casing and operable within a fume hood
![Page 10: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/10.jpg)
Design Matrix Adjustment
Several variables controlled by many independent components
Sprayer Method of application Enclosure Orientation
Created a component matrix to decide final design
![Page 11: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/11.jpg)
Component Matrix
VerticalHorizontalOrientation
Detached covering
Integrated covering
Fume hoodEnclosure
Stationary sprayer,Movable
plate
Movable sprayer,
Stationary plate
Method of Application
Produce irrigation system
Pneumatic sprayer
Nebulizer
Nozzle-valve with pressure vessel
Automatic Spray Gun Sprayer
Ideas Feature
![Page 12: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/12.jpg)
Criteria for ranking
Reliability Adjustability User interaction Cost Maintenance Ease of manufacture
Ranking System (1-5), highest = best
![Page 13: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/13.jpg)
Mega MatrixSpraying Component
2.354.104.303.452.251.00Total
214310.05Ease of
Manufacture
243310.10Maintenance
315340.15Cost
354450.20User
Interaction
255410.25Adjustability
254310.25Reliability
Produce irrigation system
Pneumatic sprayer
Automatic Spray Gun
Nozzle-valve and
pressure vessel
NebulizerRankCriteria
![Page 14: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/14.jpg)
Mega MatrixMethod of Application
3.252.851.00Total
130.05
Ease ofManufacture
330.10Maintenance
420.15Cost
430.20
User Interaction
330.25Adjustability
330.25
Reliability
Movable Plate Stationary Sprayer
Stationary Plate Movable Sprayer
RankCriteria
![Page 15: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/15.jpg)
Mega MatrixEnclosure
2.452.91.551.00Total
4350.05Ease of
Manufacture
5410.10
Maintenance
3250.15
Cost
4410.20User
Interaction
2510.25
Adjustability
---0.25Reliability
Removable CoverIntegrated CoverFume HoodRankCriteria
![Page 16: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/16.jpg)
Mega MatrixOrientation
1.25.51.00Total
--0.05Ease of
Manufacture
--0.10Maintenance
--0.15Cost
--0.20User
Interaction
--0.25
Adjustability
520.25
Reliability
Vertical SprayHorizontal
SprayRankCriteria
![Page 17: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/17.jpg)
Final Design Model
1. Automatic Spray Gun
2. Conveyor
3. DC Timing Motor
4. Integrated Polyethylene Box
14
3
2
![Page 18: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/18.jpg)
Budget Evaluation
$300.00 budget limit Breakdown
Air Spray Gun: $150.00 Motor/Conveyor: $45.00 Enclosure material: $45.00 Other small parts: $40.00
Total Estimate: $280.00 (under budget)
![Page 19: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/19.jpg)
Future Work
Final Design modifications Order materials Build prototype Test in client’s lab
![Page 20: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/20.jpg)
References Ashcroft, Alison. University of Leeds Astbury Centre for Structural
Molecular Biology. “An Introduction to Mass Spectrometry.” <http://www.astbury.leeds.ac.uk/>
Barrett-Wilt, Greg. University of Wisconsin Biotechnology Center. Personal interviews.
Caprioli Research Laboratory Center. Vanderbilt Medical Center.
<http://www.mcvanderbilt.edu> Khatib-Shahidi, Sheerin. “Direct Molecular Analysis of Whole-Body
Animal Tissue Section by Imaging MALDI Mass Spectrometry.” Analytical Chemistry. 2006, 78(18) 6448-6456.
Sugiura, Yuki. “Two-Step Matrix Application Technique to Improve Ionization Efficiency for MALDI-MSI.” Analytical Chemistry. 2006, 78(24) 8227-8235.
![Page 21: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center](https://reader035.vdocuments.net/reader035/viewer/2022062515/56649cc05503460f949874dc/html5/thumbnails/21.jpg)
Any Questions?