InfraSpecNR800 Fourier TransformNear-Infrared Analyzers
Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 JapanTel.: 81-422-52-5617 Fax.: 81-422-52-6792
GS 12Y03A03-01E
GS 12Y03A03-01E©Copyright Aug. 2001
5th Edition Mar. 18, 2014
n OverviewThe NR800 Fourier Transform Near-Infrared Analyzers offers high-resolution, high S/N (signal-to-noise) ratio, and wide wavelength scanning range measurement with its newly developed interferometer and detector. The NR800 also offers high stability, vibration resistance, and durability, inheriting features from NR500 earlier successful models. The NR800 allows online, real-time, continuous, multiple, and simultaneous measurement for properties and component concentration of various processes.
nFeatures(includingoptions) Newlydevelopedinterferometeranddetector
Usedforavarietyofapplications,rangingfromovertocombinationtone.• Highresolution:Upto4cm–1, user selectable
setting• HighS/Nratio:2250:1(RMS,4cm-1 resolution,
4100 to 4200 cm-1 , 1 sec.)• Widewavelengthsscanningrange900to2500nm(11,000to4000cm-1)
• Wavelengthreproducibility:0.007cm-1
• Wavelengthaccuracy:0.04cm-1
Enhancedenvironmentalresistance,durability,andreliability• Provideshighvibrationresistancebyaunique
design free of sliding parts• Featuresamulti-channelmeasurementopticalsystemfreeofmovingparts
• EliminatestheneedforaPCforcontinuousoperation.
APCisnowonlynecessaryforgeneratingthecalibration model and loading data.
OutstandingwavelengthaccuracyallowscalibrationmodeltransportbetweenNR800s
Outlierdetectionandself-diagnosticfeaturescomeasstandardfunction
Variousstandardfeaturesandavailableoptionsforoptimalsystemconfiguration• Dust-proofanddrip-proof:IP53equivalentorNEMATYPE3R
• OptionalExplosion-proofEnclosure:FM,CSA,ATEX,TIIS(*1),KOSHA
• LineupwithCEmarking• Non-movingmulti-channelmeasurement:
Expandable to 4 channels• Upto12itemscanbemeasuredperstream:limitedto64items/unit
• TheoptionalI/Ounitoffersavarietyofinputs/outputs:
Analogueoutput(upto40points),analogueinput,andcontactinput/output
• Communicationoutput:RS-422• FastEthernetcommunicationbetweenEngineeringPCandtheanalyzer
(*1) TIIS:TheTechnologyInstitutionofIndustrialSafety
ExplosionproofModel
2
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E Mar. 18, 2014-00
1.Configuration1.1 Configurationexamples
• Modbusoutputwithsamplingunit(explosion-proof,1measuringchannel)
RS-422/RS-232Cconverter *2
RS-232C(MODBUS)
(15m max.)(1km max.)
Sampleejection
Process
Fiber-optic cable *2
Clean, dry air
Hazardous area Non-hazardous area
Alarm contact output
Optical/electricalsignal converter
Monitor/ControlSystem
Engineering PC
Samplingunit
Ethernet(Fiber-optic cable *1)
RS-422(MODBUS)
Ethernet(10 Base-T)
InfraSpecNR800
explosion-proofmodel
Measurement
cell
• Analogueoutputwithsamplingunit(explosion-proof,2measuringchannels)
I/O unit*2
RS-422Process 2
Process 1
Sampleejection
Sampleejection
Clean, dry air
Fiber-optic cable *2
Hazardous area Non-hazardous area
Optical/electricalsignal
Alarm contact output
Engineering PC
Ethernet(10 Base-T)
Analog signal
Contact signal
Contact signal
Monitor/ControlSystem
Sampling cell 1
Sampling cell 2
Ethernet(Fiber-optic cable *1)
InfraSpecNR800
explosion-proofmodel
Measurementcell 1
Measurementcell 2 (1km max.)
*1: Alwaysuseaspecialfiber-opticcable(seesection2.6)foranexplosion-proofmodel,anelectriccablecannotbeused. For a general purpose model, the type of cable depends on the length. •Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Anelectriccablecannotbeused.*2: OnlyusetheRS-422outputofanexplosion-proofmodelincombinationwithanI/Ounit(seesection2.2,selectSuffixcode
toExplosion-proofmodel)orRS-422/RS-232Cconverter(seesection2.5).TheseunitswillblockthecommunicationsignaluponreceivingaPurgefailuresignalfromtheAnalyzer.
Andusethededicatedfiber-opticcableformeasurement(seesection2.4)andEthernetcommunication(seesection2.6). Thus, ensuring the explosion-proof integrity of the analyzer.
3<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
1.2 Component,softwareandcalibrationmodelgeneration
Item Require-ment(*1) Model Description Reference
1 InfraSpec NR800 analyzer A
NR801AG General purpose model, CSA standard
1.32.1
NR805AG Explosion-proofmodel,FMcertified(excludingNR805AG/CSA),CSAcertified(onlyforNR805AG/CSA)
NR801EG General purpose model with CE markingNR805EG Explosion-proofmodel,ATEXcertifiedandCEmarkingNR801JG Generalpurposemodel,Japaneseversion
NR805JG Explosion-proofmodel,TIIScertified(excludingNR805JG/K),KOSHAcertified(onlyforNR805JG/K)
2 I/Ounit BNR893AG To be used with NR801,805AG, CSA standard
1.32.1NR893EG To be used with NR801,805EG, and CE marking
NR893JG To be used with NR801,805JG
3 Measurement cell A
NR510 Flow through cell 2.3.1NR512 Flow through cell with constant temperature water tube 2.3.2NR511 Special measurement cell 2.3.3FIR200 In-situ probe 2.3.4
4 Fiber-optic cable for measurement A
NR821 Applicablewavelengthrange:900to2100nm,Silica,single2.4.1
NR822 Applicablewavelengthrange:900to2100nm,Silica,dualNR823 Applicablewavelengthrange:900to2500nm,Fluoride,single
2.4.2NR824 Applicablewavelengthrange:900to2500nm,Fluoride,dual
5 RS-422/RS-232Cconverter B
K9404LAor
K9404LDConvertstheRS-422signalfromtheanalyzerintoRS-232C. 2.5
6 Ethernetcable(*2) BNR895 Dedicatedfiber-opticcable(*3) 2.6
- Electriccable,providedbyuser 6.2
7 SoftwareA NR831 SPECTLAND2managementandmaintenancesoftware 2.7B NR530 Chemometrics software 2.8
8 Sampling unit B J439 Yokogawa will propose an optimum unit based on sample pressure, temperature, properties, and measurement items 2.9
9On-siteguidanceof calibration model generation
B J964 Hands-onpracticeandguidanceformodelgenerationonsite. 5.1
10 Calibration model generation B J965 Calibration model generation by Yokogawa based on user-
providedsamplewithlaboratoryanalysisresults 5.2
11 EngineeringPC A - Providedbyuser.Seerecommendedspecifications. 6.1
12Optical/electricalsignalconverterforEthernet
B -ConvertsopticalsignalsforanEthernetoutputintoelectricalsignalsforengineeringPCInterface.Providedbyuser.Seerecommendedspecifications
6.2
13 Customer inspection
B J963 For Instrument except following J443 -B J443 For J439 Sampling unit and FIR200 In-situ probe -
14 Equipmentstart-up B - Start-up work for analyzers and sampling units -
Notes*1: A:Required,B:optional*2: AnEthernetcableisrequired.Chooseeitheraspecialfiber-opticcableorelectriccabledependingonthefollowing
conditions: Alwaysuseaspecialfiber-opticcable(seesection2.6)foranexplosion-proofmodel,anelectriccablecannotbeused. For a general purpose model, the type of cable depends on the length. •Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Anelectriccablecannotbeused.*3: Foratotalcabledistanceoflongerthan20m,anadditionalfiber-opticcable(seesection2.6)shallbeprovidedbytheuser.
May 15, 2014-01
4
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
1.3Safetystandard,EMCstandardModel SafetyStandard EMCstandard
FT-NIRanalyzer I/OunitNR801AG, NR805AG, NR805AG/CSA*2
NR893AG*2
CSAC22.2No.61010-1*1FMClass3810(excludeNR805AG/CSA)ANSI/ISA61010-1(excludeNR805AG/CSA)ANSI/ISA-82.02.02(IEC61010-2-031)(excludeNR805AG/CSA)
-
NR801EG, NR805EG*2,*3
NR893EG*2,*3
EN61010-1*1 EN61326-1:ClassA,Table2(Foruseinindustrial location)EN 61326-2-3EN 61000-3-2 +A1 +A2EN 61000-3-3KoreaElectromagneticConformityStandardRCM(C-tick)mark
NR801JG, NR805JG(excludingNR805JG/K)
NR893JG - -
NR805JG/K - - KoreaElectromagneticConformityStandard
*1: •Installationcategory(Overvoltagecategory)II Describesanumberwhichdefinesatransientovervoltagecondition.Itimpliestheregulationforimpulsewithstand
voltage. “II”appliestoelectricalequipmentwhichissuppliedfromfixedinstallationslikedistributionboards. •PollutionDegree2 Describesthedegreetoasolid,liquid,orgaswhichdeterioratesdielectricstrengthorsurfaceresistivityisadhering. “2”appliestonormalindooratmosphere.Normally,onlynon-conductivepollutionoccurs.*2: Usea5Abreakerinthepowersupplycircuit.*3: ThisinstrumentisaClassAproduct,anditisdesignedforuseintheindustrialenvironment.Pleaseusethisinstrumentin
theindustrialenvironmentonly.
