features benefits - teledyne princeton instruments · 2020-06-24 · nirvana: 640 rev. t2 page 1 of...
TRANSCRIPT
Page 1 of 6NIRvana: 640 Rev. T2Teledyne Princeton Instruments |+1 609-631-4000 | www.princetoninstruments.com | [email protected]
The NIRvana: 640 from Teledyne Princeton Instruments is the world’s first scientific grade, deep-cooled, large format InGaAs camera for low-light scientific SWIR imaging and spectroscopy applications. The camera uses a 640 x 512 InGaAs array with response from 0.9 µm to 1.7 µm. The detector is Peltier cooled to -85°C to minimize thermally generated noise and to improve signal-to-noise ratio for the most demanding SWIR applications. It offers 16-bit digitization and low read noise for outstanding dynamic range.
FEATURES BENEFITS640 x 512 InGaAs array High resolution imaging in the SWIR region;
Offers 4X more pixels over 320 x 256 sensor20 mm x 20 mm pixels High spatial resolutionResponse from 0.9 mm to 1.7 mm with 85% (typical) peak quantum efficiency
Excellent SWIR sensitivity for demanding imaging applications
Thermo-electric (Peltier) cooling Minimizes dark noise and allows extended integration times without the need for liquid nitrogen; Built-in air or liquid cooling option for vibration sensitive environments
Integrated cold-shield Special “cold shield” limits the ambient thermal background Electronic shutter Provides integration times from 2 ms to many minutesC-mount Standard lens interface compatible with numerous lenses and microscopes
(spectrometer adapters available)Fused silica window AR coated, fused silica window for high transmission in SWIR rangeGigE interface Industry standard for fast data transfer over long distances, up to 50 mOptional: LightField® (for Windows 8/7, 64-bit) Or WinView/Spec (for Windows 8/7/XP, 32-bit)
Flexible software packages for data acquisition, display and analysis; LightField offers intuitive, cutting edge user interface, IntelliCal® and more
SITK® for LabVIEW™ Easy integration into complex experimental setupPICAM (64-bit)/PVCAM (32-bit) software development kits (SDKs)
Compatible with Windows 8/7/XP, and Linux;Universal programming interfaces for easy custom programming.
NIRvana: 640
Applications: Nanotube fluorescence, emission, absorption, non-destructive testing and singlet oxygen detection
NOTE: Export of this camera outside of the United States is prohibited by law unless accompanied by a valid Export License as issued by the United States Department of Commerce.
NIRvana: 640 shown with lens, sold separately.
Page 2 of 6NIRvana: 640 Rev. T2Teledyne Princeton Instruments |+1 609-631-4000 | www.princetoninstruments.com | [email protected]
SPECIFICATIONS
NIRvana: 640
CCD image sensor 2D InGaAs focal plane array
CCD format 640 x 512 imaging pixels
CCD image area 12.8 mm x 10.24 mm
Pixel size 20 µm x 20 µm
Pixel well capacity (e-)
High Gain 45,000
High Capacity600,000
Nominal gain (e-/count) 1.2 14
Typical system read noise (e-) 80* (High Gain)
Response nonlinearity: High gain < 2** %
Cooling temp. @ +20°C ambient -80°C (guaranteed), -85°C (typical with Air +15°C liquid circulation)
Cooling method Air only, liquid only, or a combination of air and liquid
Dark current (e-/p/sec) @ -80°C 75***
Blemish specification Grade A: <2% defects For detailed blemish specifications, contact Princeton Instruments
Digitization 16 bits
Scan rate 2 MHz, 5 MHz, 10 MHz
Frame rate 22 fps @ 2 MHz55 fps @ 5 MHz110 fps @ 10 MHz
Binning and ROI Software only
Cold shield f#/1.5
Exposure time < 2 ms to >10 min
Window material Fused Silica (AR coated)
Thermostating precision ±0.05°C across entire temperature range
Operating temperature 0°C to +30°C
Certification CE
Weight / Dimensions (C-mount) 9.5 lbs (4.3 kg) / 7.40” x 5.8” x 5.8” (187.96 mm x 147.32 mm x 147.32 mm) (L x W x H)
Weight / Dimensions (SP-mount) 9.5 lbs (4.3 kg) / 7.297” x 5.8” x 5.8” (185.34 mm x 147.32 mm x 147.32 mm) (L x W x H)
All specifications are typical. Subject to change without notice.
