Hamilton Venus Software Intelligent Liquid Sensing
2
Intelligent Liquid Sensing
1. Capacitive Liquid Level Detection (cLLD)
2. Pressure Liquid Level Detection (pLLD)
3. Anti Droplet Control (ADC)
4. Capacitance Clot Detection
5. Monitored Air Displacement (MAD)
6. Total Aspiration and Dispense Monitoring (TADM)
3
Intelligent Liquid Sensing
Conditions: Black tips or needles Conductive liquids only Clean or used tips All channels 96 head - A1 and H12 only 384 Head – A5 and P24 only Aspirate and or dispense steps
Capacitance Liquid Level Detection
More Details
6
Intelligent Liquid Sensing
1. Capacitive Liquid Level Detection (cLLD)
2. Pressure Liquid Level Detection (pLLD)
3. Anti Droplet Control (ADC)
4. Capacitance Clot Detection
5. Monitored Air Displacement (MAD)
6. Total Aspiration and Dispense Monitoring (TADM)
7
Intelligent Liquid Sensing
Pressure Liquid Level Detection
More Details
Conditions: Black / clear tips Any liquid Clean tips, aspirate only All channels
10
Intelligent Liquid Sensing
When to use cLLD & pLLD
Bubbles on liquid surface Use pLLD to find the liquid
surface cLLD could be triggered by the
bubbles
Aqueous layer under organic layer Use pLLD to find the non-
conductive organic layer Use cLLD to find the conductive
aqueous layer
11
Intelligent Liquid Sensing
1. Capacitive Liquid Level Detection (cLLD)
2. Pressure Liquid Level Detection (pLLD)
3. Anti Droplet Control (ADC)
4. Capacitance Clot Detection
5. Monitored Air Displacement (MAD)
6. Total Aspiration and Dispense Monitoring (TADM)
12
Intelligent Liquid Sensing
With ADC – the pressure increase inside the channel is measured and the piston moved to compensate.
Anti-Droplet Control
Without ADC – high vapour pressure of volatile liquid results in a pressure increase in the tip forcing the liquid out.
Plunger position
13
Intelligent Liquid Sensing
Prevent droplets on tips when pipetting highly volatile liquids
This is a standard feature in VENUS
Anti-Droplet Control
Conditions: Black / clear tips or needles Channels only Pre-wetting tips improves
performance
More Details
18
Intelligent Liquid Sensing
1. Capacitive Liquid Level Detection (cLLD)
2. Pressure Liquid Level Detection (pLLD)
3. Anti Droplet Control (ADC)
4. Capacitance Clot Detection
5. Monitored Air Displacement (MAD)
6. Total Aspiration and Dispense Monitoring (TADM)
19
Intelligent Liquid Sensing
The conductive signal after aspiration is measured. gap between tip and liquid - the capacitance
signal will vanish. clot - it bridges the gap and the signal remains,
resulting in an error message
Capacitance Clot Detection
More Details
Conditions: Black tips or needles Conductive liquids only Clean or used tips All channels
23
Intelligent Liquid Sensing
1. Capacitive Liquid Level Detection (cLLD)
2. Pressure Liquid Level Detection (pLLD)
3. Anti Droplet Control (ADC)
4. Capacitance Clot Detection
5. Monitored Air Displacement (MAD)
6. Total Aspiration and Dispense Monitoring (TADM)
24
Intelligent Liquid Sensing
Monitored Air Displacement
Tip Blocked
Pressure
Time
Correct Aspiration
Pressure
Threshold
Empty tip
Tip filling with liquid under vacuum
Tip filled with correct volume of liquid
Tip in liquid start aspirating
Tip blocked
Plunger stops moving, tip continues to fill
25
Intelligent Liquid Sensing
Monitored Air Displacement
Conditions: Black / clear tips Clean tips, aspirate only All channels Lower limit varies but >10µl MAD has only been optimised for
aqueous liquids
MAD can be switched on for individual aspiration steps in a method
The pressure threshold can be varied to detect whether each pipetting step was successful
More Details
28
Intelligent Liquid Sensing
1. Capacitive Liquid Level Detection (cLLD)
2. Pressure Liquid Level Detection (pLLD)
3. Anti Droplet Control (ADC)
4. Capacitance Clot Detection
5. Monitored Air Displacement (MAD)
6. Total Aspiration and Dispense Monitoring (TADM)
29
Intelligent Liquid Sensing
Total Aspiration and Dispense Monitoring
TADM increases the safety and robustness of pipetting processes by recording and monitoring pressure throughout the pipetting cycle. It’s main use in in clinical settings where a permanent record showing correct pipetting of every sample is needed.
Conditions: Black / clear tips Clean tips All channels 96 TADM head Requires additional
TADM software
30
Intelligent Liquid Sensing
Total Aspiration and Dispense Monitoring
Various pipetting errors change the pressure curves in interpretable ways
Aspirate
Dispense
31
Intelligent Liquid Sensing
Aspirate Dispense
Total Aspiration and Dispense Monitoring
The user sets both TADM tolerance bands and the required response if they are crossed
Bands are set , using test data, for every liquid, tip type and volume used