experimentalshodhganga.inflibnet.ac.in/bitstream/10603/7075/7/07...for reasons defined by electronic...
TRANSCRIPT
42
EXPERIMENTAL
This chapter covers the details regarding the materials, their purification and
experimental techniques used for determination of various physicochemical properties
such as apparent molar volumes, viscosities and apparent molar adiabatic compressibilties.
The theoretical background of various physicochemical properties and uncertainties in
various quantities have also been included.
3.1 Materials
The various chemicals used for the present study are listed in Table 3.1 along with their
source, grade and molar mass.
3.1.1. Purification of the Chemicals
The saccharides and methyl glycosides of highest purity grade were used as
received without further purification; however these chemicals were dried over anhydrous
CaCl2 in a vacuum desiccator for 48 h before use. Potassium chloride was used after
drying for 24 h in an oven. Magnesium chloride and lithium chloride, were used as such
and kept stored in a vacuum desiccator. Conductivity water was prepared by distilling
deionised water first over acidic potassium permanganate and then over alkaline potassium
permanganate. It was further degassed before use by boiling for about 10 minutes to avoid
micro air bubbles in the solutions. The specific conductivity of the water used for studies
was less than 1.29 10-6
-1
cm-1
. All the solutions were prepared afresh on weight basis
using a Mettler balance with a precision of 0.01 mg.
3.2 Experimental Techniques
The apparent molar volumes, viscosities and apparent molar adiabatic
compressibilities have been obtained from the measurements of densities, flow time and
speed of sound using densimeter, Ubbelohde type capillary viscometer and ultrasonic
interferometer, respectively.
3.2.1 Vibrating-Tube Digital Densimeter
There are several methods of measuring density such as magnetic float density cell,
differential hydrostatic balance, pycnometer, vibrating-tube digital densimeter, etc.
The vibrating-tube digital densimeter (Model DMA 60/602, Anton Paar, Austria)
was used for density measurements and the schematic diagram for densimeter unit is
shown in Fig. 3.1.
43
Table 3.1: Specifications of chemicals used.
Compound Source Grade/
Mass fraction
Purity
Molar mass
(g mol-1
)
D-(+)-Xylose SRL* AR/(0.999) 150.13
D-(–)-Arabinose SRL AR/(0.999) 150.13
D-(–)-Ribose SRL AR/(0.999) 150.13
D-(+)-Mannose SRL AR/(0.999) 180.16
D-(–)-Fructose SRL AR/(0.994) 180.16
D-(+)-Galactose SRL AR/(0.997) 180.16
D-(+)-Glucose SRL AR/(0.994) 180.16
D-(+)-Melibiose SRL AR/(0.994) 342.30
D-(+)-Cellobiose SRL AR/(0.994) 342.30
D-(+)-Maltose monohydrate SRL AR/(0.994) 360.31
Sucrose Lancaster AR/(0.99) 342.30
D-(+)-Lactose monohydrate SRL AR/(0.999) 360.31
D-(+)-Trehalose dihydrate SRL AR/(0.998) 378.33
D-(+)-Raffinose pentahydrate SRL AR/(0.999) 594.53
(+)-methyl α-D-glucopyranoside S.D. Fine
Chem. Ltd.
AR/(0.989) 194.18
Methyl α-D-xylopyranoside S.D. Fine
Chem. Ltd.
AR/(0.98) 164.20
Methyl β-D-xylopyranoside S.D. Fine
Chem. Ltd.
AR/(0.98) 164. 20
Potassium Chloride(KCl) Qualigens AR/(0.999) 74.55
Magnesium Chloride
(MgCl2.6H2O)
Qualigens AR/(0.999) 203.31
Lithium Chloride
(LiCl)
CDH# AR/(0.999) 42.39
*Sisco Research Laboratories, India.
#Central Drug House Pvt. Ltd., India.
44
Fig
3.1
Sch
emati
c d
iagra
m o
f th
e d
ensi
tym
ete
r (D
MA
60/6
02)
45
(i) Principle
It is based on the principle1-2
of time-lapse measurement for a certain number of
oscillations of vibrating U-shaped sample tube, which is filled with liquid sample. The
oscillation device, which is a hollow glass oscillator, is electronically excited, in an
undamped fashion. The direction of oscillation is perpendicular to a plane passing through
the inlet and outlet opening of the sample tube. The natural frequency of the oscillating
device will be influenced only by the volume of the sample that fills it up to the nodal
points. Therefore, it is necessary to fill the oscillating sample tube at least past its nodal
points, which are fixed and not oscillating.
