eeg383 measurement - chapter 2 - characteristics of measuring instruments

Upload: -

Post on 03-Jun-2018

224 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    1/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms1 of 26

    EEG 383 Electrical Measurements andInstrumentations

    CHAPTER 2

    Characteristics of Measuring Instruments

    Prof. Fawzy Ibrahim

    Electronics and Communication Department

    Misr International University (MIU)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    2/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms2 of 26

    Chapter Contents

    2.1 Static characteristics of instruments

    2.1.1 Accuracy and inaccuracy (measurement uncertainty)2.1.2 Precision/repeatability/reproducibility

    2.1.3 Tolerance

    2.1.4 Range or span

    2.1.5 Linearity

    2.1.6 Sensitivity of measurement

    2.1.7 Threshold2.1.8 Resolut ion

    2.1.9 Sensitiv ity to disturbance

    2.1.10 Hysteresis effects

    2.1.11 Dead space

    2.2 Dynamic characteristics of Instruments2.2.1 Zero order instrument

    2.2.2 First order instrument

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    3/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms3 of 26

    2.1 Static characteristics of Instruments If we have a thermometer in a room and its reading shows a temperature of

    20C, then it does not really matter whether the true temperature of the room is

    19.5C or 20.5C. Such small variations around 20C are too small to affectwhether we feel warm enough or not. Our bodies cannot discriminate between

    such close levels of temperature and therefore a thermometer with an

    inaccuracy of 0.5C is perfectly adequate.

    If we had to measure the temperature of certain chemical processes, however,

    a variation of 0.5C might have a significant effect on the rate of reaction or

    even the products of a process.

    A measurement inaccuracy much less than 0.5C is therefore clearly required. Accuracy of measurement is thus one consideration in the choice of instrument

    for a particular application.

    Other parameters such as sensitivity, linearity and the reaction to ambient

    temperature changes are further considerations.

    These attributes are collectively known as the static characteristics of

    instruments, and are given in the data sheet for a particular instrument. It isimportant to note that the values quoted for instrument characteristics in such a

    data sheet only apply when the instrument is used under specified standard

    calibration conditions.

    The static characteristics of measuring instruments are connected only with the

    static reading that the instrument settles down to.

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    4/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms4 of 26

    2.1 Static characteristics of Instruments

    Definitions: Measurand:the physical quantity being measured is called the measurand.

    Measuring system is the assembly that is used to measure the measurand

    (Method and Instruments).

    Is-Value:The real measured value (XIS).

    Should-Value:The value that it should be (VSH), reference value.

    Error is the A deviation between the Is-Value and the Should-Value of a

    measurand or Also it can be defined as the difference between the measured

    and the true value (as per standard). Arithmetic mean (or simply mean):Sum of all values divided by the number

    of quantities and is given by.

    n

    xxxX n

    .....21

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    5/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms5 of 26

    2.1 Static characteristics of Instruments (Continued)

    2.1.1 Accuracy and inaccuracy (measurement uncertainty)

    The accuracyof an instrument is a measure of how close the output reading of theinstrument is to the correct value or How close is the mean measurement of a

    series of trials to the true value?

    Where xnis the nth measured value .

    In practice, it is more usual to quote the inaccuracy figure rather than theaccuracy figure for an instrument.

    Inaccuracyis the extent to which a reading might be wrong, and is often quoted as

    a percentage of the full-scale (f.s.) reading of an instrument.

    If, for example, a pressure gauge of range 010 bar has a quoted inaccuracy of

    1.0% f.s. ( 1% of full-scale reading), then the maximum error to be expected

    in any reading is 0.1 bar. This means that when the instrument is reading 1.0bar, the possible error is 10% of this value.

    Thus, if we were measuring pressures with expected values between 0 and 1

    bar, we would not use an instrument with a range of 010 bar. The term

    measurement uncertainty is frequently used in place of inaccuracy.

    %100)(

    1

    sh

    shn

    X

    XxAccuracy

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    6/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms6 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.2 Precision/repeatability/reproducibilityPrecisionHow much do the measurements vary from trial to trial?

