Updated: Master Manual of Procedures for Human Testing Center Institute of Applied Life Sciences Life Science Laboratories S360 240 Thatcher Road, Amherst, MA 01003 at the University of Massachusetts, Amherst

Upload: lamdiep

Post on 20-May-2018

217 views

Category:

Documents


1 download

TRANSCRIPT

Updated:

MasterManualofProceduresfor

HumanTestingCenter

InstituteofAppliedLifeSciences

LifeScienceLaboratoriesS360240ThatcherRoad,

Amherst,MA01003

atthe

UniversityofMassachusetts,Amherst

Updated:16October2017 ii

HumanTestingCenter

Chapter1:CoreAccess,Training,Scheduling,PricingandBilling

ExerciseInterventionandOutcomesCore

Chapter1:ParvoMedicsTrueOne2400MetabolicCart

Chapter2:VelotronBicycleErgometer

Chapter3:Monark828EBicycleErgometer

Chapter4:Monark874EBicycleErgometer

Chapter5:BiodexSystem4ProDynomometer

Chapter6:QuintonQ-Stressw/TM55Treadmill

Chapter7:GECarescapePatientMonitors

Chapter8:GELunariDXAa

LivingScienceCore

Chapter1:OxyconMobileMetabolicSystem

Chapter2:ActigraphLinkActivityMonitor

Chapter3:ActigraphGT3X+BTLEActivityMonitor

Chapter4:Actilife&ActilifeLiteSoftware

HumanMotionCore

Chapter1:QualysisOQUS700+/500+w/AMTIForcePlatformIntegration

Chapter1.1–CreatinganewprojectinQTM

Chapter1.2–QTMDataCollection

Chapter1.3-QTMDataProcessing

Chapter1.4–ExampleMarkersets

Chapter1.5–Visual3DTutorial

Chapter2:DelsysTrignoEMG/IMUunits–MotionCaptureIntegration

RoomCalorimeterCore

Chapter1:QuickGuide forValidation/Calibration

Updated:16October2017 HumanTestingCenter 1

MANUALOFPROCEDURES

HumanTestingCenter

InstituteofAppliedLifeSciences

LifeScienceLaboratoriesS360240ThatcherRoad,

Amherst,MA01003

atthe

UniversityofMassachusetts,Amherst

Updated:16October2017 HumanTestingCenter 2

UNIVERSITYOFMASSACHUSETTSAMHERST

INSTITUTEFORAPPLIEDLIFESCIENCES

HumanTestingCenter

CoreAccess,Training,Scheduling,Pricing,andBilling

MostrelevantinformationforgainingaccesstotheCoreswithintheHumanTestingCentercanbefoundontheCorewebsites.Accesswillonlybegiventousersonceallstepshavebeencompleted.

a) CoreAccess:

InternalUser.

IfyouareaUMassAmherst,UMasssystem,orFiveCollegebasedresearcherwhoisdoingresearchthatwouldreceive‘internal’pricing.ThefollowingstepsoutlinehowyouwillgainaccesstotheCoreFacilities.

i) ContactMichaelBusatheCoreFacilityDirectoranddiscussyourprojectsneedsandtimelines([email protected],413-755-0574).

ii) IfyouarenewtoIALSaMOUandpaymentauthorizationformmustbecompletedforeachcorefacilityyouwouldliketouse.ThePI’sStartUppaperworkcanbefoundintheFormstabofthewebpage.

ExternalUser

Ifyouareanexternaluserfromanacademicorindustrypartner,pleasecontactMichaelBusa,theCoreDirector,todiscussyourneeds([email protected],413-577-0574).

b) Training:IndividualswhowouldliketouseanypieceofequipmentintheHumanTestingCentermustcompleteatrainingsessionwithCorestaff(CoreRatesApply).ThesetrainingsserveasbothatechnicaltutorialforequipmentuseandanorientationforuserstotheHumanTestingCenter.Foruserswhoalreadyhaveexpertiseintheuseofaparticularpieceofequipment,theycanrequestthattheyjustdemonstrateproficiencyinuseandskipthetutorial.PleasenotethateventhoughausermaybeproficientinthetechnicalaspectsofacertainpieceofequipmentthattheintegrationofthebillinganddatastoragesystemsintheIALScoresdiffersfromstandalonepiecesofequipment.Beingtrainedontheseaspectswillensurethatdataisproperlystoredandbillingrecordsareaccurate.

Aspartofthetraining,theprospectiveusermustalsoprovidetheCoreStaffwithproofoftherequireEHStrainings.AllrequiretrainingsarelocatedatthebeginningofeachinstrumentsStandardOperatingProcedure.

c) Scheduling:AllschedulingofequipmenttimeistobedonethroughtheFacilitiesOnline

Managementsystem(FOM).i) Supervisors(i.e.PIs)musthavesupervisoraccountscreatedpriortohavingtheirRA’screate

profiles,theCoreDirectorwilldothisforthemfollowinganinitialemailorphonecallindicatingtheirintentiontouseacore.

ii) AllusersmustcreatetheirownFOMaccountandlinkittotheirsupervisor.

Updated:16October2017 HumanTestingCenter 3

(1) Theymustthenclickontheindividualcapacitythattheywanttogainaccessto,andrequestatrainingtimewiththeCoreStaff.

(2) Duringregistration,theymustusethespeedtypethatmatchesthePaymentAuthorizationFormthatissubmittedaspartofthePI’sStartUppaperwork.Additionalspeedtypescanbeaddedbyusersinthecasewhentheyworkonmultipleprojects.

d) Pricing/Usage:PricingforfacilityusearepostedontheCorewebsites.i) Ratesaresetonanannualbasisandaresubjecttochange.ii) PricesaresetinFOM.ChargesforusearebasedonFOMactivetimeroundeduptothe

nextcompleteschedulingperiod.

e) Billing:BillingiscarriedoutbytheIALSaccountingstaff.i) Statementsaresentouttosupervisorsandtheirdesignatedrepresentative.ii) Disputesorchangesinspeedtypetotheinvoicemustbemadewithin10businessdays,

otherwisetheselectedspeedtypewillbeautomaticallychanged.

Updated:16October2017 ExerciseIntervention&OutcomesCore 1

ManualofProcedures

ExerciseIntervention&OutcomesCore

InstituteofAppliedLifeSciences

LifeScienceLaboratoriesS360240ThatcherRoad,

Amherst,MA01003

atthe

UniversityofMassachusetts,Amherst

Updated:16October2017 ExerciseIntervention&OutcomesCore 2

UNIVERSITYOFMASSACHUSETTSAMHERST

INSTITUTEFORAPPLIEDLIFESCIENCE

ExerciseInterventionandOutcomesCore

Chapter1:ParvoMedicsTrueOne2400

1.1Introduction:TheTrueOne®2400MetabolicMeasurementSystem(ParvoMedics,INC.,Sandy,UT,USA)sanintegratedmetabolicmeasurementsystemformaximalO2consumptiontestingandindirectcalorimetryassessment.TheTrueOne®2400'sAnalyzerModule,withparamagneticoxygenandinfra-redcarbondioxideanalyzers,isdesignedtobeaccurate,reliable,andeasytouse.Theuniqueflowmetercalibrationalgorithm,utilizinganimagereconstructiontechnique,correctsthenon-linearityofthepneumotachandprovideshighlyaccurateflowmeasurement.

Features:

•HighPrecisionFlowmeterCalibration

•CompactAnalyzerModule

•HighlyReliableMixingChamberPrincipleRecommendedApplications

•AthleticTesting

•ExercisePhysiologyTeachingandResearch

•SportsMedicineResearch

•EnergyExpenditureAssessment

•DifferentialDiagnosisinCardiopulmonaryDiseases

Updated:16October2017 ExerciseIntervention&OutcomesCore 3

1.2HowtoCalibrate:Belowisastep-by-stepoutlineforcalibratingtheunit.

PreliminaryCalibrationSteps1. LocatethepowerswitchontherearoftheAnalyzerModuleandturniton.

a. Allowit30minutestowarmup

2. TurnonthecomputerandmonitorandlaunchTrueOnesoftware

a. Theydonotrequireawarmuptimeofmorethan2-minutes3. Assemblethewater-trapconnection

a. ConnectthefiltertotheT-connectorusingthegreyconnectionpiece.Gentlypushthetwotogetherinacircularmotion.

Updated:16October2017 ExerciseIntervention&OutcomesCore 4

4. Attachthisassemblytothepneumotach.Attachbypushingthewatertraptothepneumotachwithslightpressure.Ensurethatthemixingchamber(tintedrectangularbox)doesnotfalloffthecart.

5. Ontheotherend(labeledtube)attachedoneendofthehose.

Updated:16October2017 ExerciseIntervention&OutcomesCore 5

6. Next,assembletheT-mouthpiece3-Lsyringecalibrationassembly.Makesurethesmallwhitemembranesarefacingthesamedirection.Lookingatimage#3below,whenholdingtheunitwithupright,thesideofthebodywiththeair-flowarrowfacingup(smallarrowonthecenterbodypiece)themembraneonthissideshouldhavesurfaceparallelwiththelongwaterchamber.Themembraneontheoppositesideshouldbefacingthesameway.Ifyouarehavingtroubleusethisresource:https://www.youtube.com/watch?v=YSg65afxMCM

FINALPRODUCTforT-mouthpiece3-Lsyringecalibrationassembly

Updated:16October2017 ExerciseIntervention&OutcomesCore 6

7. Taketheotherendofthehose(endnotattachedtothewatertrap)andconnectittotheT-mouthpiece3-Lsyringecalibrationassembly.Makesurethehoseisattachedtotheoutputvalve(endwith3metalrings).

8. Next,grabthe3-Lsyringe,ensuringithasthegreyconnectorattached

Updated:16October2017 ExerciseIntervention&OutcomesCore 7

9. InterfacetheT-mouthpiece3-Lsyringecalibrationassemblywiththegreyendofthe3-Lsyringe.Makesurethattheconnectionistight,howeverdonotforcefullypushthemtogether.Rather,gentlymakesmallcircularmotionstoconnectthem.

10. Next,prepareforgascalibrationbyopeningthegascylinder.Turnthemetalknobonthetopofthe

cylindercounterclockwise1and½turns.TheCalGascalibrationiskeytoaccuratetestingresults.Youcanrunthegascalibrationasmanytimesasneededduringcalibration.Itisrecommendedtorunthecalgascalibrationonelasttimejustbefore(getsonthetreadmillorergometer)testingthesubject.WeneedtoensuretheCO2levelstheParvocartisreadingforroomairhavenotfloatedabove3-4%.Ifthishappens,theRERvalueinthetestwindowwithbeinflatedwhentheteststarts.

Updated:16October2017 ExerciseIntervention&OutcomesCore 8

11. Nowyoucanattachthecalibrationgashose/whiteconnectorfromthecylindertotheCalGasPortonthemetaboliccart.Thisislocatedonthebackoftheunit.ItistotheleftoftheRoomAirportandjustabovetheO2Rangeswitch.Youmustholdthefrontoftheunit,soitdoesnotfalltothefloor,whilebringingthecalibrationgashose/whiteconnectortotheCalGasport.Stayparalleltothefloorandpushtheconnectorstraightintheportconnector.Toremove,pushdownlightlyonthemetalclaspandpullthecalibrationgashose/whiteconnectoraway.

12. Now,touchpneumotach,itshouldbewarm,indicatingitisshouldbewarmedup.Ensurethat30-

minuteshaspassed,andcontinewithcalibration

13. Thesystemisnowreadyforthethreestepcalibration.Oncethesystemhasbeensetup,thecalibrationprocessisveryquickandeasy.Thesystemshouldbecalibratedagaininbetweenmultiplesubjectsorifithasbeensittingturnedonbutnotusedforalongperiodoftime.Theonlycalibrationsteptheoreticallyneededafterinitialcalibrationistoapplythecalibrationgasagain.However,theenemyofthepneumotachismoisture.Excessmoisturewillelectricallydistortwhatthepneumotachissensing;thiswillbereflectedinerroneousreadingofthe3litersyringeas9,12,15litersetc...Thescreeninsidetheroundovenneedstobeasdryaspossible,soyoumayhavetousethesyringetoinputroomairacrossthesensorandintothemixingchamberifyouaretestingmultiplesubjects.ThepneumotachisverysensitivetosubtlechangesinCO2.Thisisthereasonwhyweneedtoletthepneumotachheatuptooperatingtemperature.DoingatestwithacoldpneumotachwillhavedistortedCO2results–theRERandVO2kgreadingswillbewayoff.

Updated:16October2017 ExerciseIntervention&OutcomesCore 9

1.3FLOWMETERCALIBRATION

1. OpenTrueOnesoftwareandclick“FlowmeterCalibration”2. Enterroomtemperature,relativehumidity,andbarometricpressurereadingsfromtheVantage

VUEintodialogueboxonTrueOne

3. ClickandopenSampleBaselineintheupperrightcornerofthedialoguebox4. TheFlowmeterCalibrationProcessingwindowwillopen.Allinputstrokeswillbedonewiththe

3-Lsyringe.

5. Detectionstroke-oneinputstrokeindicatesairisflowingintothesystem(mouthpieceandallassembliesareproperlyconfigured).*NoticetheFlowindicatorwillshowtherealtimeinput

Updated:16October2017 ExerciseIntervention&OutcomesCore 10

velocity,StrokePeakwithshowthepeakinrealtimeandintheblankbluefieldbesidetheCancelbutton,thevolumeofthesyringewillbedisplayedattheendofanystroke.Thisvalueshouldbearound3liters.Ifitreadshigh–9,12,15litersetc..thepneumotachhasamoistureissue(iftestingmultiplesubjectsoverashortperiodoftime)andmustbeallowedtodry.YoucanusethesyringesetuptodotheFlowmetercalibration,butdon'topenthesoftware.Youarejustphysicallydrivingairacrossthepneumotach,throughthemixingchamberandintothesystem

6. FlushStroke–fourinputstrokestopushfreshairintothesystem7. Thepneumotachcansense800lpm.Ofthefivecalibrationstrokesinputnext,twomustbeless

than80lpm.ThenumberwillfluctuatebesidetheFlownumericalindicator(totheleftofStrokePeak.ForExerciseStrokevelocityisveryveryslow,veryslow,slow,fast,faster.