To derive the mathematical relationship between sample density and natural
frequency or period it is possible to consider an equivalent system consisting of a hollow
body with the mass, M, which is suspended on a spring with a spring constant C. The
volume of hollow body is assumed to be filled by a sample of density, d and natural
frequency, f of such a system is thus given by equation (3.1)
VdM
Cf
2
1 (3.1)
Therefore, time period of oscillation, T is given as:
C
VdMT
2 (3.2)
By squaring and simplifying the equation (3.2):
C
Vd
C
MT
222 44
(3.3)
On substituting CVA /4 2 and CMB /4 2 into equation (3.3), it becomes:
BAdT 2 (3.4)
where A and B are the constants of the oscillator which have the contribution from the
spring constant of oscillator, empty oscillator's mass and volume of the sample which
participates in the oscillations. These constants are therefore instrument's constants for
each individual oscillator and can be determined by time calibration measurements with
sample of known density e.g. dry air and distilled water.
46
Therefore, the following relation can be used for the density difference of the
sample.
)( 2
2
2
121 TTKdd (3.5)
where K=1/A, d1 and d2, T1 and T2 are densities and time periods of sample 1 and 2,
respectively. Hence
)( 2
2
2
121 TTKdd (3.6)
To obtain a precision of + 10-n
in determining densities, it is necessary to measure a time
lapse at least 10n+1
times a chosen unit of time. If the resonance period of the oscillator is of
the order of 2-5 ms selected as the time unit, then a measurement time of 5-6 hours would
be required. Since this is impractical, the total time lapse is measured by independent
means in small but precisely defined units for a preset number of oscillations of the sample
holder. For reasons defined by electronic circulatory, the smallest time unit which can be
measured is 10 s. Thus precision of 10-5
-10-6
g cm-3
in density measurements is thereby
attained by measuring the time lapse in the range of 10 to 100 s.
(ii) DMA 602 Density Measuring Cell
DMA 602 density measuring cell is contained in its own separate housing, complete
with oscillator counter mass and thermostat connectors. It consists of a U-shaped
borosilicate oscillator sample tube fused into a dual wall glass cylinder. The space between
the U-shaped sample tube and the inner wall of the dual-wall cylinder is filled with gas of
high thermal conductivity to facilitate a rapid thermal equilibrium of the sample inside the
oscillator with the thermostat liquid which flows through the dual wall cylinder around the
sample tube. An additional shorter capillary tube inside the inner space of the dual wall
cylinder determines temperature by means of a temperature sensor. This capillary tube of
about 2 mm inside diameter has a very thin wall thickness of about 0.2 mm to assure good
heat transfer. The rest of the instrument consists of the electronic excitation system for the
oscillation and the electronics, which assure an interface-free transmission of the period
signal to the processing unit. Built-in air pump with a sintered glass filter for drying the
sample tube, as well as the sample tube illumination system have also been provided in the
densimeter.
47
(iii) DMA 60 Electronic Processing Unit
Analog signal generated in synchronism with the oscillating sample tube in the
density-measuring cell is transmitted into the electronic processing unit DMA 60. This
signal is amplified and displayed digitally through the light emitting diodes. Details of its
working have been described1-2
elsewhere.
(iv) Temperature Bath
An efficient constant temperature bath (Heto Birkerod/Denmark) with stability
within ± 0.01 K was used to control the temperature of the water circulating around the
densimeter cell. It has both cooling and heating arrangement. The water was made to flow
with the help of inbuilt water circulating pump. The length of the water circulating tubes
was reduced to minimum and these were insulated properly to avoid any heat loss.
Absoluteness of the temperature was checked by putting a Beckmann thermometer, which
was calibrated by determining the transition temperature of freshly crystallized sodium
sulphate with the help of 1°C (1/1000°C) Beckmann thermometer.
(v) Experimental Procedure for Density Measurements
Initially the instrument was switched on for about 30 minutes. The sample tube was
washed several times with deionised water and then with double distilled water. It was
then dried. The liquid whose density was to be determined, was injected continuously into
the lower opening of the U-shaped vibrating tube, until the excess liquid flows out of the
upper part. Liquid should be filled in the tube very slowly to enable the liquid to properly
wet the walls of the sample tube. Care was taken to avoid trapping of micro air bubbles.