    If a large number of readings are taken of the same quantity by a high precision

    instrument, then the spread of readings will be very small.

    Precision is often, though incorrectly, confused with accuracy. High precision

    does not imply anything about measurement accuracy. A precise system is notnecessarily an accurate system and vice versa.

    Repeatability describes the closeness of output readings when the same input is

    applied repetitively over a short period of time, with the same measurement

    conditions, same instrument and observer, same location and same conditions

    of use maintained throughout.

    Reproducibility describes the closeness of output readings for the same input

    when there are changes in the method of measurement, observer, measuring

    instrument, location, conditions of use and time of measurement.

    Both terms thus describe the spread of output readings for the same input. This

    spread is referred to as repeatability if the measurement conditions are

    constant and as reproducibility if the measurement conditions vary.

    valuemeantheisX

    X

    Xxecision n

    %100

    )(1Pr

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    7/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms7 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.2 Precision/repeatability/reproducibility Figure 2.1 shows the results of tests on three industrial robots that were

    programmed to place components at the center of the concentric circles shown,

    and the black dots represent the points where each robot actually deposited

    components at each attempt.

    Both the accuracy and precision of Robot 1 are shown to be low in this trial.

    Robot 2 consistently puts the component down at approximately the same

    place but this is the wrong point. Therefore, it has high precision but low

    accuracy. Finally, Robot 3 has both high precision and high accuracy, becauseit consistently places the component at the correct target position.

    Fig. 2.1 Comparison of accuracy and precision (a) Robot 1; (b) Robot 2 and (c) Robot 3.

    (a) Low precision, low accuracy (b) High precision, low accuracy(c) High precision, high accuracy

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    8/26

    First digit (A)

    Second digit (B)

    Multiplier (C)

    Tolerance (D)

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms8 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.3 Tolerance Tolerance describes the maximum deviation of a manufactured component

    from some specified value.

    For instance, crankshafts are machined with a diameter tolerance quoted as so

    many microns (10-6 m).

    Another example, electric circuit components such as resistors have tolerances

    of perhaps 5%. One resistor chosen at random from a batch having a nominal

    value 1000W and tolerance 5% might have an actual value anywhere between

    950W and 1050 W. The colored bands that are found on a resistor as shown in Fig. 2.2 can be

    used to determine its resistance. The first and second bands of the resistor give

    the first two digits of the resistance, and the third band is the multiplier which

    represents the power of ten of the resistance

    value. The final band indicates what tolerance

    value (in %) the resistor possesses. The

    resistance value written in equation form is:

    Fig. 2.2 The resistors

    representation

    D%10AB C (2.1)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    9/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms9 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.2 Tolerance

    Fig. 2.2 (b) The color code for resistors.

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    10/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms10 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.4 Range or span

    The range or span of an instrument

    defines the minimum and maximum

    values of a quantity that the

    instrument is designed to measure.

    2.1.5 Linearity Linear device is the one in which the

    relation between its output reading is

    linearly proportional to the quantitybeing measured as shown Fig. 2.3 .

    The Xs marked on Fig. 2.3 show a

    plot of the typical output readings of

    an instrument when a sequence of

    input quantities are applied to it.

    The non-linearity is then defined asthe maximum deviation of any of the

    output readings marked X from this

    straight line.

    Non-linearity is usually expressed as a

    percentage of full-scale reading.

    Fig. 2.3 Instrument output characteristic.

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    11/26

    2.1.6 Sensitivity of measurement The sensitivity of measurement is a measure of the change in instrument output

    that occurs when the quantity being measured changes by a given amount.Thus, sensitivity is the ratio:

    The sensitivity of measurement is therefore the slope of the straight line drawn

    on Fig. 2.3.

    Example 2.1The following resistance values of a platinum resistance thermometer were

    measured at a range of temperatures. Determine the measurement sensitivity

    of the instrument in ohms/C.