8. Stroke1→approximately50-80lpm(very,veryslow)9. Stroke2→approximately100+lpm(veryslow)10. Stroke3→approximately200+lpm(slow)11. Stroke4→approximately300+lpm(fast)12. Stroke5→approximately400+lpm(faster)13. ForRestingStrokevelocitycanstayat50-80lpm14. Theresultissomethinglikethiscurve

Updated:16October2017 ExerciseIntervention&OutcomesCore 11

1.4GASCALIBRATION

1. Onthehomescreen,click“GasCalibration”2. TheGasCalibrationwindowwillopen,likewithFlowmetercalibration,correctlyinput

temperature,humidity,andpressureintothedialoguebox.ClickOK.3. Theunitwillperformgascalibrationandafterabout20seconds.4. Apop-upwillappearpromptingyoutoturnoffthecalibrationgas.5. Asummarywindow(similartotheonepicturedbelow)willarise.Itisrecommendedtodothe

CalGascalibrationuntilyougettheexactnumbersonthecylinder.Also,youmaydothisstepjustbeforeyoursubjectisreadyfortesting

6. Click“Save”.7. Whendone,turnoffthecalibrationgascylinderbyturningclockwisehandtight.Thenbleedof

theexcessleftintheregulatorandthecalibrationgashose/whiteconnectorbygentlypushingontheplasticconnector/openingattheend.Bleedoffuntilbothindicatorsreadzero

Updated:16October2017 ExerciseIntervention&OutcomesCore 12

1.5TESTINGwiththeParvo2400:1. Fitparticipantwiththeheartratemonitor,theheartratetransmitterislocatedonthereticulated

arm.Besuretowetthestraporapplyelectrodegelbeforethemonitorisused.2. Ifyouareusingareusingtheheadsupportsystem,attachedsweatguardtothefrontofthesystem.

Updated:16October2017 ExerciseIntervention&OutcomesCore 13

3. Whiletheparticipantisseated,fittheheadsupportsothefrontportionisjustabovetheeyebrows.

Ensureasnugfit.Tightentheentirebandaroundtheheadusingtheknobontheback.Youwillbeusingamouthpieceandnoseclipwiththissetup.Attacharubbermouthadaptertothesmallwhitecylinderofthemouthpieceyoumadeearlier.Youwanttoavoidthesubjectbreathingoutofthecorneroftheirmouth(whenatmaximalexertion).Wewantatightsealregardlessofwhatdeviceisused-headsupportorfullmask.

4. Givethesubjectanosecliptoputon5. Theyarereadyforthetreadmillorergometer.Attachthehosetotheoutputendofthe

mouthpiece.Donotletthesubjectinserttheadaptoryet.Whenitistimetostartthetest,thesubjectcanjustspintheadaptortowardtheirface,placetheadaptorintheirmouthandbegin

Updated:16October2017 ExerciseIntervention&OutcomesCore 14

mouthbreathingintothevalves.Wemustwaituntiltheappropriatetestwindowopensforthesubjecttodothis.Theyarenottobreatheintothehoseatthistime.

6. Thefullfacemaskdoesnotrequireanosecliptobeused.Tousethispiece,switchoutthesmall

whitecylinderforthesmallerclearadapter(picturedbelow).Threadtheendofthesmallclearadapterintothefrontholeofthebluefacemask.Facialhairmayinterferewithatightseal.Donothookupthehosetothefullfacemaskuntiltheteststarts.Thestrapswraparoundtheheadandthentheblackconnectorslockintotheclearbracketsofthemask.ThestrapsareadjustablebyslidingthestrapthroughtheblackconnectortheusingVelcrotokeepinplace.Themaskcoversthemouthandnose–sothestrapsneedtobetighttoensureagoodseal

Updated:16October2017 ExerciseIntervention&OutcomesCore 15

1.6METABOLICTESTING1. Onthehomescreen,clickonVO2/MetabolicTesting.2. Inputparticipantinformation.Updatetemperature,humidityandbarometricpressureifithas

changed.Choosewhichdeviceyouareusing.

3. Ifyouusethetreadmillyouneedtochooseaprotocol.Youcancreateanewoneandnameitwhateveryouwant.Or,modifyanexistingoneandsaveit.Usewhatistherealreadyfromthepulldownmenu.OrmodifyaprotocolandnotsaveitbyjustclickingOK.

4. Treadmillprotocolscanhavearesting,warmup,exerciseandcooldownphase.Ifyoudonotwant

restingorwarmupphases,maketheduration=0.

Updated:16October2017 ExerciseIntervention&OutcomesCore 16

5. ClickOK6. Ifyouusetheergometeryoualsoneedtochooseaprotocol.Youcancreateanewoneandnameit

whateveryouwant.Modifyanexistingoneandsaveit.Usewhatistherealreadyfromthepulldownmenu.OrmodifyaprotocolandnotsaveitbyjustclickingOK.

7. YouhavesimilaroptionswhenusingabikechooseaprotocolandpressOK.8. Nowthetestingwindowswillappear.Ifusingtheheadsupport,youcanattachthehosefromthe

reticulatedarm/T-connectortotheoutputconnectoronthemouthpiece.Thenoseclipcanbeworn.Donotletthesubjectbreathintothemouthpieceyet.Ifusingthefullfacemask,donotattachthehosefromthereticulatedarm/T-connectortothemaskyet.

9. Aseriesofsoftwarewindowswillopenandclose.10. Thenthiswindowwillopen.Youwillseetheblanktestresultswindowopenaswell.Thefieldswill

beblankbecausethesubjectisnotbreathingintothesystemyet.WaituntiltheCO2levelsexceeds0.03%,thenclickOK.

11. Ifusingtheheadgear,thesubjectcanspinthemouthpiecetowardtheirface.Theadaptorcanbe

placedinthemouthwhilethenoseclipison.Ifusingthefullfacemask,thehosefromthereticulatedarm/Tcannowbeattachedtotheinputconnectoronthefullfacemask.

12. Therewillbeadelayofapproximately15–20secondsbeforetheMetabolicMeasurementTesting

windowupdateswithdata.Thewhitespacebelowwillbefilledwithdataevery5seconds(thisresolutionlevelcanbechanged)

Updated:16October2017 ExerciseIntervention&OutcomesCore 17

13. TheEventsbuttontotherightcanbeusedtoenterRPEvalues,increasedwattage,etc...apop-upwindowwillappear.SimplyentertheRPEvaluetotheemptydatafieldinthepop-up.Thisneweventwillbeaddedtothedatalinesinthewhitespacebelow

14. Welldocumentedreportsareavailableaftertesting.TotherightoftheCancelbuttonisasetoftimers.Thiswillshowoveralltesttime,restphasetimeleft,warm-upphasetime,andexercisephasetimeleftasyougothroughthestagesoftheprotocolyouareusing.Afterthetestyoucanviewdataunderthe“Reports”tabonthehomescreen.

Updated:16October2017 ExerciseIntervention&OutcomesCore 18

1.7CLEANINGEQUIPTMENT1.) Whilewearinggloves,removeheadgearorthefullfacemaskfromthesubject.Rememberto

keepthesubjectattachedtothecartforabout20secondsafterthetesthasfinished.Thisallowssoftwaredelaytofillintheremainingdata

2.) Takeallcomponentsoffandbringtogreycart/sinkarea3.) Detachmouthpiecefromheadgear4.) Rinsethesweatguardunderwater.Theheartratestrapandtransmittercanberinsedaswell.

Letthesepartsairdryonthepeggedrack5.) DisassemblethemouthpieceandrinseallcomponentsinsinkwithJohnsonbabysoap

6.) Rinsein______________solution

1.8Contacts1. ParvoMedics

8152South1715EastSandy,UT84093,USA

1800-942-72552. MichaelBusa–CoreDirector

[email protected]

Updated:16October2017 ExerciseIntervention&OutcomesCore 19

Chapter3:Monark828E,weightergometer3.1Introduction

MonarkErgometermodel828Eistheworld’smostcommonlyusedexerciseergometer.TheadjustablebrakesystemandtheforcecanbesetandreadKiloponds(kp)andNewton(N).Theresultoftheresistanceandcadenceisdisplayedinpower(Watts)ontheelectronicmeter.Thecycleisequippedwithanelectronicmetershowingpedalrevolutionsperminute(RPM),heartrateinbpm(HR),exercisetimeinminutesandseconds(TIME),speedinkmperhourormilesperhour(SPEED),covereddistanceinkmormile(DISTANCE),burnedCalories(CAL)andthepoweronthecycle(WATT).Thewattleveldependsonpedallingspeed,itcanbefinetunedbyincreasingordecreasingthespeedorpedalrpm.Each828Eiscalibratedatthefactory.ThismeansthatyoucanbegintousetheErgometerdirectlyafterassembly.However,iftheuserwishestoverifythescale,pleasereadtheinstructionfor“Calibration”inthismanual.

Quickfactsaboutthe874E:

•Large,well-balancedflywheel20kg(44lbs)

•Pendulumscale,easytocalibrate

•Adjustableseatheight

•Adjustablehandlebarwithquickreleaselever

•Stableframe,solidsteeltube

•Electronicdisplaywithheartrate

Corecompetencies:

1) Understandingthebrakedeviceandbasicsofthetensioncenter2) BefamiliarwiththeFitnesscomputerandwhatmetricsitgenerates3) Reviewserviceactionsandunderstandwhatactionsyoumightneedtotakebefore/afterusing

thisequipment.3.2UserInstructions

CycleadjustmentsSeatheightshouldbeadjustedtoacomfortableposition.Theappropriateheightistohavethekneeslightlybentwhenthesoleofthefootiscenteredoverthepedalaxlewiththepedalinthebottomposition.Toadjusttheseatheight,loosenthelever(1)ontheseattube.SeeFig:Adjustments.Alsonotethatthehandlebarstemshouldbeinsertedintotheframetubeatleast3inches.Thismeasureismarkedwith“MAX”onthestem(3).

Updated:16October2017 ExerciseIntervention&OutcomesCore 20

FitnesscomputerTheMonark874EisequippedwithaFitnesscomputerthatdisplaysthefollowingmetrics:

1. Revolutionsperminute(RPM)2. Heartrateinbpm(HR)

a. Atthedisplayforheartrate(HR)aheartsymbolislitwhichmeansthatthemeteristryingtofindapulsesignalfromanexternalsource,chestbeltwithelectrodes,IfthemetercannotfindsuchasignaltheHRfunctionisautomaticallyturnedoffafter30seconds.Whenthefunctionisturnedofftheheartsymbolisnotlitanymore.Theheartratefunctioncanbeturnedonagainbypressingakey.

3. Exercisebouttimeinminutesandseconds(TIME)a. Thetimerstartsautomaticallywhenpedalsaremoved.MetervaluesforTime,Distanceand

CaloriescanbesettozerobypressingtheRESETbutton(seeimagebelowforRESTbuttonlocation)formorethantwoseconds.

4. Cyclingspeedinkmperhourormilesperhour(SPEED)5. Covereddistanceinkmormiles(DISTANCE)6. Caloriesburned(CAL)andpower(WATT)arealsodisplayedonthecomputer.Togetcorrect

readingsforcaloriesandwatts,thekp-valueontheelectronicmeter(Fitnesscomputer)hastobesettothesamevalueasthependulum,orthekpwindowshowntotheleftoftheelectronicmeter.

a. Forexample,thependulumandthekp-windowisshowing2kp.Pressthekp-keytotheleftonthemeter.Thelowerdisplaywindowisnowshowingfiguresinkp.Increaseordecreaseinstepsof0.1kpbypressingthekp-key-(arrowup)ortheRESET-key-(arrowdown)untilthereadingcorrespondswiththeactualordesiredkp-valuesonthependulumscaleorinthekp-window.AfterthatpresstheCAL/WATTbuttontoeithershowtheCAL-orWATT-figures.Thewattreadinginthedisplayisdependingonthepedallingspeed.Thewattreadingcanthenbefinetunedthroughincreaseordecreaseofthepedallingspeed.Caloriesarecalculatedallthetime.

Comment[MB1]:Makeavideo?

Comment[GP2R1]:Made.WaitingonYouTubeaccount.

Updated:16October2017 ExerciseIntervention&OutcomesCore 21

Toactivateanyofthefunctionsdescribedabove,pressanykeyormovepedals.

Toreset,holdtheRESETkeyfortwoseconds.

BelowaresampleandimagesoftheFitnesscomputerandatableoutliningmetricsthecomputerdisplayswithcorrespondingunitsandranges.

Updated:16October2017 ExerciseIntervention&OutcomesCore 22

Scale-zeroadjustmentLoosentensiondevicesothatthebrakebeltfeelsloose.Checkthatthependulumwillhanginverticalposition.

Scaleboard:Checkthattheindexonthependulum(2)weighisalignedwiththeindexatthe0-positiononthescaleboard.Ifadjustmentisnecessary,firstloosenthelocknut(1)andthenchangethepositionoftheboard.Tightenthelock-nutaftertheadjustment.Seefig:Scaleadjustments.

Kp-scale:Atthesametime,checkthatthekilopond-scale(5)totheleftoftheFitnesscomputer,shows0inlinewiththeindex.Toadjustthekp-scaleloosenthelockscrewforscaleindicator(9).Tightenthescrewfirmlyaftertheadjustment.Seefig:Calibration,Adjustmentkp-scalewindow.

Updated:16October2017 ExerciseIntervention&OutcomesCore 23

CalibrationAlthoughallErgometersarecalibratedatthefactory,theusermaywishtoverifythisbyperformingamechanicalscalecalibration.Tocalibratefollowthestepsbelow:

1. Removethebluebrakebeltfromthespring2. Leanthebikeforwardsothatthecalibrationweight,4kg(4)ishangingindependentlyfromthe

framecover3. Besurethatthescaleboardissettozerobeforetheweightisputon,seesection”Scale-zero

adjustment”4. Putthe4kgweightonthespring(6).Whencorrectlyset,itshouldbepossibletoreadthisweight

fromthecorrespondingplaceonthescaleboard(7).Seefig:Calibration

Shouldtherebeadeviationfromthecalibrationweight,adjustthependulumtothecorrectweightonthescalebymeansoftheadjustingweightinsidethependulum(3).Seefig:Scaleadjustment.Tochangethepositionoftheadjustingweight,loosenthelockscrew(8)onthebackofthependulumweight.Shouldtheindexofthependulumweightbetoolow,movetheadjustingweightupwardsintheweight.Iftheindexshouldbetoohigh,theadjustingweightismovedsomewhatdownwardsandlockedinthenewposition.Repeatuntilthecorrectreading(4kg)isachieved.

Checkthecalibrationofthependulumweightonceayearorwhenneeded

Updated:16October2017 ExerciseIntervention&OutcomesCore 24

SampleDataSheetforGradedExerciseTest

TIME DISTANCE SPEED HR RPM kP/meter/minàWatts(1kP/m/min=0.1634watts)

1:00

2:00

3:00

4:00

5:00

6:00

7:00

Updated:16October2017 ExerciseIntervention&OutcomesCore 25

3.3Generaltroubleshootingguide

MaintenanceSchedule:

Serviceaction:

• Manufacturersuggestusingisopropylalcohol.Useadampbut,notwetclothtocleanthesurfaceyouwishtodisinfect.

