efecto de la luminancia en la anisocoria y en el

1
EFECTO DE LA LUMINANCIA EN LA ANISOCORIA Y EN EL DIAMETRO PUPILAR MESÓPICO Y ESCOTÓPICO SÁNCHEZ- RAMOS C, PUELL MARIN M C, PÉREZ CARRASCO MJ, LANGA MORAGA A.,PELAEZ MARTINEZ T , JIMÉNEZ GARCIA I, DEL MORAL MARTINEZ I DEPARTAMENTO DE ÓPTICA. ESCUELA UNIVERSITARIA DE ÓPTICA. UNIVERSIDAD COMPLUTENSE DE MADRID 80 Congreso Internacional de la Sociedad Española de Oftamología. Córdoba. Sep-Oct 2004. E-mail: [email protected] BIBLIOGRAFÍA 1. Kaufman PL. Adler Fisiología del ojo. Aplicación Clínica. Mosby, Elsevier . 2003 2. Loewenfeld IE Newsome DA: Iris mechanics I. Influence of pupil diameter on dynamics of pupillary movements, Am J Ophthalmol 71: 347, 1971. 3. Loewnstein O Kawabata H, Loewenfield I: the pupil as indicator of retinal activity, Am J Ophthalmol 57:569, 1964. 4. Loewenfeld IE: The reaction to near vision. InThe pupil anatomy physiology and clinical application, vol I. Ames, Iowa and Detroid, Mich, 1993, Iowa State University Press and Wayne State University Press. INTRODUCCIÓN El movimiento de la pupila en respuesta a los cambios de intensidad de la luz optimiza la iluminación de la retina para maximizar la perceción visual. El diámetro de la pupila contribuye a mejorar la calidad de la imagen retiniana cuando este, en estado normal, es pequeño (1) . Clínicamente, la pupila actúa como un indicador subjetivo de la cantidad de luz trasducida al sistema visua, se utiliza para monitorizar la sensibilidad de la retina a la luz. La contracción pupilar transitoria en respuesta a un estímulo lumínico puede reflejar la integridad de la retina y del nervio óptico y sirve para el diagnostico objetivo de patologias visuales. Cualquier lesión en alguno de los puntos de la vía visual reducirá la amplitud del movimiento pupilar en respuesta a la luz (1, 2, 3) . Según Loewnfeld (2, 4) , la amplitud del movimiento pupilar aumenta de forma proporcional al logarítmo de la intensidad lumínica del estímulo, mientras que, el tiempo de latencia del reflejo fotomotor (tiempo desde el inicio del estímulo hasta el inicio de la contracción pupilar) se acorta (1) . En condiciones de iluminacion mesópica y escotópica, la dilatación de la pupila maximiza la incidencia de fotones en la retina, lo que además complementa los mecanismos lentos de adaptación a la oscuridad que implican el control retiniano de los fotorreceptores y de las células bipolares (1) . Los valores medios de los diámetros pupilares y las desviaciones estándar se muestran en la Tabla2 .La Gráfica 1 representa los valores del diámetro pupilar para cada uno de los ojos para los tres niveles de iluminación. CONCLUSIÓN Se presentan diferencias significativas del tamaño pupilar entre los distintos niveles de iluminación mesópico y escotópico estudiados. Las respuestas de ambos ojos son equilibradas no habiendose encontrado cambios significativos en la anisocoria. AGRADECIMIENTO Este trabajo forma parte del proyecto de investigación 5280015 patrocinado por el INSTITUTO MAPFRE -SEGURIDAD VIAL. Caracterización de la función visual de conductores de vehículos” OBJETIVO Determinar los cambios del diámetro pupilar y de la anisocoria en condiciones mesópicas y escotópicas en una amplia población MATERIAL Y MÉTODO Se ha realizado un estudio prospectivo del diámetro pupilar de 64 pacientes (20 3 años) utilizando el pupilómetro binocular de infrarrojos P2000 (Procyon Instruments Ltd).. El grado de anisocoria para las tres condiciones de iluminación ha sido: 3,85% para condiciones escotópicas (0,02 lux) 4,98% para el nivel mesópico bajo (----lux) 4,25%para las condiciones mesópicas altas ( 10,6 lux) No hay diferencias significativas entre las anisocorias para los tres niveles de iluminación. En la Figura podemos observar el diámetro pupilar obtenido para cada uno de los ojos en los diferentes niveles de iluminación Tabla1 : Diámetros pupilares para las diferentes condiciones de iluminación Figura 2 : Pupilómetro binocular de infrarrojos P2000 (Procyon Instruments Ltd).. Figura 1 : Imagen de la pantalla visualizada para cada componente de la muestra Niveles de iluminación (lux) 0,02 (x σ) 0,15 (x σ) 10,5 (x σ) Ojo Derecho 6,72 0,84 5,24 1.07 3,73 0.60 Ojo Izquierdo 6,63±0,86 5,22 0.84 3,84 0.89 Tabla : valores medios y desviaciones estándar de los diámetros pupilares en los tres niveles de iluminación. Grafica 1: Diámetro pupilar (mm) para cada uno de los ojos en los tres niveles de iluminación Al aumentar la luminancia hasta 0.15 lux (condiciones mesópicas bajas) se produjo una miosis del 22.02% y del 22.44% para el ojo derecho e izquierdo respectivamente. La miosis desde el nivel anterior fue del 28.03% y del 26.44% para el ojo derecho y el ojo izquierdo respectivamente. Existen diferencias significativas del diámetro pupilar en función de la iluminación ambiente. No hay diferencias en las miosis generadas por el nivel de iluminación entre los dos ojos. No se presentaron diferencias por sexo para ninguna de las variables investigadas. Figura 3 : Grafico de las diez medidas realizadas para cada ojo con indicación de cada anisocoria Las medidas obtenidas de cada paciente se han realizado a 3 niveles de iluminación: 0.02 lux (escotópico); 0.15 lux (mesópico bajo) y 10.6 lux (mesópico alto). En cada toma de datos se han realizado diez imágenes consecutivas en 2 segundos Figura 4:Ejemplo de las medidas de diametros pupilares: Media de las diez valoraciones para cada ojo y para nivel de iluminacion

