DULCOTEST ® sensor ICT 5 and ICT 5-IMA Sensor for electrical conductivity Assembly and operating instructions A2632 EN/DE 982358 BA DT 157 10/18 DE/EN Please carefully read these operating instructions before use. · Do not discard. The operator shall be liable for any damage caused by installation or operating errors. The latest version of the operating instructions are available on our homepage.

Upload: others

Post on 15-Feb-2021

9 views

Category:

Documents


0 download

TRANSCRIPT

  • DULCOTEST® sensor ICT 5 and ICT 5-IMASensor for electrical conductivity

    Assembly and operating instructions

    A2632

    EN/DE

    982358 BA DT 157 10/18 DE/EN

    Please carefully read these operating instructions before use. · Do not discard.The operator shall be liable for any damage caused by installation or operating errors.

    The latest version of the operating instructions are available on our homepage.

  • Overall Table of Con‐tentsEN DULCOTEST® sensor ICT 5 and ICT

    5-IMA Sensor for electrical conduc‐tivity..................................................... 41 Introduction.................................... 81.1 Measuring principle.................... 81.2 Construction and function of

    the sensor................................... 91.3 Nameplate................................ 121.4 Scope of delivery...................... 122 Safety and responsibility............. 132.1 Labelling of Warning Informa‐

    tion............................................ 132.2 User qualification...................... 152.3 General safety information....... 162.4 Intended use............................. 172.5 Information in the event of an

    emergency................................ 173 How to store and transport the

    sensor.......................................... 183.1 Storage..................................... 183.2 Transport.................................. 184 Assembly and installation............ 194.1 Installation of sensor type ICT

    5............................................... 204.2 Installation of sensor type ICT

    5-IMA........................................ 234.3 Electrical installation................. 245 Maintenance................................ 266 Rectifying faults and malfunc‐

    tions............................................. 277 Disposal of used parts................. 288 Accessories................................. 299 Technical data............................. 309.1 General data............................. 309.2 Certification/Test marks............ 31

    9.3 Permitted process pressure..... 3210 Dimensions................................ 33

    DE DULCOTEST® Sensor ICT 5 undICT 5-IMA Sensor für elektrolytischeLeitfähigkeit....................................... 361 Einleitung..................................... 401.1 Messprinzip.............................. 401.2 Aufbau und Funktion des Sen‐

    sors........................................... 411.3 Typenschild.............................. 441.4 Lieferumfang............................ 442 Sicherheit und Verantwortung..... 452.1 Kennzeichnung der Warnhin‐

    weise........................................ 452.2 Benutzer-Qualifikation.............. 472.3 Allgemeine Sicherheitshin‐

    weise........................................ 482.4 Bestimmungsgemäße Verwen‐

    dung......................................... 492.5 Angaben für den Notfall............ 493 So lagern und transportieren Sie

    den Sensor.................................. 503.1 Lagerung.................................. 503.2 Transport.................................. 504 Montage und Installation............. 514.1 Montage des Sensortyps ICT

    5............................................... 524.2 Montage des Sensortyps ICT

    5-IMA........................................ 554.3 Elektrische Installation.............. 565 Wartung....................................... 586 Fehler und Störungen beheben... 597 Altteileentsorgung........................ 608 Zubehör....................................... 619 Technische Daten....................... 629.1 Allgemeine Daten..................... 629.2 Zulassungen/Prüfzeichen......... 639.3 Zulässiger Prozessdruck.......... 64

    Overall Table of Contents

    2

  • 10 Abmessungen........................... 65

    Overall Table of Contents

    3

  • DULCOTEST® sensor ICT 5 and ICT 5-IMASensor for electrical conductivity

    Assembly and operating instructions

    A2632

    EN

    982358 BA DT 157 10/18 DE/EN

    Please carefully read these operating instructions before use. · Do not discard.The operator shall be liable for any damage caused by installation or operating errors.

    The latest version of the operating instructions are available on our homepage.

  • General non-discriminatory approach In order to make it easier to read, this docu‐ment uses the male form in grammatical struc‐tures but with an implied neutral sense. It isaimed equally at both men and women. Wekindly ask female readers for their under‐standing in this simplification of the text.

    Supplementary information

    Please read the supplementary information in its entirety.

    Information

    This provides important information relating to the correct operation of the unit or is intendedto make your work easier.

    Warning information

    Warning information includes detailed descriptions of the hazardous situation, see Ä Chapter 2.1‘Labelling of Warning Information’ on page 13.

    The following symbols are used to highlight instructions, links, lists, results and other elements inthis document:

    Tab. 1: More symbols

    Symbol Description

    Action, step by step.

    ⇨ Outcome of an action.

    Links to elements or sections of these instructions or other applicable docu‐ments.

    n List without set order.

    [Button] Display element (e.g. indicators).

    Operating element (e.g. button, switch).

    Supplemental directives

    5

  • Symbol Description

    ‘Display /GUI’ Screen elements (e.g. buttons, assignment of function keys).

    CODE Presentation of software elements and/or texts.

    Supplemental directives

    6

  • Table of contents1 Introduction............................................................................................................................. 8

    1.1 Measuring principle........................................................................................................ 81.2 Construction and function of the sensor......................................................................... 91.3 Nameplate.................................................................................................................... 121.4 Scope of delivery.......................................................................................................... 12

    2 Safety and responsibility....................................................................................................... 132.1 Labelling of Warning Information.................................................................................. 132.2 User qualification.......................................................................................................... 152.3 General safety information........................................................................................... 162.4 Intended use................................................................................................................. 172.5 Information in the event of an emergency.................................................................... 17

    3 How to store and transport the sensor.................................................................................. 183.1 Storage......................................................................................................................... 183.2 Transport...................................................................................................................... 18

    4 Assembly and installation...................................................................................................... 194.1 Installation of sensor type ICT 5................................................................................... 204.2 Installation of sensor type ICT 5-IMA........................................................................... 234.3 Electrical installation..................................................................................................... 24

    5 Maintenance.......................................................................................................................... 26

    6 Rectifying faults and malfunctions......................................................................................... 27

    7 Disposal of used parts........................................................................................................... 28

    8 Accessories........................................................................................................................... 29

    9 Technical data....................................................................................................................... 309.1 General data................................................................................................................. 309.2 Certification/Test marks................................................................................................ 319.3 Permitted process pressure......................................................................................... 32

    10 Dimensions........................................................................................................................... 33

    Table of contents

    7

  • 1 IntroductionThese operating instructions provide informa‐tion on the technical data and functions of theDULCOTEST® sensor for inductive conduc‐tivity, ICT 5 and ICT-IMA.

    Sensor type ICT 5 is designed for continuousflow operation.

    Sensor type ICT 5-IMA is integrated in animmersion tube for immersion in a storage tankand in an open channel.

    The sensor records the electrolytic conductivityof an aqueous process liquid online. Both typesof sensor work on the basis of an inductivemeasuring principle. The sensor is almostmaintenance-free due to the inductive meas‐uring process. Solid deposits and fatty or oilyfilms on the surface of the sensor have noimpact on its measuring precision. The sensoris particularly suitable for measuring high con‐ductivities of up to 2000 mS/cm. A stand-alonetemperature sensor (Pt1000) with a shortresponse time records the process tempera‐ture. The temperature sensor is enclosed withina protective stainless steel sleeve. Tempera‐ture correction of the conductivity measurementin the connected controller is provided for bythe temperature sensor (Pt1000).

    Typical applications

    The sensor is predominantly designed for appli‐cations in water treatment and for water moni‐toring, e.g.:

    n Contaminated waste water and processwater

    n Air conditioning plants and chillersn Blowdown control in cooling towers,n Flushing baths,n Vehicle washing systems,n Salt water desalination (feed),n Swimming pool water control.

    You can also use both types of sensor in conta‐minated water or in aggressive media, which ischemically inert to PVC/EPDM and/or PP/EPDM (depending on the design of the sensor).