• Periodicallywipethesurfacewitharustpreventative,especiallywhenithasbeencleanedandthesurfaceisdry.Thisisdonetoprotectthechromeandzincpartsaswellasthepaintedparts(4timesperyear).

• Checknowandthenthatbothpedalsarefirmlytightened.Ifnot,thethreadinginthepedalarmswillbedamaged.Also,checkthatpedalarmsarefirmlytightenedonthecrankaxle,tightenifnecessary.WhentheErgometerisnewitisimportanttotightenthepedalsafter5hoursofpedalling(4timesperyear).

• Checkthatthepedalcrankissecuretothecrankaxle(4timesperyear).• Besurethatthepedalsaremovingsmoothly,andthatpedalaxleisclearofdirtandfibres(4

timesperyear).• Whencleaningandlubricatingbesuretocheckthatallscrewsandnutsareproperlytightened

(2timesperyear).• Checkthatthechainissnugandthereisnoplayinthepedalcrank(2timesperyear).• Checkthatpedals,chainandfreewheelsprocketarelubricated(2timesperyear).• Besurethatthebrakebeltdoesnotshowsignificantsignsofwear(2timesperyear).• Checkthatthehandlebarsandseatadjustmentscrewsarelubricated(2timesperyear).• Besurethatallmovingpartsascrankandflywheelareworkingnormalandthatnoabnormal

playorsoundexists.(I.e.playinbearingscausesfastwearingandwiththatfollowsahighlyreducedlifetime.)

• Checkthattheflywheelisplacedinthecenterandwithplanerotation.• Thecrankandflywheelbearingsarelongtermgreasedanddonotrequiresupplementary

lubrication.Ifproblemariseswithcrankorflywheelbearings,contactMonarkdealer.• Replacebatteriesifthecomputerisnoton/seemstobeproducingincorrectmeasurements

Updated:16October2017 ExerciseIntervention&OutcomesCore 26

3.4Contacts

1.)MonarkExerciseAB(Vabsbro,Sweeden)+46(0)281594940

2.) MichaelBusa–[email protected]

Updated:16October2017 ExerciseIntervention&OutcomesCore 27

Chapter4:Monark874E,weightergometer

4.1Introduction:MonarkErgometermodel874Eisasafe,easy-tousebikeforfitnesstestingandworktests.Ithasabrakesystemwheretheworkloadisdeterminedbyweightsaddedintheweightbasket.Thepatentedweightbasketsystemdoesnotrequirecalibration,theprecisionoftheweightsensuresthattheworkloadiscorrect.

Quickfactsaboutthe874E:

•Large,well-balancedflywheel20kg(44lbs)

•Adjustablesaddleandhandlebar

•Stableframe,solidsteeltube

•Wheelsforeasytransport

•Electronicmeterwithheartrate

Corecompetencies:

1) Understandingthebrakedeviceandbasicsofthetensioncenter2) BefamiliarwiththeFitnesscomputerandwhatmetricsitgenerates3) Reviewserviceactionsandunderstandwhatactionsyoumightneedtotakebefore/afterusing

thisequipment

Updated:16October2017 ExerciseIntervention&OutcomesCore 28

4.2Userinstructions

CycleadjustmentsSeatheightshouldbeadjustedtoacomfortableposition.Theappropriateheightistohavethekneeslightlybentwhenthesoleofthefootiscenteredoverthepedalaxlewiththepedalinthebottomposition.Toadjusttheseatheight,loosenthelever(1)ontheseattube.SeeFig:Adjustments.Alsonotethatthehandlebarstemshouldbeinsertedintotheframetubeatleast3inches.Thismeasureismarkedwith“MAX”onthestem(3).

FitnesscomputerTheMonark874EisequippedwithaFitnesscomputerthatdisplaysthefollowingmetrics:

1. Revolutionsperminute(RPM)2. Heartrateinbpm(HR)

a. Atthedisplayforheartrate(HR)aheartsymbolislitwhichmeansthatthemeteristryingtofindapulsesignalfromanexternalsource,chestbeltwithelectrodes,IfthemetercannotfindsuchasignaltheHRfunctionisautomaticallyturnedoffafter30seconds.Whenthefunctionisturnedofftheheartsymbolisnotlitanymore.Theheartratefunctioncanbeturnedonagainbypressingakey.

3. Exercisebouttimeinminutesandseconds(TIME)a. Thetimerstartsautomaticallywhenpedalsaremoved.MetervaluesforTime,Distanceand

CaloriescanbesettozerobypressingtheRESETbutton(seeimagebelowforRESTbuttonlocation)formorethantwoseconds.

4. Cyclingspeedinkmperhourormilesperhour(SPEED)5. Covereddistanceinkmormiles(DISTANCE)6. Workload(kp=weightbasket+weightinkg)canbesetwhichgivesareadingofcaloriesburned

(CAL)andpower(WATT).EnergyisexpressedinkJ(kilojoule)orcal(kilocalories,kcal).Conversion:1

Updated:16October2017 ExerciseIntervention&OutcomesCore 29

kCal=4,2kJ.Powerisdependentoncadence,orpedalingspeed,whichmakesitpossibletoadjusttheworkload/powerbyincreasingordecreasingcadence.

a. Togetcorrectreadingsforcaloriesandwattsthekpvalueontheelectronicmeterhastobesettothesamevalueastheworkloadthatistheweightofthebasketincludingtheweightsinit.Therubberplatesareincludedinthecalibratedweightofthe1kgweightbasket.

i. Forexample,theworkloadis3kg(weightbasket0,1kg+2x1kgweight).Pressthekpkeytotheleftonthemeter.Thelowerdisplaywindowisnowflashingandshowingfiguresinkp.Increaseordecreaseinstepsof0.1kpbypressingthekpbutton(arrowup)ortheRESETbutton(arrowdown)untilthereadingiscorrespondingwiththeactualordesiredkpvalues(workload)fromtheweightbasket.Afterthat,presstheCAL/WATTbuttontoeithershowtheCALorWATTfigures.Next,presstheCAL/WATTbuttontoeithershowtheCALorWATTfigures.Refertosampleimagesbelow.

ii. Thewattreadingsinthedisplayaredependentoncadence.Thewattscanaccordinglybeadjustedbyincreasingordecreasingcadence.Caloriesarecalculatedinreal-time.

Toactivateanyofthefunctionsdescribedabove,pressanykeyormovepedals.

BelowaresampleandimagesoftheFitnesscomputerandatableoutliningmetricsthecomputerdisplayswithcorrespondingunitsandranges.

Updated:16October2017 ExerciseIntervention&OutcomesCore 30

HeartrateUsingtheMonarkchest-strapheartratebeltresearcherscanmonitorparticipantheartrate.Thistechnologysenseselectricaloutputofthehearttocomputeheartrate.Properlyfittingthechest-strapisintegraltoprovidingaccuratereadings.

Beforeharnessingthechest-strap,cleanskinwherethebeltistobeplaced.Additionally,electrodesonthebacksideofthestrapshoulidebemoistenedwithwaterorspecializedgel.Thechest-strapshouldbesecuredatacomfortabletensionaroundtheparticipant’ssternum,seeFig:Placementofchestbeltforvisitation.Thebeltshouldbeplacedsuchthattheword“Monark”isoutwardfacing,upright,andcenteredontheparticipant’schest.InordertoproperlyconnectwiththeHRreceiveronthebike,thestrapshouldnotexceed100cmindistancefromthereceiver.Itisespeciallyimportantwhenfirstpairingthechestbeltwiththereceiver,thetwounitsarewithin60cmofeachother.

Figure:Placementofchestbelt

WingatetestingWhenusingthispieceforwingatetesting,resistanceintheweightbasketshouldbesetat0.075kgperkilogramoftheparticipant’sbodyweight.Whentestinghighly-trainedathletesconsiderusingahigherresistance,forexample1.0-1.3kgperkilogramofparticipantbodyweight.

Aftertheparticipanthascompleteda3-5minutewarmup,thetestshouldbegin.Afterthreesecondsofalloutpedaling,thebrakecordistobereleased,dosobylettinggoofcord.Thiswillapplyresistancetothesystem.The30-secondstestallowsforcalculatingthefollowingvariables:

1.) Peakpower(PP),inwatts2.) Relativepeakpower(RPP),inW/kg3.) Anaerobicfatigue(AF)4.) Anaerobiccapacity,inkilogram-Joules

Updated:16October2017 ExerciseIntervention&OutcomesCore 31

Adjustingthebrakecord/belttensionImportantnote:Toreceivecorrectworkloadmetrics,itisimportanttoplacetheweightbasketaccordingtothedescriptionthatfollows.Ifthebaskethangstoolow,itmaytouchtheflywheel.Ifthebasketistoohigh,wrongworkloadinformationwillbedisplayed.

First,checkthatthebrakebeltislyingcorrectlyontheflywheelbrakesurface,seeimage1belowforproperalignment.Next,put4kgintheweightbasket(4),seeFig:Brakedevice,below.Whenrotatingtheflywheelbyhand,orwhenpulling/releasingthebrakecord,thebasketshouldlift/dropbetween40mmand60mm.Ifthisisnotthecase,thebrakebelthastobeloosenedortightenedalittleatthetensioncenter.

Loosenthecordbracket(#2inFigureBrakedevice)sothatthecordlengthcanbeadjusted.Ifthebasketistoolow,shortenbytensioningthecord.Alternatively,ifthebasketistoohigh,lengthenthecordbyeasingtension.

Toshortenorlengthentension,turnthetensioncenter(3)approximately45degrees,andmakeappropriatemodificationstocordlength.Afterthattightenthebracketagain,andcheckthatthedistancebetweentheweightbasketandflywheelisbetween40and60mmwhenyoupullonbrakecord.

Image1:Brakebeltpositioningontheflywheel

Updated:16October2017 ExerciseIntervention&OutcomesCore 32

ReplacingthebrakebeltIfthebrakebeltneedstobereplaced,loosenthebrakecordandsetthebaskettoitsupperposition.Loosenthelockwasherthatisholdingthecord(#2)andremoveitfromthetensioncenter.Loosenorcutofftheknotontheotherendofthecordandthenremovethewholecordfromthebike.Whenassemblinganewbrakecord,firstenteroneendintotheholeinthetensioncenter,andtieaknotandlettheknotfallintothebiggerpartofthehole.Locktheendofthecordwiththelockwasher.

Updated:16October2017 ExerciseIntervention&OutcomesCore 33

SampleDatasheetforWingateTest

Time RPM

0:05

0:10

0:15

0:20

0:25

0:30

4.3Generaltroubleshootingguide

Updated:16October2017 ExerciseIntervention&OutcomesCore 34

4.4MaintenanceSchedule:

Serviceaction:

• Manufacturersuggestusingisopropylalcohol.Useadampbut,notwetclothtocleanthesurfaceyouwishtodisinfect.

• Periodicallywipethesurfacewitharustpreventative,especiallywhenithasbeencleanedandthesurfaceisdry.Thisisdonetoprotectthechromeandzincpartsaswellasthepaintedparts(4timesperyear).

• Checknowandthenthatbothpedalsarefirmlytightened.Ifnot,thethreadinginthepedalarmswillbedamaged.Also,checkthatpedalarmsarefirmlytightenedonthecrankaxle,tightenifnecessary.WhentheErgometerisnewitisimportanttotightenthepedalsafter5hoursofpedaling(4timesperyear).

• Checkthatthepedalcrankissecuretothecrankaxle(4timesperyear).• Besurethatthepedalsaremovingsmoothly,andthatpedalaxleisclearofdirtandfibers(4

timesperyear).• Whencleaningandlubricatingbesuretocheckthatallscrewsandnutsareproperlytightened

(2timesperyear).• Checkthatthechainissnugandthereisnoplayinthepedalcrank(2timesperyear).• Checkthatpedals,chainandfreewheelsprocketarelubricated(2timesperyear).• Besurethatthebrakebeltdoesnotshowsignificantsignsofwear(2timesperyear).• Checkthatthehandlebarsandseatadjustmentscrewsarelubricated(2timesperyear).• Besurethatallmovingpartsascrankandflywheelareworkingnormalandthatnoabnormal

playorsoundexists.(I.e.playinbearingscausesfastwearingandwiththatfollowsahighlyreducedlifetime.)

• Checkthattheflywheelisplacedinthecenterandwithplanerotation.• Thecrankandflywheelbearingsarelongtermgreasedanddonotrequiresupplementary

lubrication.Ifproblemariseswithcrankorflywheelbearings,contactMonarkdealer.• Replacebatteriesifthecomputerisnoton/seemstobeproducingincorrectmeasurements

4.5Contacts

1.)MonarkExerciseAB(Vabsbro,Sweeden)+46(0)2815949402.)MichaelBusa–[email protected]

Updated:16October2017 ExerciseIntervention&OutcomesCore 35

Chapter5:VeltonCycleErgometerbyRacermate

5.1 Introduction-TheVelotronisalaboratory-gradecycleergometerthatdelivershighaccuracyandrepeatability.BuiltwithRacerMate’ssignaturedurability,enjoythemostrealistic,sciencegradetrainingexperienceontheworld’smostadvancedcycleergometer.Velotronisusedworldwidebythefinestuniversities,sportssciencelabs,andcoachingcenters,ofteninconjunctionwithpowerfulWingatesoftware.TheonlybicycleergometereverapprovedforuseintheUSACyclingNationalTalentSearch,Velotronisalsotheultimatecyclingtrainer.

TheVelotronisequippedwithaprecisionloadgenerator,providingsmoothvirtualshifting-pedalthroughthegearshift,withnowaitorfriction.Secondly,theradicalflywheeldesignallowsforendlesscoastingforatrueroadfeel,withacurrentdrivepowerfulenoughtodeliverthewidestspectrumofloadresistance.

POWERGENERATOR

FLYWHEEL

Updated:16October2017 ExerciseIntervention&OutcomesCore 36

5.2CoreCompetenciesGettingstartedwithVelotronCoachingSoftware

1.) ConnectEXTPCcabletotheappropriatespotonthebackofthepowergenerator

2.) Attachoppositeend(stereotoUSBadaptor)toUSBonExercisetestinglaptop

3.) OpenVelotronCoachingSoftware(CS)2008fromthestartmenu

4.) GotoSourcemenuandselectReal-time5.) Ifthisisyourfirstuse,select“New”onsub-directoryofReal-timeinSourcemenu

a. Createauserdatafile,enteruserinformationandsaveàOK.OnsubsequentusesselectOpentochoosethisuserfile.Ifthereismorethanoneuser,createaUserDataFileforeachpersonwhowillusetheVelotron.

Updated:16October2017 ExerciseIntervention&OutcomesCore 37

6.) GotoUtilitiesmenu,select“CommPorts”todeterminewhatserialportVelotronisconnected

to.Withthisinformation,selectSource-Real-time-CommPortandverifythecorrectportfoundinthetesthasbeenset.