Upload: vodieu

Post on 10-Jan-2017

236 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: EFECTO DE LA LUMINANCIA EN LA ANISOCORIA Y EN EL

EFECTO DE LA LUMINANCIA EN LA ANISOCORIA Y EN EL

DIAMETRO PUPILAR MESÓPICO Y ESCOTÓPICOSÁNCHEZ- RAMOS C, PUELL MARIN M C, PÉREZ CARRASCO MJ, LANGA MORAGA A.,PELAEZ MARTINEZ T , JIMÉNEZ GARCIA I, DEL MORAL

MARTINEZ I

DEPARTAMENTO DE ÓPTICA. ESCUELA UNIVERSITARIA DE ÓPTICA.

UNIVERSIDAD COMPLUTENSE DE MADRID

80 Congreso Internacional de la Sociedad Española de Oftamología. Córdoba. Sep-Oct 2004. E-mail: [email protected]

BIBLIOGRAFÍA1. Kaufman PL. Adler Fisiología del ojo. Aplicación Clínica. Mosby, Elsevier . 2003

2. Loewenfeld IE Newsome DA: Iris mechanics I. Influence of pupil diameter on dynamics of pupillary movements, Am J Ophthalmol 71:

347, 1971.

3. Loewnstein O Kawabata H, Loewenfield I: the pupil as indicator of retinal activity, Am J Ophthalmol 57:569, 1964.

4. Loewenfeld IE: The reaction to near vision. InThe pupil anatomy physiology and clinical application, vol I. Ames, Iowa and Detroid,

Mich, 1993, Iowa State University Press and Wayne State University Press.

INTRODUCCIÓN

El movimiento de la pupila en respuesta a los cambios de intensidad de la luz optimiza

la iluminación de la retina para maximizar la perceción visual. El diámetro de la pupila

contribuye a mejorar la calidad de la imagen retiniana cuando este, en estado normal, es

pequeño (1).

Clínicamente, la pupila actúa como un indicador subjetivo de la cantidad de luz

trasducida al sistema visua, se utiliza para monitorizar la sensibilidad de la retina a la luz. La

contracción pupilar transitoria en respuesta a un estímulo lumínico puede reflejar la

integridad de la retina y del nervio óptico y sirve para el diagnostico objetivo de patologias

visuales. Cualquier lesión en alguno de los puntos de la vía visual reducirá la amplitud del

movimiento pupilar en respuesta a la luz(1, 2, 3).