    1.1 Measuring principleThe electrolytic conductivity is measured withan inductive sensor. The sensor contains atransmission and receiver coil. Sine alternatingvoltage feeds the transmission coil. A sampleflow is induced into the receiver coil dependingon the conductivity of the liquid to be meas‐ured. The sample flow is proportional to theconductivity of the medium.

    Introduction

    8

  • 1.2 Construction and function of the sensorThe inductive transducer of the two types of sensor ICT 5 and ICT 5-IMA consists of a hermeticallysealed polypropylene (PP) body. A transmission coil and a receiver coil are incorporated in thetransducer. A flow opening in the sensor (sensor eye) permits the measuring medium to flowthrough. There is galvanic isolation between the sample medium and the signal output. A stand-alone temperature sensor (Pt1000) with a short response time records the process temperature. Thetemperature sensor is enclosed within a protective stainless steel sleeve.

    A2617

    4

    3

    2

    15

    Fig. 1: Components

    1 PP sensor body2 Transmission coil3 Sensor eye through which the sample

    medium flows4 Receiver coil5 Temperature sensor, stand-alone, enclosed

    within a stainless steel sleeve

    Introduction

    9

  • Sensor type ICT 5

    The sensor type ICT5 is connected to a standard DN40 T-piece (on site) using a threaded socketand a union nut for operation in the flow.

    A2641

    Fig. 2: Sensor type ICT 5

    1. Threaded socket, G 1 1/2” (PVC or PP)2. Union nut (PVC or PP)

    Introduction

    10

  • Sensor type ICT5-IMA

    The sensor type ICT5-IMA is immersed by the fully assembled immersion tube (1 m) in the storagetank and channel.

    A2642

    Fig. 3: Sensor type ICT 5-IMA

    1. Fixed cable, 10 m2. Threaded connector M16, degree of protec‐

    tion, IP68 (up to 0.2 m)3. PP4. PP

    Introduction

    11

  • 1.3 Nameplate

    The cell constant K is needed for calibration on the transmitter/controller.

    A2637

    Fig. 4: ICT5 nameplate

    A2638

    Fig. 5: ICT5-IMA nameplate

    1.4 Scope of deliveryScope of delivery of sensor type ICT 5:

    n Sensor type ICT 5, order number 1095248n Straight solvent union PVC for adaptation

    of the sensor to a standard T-piece DN40,PVC (T-piece not included in the scope ofdelivery)

    n Assembly and operating instructions

    Scope of delivery of sensor type ICT 5-IMA:

    n Sensor type ICT 5-IMA, order number1095249

    n Assembly and operating instructions

    Introduction

    12

  • 2 Safety and responsibility2.1 Labelling of Warning Infor‐

    mationIntroduction

    These operating instructions provide informa‐tion on the technical data and functions of theproduct. These operating instructions providedetailed warning information and are providedas clear step-by-step instructions.

    The warning information and notes are categor‐ised according to the following scheme. Anumber of different symbols are used to denotedifferent situations. The symbols shown hereserve only as examples.

    DANGER!

    Nature and source of the dangerConsequence: Fatal or very seriousinjuries.

    Measure to be taken to avoid thisdanger.

    Description of hazard

    – Denotes an immediate threateningdanger. If the situation is disre‐garded, it will result in fatal or veryserious injuries.

    WARNING!

    Nature and source of the dangerPossible consequence: Fatal or veryserious injuries.

    Measure to be taken to avoid thisdanger.

    – Denotes a possibly hazardous sit‐uation. If the situation is disre‐garded, it could result in fatal orvery serious injuries.

    CAUTION!

    Nature and source of the dangerPossible consequence: Slight or minorinjuries. Material damage.

    Measure to be taken to avoid thisdanger.

    – Denotes a possibly hazardous sit‐uation. If the situation is disre‐garded, it could result in slight orminor injuries. May also be usedas a warning about materialdamage.

    Safety and responsibility

    13

  • NOTICE!

    Nature and source of the dangerDamage to the product or its surround‐ings.

    Measure to be taken to avoid thisdanger.

    – Denotes a possibly damaging sit‐uation. If the situation is disre‐garded, the product or an object inits vicinity could be damaged.

    Type of information

    Hints on use and additional information.

    Source of the information. Additionalmeasures.

    – Denotes hints on use and otheruseful information. It does not indi‐cate a hazardous or damaging sit‐uation.

    Safety and responsibility

    14

  • 2.2 User qualification

    WARNING!

    Danger of injury with inadequately qualified personnelThe operator of the system / equipment is responsible for ensuring that the qualifications arefulfilled.

    If inadequately qualified personnel work on the unit or loiter in the hazard zone of the unit,this could result in dangers that could cause serious injuries and material damage.

    – All work on the unit should therefore only be conducted by qualified personnel.– Unqualified personnel should be kept away from the hazard zone.

    The pertinent accident prevention regulations, as well as all other generally acknowledgedsafety regulations, must be adhered to.

    Training Definition

    Instructed personnel An instructed person is deemed to be a person who has beeninstructed and, if required, trained in the tasks assigned to him and pos‐sible dangers that could result from improper behaviour, as well ashaving been instructed in the required protective equipment and protec‐tive measures.

    Trained user A trained user is a person who fulfils the requirements made of aninstructed person and who has also received additional training specificto the system from the manufacturer or another authorised distributionpartner.

    Trained, qualified per‐sonnel

    A trained, qualified employee is deemed to be a person who is able toassess the tasks assigned to him and recognize possible hazardsbased on his training, knowledge and experience, as well as knowledgeof pertinent regulations. A trained, qualified employee must be able toperform the tasks assigned to him independently with the assistance ofdrawing documentation and parts lists. The assessment of a person'stechnical training can also be based on several years of work in the rel‐evant field.

    Safety and responsibility

    15

  • Training Definition

    Electrical technician An electrical technician is able to complete work on electrical systemsand recognise and avoid possible dangers independently based on histechnical training and experience as well as knowledge of pertinentstandards and regulations. An electrical technician must be able to per‐form the tasks assigned to him independently with the assistance ofdrawing documentation, parts lists, terminal and circuit diagrams. Theelectrical technician must be specifically trained for the working environ‐ment in which the electrical technician is employed and be conversantwith the relevant standards and regulations.

    Service The Service department refers to service technicians, who havereceived proven training and have been authorised by the manufacturerto work on the system.

    2.3 General safety information

    WARNING!

    Unauthorised access!Possible consequence: Fatal or veryserious injuries

    – Measure: Ensure that there can beno unauthorised access to thedevice

    – Only trained personnel may fit,install, service and operate thissensor

    CAUTION!

    Functional limitationsPossible consequence: Slight or minorinjuries. Material damage

    – Check the sensor regularly for dirtand contamination

    – Observe all applicable nationalregulations relating to mainte‐nance, service and calibrationintervals

    Safety and responsibility

    16

  • 2.4 Intended usen Only use the sensor to measure and regu‐

    late electrolytic conductivity in aqueousmedia, which are used in the applicationsdescribed in these operating instructions.

    n All other uses or modifications are pro‐hibited.

    n The sensor is not a safety component inthe sense of DIN EN ISO13849-1:2008-12. If there is a criticalprocess in your measurement and controlcircuit, then it is your responsibility to makesure this process is safe.

    2.5 Information in the event ofan emergency

    n In the event of an emergency, switch offthe controller

    n If liquid escapes from the continuous flowgauge, close the stopcocks on the inletand outlet installed by the customer

    n Observe the plant operator's safety infor‐mation before opening the continuous flowgauge

    Safety and responsibility

    17

  • 3 How to store and transport the sensorOriginal packaging

    User qualification: instructed user, seeÄ Chapter 2.2 ‘User qualification’ on page 15

    Damage to the product.

    n Only transport, ship and store the sensorin its original packaging.

    n Retain the packaging in its entiretyincluding the polystyrene inserts.

    3.1 StorageStorage and transport temperature:

    n complete sensor -20 °C ... 75 °C.