7.) SelecttheOptionsmenutosetupyourManualErgosettings,ScreenCaptureOptions,Drag

Factor™settings,Reports,anddecidewhetheryouwanttorunthetestinMilesPerHourorKilometersPerHour.

8.) SelectStart,thenCharts,ManualErgo,Mode,orSpinScan.ByopeningChartsaselectionwindowforthecoursesortestsyouwanttouseforyourtestwillappear.Selectacoursefromthe“Courses”folderfromtheVelotronCS2008directory.ThenclickonOpen,theChartscreenwillappearwiththisselection.

9.) ByselectingacoursethiswillopenthedesiredcourseintheChartsScreen,clicktheredorgreenbuttonsontheleft-handsideofthechartwindowtoenableordisablethevariouschannelsofdata.

10.) HitF1ontheVelotronHandlebarControllertostartthetimer.Thecourseortestwillstart.11.) Belowisageneraloverviewofthechartfeatures:

Updated:16October2017 ExerciseIntervention&OutcomesCore 38

a. FullarrayofdatatakendirectlyfromaVelotronsavedperformancefile.b. Current,PeakandAveragedisplayofdata.HRzone%andPulsePower™,Gearand

Gearing.c. Userselectablecolorgraphsthatcorrespondtothedatasection.d. TimeIndicatorBartopinpointexacttimecoordinates.e. Course/programprofileforRealTimeusetoindicatecurrentposition,orworkload.f. MultipleZoomLevelsinFileModeoperation.g. RPMzonepointertohelpmaintainafixedRPMduringanytest.Itisidealforfixed

wattagetests.Concludingatest

1.) Whenyouhavecompletedyourtest,youcaneitherresetthetimerandrunanewtest,orsavetheperformancedataasapermanentrecord.

a. Attheconclusionofaknownlengthtestadialogwillappearaskingyouto“ResetorSave.”

b. Ifthetestisopen-endedyouwouldpresstheF5(reset)buttontoprompta“Save,Reset,orCancel”(CanceladdedincaseyoupressedF5bymistake).ThesequestionscanbeansweredusingtheHandlebarControllerkeysshowninthedialogtoeliminatehavingtodismountthebike

Saving/Viewing/Exportingatest

PerformancefilesinCSaresavedas.velfiles.VelfileextensionsarecreatedwhetherthetestwasanERGfile,CRSfile,orsavedSpinScansession.TheperformancefileswillbesavedwithaUserName/Courseortesttype/Date/Time.velformat.1.) PressRESETafterthecompletionofatesttodisplaythe“Doyouwanttosave”dialog.

AnswerYes.2.) UsingtheHandlebarKeys,pressF1toSaveorF2nottosave.3.) Exportingthedata.SavedRaceFilesfromanyVelotronPCprogram(availableonyourhard

drive)canbeexportedasASCIItextfiles.Thiscanbedoneautomaticallyormanually.Ifyou

Updated:16October2017 ExerciseIntervention&OutcomesCore 39

alwayswantanexportperformancefilesafteratest(forusein3rdpartyapplications)checktheAutoExportoptionintheExportOptionsmenu.FormanualexportchooseFile,thenselecteithertheExportfunctionorchangetheExportOptionsasneeded(check-markboth”shortformat”&“commadelimited”intheoptionsifyouplantousetheseexportfilesinaspreadsheet).Youwillbeaskedtoselecttheoriginalsavedperformancefile.Youcanthenbrowsetoanyfolderwherethefilemayreside.Onceexported,thesefileshavethesamenameastheoriginalfile,butwithanew.txtfileextension.

4.) Toopenaperformancereport,selectFile,OpenReport,andnavigatetothedesiredfileandselect.

FileMode,viewingsavedracefiles

1.) Select“SavedFile”fromtheSourcemenu.2.) SelectOptionsandcheckoruncheckMetricasdesired.3.) SelectStart,thenCharts,andyouwillbepresentedwithawindowwhereyoucanbrowseto

anyfolderfromthe“LookIn”dropdownwindowtoselectasavedperformancefile.Thiswillopeninthebackground,thereforeminimizeCS2008andopenthisdialoguebox.Oncethefileisfound,highlightitandselectOpen.Thechartscreenwillappearwiththeracedatashowinginit’sentirety.Toopenanotherfile,pressCtrl-O.

4.) Clicktheredorgreenbuttonsontheleft-handsideofthechartwindowtoenableordisablethevariouschannelsofdata.

Updated:16October2017 ExerciseIntervention&OutcomesCore 40

5.) Youcannowmovethemouseleftandrightoverthedataandzoominandoutwiththeleftandrightmousebuttons.Right-clickyourmouseforFullZoomIn,andscrollforwardandbackthroughthedatabymovingtheTimeIndicatorBarfromtheleft-handorright-handsideoftheChartWindow,orestablishaZoomWindowusingtheleft/rightandup/downarrowkeysofyourPCkeyboardandhitEntertoactivatethislevelofzoom

Creatinganewcourse/programBecausecoursefilesmustretaintheformatasshownintheimageonthepreviouspage,itisrecommendedtoeditoneofthepreexistingcourseswhenmakinganewcourse.Thiswillalmostguaranteeausablecourseeverytime.

1.) TocreateanewcoursefromanexistingcoursewhileinNotepad,firstopenapreexistingcourse.

a. Toopenapreexistingcourse,openNotepadfromtheStartmenub. InNotepad,clickFile,thenOpenc. NavigatetoVelotronCS2008folderontheCdrive,thenopentheCoursessubdirectoryd. BydefaultNotepadlooksonlyfor“textfiles”,butifyouleft-clickonthedownarrowin

the“Filesoftype:”sectionandchoose“AllFiles(*.*)”,thenallthefilesinthisfolderwillappear.SelectthecoursetheyoudesiretoamendandclickOpen

Updated:16October2017 ExerciseIntervention&OutcomesCore 41

e. WhenthecourseyouwanttoeditisopeninNotepad,renameitbyclickingFile,Save

As.TypeanewnameforthiscourseandchooseAllFiles(*.*)extensionwhensavingnewcourse.

f. LeftclickSavetomakea“carboncopy”of,andclose,thefirstopenedcourse.g. Nowyoucanproceedwithchangestomaketoyournewcourse:

i. Generally,toamendacoursechangeparametersfromtheoriginalcourse.Followtheformattinginstructionsoutlinedbelow:

h. Whenfinishedediting,clickFile/Save.

Updated:16October2017 ExerciseIntervention&OutcomesCore 42

5.3GettingstartedwithVelotronWingateSoftware

1.) GototheStartmenuandopenVelotronWingate

2.) GotoClientmenuandselectNewDatabase

Updated:16October2017 ExerciseIntervention&OutcomesCore 43

3.) Next,gotoClientmenuandselectNewClienta. Enterthedatabasethatyouwantdatatobesavedto,entersubjectantropometric

measuresandapplyoneofthefollowing:1.)torquefactor,2.)Torquevalue,3.)Droppedweight

b. OKtosave

4.) Now,gotoWingatemenu,andselectCreatenewprotocol

a. Createanameforprotocolandedittestingparametersb. Oktosavetodesiredpathway.Suggestedtosaveinclientdatafolder

Updated:16October2017 ExerciseIntervention&OutcomesCore 44

5.) SelectWingatemenuandclickGraphScaling.EditWattsandHR/CadenceScaling6.) CheckthatVelotronispluggedintocomputer(seeabove).SelectSetupmenuandclickrunTest

CommPorts.NotewhichComminconnectedtoVelotron.ClickOK.Next,clickSetupmenu,andselectCommPort,verifythatthissettingmatchesTestCommPortsresults.Seeabovesectionforimages.

7.) GotoClientmenu,selectLoadClientandnavigatetoclientdata.a. GotoC-drive,VeloTronWingate2008folder,datafolder,selectyourdesireddatabase

folder,andselectandopenclientfolder

Updated:16October2017 ExerciseIntervention&OutcomesCore 45

8.) Now,selectLoadProtocolfromWingatemenu.Navigatetotheprotocolyoumadeinstep4.

9.) Next,gotoWingatemenuandclickRunWingatea. TheVelotronisnowis“Warm-upmode.”Theclientcanwarm-upusingthepreselected

load-or-byusingtheUp/DownbuttonsontheWingateScreenorthoseontheVelotronHandlebarControlleradjusttheworkloadtosuitthewarm-uprequirements.Duetox-axisscalinglimitationsthereisonlyabout20secondsofwarm-upshown,butthewarm-upperiodpriortostartingtheclockisopen-ended.

Updated:16October2017 ExerciseIntervention&OutcomesCore 46

10.) Whenyouarereadytobeginthetest,selecttheStartWingatebutton,thefirsticononthetoolbar

STARTa. Thewarm-upphasethatyouselectedwillbegin,thenaredcountdowntimerwill

appearintheupperrighthandcorner,indicatingwhentheloadwillbeappliedb. Attheendofthetestaplotoftheclient’sperformancewillappear.ClicktheReset

button,fourthbuttonfromtheleftonthetoolbar.Selectthepathwaywhereyouliketosavetheperformance.Suggestedsavinglocationistheoriginalclientdatafolder

RESETFilemode,viewingandexportingsaveddata

1.) ReturntotheWingatesoftwarehomepage2.) Toviewdatainfilemode,intheTree-viewdiagramontherightsideofthepageselectyour

database,thenclient,andfinallytheclientperformancefilea. Rightclickontheperformancefileandchooseplot.APlotwillappearintherightpanel.

Updated:16October2017 ExerciseIntervention&OutcomesCore 47

3.) Toexportdata,returntothetree-viewdiagramandrightclick,butselectexport.a. Adialogueboxwillappearthatnoteswherethefilehasbeensaved

b. Theprogramwillexporttheperformancefileincommaseparatedtextformatintothesame

databasefoldertheoriginalperformancefilewasfound.Thefilewillhavethesamefilename,butwithaTXTfileextension.ThesefilescanbeusedMicrosoftexcel.

5.4Contacts

1.) Velotronforums:https://www.racermateinc.com/help-center-faq/2.) MichaelBusa–CoreDirector

413-577-0574 [email protected]

Updated:16October2017 ExerciseIntervention&OutcomesCore 48

Chapter6:Quinton-55EKGStressSystem

6.1.Introduction-ImportanceofElectrocardiogramsAstheheartisamuscularpumpcomprisedoffourchambers,itrequireselectricalstimulationtocontrolthemuscularcontractionsandflowofbloodthroughoutthebody.Anelectrocardiogramisamethodtypicallyusedtoassesselectricalconductionintheheart.Thissimplenon-invasivetestallowstocheckfornormalrhythms,ratesandproperconductionofelectricalimpulsethroughoutthecardiactissue.ThismanualwillallowaresearchertounderstandandutilizetheEKGsysteminordertoobtainEKG’sformedicalinterpretation.Forallsubjects>70yearsofageorifasubjecthas≥2cardiovascularriskfactors,anEKGmustbecompletedpriortobeginningexerciseprotocols.

6.2EKGSettingsUnlessspecificallyrequestedotherwise,EKGsshouldberunwiththefollowingstandardsettings:PaperSpeed:25mm/sec

CalibrationStandard:1mV=10mmRhythmLead:DefaultLeadII

6.2.1SkinPrepPatientswithexcessivechesthairshouldbeshavedattheelectrodesite.Abradetheseareaswithfinesandpaperoranabrasivepad,andthencleanwithalcohol-saturatedgauze.Allowtheskintoairdrybeforeplacingelectrodes.Besurethesubjectdoesnothaveanyjewelrythatcouldinterferewiththetracings(necklaces,piercingsetc.).

6.3LimbLeadPlacementRA(White)–RightArmelectrodeisplacedonadistalportionoftherightlateralsideoftheupperarmbelowtheshoulder.LA(Black)–LeftArmelectrodeisplacedonadistalportionoftheleftlateralsideoftheupperarmbelowtheshoulder.RL(Green)–RightLegelectrodeisplacedontheinsidecalf,midwaybetweenkneeandankle.LL(Red)–LeftLegelectrodeisplacedontheinsidecalf,midwaybetweenkneeandankle.

6.4PrecordialLeadPlacementV1(COLOR)–4thintercostalplaceattherightmarginofthesternum.V2(COLOR)–4thintercostalplaceattheleftmarginofthesternum.V3(COLOR)–MidwaybetweenV2andV4(ontopofthe5thrib).V4(COLOR)–5thintercostalplaceattheleftmid-clavicularline.V5(COLOR)–AtthehorizontallevelofV4,attheleftanteriorline.V6(COLOR)–AtthehorizontallevelofV4,atthemid-axillaryline.

Updated:16October2017 ExerciseIntervention&OutcomesCore 49

NOTE:LeadplacementdoesaffecttheEKGwaveform.Whenthelimbleadsareplacedonthetorso,waveformchangesmightbeseenintheQRSamplitude,axisshiftoccurs,Qwavescanbeseen,andTwavesmightappearflippedorflattened.Thesechangesareclinicallysignificantinthattheyareassociatedwithcardiacischemia.Ifanon-standardleadplacementisused,notethevariationintheEKGcommentfield.

6.5 InputtingaNewPatient

a. ToinitializetheQ-stressprogram,clickontheicononthedesktop

Updated:16October2017 ExerciseIntervention&OutcomesCore 50

b. Tofirstsearchforpreviouspatients,clickthepatienticon

c. SearchforthepatientnameorMRN(MedicalRecordNumber)

Updated:16October2017 ExerciseIntervention&OutcomesCore 51

d. Ifpatientinformationisnotfound,Select“New”andinputpatientdata

e. Thefollowingscreenwillallowyoutoenterallpossibledemographicinformation.ClickOKwhencompletedtocompletethedataentry.Alsonotethereshouldbenochangesmadetoprotocol,procedure,orreportunlessinstructed.

Updated:16October2017 ExerciseIntervention&OutcomesCore 52

6.6 CheckLeadPlacementandArtifact

a. Afterpatientdataiscomplete,clickonthelightblueicontocheckleadplacementandconnections.

b. Click“Start”andwaitfortheiconstochangecolor

NOTE:Leadsshouldbeinthe“good”range

c. IfLeadsarenotwithinacceptable“good”range,removeleadsandpreptheelectrodesitesagain

Updated:16October2017 ExerciseIntervention&OutcomesCore 53

d. Clickexitafterallleadsareproperlyplaced.