Según Loewnfeld(2, 4), la amplitud del movimiento pupilar aumenta de forma

proporcional al logarítmo de la intensidad lumínica del estímulo, mientras que, el tiempo de

latencia del reflejo fotomotor (tiempo desde el inicio del estímulo hasta el inicio de la

contracción pupilar) se acorta (1).

En condiciones de iluminacion mesópica y escotópica, la dilatación de la pupila

maximiza la incidencia de fotones en la retina, lo que además complementa los mecanismos

lentos de adaptación a la oscuridad que implican el control retiniano de los fotorreceptores y

de las células bipolares (1).

Los valores medios de los diámetros pupilares y las desviaciones estándar se

muestran en la Tabla2 .La Gráfica 1 representa los valores del diámetro pupilar

para cada uno de los ojos para los tres niveles de iluminación.

CONCLUSIÓN

Se presentan diferencias significativas del tamaño pupilar entre los

distintos niveles de iluminación mesópico y escotópico estudiados. Las

respuestas de ambos ojos son equilibradas no habiendose encontrado cambios

significativos en la anisocoria.

AGRADECIMIENTO Este trabajo forma parte del proyecto de investigación 5280015 patrocinado

por el INSTITUTO MAPFRE -SEGURIDAD VIAL. “ Caracterización de la función visual de conductores de

vehículos”

OBJETIVO

Determinar los cambios del diámetro pupilar y de la anisocoria en

condiciones mesópicas y escotópicas en una amplia población

MATERIAL Y MÉTODO

Se ha realizado un estudio prospectivo del diámetro pupilar de 64

pacientes (20 3 años) utilizando el pupilómetro binocular de infrarrojos

P2000 (Procyon Instruments Ltd)..

El grado de anisocoria para las tres condiciones de iluminación ha sido:

•3,85% para condiciones escotópicas (0,02 lux)

•4,98% para el nivel mesópico bajo (----lux)

•4,25%para las condiciones mesópicas altas ( 10,6 lux)

No hay diferencias significativas entre las anisocorias para los tres

niveles de iluminación.

En la Figura podemos observar el diámetro pupilar obtenido para cada

uno de los ojos en los diferentes niveles de iluminación

Tabla1 : Diámetros pupilares para las diferentes condiciones de

iluminación

Figura 2 : Pupilómetro binocular de

infrarrojos P2000

(Procyon Instruments Ltd)..

Figura 1 : Imagen de la pantalla

visualizada para cada componente de

la muestra

Niveles de iluminación (lux)

0,02

(x σ)

0,15

(x σ)

10,5

(x σ)

Ojo Derecho 6,72 0,84 5,24 1.07 3,73 0.60

Ojo Izquierdo 6,63±0,86 5,22 0.84 3,84 0.89

Tabla : valores medios y desviaciones estándar de los diámetros pupilares en los

tres niveles de iluminación.

Grafica 1: Diámetro pupilar (mm) para cada uno de los ojos en los tres

niveles de iluminación

Al aumentar la luminancia hasta 0.15 lux (condiciones mesópicas bajas) se produjo

una miosis del 22.02% y del 22.44% para el ojo derecho e izquierdo respectivamente. La

miosis desde el nivel anterior fue del 28.03% y del 26.44% para el ojo derecho y el ojo

izquierdo respectivamente.

Existen diferencias significativas del diámetro pupilar en función de la iluminación

ambiente. No hay diferencias en las miosis generadas por el nivel de iluminación entre los

dos ojos. No se presentaron diferencias por sexo para ninguna de las variables

investigadas.

Figura 3 : Grafico de las diez medidas realizadas para cada ojo con

indicación de cada anisocoria

Las medidas obtenidas de cada paciente se han realizado a 3 niveles de

iluminación: 0.02 lux (escotópico); 0.15 lux (mesópico bajo) y 10.6 lux

(mesópico alto). En cada toma de datos se han realizado diez imágenes

consecutivas en 2 segundos

Figura 4:Ejemplo de las medidas de diametros pupilares: Media de las diez

valoraciones para cada ojo y para nivel de iluminacion