ftir & uv vis theory
TRANSCRIPT
-
8/3/2019 Ftir & Uv Vis THEORY
1/12
THEORY
Fourier Transform Infrared (FTIR)
Fourier Transform Infrared (FTIR) spectrometry was developed in order to overcome the
limitations encountered with dispersive instruments. The main difficulty was the slow scanning
process. A method for measuring all of the infrared frequencies simultaneously, rather than
individually, was needed. A solution was developed which employed a very simple optical
device called an interferometer. The interferometer produces a unique type of signal which has
all of the infrared frequencies fixed into it. The signal can be measured very quickly, usually on
the order of one second or so. Thus, the time element per sample is reduced to a matter of a few
seconds rather than several minutes. Most interferometers use a beam splitter which takes the
incoming infrared beam and divides it into two optical beams. One beam reflects off of a flat
mirror which is fixed in place. The other beam reflects off of a flat mirror which is on a
mechanism which allows this mirror to move a very short distance (typically a few millimeters)
away from the beam-splitter. The two beams reflect off of their respective mirrors and are
recombined when they meet back at the beam-splitter.
Because the path that one beam travels is a fixed length and the other is constantly
changing as its mirror moves, the signal which exits the interferometer is the result of these two
beams interfering with each other. The resulting signal is called an interferogram which has
the unique property that every data point (a function of the moving mirror position) which makes
up the signal has information about every infrared frequency which comes from the source. This
means that as the interferogram is measured; all frequencies are being measured simultaneously.
Thus, the use of the interferometer results in extremely fast measurements. Because the analyst
requires a frequency spectrum (a plot of the intensity at each individual frequency) in order to
make identification, the measured interferogram signal cannot be interpreted directly. A means
of decoding the individual frequencies is required. This can be accomplished via a well -known
mathematical technique called the Fourier transformation. This transformation is performed by
the computer which then presents the user with the desired spectral information for analysis.
Infrared spectrophotometer is a powerful tool for identifying pure organic and inorganic
compounds because, with the exception of a few same mononuclear molecules such as O 2, N2
and Cl2, all molecular species absorb infrared radiation. Each molecular species has a unique
-
8/3/2019 Ftir & Uv Vis THEORY
2/12
infrared absorption spectrum. Therefore, exact match between the spectrum and of a compound
of known structure and the spectrum of analyte unambiguously identifies the analyte. Infrared
spans a section of the electromagnetic spectrum by wavenumbers, , in the range of 12,800 to 10
cm-1
and the wavelength, , ranging from 0.78 - 1000m. Both wavenumber and wavelength
represent IR absorption. Wavenumber unit is more commonly used in modern IR instruments
that are linear in the cm-1
scale. In the contrast, wavelengths are inversely proportional to
frequencies. Wavenumber and wavelength can be interconvert using following equation:
( )
( ) (eq. 1)
Transmittance, T, is the ratio of radiant power transmitted by the sample (I) to the radiant
power incident on the sample (I0). Absorbance, A, is the logarithm to the base 10 of the
reciprocal of the transmittance.
( ) (eq. 2)
The transmittance spectra provide better contrast between intensities of strong and weak
bands because transmittance ranges from 0 to 100% T whereas absorbance ranges from infinity
to zero. The 0% T and 100% T adjustments should be made immediately before each
transmittance or absorbance measurement. To obtain reproducible transmittance measurements,
it is essential that the radiant power of the source remain constant while the 100% T adjustment
is made and the % T is read from the meter.
Like as stated above before, the infrared region of the spectrum encompasses radiation
with wave numbers ranging from about 12,800 to 10 cm -1 or wavelengths from 0.78 to 1000 m.
From the perspective of both application and instrumentation, the infrared spectrum is
conveniently divided into near-, mid- and far- infrared radiation; rough limits of each are shown
in table below.
-
8/3/2019 Ftir & Uv Vis THEORY
3/12
Table 1: Infrared Spectral Regions
RegionWavelength ()
Range, m
Wavenumber ()
Range, cm-1
Frequency () Range,
Hz
Near 0.78 to 2.5 12,800 to 40003.8 x 10
14to 1.2 x
1014
Mid 2.5 to 50 4000 to 2001.2 x 1014 to 6.0 x
1012
Far 50 to 100 200 to 106.0 x 10
12to 3.0 x
1011
Most used 2.5 to 15 4000 to 6701.2 x 10
14to 2.0 x
1013
Figure 1: Instrument Diagram of Basic FTIR
Adapted from http://binoybnair.blogspot.com/2007_10_01_archive.html
-
8/3/2019 Ftir & Uv Vis THEORY
4/12
Figure above shows the instrument diagram for a basic FTIR spectrometer. Radiation of
all frequencies from the IR source is reflected into the interferometer where it is modulated by
the moving mirror on the left. The modulated radiation is then reflected from the two mirrors on
the right through the sample in the compartment at the bottom. After passing through the sample,
the radiation falls on the detector. A data acquisition system attached to the detector records the
signal and stores it in the memory of a computer as an interferogram. (Courtesy of Thermo
Electron Corp., Madison, WI).