6.7 BaselineWaveforma. Aftertheleadsareplacedandthepatientdataisinput,clickontheresticontobeginrestingwaveforms.

Updated:16October2017 ExerciseIntervention&OutcomesCore 54

b. ThescreenwillbegintodisplayheartrateandEKGwaveforms.Allowsabout10secondstopassbeforecollectingarestingEKG.

c. PressF5toprintoutafull12-leadrestingEKGd. PressF6toprintthecurrent3-leadscreene. Tobeginexerciseprotocolsclickonthe“startbelt”iconfollowedbythestartexercisebutton

Updated:16October2017 ExerciseIntervention&OutcomesCore 55

f. Toendexerciseprotocolsclickthestartrecoverybuttonfollowedbythe“stopbelt”button

g. ToreviewthecollectEKGsclickthereviewtabandclick“Yes”

Updated:16October2017 ExerciseIntervention&OutcomesCore 56

h. Clickyeswhenenteringreviewmode.Thiswillendthetest.

i. Whenthetesthasended,ifareportisneedclick“PartialReport”

Updated:16October2017 ExerciseIntervention&OutcomesCore 57

Thiswillprintasummaryreportoftheprocedure

j. Whenfinishedreviewing,click“Exit”tosavethetest.Whenprompted,besurePDFisclickedyes,andpress“OK”

k. Toexitcurrentpatient,press“Exit”again

Updated:16October2017 ExerciseIntervention&OutcomesCore 58

Afterobtainingabaselinewaveform,theEKGshouldbeinterpretedbyamedicalprofessional.Theinterpretationwillbecompletedbyeitheracertifiedclinicalexercisephysiologist(CCEP)ormedicaldoctor(MD)withinthedepartment.Anexercisephysiologistmayinterpretthebaselinewaveformswithsomerestriction.IftheEKGisdeterminedtobesatisfactoryandwithoutcontraindicationstoexercise,theexerciseportionwillresumewithoutreferralfromthemedicaldoctor.However,iftheexercisephysiologistdeterminesthatthebaselinewaveformisquestionableinanyway(basedoncurrentACSMrecommendationstoexercise),themedicaldoctorshouldbeconsultedtoverifytheEKG.Thedoctorwillthendeterminewhetherornottoproceedwithtesting.QuestionablebaselineEKG’scanincludebutarenotlimitedto:frequentectopy,abnormalheartrates(<40bpmor>100bpmatrest),STsegmentabnormalities<2mmdepressionorconductiondelays.Subjectswithsignificantabnormalitieswillbeexcludedfromthestudy.SignificantEKGabnormalitiesmayincludebutarenotlimitedto:dysrhythmias,electronicallypacedrhythms,conductionblocks,>1mmSTsegmentelevationsor>2mmSTsegmentdepressions.

6.8 ExerciseProtocols

AsthereisnocontinuousEKGmonitoringduringCADENCE-Adults,protocolisnotsignificantwhencollectingEKG’s.EKGwillbean“asneeded”proceduretoensuresafety.

6.9 Follow-Up

Ifatanytimeduringtestingprotocols,asubjectbecomesdizzy,lightheadedorcomplainsofchestpain,afollow-upEKGshouldbeperformedalongwithbloodpressures.InterpretationoftheEKGshouldbeimmediatelyreferredtotheExercisePhysiologistorMDonsite.

Updated:16October2017 ExerciseIntervention&OutcomesCore 59

6.10 Cleaning

Allwiresandconnectionsshouldbewipeswithdisinfectingwipes.Usedelectrodesshouldbethrownaway.Thetreadmillandanyotherequipmentassociatedwiththetestshouldbethoroughlycleanedbetweenparticipantswithalcoholwipes.

Updated:16October2017 ExerciseIntervention&OutcomesCore 60

ChapterX:

BiodexSystem4ProDynamometer

Turningthingson:

1.MakesuretheBiodexispluggedin.

2.Turnonthemainpowerswitch(locatedonthebottomleftofthebackoftheBiodexcart)byflippingitfromOtoI.

3.Turnonthepowerforthecomputeranddynamometer(greenswitchesonthebottomrightofthebackoftheBiodexcart)byflippingthemfromOtoI.

4.Turnonthecomputertower.

BasicsofsettinguptheBiodex

ThedynamometerandtheBiodexchaircanbetranslatedinspacesidetosideandverticallyandrotated.Thefollowingimagespointouthowtomovethechairanddynamometer.

Updated:16October2017 ExerciseIntervention&OutcomesCore 61

Tiltthedynamometerheadbylooseningthisknob.

Comfortstopbutton.Pushingthisbuttonwillimmediatelystopatestandexitoutoftestingmode.Useasakillswitchonly.

Updated:16October2017 ExerciseIntervention&OutcomesCore 62

Thisknobisusedtotilttheseatback

Thisleverisusedtorotatetheseat

Whenputtingattachmentsondynamometer,thesedotsmustalign.Youcanrotatethedynamometerdotusingthebluebuttonsonthedynamometerhead

Watchout,fingers/handsarecrushable!

Attachmentswithoutanutonthemwillusetheattachmenthandlestoredonthedynamometerhead(seeabove)Lettercorrespondstoside

(left/right)you’retesting

Updated:16October2017 ExerciseIntervention&OutcomesCore 63

RunningaTestinBiodexAdvantage:

TheseinstructionswilltakeyouthroughrunningatestontheBiodex.IfyouwishtousetheExerciseorBiofeedbackmodes,pleaseseetheinstructionsintheBiodexusermanualorhelpdocumentation.

1.Whenthecomputerbootsup,itwillaskyoutoinitializethedynamometer.Removeanyattachments(theattachmentpostwillrotate)andclickOK.Thedynamometerwilltheninitialize.

2.BiodexAdvantageusuallybootsuptothelasttestthatwasrun.

A.Ifyouneedtoseehowtosetupthebiodexattachmentsforagiventest,clickthe“Anatomical”button.Inthewindowthatpopsup,youcanselectthejointandjointmotionyou’retestingtoseeavideoofhowtosetupthemachine.

Updated:16October2017 ExerciseIntervention&OutcomesCore 64

B.Beforestartinganewtestyouneedtoloada“patient”andselectaprotocol.Follow thebuttonsontheleftfrom1-4.Asyougothroughthesteps,theactivewindowwillbe yellow.

1.Selectajointandmotion2.Selectamethodtoviewtheinstructions

3.Clickview

Updated:16October2017 ExerciseIntervention&OutcomesCore 65

3.ThePatientwindowgivesyouoptionstoopenanexistingpatient,addapatient,oreditanopenpatient.

Tostartatest,clickherefirst

Toviewinstructionsforhowtosetupagiventest,clickhere

Notreadybecauseyouhavetogothroughthetestsetupfirst

Updated:16October2017 ExerciseIntervention&OutcomesCore 66

A.Ifyouclick“Open,”youcanseealistofexistingentriesandselectonetocreateanewtestingsessionfor.Ifyouselectapatientfromthislist,youcanbeginanewtestbyclickingthe“NEW”buttonorEDITthepatient’sinformation.Ifapatienthasexistingtests,youcanREPEAToneofthese,printareportforanexistingtest,orDELETEanexistingtest.

B.Ifyouclick“AddPatient,”thewindowwillchangetotheviewbelow.Atminimum, youmustfilloutanyinformationboxesthatareinred.NOTE:Personalidentifying informationshouldneverbeenteredintotheBiodexsoftware.Onlyenterstudy participantIDsintonamefields.Neverenternames,birthdates,age,address,phone number,orotherinformationthatcouldidentifyanindividual.Weightandgenderare requiredbyBiodexandcanbeentered.

Updated:16October2017 ExerciseIntervention&OutcomesCore 67

C.Afterenteringallnecessarypatientinformation,click“Save,”andthenclickon “Protocol.”

Updated:16October2017 ExerciseIntervention&OutcomesCore 68

4.Hereyoucanloadaprotocol,addaprotocol,oreditaselectedprotocol.

Optionbuttons

Updated:16October2017 ExerciseIntervention&OutcomesCore 69

A.Ifyouclickthe button,thefollowingwindowwillpopup.Hereyoucanviewbuilt-inorpreviouslydesignedtestingprotocols.

Updated:16October2017 ExerciseIntervention&OutcomesCore 70

B.Ifyouwanttocreateanewprotocol,clickthe button.Thiswillactivatethe editableportionsoftheprotocolpane.

i.Dependingonthe“StudyType”and“Mode”optionsyouchoose,theotheroptionsintheblueboxabovewillchange.Forexample,ifyouchangefromIsometricmodetoIsokineticmode,insteadof“#Positions,”thegreenboxwillbelabeled“#Sets.”

Ifyouclickthis,theseoptionsbecomeavailable

Updated:16October2017 ExerciseIntervention&OutcomesCore 71

5.Afteryouselectaprotocol,clickonthe“SetROM”button.Atthispoint,thechairanddynamometershouldbeinposition,thecorrectattachmentshouldbeonthedynamometer,andtheparticipantshouldbeseatedandstrappedin.

A.Belowisthe“SetROM”window.Followtheblinkingredtextthrougheachstep.

1.First,selectthesidebeingtested

Usingthisoptionisnotrecommended(generallyonlyavailablewhenswitchingfromonesidetotheotherinbilateralprotocols)

2.ClickCLEARLIMITS.SetAWAYandTOWARDlimitbymovingparticipant’slimbtothesepositions,holdingthere(seebelow),andclickingtheSETbutton.Notethatthissetstherangethattheparticipantwillbetestedin.

3.MovelimbtoAnatomicalReferenceposition,holdthere,andclicktheCalibratePositionbutton.AnatomicalReferencecanbechangedbyclickingonthebuttonthatreads“90”here.

4.MovelimbtoAWAYlimit,holdthere(usingholdbutton),haveparticipantcompletelyrelaxlimb,andclicktheLimbWeightbutton.Ifyougetanerroronthis,seetextbelowini.

5.ClickContinueViewSetupbuttonwillshowaschematicofwhat“AWAY”and“TOWARD”are,aswellashowanglesarereferenced

Updated:16October2017 ExerciseIntervention&OutcomesCore 72

i.Clicking“LimbWeight”maygivethefollowingerror:

ii.IfthisoccursandthelimbwasweighedattheAWAYlimit,youwillneedtoincreasetheAWAYlimit,movethelimbtothisposition,andweighagain.IftheAWAYlimitisalreadyattheparticipant’slimitforthegiventest.

6.Afterhitting“Continue”intheSetROMwindow,theprogramwillmoveontotesting.TheprogramwillswitchtothefollowingwindowandwaitthereuntilyouhitSTART.IfStartisnotanoption,itmeansyoumissedsomethingbetweenstep3andthisstep.Whenyouarereadytobeginyourtest,clickSTART.

A.Ifyouwanttochangethevisualizationforatest(e.g.,bargraphinsteadofline),clickonthe

options button.

Holdbuttoncanholdparticipant’slimbinaposition.

Updated:16October2017 ExerciseIntervention&OutcomesCore 73

B.First,atrialrepswindowwillappear.Here,theparticipantcan practicethemotionandspeedofthetestbeforetheprogramstarts collectingdata.Whenyou’rereadytostarttherealreps/sets,have theparticipantholdtheirlimbinthestartposition(usually TOWARDlimit).

C.Oncetheparticipantholdsatthestartpositionforasufficient time,thetestwillbegin.Youwill seethiswindow:

Startbutton

Updated:16October2017 ExerciseIntervention&OutcomesCore 74

D.Oncethetestbegins,youwillseereal-timedataofthetorquetheparticipantis generating,aswellasprogressthroughthecurrentsetviathePieChart.

E.Duringrestperiodsbetweensetsorsectionsofaprotocol,the windowtotherightwillcountdowntothenextset.

7.Whenatestisdone,youcanclickontheReportbuttonortheCurvebuttontoviewresults.

A.Thereportwindowgivesoptionsforgeneratingstandardreports.Thisisaneasywaytogiveparticipantsfeedbackontheirperformance.Hereyoucanpreviewareportorprintareport.Ifyouwillnotbeprintingareportandalltestsarecomplete,thetestingsessionisover.

Updated:16October2017 ExerciseIntervention&OutcomesCore 75

B.ToanalyzeresultsinBiodexAdvantageorexportdatatoanASCIIformat,clickontheCurvebutton.

i.Therearemanyoptionsforreal-timeanalysisinthecurvewindow.Thesearepointedoutbelow.

ii.DataexportedfromtheCurvewindowcanbetakenoffthecomputeronaUSBdriveandanalyzedusingothersoftware(e.g.,MatLab,Excel).

Clickheretoexportdata

Enablecursors(bluelinesingraphabove)andusebuttonsheretoselectrangesofdata

Optionsforadjustinggraphdisplayandperformingsomereal-timeanalysis.Wedonotrecommendusingthe“Filtered”optionexceptfordemonstrationpurposes

Changebetweentestshere

Updated:16October2017 RoomCalorimeterCore 76

MANUALOFPROCEDURES

HumanMotionCore

InstituteofAppliedLifeSciences

LifeScienceLaboratoriesS360240ThatcherRoad,

Amherst,MA01003

atthe

UniversityofMassachusetts,Amherst

Updated:16October2017 RoomCalorimeterCore 77

Chapter1QualisysOqus700+/500+w/AMTIForcePlatformintegration

Chapter1.1-CreatingaNewProjectinQTM

WhenbeginninganewstudyyouwillneedtocreateanewprojectinQTM.Thisprocessensuresthatyoudon’tinadvertentlyalteranotherstudy’sprojectsettingsandalsocreatesadesignateddirectoryforyourstudy’sdata.

1.First,openQTM

2.AwindowdisplayingcurrentprojectswillpopupwhenyouopenQTM.Whencreatinganewproject,click“Newproject…”

3.Thefollowingwindowwillpopup.In“Projectname:”typethenameofyourstudyorthenameyouwouldliketobeappliedtoyourdirectory.Notethatwhenyouchangethisname,theendofthe“Projectfolderpath:”willchange.Ifyouwouldliketheprojectfoldertobelocatedsomewhereotherthanwhereisindicatedbythe“Projectfolderpath:”youcanbrowsetothatlocationusingthe“Browse…”button.Whenyournameandpathareset,select“Settingsimportedfromanotherproject”andhit“OK.”

Updated:16October2017 RoomCalorimeterCore 78

4.Anotherdialogwindowwillopenaskingyoutoselecta“Settings”filetobaseyournewprojecton.NavigatetoC:\Users\Qsys02\Documents\Project_Starterandselectthe“Settings”filefoundthere.Theseprojectsettingshavebeencreatedspecificallytoactasastartingpointfornewprojects.

Typeyourdesiredprojectnamehere

Checkthispath

Usethisoption

Updated:16October2017 RoomCalorimeterCore 79

5.Afterclicking“Open,”QTMshouldstartup.EnsurethatyouareinyournewprojectbycheckingthetopbarontheQTMwindow.Itshouldnowread[Project:Yourprojectname]

Additionally,thereshouldnowbeafolderatthepathyoupreviouslydesignatedthathasthefollowingfolders/filesinit:

6.Afterverifyingthatyournewprojecthasbeencreatedcorrectly,youcanbeginchangingsettings.Todo

this,firstclickontheprojectoptions button.Thereareseveralprojectsettingsthatyoumaywishtochange(NOTE:afteranychangeismade,youmusthit“Apply”).Thissectioninthismanualisnotall-encompassing;forfurtherdetailsseetheonlineQualisysusermanualsection“Projectoptions”byclickingHelp>Helptopics.