    Humidity: maximum 90 % relative air humidity,non-condensing.

    Other: No dust, no direct sunlight.

    3.2 TransportThe sensor should be transported in its originalpackaging and in compliance with the permis‐sible environmental conditions. No further spe‐cial conditions have to be observed in relationto transport.

    How to store and transport the sensor

    18

  • 4 Assembly and installationn User qualification, mechanical installation:

    trained and qualified personnel Ä Chapter2.2 ‘User qualification’ on page 15

    n User qualification, electrical installation:electrical technician Ä Chapter 2.2 ‘Userqualification’ on page 15

    WARNING!

    Danger from hazardous substances!Possible consequence: Fatal or veryserious injuries.

    Please ensure when handling haz‐ardous substances that you have readthe latest safety data sheets providedby the manufacture of the hazardoussubstance. The actions required aredescribed in the safety data sheet.Check the safety data sheet regularlyand replace, if necessary, as thehazard potential of a substance can bere-evaluated at any time based on newfindings.

    The system operator is responsible forensuring that these safety data sheetsare available and that they are kept upto date, as well as for producing anassociated hazard assessment for theworkstations affected.

    Installation site

    Ensure that there is unimpeded access for sub‐sequent calibration.

    The mounting must be secure and low-vibra‐tion.

    Avoid direct sunlight.

    Make sure that the medium flows correctlythrough and around the sensor.

    Maintain a minimum distance of 20 mm fromthe sensor to the tube wall when installing inpipework.

    Minimum compensation will be achieved withthe “Installation factor” parameter if you cannotadhere to these minimum spacings.

    Provide a representative installation site for typ‐ical conductivity and/or concentration whenimmersing in a tank.

    Assembly and installation

    19

  • 4.1 Installation of sensor type ICT 5Fitting position

    You can operate the sensor at any angle (0 ... 360°). The sensor is installed in a DN40 standard T-piece, as shown in Fig. 6 with regard to its flow and the minimum distances from the walls.

    1

    1

    3

    2

    A2618

    Fig. 6: Fitting position of sensor type ICT5 in the T-piece (DN40)

    1 ≧ 20 mm2 Flow direction3 Sensor

    Assembly and installation

    20

  • Hydraulic installation situation

    Please note the hydraulic fitting position indicated in Fig. 7.

    It is essential that the following are avoided:

    n Formation of gas bubblesn The device running dryn Stallingn Turbulence

    A2619

    Fig. 7: Hydraulic fitting position / tick = OK, cross = not OK

    Assembly and installation

    21

  • Bonding the straight solvent union into a T-piece

    Installing the sensor type ICT5 into PVC pipework. You can install the sensor into a PVC sectionusing the PVC straight solvent union supplied

    Colour Material

    Beige PP

    Grey PVC

    Use suitable adhesive, resistant to the media, for the materials to be bonded.

    A2639

    5

    4

    3

    2

    1

    Fig. 8: Sensor and straight solvent union in a T-piece

    1. Bond the straight solvent union (2) supplied into a standard T-piece (3) (on site).

    When doing so, adhere to the minimum distances from the walls.

    Assembly and installation

    22

  • Align the straight solvent union so that the sample medium flows through the sensor eye (5).

    Align the guide lip (1) of the adhesive connector towards the T-piece in the flow direction.

    Installing the sensor type ICT5 into polypropylene (PP) pipework

    2. You can install the sensor directly into an optionally available PP T-piece (order number1096349, not included in the scope of delivery). The PP T-piece already includes the sensoradapter.

    Assembling the sensor into the T-piece

    3. Place the sensor (5) into the T-piece so that the groove (4) on the sensor shaft engages withthe guide lip of the sensor adapter.

    4. Manually screw the union nut tightly onto the thread of the sensor adapter.

    5. Check the leak-tightness.

    4.2 Installation of sensor typeICT 5-IMA

    Installation place

    The transducer of sensor type ICT 5-IMA isfully integrated in the immersion pipe. Theimmersion pipe can either be connected to astorage tank or channel using the holes pro‐vided or using pipe clamps (on site).

    Select a suitable installation site for the repre‐sentative sample measurement.

    Assembly and installation

    23

  • 4.3 Electrical installation

    Do not shorten or mechanically damage the connecting cable.

    Once connected, calibrate the inductive conductivity and temperature sensor.

    Tab. 2: Terminal assignment, Dulcometer Compact

    Terminal Function Sensor ICT5 lead colours

    X1.1 Transmission coil, signal [bn], brown

    X1.2 Transmission coil, signal, earth [wh], white

    X2.1 Receiver coil, signal [pk], pink

    X2.2 Receiver coil, signal, earth Cable shield for X2.1, transparent

    XE2.2 Earth Complete cable shield, black

    XE4.1 Pt100x [gn] green

    XE4 Pt100x [ye] yellow

    The descriptions of the terminals refer to the ProMinent measuring and control unit, type DCCa. Theinformation about the colours relates to the leads of the sensor’s fixed data line.

    Connect the data line to the terminals of your ProMinent measuring and control unit, type DCCa. Indoing so, refer to the relevant operating instructions

    Connect the data line to the terminals of your measuring/control unit. Refer to the operatinginstructions for your measuring/control unit

    Assembly and installation

    24

  • Tab. 3: General assignment for further measuring/control units

    Function Sensor ICT5 lead colours

    Transmission coil, signal [bn], brown

    Transmission coil, signal, earth [wh], white

    Receiver coil, signal [pk], pink

    Receiver coil, signal, earth Cable shield for X2.1, transparent

    Earth Complete cable shield, black

    Pt100x [gn] green

    Pt100x [ye] yellow

    Connect the data line to the terminals of your measuring/control unit. Refer to the operatinginstructions for your measuring/control unit

    Assembly and installation

    25

  • 5 Maintenance

    WARNING!

    Danger from hazardous substances!Possible consequence: Fatal or veryserious injuries.

    Please ensure when handling haz‐ardous substances that you have readthe latest safety data sheets providedby the manufacture of the hazardoussubstance. The actions required aredescribed in the safety data sheet.Check the safety data sheet regularlyand replace, if necessary, as thehazard potential of a substance can bere-evaluated at any time based on newfindings.

    The system operator is responsible forensuring that these safety data sheetsare available and that they are kept upto date, as well as for producing anassociated hazard assessment for theworkstations affected.

    NOTICE!

    SolventOnly use water rather than a volatileorganic solvent to clean the sensor.

    1. Clean the sensor with water and a softbrush.

    2. Loosen tough films or deposits withdiluted hydrochloric acid and removewith a soft brush.

    3. Check the complete measuring chain(sensor, connecting line and transmitter/controller) should there be deviations ofthe measured values with an unknowncause.

    Maintenance

    26

  • 6 Rectifying faults and malfunctionsProblem Possible cause Measure

    No display of measured valueor signal output

    No power supply Check power supply, checkterminals

    Display of measured value is000 or signal output is 0% (e.g.4 mA)

    Sensor not immersed in themedium; storage tank level toolow

    Top up storage tank

    Flow fitting blocked Clean flow fitting

    Sensor faulty

    Incorrect or fluctuating meas‐ured value display

    Sensor not immersed deeplyenough

    Top up storage tank

    No mixing Ensure that the sample ismixed well, ensuring that thereis approx. 18 mm of free circu‐lation on all sides of the sensor

    Air bubbles Refer to the installation instruc‐tions

    Rectifying faults and malfunctions

    27

  • 7 Disposal of used partsn User qualification: instructed user, see

    Ä Chapter 2.2 ‘User qualification’on page 15

    NOTICE!

    Regulations governing the disposal ofused parts– Note the national regulations and

    legal standards that currentlyapply in your country

    The manufacturer will take back decontami‐nated used devices providing they are coveredby adequate postage.

    Decontaminate the device before returning it forrepair. To do so, remove all traces of haz‐ardous substances. Refer to the Material SafetyData Sheet for your feed chemical.

    A current Declaration of Decontamination isavailable to download on the ProMinent web‐site.