* Measured with 1 µs exposure time** Linearity for exposure > 20 ms *** Measured with a cold target at -174°C (+99°K)
Page 3 of 6NIRvana: 640 Rev. T2Teledyne Princeton Instruments |+1 609-631-4000 | www.princetoninstruments.com | [email protected]
SENSOR QUANTUM EFFICIENCY CURVE
0
10
20
30
40
50
60
70
80
90
100
0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80
Qua
ntum
Effi
cien
cy (%
)
Wavelength (µm)
NOTE: GraphshowstypicalQuantumEfficiency(QE)datameasuredat+25°C.Longercut-offwavelengthshiftstowardsblueby8nmperevery10°Cofcooling.
Page 4 of 6NIRvana: 640 Rev. T2Teledyne Princeton Instruments |+1 609-631-4000 | www.princetoninstruments.com | [email protected]
APPLICATION IMAGES
SWIR Angiography
Single walled nanotube (SWNT)
Topographical rendering of energy-momentum-resolved spectroscopy
Images courtesy of Dr. R. Bruce Weisman, Professor of Chemistry, The Richard E. Smalley Institute for Nanoscale Science and Technology, Rice University, Houston, TX USA
Image courtesy of Dr. Rashid Zia, Manning Assistant Professor, School of Engineering, Brown University, Providence, RI USA
Image courtesy: Bawendi Lab, Massachusetts Institute of Technology, Cambridge, MA. Fromhigh-speedintravitalimagingforangiographyinabraintumormodeltohighresolutionandhigh-speedSWIRintravitalimagingtogenerateflowmapsofmicrovascularnetworksusing QD composite particles.
Page5of6NIRvana: 640 Rev. T2Teledyne Princeton Instruments |+1 609-631-4000 | www.princetoninstruments.com | [email protected]
OUTLINE DRAWING WITH C MOUNT
Spectrometer adapter is available upon request.
0
0
0.75019.05
2.35059.69
2.90073.66
0.75019.05
2.35059.69
2.90073.66
2.
350
59.6
9
2.
900
73.6
6
3.
906
99.2
0
2.
350
59.6
9
2.
900
73.6
6
00.
351
8.91
WIN
DO
W F
RO
NT
SU
RFA
CE
0.47
612
.08
WIN
DO
W IN
VA
CU
UM
SU
RFA
CE
0.69
FP
A L
OC
ATI
ON
1.
150
29.2
1
3.13
79.5
3
7.40
187.
84
7.62
193.
66
0
0
0.62
515
.88
0.62
015
.74
1.72
543
.82
1.75
544
.58
0.50512.83
1.25531.88
1.79645.62
0.3168.03
NOTES:
WEIGHT: 9.5 LBS / 4.3 KGS2.
WINDOW THICKNESS = 0.125" [3.18MM]3.
MECHANICAL SHUTTER NOT AVAILABLE4.
MAX. POWER CONSUMPTION = 135W5.
C-MOUNT THREAD1.00-20 UN
4.063 20 UN THREAD
LIQUID COOLANTINLET/OUTLET
ALL DIMENSIONS IN INCHES [MM]1.
Page 6 of 6NIRvana: 640 Rev. T2Teledyne Princeton Instruments |+1 609-631-4000 | www.princetoninstruments.com | [email protected]
OUTLINE DRAWING WITH SPECTROMETER MOUNT
35°
17.5° 3.600
91.45 B.C.
0
0
0.75019.05
2.90073.66
0.75019.05
2.90073.66
2.
900
73.6
6
3.
906
99.2
0
2.
900
73.6
6 0
0
0.25
26.
41FR
ON
T S
UR
FAC
EO
F V
AC
UU
M W
IND
OW
0.59
014
.99
FPA
DIS
TAN
CE
1.
050
26.6
7
3.
032
77.0
2
7.
297
185.
33
7.
526
191.
16
0.0
63
1.12528.58
1.12528.58
0
0
0.50512.83
1.25531.88
1.79645.62
0.3168.03
0.
625
15.8
8
0.
620
15.7
4
1.
725
43.8
2
1.
755
44.5
8
NOTES:
WEIGHT: 9.5 LBS / 4.3 KGS2.
WINDOW THICKNESS = 0.125" [3.18MM]3.
MECHANICAL SHUTTER NOT AVAILABLE4.
MAX. POWER CONSUMPTION = 135W5.
4.200106.68
COOLANT INLET/OUTLET
2X 10-32 UNF TYP. ALL FOUR SIDES
ALL DIMENSIONS IN INCHES (MM)1.
C-mount adapter is available upon request.