InputDevices.Hereyoucanchangewhichdevicesyouwillbecollectingwith.Oquscamerascannotbedisabled,butinthissettingwindowyoucanselectordisableforceplatecollection(USB-2533)andEMGinputs(MEGA,Noraxon).

Updated:16October2017 RoomCalorimeterCore 80

Markercapturefrequency.Changethisnumbertoadjusthowmanyframesofmarkerdataarecollectedpersecond.Capturefrequencyneededdependsonspeedofmovementbeingcaptured.

Timingoptions.Usethesesettingstoactivateanexternaltrigger.Thismaybedesiredifyouaresynchronizingmultipledataacquisitionsoftwares(e.g.,QTMandEMGsoftware)orifyouwishtotriggeradatacapturefrominthecollectionspaceinsteadoffromthecomputer.

Updated:16October2017 RoomCalorimeterCore 81

Camerasettings.Inthistabyoucanadjustcapturesettingsforindividualcameras.NotethatyoucanadjustsettingsformultiplecamerasatatimebyholdingCtrlwhileclickingoncameranumbersoryoucanadjusteachcameraindividually.Ingeneral,camerasettingsshouldnotneedtobechanged.Instanceswhentheymayneedtobeadjustedincludetheuseofadeviceinthecapturespacethatisveryshinyorverysmallorverylargemarkers.IfyouwishtochangecamerasettingsitisrecommendedthatyoureadtheQualisyshelptopics.

Updated:16October2017 RoomCalorimeterCore 82

Analogsettings.Hereyoucanchangethecapturerateforanyanalogdevicesyouincludeinyourproject.Sampleratesmustbeamultipleofthecameraframerateandcanbechangedintheboxoutlinedbelow.Therearealsosettingshereforremovingoffsetfromanalogdata(yellowbox).Usecautionifactivatingtheseoptionsasthesoftwareoffsetremovalassumesthatthereisnothingontheforceplateswhenitremovesanybaselineoffset,afeaturethatmaybeproblematicifsomeonehappenstocrossoveraforceplateduringtheoffsetremoval.

Updated:16October2017 RoomCalorimeterCore 83

3Dtracking.Hereyoucanadjustsettingsforerrortolerancein3Dmarkertracking.

“Predictionerror”isthemaximumdistancebetweenamarker’spredicted locationandthecapturedlocationofthemarker.Ifthecapturedlocationis outsideofthispredictedlocation,thesoftwarewillnotreconstructit.Increasethis valuetoincreasethetoleranceinreconstruction.

“Maximumresidual”isthemaximumdeviationincamera2Dviewsofamarkerthatcanbeconvergedintoasinglemarkertrajectory.Qualisysrecommendsthatthisvaluebe2-5timestheaverageresidualvaluesfromthecalibrationresults.

“Minimumtrajectorylength”istheminimumnumberofframesofdatato createatrajectory.Dataseenforfewerframeswillnotbeassignedtoatrajectory. Theminimumvalueforthisoptionis2.

“Autojoin”automaticallyjoinstrajectoriesthatappeartobelongtothesame markerin3Dspace.Thismayreducethenumberofpartsinatrajectoryandmay maketrackingeasier.Thisdoesnotautomaticallyfillgaps.

Trajectories.Hereyoucanadjustsettingsforautomaticallyfillinggaps.

Updated:16October2017 RoomCalorimeterCore 84

ForceData.Inthesesettingsyoucanselectwhichforceplatesyou’dliketocollectwithandcheckadvancedforceplateoptions.

GUI.InGUI,2DViewSettings,and3DViewSettingsyoucanadjustsettingssuchasthereal-timescreenrefresh,andtheappearanceofthescreenin2Dand3Dmodes.

Miscellaneous.HereyoucanadjustfoldersettingsandQTMstartupsettings.Pleasedo notadjustthe“Globaldirectories”options(inFolderOptions)orthe“Automaticallyloadaprojectatstartup”options(inStartup)asthesechangedefaultsforallusers.

ProjectoptionsareautomaticallysavedtoyourprojectSettingsfilewhenyouclick“Apply”andthen“OK.”Itisrecommendedthatyoualwayspilotdatacollection,processing,andanalysisusingyournewprojectsettings.Thiswillensurethatyoursettingsaresufficienttocapturetheinformationyouneed.

Updated:16October2017 RoomCalorimeterCore 85

Chapter1.2

QTMDataCollection

Ifcollectingdataonaparticipant,turningthesystemonandcalibrationstepsshouldbeperformedbeforetheparticipantarrives.Camerasandforceplatesshouldbeallowedtowarmupforatleast20minutesbeforecalibrating.Turningthingson:

1.Turnondatacollectioncomputer.

2.Turnonpowerstripforcameras(locatedontowertoleftofdatacollectioncomputer)

3.Makesureforceplatesareon(3boxesdirectlytotheleftofthecomputermonitors).

4.Find/addnewparticipantfoldertoprojectfolder(C:\Users\Qsys02\Documents\yourprojectfolder\Data).

5.OpenQTM

Powerswitch

Manualzeroingbutton

Updated:16October2017 RoomCalorimeterCore 86

6.Loadyourproject

7.Ifyouneedtocreateanewproject,followinstructionsinChapterX.X

Updated:16October2017 RoomCalorimeterCore 87

Calibration:

1.Hitthe“new”button (topleftoftoolbar)

2.Clickon2Dview,checkcameraviewstomakesureeachcameraisseeing0markers.

3.PlaceLframeonforceplate1(endsmallFP)withthelongarmpointingtowardsthebigforceplate.Makesureitfitstightlytotheforceplatecorner(usedrop-downlegs),alignswiththeforceplateedges,andislevel.IfthelevelsareoffthetiltoftheLcanbeadjustedusingtheroundscrewsonthearms.AdjustthemuntiltheLislevelinbothplanes.

Updated:16October2017 RoomCalorimeterCore 88

4.Putthecalibrationwandtogether.Clickthecalibrationbutton .

5.Adjustcalibrationsettingsaspreferred.Forcalibratingavolumeforgait(all3forceplates+space)60secondsmaybenecessary.Usingacalibrationdelaywillgiveyoutimetohit“OK”andgetsetinthevolumebeforethestartofcalibrationcapture.Whenyouarereadytocalibrate,hit“OK.”

6.Calibratethevolume.It’sgenerallyrecommendedtodothisinasystematicmethod,coveringtheentiretyofthedesiredcapturevolumewiththecalibrationwandinawaythatallcamerasviewthewandoverthefullvolume(i.e.,don’thaveyourbacktoonewallofcamerasthroughouttheentirecalibration).

Updated:16October2017 RoomCalorimeterCore 89

7.Calibrationwillendwhenthesettimehaselapsed.Thiswindowwillpopup:

Standarddeviationofwandlengthshouldbelessthan1mm.Ifthiscriteriaismet,hit“OK”andputawaycalibrationequipment.

8.Calibrationwilloccasionallyfail.Ifcalibrationfails,checkcamera2DviewstomakesurenomarkersotherthantheLandthecalibrationwandarevisible.Ifnoothermarkersorreflectionsarevisible,startcalibrationoveragain.

Note:cameraswillpickupanythingreflective,includingrunningshoesandreflectivetagsonclothing.Makesureyoudonothaveanythingreflectiveonwhencalibrating.

Updated:16October2017 RoomCalorimeterCore 90

DataCollection:

1.Withnothingontheforceplates,zeroforceplates(orangebuttonsonforceplateboxes).Whenplatesarezeroed,theorangelightsshouldbeoff.

2.Clickthe“capture”button. Thiswindowwillpopup:

3.AdjusttheCapturePeriodtothedesiredlengthoftime.Changingthetimeboxwillautomaticallychangethenumberofmarkerframesbox.

4.Checkthe“Savecapturedandprocessedmeasurementautomatically”box.Makesurefolderthatisdirectedtoiswhereyouwouldlikeyourdatatobestored.

5.Changethe“Name”boxtowhateveryouwouldliketheprefixofyourfilestobe.Ifyouwouldlikefilestobeautomaticallynumberedasyoucollectthem,check“Addcounterstarting…”

6.Click“Options…”tocheckCameraSystemSettingstomakesureframerateandtriggersettingsarecorrect.

Updated:16October2017 RoomCalorimeterCore 91

7.Whenyouarereadytocollectatrial,hit“Start.”

8.Attheendofthecaptureperiod,thetrialwillopenin3Dview.Playthroughthetrialusingthebuttonsinthe

toolbar:

orbydraggingthetoparrowalongthebottomtimebar:

9.Labelingdata(tocreateanAIMmodel,useamotiontrial,notastatictrial):

Scanthroughthetrialtomakesureallnecessarymarkerandforcedatawerecaptured.Movedatatoatimepointwhereallmarkersarevisibleinthe3Dview.

Tolabeldata,firstopentrajectorywindowsbyclickingView>TrajectoryInfo>All.

Right-clickinthe“Labeledtrajectories”windowandselect“Loadlabellist…”

Selectyourmarkerlist.

Ifyouhavenotpreviouslycreatedalabellist,right-clickontheword“Trajectory”andselect“Addnewlabel.”Thiswillcreatea“New####”label.

Addasmanylabelsasmarkersinyourmarkersetandnamethemappropriately.

Whenalllabelsarecreated,right-clickinthe“Labeledtrajectories”windowandselect“Savelabellist…”

II.Tobeginlabelingmarkers,selectthefirstmarkerinthelabellistandclickthe

button(leftverticaltoolbar).

III.Inthe3Dview,clickonthemarkercorrespondingtothefirstlabel.InID mode,theprogramwillprogresstothenextlabelinthelabellistwhenamarkerisclickedon.

IV.Continueclickingonthemarkersinthe3Dviewthatcorrespondtothelabel highlightedinthelabellistuntilyoureachtheendofthelabellist.

V.Clickthe buttontoturnoffIDmode.

C.Trimtrialsothatonlylabeleddataareincludedusingthesidearrowsofthetimebar.Onlylabeleddataareincludedwhenallmarkerfilllevelsareat100%(inlabeledtrajectorywindow).

D.Click“GenerateAIMmodel”button

Updated:16October2017 RoomCalorimeterCore 92

E.Select“Createnewmodel”inthewindowthatpopsupandthenclick“Next.”

F.Inthenextwindow,verifythatthemodellookscorrect,andthenclick“Next.”

G.EnterafilenameforyourAIMmodelandclick“OK.”Inthefinalwindow,click“Finish.”

10.TohaveAIMappliedasyoucollectdata,inProjectOptions,gotoAIMandaddthemodelyoujustcreatedtothe“Appliedmodels:”window.Hit“Apply,”then“OK.”

11.Continuecollectingtrialsuntilfinished.

12.Ifyouwishtooverwriteatrial,settheNameandcountertomatchthetrialyouwishtooverwrite.Whenyouhit“Start,”awindowwillpopupaskingifyouwishtooverwritethetrial.Datawillbegincollectingwhenyouhit“Yes.”

13.Whendatacollectioniscompleted,turnoffcamerasandcopydatatoappropriatelocation(e.g.,externalharddrive,networkfolder,etc.).

Updated:16October2017 RoomCalorimeterCore 93

Chapter1.3-QTMDataProcessing

DataProcessinginQTM

Afterdataarecollected,theymustbetrackedandexportedfromQTMforfurtheranalysis.

Trackingmarkerdata

1.Navigatetolocationofmotioncapturedataonthecomputer.QTMfilesendin*.qtmandcanbeopenedonacomputerwithQTMsoftwareeitherbydouble-clickingonthefilefromWindowsExplorerorusingthe

“Openfile” buttoninQTM.

2.Beforetrackingyourdata,youmaywishtotrimthetrialtoonlytherangeofframesofinterest.Forexample,youmaywishtodothisifyouwantasinglestrideofdatafromatrialwhichincludesdataofyourparticipantwalkingallthewayacrossthedatacapturespace.

A.Toshortenatrial,clickanddragthearrowbuttonsoneitherendofthetimebar.

B.Forexample,hereIhavetrimmedoffthebeginningandendofatrialofdata.

C.Onceyoutrimatrial,markerinformationwillonlybedisplayedfortheframesinside thetimebararrows.Notethatthisrangecanalwaysbechangedlateronifyoudecideyou wanttotrackmoreofyourtrial.

3.Ifmarkersarealreadylabeled,skiptostep5.

Updated:16October2017 RoomCalorimeterCore 94

4.Ifmarkersarenotyetlabeled,alabellistwillneedtobecreated.Un-labeledmarkersappearredandalltrajectorieswillbelistedinthe“unidentifiedtrajectories”box.Ifyoudon’tseethetrajectoryboxes,clickView>TrajectoryInfo>Alltoopentheseboxes.

A.Right-clickinthe“Labeledtrajectories”windowandselect“Loadlabellist…” Selectyourmarkerlist.

I.Ifyouhavenotpreviouslycreatedalabellist,right-clickandselect “Addnewlabel.”Thiswillcreatea“New####”label.

II.Addasmanylabelsasmarkersinyourmarkersetandnamethem appropriately.

III.Whenalllabelsarecreated,right-clickinthe“Labeledtrajectories” windowandselect“Savelabellist…”

Updated:16October2017 RoomCalorimeterCore 95

B.Tobeginlabelingmarkers,selectthefirstmarkerinthelabellistandclickthe

button.

C.Inthe3Dview,clickonthemarkercorrespondingtothefirstlabel.InID mode,theprogramwillprogresstothenextlabelinthelabellistwhenamarkeris clickedon.

D.Continueclickingonthemarkersinthe3Dviewthatcorrespondtothelabel highlightedinthelabellistuntilyoureachtheendofthelabellist.

E.Clickthe buttontoturnoffIDmode.

F.Trimtrialsothatonlylabeleddataareincludedusingthebottomarrowofthetimebar:

G.Click“GenerateAIMmodel”button

E.Select“Createnewmodel”inthewindowthatpopsupandthenclick“Next.”

F.Inthenextwindow,verifythatthemodellookscorrect,andthenclick“Next.”

G.EnterafilenameforyourAIMmodelandclick“OK.”Inthefinalwindow,click “Finish.”

H.AftercreatinganAIMmodel,themodelcanbeappliedtonew,unlabeledfilesofthesame

participant/datacollectionsessionusingthe button.

Updated:16October2017 RoomCalorimeterCore 96

5.Ifmarkersarealreadylabeled,thenextstepistoclean-upthedata.Generally,thegoalofcleaning-upthedataistomakesuretherearenogapsinmarkerdata.

