fresenius medical care peritoneal dialysis connections baxter to stay safe connection
TRANSCRIPT
Fresenius Medical Care Peritoneal Dialysis Fresenius Medical Care Peritoneal Dialysis ConnectionsConnections
Baxter to staysafe Connection
Connections using Fresenius Medical Connections using Fresenius Medical Care productsCare products
This inservice is not intended to replace the advice or training from your physician. Refer to the Instructions For Use (IFU) for detailed device description, instructions, contraindications, warnings, and precautions.
Fresenius Medical Care, DEL-CLAMP, Stay●Safe, Safe-Lock, DELFLEX and FREEDOM are trademarks of Fresenius Medical Care Holdings Inc. or its affiliated companies. All other trademarks are the property of their respective owners.
Fresenius ProductsFresenius Products
Baxter is a trademark of Baxter International Inc. or its affiliates.
Instructions on how to use Baxter products were reproduced from the manufacturer’s literature and were current at the time of print. Instructions are included for the patient's convenience but Fresenius Medical Care North America does not specifically endorse the use of these products or assume any liability related to their use. Should there be any differences between the instructions listed in this manual and the instructions for use provided with the medical devices referenced, the instructions provided with the medical device should be taken as accurate."
Connections using Fresenius Medical Connections using Fresenius Medical Care productsCare products
Note: for clarity the model is shown not wearing personal protective equipment (PPE). Follow your facility’s policies and procedures for use of personal protective equipment.
It is vital to maintain sterility of the connectors during connection procedures.
Baxter patient to Baxter patient to staystaysafesafe system system
Baxter PD patient staysafe System
staysafe Extension Set with Luer-Lock patient connector
A staysafe Extension Set with Luer-Lock patient connector will allow the patient with a Baxter PD system to use staysafe supplies while in the hospital.
Baxter patient to Baxter patient to staystaysafesafe system system
Gather supplies
staysafe solution set and related supplies for transfer set connection and exchange procedure.
staysafe Extension Set with Luer-Lock patient connector 12 inch (P/N # 050-95004)
Or18 inch (P/N # 050-95005)
Baxter patient to Baxter patient to staystaysafesafe system system1. Remove the blue sterility protector from
the end of the Fresenius staysafe Extension Set .
2. Remove the mini-cap from the Baxter patient extension set.
3. Attach the Fresenius staysafe Extension Set to the end of the Baxter extension set. Proceed with the staysafe CAPD or APD exchange.
1ent extension set
Baxter Extension Set
stay safe Extension Set
3
2
Titanium luer lock adapter to Titanium luer lock adapter to staystaysafesafe systemsystem
staysafe solution set and related supplies for transfer set connection and exchange procedure.
staysafe Extension Set with Luer-Lock patient connector 12 inch (P/N # 050-95004)
Or18 inch (P/N # 050-95005)
Remove the blue sterility protector from the end of the Fresenius stay● safe extension set and attach to the luer lock catheter adapter.
DISCHARGE INSTRUCTIONSDISCHARGE INSTRUCTIONSFOR ALL CONNECTIONSFOR ALL CONNECTIONS
• The patient should return to his/her home or clinic with the additional extension in place. This practice will allow the patient to return home and resume dialysis without interruption.
• Instruct the patient to remove any adaptors added to their extension set at the beginning of the next exchange or as directed by the PD clinic nurse.
Select the catheter or transfer set used by the patient.
Select the System
the Hospital
Uses
Catheteror
Transfer Set or or
With Luer-Lock Adapter
Titanium
With Safe-Lock Adapter
With Baxter Transfer Set
With Fresenius Safe-Lock
transfer Set
With Fresenius Stay-Safe
Transfer Set
Hospital Uses Baxter Supplies
Add Baxter Transfer
Set5C4484
Add FMC Universal Connector 050-30026 or pre-filled050-30031
No Change Needed
Add FMC Convertible Adapter 050-40393 and Baxter 5C4484
AddStay●Safe Luer Lock Adapter 4”
050-95003
Hospital Uses Fresenius (FMC) Safe-Lock Supplies
Add Safe-Lock Transfer Set with Luer Lock Adapter
050-30034
No Change Needed OR
Add a FMC Safe-Lock Transfer Set with a Safe-Lock
Adapter
Add Convertible Extension050-40026 No Change
Needed
Add4” Stay●Safe
Safe-Lock Adapter050-95006
Hospital Uses FMC Stay-Safe Supplies
AddStay● Safe Extension Set with Luer lock Patient Connector050-95004
AddStay● Safe Extension Set with Safe lock Patient Connector050-95001
Add Stay● Safe Luer Lock Adapter 050-95003 and a Baxter Transfer Set 5C4484
Add Stay● Safe Extension set with Safe-Lock Patient connector 050-95001
No Change Needed
Choose the type of catheter or transfer set and the type of system utilized. Look down the column and across the row to the adapter needed. Click on
the icon in the box to take you to the procedure
Please contact your Fresenius Clinical Sales Representative at 1-800-662-1237 if you have questions regarding the use of a Fresenius PD system or adapter.
CD Available through Customer Service
800-323-5188 (P/N # 100676-01)