Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center Brenda Ogle, PhD Department of Biomedical Engineering

Upload: alfonso-bugg

Post on 16-Dec-2015

218 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Device to apply a coating on tissue

samples for MALDI imaging

Holly LiskeLaura PiechuraKellen SheedyJenna Spaeth

Amy Harms, PhDUW Biotechnology Center

Brenda Ogle, PhDDepartment of Biomedical Engineering

Page 2: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Presentation Overview

Background of MALDI imaging Project motivation Problem statement Design criteria Design matrices Final design Budget evaluation Future work

Page 3: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

MALDI-MSI

www.mc.vanderbilt.edu

Page 4: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Data Acquisition from MALDI-MSI

Alanine, neuropeptide in crab brain m/z = 1474.6

Valine, m/z = 1502.7

Alanine

Valine

Greg Barrett-Wilt, PhD., UW Biotechnology Center

Page 5: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Benefits of MALDI-MSI Analysis of entire

sample in one reading

Previous knowledge of molecular composition is not necessary

Allows for investigation of disease formation, progression, and treatment

www.maldi-msi.org

Page 6: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Matrix Application

Matrix application is vital for quality image resolution

Must contact sample as fine, liquid mist

Current procedure involves manual application with airbrush

100µm raster step

Greg Barrett-Wilt, PhD., UW Biotechnology Center

Page 7: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Problem Statement

Sample preparation methods MALDI-MSI are difficult to control

Requires accuracy and precision

A device to apply a fine, uniform coating of light-absorbing compounds

Page 8: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Project Motivation Objective 1:

Simplify matrix application, facilitate MALDI

Objective 2: Centralize a reliable

tool at UW Biotechnology Center

Implications for campus-wide, even global, research

www.buswire.ocr.wisc.edu

Page 9: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Design Criteria

Spray an even coating of matrix over an 81cm x 123 cm tissue sample

Adjustable spray aperture, air pressure and positioning of the plate and sprayer

Enclosed in a casing and operable within a fume hood

Page 10: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Design Matrix Adjustment

Several variables controlled by many independent components

Sprayer Method of application Enclosure Orientation

Created a component matrix to decide final design

Page 11: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Component Matrix

VerticalHorizontalOrientation

Detached covering

Integrated covering

Fume hoodEnclosure

Stationary sprayer,Movable

plate

Movable sprayer,

Stationary plate

Method of Application

Produce irrigation system

Pneumatic sprayer

Nebulizer

Nozzle-valve with pressure vessel

Automatic Spray Gun Sprayer

Ideas Feature

Page 12: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Criteria for ranking

Reliability Adjustability User interaction Cost Maintenance Ease of manufacture

Ranking System (1-5), highest = best

Page 13: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Mega MatrixSpraying Component

2.354.104.303.452.251.00Total

214310.05Ease of

Manufacture

243310.10Maintenance

315340.15Cost

354450.20User

Interaction

255410.25Adjustability

254310.25Reliability

Produce irrigation system

Pneumatic sprayer

Automatic Spray Gun

Nozzle-valve and

pressure vessel

NebulizerRankCriteria

Page 14: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Mega MatrixMethod of Application

3.252.851.00Total

130.05

Ease ofManufacture

330.10Maintenance

420.15Cost

430.20

User Interaction

330.25Adjustability

330.25

Reliability

Movable Plate Stationary Sprayer

Stationary Plate Movable Sprayer

RankCriteria

Page 15: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Mega MatrixEnclosure

2.452.91.551.00Total

4350.05Ease of

Manufacture

5410.10

Maintenance

3250.15

Cost

4410.20User

Interaction

2510.25

Adjustability

---0.25Reliability

Removable CoverIntegrated CoverFume HoodRankCriteria

Page 16: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Mega MatrixOrientation

1.25.51.00Total

--0.05Ease of

Manufacture

--0.10Maintenance

--0.15Cost

--0.20User

Interaction

--0.25

Adjustability

520.25

Reliability

Vertical SprayHorizontal

SprayRankCriteria

Page 17: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Final Design Model

1. Automatic Spray Gun

2. Conveyor

3. DC Timing Motor

4. Integrated Polyethylene Box

14

3

2

Page 18: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Budget Evaluation

$300.00 budget limit Breakdown

Air Spray Gun: $150.00 Motor/Conveyor: $45.00 Enclosure material: $45.00 Other small parts: $40.00

Total Estimate: $280.00 (under budget)

Page 19: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Future Work

Final Design modifications Order materials Build prototype Test in client’s lab

Page 20: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

References Ashcroft, Alison. University of Leeds Astbury Centre for Structural

Molecular Biology. “An Introduction to Mass Spectrometry.” <http://www.astbury.leeds.ac.uk/>

Barrett-Wilt, Greg. University of Wisconsin Biotechnology Center. Personal interviews.

Caprioli Research Laboratory Center. Vanderbilt Medical Center.

<http://www.mcvanderbilt.edu> Khatib-Shahidi, Sheerin. “Direct Molecular Analysis of Whole-Body

Animal Tissue Section by Imaging MALDI Mass Spectrometry.” Analytical Chemistry. 2006, 78(18) 6448-6456.

Sugiura, Yuki. “Two-Step Matrix Application Technique to Improve Ionization Efficiency for MALDI-MSI.” Analytical Chemistry. 2006, 78(24) 8227-8235.

Page 21: Device to apply a coating on tissue samples for MALDI imaging Holly Liske Laura Piechura Kellen Sheedy Jenna Spaeth Amy Harms, PhD UW Biotechnology Center

Any Questions?