Disposable, polyethylene syringe about 2 cm3 in volume was used for introducing the
liquid into the U-tube. After filling the sample tube completely, its openings were closed
with teflon stoppers. Then, the illumination was switched off and start point was pressed
on the DMA 60 processing unit. After observing few read out cycles, a constant value of
the time period, T was noted. Each observation is the mean of triplicate measurements.
Instrument constant, K was calculated after measuring the T values for dry air and
double distilled degassed water in the U-tube, which were taken as standards. The density
of dry air, dT,P (g cm-3
), at temperature T(C) and atmospheric pressure, P (torr) was
calculated3 by using following equation:
dT,P = [0.0012390/1 + 0.00367 T] [P/760] (3.7)
48
Density of the air free pure water was taken from the Kell's data4. The K values
were checked regularly. The densimeter was tested by measuring the densities of the
aqueous solution of sodium chloride at different concentrations at 298.15 K. The results
obtained agreed well with the literature values5-8
within the combined uncertainties of
0.1%.
3.2.3. Multifrequency Ultrasonic Interferometer
The sound velocity, u was determined using Multifrequency Ultrasonic
Interferometer (Model: M-82, Mittal Enterprises, India) which is a direct and simple device
for the measurement of the sound velocities of the liquids with a high degree of accuracy.
The schematic diagram for multifrequency ultrasonic interferometer unit and the measuring
cell are shown in Figs. 3.2-3.3.
(a) Principle
The principle used in the measurement of sound velocity, u is based on the accurate
determination of the wavelength, of sound waves in the medium. Sound waves of known
frequency, f are produced by a quartz crystal fixed at the bottom of the cell. These waves
are reflected by a movable metallic plate kept parallel to the quartz crystal. If the
separation between these two plates is exactly a whole multiple of the sound wavelength,
standing waves are formed in the medium. This acoustic resonance gives rise to an
electrical reaction on the generator which excites the quartz crystal and the anode current
of the generator becomes maximum.
If the distance is now increased or decreased, and the variation is exactly one half
of the wavelength, /2 or multiple of it, anode current becomes maximum. From the
knowledge of wavelength, the sound velocity can be obtained as:
u = f (3.8)
where f is frequency of waves.
(b) Design and Description of Interferometer
The ultrasonic interferometer consists of the following parts:
(i) The high frequency generator.
(ii) The measuring cell.
49
Fig. 3.2 Multifrequency Ultrasonic Interferometer (Model: M-82)
50
MICROMETER
REFLECTOR
EXPERIMENTAL
LIQUID
TO CONST. TEMP
WATER BATH
QUARTZ CRYSTAL
R.F.INPUT
THE MEASURING CELL
Fig. 3.3 The Measuring Cell
51
(i) The High Frequency Generator
It is meant to excite the quartz crystal fixed at the bottom of the measuring cell at its
resonant frequency to generate ultrasonic waves in the experimental liquid filled in the
measuring cell. The panel of the body having high frequency generator is provided with a
microammeter to observe the changes in current and two controls marked 'Adj' and 'Gain'
are also provided, which are used for initial adjustment of the needle of the ammeter and
sensitivity regulation, respectively.
(ii) Measuring Cell
It is a specially designed, double walled cell, which is made of steel with gold plating
at the bottom covering the quartz crystal and chromium plating on inner walls of the cell
for maintaining the temperature of the liquid during the experiment. A fine micrometer
screw has been provided at the top that can lower or raise the reflector plate in the cell
through a known distance.
The quartz crystal, which generates the ultrasonic waves, is fixed at its bottom. The
temperature of the experimental liquid in the cell was maintained at constant temperature
by circulating the water from constant temperature circulator (Model: Julabo F 25) with
an accuracy of 0.01 K.
(c) Working of Ultrasonic Interferometer
Instrument was operated in the following manner to measure the sound velocities:
1. The cell was inserted in the square base socket and clamped to it in the square base
socket with the help of a screw provided on one of its side.
2. Unscrewed the knurled cap of the cell and lifted it away from double walled
construction of the cell. In the middle portion of it, experimental liquid was poured
and cap was screwed again.
3. Two chutes in double walled construction were provided for circulation of water to
maintain the desired temperature.
4. The high frequency generator was connected with cell by co-axial cable provided
with the instrument.