    Sign indicating EU collection system

    In accordance with the European Directive2012/19/EU on waste electrical and electronicequipment, this device features the symbolshowing a waste bin with a line through it. Thedevice must not be disposed of along withdomestic waste. To return the device, use thereturn and collection systems available andobserve the local legal requirements.

    Disposal of used parts

    28

  • 8 AccessoriesDescription Part

    number

    Straight solvent union (PVC) + O-ring, for installation of sensor type ICT5 into a PVCT-piece DN40

    1096348

    Union nut G 1 1/2”, PVC 1096351

    T-piece, PP, DN40 with integral sensor adapter* 1096349

    Union nut G 1 1/2”, PP 1096353

    * Features an anti-turn locking device; the sensor can only be installed in the right direction.

    Accessories

    29

  • 9 Technical data9.1 General data

    Parameter ICT 5 ICT 5-IMA

    Design For operation in the flow For immersion

    Measuring principle inductive

    Measuring range 0 ... 2000 mS/cm

    Precision within the measuring range

    0 ... 1 mS/cm ≤ 1%

    0 ... 10 mS/cm ≤ 0.5%

    0 ... 50 mS/cm ≤ 0.5%

    0 ... 100 mS/cm ≤ 0.5%

    0 ... 1000 mS/cm ≤ 1%

    0 ... 2000 mS/cm ≤ 1%

    Cell constant k = 6.25 1/cm

    Temperature sensor Pt1000, Class A

    t90 Temperaturea ≤ 6 seconds

    Permissible ambient temperature -10 ... +60 °C

    Permissible storage temperature -20 ... +75 °C

    Degree of protection b IP68

    Permitted medium temperaturewhen operating in pipework madeof:

    PVC: -10 ... +60 °C

    PP: -10 ... +80 °C

    -10 ... +60 °C

    Maximum process pressure with PP design

    at 20 °C 10 bar 0 bar

    at 60 °C 6 bar 0 bar

    at 80 °C 0 bar

    Technical data

    30

  • Parameter ICT 5 ICT 5-IMA

    at -10 ... +80 °C minimum -0.1 bar minimum -0.1 bar

    Sensor material, wetted

    Sensor body PP

    Protective sleeve, temperaturesensor

    Stainless steel 1.4571

    Seal EPDM

    Electrical connection Fixed cable, 10 m, PUR, lead

    Threaded cable connector M16, PBT/PA

    Length of immersion fitting 1000 mm

    a DIN EN 60751b DIN EN 60529

    9.2 Certification/Test marksThe materials wetted by the sample mediumare approved for use with food, physiologicallyharmless and listed by the FDA.

    Technical data

    31

  • 9.3 Permitted process pressureTemperature, pressure and the sample medium have an impact on the service life of the sensor.

    0

    8042

    10

    50403020100-10 60 70 80

    I

    IIA2624

    Fig. 9: Permitted process pressure type ICT 5 fitted in PP pipework

    I Pressure in bar.II Temperature in °C.

    Technical data

    32

  • 10 Dimensions

    I. II. III. IV.

    Ø40 Ø44,5

    Ø40

    22

    21 120 ± 10 ELA2620

    Fig. 10: Dimensions, all dimensions in mm. Type ICT 5-IMA... Immersion version

    I Fixed cable, 10 mII Threaded connector M16, degree of protec‐

    tion IP68 (up to 0.2 m), PBT/PAIII PPIV Stainless steel 1.4301, AISI 304

    Dimensions

    33

  • I.

    II.

    145

    ~811

    18

    36,5

    A2622

    Ø41Ø44,5

    67

    Fig. 11: Dimensions, all dimensions in mm. Type ICT 5

    I Threaded connector M16, degree of protec‐tion IP68 (up to 0.2 m), PBT/PA

    II PP

    Dimensions

    34

  • 982358, 1, en_GB

    © 2018

    ProMinent GmbHIm Schuhmachergewann 5 - 1169123 HeidelbergTelephone: +49 6221 842-0Fax: +49 6221 842-419Email: [email protected]: www.prominent.com

  • DULCOTEST® Sensor ICT 5 und ICT 5-IMASensor für elektrolytische Leitfähigkeit

    Montage- und Betriebsanleitung

    A2632

    DE

    982358 BA DT 157 10/18 DE

    Betriebsanleitung bitte zuerst vollständig durchlesen. · Nicht wegwerfen.Bei Schäden durch Installations- oder Bedienfehler haftet der Betreiber.

    Die neueste Version einer Betriebsanleitung ist auf unserer Homepage verfügbar.

  • Allgemeine Gleichbehandlung Dieses Dokument verwendet die nach derGrammatik männliche Form in einem neutralenSinn, um den Text leichter lesbar zu halten. Esspricht immer Frauen und Männer in gleicherWeise an. Die Leserinnen bitten wir um Ver‐ständnis für diese Vereinfachung im Text.

    Ergänzende Anweisungen

    Lesen Sie bitte die ergänzenden Anweisungen durch.

    Infos

    Eine Info gibt wichtige Hinweise für das richtige Funktionieren des Geräts oder soll IhreArbeit erleichtern.

    Warnhinweise

    Warnhinweise sind mit ausführlichen Beschreibungen der Gefährdungssituation versehen, sieheÄ Kapitel 2.1 „Kennzeichnung der Warnhinweise“ auf Seite 45.

    Zur Hervorhebung von Handlungsanweisungen, Verweisen, Auflistungen, Ergebnissen und anderenElementen können in diesem Dokument folgende Kennzeichnungen verwendet werden:

    Tab. 1: Weitere Kennzeichnung

    Kennzeichen Beschreibung

    Handlung Schritt-für-Schritt.

    ⇨ Ergebnis einer Handlung.

    Links auf Elemente bzw. Abschnitte dieser Anleitung oder mitgeltende Doku‐mente.

    n Auflistung ohne festgelegte Reihenfolge.

    [Taster] Anzeigeelemente (z. B. Signalleuchten).

    Bedienelemente (z. B. Taster, Schalter).

    Ergänzende Anweisungen

    37

  • Kennzeichen Beschreibung

    „Anzeige/GUI“ Bildschirmelemente (z. B. Schaltflächen, Belegung von Funktionstasten).

    CODE Darstellung von Softwareelementen bzw. Texten.

    Ergänzende Anweisungen

    38

  • Inhaltsverzeichnis1 Einleitung.............................................................................................................................. 40

    1.1 Messprinzip.................................................................................................................. 401.2 Aufbau und Funktion des Sensors............................................................................... 411.3 Typenschild.................................................................................................................. 441.4 Lieferumfang................................................................................................................ 44

    2 Sicherheit und Verantwortung............................................................................................... 452.1 Kennzeichnung der Warnhinweise............................................................................... 452.2 Benutzer-Qualifikation.................................................................................................. 472.3 Allgemeine Sicherheitshinweise................................................................................... 482.4 Bestimmungsgemäße Verwendung............................................................................. 492.5 Angaben für den Notfall................................................................................................ 49

    3 So lagern und transportieren Sie den Sensor....................................................................... 503.1 Lagerung...................................................................................................................... 503.2 Transport...................................................................................................................... 50

    4 Montage und Installation....................................................................................................... 514.1 Montage des Sensortyps ICT 5.................................................................................... 524.2 Montage des Sensortyps ICT 5-IMA............................................................................ 554.3 Elektrische Installation.................................................................................................. 56

    5 Wartung................................................................................................................................. 58

    6 Fehler und Störungen beheben............................................................................................ 59

    7 Altteileentsorgung................................................................................................................. 60

    8 Zubehör................................................................................................................................. 61

    9 Technische Daten................................................................................................................. 629.1 Allgemeine Daten......................................................................................................... 629.2 Zulassungen/Prüfzeichen............................................................................................. 639.3 Zulässiger Prozessdruck.............................................................................................. 64

    10 Abmessungen....................................................................................................................... 65

    Inhaltsverzeichnis

    39

  • 1 EinleitungDiese Betriebsanleitung beschreibt die techni‐schen Daten und Funktionen desDULCOTEST® Sensors für Induktive Leitfähig‐keit, ICT 5 und ICT-IMA.