2.Componentspecifications(includingoptions)
2.1 NR800FourierTransformNear-InfraredAnalyzers
2.1.1 HardwarespecificationsPrinciple: Fourier-transformremotemeasurement
viafiber-opticcableMeasurementmethod: Opticaltransmission
absorptionMeasuredsample: LiquidBeamsource: Halogenlamp(recommended
replacementintervalforcontinuousoperation: 5000 hours)
Detector: InGaAs(indiumgalliumarsenide)photodiode,effectivewavelengthrange:900 to 2500 nm
Numberofmeasuringchannels: 1to4(non-moving)Housingstructures: Fieldsuitablewithfull
hingedfrontcoverFiber-optic cable entry: Cable glandFiber-optic cable connectors:
For measurement; FCFor Ethernet; ST
Display: LEDKeyboard: Coveredwithwater-proofsheetingOperatinglocationrequirements: Seechapter4.Grounding type: Independent, 100ΩorlowerInsulationresistance: 10MΩorgreater(500VDC)Withstandingvoltage:
PowerSupply WithstandingVoltage100VAC,115VAC 1000VACfor1min.200VAC,230VAC 1500VACfor1min.
Coating: Epoxy resin coatingPaintcolor:
ModelNR801AG,NR801EG,NR801JG
NR805AG,NR805AG/CSA,
NR805EG,NR805JG
Flame proof enclosure - Lampblack
(equivalenttoMunsell0.8Y2.5/0.4)Airregulatorcover -
Case Frostywhite(equivalenttoMunsell2.5Y8.4/1.2)
Displaypanel Lampblack(equivalenttoMunsell0.8Y2.5/0.4)
Wallmountbracket,Free standing rack
Deepseamossgreen(equivalenttoMunsell0.6GY3.1/2.0)
Mar. 18, 2014-00
5<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
Model NR801AG NR801EG NR801JGNR805AG(excluding
NR805AG/CSA)
NR805AG/CSA NR805EG
NR805JG(excludingNR805JG/K)
NR805JG/K
Explosion proof General purpose type(nonexplosion-proof) FM*1,*2 CSA*3 ATEX*2,*4 TIIS*5 KOSHA
*8CE marking - A*6 - - - A*6 - -Housingstructure IP53,NEMA3RorCSAType3RAir(purgegas)connection - 1/4NPT(F) Rc1/4or1/4NPT(F) Rc1/4or1/4NPT(F)
Cable connection Packinggland 3/4NPT(F) G3/4(F),3/4NPT(F),M25x1.5(F),orPG21(F)
Flame-proof packing glandorsealingfitting
Weight(kg)*7
Withoutbracket Approx. 50 Approx. 65
Withwallmount bracket Approx. 55 Approx. 70
Free stand rack Approx. 85 Approx. 100
*1: TypeXpurgingandexplosionproofforClassI,Division1,GroupsB,CandD.TemperatureClassT5. Allwiringshallcomplywiththemanufacture’sinstructions,NationalElectricalCodeANSI/NFPA70andLocalElectricalCodes.*2: Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.*3: TypeXpurgingandexplosionproofforClassI,Division1,GroupsB,CandD.TemperatureClassT5. Allwiringshallcomplywiththemanufacture’sinstructions,CanadianElectricalCodePartIandLocalElectricalCodes.*4: KEMA02ATEX2175,II2G,EExpdIIB+H2T5 Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.*5: ExpdIIB+H2T5X Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.*6: A: applicable*7: Notincludingweightofthepackagingmaterials.*8: KOSHA,ExdpIIB+H2T5 Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.
2.1.2 PerformanceWavelengthscanningrange: 900to2500nm
(11000to4000cm–1)Wavelengthresolution: 4,8,16,32,and64cm-1
(userselectable)Wavelengthreproducibility: 0.007cm-1Wavelengthaccuracy: 0.04cm-1S/Nratio 2250:1(RMS,resolution4cm -1, 11000 to
4000 cm -1, 1 sec.)
2.1.3 Inputs/Outputs(seealsosection2.2I/OUnit)
(1)CommunicationInterface:EngineeringPC: 1channel(Ethernet)DCS/I/Ounit: 2channels(RS-422) 1channelforDCS(Modbus)and
anotherforI/Ounit;or2channelsforI/Ounit
(2)Contactoutputs:Specification: NC/NOselectable,2channelsRating: 0.5A,30VDCorlessAction:
General purpose model:NR800analyzerpower”ON/OFF”;Powerfailsignal
StatusDO1TerminalNumber1to2 2to3
Power“off” Short OpenPower“on” Open Short
Mar. 18, 2014-00
NR800Systemalarm
StatusDO2TerminalNumber1to2 2to3
Power“off” Short OpenAlarm occurred Short OpenAlarm not occurred Open Short
Explosion proof model:NR800Systemalarm
StatusTerminalNumber
D1toD2 D2toD3Power“off” Open ShortAlarm occurred Open ShortAlarm not occurred Short Open
Annunciatorsignaloutput
StatusTerminalNumber
A1toA2 A2toA3Power“off” Open ShortDuringpurging Open ShortOverrideswitchpressed* Open ShortAfter purging Short Open
*: Overriding Whenopeningthedoorofthepressurized
explosion-protected enclosure under power on, the inner-pressure protection system will automatically turn off the power. The function that disables this automatic power off is called “overriding.”
Theoverrideswitchislocatedinsidetheanalyzer(flameproof).Youcanuseoverridingtoopenthedoor of the explosion-protected section.
Whenopeningthedoor,alwaysmakessuretheambient atmosphere is non-hazardous using, for example, a gas sensor.
6
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.1.4 Operatingmodes*1
Basicoperatingmodeanddescription
Channeloperatingmode
No. Auto/Manual.*3
Maintenance *2
Spectrum analysis of a reference sample,equipmentmaintenance
- -
Run
On-linemeasurement(allowsspectrum analysis on selected channels)
1AUTO
MANUAL
2AUTO
MANUAL
3AUTO
MANUAL
4AUTO
MANUAL
Notes*1: Whenthepoweristurnedon,theanalyzerstarts
accordingtoapredefinedoperatingmode.*2: Canperformspectrumanalysis(notcontinuous
measurement).*3: Auto: Performscontinuousmeasurement Manual: Canperformspectrumanalysis(not
continuous measurement).
2.1.5 ChangingandSettingOperatingModeSome operations are prohibited depending on the user level.
UserlevelDescription Changing/
settingmodeAbbreviationUserA UA For operator Unauthorized
UserB UB Forequipmentsupervisor Authorized
UserC UC For maintenance Authorized
•AuserlevelcanbeswitchedonthefrontconsolepaneloftheanalyzerorfromSPECTLAND2screenofanengineeringPC.
•ApasswordisrequiredtoswitchlevelsfromUAtoUBandswitchfromUAorUBtoUC.
2.1.6 SampleMeasurement
(1)ContinuousmeasurementNumber of measuring channels: 1to4(Specificationisrequired.)Dataupdatingperiod: 4secondsorlonger(dependingonthe
numberofaveragingandmeasuringchannels, as well as measurement items)
Number of measurement items: Upto12perchannel(48max.per
analyzer)Number of outlier detection items: Upto12perchannel(48max.per
analyzer) Separatemaintenanceisavailableforeachchannel(exceptforcommonhardware).
(2)Streamswitchingbyinputsignal(intermittentmeasurement)
Calibration model set for the each channel can be changed by external input signal. This function is used for multi-stream application or multi sample application.
Stream switchingNumberofstreamstoswitch: Upto16Switching patterns: See table 1. “Stream numbers
assignable to measuring channels corresponding to switching patterns”
Contact inputs: Seesection“2.2I/OUnit”RS-422 communication: Seesection“2.1.8MODBUS”
Dataupdatingperiod: 8secondsorlonger(dependingonthe
switchingpattern,numbersofaveragingand measuring channels, and stream configuration)
Number of measurement items: Upto12perstream(64max.per
analyzer)Number of outlier detection items: Upto12perstream(64max.per
analyzer)Separatemaintenanceisavailableforeachchannel
(exceptforcommonhardware).InputsignalsandmeasurementcomputationsequenceStreamidentification: Usedtoidentifytheselectedsample.
The analyzer will choose a calibration modelsettosuitefortherelevantstream based on this signal.
Measurementenabled(validsample): Whenclosed,theanalyzerperforms
measurement using the calibration modelspecifiedbythesampleidentificationsignalabove.
Schematictimingchartofmeasurementandcomputationsequence(e.g.2streams)
Stream 1 Stream 2 Stream 1
Computationperformed
Computationperformed
Computationperformed
OpenClosed
Closed
Open
Stream identification*
Measurement enabled
Notes(*) Astreamnumberisdefinedbyacombinationof
open/closedstatesofanidentificationcontactsignal.
Mar. 18, 2014-00
7<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
Table1 Streamnumbersassignabletomeasuringchannelscorrespondingtoswitchingpatterns
Streamswitch MeasuringchannelNo.*3Totalstreams Applicable
channelNo.Case PatternNo. 1 2 3 4None 0 1 2 3 4 4 max 1 to 4
Stream switching per channel*2
1*1 1to(17-N) 18-N 19-N 20-N 16 max 1 to 42 1 to 8 9 to 16 None None 16 max 1 to 23 1 to 4 5 to 8 9 to 12 13 to 16 16 max 1 to 4
Notes*1: N:Maximumnumberofmeasuringchannelsincludedwithintheanalyzer.*2: Thestreamnumberforameasuringchannelthatdoesnotswitchpathsmustbethesmallestnumberintherelevant
column.*3: Measuringchannelnumbersthatequalthenumberofmeasuringchannelsincludedwithintheanalyzerorsmallerarevalid.