5. Initially temperature bath was switched on for about 30 minutes.
Now switched on the generator and the position of needle on the ammeter was
adjusted with the knob marked as „Adj‟ and knob marked „Gain‟ was used to increase the
52
sensitivity of the instrument for greater deflection, if desired. The ammeter was used to
notice the number of maximum deflections while micrometer was moved up and down in
liquid.
After the equilibration, the micrometer was slowly moved upward till the anode
current was maximum. The initial reading on micrometer scale was noted. Similarly 20
maxima readings on the ammeter were passed and micrometer reading at twentieth
maximum was again noted. The difference between initial and final reading gives the
distance, 'D' travelled by reflector. The average of 10 readings was treated as final value of
the distance travelled by micrometer. The generator was switched off after measurements
were completed and liquid was taken out of the cell. The upper part of cell was removed
and cell was cleaned with distilled water and then rinsed with double distilled water and
filled with the next experimental liquid.
The following relation relate the distance and wavelength of the ultrasonic sound
waves
D = n /2 (3.9)
The sound velocity, u was calculated using equation (3.9). The uncertainties in
sound velocities were ± 0.5 ms-1
, while these were precise within 0.1 ms-1
. The measured
value for u in water at 298.15 K (1496.66 ms-1
) agrees well with the literature9-10
value
(1496.69 ms-1
).
3.2.2. Ubbelohde Type Capillary Viscometer
There are ranges of viscometers available commercially for the measurement of
viscosity. The selection of viscometer depends on a number of criteria i.e. magnitude of
viscosity to be measured, whether the liquid or solution are elastic, whether they are
transparent or opaque and the temperature dependence of viscosity. Some types of
viscometers are: capillary, rotational and moving body,11
etc. In the present work, the
viscosities of the solutions have been measured using Ubbelohde type capillary
viscometer as shown in Fig. 3.4.
(i) Design of Viscometer
The viscometer consists of three arms A, B, C and bulb D of capacity about 60 cm3.
Arm A is fixed to one side of the bulb D. To the upper part of bulb D, another small bulb G
53
Fig. 3.4 Ubbelohde type capillary viscometer
54
is fused to one side of which the arm C is fixed. The lower end of the bulb G is drawn into
a jet just touching the bottom of bulb E. To the upper part of bulb G, arm B of viscometer
is fixed. The arm B contains a fine capillary H, a small bulb I, again a capillary J with the
wide bore, bulb K of capacity of about 10 cm3, again a capillary L of the same bore as J
and another small bulb M, extending into the wide tube T. On the capillaries J and L the
marks X1 and X2 were etched, within which the flow times of the solutions were
determined.
The purpose of bulbs I and M and capillary is to ensure exact noting of the time of
start and finish and thus ensure reproducible times. The absence of the bulb I gives rise to
the bubble formation on the exhaustion of the liquid in the bulb K and thus do not allow
the finish to be noted accurately.
(ii) Calibration of the Viscometer
The viscometer was calibrated using deionised, doubly distilled and degassed water.
The flow time of water between reference marks X1 and X2 was noted with the help of a
stopwatch at different four temperatures. The flow times along with literature values of
densities and viscosities were fitted to the equation (3.10)
η = ρ (at-b/t) (3.10)
where a and b are the constants of the viscometer, ρ and t are the density and flow time of
water, respectively.
(iii) Working of Viscometer
The viscometer was cleaned using freshly prepared chromic acid. The cleaned
viscometer was washed with distilled water, rinsed with acetone and finally dried. Then
the sample solution was transferred into the viscometer and it was immersed into the
thermostat and clamped vertically. After equilibrium for about 45 minutes at the desired
temperature, the arm B of the viscometer was connected to a vacuupet and the solution
from bulb E was filled to arm B when the solution has just crossed the mark X1 on the
capillary L, the vacuupet was released and solution was made to fall from mark X1 to mark
X2. The flow time was recorded with the stopwatch of the resolution of ± 0.01s. The
average of at least four-five readings reproducible within 0.l s was used as final flow time.
The constant temperature water bath (Model: INSREF/India) was used to control the
temperature within 0.01 K. The working of viscometer was checked by measuring the
55
viscosities of aqueous solution of sodium chloride and results were found to agree well
with the literature values12
.
3.3 Theoretical Background of Various Thermodynamic and Transport Properties
The theoretical background of various thermodynamic and transport properties have
been discussed as follows:
3.3.1. Partial Molar Volume
The partial molar volume of a solute, 2V can be defined, as the change in the
volume of solution that results on the addition of one mole of solute to such a large
quantity of the solution that there is no appreciable change in the overall concentration of
the solution when the temperature, T, pressure, P, and number of moles of the other
components if present, remains unchanged.