Note:itisrarethattherearenomarkersin“Unidentifiedtrajectories,”evenwhenallmarkersarelabeled.

Updated:16October2017 RoomCalorimeterCore 97

6.First,checkthatallmarkersarecorrectlylabeled(e.g.,theRightASISisactuallylabeledasRASIratherthanasRSHForsomethingelse).Tocheckthis,youcanclickonamarkerinthe“Labeledtrajectories”boxanditwillbehighlightedinthe3Dview(andviceversa).Alsonotethatlabelcolorsinthe“Labeledtrajectories”boxmatchthecolorofthemarkersinthe3Dview.

A.The“FillLevel”columnforeachmarkertellsyouwhatpercentofthetrialcontainsdataandalabelforthismarker.The“Range”columntellsyouduringwhichframesthelabeledmarkerappears.Attheendofdatatrackingmarkerdatayouwantallmarkersofinteresttobeat100%fortherangeofframesofinterest.

B.The+signnexttoamarkerindicatesthatthismarkertrajectoryismadeupofmultiple parts.Ifyouexpandbyclickingonthe+sign,youcanseehowmanypartsareinagiventrajectoryandwhatframestheyappearduring.

Updated:16October2017 RoomCalorimeterCore 98

7.Whenyouhaveconfirmedthatallmarkersarelabeledcorrectly,youcanbeginfixinggapsindata.Thereare2generalmethodsforfixinggapsindatawhicharedetailedbelow.

A.Option1:fillingagapinamarkertrajectorywhenthereisanunlabeledtrajectory available(bestoptionifavailable)

I.SometimesQTMdoesnotdoaperfectjobofidentifyingwhichmarkeris which,evenafteryoulabeldata.Thiswillresultinsectionsofdatawherethereis amarkerinthe3Dviewbutitisunlabeled(orinsomecases,islabeled incorrectly).Below,weseethattheLeftPSIS(LPSI)islabeledfor90%ofthe trial,exceptfortheframerange1960-2559.Thereisanunidentified trajectorythatcoincideswithmostofthisframerange.

II.ToaddthisunidentifiedtrajectorytotheLPSI,clickandholdoneitherthe unidentifiedmarkerinthe3DviewortheUnidentifiedtrajectoryinthe “Unidentifiedtrajectories”box,anddragittoLPSIinthe“Labeledtrajectories” box.

Updated:16October2017 RoomCalorimeterCore 99

III.NowtheunidentifiedtrajectoryhasbeenaddedtoLPSI.

IV.Sometimestheunidentifiedtrajectoryyouaretryingtoaddtofillagapina labeledmarkeroverlapswithexistingdataandyouwillgetthefollowing message:

a.Ifyouhit“OK,”aboxshowing3Dtrajectorieswillpopupinwhich youcanselecttherangeoftheunidentifiedmarkeryou’dliketokeep. Thistoolcanbeunwieldyanddifficulttouseattimes.

b.Ifyouhit“Cancel,”youcaninsteadmanuallyidentifytheoverlapping data.

Updated:16October2017 RoomCalorimeterCore 100

c.Intheexamplebelow,youcanseethattheunidentifiedtrajectorycontains2partsofdatawhichcovertherange1961-2558.LPSIalreadyhasdataforframes2553-2559.Todeterminewhichdatatokeep(thedataalreadyinLPSIfortheseframesorthedatafromtheunidentifiedtrajectoryfortheseframes),navigatetothisframerangeinthe3Dviewusingthetimebaratthebottomofthescreen.

d.Duringtheoverlappingframes,youshouldbeabletoseetwomarkers. Thesemayoverlapverycloselyinspace(asshownbelow)ortheymay not.

e.Atthispointyouneedtochoosewhichdatayouwanttokeep.Ifyouwishtokeepsectionsofonemarkerwhichgobeforeandafteranother(inthiscase,saywewanttotaketheunidentifiedmarkerfortherangebeforeandaftertheoverlappingLPSIdata),youmayneedtosplitonetrajectory.Todothis,navigatetothelasttimepointyou’dliketokeepofthetrajectoryofinterest,right-clickonthetrajectoryinthe“Labeledtrajectories”or“Unidentifiedtrajectories”box,andclick“Splitpartaftercurrentframe.”Thiswillcuttheselectedtrajectoryinto2parts.Youmayneedtodothisagainattheendoftherangeofinterest.Thenthesplitpartorpartscanbedraggedandaddedtothelabeledtrajectory.

B.Option2:gap-fillingagapinamarkerwhenthereisnounidentifiedtrajectorytoswapin

I.Below,weseeanexampleofaframeofdatawherethereisnoLPSI,andnounlabeledtrajectorytoswapintoitsplace.Incaseslikethis,youhavetoperformagapfill.

Updated:16October2017 RoomCalorimeterCore 101

II.Togapfilldata,selectthePartsoneithersideofthegap(below,therearedata from1-1959and1961-2552,butnodataforframe1960),right-click,andselect “Gapfilltrajectorywithpreview”

Updated:16October2017 RoomCalorimeterCore 102

III.Awindowdisplayingthe3Dtrajectoryoftheselectedmarkerwillpopupandshowhowthegapistobefilled.Followingtheinstructionsinthebottomofthiswindow,youcanzoominonthegapandadjusttherangeofframesyouwantthegapfilltospan.Notethatgapfillsmustbeginandendatlabeledmarkerdataasthefillisconstructedusingapolynomialfitbasedonexistingdata.

Gaprangeintrajectories(X,Y,andZ)

Rangeofgapbeingshown

Updated:16October2017 RoomCalorimeterCore 103

IV.Generalguidelinesforgapfillingdata:

a.Usecautionwhenfillinglargergaps(rangedependsonfrequencyof datacapture,speedofmovement,etc.).Gapfilleddataisnotrealdataand largegapfillsmaybeadistortedrepresentationofthetruemotionofa marker.

b.Adjustgapfillframessothattherearenotabruptshifts/discontinuitiesinthe3Dtrajectories.Oftentimesbeforeamarkerdropsoutandcreatesagap,thecamerasarehavingdifficultycalculatingitspositionandsothelast/firstframeortwoofexistingdatawillbeshifted.

c.Foranygapfilllargerthanafewframes,alwaysviewthe3Drepresentationofthegapfillbyplayingthroughthecorrespondingframeswhilewatchingthe3Dviewofyourdatatocheckforvisibleerrorsinthegapfilleddata.

8.Whenallmarkersarelabeledat100%,datatrackingiscomplete.

9.Afterdatatracking,datamustbeexportedfromQTMforfurtherprocessing.Filescanbeexportedinthefollowingformats:

-C3D:Visual3Dfileformat

-TSV:tab-delimiteddatathatcanbeviewedinexcelandreadintoavarietyof programs

-MAT:MatLabfileformat

-AVI:videofileofmarkerdata

A.Filescanbeexportedone-at-a-timeorinbulk.

I.Toexportfilesinbulk,clickFile>BatchProcess

a.WhenusingBatchProcess,thefirstwindowtopopupasksyouto selectthedesiredfiles.Afterselectingthedesiredfilesandclicking “Open,”awindowsimilartotheonebelowwillpopup.

Updated:16October2017 RoomCalorimeterCore 104

b.Inthe“Processing”tab(shownbelow),youcanselectwhichfileformatyouwanttoexporttoandalsosetadditionalprocessingoptions.Ifyouwanttoexportthedataintheformthatyoucurrentlyhave(e.g.,markerstracked,trialstrimmed,etc.),DONOTselectanyoptionsexceptforyourexportingfileformat.Otheroptionsmayundoworkthatyouhavealreadydone(e.g.,selecting“Trackthemeasurement”willre-trackallfilesinraw,unlabeledformat).

c.InthesecondaryBatchProcessingtab(above,thiswouldbethe“C3DExport”tabonthetopleft),youcanselectoptionsspecifictothatfileformat.

II.Toexportfilesone-at-a-time,clickFile>Export>[desiredfiletype].Using thisdialog,onlythesecondarywindowoptionsdescribedaboveinCwillpopup.

10.Afterexportingdata,processinginQTMiscomplete

Updated:16October2017 RoomCalorimeterCore 105

Chapter1.4-ExampleMarkerSets

TheHumanMotionCorehassomestandardmarkersetsthatmaybeusedformotioncaptureof

certaintasks.Thesemarkersetsincludefull-body,rightlowerextremity,leftlowerextremity,andheadwithFrankfortplane.Userscanfeelfreetocopythesemarkersetsandmodifyasneeded.Ingeneral,thefullbodyandlowerextremitymarkersetsareoptimizedforcalculationoflowerextremityjointmechanics.Amoredetailedmarkersetthanthefullbodymarkersetwouldbeneededtocalculateupperextremitymechanics,butthecurrentmarkersetcanbeusedtoaccountforsegmentsizes/massesforcenter-of-masscalculations.

Belowareimagesofthestandardmarkersetswithmarkerslabeled.Attheendofthisdocumentisadefinitionofthemarkerlabels.Notethatinthestandardmarkersets,markerlabelsareconsistentfrommarkersettomarkerset.

Updated:16October2017 RoomCalorimeterCore 106

UnilateralLowerExtremityThismarkersetexistsinarightsideandleftsideversion.Therightsideisshownbelow.Therearesomedifferencesinthemarkersonthelowerextremitybetweenthismodelandthefull-bodymodel.Thismodelhasafootmarkersetthatcanbeusedtocalculatewholefootandrearfootangles.Additionally,thismodelhasdifferentpelvicmarkerscomparedtothewholebodymodelwhichwillleadtodifferencepelvicsegmentdefinitionsbutnosubstantialdifferencesinsegmentmotioncalculations.

RASI LASI

LGRT RGRT

LPSI RPSI

RMCO RLCO

RMMA RLMA

R1MH R5MH

RTOE

RTHF RTHU

RTHL RTHR

RSHF RSHU

RSHL RSHR

RHLA

RHPR RTOE

RPSI RASI

RGRT

RHPR

RHLA RHMD

RMMA RLMA

R5MH

RTOE

Frontalviewofunilateralmarkerset

Closeupofrearfootofunilateralmarkerset

Posteriorviewofunilateralmarkerset

Updated:16October2017 RoomCalorimeterCore 107

RASI LASI LILC RILC

SLJ

RACJ LACJ

LLCO LMCO

RMCO RLCO

RMMA RLMA LLMA LMMA

L5MH L1MH LTOE RTOE

R1MH R5MH

FLH FRH

BRH BLH

TOPH

LELB RELB

RDUL

RDRA LDRA

LDUL LBTH

LHAN

LFIN RFIN

RHAN

RBTH

FullBody

Thismarkersetincludesthesamegenerallowerextremitymodelasthelowerextremitymarkersets,butaddsatrunk,arms,andahead.Notethatthismarkersetdoesnotincludethe3-markerrearfoot.Thiscanbeeasilyaddedasshowninthelower-extremityimage.

Frontalviewoffullbodymarkerset

Updated:16October2017 RoomCalorimeterCore 108

Viewofright/frontsideoffull-bodymarkerset

LTHF

LTHL

LTHU

LTHR

RTHF

RTHL

RTHU

RTHR

LSHF

LSHL

LSHU

LSHR RSHF

RSHL

RSHU

RSHR

Updated:16October2017 RoomCalorimeterCore 109

PosteriorViewoffull-bodymarkerset

RARF RARU

RARL RARR

RFAF

RFAL RFAU

RFAR LFAR LFAL LFAF

LARU

LFAU

LARF

LARR LARL

BLH

RACJ

SLJ

RHEEL LHEEL

LACJ

BRH

TOPH

Updated:16October2017 RoomCalorimeterCore 110

HeadMarkersetforGazeTracking

FLH FRH

BRH BLH

TOPH

LAUM

RZYA LZYA

RAUM LAUM RZYA

LZYA

RAUM

TOPH

BLH FLH

FRH

Updated:16October2017 RoomCalorimeterCore 111

MarkernamedefinitionsMarkerlabelsarelistedbysegment

Head:Crown/helmetTOPH–topofheadFRH–frontrightofheadBRH–backrightofheadFLH–frontleftofheadBLH–backleftofheadFrankfortplaneR/LAUM–right/leftauditorymeatusR/LZYA–right/leftzygomaticarchTrunk:R/LACJ–right/leftacromioclavicularjointUpperBackTrunkClusterLowerBackCluster(ifyouwantbothlumbar&thoracicspine)Arm:UpperarmrigidclusterR/LARU–right/leftarmupperR/LARF–right/leftarmfrontR/LARL–right/leftarmlowerF/LARR–right/leftarmrearForearmrigidclusterR/LFAU–right/leftforearmupperR/LFAF–right/leftforearmfrontR/LFAL–right/leftforearmlowerF/LFAR–right/leftforearmrearR/LDRA–right/leftdistalradius(radialstyloid)R/LDUL–right/leftdistalulna(ulnarstyloid)R/LHAN–right/lefthandR/LBTH–right/leftbaseofthumbR/LFIN–right/leftsecondfingerPelvis:R/LASI–right/leftanteriorsuperioriliacspine(ASIS)R/LILC–right/leftiliaccrestR/LPSI–right/leftposteriorsuperioriliacspine(PSIS)SLJ–sacrolumbarjointR/LGRT–right/leftgreatertrochanterLowerextremity:Thighrigidcluster

Updated:16October2017 RoomCalorimeterCore 112

R/LTHU–right/leftthighupperR/LTHF–right/leftthighfrontR/LTHL–right/leftthighlowerR/LTHR–right/leftthighrearShankrigidclusterR/LSHU–right/leftshankupperR/LSHF–right/leftshankfrontR/LSHL–right/leftshanklowerR/LSHR–right/leftshankrearR/LLCO–right/leftlateralfemoralepicondyleR/LMCO–right/leftmedialfemoralepicondyleR/LLMA–right/leftlateralmalleolusR/LMMA–right/leftmedialmalleolusR/L1MH–right/leftfirstmetatarsalphalangealjointR/L5MH–right/leftfifthmetatarsalphalangealjointR/LTOE–right/lefttoeR/LHEEL–right/leftheelR/LHPR–right/leftheelproximalR/LHMD–right/leftheelmedialR/LHLA–right/leftheellateral

CalibrationMarkersBothCalibration&TrackingMarkersTrackingMarkers

Updated:16October2017 RoomCalorimeterCore 113

CHAPTER1.5-Visual3DTutorial

Visual3D(V3D)isamotioncaptureanalysissoftwarepackagethatcanmakedataanalysisrelativelystraightforwardandstreamlined.C-Motion,thedevelopersofV3D,havecreatedahigh-qualitysetofdocumentationforthissoftware(https://www.c-motion.com/v3dwiki/index.php/Visual3D_Overview).Ifyouareafirst-timeuserofV3DitishighlyrecommendedthatyougothroughtheV3Dtutorials(https://www.c-motion.com/v3dwiki/index.php/Visual3D_Tutorials)tofullyappreciatethetoolsincludedinthisprogram,aswellastheimplicationsofincorrectlymodeling/computingyourresults.