2.1.7 ConsolePanel
(1)Displayitems
OperatingmodeLEDs(MODE) ThefollowingLEDsindicatethecurrent
basic operating mode:RUN: LitwhenintheRunmode.MAINT: LitwhenintheMaintenancemode. The operating modes of each channel
aredisplayedbyLEDswhenintheRunmode.
TwoLEDsareprovidedforeachchannel,amountingtoatotalof8LEDs.
AUTO: LitwhenintheAutomode.MAN: LitwhenintheManualmode.
Self-diagnosisLEDs(DIAGNOSIS)ALARM: Litwhenanalarmoccurs.LAMP: Litwhenalamphasburnedoutorafter
atimeperioddefinedbytheservicelifesetting elapses.
LASER: Litwhenthelaserhasburnedoutorafteratimeperioddefinedbytheservicelifesettingelapses.
LEDdisplay(16digits,STREAM/CONSTITUENT/VALUE)
The content depends on the operating status or operation.
Operationkeys Thefollowingsixkeysareprovided:ITEM: Usedtochangeitems.ENT: Usedtoconfirmtheentry.Arrowkeys: Usedtomovethecursororchange
display.
(2)Analyzerbehaviorforeachbasicoperatingmode
Maintenance modeBasicoperatingmodeLEDs: OnlyMAINTlightsup.ChanneloperatingmodeLEDs: Litinaccordancewitheachsetting.
Self-diagnosisLEDs: Litinaccordancewiththeresultsof
self-diagnosis.LEDdisplay: The display depends on the operating
status.Run modeBasicoperatingmodeLEDs: OnlyRUNlightsup.ChanneloperatingmodeLEDs: Litinaccordancewitheachsetting.Self-diagnosisLEDs: Litinaccordancewiththeresultsof
self-diagnosis.LEDdisplay: The display depends on the operating
status.
2.1.8 MODBUS
(1)Communicationspecifications
Item SpecificationCommunication standard*
RS-422orRS-232C(aconverterisrequired)
Start bit 1Number of data bits 7(ASCIImode)or8(binarymode)Paritybit 1Stop bit 1Communication speed 4800,9600or19200(selectable)
Error detection Oddnumberparity,EvennumberParityornone(selectable)
*: OnlyRS-232CcanbeusedinNR805*Gexplosion-proof type.
See section 1.2 and 2.5
(2)Serialcommunication
Item ASCIImode RTUmodeNumber of data bits 7(ASCII) 8(binary)
Message starting character Colon “:” None
Message ending character
Carriagereturn/Linefeed“<CR><LF>” None
Error detection LRC(LogicalRedundancy Check)
CRC-16(CyclicRedundancy Check)
Inter-character time out None 100msec
Mar. 18, 2014-00
8
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
(3)SupportfunctionintheMODBUSprotocol
FunctionNo. Description
01 ReadstheON/OFFstatusofcoils.02 ReadstheON/OFFstatusofinputrelays.03 Readsthecurrentvalueofholdingregisters.04 Readsthecurrentvalueofinputregisters.05 Forcibly changes the status of a coil.06 Writeavaluetoaholdingregister.
08Loop-backtestSends back the same message as the command message.
(4)Recommendedvalueofmasterside
Item RecommendedvalueTransmissionspeed(bps) 4800/9600/19200Time monitor of Inter-character gap 1000msNo reply time 2 sec.Number of transmission retries 5 timesCommunicationRecoveryPeriod 30 sec.
Mar. 18, 2014-00
2.1.10ModelandSuffixCodes
(1)NR801AG,Generalpurposemodel
Model SuffixCode OptionCode DescriptionNR801AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800FT-NIRAnalyzer,GeneralPurposemodelLanguage -E ••••••••••••••••• EnglishPowersupply 1
346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelengthscanningrange W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDualcable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1
2
•••••••••••••••••
•••••••••••••••••
Electric cable, only for general purpose model and less than 40 mFiber-optic cable
Mounting ABC
•••••••••••••••••••••••••••••••••••••••••••••••••••
WithoutbracketsWithwall-mountingbracketsWithfreestandingrack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS Withstreamswitchinputfunction
2.1.9 OtherFunctions
(1)Baselinecompensation:Upto10points
(2)Outputoftheresultcompensatedusinganalogueinputs
The following operation result can be output.Compensation operation processing:
P2=C1*P1+C0+D D=An2 *D2+An*D1+D0
An: Analogueinput(n:4orless)P1: Measurementvaluebeforethe
compensationP2: Measurementvalueafterthe
compensationD0,D1,D2,C0,C1: Coefficient
(3)OutputsmoothingSimplemovingaverage(Upto10points)Exponentialmovingaverage
(4)On-linemeasurementspectrasaving
9<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
(2)NR805AG,TypeXpurgedandFMapproved,CSAapprovedexplosion-proofmodel
Model SuffixCode OptionCode DescriptionNR805AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion proof model, FM
approvedversion(*1)Language -E ••••••••••••••••• EnglishPowersupply 1
346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N"Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelengthscanningrange W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDualcable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A
BC
•••••••••••••••••••••••••••••••••••••••••••••••••••
WithoutbracketsWithwall-mountingbracketsWithfreestandingrack
- -N ••••••••••••••••• Always “-N"- -N ••••••••••••••••• Always “-N"Cable entrance -2 ••••••••••••••••• Female3/4NPTPurgeairconnection 2 ••••••••••••••••• Female1/4NPTOption /SS
/CSAWithstreamswitchinputfunctionCSAapprovedversion(*1)
(*1) Optioncode"/CSA"isspecifiedCSAapproved Optioncode"/CSA"isnotspecifiedFMapproved
Mar. 18, 2014-00
10
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
(3)NR801EG,Generalpurposemodel,withCEmarking
Model SuffixCode OptionCode DescriptionNR801EG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800FT-NIRAnalyzer,GeneralPurposemodel,
CE markingLanguage -E ••••••••••••••••• EnglishPowersupply 1
346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelengthscanningrange W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDualcable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1
2
•••••••••••••••••
•••••••••••••••••
Electric cable, only for general purpose model and less than 40 mFiber-optic cable
Mounting ABC
•••••••••••••••••••••••••••••••••••••••••••••••••••
WithoutbracketsWithwall-mountingbracketsWithfreestandingrack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS Withstreamswitchinputfunction
Mar. 18, 2014-00
11<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
(4)NR805EG,ATEXapprovedexplosion-proofmodel,withCEmarking
Model SuffixCode OptionCode DescriptionNR805EG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion-proof model,
ATEXapproved,CEmarkingLanguage -E ••••••••••••••••• EnglishPowersupply 1
346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelengthscanningrange W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDualcable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A
BC
•••••••••••••••••••••••••••••••••••••••••••••••••••
WithoutbracketsWithwall-mountingbracketsWithfreestandingrack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Cable entrance -3
-4-5-6
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
FemaleG3/4Female3/4NPTFemale M25x1.5FemalePG21
Purgeairconnection 12
••••••••••••••••••••••••••••••••••
FemaleRc1/4Female1/4NPT
Option /SS Withstreamswitchinputfunction
Mar. 18, 2014-00
12
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
(5)NR801JG,Generalpurposemodel
Model SuffixCode OptionCode DescriptionNR801JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800FT-NIRAnalyzer,GeneralPurposemodelLanguage -E
-J••••••••••••••••••••••••••••••••••
EnglishJapanese
Powersupply 1346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelengthscanningrange W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDualcable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1
2
•••••••••••••••••
•••••••••••••••••
Electric cable, only for general purpose model and less than 40 mFiber-optic cable
Mounting ABC
•••••••••••••••••••••••••••••••••••••••••••••••••••
WithoutbracketsWithwall-mountingbracketsWithfreestandingrack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS Withstreamswitchinputfunction
Mar. 18, 2014-00
13<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
(6)NR805JG,TIISapprovedexplosion-proofmodel*1
Model SuffixCode OptionCode DescriptionNR805JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion-proof model,
TIISapproved*1Language -E
-J••••••••••••••••••••••••••••••••••
EnglishJapanese*1
Powersupply 1346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelengthscanningrange W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDualcable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A
BC
•••••••••••••••••••••••••••••••••••••••••••••••••••
WithoutbracketsWithwall-mountingbracketsWithfreestandingrack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Cable entrance -1
-2-3-4
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Metalconduit(G3/4)Metalconduit(3/4NPT)Flameproofpacking(G3/4)Flameproofpacking(3/4NPT)
Purgeairconnection 12
••••••••••••••••••••••••••••••••••
FemaleRc1/4Female1/4NPT
Option /SS/K
WithstreamswitchinputfunctionKOSHAapprovedexplosionproofmodel*1
*1: “TIIS”approvalisnotapplicablewhenoptioncode“/K”isselected.“Japanese”cannotbeselectedwithoptioncode“/K.”