    Der Typ ICT 5 ist für den Durchflussbetriebausgelegt.

    Der Typ ICT 5-IMA ist in ein Eintauchrohr integ‐riert, für das Eintauchen im Behälter und imoffenen Gerinne.

    Der Sensor erfasst online die elektrolytischeLeitfähigkeit einer wässrigen Prozessflüssig‐keit. Beide Sensortypen arbeiten nach deminduktiven Messprinzip. Durch das induktiveMessverfahren ist der Sensor nahezu war‐tungsfrei. Feste Ablagerungen und Fett- oderÖlfilme auf der Sensoroberfläche haben kaumEinfluss auf die Messgenauigkeit. Der Sensorsind insbesondere zum Messen von hohenLeitfähigkeiten bis 2000 mS/cm geeignet. Einfreistehender Temperatursensor (Pt1000) mitkurzer Ansprechzeit erfasst die Prozesstempe‐ratur. Der Temperatursensor ist in einer Edel‐stahlhülse geschützt. Durch den Temperatur‐sensor (Pt1000) kann die Temperaturkorrekturder Leitfähigkeitsmessung im angeschlossenenMess- und Regelgerät erfolgen.

    Typische Anwendungsgebiete

    Der Sensor ist vornehmlich für die Anwendungin der Wasseraufbereitung und zur Wasserü‐berwachung konzipiert, z. B.:

    n verschmutzte Ab- und Brauchwassern Klima- und Kühlanlagenn Absalzkontrolle in Kühltürmen,n Spülbäder,n Fahrzeugwaschanlagen,n Meerwasserentsalzung (Zulauf),n Schwimmbadwasserkontrolle.

    Sie können die beiden Sensortypen auch inverschmutzten Wässern oder in aggressivenMedien einsetzen die sich gegenüber PVC/EPDM bzw. PP/EPDM (ja nach Ausführung desSensors) chemisch inert verhalten.

    1.1 MessprinzipDie Messung der elektrolytischen Leitfähigkeiterfolgt mit einen induktiven Sensor. Der Sensorenthält im Sensorkopf eine Sende- und Emp‐fangsspule. Eine Sinus-Wechselspannungspeist die Sendespule. In Abhängigkeit von derLeitfähigkeit der zu messenden Flüssigkeit wirdein Messstrom in die Empfangsspule induziert.Der Messstrom ist proportional zur Leitfähigkeitdes Mediums.

    Einleitung

    40

  • 1.2 Aufbau und Funktion des SensorsDer induktive Messwertaufnehmer der beiden Sensortypen ICT 5 und ICT 5-IMA besteht aus einemhermetisch verschlossenen Körper aus Polypropylen (PP). Im Innerem des Messwertaufnehmersbefindet sich eine Sende- und eine Empfängerspule. Eine Durchflussöffnung in dem Sensor (Sen‐sorauge) erlaubt die Durchströmung mit dem Messmedium. Bedingt durch das Messprinzip bestehteine galvanische Trennung zwischen Messmedium und Signalausgang. Ein freistehender Tempera‐tursensor (Pt1000) mit kurzer Ansprechzeit erfasst die Prozesstemperatur. Der Temperatursensorist in einer schützenden Edelstahlhülse integriert.

    A2617

    4

    3

    2

    15

    Abb. 1: Bauteile

    1 Sensorkörper aus PP2 Sendespule3 Sensorauge, durchströmt vom Messme‐

    dium4 Empfängerspule5 Temperatursensor, freistehend, integriert in

    eine Edelstahlhülse

    Einleitung

    41

  • Sensortyp ICT 5

    Der Sensortyp ICT5 wird über eine Gewindebuchse und Überwurfmutter an ein Standard T-StückDN40 (bauseits) zum Betrieb im Durchfluss angebunden.

    A2641

    Abb. 2: Sensortyp ICT 5

    1. Gewindebuchse G 1 1/2“ (PVC bzw.PP)2. Überwurfmutter (PVC bzw. PP)

    Einleitung

    42

  • Sensortyp ICT5-IMA

    Der Sensortyp ICT5-IMA wird über das komplett montierte Tauchrohr (1 m) in Behälter und inGerinne eingetaucht.

    A2642

    Abb. 3: Sensortyp ICT 5-IMA

    1. Festkabel, 10 m2. Kabelverschraubung M16, Schutzart IP68

    (bis 0,2 m)3. PP4. PP

    Einleitung

    43

  • 1.3 Typenschild

    Die Zellenkonstante K wird für die Kalibrierung am Messumformer/Regler benötigt.

    A2637

    Abb. 4: Typenschild ICT5

    A2638

    Abb. 5: Typenschild ICT5-IMA

    1.4 LieferumfangLieferumfang Sensortyp ICT 5:

    n Sensor Typ ICT 5, Bestellnummer1095248

    n Klebemuffe PVC zur Adaption des Sen‐sors, an ein Standard-T-Stück DN40, PVC(T-Stück nicht im Lieferumfang enthalten)

    n Montage- und Betriebsanleitung

    Lieferumfang Sensortyp ICT 5-IMA:

    n Sensor Typ ICT 5-IMA, Bestellnummer1095249

    n Montage- und Betriebsanleitung

    Einleitung

    44

  • 2 Sicherheit und Verantwortung2.1 Kennzeichnung der Warn‐

    hinweiseEinleitung

    Diese Betriebsanleitung beschreibt die techni‐schen Daten und Funktionen des Produktes.Die Betriebsanleitung gibt ausführliche Warn‐hinweise und ist in klare Handlungsschritte auf‐gegliedert.

    Warnhinweise und Hinweise gliedern sich nachdem folgenden Schema. Hierbei kommen ver‐schiedene, der Situation angepasste, Piktogr‐amme zum Einsatz. Die hier aufgeführten Pik‐togramme dienen nur als Beispiel.

    GEFAHR!

    Art und Quelle der GefahrFolge: Tod oder schwerste Verlet‐zungen.

    Maßnahme, die ergriffen werden muss,um diese Gefahr zu vermeiden.

    Beschriebene Gefahr

    – Bezeichnet eine unmittelbar dro‐hende Gefahr. Wenn die Situationnicht gemieden wird, sind Tododer schwerste Verletzungen dieFolge.

    WARNUNG!

    Art und Quelle der GefahrMögliche Folge: Tod oder schwersteVerletzungen.

    Maßnahme, die ergriffen werden muss,um diese Gefahr zu vermeiden.

    – Bezeichnet eine möglicherweisegefährliche Situation. Wenn dieSituation nicht gemieden wird,können Tod oder schwerste Ver‐letzungen die Folge sein.

    VORSICHT!

    Art und Quelle der GefahrMögliche Folge: Leichte oder geringfü‐gige Verletzungen. Sachbeschädigung.

    Maßnahme, die ergriffen werden muss,um diese Gefahr zu vermeiden.

    – Bezeichnet eine möglicherweisegefährliche Situation. Wenn dieSituation nicht gemieden wird,können leichte oder geringfügigeVerletzungen die Folge sein. Darfauch für Warnung vor Sach‐schäden verwendet werden.

    Sicherheit und Verantwortung

    45

  • HINWEIS!

    Art und Quelle der GefahrSchädigung des Produkts oder seinerUmgebung.

    Maßnahme, die ergriffen werden muss,um diese Gefahr zu vermeiden.

    – Bezeichnet eine möglicherweiseschädliche Situation. Wenn dieSituation nicht gemieden wird,kann das Produkt oder etwas inseiner Umgebung beschädigtwerden.

    Art der Information

    Anwendungstipps und Zusatzinforma‐tion.

    Quelle der Information. ZusätzlicheMaßnahmen.

    – Bezeichnen Anwendungstipps undandere besonders nützliche Infor‐mationen. Es ist kein Signalwortfür eine gefährliche oder schäd‐liche Situation.