    Solution

    If these values are plotted on a graph,

    the straight-line relationship betweenresistance change and temperature

    change is obvious, so it is a linear device.

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms11 of 26

    2.1 Static characteristics of Instruments (Continued)

    valuemeasuredinChange

    readingoutputinChangeySensitivit

    CySensitivit o/30

    7

    200230

    307314

    (2.2)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    12/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms12 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.7 Threshold If the input to an instrument is gradually increased from zero, the input will have

    to reach a certain minimum level before the change in the instrument outputreading is of a large enough magnitude to be detectable. This minimum level of

    input is known as the threshold of the instrument.

    Manufacturers may specify threshold for instruments as absolute values or as a

    percentage of full-scale readings.

    As an illustration, a car speedometer typically has a threshold of about 5 km/h.

    This means that, if the vehicle starts from rest and accelerates, no outputreading is observed on the speedometer until the speed reaches 5 km/h.

    2.1.8 Resolution How finely can we and/or the instrument separate one value from another thats

    close to it?

    Resolution is specified as an absolute value or a percentage of f.s. deflection.

    One of the major factors influencing the resolution of an instrument is how finely

    its output scale is divided into subdivisions.

    For instance, if we have 4-bit and 8-bit multimeters to show the range from 0-

    10V, the best resolution for both of them:

    for 4-bit the resolution is: for 8-bit8

    1039.0625 .

    2mVmV625

    2

    104

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    13/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms13 of 26

    2.1 Static characteristics of Instruments (Continued) Using a car speedometer as an example again, this has subdivisions of

    typically 20 km/h. This means that when the needle is between the scale

    markings, we cannot estimate speed more accurately than to the nearest 4km/h. This figure of 4 km/h thus represents the resolution of the instrument.

    2.1.9 Sensitivity to disturbance All calibrations and specifications of an instrument are only valid under

    controlled conditions of temperature, pressure etc. These standard ambient

    conditions are usually defined in the instrument specification.

    As variations occur in the ambient temperature, pressure, etc., the sensitivity todisturbance is a measure of the magnitude of this change.

    Such environmental changes affect instruments in two main ways, known as

    zero drift and sensitivity drift.

    1. Zero drift or biasdescribes the effect where the zero reading of an instrument

    is modified by a change in ambient conditions. This causes a constant error

    that exists over the full range of measurement of the instrument. The mechanical form of bathroom scale is a common example of an instrument

    that is prone to bias. It is quite usual to find that there is a reading of perhaps

    1 kg with no one stood on the scale.

    If someone of known weight 70 kg were to get on the scale, the reading would

    be 71 kg, and if someone of weight 100 kg, the reading would be 101 kg.

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    14/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms14 of 26

    2.1 Static characteristics of Instruments (Continued)

    2.1.9 Sensitivity to disturbance

    1. Zero dr ift or bias If a thumbwheel is usually provided that can be turned until the reading is zero

    with the scales unloaded, thus removing the bias.

    Zero drift is also commonly found in instruments like voltmeters that are

    affected by ambient temperature changes. Typical units by which such zero drift

    is measured are volts/C.

    A typical change in the output characteristic of an instrument subject to zero

    drift is shown in Figure 2.4(a).

    2. Sensitivity drift(also known as scale factor drift) defines the amount by which

    an instruments sensitivity of measurement varies as ambient conditions

    change.

    Figure 2.4(b) shows what effect sensitivity drift can have on the output

    characteristic of an instrument. Sensitivity drift is measured in units of the form(angular degree/bar)/C.

    If an instrument suffers both zero drift and sensitivity drift at the same time, then

    the typical modification of the output characteristic is shown in Figure 2.4(c).

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    15/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms15 of 26

    2.1 Static characteristics of Instruments (Continued)

    2.1.9 Sensitivity to disturbance

    Fig. 2.4 Effects of disturbance: (a) zero drift; (b) sensitivity drift; (c) Both drifts.

    (a)

    (b)

    (c)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    16/26

    Example 2.2

    The A spring balance is calibrated in an environment at a temperature of 20Cand has the deflection/load characteristic (Table 1).