Mar. 18, 2014-00
14
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.1.11Externaldimensions
(1)Generalpurposemodel(NR801*Gtype)•Withoutbrackets(Suffixcode:A)
6282 60 60 60 60 62 62
68±5
30±5
480
92
60430
277
665
355550
Mounting screws4-M8
2-cable outlets (For NR801AG, NR801EG)(Packing glands, acceptable cables: ø10 to 13)
Earth terminal(M4 screw)
Power supply cable connection inlet (For NR801AG, NR801EG)(Packing gland, acceptable cable: ø10 to 15.9)
3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)
Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)
4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)
Unit: mm
For NR801AG, NR801EG
Part A
Part A
Note: The wall construction should be withstood a force of four times the weight of the NR800.
Mar. 18, 2014-00
15<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
•Wall-mountingmodel(Suffixcode:B)
30±5
665
550
60 430
50 50
603086
0
355 20
920
6282 60 60 60 62 6260
277
68±5
Mounting holes4-ø10.5 holes
Earth terminal(M4 screw)
Power supply cable connection inlet (For NR801AG, NR801EG)(Packing gland, acceptable cable: ø10 to 15.9)
2-cable outlets (For NR801AG, NR801EG)(Packing glands, acceptable cables: ø10 to 13)
3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)
Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)
4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)
Wall mounting bracket
Unit: mm
For NR801AG, NR801EG
Part A
Part A
Note: The wall construction should be withstood a force of four times the weight of the NR800.
Mar. 18, 2014-00
16
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
•Free standing rack model(Suffixcode:C)66
5
1500
70
350 8080
30±5
68±5
550 355 50
630
550
490
56827
7
6282 60 60 60 62 6260
Mounting holes 4-ø17 (for anchor bolts)
4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)
Earth terminal(M4 screw)
Unit: mm
Power supply cable connection inlet (For NR801AG, NR801EG)(Packing gland, acceptable cable: ø10 to 15.9)
2-cable outlets (For NR801AG, NR801EG)(Packing glands, acceptable cables: ø10 to 13)
3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)
Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)
For NR801AG, NR801EG
Part A
Part A
Mar. 18, 2014-00
17<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
(2)Explosion-proofmodel(NR805*Gtype)•Withoutbrackets(Suffixcode:A)
62
82180
93
36
6262332
665
355550
App
rox.
300
200
430 60
9248
0
4-M8Mounting screws
95
200
32±5
Connection: M25x1.5
Connector *2
For NR805EG
Internal-pressure protection unit(flameproof enclosure)Cable connectors(with sealing fitting orflameproof packing gland *1)
Air (purge gas) connection: Rc1/4 or 1/4NPT(only 1/4NPT for NR805AG)
Connector *4
Connection:3/4NPT
For NR805AG
Earth terminal(M4 screw)
86
277
216
65
4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)
2-cable outlets *3 *4(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)
Special Ethernet fiber-optic cable outlet
Unit: mm
For NR805AG, NR805EG
Part B
Part A
Part A Part A
Part B
*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: Selecttheconnectorfromthefollowing. 3/4NPT,PG21orG3/4*3: InthecaseofNR805AGorNR805EG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*4: InthecaseofNR805EG,selecttheconnectorfromthefollowing. 3/4NPT,PG21,G3/4orM25x1.5 TheconnectorofNR805AGis3/4NPT.Note: The wall construction should be withstood a force of four times the weight of the NR800.
Mar. 18, 2014-00
18
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
•Withwall-mountingbrackets(Suffixcode:B)
62
App
rox.
300
36
82180
6093
6262332
665
550
60 430
50 50
3086
0
200
200
32±5
For NR805AG
Connection: 3/4NPT
Connector *2
Connection: M25x1.5
Mounting holes 4-ø10.5 holes
For NR805EG
Internal-pressure protection unit(flameproof enclosure)
Cable connectors(with sealing fitting or flameproof packing gland *1)
Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)
4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)
95
Earth terminal(M4 screw)
86
277
21665
Special Ethernet fiber-optic cable outlet
2-cable outlets *3 *4(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)
920
355 20
Wall mounting bracket *5
Connector *4
Unit: mm
For NR805AG, NR805EG
Part A
Part A
Part A
Part B
Part B
*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: Selecttheconnectorfromthefollowing. 3/4NPT,PG21orG3/4*3: InthecaseofNR805AGorNR805EG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*4: InthecaseofNR805EG,selecttheconnectorfromthefollowing. 3/4NPT,PG21,G3/4orM25x1.5 TheconnectorofNR805AGis3/4NPT.*5: ThewallconstructionshouldbewithstoodaforceoffourtimestheweightoftheNR800.
Mar. 18, 2014-00
19<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
•Withfreestandingrack(Suffixcode:C)
568
490
550
630
70
1500
665
App
rox.
300
80
200
550
Mounting holes 4-ø17 (for anchor bolts)
3693
Connection: 3/4NPT
For NR805AG
For NR805AG, NR805EG
32±5
200
Connector *2
Connection: M25x1.5
Internal-pressure protection unit(flameproof enclosure)
Cable connectors(with sealing fitting or flameproof packing gland *1)
Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)
For NR805EG
82180
350 80
95
Connector *4
50355
Bottom of analyzer
277
216
8665
Earth terminal (M4 screw)
62332 62 62
4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)
2-cable outlets *3 *4(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)
Special Ethernet fiber-optic cable outlet
Unit: mm
Part B
Part A
Part A
Part A
Part B
*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: Selecttheconnectorfromthefollowing. 3/4NPT,PG21orG3/4*3: InthecaseofNR805AGorNR805EG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*4: InthecaseofNR805EG,selecttheconnectorfromthefollowing. 3/4NPT,PG21,G3/4orM25x1.5 TheconnectorofNR805AGis3/4NPT.
Mar. 18, 2014-00
20
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.2 I/OUnit
2.2.1 OverviewTheI/Ounitisaninput/outputinterfacebetweentheanalyzerandexternalmonitor/controlsystemsuchasDCS.TheI/OunitconnectsNR800analyzertoserialcommunications(RS-422).UptotwoI/Ounitscanbe connected to the Analyzer, and one of which can incorporate the contact input for stream switching and analogue input for compensation options.
2.2.2 SpecificationsPowersupply: Seesection“3.UtilitySpecifications”.Insulationresistance: 5MΩormore,500VDCWithstandvoltage: 1500VACfor1min.Analogue output: Outputdata: Measurementresults(Propertiesand
concentration)Number of outputs: 0 to 40Outputspecifications:
Item Description
Outputrange 4to20mADC(3.0to21.0mADC,Floating-common type)
Isolation method
Between output terminals and internal circuit:PhotocouplerisolationBetween output terminals: non-isolated, Commonnegative
Withstandvoltage 500VDCfor1min.Allowable load resistance 600Ωorless
12-bitD/Aconverterresolution 5.7 µA
System accuracy ±0.5%offullscale
Separately,24VDCpowersupplyisrequired(seesection“2.2.3.ModelandSuffixcodes”and“3.UtilitySpecifications”).
Contact output:
Alarm Quantityofoutputs Alarm Quantity
ofoutputsGeneral 1 Outlier 4Communication failure 1 I/Ounitfailure 1
Operatingmode 5 - -
Contactoutputspecifications:Item Description
I/Ounitfailure
Rating 24VDC.0.3A
Action and number1 normally open and 1 normallyclosed(sharedcommon)
Alarm, operating mode, and outlier detection
Insulation method Mechanical isolationRated load voltage
DC 24VAC 100to240V
Maximum load current 2Amps/point,8Amps/common
Servicinglife
Mechanical At least 20 million actions
Electrical At least 100 thousand actions
Surge killer NoneNumber per common 8points/commonExternal power supply Notrequired.
Contact input:Description:Streamidentificationformulti-streamsampling
unit : 8 points, status signals to identify
sample streams that pass through measurement cells.
Measurementenabled(streamvalid): 4points,statussignalstoconfirmthat
samples inside the measurement cells are ready for measurement.
Inputspecifications:Item Description
Input type VoltagefreecontactCommon terminal Common to 8 pointsIsolation method Transformer isolation
WithstandvoltageBetween external connectors collectivelyandinternalcircuit:500VDCfor1min.
Off-stateopen-circuitvoltage 5to7V
On-stateloadcurrent 1 to 3 mA
On-stateloadresistance 200Ωorsmaller
Off-stateloadresistance 100Ωorlarger
Provideexternalcontact-inputsignalsmeetingtheaboverequirements.
Analogue input:Input data and number: Analogue output
compensation signal, 4 pointsInputspecifications
Item DescriptionInputrange(actual) 1to5VDC(-0.25to5.25VDC)
Isolation method
Between input terminals and internal circuit: photo coupler isolationBetween input terminals: non-isolated, negativecommon
Withstandvoltage 500VDCfor1min.Input resistance 1MΩ12-bitA/Dconverterresolution 1to5VDC:1.4mV
System accuracy ±0.5%offullscaleat0to40°C
Installationrequirements:Seechapter4.Mounting: WallmountingExternal connection terminal: M3.5 screwDrip-proofanddust-proofconstruction: IP4X(withenclosure)
Mar. 18, 2014-00
21<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
2.2.3 ModelandSuffixcodes
(1)NR893AG:forNR801AGandNR805AGanalyzer
Model SuffixCode OptionCode DescriptionNR893AG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IOunit(nonexplosion-proof)Language -E ••••••••••••••••• EnglishAnalyzer 1
2••••••••••••••••••••••••••••••••••
NR805AG(Explosion-proofmodel) (*1)NR801AG(GeneralPurposemodel)
Contact input for stream switching
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Analogue input for compensation
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Number of analogue outputs -00-04-08-12-16-20-24-28-32-36-40
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
None481216202428323640
Enclosure -A-B-C
•••••••••••••••••••••••••••••••••••••••••••••••••••
Number of analog outputs: 0 to 12Number of analog outputs: 0 to 28Number of analog outputs: 0 to 40
- -N ••••••••••••••••• Always “-N”
*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.