Mathematically, the partial molar volume of a solute, 2V can be represented as:
2V =
.......n,P,T21
n
V
(3.11)
The total volume of solution, V can be expressed as:
(3.12)
where 1V is the partial molar volume of solvent and n1 and n2 are the number of moles of
solvent and solute, respectively. 2V is generally determined from the apparent molar
volume, V2, which may be defined by the relation:
V2, = 2
011
n
VnV (T and P are constants) (3.13)
or
011,22 VnVnV (3.14)
where 0
1V is the molar volume of pure water (solvent) under identical conditions of T and
P. The relationship between V2, and 2V can be obtained by using the equations (3.12) and
(3.14):
2211 VnVnV
56
2V =
........,,2
,2
2,2
,,211 nPTnPT
n
VnV
n
V
(3.15)
....,,2
,22
2
0
11
11
221
1
1
nPTn
VnVn
nn
VnVV
(3.16)
In terms of the experimentally measured densities, d0 and d of the solvent and
solution, and the molar masses, M1 and M2 of solvent and the solute, the V2, is given by
the equation:
(3.17)
when the concentration is expressed in molality, then n2 = m, the molality of the solution
and n1 is equal to the number of moles of solvent in 1000 g of the solvent, then V2,
becomes:
0
2,2
100010001
dd
mM
mV (3.18)
or
0
02,2
1000)(
mdd
dd
d
MV (3.19)
when the molar concentration is used, then n2 = c, the molarity, then V2, becomes:
0
02,2
)(1000
cd
dd
d
MV
(3.20)
To obtain reliable values of apparent molar volumes, V2, it is necessary to measure
the densities with a great precision. At infinite dilution, the apparent molar volume and the
partial molar volume become equal (V0
2, =V0
2).
The extrapolation of apparent molar volume to infinite dilution has been made by
the following equations:
1. Masson equation13
2. Redlich-Meyer equation14
3. Owen-Brinkley equation15
0
112211
2
,21
Vnd
MnMn
nV
57
Masson13
found that V2, of electrolyte varies linearly with the square root of their
molar concentration, c as follows:
V2, = cSV v0
2 (3.21)
where Sv is the experimental slope.
Redlich and Rosenfeld16
on the basis of Debye-Huckel limiting law predicted a
constant slope for a given ionic charge. On this basis, Redlich and Meyer14
suggested the
following extrapolation function
V2, = cbcSV vv 0
2 (3.22)
where Sv is the theoretical limiting slope and bv is an empirical constant determined from
experimental results.
Owen-Brinkley‟s equation15
can be used in the extrapolation and to represent
concentration dependence of V2, as:
V2, = cKcKaWcKaSV vvv 21)()(0
2 (3.23)
where Kv is an empirical constant and „a‟ is the distance of closest approach. The ideal
method to determine V2, value is to make the measurements for very dilute solutions,
where the deviation from limiting law is very small.
In case of non-electrolytes and zwitterionic solutes, V2, versus molality, m plots
are linear and V0
2 values are obtained by least squares fitting of the following equation:
V2, = V0
2+SvmA (3.24)
For the present study, the equation (3.24) has been used to obtain V0
2 values.
3.3.3. Adiabatic Compressibility
The measure of a change in volume of liquid as a response to a change in pressure,
relative to volume at constant temperature is called isothermal coefficient of
compressibility (KT).
58
Mathematically,
KT = TP
V
V
1 (3.25)
where V is volume of solution, P and T are the pressure and temperature, respectively. The
minus sign indicates that volume decreases with increase of pressure and vice-versa. As KT
is the coefficient of compressibility of the bulk solutions, it is not particularly informative
with regard to either solute-solvent or solute-solute interactions. Therefore, it is better to
determine partial molar quantities to have information on these interactions17-18
. Hence,
partial molar isothermal compressibility of a component in a solution is defined as:
T
i
iTP
VK
, (3.26)
It is generally difficult to calculate precise values of partial molar compressibilities
from a direct determination of the derivative (Vi /P)T. A more convenient method is to
determine the apparent molar compressibility from which the partial molar property can be
derived. The apparent molar isothermal compressibility of a solute in a solution is given by
the equation:
T
T
P
VK
,2, (3.27)
where V2, is the apparent molar volume of solute (equation 3.19).