Figure 2: IR Spectroscopy Apparatus
Adapted from
http://chemwiki.ucdavis.edu/Wikitexts/UCD_Chem_205:_Larsen/ChemWiki_Module_Topics/H
ow_an_FTIR_Spectrometer_Operates
In most UV-VIS, the cell is located between the monochromator and the detectors in
order to avoid photodecomposition of the sample, which may occur of samples, are exposed to
the full power of the source. Photodiode array instruments avoid this problem because of the
short exposure time of the sample to the beam. In IR radiation, in contrast, is not sufficiently
energetic to bring about photodecomposition. Most samples are good emitters of IR radiation.
Because of this, the cell compartment is usually located between the source and the
monochromator in IR instrument.
http://upload.wikimedia.org/wikipedia/en/f/fc/IR_spectroscopy_apparatus.jpeg -
8/3/2019 Ftir & Uv Vis THEORY
5/12
As stated before, the components of IR instruments differ significantly from UV-VIS
instruments. Thus, IR sources are heated solids and IR detectors respond to heat rather than
photons. FTIR which is widely used nowadays, detect all the wavelengths all the time. They
have greater light-gathering power than dispersive instruments and consequently better precision.
It contains no dispersing element and all wavelengths are detected and measured simultaneously.
Instead of monochromator, an interferometer is used to produce interference patterns that contain
the infrared spectral information. To obtain radiant power as a function of wavelength, the
interferometer modulates the source signal in such a way that it can be decoded by the
mathematical technique of Fourier transformation.
The energy of infrared radiation can excite vibrational and rotational transitions, but it is
insufficient to excite electronic transitions. Variations in rotational levels may give rise to a
series of peaks for each vibrational state. The number of molecule can vibrate is related to the
number of atoms and the number of bonds, it contains. Infrared absorption occurs not only with
organic molecules but also with covalently bonded metal complexes, which are generally active
in the longer wavelength infrared region.
The major type of molecular vibrations are stretching and bending. Infrared radiation is absorbed
and the associated energy is converted into these types of motion. The absorption involves
discrete, quantized and energy levels. However, the individual vibrational motion is usually
accompanied by other rotational motions. Figure below shows major vibrational modes.
Vibrations as stated above fall into stretching and bending. Stretching involves
continuous change in the interatomic distance along the axis of the bond between two atoms.
While bending are characterized by a change in the angle between two bonds and are of four
types; scissoring, rocking, wagging and twisting. Infrared radiation is not energetic enough to
bring about the kinds of electronic transitions that have encountered in UV-VIS. Absorption of
infrared radiation is thus confined largely to molecular that have small energy differences
between various vibrational and rotational states. In order to absorb infrared radiation, a
molecule must undergo a net change in dipole moment due to vibrational and rotational motion.
However, no net change in dipole moment occurs during the vibration or rotation of
homonuclear species such as O2, N2 and Cl2. Consequently, such compounds cannot absorb in
infrared. Absorption causes increase in vibration amplitude/rotation frequency.
-
8/3/2019 Ftir & Uv Vis THEORY
6/12
Figure 3: Types of Molecular Vibrations of CH2
As noted that the approximate frequency (or wave number) at which organic functional
group, such as absorbs infrared radiation can be calculated
from the masses of the atoms and the force constant of the bond between them. These
frequencies, called group frequencies, are seldom totally invariant because of interactions with
other vibrations associated with one or both of the atoms composing the group. Table below lists
group frequencies for several common functional groups.
-
8/3/2019 Ftir & Uv Vis THEORY
7/12
Functional Group NamesAbsorption Ranges
Frequency (cm--1)Type of Vibration
Alkanes
3000-2800
1500-1450
Alkenes
3100-3000
1675-1600
Alkynes3300-3200
2200-2100
Aromatic Rings
3100-3000
1600-1580
1500-1450
Alcohols, Phenols3600-3100
(Note: Phenols MUST have
Aromatic Ring Absorptions
too)
1300-1000
Nitriles
2300-2200
-
8/3/2019 Ftir & Uv Vis THEORY
8/12
Ketones
1750-1625
Aldehydes 1750-1625
2850-2800
2750-2700
Carboxylic Acids
3400-2400
(Note: This peak always
covers the entire region
with a VERY BROAD
peak.)