    It is then used in an environment at a temperature of 30C and the following

    deflection/ load characteristic is measured.

    Determine per C change in ambient temperature the following:

    (a) The zero drift. (b) The sensitivity drift.

    Solution

    At 20C, deflection/load characteristic is a straight line. Sensitivity = 20 mm/kg.

    At 30C, deflection/load characteristic is still a straight line. Sensitivity = 22

    mm/kg.

    Bias (zero drift) = 5mm (the no-load deflection)

    Sensitivity drift = 2 mm/kg

    (a) Zero drift/C = 5/10 = 0.5 mm/C

    (b) Sensitivity drift/C = 2/10 = 0.2 (mm per kg)/C

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms16 of 26

    2.1 Static characteristics of Instruments (Continued)

    2.1.9 Sensitivity to disturbance

    Table 1 at 20C Table 2 at 30C

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    17/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms17 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.10 Hysteresis effects If the input measured quantity to the instrument is steadily increased from a

    negative value, the output reading varies in the manner shown in curve (a) ofFig. 2.5.

    If the input variable is then steadily decreased, the output varies in the manner

    shown in curve (b) of Fig. 2.5.

    The non-coincidence between

    these loading and unloading

    curves is known as hysteresis. Two quantities are defined,

    maximum input hysteresis and

    maximum output hysteresis,

    as shown in Fig. 2.5. These

    are normally expressed as a

    percentage of the full-scale

    input or output reading

    respectively.

    Fig. 2.5 Instrument characteristic with hysteresis.

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    18/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms18 of 26

    2.1 Static characteristics of Instruments (Continued)2.1.11 Dead space Dead space is defined as the range of different input values over which there is

    no change in output value. Any instrument that exhibits hysteresis also displaysdead space, as marked on Figure 2.6.

    Some instruments that do not

    suffer from any significant

    hysteresis can still exhibit a

    dead space in their output

    characteristics.

    Fig. 2.6 Instrument characteristic with dead

    space.

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    19/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms19 of 26

    2.2 Dynamic characteristics of Instruments The dynamic characteristics of a measuring instrument describe its behavior

    between the time a measured quantity changes value and the time when the

    instrument output attains a steady value in response. As with static characteristics, any values for dynamic characteristics quoted in

    instrument data sheets only apply when the instrument is used under specified

    environmental conditions.

    Outside these calibration conditions, some variation in the dynamic parameters

    can be expected.

    In any linear, time-invariant measuring system, the following general relation

    can be written between input (x) and output (y) for time t > 0:

    where x is the measured quantity, y is the output reading and a0. . . an, b0. . .

    bmare constants.

    If we limit consideration to that of step changes in the measured quantity only,

    then equation (2.3) reduces to:

    xbdt

    dxb

    td

    xdb

    td

    xdb

    td

    xdb

    yadt

    dya

    td

    yda

    td

    yda

    td

    yda

    m

    m

    mm

    m

    mm

    m

    m

    n

    n

    nn

    n

    nn

    n

    n

    012

    2

    21

    1

    1

    012

    2

    21

    1

    1

    ....

    ....

    (2.3)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    20/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms20 of 26

    2.2 Dynamic characteristics of Instruments (Continued)

    2.2.1 Zero Order Instrument If all the coefficients a1. . . another than a0in equation (2.4) are assumed zero,

    then:

    where K is a constant known as the instrument sensitivity as defined earlier. Any instrument that behaves according to equation (2.5) is said to be of zero

    order type.

    Following a step change in the measured quantity at time t, the instrument

    output moves immediately to a new value at the same time instant t, as shown

    in Figure 2.7.

    A potentiometer, which measures motion, is a good example of such aninstrument, where the output voltage changes instantaneously as the slider is

    displaced along the potentiometer track.

    xbya

    dt

    dya

    td

    yda

    td

    yda

    td

    yda

    n

    n

    nn

    n

    nn

    n

    n 0012

    2

    21

    1

    1 ....