(2)NR893EG:forNR801EGandNR805EGanalyzer
Model SuffixCode OptionCode DescriptionNR893EG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IOunit,CEmarking(nonexplosion-proof)Language -E ••••••••••••••••• EnglishAnalyzer 1
2••••••••••••••••••••••••••••••••••
NR805EG(Explosion-proofmodel) (*1)NR801EG(GeneralPurposemodel)
Contact input for stream switching
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Analogue input for compensation
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Number of analogue outputs -00-04-08-12-16-20-24-28-32-36-40
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
None481216202428323640
Enclosure -C ••••••••••••••••• Always “-C”- -N ••••••••••••••••• Always “-N”
*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.
Mar. 18, 2014-00
22
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
(3)NR893JG:forNR801JGandNR805JGanalyzer
Model SuffixCode OptionCode DescriptionNR893JG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IOunit(nonexplosion-proof)Language -E
-J••••••••••••••••••••••••••••••••••
EnglishJapanese
Analyzer 12
••••••••••••••••••••••••••••••••••
NR805JG(Explosion-proofmodel) (*1)NR801JG(GeneralPurposemodel)
Contact input for stream switching
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Analogue input for compensation
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Number of analogue outputs (*2)
-00-04-08-12-16-20-24-28-32-36-40
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
None481216202428323640
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Option /C Withenclosureincluded24VDCpowersupplyunit
foranalogoutput.(*2)
*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.
*2: Separate24VDCpowersupplyrequiredwhenusingananalogueoutputandoptioncode“/C”arenotspecified.(seechapter 3).
Mar. 18, 2014-00
23<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
2.2.4 Externaldimensionsandweight:
(1)NR893AG,NR893EGPaintcolor:Lightbeige(Munsell5Y7/1orequivalent)
B D F
CA
E G
H
75 17
16118°
1743
5028
0
400
68.5
200
6100 505050505050
L + 50
L50 50
4-ø14 holesfor mounting *1
ø20 holefor protective ground wiring
Number ‘N’ of wiring holesSuitable cable OD *2 A: ø10 to 15.9 (to the power supply) B to H: ø10 to 18.5
Unit: mm
Enclosurecode
Numberofanalogoutputs L(mm) N Approximate
WeightApplicablemodelname
-A 0 to 12 400 6(AtoF) 15.5 kg NR893AG-B 0 to 28 500 7(AtoG) 17.5 kg NR893AG
-C*3 0 to 40 600 8(AtoH) 20.0 kg NR893AGNR893EG
*1: The wall construction should be withstood a force of four times the weight.*2: Sinceattachedferritecoresareused,thepracticalcableODareuptoø13mm.*3: InthecaseofNR893EG,theenclosurecodeisonly-Cirrespectiveofthenumberofanalogoutputs.
Mar. 18, 2014-00
24
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
(2)I/OunitWithoutenclosure(onlymodelNR893JG)•ForExplosionproofanalyzerunit•ThesizeofIOunitforgeneralpurposeanalyzerunitisthesameasabovefigurebutRS-422/232Cconverterpowersupplycaseremoved.Paintcolor:Baseplate,RS-422/232Cconverterpowersupplycase;Modulecolors: Light cobalt blue (Munsell 6.2PB4.6/ 8.8 or equivalent), lamp black (Munsell 0.8Y2.5/0.4 or
equivalent)
215
199
L
303 151510
(101)Approx. 120
N-ø7.2 holesfor mounting *1
Safety barrier power supply caseused only with NR805JG explosion-proof analyzer
Unit: mm
Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 333 4 6.5 kg-16, -20, -24, -28 16, 20, 24, 28 450 6 7.0 kg
-32, -36, -40 32, 36, 40 537 6 7.5 kg
*1: Thewallconstructionshouldbewithstoodaforceoffourtimestheweight.
(3)Withenclosure(Optioncode:/C)Paintcolor:Lightbeige(Munsell5Y7/1orequivalent)
C E
D
A
F H
G
5050
B
5050 50505075 17
10118°
1743
5028
0
400
App
rox.
30
200
6
L + 50
L50 50
4-ø14 holesfor mounting *1
ø20 holefor protective ground wiring
Number ‘N’ of wiring holesSuitable cable OD: ø10 to 18.5
Unit: mm
I/O Unit (inside)
Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 400 6(AtoF) 15.5 kg-16, -20, -24, -28 16, 20, 24, 28 500 7(AtoG) 17.5 kg
-32, -36, -40 32, 36, 40 600 8(AtoH) 20.0 kg
*1: Thewallconstructionshouldbewithstoodaforceoffourtimestheweight.
Mar. 18, 2014-00
25<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
2.3MeasurementCells
2.3.1 FlowthroughCell
(1)SpecificationsOpticalpathlength: 1,2,5,10,or20mmFiber-optic connector: FC typeWettedpartmaterial: Fluorinerubber,Kalrez4079,316SS,PTFE,Borosilicatecrownglass,orSapphireSamplepressurerange: Atmosphericpressureupto1.9MPaGSampletemperaturerange: –5°Cto+80°C(ItvarybythematerialsofO-ring.ContactYokogawaforfurther
information.)Sampleconnection: FemaleRc1/4Installationangle: VerticalInstallationlocationrequirements:Seesection4.Weight: Approx.3kg
(2)ModelandSuffixCodes
Model SuffixCode OptionCode DescriptionNR510 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cellWindowmaterial and optical path length
-B00
-B01-B02-B05-B10-B20-S00
-S01-S02-S05-S10-S20
•••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Borosilicatecrownglass,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm
Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ringmaterial -B
-K••••••••••••••••••••••••••••••••••
Fluorine rubberKalrez4079
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”
(3)ExternalDimensions
Unit: mm(96)55 29
102
92
L4-ø5.2 holes
2-Rc1/4
Optical path length (mm)
1251020
L (mm)
64±2
65±2
68±2
73±2
83±2
Fiber-optic connection(in the other side)
Mar. 18, 2014-00
26
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.3.2 FlowthroughCellwithConstantTemperatureWaterTube
(1)SpecificationsOpticalpathlength: 1, 2, 5, 10, or 20 mmFiber-optic connector: FC typeWettedpartmaterial: Fluorinerubber,Kalrez4079,316SS,PTFE,Borosilicatecrownglass,orSapphireSamplepressurerange: Atmosphericpressureupto1.9MPaSampletemperaturerange: +5°Cto+80°C(ItvarybythematerialsofO-ring.ContactYokogawaforfurther
information.)Constantwatertemperaturerange:+5°Cto+80°C(ItvarybythematerialsofO-ring.ContactYokogawaforfurther
information.)Sampleconnection: FemaleRc1/4Connection for water with constant temperature: 6-mm outside diameter, copper tubeInstallationangle: VerticalInstallationlocationrequirements:Seesection4.Weight: Approx.3kg
(2)ModelandSuffixCodes
Model SuffixCode OptionCode DescriptionNR512 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cell with constant temperature water
tubeWindowmaterial and optical path length
-B00
-B01-B02-B05-B10-B20-S00
-S01-S02-S05-S10-S20
•••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Borosilicatecrownglass,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm
Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ringmaterial -B
-K••••••••••••••••••••••••••••••••••
Fluorine rubberKalrez4079
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”
(3)ExternalDimensions
Unit: mm
4-ø5.2 holes
2-Rc1/4
L29
Approx. 160
App
rox.
160
102
92
B-SIDEB-SIDE A-SIDE A-SIDE
Thermal insulation jacket
Approx. 145
70111
Optical path length (mm)
1251020
L (mm)
64±2
65±2
68±2
73±2
83±2Fiber-optic connection (in the other side)
Mar. 18, 2014-00
27<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
2.3.3 SpecialmeasurementcellAPFAflowthroughtypespecialcellisdescribedhere.
(1)SpecificationsOpticalpathlength: 1,2,5,or10mmFiber-optic connector: FC typeWettedpartmaterial: PFA,Kalrez4079orMoriseiperflioro(MP4275B),orSapphire (Stainlesssteelispartlyusedasnon-wettedparts.)Samplepressurerange: Atmosphericpressureupto0.3MPaGSampletemperaturerange: 5°Cto80°CSampleconnection: 1/4inchs(Ø6.35xØ3.95mm)PFAtube,orØ6xØ4mmPFAtube (length:3000mm)Installationangle: VerticalInstallationlocationrequirements:Seesection4.Weight: Approx.1.5kgOthers: O-ringsandPFAbodymustbeexchangedsimultaneously.