    Sicherheit und Verantwortung

    46

  • 2.2 Benutzer-Qualifikation

    WARNUNG!

    Verletzungsgefahr bei unzureichender Qualifikation des PersonalsDer Betreiber der Anlage/des Gerätes ist für die Einhaltung der Qualifikationen verantwort‐lich.

    Wenn unqualifiziertes Personal Arbeiten an dem Gerät vornimmt oder sich im Gefahrenbe‐reich des Gerätes aufhält, entstehen Gefahren, die schwere Verletzungen und Sachschädenverursachen können.

    – Alle Tätigkeiten nur durch dafür qualifiziertes Personal durchführen lassen.– Unqualifiziertes Personal von den Gefahrenbereichen fernhalten.

    Die einschlägigen Unfallverhütungsvorschriften sowie die sonstigen allgemein anerkanntensicherheitstechnischen Regeln einhalten.

    Ausbildung Definition

    unterwiesene Person Als unterwiesene Person gilt, wer über die übertragenen Aufgaben undmöglichen Gefahren bei unsachgemäßem Verhalten unterrichtet underforderlichenfalls angelernt, sowie über die notwendigen Schutzein‐richtungen und Schutzmaßnahmen belehrt wurde.

    geschulter Anwender Als geschulter Anwender gilt, wer die Anforderungen an eine unterwie‐sene Person erfüllt und zusätzlich eine anlagenspezifische Schulungbei dem Hersteller oder einem autorisierten Vertriebspartner erhaltenhat.

    ausgebildete Fachkraft Als ausgebildete Fachkraft gilt, wer aufgrund seiner Ausbildung, Kennt‐nisse und Erfahrungen sowie Kenntnis der einschlägigen Bestim‐mungen die ihm übertragenen Arbeiten beurteilen und möglicheGefahren erkennen kann. Eine ausgebildete Fachkraft muss in derLage sein, die ihr übertragenen Arbeiten unter Zuhilfenahme von Zeich‐nungsdokumentation und Stücklisten selbständig durchzuführen. ZurBeurteilung der fachlichen Ausbildung kann auch eine mehrjährigeTätigkeit auf dem betreffenden Arbeitsgebiet herangezogen werden.

    Sicherheit und Verantwortung

    47

  • Ausbildung Definition

    Elektrofachkraft Die Elektrofachkraft ist aufgrund ihrer fachlichen Ausbildung, Kennt‐nisse und Erfahrungen sowie Kenntnis der einschlägigen Normen undBestimmungen in der Lage, Arbeiten an elektrischen Anlagen auszu‐führen und mögliche Gefahren selbstständig zu erkennen und zu ver‐meiden. Eine Elektrofachkraft muss in der Lage sein, die ihr übertra‐genen Arbeiten unter Zuhilfenahme von Zeichnungsdokumentation,Stücklisten, Klemmen- und Schaltplänen selbständig durchzuführen.Die Elektrofachkraft ist speziell für das Arbeitsumfeld, in dem die Elekt‐rofachkraft tätig ist, ausgebildet und kennt die relevanten Normen undBestimmungen.

    Kundendienst Als Kundendienst gelten Servicetechniker, die von dem Hersteller fürdie Arbeiten an der Anlage nachweislich geschult und autorisiertwurden.

    2.3 Allgemeine Sicherheitshin‐weise

    WARNUNG!

    Unbefugter Zugriff!Mögliche Folge: Tod oder schwersteVerletzungen

    – Maßnahme: Sichern Sie das Gerätgegen unbefugten Zugriff

    – Der Sensor darf nur durch hierfürausgebildetes Personal montiert,installiert, gewartet und betriebenwerden

    VORSICHT!

    FunktionseinschränkungMögliche Folge: Leichte oder geringfü‐gige Verletzungen. Sachbeschädigung

    – Den Sensor regelmäßig auf Ver‐unreinigungen überprüfen

    – Die gültigen nationalen Vor‐schriften für Pflege-, Wartungs-und Kalibrierintervalle einhalten

    Sicherheit und Verantwortung

    48

  • 2.4 Bestimmungsgemäße Ver‐wendung

    n Der Sensor darf nur zum Messen undRegeln der elektrolytischen Leitfähigkeit inwässrigen Medien, der in dieser Betriebs‐anleitung beschriebenen Anwendungenverwendet werden.

    n Alle anderen Verwendungen oder einUmbau sind verboten.

    n Der Sensor ist kein Sicherheitsbauteil imSinne der DIN EN ISO 13849-1:2008-12.Sollte es sich bei Ihren Mess- und Regel‐kreis um einen kritischen Prozess handeln,dann liegt es in Ihrer Verantwortungdiesen Prozess abzusichern.

    2.5 Angaben für den Notfalln Im Notfall den Regler spannungsfrei

    machenn Falls aus der Durchlaufarmatur Flüssigkeit

    austritt, die bauseitig installierten Absperr‐hähne am Zu- und Ablauf schließen

    n Vor dem Öffnen der Durchlaufarmatur dieSicherheitshinweise des Anlagenbetrei‐bers beachten

    Sicherheit und Verantwortung

    49

  • 3 So lagern und transportieren Sie den SensorOriginalverpackung

    Benutzer-Qualifikation: unterwiesene Person,siehe Ä Kapitel 2.2 „Benutzer-Qualifikation“auf Seite 47

    Schädigung des Produkts.

    n Transportieren, versenden und lagern Sieden Sensor nur in der Originalverpackung.

    n Bewahren Sie die Verpackung komplettmit den Styroporteilen auf.

    3.1 LagerungLager- und Transporttemperatur:

    n kompletter Sensor -20 °C ... 75 °C.

    Feuchtigkeit: maximal 90 % relative Luftfeuch‐tigkeit, nicht kondensierend.

    Sonstige: Kein Staub, kein direktes Sonnen‐licht.

    3.2 TransportDer Transport sollte in der Originalverpackungund innerhalb der zulässigen Umweltbedin‐gungen erfolgen. Weitere Besonderheiten sindbeim Transport nicht zu beachten.

    So lagern und transportieren Sie den Sensor

    50

  • 4 Montage und Installationn Benutzer-Qualifikation, mechanische Mon‐

    tage: ausgebildete Fachkraft Ä Kapitel 2.2„Benutzer-Qualifikation“ auf Seite 47

    n Benutzer-Qualifikation, elektrische Installa‐tion: Elektrofachkraft Ä Kapitel 2.2„Benutzer-Qualifikation“ auf Seite 47

    WARNUNG!

    Gefährdung durch einen Gefahrstoff!Mögliche Folge: Tod oder schwersteVerletzungen.

    Beachten Sie beim Umgang mitGefahrstoffen, dass die aktuellenSicherheitsdatenblätter der Gefahrstoff-Hersteller vorliegen. Die notwendigenMaßnahmen ergeben sich aus demInhalt des Sicherheitsdatenblatts. Daaufgrund neuer Erkenntnisse, dasGefährdungspotenzial eines Stoffesjederzeit neu bewertet werden kann, istdas Sicherheitsdatenblatt regelmäßigzu überprüfen und bei Bedarf zuersetzen.

    Für das Vorhandensein und den aktu‐ellen Stand des Sicherheitsdatenblattsund die damit verbundene Erstellungder Gefährdungsbeurteilung der betrof‐fenen Arbeitsplätze ist der Anlagenbe‐treiber verantwortlich.

    Montageort

    Achten Sie auf eine leichte Zugänglichkeit fürdie spätere Kalibrierung.

    Die Befestigung muss sicher und vibrationsarmsein.

    Vermeiden Sie direkte Sonneneinstrahlung.

    Achten Sie auf eine gute Durch- und Umströ‐mung des Sensors.

    Halten Sie beim Einbau in eine Rohrleitungeinen Mindestabstand von 20 mm vom Sensorzur Rohrwandung ein.