In order to obtain accurate apparent molar compressibilities, it must be possible to
measure the isothermal compressibilities of solution and solvent with great precision. It is
not an easy task to measure directly isothermal coefficient of compressibilities with the
precision necessary to give reliable apparent molar compressibilities at low pressure.
However, it is possible through speed of sound measurements to obtain very precise values
of a related quantity termed as isentropic coefficient of compressibility
KS = -(1/V)(V/P)S (3.28)
The sound velocity through a fluid, u is related to adiabatic compressibility, KS and
the density, d by the following equation:
59
KS = 1/u2d (3.29)
The relationship between KT and KS is given by equation (3.30):
KT = KS + 2 T/ (3.30)
where α is the coefficient of thermal expansion, = (1/V)/(V/T)P and is the volumetric
heat capacity. The apparent molar isentropic compressibilities, KS,2, can be obtained from
the coefficient of compressibilities and densities of solution and solvent by using the
relation:
0
0
0
2,2,
mdd
dKdK
d
MKK SSS
S
(3.31)
where KS and K0
S are the coefficients of compressibilities of solution and solvent,
respectively and M2, d, d0 and m are the same parameters as used in the previous section.
At infinite dilution, the apparent molar adiabatic compressibility is equal to partial
molar compressibility of the solute (K0
S,2, = K0
S,2).
3.3.2. Viscosity
Viscosity is the measure of the internal friction of a fluid. This friction becomes
apparent when a layer of fluid is made to move in relation to another layer. The greater the
friction, greater the amount of force required to cause this movement, which is called
“shear”. Shearing occurs whenever the fluid is physically moved or distributed, as in
pouring, spreading, mixing, etc.
In case of liquids, viscosity is defined as the ratio of the shearing stress, F' to shear rate,
S as follows:
= F'/ S (3.32)
where is the coefficient of viscosity. The force per unit area required to produce
the shearing action is termed as shear stress i.e. F' = F/A, whereas shear rate can be
described as the measure of change in speed at which the intermediate layers move with
respect to each other i.e.
S=dV/dx (3.33)
60
According to Newton‟s law, the ratio of F'/ S is a constant and is independent of the
shear rate. This law is true only for ideal or “Newtonian” solutions, whereas for non-
Newtonian solutions the ratio of F' / S is not a constant and shear rate is varied.
In case of liquids, decreases with the increase in temperature. The viscosity of
liquid is mainly dependent upon intermolecular forces. In order to flow, the molecules in a
liquid need energy to escape from their neighbours. Evidently this energy is more freely
available at higher temperature than at low temperature.
Relative viscosity, r may be defined by the relation, = /0, where and 0 are
the viscosities of solution and solvent, respectively.
Development of Jones-Dole Equation
Arrhenius19
gave the following relationship between the relative viscosity, r and
concentration, c for moderately dilute solution:
r= Ac (3.34)
where A is a constant for given salt and temperature. Various workes have shown negative
curvature for the salt solutions instead of straight at lower concentrations and at low
temperature. Rabionovich20
have concluded that deploymerization of water molecules must
be responsible for negative viscosity.
Jones and Dole21
from the special behaviour of salts concluded that there must be
some effect, which is of relatively greater importance and is responsible for curvature in
viscosity vs concentration plots. This effect always tends to increase whether the overall
effect of addition of salt is to increase or decrease the viscosity. The decrease in viscosity
was attributed to interionic forces.
Earlier Debye and Huckel observed that the effect of interionic forces in opposing
the motion of ions is proportional to the square root of the concentration in very dilute
solutions. Thus, they gave the equation:
BccA 11
(3.35)
where is the fluidity and A and b are constants, where A has negative values for the
strong electrolytes, which tend to stiffen the solutions or decrease the fluidity and zero for
61
the non-electrolytes. The B has positive values for the liquids with high fluidity and
negative for those with low fluidity.
Later Jones and Talley22
measured the viscosities of urea and sucrose solutions and
further confirmed that the values of A for non-electrolytes like urea and sucrose is zero and
the equation (3.35) reduced to:
= 1 + Bc (3.36)
This equation was extended to represent the data for solutions in the following form:
r = /o = 1 + Bc (3.37)
where B is viscosity B-coefficient, c is the molarity.