(2)ModelandSuffixCodes
Model SuffixCode OptionCode DescriptionNR511 ••••••••••••••••••••••• ••••••••••••••••• Special measurement cellWindowmaterial and optical path length
-S01-S02-S05-S10
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mm
Body material P ••••••••••••••••• PFAO-ringmaterial -K
-P••••••••••••••••••••••••••••••••••
Kalrez4079Moriseiperflioro(MP4275B)
Tube -A-B
••••••••••••••••••••••••••••••••••
1/4inchs(Ø6.35xØ3.95mm),3000mm,PFAØ6xØ4mm,3000mm,PFA
Option /Z Alwaysadd“/Z”
(3)ExternalDimensions
Unit: mm(80)(50)
90
(300
0)(3
000)
74
77(94)
4-ø5.2 holes
Fiber-optic connection(in the other side)
Mar. 18, 2014-00
28
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.3.4 In-situprobe
(1)SpecificationsFiber-optic connector: FC typeWetpartmaterial:
Body; HastelloyC-276orUNSS31603(Stainlesssteel316L)Others; Sapphire,HastelloyC-276,Gold
Samplepressurerange: Atmosphericpressureto0.98MPa*Sampletemperaturerange: Normaltemperatureto80°C*Purgegasconnection: Female1/8NPTx2Fiber-opticentrance: Female1NPTInstallationangle: Vertical(Insertportionisdownward)orHorizontalInstallationlocationrequirements:Seesection4Weight: Approx.5kg(excludingflange)*: Whenthesamplepressureistheoutsideoftherange,pleaseconsultthesalesperson.
(2)ModelandSuffixCodesFIR200/Z
(3)ExternalDimensions
L *3
( 560 ) *4 ( 80 )
Approx. 35 A *1
Flange *2
Fiber-optic connection box
Fiber-optic entrance(Exclusive packing)
Purge-gas connections2-1/8NPT
Approx. ø27
Unit: mm
*1:Opticalpathlength“A”:2/5/10mm,selectable(Dependentonsampleconditions,measurementitems,measurementrange, etc. Consult Yokogawa.)
*2:Mountingflange(HastelloyC-276/UNSS31603comparable)shouldbeprovidedbycustomer.*3:Specifylength“L”withintherange150to250mm.*4:Theshapeandsizeofconnectionboxmaychange.
Note1:Keepthetemperatureattheconnectionofthefiber-opticcableat80°Corbelow.Ifthetemperatureattheconnectionexceeds80°C,theconnectionboxshouldbepurgedwithclean,dryair,orotherappropriatemeasuresshouldbetaken.Inthiscase,theconnectionboxmaybedifferentinshapeandsizefromtheoneshownabove.Ifafiber-opticcableformeasurement is exposed to radiant heat, it should be shielded.
Note2:Dimensionsofeachpartmaychangewithoutpriornotice.
2.4 Fiber-opticCablesformeasurementFiber-optic cables connect an analyzer to measurement cell. Select appropriate cables to meet your application requirement,e.g.,wavelengthsscanningrange.
2.4.1 SilicaFiber-opticCable
(1)SpecificationsApplicablewavelengthrange: 900to2100nmConnector: Double-endFCorFC-SMAtypeStructure: Dual(forsampleandreference)orsingle(sample),two-core,protectedby
stainlessflexibletubeMinimum bending radius: 100 mm. To reduce optical attenuation, make the radiuses along the cable as
large as possible when laying cables.Installationlocationrequirements:Seesection4.Cabling: Conduit protected cabling is recommended.
Mar. 18, 2014-00
29<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
(2)ModelandSuffixCodes
a.Singlefiber-opticcableModel SuffixCode OptionCode Description
NR821 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2100nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side
Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
5 m10 m20 m30 m50 m100 m150 m200 m300 m
- -000 ••••••••••••••••• Always “-000”
b.Dualfiber-opticcableModel SuffixCode OptionCode Description
NR822 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2100nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side
Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
5 m10 m20 m30 m50 m100 m150 m200 m300 m
- -000 ••••••••••••••••• Always “-000”
(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.
• Double-endFCconnector
BA
A B
Unit: mmFC connector FC connector
(Cable length)600±50
(57) 3010
1510
• FC(analyzer)-SMA(measurementcell)connector
BA
A B
Unit: mmFC connector (analyzer) SMA connector
(measurement cell)
(Cable length)600±50
(57) 3010
600±50
(53) 8
1510
Mar. 18, 2014-00
30
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.4.2 FluorideFiber-opticCable(1)Specifications
Applicablewavelengthrange: 900to2500nmConnector: Double-endFCtypeorFC-SMAtypeStructure: Dual(forsampleandreference)orsingle(sample),two-core,protectedby
stainlessflexibletubeMinimum bending radius: Analyzer side; 100 mm, Measurement side;120 mm To reduce optical attenuation, make the radiuses along the cable as large as
possible when laying cables.Installationlocationrequirements:Seesection4.Note: Fluoride,amaterialusedintheNR823andNR824fiber-opticcables,hasthedeliquescenceproperties.Whenstoring
thesecables,takeappropriatemeasurestopreventmoisture,e.g.,acableshouldbeplacedwithdesiccantinasealed,moisture-barrier plastic bag and the desiccant replaced regularly.
(2)ModelandSuffixCodes(ModelcodetoincludefibersforAnalyzer–Cell(probe)–Analyzer)a.Singlefiber-opticcable
Model SuffixCode OptionCode DescriptionNR823 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2500nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side
Cable length -L003-L004-L005
•••••••••••••••••••••••••••••••••••••••••••••••••••
3 m4 m5 m
- -000 ••••••••••••••••• Always “-000”
b.Dualfiber-opticcableModel SuffixCode OptionCode Description
NR824 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2500nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side
Cable length -L003-L004-L005
•••••••••••••••••••••••••••••••••••••••••••••••••••
3 m4 m5 m
- -000 ••••••••••••••••• Always “-000”
(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.(Connectthelongerendtothe measuremrnt cell side.)• Double-endFCconnector
BA
A B
Unit: mmFC connector (analyzer) FC connector
(measurement cell)
(Cable length)600±50 800±50
(57) 3010
1510
• FC(analyzer)-SMA(measurementcell)connector
BA
A B
Unit: mmFC connector (analyzer) SMA connector
(measurement cell)
(Cable length)600±50
(57) 3010
800±50
(53) 8
1510
Mar. 18, 2014-00
31<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
2.5 RS-422-to-RS-232CConverter(PartNumber:K9404LAorK9404LD)ThisunitconvertsoutputsignalsoftheanalyzermainunitfromtheRS-422formattotheRS-232CformatforpersonalcomputerorDCScommunicationandalsoblockscommunicationsignalswhenitreceivesapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity.
2.5.1 SpecificationsPowersupply: Seesection3.Signal terminals: Analyzermainunitside(RS-422):M4,outputside(RS-232C):FemaleDsub25-pinGroundingtype: 100ΩorlowerInstallationlocationrequirements:Seesection4.Housingstructure: Desktop(K9404LA),Screwcramp(K9404LD)Weight: Approx.2kgNote: TheconverterdoesnotcarrytheCEMarking.Anappropriatehousingshouldbeprovidedbycustomer,ifnecessary.
2.5.2 TerminalDiagram
CN3 CN2
LN-+RDBRDASDBSDA
RS-422
From analyzer
D/I 12V AC power supply
Female D-SUB 25-pin connectorCN1
RS-232C
2.5.3 RS-232CCommunicationStraightCable
GND
123456789
CDRDSDERSGDRRSCSCI
FGCDRDSDERSGDRRSCSCI
GND183220764522
Internal connection
To converter To host
2.5.4 ExternalDimensions
4-ø4.5 holes203
126
126
39.5
15 15
9611
.5 218
203
37
K9404LA Desktop type K9404LD Screw cramp type
Unit: mm
May 15, 2014-01
32
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.6 EthernetFiber-opticCableCableforEthernetcommunicationbetweentheanalyzerandtheengineeringPC.
2.6.1 SpecificationsLength: Upto20mConnector: ST typeStructure: Two-core,protectedbystainlessflexibletubeInstallationlocationrequirements:Seesection4.Minimum bending radius: 50 mm. To reduce optical attenuation, make the radiuses along the cable as
large as possible when laying cables.
2.6.2 ModelandSuffixCodesModel SuffixCode OptionCode Description
NR895 ••••••••••••••••• ••••••••••••••••• Ethernetfiber-opticcableLength*1 -L003
-L005-L010-L020
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
3 m5 m10 m20 m
- -000 ••••••••••••••••• Always “-000”Option /JB Withjunctionbox*2
*1: Foratotalcabledistanceof20morlonger,anadditionalfiber-opticcable(fittedwithSTconnectors)mustbeprovidedbythe user.
Seechapter6.2forthecablespecificationofuserprovisions.*2: TheJunctionboxisforconnectionbetweenanNR895fiber-opticcableandtheadditionalcable.
2.6.3 ExternalDimensions(1)Fiber-opticCable
15
9.310
Blue
Yellow
Unit: mm
1000 0+100
43 30
(Cable length)
Mar. 18, 2014-00
33<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
(2)JunctionBoxforEthernetfiber-opticcables(Optioncode:/JB)
400 430
305
300
142
24120
280
60
20260
4-ø10
Handle to open
100 30Hole
2-Cable gland
Unit : mm
NR895_JB.ai
Housingstructure: IP44Coating: Epoxy resin coatingPaintcolor: Lightbeige(equivalenttoMunsell5Y7/1)Material: IronWeight: Approx.10kg
Mar. 18, 2014-00
34
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.7 SPECTLAND2MeasurementandMaintenanceSoftware
2.7.1 OverviewSPECTLAND2isanapplicationsoftwarethatcontrolsNR800operationandmonitoringonaPC.Itdisplaysthe analyzer status, and instructs measurement and sets parameters for the analyzer. To use SPECTLAND2,firstinstallitintheEngineeringPC.ThecontinuousoperationofSPECTLAND2forextended periods of time is not guaranteed.