    (2.4)

    kxyxa

    byorxbya

    0

    000

    (2.5)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    21/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms21 of 26

    2.2 Dynamic characteristics of Instruments (Continued)

    Fig. 2.7 Zero order instrument characteristic. Fig. 2.8 First order instrument characteristic

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    22/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms22 of 26

    2.2 Dynamic characteristics of Instruments (Continued)2.2.1 First Order Instrument If all the coefficients a2 . . . an except for a0 and a1 are assumed zero in

    equation (2.4) then:

    Any instrument that behaves according to equation (2.6) is known as a first

    order instrument.

    If equation (2.6) is solved analytically, the output quantity y in response to a

    step change in x at time t varies with time in the manner shown in Fig. 2.8. Thetime constant of the step response is the time taken for the output quantity, y to

    reach 63% of its final value.

    A large number of other instruments also belong to this first order class: this is

    of particular importance in control systems where it is necessary to take

    account of the time lag that occurs between a measured quantity changing in

    value and the measuring instrument indicating the change.

    Fortunately, the time constant of many first order instruments is small relative to

    the dynamics of the process being measured, and so no serious problems are

    created.

    (2.6)xbyadt

    dya 001

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    23/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms23 of 26

    2.2 Dynamic characteristics of Instruments (Continued)2.2.1 First Order Instrument

    Example 2.3

    A balloon is equipped with temperature and altitude measuring instruments andhas radio equipment that can transmit the output readings of these instruments

    back to ground. The balloon is initially anchored to the ground with the

    instrument output readings in steady state. The altitude-measuring instrument

    is approximately zero order and the temperature transducer first order with a

    time constant, of 15 seconds. The temperature on the ground, T0, is 10C and

    the temperature Txat an altitude of x meters is given by the relation:

    (a) If the balloon is released at time zero, and thereafter rises upwards at a

    velocity of V = 5 meters/second, draw a table showing the temperature and

    altitude measurements reported at intervals of 10 seconds over the first 50

    seconds of travel. Show also in the table the error in each temperature reading.

    (b) What temperature does the balloon report at an altitude of 5000 meters?

    Solution

    (a) The transducer is assumed to be a first order differential. Let the temperature

    reported by the balloon at some general time t be Tr. Then Tx is related to Trby

    the relation:

    xTTx 01.00 (2.7)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    24/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms24 of 26

    2.2 Dynamic characteristics of Instruments (Continued)2.2.1 First Order Instrument

    Example 2.3 Solution

    Where in Eq. (2.6) the output, y = Tr(the reported reading), the input, x = Tx(the temperature Txat an altitude of x ), the constants, a1= , a0= 1, and b0= 1.

    Substitute for Txfrom Eq.(2.7), we get:

    The solution of the differential (2.9) consists of two parts: The transient or

    complementary function part (TrCF) when Tx= 0, and the particular integral part

    (TrPI) in the form:

    The transient or complementary function part of the solution of Eq. (2.8) whenTx= 0 is given by:

    xrr TT

    dtdT (2.8)

    ttxTdt

    dTr

    r

    05.010)5(01.01001.010 (2.9)

    0 rr T

    dt

    dT

    15// tt

    rCF AeAeT

    rPIrCFr TTT (2.10)

    (2.11)

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    25/26

  • 8/12/2019 EEG383 Measurement - Chapter 2 - Characteristics of Measuring Instruments

    26/26

    Prof Fawzy IbrahimEEG383 Ch.2 Charst of Instrms26 of 26

    2.2 Dynamic characteristics of Instruments (Continued)2.2.1 First Order Instrument

    Example 2.3 Solution

    Thus, the whole solution is given by:

    Using Eq. (2.15) to calculate Trfor various values of t, the following table can be

    constructed:

    (b) At 5000 m, t = 1000 seconds. Calculating Tr from Eq. (2.15) we get:

    (2.15)teTTT t

    rPIrCFr 05.075.1075.0 15/

    CxeT o

    r 25.39100005.075.1075.0 15/1000