3.4 Preparations of Solutions
The solutions were prepared in standard joint flasks (50 cm3). The flasks were
cleaned with freshly prepared chromic acid, followed by washing with distilled water and
then after rinsing with acetone these were dried. The weight of dried flask was noted and
required amount of solute was added into it and then weighed. The required amount of
solvent was then added and again weighed. The molality of the resulting solution was
calculated as follows:
Weight of empty flask = w1g
Weight of empty flask + solute = w2g
Weight of empty flask + solute + solvent = w3g
Weight of solute = (w2-w1) = „a‟g
Weight of solvent = (w3-w2) = „b‟g
Molality, m = 1000 „a‟/ (M „b‟)
where M is molecular weight of solute, a and b are the weights of solute and solvent,
respectively.
3.5 Estimation of Uncertainties
(A) Uncertainty in apparent molar volume, V2,
The uncertainty in V2, may arises from (i) density (ii) molality measurements.
(i) Uncertainty in density measurements
The density has been calculated by using equation (3.6):
62
d1 = d2 + K (T12 – T2
2)
The uncertainty in density „d2‟ determination may arise due to uncertainties in calibration
constant K and T values.
By rearranging the equation (3.6):
(3.38)
where densities, d1, d2 are constants (taken from literature) and time periods T1 and T2 are
experimentally determined quantities for water and dry air, respectively.
Uncertainty in „K‟ due to uncertainty
In
122
2
2
1
12111 .
2.dT
TT
TddUT
(3.39)
in
222
2
2
1
22122 .
2.dT
TT
TddUT
(3.40)
Total uncertainty in K= 2
2
2
1 UU
For dT1 = dT2 = 1 10-6
, the estimated uncertainty in „K‟ comes out to be 4.15 10-5
.
From equation (3.6),
Uncertainty in „d2‟ due to uncertainty
in T1 = U1 = 2 KT1.dT 1 (3.41)
in T2 = U2 = 2 KT2.dT2 (3.42)
in K = U3 = (T12 – T2
2) dK (3.43)
The estimated uncertainty in „d2‟ due to uncertainty in temperature (0.01K), U4 =
0.0000003.
Thus total uncertainty in density measurement becomes
(3.44)
The estimated uncertainty in density measurements comes out to be 2.67 10-6
g cm-3
.
(ii) Uncertainty in molality
Molality, m of a solution has been calculated using the relation:
m = w2 1000/(w1 M2) (3.45)
2
2
2
1
21
TT
ddK
2
4
2
3
2
2
2
1 UUUUUd
63
where w2 is the weight of solute, w1 is the weight of solvent and M2 is the molecular weight
of solute which is constant, so uncertainty in „m‟ may arise from the uncertainty in w2 and
w1, which can be estimated as follows:
Uncertainty in „m‟ due to uncertainty
in w2 = U1 = 12
21000
wM
dw
(3.46)
= 1.07 10-6
mol kg-1
in w1=U2 = 2
12
121000
wM
dww
(3.47)
=1.22 10-7
mol kg-1
Total uncertainty in „m‟ becomes
Um = 2
2
2
1 UU (3.48)
The estimated uncertainty in „m‟ comes out to be 1.08 10-6
mol kg -1
.
Uncertainty in V2, due to uncertainty
in m = U m = dmmdmd
22
0
10001000 (3.49)
= 3.55 10-3
cm3 mol
-1
in d = Ud = ddd
M
dm
2
2
2
1000 (3.50)
= 0.020 cm3 mol
-1
Total uncertainty in V2, becomes
U(V2.) = 22
dm UU (3.51)
The estimated uncertainty in V2, comes out to be 0.020 cm3 mol
-1.
(B) Uncertainties in adiabatic compressibilities and apparent molar adiabatic
compressibilities
The uncertainties in adiabatic compressibility arise due to uncertainty in (i) sound velocity
(ii) density (iii) temperature variation.
Uncertainty in „KS‟ due to uncertainty,
in U = Ux = dUdU
..
23
(3.52)
where dU = 0.5 s
= 0.00160 10-10
kg-1
s2
64
in d = Uy = ddUd
..
122
(3.53)
= 7.31 10-18
kg-2
m4s
2
in T = Uz = dTT
K s . (3.54)
= 0.000090 10-10
kg-1
m s2 K
-1
where dT = 0.01 K.
Total uncertainty in „KS‟ becomes
U(Ks,2, ) = 222
zyx UUU (3.55)
The estimated uncertainty in „KS‟ comes out to be 0.00160 10-10
kg-1
ms-2
.