MainFeatures•Instructstheanalyzertomeasurespectra,savedata
and display. The measured spectra can be used to generate calibration models.
•Showstrendgraphsofthemeasurementsandsavesthemtofilesduringacontinuousmeasurement.
•Setsparametersforcontinuousmeasurement.•Displaysvariousdataoftheanalyzer,suchas
operating modes, alarm status, and maintenance data.
•Instructstheanalyzertoperformsuchtasksasoperation mode change, spectra measurement, and setting property information, calibration models, or measurement conditions.
2.7.2 MainWindows(1)ManualSpectrumWindow
Enables the analyzer to measure spectra, which are tobeprocessedbyChemometrics(calibrationmodelgenerationsoftware).Alsoallowsdatatobesavedtofilesanddisplayed.
(2)AutoSpectrumWindowAllows the user to upload spectra data to the EngineeringPCduringcontinuousmeasurementuponreceivingasignalatperiodicintervals,outlierdetection,orapropertyvaluevariationfailure.Thisdata is displayed for each measuring channel.
(3)PowerSpectrumWindowIt displays power spectra. ThiswindowisavailableforUBlevel(equipmentsupervisor)users.
(4)InterferogramWindowIt displays interferogram data collected. This window isavailableforUBlevelusers.
(5)Real-timeTrendWindowsDisplaymeasurementvaluesofNos.1to6andNos.7 to 12 components in two separate trend graph windowsforeachstream.Upto10windowscanbeopen at the same time.
(6)HistoricalTrendWindowsDisplayhistoricaltrenddatasaved.Trenddataof24hoursforeachstreamissavedtoafile.Upto4windows can be open at the same time.
(7)ParameterWindowDisplaysthecurrentparametersettingsfortheanalyzer.Inaddition,UBleveluserscanchangethesettings.
(8)AlarmStatus/HistoryWindowsTheAlarmStatuswindowdisplaystheactivealarmsfortheanalyzer,whiletheAlarmHistorywindowdisplays all the past alarms. The alarm history can be deleted with commands.
(9)Tab-controlledMaintenanceWindowDisplaystheA/Dreferencevalueandservo-relateddataoftheanalyzer.ThiswindowisavailableforUClevelusers.
(10) Tab-controlledCommunicationStatusWindow
Displaysthecommunicationstatusbetweenthepersonal computer and the analyzer. This window is availableforUClevel(maintenance)users.
2.7.3 ModelandSuffixCodes
Model SuffixCode
OptionCode Description
NR831 ••••••••••• ••••••••••• SPECTLAND2measurement and maintenance software
Language -E-J
••••••••••••••••••••••
EnglishJapanese
- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”Option /UP Versionup
Packagecontents: OneCD-ROM Oneuser’smanual
2.8 ChemometricsSoftware
2.8.1 Specifications•Calibrationmodelgeneratingtechnique: Partialleastsquare(PLS)andothers
2.8.2 ModelandSuffixCodes
Model SuffixCode
OptionCode Description
NR530 ••••••••••• ••••••••••• Chemometrics softwareLanguage -E ••••••••••• English- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”
Packagecontents: TwoCD-ROMs(Installdisc/User'smanual)
Onesetofuserregistrationdocument
Mar. 18, 2014-00
35<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
2.9 SamplingUnitUseofasamplingunitishighlyrecommendedtoensurecompatibilityoftheuser’sprocesssamplewiththemeasurementcell.Inaddition,itallowsseparationofthemeasurementcell(probe)apartfromtheanalyzerwithafiber-opticcableupto300minlength.Thisenablesselectionofameasurementlocationindependentoftheanalyzer.Theoptimumsamplingunitispreparedforindividualapplicationrequirements.ContactYokogawaforfurther information.
l ExampleofSamplingSystemThefollowingintroducesaconceptualsamplingunitformeasuringtheproperties(RON,RVP,Distillationetc.)ofgasolineblendinginapetroleumrefinery.Notethateachsamplingunitshouldbedesignedforeachapplication;designswillvary.
MeasurementSampleConditionsItem Condition
Fluid to be measured Gasoline
Sample inlet/outletcondition
Inlet pressure 0.3to0.9MPaGOutletpressure 0.3to0.9MPaGInlet temperature 0°Cto+40°C
Notes1: Whensamplesdonotcontainfreewatercontent,acoalescer(CF1)isnotrequired.2: Whensamplesdonothavebubbles,adegasser(DG1)isnotrequired.3: Ifthepressuredifferencebetweenthesampleinletandoutletisgreaterthan0.3MPa,asamplingpump(PU1)isnot
required.
com comcom com
PU1CHV1, 2, 3BV1, 2, 3, 4BV5, 6, 7, 8, 9V1, 2, 3, 4, 5, 6NV1, 2, 3, 4, 5FM1, 2, 3, 4PR1, 2CF1DG1P1, 2DT1TK1, 2, 3, 4CON1RN1
PUMPCHECK VALVE3-WAY BALL VALVE2-WAY BALL VALVESTOP VALVENEEDLE VALVEFLOW METERPRESSURE REGULATORCOALESCER FILTERDEGASSERPRESSURE GAUGEDETECTORTANKInfraSpec NR800RESISTANCE TEMP. SENSOR
: JIS 10K-15A-RF: Rc1/2: Rc1/4
<connection>
: SUS316TP 15A sch40: ø12/ø10 SUS316: ø8/ø6 SUS316: ø4/ø3 Nylon tube: ø6/ø4 Nylon tube: ø22/ø6 Braid tube
<piping>
outlet
Cooling waterinlet
*2PR2
Air inlet
*3
*4
*1PR1
PU1Sample inlet
P1
V1
NV1
FM1NV3NV4
BV1
DG1 BV2
CF1
BV9
Sample outlet
P2
V2
FM4FM3
Cooling water
*3 *4
*1
TK1
V3
V4
CH
V2
DT1 BV3C
HV
1
*2NV5 RN1
CON1
TK2 V5V6
a'
NV2FM2
BV4
Drain
TK4
BV
7
TK3
CH
V3
BV
8
a'
BV
5
BV
6SAMPLING UNIT
N2 gas inlet
N2 gas
Mar. 18, 2014-00
36
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
2.10 AuxiliaryParts*Therecommendedreplacementintervalsforauxiliaryparts,listedinSection2.10.1,ModelandSuffixCodes,aresetforpreventivemaintenanceanddonotguaranteetheservicelifeofparts.Actualreplacementintervalsforauxiliarypartsvarydependingonoperatingconditions,sampleconditions,andotherfactors.Besuretoperformpreventivemaintenancebasedonoperationalstatusofinstrument.Theauxiliarypartsareclassifiedintotwotypes:thosethatcanbereplacedbycustomerandthosethatmustbereplacedbyaYokogawaserviceengineer.Priortoreplacement,refertotheuser’smanualandfollowtheinstructiontherein.* Auxiliarypartsmeanlimitedlifecomponentsthatareexceptedtobeintheirwear-outfailureperiodwithin5yearsunder
normal operating or storage conditions.
2.10.1ModelandSuffixCodes
Model SuffixCode
OptionCode Description
RecommendedReplacement
IntervalNR8SP01 •••••••• ••••••••• Auxiliary parts set for NR800
-01-02-03-04-05-06
-07-08-10-11-12-13-14-15-16-17
••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Sealingpackingforanalyzerdoor(forNR801AG/NR805AG/NR801JG/NR805JG)Analyzercoolingfan(5pcs)Spectrometersupportinsulator(4pcs)Lamp(adjusted)LaserheadDesiccantforspectrometer(9pcs)
Desiccantforflow-throughcell(10pcs)*1Sealingpackingforanalyzerdoor(forNR801EG/NR805EG)O-ringforflow-throughcell(JISP38,FPM(fluorinerubber),2pcs)O-ringforflow-throughcell(JISP21,FPM(fluorinerubber),2pcs)O-ringforflow-throughcell(JISP16,FPM(fluorinerubber),2pcs)O-ringforflow-throughcell(JISP38,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP21,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP16,Kalrez4079,2pcs)Windowpaneofflow-throughcell(borosilicatecrownglass,2pcs)*3Windowpaneofflow-throughcell(sapphire,2pcs)*3
1 year3 years3 years5,000 hours12,000 hours1year(Shelflife:1year after shipment)3 months1 year*1,*2*1,*2*1,*2*1,*2*1,*2*1,*2
*1: Whenameasurementcellisdisassembledforservice/cleaning,itisrecommendedthatO-ringsshouldbereplaced.*2: Replacementintervalsvarydependingonsamplecomposition,humidity,pressure,etc.Determinereplacementintervals
based on actual operational status.*3: Ifawindowisscratchedorfogged,replacethewholesetofwindows(2pieces/set).
3.UtilitySpecifications3.1PowerSupplies
(1)NR800analyzerPowersupply: 100,115,200,230VAC,singlephase,
50/60Hz(*1)Voltagefluctuation: Rating±10%,50/60±2HzPowerconsumption: Approx.250VA(*1): Tobespecified.Seethecorrespondingmodel
andsuffixcodesfordetails.
