image quality

39
Part II Image Quality 1

Upload: triage

Post on 23-Jan-2016

36 views

Category:

Documents


0 download

DESCRIPTION

Image Quality. Part II. Image Quality. Image Quality Photographic Quality Geometric Quality. The films or images have different levels of density – different shades of gray X-rays show different features of the body in various shades of gray. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Image Quality

Part II

Image Quality

1

Page 2: Image Quality

Image Quality

Image Quality Photographic Quality Geometric Quality

2

Page 3: Image Quality

3

Why you see what you see…

The films or images have different levels of density – different shades of gray

X-rays show different features of the body in various shades of gray.

The gray is darkest in those areas that do not absorb X-rays well – and allow it to pass through

The images are lighter in dense areas (like bones) that absorb more of the X-rays.

Page 4: Image Quality

Terminology Density, contrast Detail Distortion Image Sharpness spatial resolution Lp/mm Blur Motion SID OID Focal spot size

Image receptor Magnification Penumbra Size distortion Shape distortion Focal Spot size

Page 5: Image Quality

Image quality5

Page 6: Image Quality

Resolution, ability to distinguish shapes, ability

to distinguish lines on image

detail

6

Page 7: Image Quality

RECORDED DETAIL

The degree of sharpness in an object’s borders and structural details.

How “clear” the object looks on the radiograph

7

Page 8: Image Quality

Recorded detail Image sharpness Spatial resolution Smallest separation

of two lines or edges Measured by lp/mm

8

Page 9: Image Quality

RESOLUTION TEST

TOOLS

LINE PAIRS/ MM

Depicts how well you can see the differences in structures

More lines=more detail

9

Page 10: Image Quality

10

Page 11: Image Quality

11

Page 12: Image Quality

Recorded Detail

SID Focal spot size OID Image receptor type Motion

Factors that affect detail

Page 13: Image Quality

Focal Spot Geometry

Angle of anodesize

Page 14: Image Quality

Focal Spot Geometry

Focal spot is notreally a single point

Dependent on filament size and anode angle.

Page 15: Image Quality

SID

Shine a flashlight on a 3-D object, shadow borders will appear “fuzzy”

-On a radiograph called penumbra

Penumbra (fuzziness) obscures true border

Farther the flashlight from object = sharper borders. Same with radiography.

15

Page 16: Image Quality

Recorded Detail: Penumbra and SID

Page 17: Image Quality

Recorded Detail: Penumbra and Focal Spot Size

Page 18: Image Quality

Recorded Detail: Penumbra and OID

Page 19: Image Quality

19

The position of the structure in the body will influence how magnified it will be seen on the image

The farther away – the more magnified

Page 20: Image Quality

20

Recorded Detail: OID and Penumbra

The closer the object to the film, the sharper the detail.

OID , penumbra , sharpness OID , penumbra , sharpness

Structures located deep in the body, radiographer must know how to position to get the object closest to the film.

Page 21: Image Quality

Recorded Detail: Image Receptor Type

Film/Screen Imagingexcellent spatial resolution-smallest

detail that can be detected in an imageComputed Radiography (cassettes)Digital Radiography (cassette-less)

improved contrast resolution

Page 22: Image Quality

22

Motion Can be voluntary or involuntary

Best controlled by short exposure times

Use of careful instructions to the patient

Suspension of patient respiration

Immobilization devices

Page 23: Image Quality

23

Blurring of image due to patient movement during exposure.

Page 24: Image Quality

Misrepresentation of size or shape of anatomic part; when part is distorted,

detail is reduced

distortion

24

Page 25: Image Quality

Distortion

An increase of decrease in the size of an object : magnification or reductionThree types: size, shape, placement of part in body

Page 26: Image Quality

Types of Distortion

SID-size distortion OID-size distortion Beam Angulation-

shape distortion Body Part-Beam

alignment-shape distortion

FACTOR INFLUENCING DISTORTION

Page 27: Image Quality

Distortion: SID

Page 28: Image Quality

40” SID VS 72” SID Which one is which?

28

Page 29: Image Quality

29

Page 30: Image Quality

Distortion: OID

Page 31: Image Quality

31

Page 32: Image Quality
Page 33: Image Quality
Page 34: Image Quality
Page 35: Image Quality
Page 36: Image Quality
Page 37: Image Quality

Elongation Foreshortened Normal

37

Page 38: Image Quality

38

Page 39: Image Quality