1730-1660
Ethers
1300-1000
Esters1735
1300-1000
Amines: Primary3500-3200
(TWO PEAKS!)
1640-1560
Amines: Secondary3500-3200
(ONE PEAK!)
1550-1450
Nitro Groups1600-1500
1400-1300
-
8/3/2019 Ftir & Uv Vis THEORY
9/12
Amides3500-3100
1670-1600
1640-1550
UV- VIS
The principle of UV described that molecules have ability to absorb ultraviolet or visible light.
UV-vis spectroscopy is useful as an analytical technique for two reasons which are to identify
some functional groups in molecules and determining the content and strength of a substance.
Spectrophotometers are made up of stable source of radiant energy, a transparent sample
container, a device for isolating specific wavelength, a radiation detector which converts
transmitted radiation to a usable signal and a signal processor and readout. A wave is usually
described in terms of its wavelength, , and frequency, , which is the number of peaks passing a
given point per second. Table 1 showed the different regions of wavelength. All the
electromagnetic radiation travels through a vacuum at the same velocity. This velocity (c) is
called the velocity of light which is 2.99792458 X 108 ms-2. It can be connected to wavelength
and frequency as c= .
Region Wavelength (nm)
Far ultraviolet 10 - 200
Near ultraviolet 200 - 380
Visible 780 - 3000
Near infrared 3000 - 30,000
Middle infrared 30,000 - 300,000
Microwave 300,000 - 1,000,000,000
Table 2: The regions of wavelength.
UV-vis spectrum, primarily provide structural information about the kind and extent of the
conjugation of multiple bonds in the compound being analyzed. The absorption of UV-vis
-
8/3/2019 Ftir & Uv Vis THEORY
10/12
spectroscopy is caused by transfer of energy from the radiation beam to electrons in orbitals of
lower energy, causing them to be excited into orbitals of higher energy. The excitation process
proceeds by promoting an electron from an occupied to an unoccupied molecular orbital. Every
molecule has their own ground electronic states with all electron spins in paired. Most excited
electronic states also have all paired electron spins. Excited electronic states may be have two
orbital that each of it posses only one electron. The energy of quantum of electromagnetic energyis directly related to its frequency:
E = h (eq. 3)
Where, h= Planks constant = 6.63 x 10-34
Js
=frequency (Hz)
This means, the higher energy the frequency of radiation, the greater is its energy. Since = c/,
the energy of electromagnetic radiation is inversely proportional to its wavelength.
(eq. 4)
Most spectra ofthe solution obey Beers Law. This states that the light absorbed is proportional
to the number of absorbing. The second law is Lamberts law. It tells us that the fraction of
radiation absorbed is independent of the intensity of the radiation. Combining these two laws
gives the Beer-Lambert law:
Log10I0/I= lc (eq. 5)
Where =
I0
= the intensity of the incident radiationI= the intensity of the transmitted radiation
= the molar absorption coefficient
l = the path length of the absorbing solution (cm)
c= the concentration of the absorbing species in mol dm-3
-
8/3/2019 Ftir & Uv Vis THEORY
11/12
(Log10Io/Ialso is called absorbance of optical density or A)
The absorbing groups in a molecule are called chromophores. Chromophore (literally color-
bearing) group is a functional group, not conjugated with another group, which exhibits a
characteristic absorption spectrum in the ultraviolet or visible region. Some of the more
important chromophoric groups are list in the table 2.
Name Chromophore Wavelength, (nm) Molar extinction, e
Acetylide -C C 175-180 6000
Aldehyde -CHO 210 1500
Alkynyl -C C- 160-170 8000
Azo -N=N- 285-400 3-25
Carboxyl -COOH 200-210 50-70
Ester -COOR 205 50
Ether -O- 185 1000
Ketone -C O 195 1000
Nitrile -C N 160 -
Table 3 : UV absorption maxima.
There are few types of UV-Vis those are famously and widely used namely single-beam
instrument, double-beam instrument and multichannel instruments. Figures below shows the
block diagram of each instrument:
Figure 4 : Single beam
-
8/3/2019 Ftir & Uv Vis THEORY
12/12
Figure 2 : Double beam (in space)
Figure 5 : Multichannel
For single beam instrument, 0% transmittance is set with shutter in the beam path while 100%
transmittance is set with a reference in the beam path. Measurement is made with the sample in
the beam path. While for double beam instrument, sample and reference are measured
simultaneously and the signal from reference is substracted from sample signal. Double beam
instrument is more widely used because: i) its compensate for variations in the source intensity,
ii) compensate for drift in detector and amplifier, iii) compensate for variation in intensity as a
function of wavelength.