(2)RS-422/RS-232CconverterPowersupply: 100to120VACor200to240VAC,
singlephase,50/60HzVoltagefluctuation: Rating±10%,50/60±2Hz(Don’texceed
250VAC)Powerconsumption: Approx.15VA
(3)I/OunitACPowersupply: 100to240VAC±10%,singlephase,
50/60Hz±5%Powerconsumption:
NR893AG, NR893JG with enclosure (AOpoints0to12):Approx.150VA (AOpoints16to28):Approx.190VA (AOpoints32to40):Approx.220VANR893EG (AOpoints0to40):Approx.220VANR893JG without enclosure (AOpoints0to40):Approx.100VA
DCPowersupply: 24VDC±10%Powerconsumption: Approx.180mVA/AO4points*: RequiredwhenintheNR893JGI/Ounit,option
code“/C”isnotspecified.
May 15, 2014-01
37<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
3.2 Others
(1)Clean,dryairforAnalyzerpurge(explosion-proofmodel)
Item DescriptionTemperature -10°Cto+40°CPressure 0.3to0.9MPa
Dewpoint-20°Corlower(atpressure)Compressed air should not dew at the installation ambient temperature.
Cleanliness
Mustbefreefromdust,corrosive,andtoxicelements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.
Volume; Approx.75l/min(stp.)
(2)Clean,dryairforin-situprobeItem Description
Temperature -10°Cto+40°CPressure Atmospheric pressure
Dewpoint-20°Corlower(atpressure)Compressed air should not dew at the installation ambient temperature.
Cleanliness
Mustbefreefromdust,corrosive,andtoxicelements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.
Volume;Approx.0.3to0.5l/min(stp.).Dependonsampletemperatureandambient temperature.
(3)Waterforflowthroughcellwithwatertube;Dependonmeasurementspecifications.Temperature,response speed, etc.,
(4)Utilityforsamplingunit(whenused)Dependonsamplingsystemconfiguration,sampletemperature, response speed, etc.
4.InstallationLocationRequirementsAvoidphysicalshockasitmaycausedamagetotheequipment.
4.1 NR800analyzer,measurementcell,andfiber-opticcable
Item Description
Location
Hazardousarea/non-hazardousarea.Outdoor/indoor.Avoiddirectexposuretowind,rain,sunlight and radiation heat.Altitude 2,000 m max.
Ambient temperature -10°Cto+40°CAmbient humidity 5%to95%RH(nocondensation)
VibrationInstalltheequipmentinaplacewithminimumvibration(vibrationaccelerationof2m/s2 or less)
AtmosphereMustnotcontaincorrosiveortoxicsubstances.Must be free from dust.
4.2 RS-422/RS-232CconverterandI/OunitItem Description
Location
Non-hazardous area, indoor.Avoiddirectexposuretowindandrain, sunlight, or radiation heat.Altitude 2,000 m max.
Ambient temperature +5°Cto+35°CAmbient humidity 5%to95%RH(nocondensation)
VibrationInstalltheequipmentinaplacewithminimumvibration(vibrationaccelerationof2m/s2orless)
AtmosphereMustnotcontaincorrosiveortoxicsubstances.Must be free from dust.
4.3MaintenanceSpaceforNR800AnalyzerWheninstallingtheanalyzer,tofacilitateoperationandmaintenance,providespacearounditasfollowingillustration.
561
300 577 200305
1200
1077
(Maintenance space)
50
Unit: mm
*NR800
Analyzer
* JunctionboxforEthernetfiber-opticcables.(WhenNR895/JBisselected.)Seesection2.6.3for external dimensions.
4.4 StorageRequirementsAvoidphysicalshockasitmaycausedamagetotheequipment.
Item Description
LocationIndoor.Avoiddirectexposuretowindandrain, sunlight, or radiation heat
Ambient temperature -10°Cto+50°CAmbient humidity 5%to95%RH(nocondensation)
VibrationInstalltheequipmentinaplacewithminimumvibration(vibrationaccelerationof2m/s2 or less)
AtmosphereMustnotcontaincorrosiveortoxicsubstances.Must be free from dust.
Note: Fluoride, a material used in the NR823 and NR824 fiber-opticcables,hasthedeliquescenceproperties.Whenstrongthesecables,takeappropriatemeasurestorpeventmoisture,e.g.,acableshouldbeplacedwithdesccant in a sealed, moisture-barrier plastic bag and the desiccant replaced regularly.
Mar. 18, 2014-00
38
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
5.SupportforCalibrationModelGeneration
5.1 On-siteGuidanceofCalibrationModelGeneration
AYokogawaengineerwilltrainanuser'ssitepersonnelin the procedure to generate a calibration model for one measureditemusingauser-providedsamplewithitslaboratory analysis results.
5.2 CalibrationModelGenerationYokogawa generates a calibration model using the necessaryquantityofuser-providedsampleswithlaboratoryanalysisresults.ApredefinedSEP(standarderrorofprediction)valueof1σwillbeusedasthemeasurementtargetvalue.Thetargetvalue,samplequantity,andotherdetailsaredeterminedseparatelyfor each application.
5.3 OthersOthersupportoptionsforcalibrationmodelgenerationand maintenance include:•Samplingtestforpotentialusers•Maintenancecontracts•Sampling/modelgeneration/maintenanceconsultingservice.
ContactaYokogawasalesrepresentativeforfurtherinformation,andadviceonthebestsolutionforyourneeds.
6.RecommendedSpecificationsforEngineeringPCandrelatedEquipmentprovidedbyuserthatsatisfiesthefollowingrequirements:
6.1 PC
Computer IBMPC/ATcompatible
Operatingsystem(OS)
MicrosoftWindows7Professional32bitSP1 orMicrosoftWindowsVistaBusinessEdition
CPU 1GHzorhigherRAM 1 GB or more
Harddiskspace 10GB(forprogram)and25GB(fordatastorage) or more
Ethernet adapter 10 BASE-TDisplay SVGAmode(1024x768pixelsormore)Other CD-Rdriveisrecommended
l ColorPrinterPrepareifnecessary.
l Connectioncablesandotherdevicesandconsumables
Prepareifnecessary.
6.2 EthernetCommunicationWhenfiber-opticcablesareusedforEthernetcommunication(seesection2.6)Anoptical-to-electricalsignalconverter(wavelength:850nm)foramultimodefiber-opticcablethatcomplieswithEthernetversion2.0andIEEE802.310Base-Tand10Base-FLstandardsandisfittedwithSTconnectors,fiber-opticextensioncables, and electric cables for Ethernet communication shouldbeprovidedbycustomer.WhenelectriccablesareusedforEthernetcommunication Electric cables for Ethernet communicationshouldbeprovidedbycustomer.
l AdditionalEthernetfiber-opticcableMaterialandconnector: SilicaglassfiberwithST
connectorsMode: Multi-mode GINumber of cores: 2Core/Claddiameters: 50/125µmApplicablewavelength: 850nmLength: 1000morlessintotalincludingan
NR895fiber-opticcable
l SpecificationsofelectriccableforEthernetcommunication
Type: 10Base-T, 8-core shielded, straight connection, RJ-45 connectors
NotethatwhenanNR895fiber-opticcable for Ethernet communication is not used,useanEthernetcrossovercable.Ifyou use a hub, consult Yokogawa.
Length: 3to40mFinished outside diameter: 5.0 to 7.4 mm
May 15, 2014-01
39<<Contents>> <<Index>>
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E
SpecificationSheetsforinquiryofNR800series(1/2)Thankyouforyourinquiry.Pleasefillitcustomer’srequestsinordertomakeaquotation.Tick all which apply.
1. ScopeofQuotationEssential: AnalyzerMeasurementcell/probe,Fiber-optic,Software,User’smanual(1copy)Optional: Document(Specificationforapproval copy Testcertificate copy) Factoryacceptancetest CalibrationModelimplementationbyYokogawa On-sitehands-ontrainingofCalibrationmodelimplementation
2. GeneralUser’sname: Purpose : Control Monitor/alarm OthersPlantname: Document: English Japanese
3. SpecificationsNumberofmeasuringchannel: 1ch 2ch 3ch 4ch PossibilityofthechannelexpansionAnalogueoutput: Need( points) NotneedCommunication: Need NotneedMeasurementcell/probe: No.1ch(cellprobe) No.2ch(cellprobe) No.3ch(cellprobe) No.4ch(cellprobe)Lengthoffiber-optics: No.1ch(Layingdistance: m) No.2ch(Layingdistance: m) No.3ch(Layingdistance: m) No.4ch(Layingdistance: m)
4. InstallationConditionAnalyzer: Indoors Outdoors explosion-proofnon-ex-proof Toxic/Corrosiveambient Restrictedinstallationarea( m x m)Measurementcell/probe: Indoors Outdoors Toxic/Corrosiveambient
5. UtilitiesPowersupply: VAC, HzInstrumentair: (Pressure MPa) NoneNote:Clearanddryairisrequiredforpurgingexplosion-prooftypeanalyzer.Seesection3.2.
6. SampleconditionPleasefillinnextpage.
7. SpecialnotePleasefilloutothernotes.
Mar. 18, 2014-00
40
AllRightsReserved.Copyright©2001,YokogawaElectricCorporation
<<Contents>> <<Index>>
GS 12Y03A03-01E
SpecificationSheetsforinquiryofNR800series(2/2)Samplecomposition/MeasuringConstituentList
Pleasefilloutallsampleconstituentincludingnon-measuringconstituent.No.1chSamplename:
Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Changing/measuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
No.2chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Changing/measuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
No.3chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Changing/measuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
No.4chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Changing/measuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to