    Wenn Sie diese Mindestabstände nicht ein‐halten können, dann kann mit dem Parameter„Einbaufaktor“ ein begrenzter Ausgleicherreicht werden.

    Beim Eintauchbetrieb in Becken müssen Sieeinen für die typische Leitfähigkeit bzw. Kon‐zentration repräsentativen Einbauort vorsehen.

    Montage und Installation

    51

  • 4.1 Montage des Sensortyps ICT 5Einbaulage

    Sie können den Sensor in jeder Winkelorientierung (0 ... 360°) betreiben. Der Sensor wird bezüglichseiner Anströmung und den Mindestabständen von Wandungen, wie in Abb. 6 gezeigt, in ein DN40Standard-T-Stück eingebaut.

    1

    1

    3

    2

    A2618

    Abb. 6: Einbaulage des Sensorstyps ICT5 im T Stück (DN40)

    1 ≧ 20 mm2 Strömungsrichtung3 Sensor

    Montage und Installation

    52

  • Hydraulische Einbausituation

    Beachten Sie die in Abb. 7 gezeigte hydraulische Einbausituation.

    Unbedingt vermeiden:

    n Bildung von Gasblasenn Trockenlaufenn Strömungsabrissen Verwirbelungen

    A2619

    Abb. 7: Hydraulische Einbausituation / Haken = OK, Kreuz = NOK

    Montage und Installation

    53

  • Einkleben der Klebemuffe in ein T-Stück

    Montage des Sensortyps ICT5 in einer Verrohrung aus PVC. Sie können den Sensor in ein PVC-T-Stück einbauen, unter Verwendung der mitgelieferten PVC-Klebemuffe

    Farbe Material

    Beige PP

    Grau PVC

    Verwenden Sie für die zu verklebenden Materialien geeignete und medienbeständige Kleber.

    A2639

    5

    4

    3

    2

    1

    Abb. 8: Sensor und Klebemuffe in Standard-T-Stück

    1. Kleben Sie die beiliegende Klebemuffe (2) in ein Standard-T-Stück (3) (bauseits) ein.

    Halten Sie dazu den Mindestabstand zu den Wandungen ein.

    Montage und Installation

    54

  • Richten Sie dazu die Klebemuffe so aus, dass das Messmedium durch das Sensorauge (5)fließt.

    Richten Sie dazu die Führungszapfen (1) des Klebestutzens in Durchflussrichtung zum T-Stück aus.

    Montage des Sensortyps ICT5 in einer Verrohrung aus Polypropylen (PP)

    2. Sie können den Sensor direkt in das optional erhältliche PP-T-Stück einbauen (Bestell‐nummer 1096349, nicht im Lieferumfang). Das PP-T-Stück enthält bereits den Sensora‐dapter.

    Montage des Sensors in das T-Stück

    3. Setzen Sie den Sensor (5) so in das T-Stück ein, dass die Nut (4) am Sensorschaft in denFührungszapfen des Sensoradapters greift.

    4. Schrauben Sie die Überwurfmutter per Hand fest auf das Gewinde des Sensoradapters auf.

    5. Überprüfen Sie die Dichtigkeit.

    4.2 Montage des SensortypsICT 5-IMA

    Einbauort

    Der Messwertaufnehmer des Sensortyps ICT5-IMA ist vollständig in das Eintauchrohr integ‐riert. Das Eintauchrohr kann entweder über dievorhandenen Bohrungen oder mit Rohrschellen(bauseits) an einen Behälter oder an einGerinne angebunden werden.

    Wählen Sie einen für die repräsentative Probe‐messung einen geeigneten Installationspunkt.

    Montage und Installation

    55

  • 4.3 Elektrische Installation

    Sie dürfen das Anschlusskabel nicht kürzen oder mechanisch beschädigen.

    Nach dem Anschluss müssen Sie den induktiven Leitfähigkeits- und Temperatursensorkalibrieren.

    Tab. 2: Klemmenzuordnung, Dulcometer Compact

    Klemme Funktion Sensor ICT5 Leiterfarben

    X1.1 Sendespule, Signal [bn], Braun

    X1.2 Sendespule, Signal, Erde [ws], Weiß

    X2.1 Empfangsspule, Signal [pk], Pink

    X2.2 Empfangsspule, Signal, Erde Kabelabschirmung für X2.1, trans‐parent

    XE2.2 Masse Kabelabschirmung gesamt,schwarz

    XE4.1 Pt100x [gn] Grün

    XE4 Pt100x [ge] Gelb

    Die Bezeichnungen der Klemmen beziehen sich auf das ProMinent-Mess-/Regelgerät, Typ DCCa.Die Angaben der Farben beziehen sich auf die Adern der fest installierten Datenleitung des Sen‐sors.

    Verbinden Sie die Datenleitung mit den Klemmen Ihres ProMinent-Mess-/Regelgerät, Typ DCCa.Beachten Sie hierbei die entsprechende Betriebsanleitung

    Verbinden Sie die Datenleitung mit den Klemmen Ihres Mess-/Regelgerätes. Beachten Siehierbei die Betriebsanleitung Ihres Mess-/Regelgerätes

    Montage und Installation

    56

  • Tab. 3: Allgemeine Zuordnung für weitere Mess-/Regelgeräte

    Funktion Sensor ICT5 Leiterfarben

    Sendespule, Signal [bn], Braun

    Sendespule, Signal, Erde [ws], Weiß

    Empfangsspule, Signal [pk], Pink

    Empfangsspule, Signal, Erde Kabelabschirmung für X2.1, transparent

    Masse Kabelabschirmung gesamt, schwarz

    Pt100x [gn] Grün

    Pt100x [ge] Gelb

    Verbinden Sie die Datenleitung mit den Klemmen Ihres Mess-/Regelgerätes. Beachten Siehierbei die Betriebsanleitung Ihres Mess-/Regelgerätes

    Montage und Installation

    57

  • 5 Wartung

    WARNUNG!

    Gefährdung durch einen Gefahrstoff!Mögliche Folge: Tod oder schwersteVerletzungen.

    Beachten Sie beim Umgang mitGefahrstoffen, dass die aktuellenSicherheitsdatenblätter der Gefahrstoff-Hersteller vorliegen. Die notwendigenMaßnahmen ergeben sich aus demInhalt des Sicherheitsdatenblatts. Daaufgrund neuer Erkenntnisse, dasGefährdungspotenzial eines Stoffesjederzeit neu bewertet werden kann, istdas Sicherheitsdatenblatt regelmäßigzu überprüfen und bei Bedarf zuersetzen.

    Für das Vorhandensein und den aktu‐ellen Stand des Sicherheitsdatenblattsund die damit verbundene Erstellungder Gefährdungsbeurteilung der betrof‐fenen Arbeitsplätze ist der Anlagenbe‐treiber verantwortlich.

    HINWEIS!

    LösungsmittelVerwenden Sie für die Reinigung desSensors keine flüchtigen organischenLösungsmittel, sondern ausschließlichWasser als Lösungsmittel.

    1. Reinigen Sie den Sensor mit Wasserund einer weichen Bürste.

    2. Hartnäckige Beläge bzw. Ablagerungenkönnen Sie z. B. mit verdünnter Salz‐säure anlösen und mit einer weichenBürste entfernen.

    3. Sollten Messwertabweichungen unbe‐kannter Ursache auftreten, müssen Siedie komplette Messkette (Sensor,Anschlussleitung und Messumformer/Regler) überprüfen, Ä Kapitel 6 „Fehlerund Störungen beheben“ auf Seite 59.