Uncertainty in apparent molar adiabatic compressibility, KS.2, arises due to the uncertainty,
in KS = U1 = (M2 + 1/m) 1/d dKS (3.56)
in d = U2 = (M2 + 1/m) KS/d2.dd (3.57)
in m = U3 = (K0
S/d0 – Ks/d) 1/m2 dm (3.58)
Total uncertainty in apparent molar adiabatic compressibility
U(Ks,2.) = 2
3
2
2
2
1 UUU (3.59)
Estimated uncertainty in KS.2. comes out to be 0.063 10-15
m3 mol
-1 Pa
-1.
(C) Uncertainty in viscosity
The uncertainty in viscosity, may arise due to uncertainty in (i) constants a and b of the
equation (3.10) i.e. = d
t
bat , (ii) measured flow time and (iii) density.
Uncertainty in „a‟ due to uncertainty
in t = Ua = (/dt2 + 2b/t
3) dt (3.60)
where dt = 0.01
= 2.43 10-7
Uncertainty in „b‟ due to uncertainty
in t = Ub = (2 at - /d) dt (3.61)
= - 0.0257
Uncertainty in „‟ due to uncertainty
in a = U1 = t d.da (3.62)
= 9.86 10-5
65
in b = U2 = d/t.db (3.63)
= -9.78 10-4
in t = U3 = a d.dt + bd/t2 dt
= dt
2t
bdad (3.64)
= 5.128 10-8
in d = U4 = at.dd – b/t .dd
= dd
t
bat
where dd = 2.667 10-6
= 3.4590 10-8
Thus, total uncertainty in viscosity becomes
U = 2
4
2
3
2
2
2
1 UUUU (3.65)
The estimated uncertainty in viscosity comes out to be 0.00014 mPa s.
66
References
1. P. Picker, E. Tremblay, C. Jolicoeur, J. Sol. Chem. 3, 1974, 377.
2. J. P. Elder, Methods of Enzymology, C. H. W. Hirs, S. N. Timasheff, eds., Academic
Press, New York, 61, 1979, 12.
3. CRC Handbook of Chemistry and Physics, R. C. Weast, ed., Chemical Rubber
Company, Cleveland, Ohin, 1978.
4. G. S. Kell, J. Chem. Eng. Data 20, 1975, 97.
5. H. E. Wirth, J. Am. Chem. Soc. 62, 1940, 1128.
6. F. Franks, H. T. Smith, Trans. Faraday Soc. 63, 1967, 2586.
7. F. J. Millero, J. Phys. Chem. 74, 1970, 356.
8. D. G. Archer, J. Phys. Chem. Ref. Data 21, 1992, 793.
9. F. J. Gucker, Jr. F. W. Lamb, G. A. Marsh, R. M. Hagg, J. Am. Chem. Soc. 72, 1949,
310.
10. F. J. Millero, A. L. Surdo, C. Shin, J. Phys. Chem. 82, 1978, 784.
11. H. D. B. Jenkins, Y. Marrcus, Chem. Rev. 95, 1995, 2695.
12. (a) K. B. Belibagli, E. Ayranci, J. Sol. Chem. 19, 1990, 867 (b) Z. Yan. J. Wang, W.
Liu, J. Lu, Thermochim. Acta 334, 1999, 17.
13. D. O. Masson, Phil. Mag. 8, 1929, 218.
14. O. Redlich, D. M. Meyer, Chem. Rev. 64, 1964, 221.
15. B. B. Owen, S. R. Brinkley, Ann. N. Y. Acad. Sci. 51, 1949, 753.
16. O. Redlich, P. Z. Rosenfeld, Electrochem. 37, 1931, 705.
17. A. P. Sarvazyan. Ann. Rev. Biophys. Biophys. Chem. 20, 1991, 321.
18. H. Hoiland, G. R. Hedwig, Biochemical and Physical Properties: Structure and
Physical Data I, Landott-Bornstein Group VIII Biophysics, Springer Verlag,
Heideberg, Germany, 2003.
19. S. Arrhenius, Z. Physik, Chem. 1, 1887, 285.
20. A. I. Rabionovich, J. Am. Chem. Soc. 44, 1922, 954.
21. G. Jones, M. Dole, J. Am. Chem. Soc. 51, 1929, 2950.
22. G. Jones, K. Talley, J. Am. Chem. Soc. 55, 1933, 624.