Chest
Stand
ControlUnit
Table Centre
Column Center
Spot Film Device
Tube ColumnStand
X- Ray Examination Room
400 cm
450 cm
All wall of the Examination Roomare 23 cm (9”) thick made of bricks.
Single Leaf Doorlined with 2.0 mm lead.
X-RayTube Head
MPB With Lead Glassviewing Window of 2.0mm Lead equivalence
ExaminationTable
UnitElectronics
SALIENT FEATURES
GENUINE X-RAY & RADIOLOGICALEQUIPMENTS PVT. LTD.
Plot No. 3, Radhe Industrial Estate, Block No. 497 (P),Tajpur Road, Sarkhej - Bavla Road, Changodar, Ta. Sanand,Ahmedabad - 382 213. Gujarat, IndiaPhone : +91 - 2717 - 298934
E-mail :
Website :
[email protected]@genuinexray.com
www. genuinexray.com
REVOLUTION IN RADIOLOGY
Autho. Dealer :
All Machine Design / Technical Specifica�ons shown in catalogue are subject to change without any prior no�ce as standard procedure and constant improvement.
CLINICAL OUTPUT
ROOM Dimensions as per AERB Guidelines - L 4.5 x B 4.0 x H 2.75 meters.
CONTROL PANEL
GOVERNMENT OF INDIA
AERB APPROVEDIS:7620
PART-1CM/L - 3950571 REVOLUTION IN RADIOLOGY
Certified & Approved by :
RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS50 kW, 65 kW, 80 kW, 100 kW
HIGH FREQUENCY
Cost effective for all Radiographic applications (Hospitals, Medical Clinics, Chiropractic and Veterinary applications)
Field upgradable for your Future Imaging requirements.
Small, Compact & Light Weight.
Anatomical Programs available in many languages that are easily edited by the Operator Membrane Control Console.
Technique Selection :
1. kV with AEC
2. kV / mAs
3. kV / mA / Time
4. Patient / Body Part Thickness
(Membrane Control Console)
Low kV ripple - maximizes image quality by providing constant potential performance.
Automatic Tube Calibration
Service Software allows fast and easy set up as well as system diagnostics for off-site analysis
Collimator Power Supply
-
HF Generator Series offers advanced high quality Radiographic capabilities with unmatched performance and reliability resulting in the lowest cost of ownership over the product lifetime. This Generator is suitable for Film and CR - Based Radiographic systems and features leading edge micro-processor based controls to minimize patient doze and maximize image quality while demonstrating excellent reproducibility with user-friendly operator controls.
GXL-HF 50 / GXL-HF 65 / GXL-HF 80 / GXL-HF 100
LED MONITOR
TECHNOLOGY
-15 Degree
RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS
45 Degree 90 Degree
HIGH FREQUENCY RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS
POSITIONS
All R & F Systems come with Floor to Ceiling Column Tube Stand fully counter balanced.
Optional Ceiling Free Column Tube Stand available on demand. Control Unit Control Panel
X - Ray Tube
H.V. Cable
OPTIONALX- RAY GENERATOR
Vertical Bucky
GENERAL SPECIFICATION 100 kW
100 kW
100 kW
80 kW
80 kW
80 kW
65 kW
65 kW
65 kW
50 kW
50 kW
50 kW
Generator Type
Rotor Supply
Input Phase / Voltage
Power De-Ra�ng
Compa�ble X-Ray Tubes
Anatomical Programs (APR)
Image Receptors
Technique Selec�on
Communica�on Ports
Auxiliary Collimator Power Supply
GenWare Service So�ware
RADIOGRAPHY
kVp Range
kVp Accuracy
High Voltage Ripple
Rise Time (10% - 90%)
mA Range / Steps (1 mA / 0.1 mA
Steps op�onal)
mA Accuracy
Exposure Timer Range
Exposure Timer Accuracy
(measured at 75% points of kV waveform)
mAs Range (non-AEC) / Steps
mAs Accuracy
FLUOROSCOPY
kVp Range
kVp Accuracy
High Voltage Ripple
mA Range / Steps
mA Accuracy
High Frequency Output (Maximum 400 kHz)
Low Speed Starter (Op�onal Dual Speed Starter)
Ф3 / 400 - 480 VAC
300 tube models
1024 / 20,000 + techniques
Up to 6
kV / AEC, kV / mAs, kV / mA / ms or kV / cm thickness (requires standard console)
USB, CAN, Ethernet, RS485, RS232, RS422
Standard Feature
WEB - based diagnos�c and technical support
40 - 150 kV in 1 kV increments