    Wartung

    58

  • 6 Fehler und Störungen behebenProblem mögliche Ursache Maßnahme

    Keine Messwertanzeige bzw.Signalausgang

    Spannungsversorgung fehlt Spannungsversorgung prüfen,Klemmen überprüfen

    Messwertanzeige 000 bzw.Signalausgang 0 % (z. B.4 mA)

    Sensor nicht in Medium einge‐taucht; Behälterniveau zuniedrig

    Behälter auffüllen

    Durchflussarmatur verstopft Durchflussarmatur reinigen

    Sensor defekt

    Falsche oder schwankendeMesswertanzeige

    Sensor nicht tief genug einge‐taucht

    Behälter auffüllen

    keine Durchmischung für gute Durchmischungsorgen, beim Sensor auf all‐seitig ca. 18 mm freie Umspü‐lung achten

    Luftblasen Einbauhinweise beachten

    Fehler und Störungen beheben

    59

  • 7 Altteileentsorgungn Benutzer-Qualifikation: unterwiesene

    Person, siehe Ä Kapitel 2.2 „Benutzer-Qualifikation“ auf Seite 47

    HINWEIS!

    Vorschriften Altteileentsorgung– Beachten Sie die zurzeit für Sie

    gültigen nationalen Vorschriftenund Rechtsnormen

    Der Hersteller nimmt die dekontaminierten Alt‐geräte bei ausreichender Frankierung der Sen‐dung zurück.

    Bevor Sie das Gerät einschicken, müssen Siedas Gerät dekontaminieren. Dazu müssen Siealle Gefahrenstoffe restlos entfernen. BeachtenSie dazu das Sicherheitsdatenblatt ihresDosiermediums.

    Eine aktuelle Dekontaminationserklärung stehtals Download auf der Homepage zur Verfü‐gung.

    Hinweis auf Sammelsystem EU

    Dieses Gerät ist entsprechend der europä‐ischen Richtlinie 2012/19/EU über Elektro- undElektronik-Altgeräte mit dem Symbol der durch‐gestrichenen Mülltonne gekennzeichnet. DasGerät darf nicht über den Hausmüll entsorgtwerden. Nutzen Sie für die Rückgabe die Ihnenzur Verfügung stehenden Rückgabe- und Sam‐melsysteme und beachten Sie die örtlichengesetzlichen Vorgaben.

    Altteileentsorgung

    60

  • 8 ZubehörBezeichnung Teile‐

    nummer

    Klebemuffe (PVC) + O-Ring, zum Einbau des Sensortyps ICT5 in ein PVC T-StückDN40

    1096348

    Überwurfmutter G 1 1/2", PVC 1096351

    T-Stück aus PP, DN40 mit integriertem Sensoradapter* 1096349

    Überwurfmutter G 1 1/2", PP 1096353

    * Mit Verdrehsicherung; der Sensor kann nur in der richtigen Ausrichtung eingebaut werden.

    Zubehör

    61

  • 9 Technische Daten9.1 Allgemeine Daten

    Parameter ICT 5 ICT 5-IMA

    Ausführung Für Betrieb im Durchfluss Für Eintauchen

    Messprinzip induktiv

    Messbereich 0 ... 2000 mS/cm

    Genauigkeit bei Messbereich

    0 ... 1 mS/cm ≤ 1 %

    0 ... 10 mS/cm ≤ 0,5 %

    0 ... 50 mS/cm ≤ 0,5 %

    0 ... 100 mS/cm ≤ 0,5 %

    0 ... 1000 mS/cm ≤ 1 %

    0 ... 2000 mS/cm ≤ 1 %

    Zellenkonstante k = 6,25 1/cm

    Temperatursensor Pt1000, Klasse A

    t90 Temperatura ≤ 6 Sekunden

    zulässige Umgebungstemperatur -10 ... +60 °C

    zulässige Lagertemperatur -20 ... +75 °C

    Schutzartb IP68

    zulässige Mediumstemperatur beiBetrieb in Rohrleitungen aus:

    PVC: -10 ... +60 °C

    PP: -10 ... +80 °C

    -10 ... +60 °C

    maximaler Prozessdruck bei Ausführung PP

    bei 20 °C 10 bar 0 bar

    bei 60 °C 6 bar 0 bar

    bei 80 °C 0 bar

    Technische Daten

    62

  • Parameter ICT 5 ICT 5-IMA

    bei -10 ... +80 °C minimal -0,1 bar minimal -0,1 bar

    Sensorwerkstoff, medienberührt

    Sensorkörper PP

    Schutzhülse, Temperaturfühler Edelstahl 1.4571

    Dichtung EPDM

    Elektrischer Anschluss Festkabel, 10 m, PUR, Litze

    Kabelverschraubung M16, PBT/PA

    Länge der Eintaucharmatur 1000 mm

    a DIN EN 60751b DIN EN 60529

    9.2 Zulassungen/PrüfzeichenDie vom Messmedium berührten Materialiensind für Lebensmittel zugelassen, physiologischunbedenklich und von der FDA gelistet.

    Technische Daten

    63

  • 9.3 Zulässiger ProzessdruckTemperatur, Druck und Messmedium beeinflussen die Lebensdauer des Sensors.

    0

    8042

    10

    50403020100-10 60 70 80

    I

    IIA2624

    Abb. 9: Zulässiger Prozessdruck Typ ICT 5 montiert in PP-Rohrleitungen

    I Druck in bar.II Temperatur in °C.

    Technische Daten

    64

  • 10 Abmessungen

    I. II. III. IV.

    Ø40 Ø44,5

    Ø40

    22

    21 120 ± 10 ELA2620

    Abb. 10: Abmessungen, alle Maße in mm. Typ ICT 5-IMA... Eintauchversion

    I Festkabel, 10 mII Kabelverschraubung M16, Schutzart IP68

    (bis 0,2 m), PBT/PAIII PPIV Edelstahl 1.4301, AISI 304

    Abmessungen

    65

  • I.

    II.

    145

    ~811

    18

    36,5

    A2622

    Ø41Ø44,5

    67

    Abb. 11: Abmessungen, alle Maße in mm. Typ ICT 5

    I Kabelverschraubung M16, Schutzart IP68(bis 0,2 m), PBT/PA

    II PP

    Abmessungen

    66

  • 982358, 1, de_DE

    © 2018

    ProMinent GmbHIm Schuhmachergewann 5 - 1169123 HeidelbergTelefon: +49 6221 842-0Telefax: +49 6221 842-215E-Mail: [email protected]: www.prominent.com

  • 982358, 1, en_GB

    © 2018

    ProMinent GmbHIm Schuhmachergewann 5 - 1169123 HeidelbergTelephone: +49 6221 842-0Fax: +49 6221 842-419Email: [email protected]: www.prominent.com

    Overall Table of ContentsTable of contents1 Introduction1.1 Measuring principle1.2 Construction and function of the sensor1.3 Nameplate1.4 Scope of delivery

    2 Safety and responsibility2.1 Labelling of Warning Information2.2 User qualification2.3 General safety information2.4 Intended use2.5 Information in the event of an emergency

    3 How to store and transport the sensor3.1 Storage3.2 Transport

    4 Assembly and installation4.1 Installation of sensor type ICT 54.2 Installation of sensor type ICT 5-IMA4.3 Electrical installation

    5 Maintenance6 Rectifying faults and malfunctions7 Disposal of used parts8 Accessories9 Technical data9.1 General data9.2 Certification/Test marks9.3 Permitted process pressure

    10 DimensionsInhaltsverzeichnis1 Einleitung1.1 Messprinzip1.2 Aufbau und Funktion des Sensors1.3 Typenschild1.4 Lieferumfang

    2 Sicherheit und Verantwortung2.1 Kennzeichnung der Warnhinweise2.2 Benutzer-Qualifikation2.3 Allgemeine Sicherheitshinweise2.4 Bestimmungsgemäße Verwendung2.5 Angaben für den Notfall

    3 So lagern und transportieren Sie den Sensor3.1 Lagerung3.2 Transport

    4 Montage und Installation4.1 Montage des Sensortyps ICT 54.2 Montage des Sensortyps ICT 5-IMA4.3 Elektrische Installation

    5 Wartung6 Fehler und Störungen beheben7 Altteileentsorgung8 Zubehör9 Technische Daten9.1 Allgemeine Daten9.2 Zulassungen/Prüfzeichen9.3 Zulässiger Prozessdruck

    10 Abmessungen