I.V3DBasics

1.V3Dworkswithmotionfilesthatarein.c3dformat.MotioncapturedatacanbeexportedfromQTMinthisformatwhichwillincludeanymarker,analog,andforcedatacapturedinatrial.NOTE:V3Dcannotworkwithfiles/commands/tagsthathavespacesinthename.Neverusespacesinfilenames,markernames,etc.

2.ToloadrawdatafilesinV3D,clicktheopenbutton andselectthec3dfilesyouwishtoview.Whenfilesareloaded,theywillappearinthe“Workspace”tabinV3D.

Updated:16October2017 RoomCalorimeterCore 114

3.Toviewthesedatafiles,switchtothe“SignalsandEvents”tab.Hereyoucanselectwhichfiletoviewusingthedrop-downmenu,playorscrollthroughatrialusingthetoolsatthebottom,orviewrawTARGET(marker),ANALOG,orFORCEdata.

Drop-downboxtoviewtrials

Playbackcontrols

ORIGINAL(raw)targetdata

3Ddataview

Updated:16October2017 RoomCalorimeterCore 115

4.Ifyouleft-clickonanydataitem,aDataViewwindowwillpopuptoshowyoutrajectoriesandvaluesforthatdataitem.Youcanalsoright-clickonadataitemtoplotit.

5.Viewingrawdatahaslimitedutility.Generallyyouwillwanttobuildamodelandapplythismodelyourdatatrials.

II.ModelsinV3D

ModelsinV3Darethebasefilesfromwhichskeleton/segmentcharacteristicsareappliedtomotionfiles.Inordertocreateamodeltoapplytoyourmotionfiles,youmusthaveastaticcalibrationtrial.Forhumanmotion,especiallygait,thestatictrialisgenerallycapturedwhileanindividualstandsstillinanatomicposition.Theseinstructionswilltakeyouthroughsomegeneralstepsforcreatingamodelfromastatictrial.Thisisnottheonlywaytocreateamodel;furtherinstructionscanbefoundontheV3Dwiki.

TherearedemoV3Dfilesthatyoucanusetoexplorethefollowinginstructions.AcompletedanalysiscanbeviewedbyopeningR_LE_walk.cmo.Thefilesusedtobuildthisanalysisincludeonestatictrial(R_LE_static.c3d),twotreadmillwalkingtrials(R_LE_walk01.c3dandR_LE_walk02.c3d)amodelfile(R_LE_model.mdh),andapipeline(R_LE_walk_pipeline.v3s).ThesefilesarefordemonstrationonlyandcanbeusedtoexploresomecapabilitiesofV3DbutarenotinclusiveofalltasksthatV3Discapableof.

Therearesomepre-createdmodelsavailableforgeneraluse.Theseincludeunilaterallowerextremitymodelsforboththerightandleftlimbsandafull-bodymodel.Thesemodelsarebasedonthemarkersetsthatarealsoavailableforgeneraluseandcanbeusedandmodifiedasneeded.Thesemodelsarenotmeanttobeall-inclusivebutratherprovideexamplesandstartingpointsfordevelopmentofproject-specificmodels.NoteespeciallythattheupperextremitiesinthefullbodymodeldonotfollowISB

X,Y,Ztrajectory

Dataresidual

Dataidentity

Updated:16October2017 RoomCalorimeterCore 116

recommendationsfortheshoulder,elbow,orforearm.UsersshouldalwaysreviewtheV3Dwikirecommendationsformodelingofsegmentsandjointsabovethepelvis.

Generalfactsaboutmodels

ModelsinV3Dconsistofdefinitionsofsegmentsandlandmarksusedtocreatethesesegments.Thesedefinitionsareneededsothatthe3Dorientationofsegmentscanbetrackedduringamovement.Ingeneral,segmentsaredefinedusingrealorvirtually-definedmarkersattheproximalanddistalends,aswellasradiioftheproximalanddistalends.Atleastthreemarkersareneededtodefinetheorientationofasegment.Whenasegmentiscreated,itisassignedacoordinatesystem(seered,green,andblueorthogonalaxesinimagebelow)whichisrigidlyassociatedwiththesegment’sproximaljoint(e.g.,thethighcoordinatesystemwillbeanchoredatthehipjoint).Modelsmustalsoincludeassignmentoftrackingmarkers,ormarkersthatareregisteredtothesegmentdefinitioninacalibrationtrialandarethenusedtotrackthemovementofthesegmentduringmovementtrials.Thesegmentcoordinatesystemismodeledasrigidlyrelatedtotheorientationofthetrackingmarkersandforthisreasontrackingmarkersareoftenplacedonplasticshells(e.g.,thighandshankclusters)oronanatomicallocationswithlittleskinmotionartifact(e.g.,acromio-clavicularjointsduringlowerextremitytasks).Inadditiontosegment-specificdefinitions,modelsrequireanindividual’sheightandbodymassforcalculationofsegmentinertialpropertiesandjointkinetics.

Theorientationofadjacentsegmentcoordinatesystemswithrespecttoeachotherprovidesthejointangleorjointtranslationalposition.Forthisreason,itisimportanttopayattentiontotheorientationofsegmentcoordinatesystemsastheyarecreated.Intheimagebelow,youcanseethatallvisiblesegmentshavethered(X)axisorientedtothemodel’sleft.Thismatchingorientationprovideseaseofunderstandingforjointdefinitions.Forexample,rotationsaboutXbetweentheshankandthigharesagittalplanejointangles,thoseaboutYarefrontalplanejointangles,andthoseaboutZaretransverseplanejointangles.Inordertoobtainaccurateandmeaningfuldata,itiscriticalthatyoucorrectlydefineandconstructyourmodel.RecommendationsformodelingsegmentsandjointsofinterestcanbefoundontheV3Dwiki,aswellasinpublishedrecommendationsfromtheInternationalSocietyofBiomechanics.

Applyinganexistingmodeltomovementtrials

InanewV3Dworkspaceorwindow,onthemenu,click“Model,”then“Create(AddStaticCalibrationTrial),”and“HybridModelfromC3DFile.”

A.Inthewindowthatpopsup,selectyourstaticcalibrationtrial.

B.V3Dwillautomaticallyswitchtothe“Models”tab.

C.ClicktheLoadModelTemplatebutton

i.Selectthemodelfile(*.mdh)youwouldliketouse.

ii.Themodelshouldnowbeappliedtoyourstatictrial.

Updated:16October2017 RoomCalorimeterCore 117

iii.Insomecases,therewillbeerrorsinbuildingthemodel.Thesewillbedetailedina“BuildModelResults”window.

Updated:16October2017 RoomCalorimeterCore 118

iv.Ifthisoccurs,youcanscrollthroughtheerrorsand/orwarningstodetermine whatissuesarepresent.Fixingtheseissuesmayinvolvefindingacorrect, completestatictrial,correctingthenamesofmarkers,ormodifyingthemodelto correspondtodifferentsegment/landmarkdefinitions.

D.Definitionsofhowmodelsegmentsandlandmarksarecreatedcanbefoundbydouble-clickingonasegmentorlandmarkintheleft-handwindowintheModelstab.

Updated:16October2017 RoomCalorimeterCore 119

i.Forexample,ifyoulookattherightthighinthismodel,youwillseethefollowingdefinitions.

ii.Notethatsomeofthemarkersusedfordefiningtheproximalanddistaljointsarerealmarkersthatwereplacedonanindividual(e.g.,RLCO,RMCO)whileothermarkersare“landmarks,”orvirtualmarkerscreatedinV3Dbasedonthepositionofothermarkers.Here,theF_RHIPisalandmark.

Updated:16October2017 RoomCalorimeterCore 120

iii.IfwelookatF_RHIPinthe“Landmarks”tab,wecanseehowthislandmarkwascreated.Thislandmarkisahipjointcenterdefinitionbasedonliteratureregressionequations.

E.Afterapplyingyourmodel,theworkspaceshouldlooksomethinglikethis:

Updated:16October2017 RoomCalorimeterCore 121

F.Nowyoumustaddmotionfiles.Todothis,clicktheopenbutton andselect yourmotiontrials.Awarningwindowwillpopup,select“Insertnewfilesintoyour currentlyopenworkspace.”

i.Afterselectingyourmotiontrials,yourworkspacewilllooksomethinglikethis:

G.Youcannowapplyyourmodeltoyourmotionfiles.Todothis,clickModel>Assign ModeltoMotionFiles.Selectthetrialswhichyouwouldliketoapplythemodelto

Updated:16October2017 RoomCalorimeterCore 122

i.Nowyourworkspaceshouldshowyourmotionfileswithintheirassigned model.

ii.Ifyouviewyourmotiontrialsinthe“SignalsandEvents”tab,youwillseethe modelappliedtothetrials(assumingyourdataweretrackedcorrectlywith markernamesthemodelexpected.)Notethatthemodelisnotnecessarilytoscale orcompletelycorrectanatomically(e.g.,modelfemoralheadsareoftennotinthe acetabulum),themodelismeantasavisualaid.

iii.Ifasegmentismissingorincorrectlyorientedinamotiontrialitisdueto incorrectlylabeleddata.Checkthatalltrackingmarkersarepresentinthemotion trialandthatthemarkernamesassignedtothemotiontrialmatchthoseinthe model.

Updated:16October2017 RoomCalorimeterCore 123

III.DataAnalysisandPipelines

OnceyouhavedataloadedintoV3Dandamodelassignedtoyourmotionfiles,youcanprocessyoursignals,createmotionevents,andcomputeoutcomes(e.g.,jointkinematicsandkinetics).Thiscanbedoneusingsomepoint-and-clickstepsbutismoreefficientlyimplementedusingpipelines.Pipelinesconsistofaseriesofcommandswhichcanbesavedandusedformultipleparticipants/conditions/iterationsinastudyandthusassistwithbatchprocessingandcalculationconsistency.Thefollowinginstructionswillgothroughsomebasicpipelinecommandsthatmaybeusedduringanyanalysis.Asyouworkthroughpipelinecommandsitwillbecomeclearthattherearemanytoolsavailableinpipelines.FurtherinformationabouteachcommandcanbefoundintheV3Dwikidocumentation.Foranexamplepipeline,intheCPHMdemonstrationdataprovided,theanalysisoftheR_LE_walk.cmofilewasperformedusingtheR_LE_walk_pipeline.v3sfile.

1.Tousethepipelinetool,clickonthepipelinebutton. Thefollowingwindowwillpopup.

2.Inthelefthandpaneofthiswindowareallavailablecommands,organizedintosubgroups.Thecenterpaneshowscommandsinthecurrentpipeline,andtherighthandpaneshowsapreviewofacommandinthecurrentpipelinewhenitisselected.Forexample,ifyouweretostartcreatingapipelinewithacommandtocreateamodelfromastatictrial,thewindowwouldlooklikethis:

Updated:16October2017 RoomCalorimeterCore 124

3.Whenyouhaveacommandinacurrentpipelinehighlighted,theinitialtextofthatcommandaswellasadescriptionofthecommandwillappearintherighthandpane.Toeditthecommand,double-clickonthecommandinthemiddlepane.Forthe“Create_Hybrid_Model”command,double-clickingonthecommandwillbringupawindowoftextthatcanbeedited.Inthesetypesofwindows,“!”indicatesthatthepipelineshouldskipthislineofcodeorusethedefaultvalue.Inthewindowbelow,thismeansthatthereisnospecificfilethatthepipelineispulling,itwillinsteadopenadialogwindowaskingtheuserto“Openstandingfile.”Ifthereisaspecificfileyou’dliketoloadusingthiscommand,thepathtothisfilecanbeenteredinthe“/CALIBRATION_FILE=”lineandthe“!”canberemovedfromthisline.

Textofcommand

Descriptionofcommand

Updated:16October2017 RoomCalorimeterCore 125

4.Insomecases,therearetwodifferentformatsinwhichtoeditacommand.Forexamplewhenyoudouble-clickonthe“Event_Copy”command,itopensaGUI-typewindow.Ifyouprefertoeditthecommandintext

format,youcanclickonthecommandinthemiddlepaneofthepipelinewindowandclickthe button.

Updated:16October2017 RoomCalorimeterCore 126

5.Whenapipelineiscreatedandsaved,the.v3sfilecanbeopenedasatextdocumentforeaseofbatchediting(e.g.,changingexportfilenamesonasubject-by-subjectbasis).

6.Pipelinewindowbuttons:

A.“OpenPipeline…”allowsyoutoopenapreviously-createdpipeline.Ifyouclickthis buttonwhencommandsarealreadyinthemiddlepaneofthepipelinewindow,youwill begiventheoptiontoreplacethecurrentcommandsorappendthepipelineyouare openingtotheendofthecurrentcommands.

B.“SavePipelineAs…”allowsyoutosaveyourpipeline.Pipelinesdonotautomatically save.Ifyouclearthepipelineorreplacethepipelinebyopeninganotherpipeline,the pipelinewillnotbesavedandyouwillnotbepromptedtosaveit.

C.“ClearPipeline”clearsallcommandsfromthemiddlepaneofthepipelinewindow.

Updated:16October2017 RoomCalorimeterCore 127

D.“ClearParameters”willclearanyparametersenteredinpipelinecommands.

E.“Step”willexecutethecommandcurrentlyhighlightedinthemiddlepane.

F.“ExecutePipeline”willexecutetheentiretyofthecommandsinthemiddlepaneofthe pipelinewindowfromthefirstcommandtothelast.

G.Notethatthereisno“undo”buttonanywhereinV3D.

7.Keypipelinecommands.Userscanfollowalonginthe“R_LE_walk_pipeline”forexamplesoftheimplementationofeachofthesecommands.

A.Create_Hybrid_Model+Apply_Model_Template:whilemodelscanbecreatedandappliedusingthepointandclickstepsdetailedabove,itisoftenusefultoautomatethesestepsonceamodelhasbeencreatedandpiloted.The“Create_Hybrid_Model”commandwillpromptausertoopenastandingcalibrationtrialandloadthistrialintotheV3Dworkspace.The“Apply_Model_Template”commandwillprompttheusertoselectamodel(.mdh)fileandthenapplythismodeltothestandingcalibrationtrial.

B.File_Open:thiscommandpromptstheusertoopenfileswhichwillbeloadedintothe V3Dworkspace.Inthe“R_LE_walk_pipeline,”thiscommandisusedtoloadmovement trialsafterthecreationofthemodel.

C.Assign_Model_File:thiscommandwillprompttheusertoassignthemodelfileto movementtrials.

D.Lowpass_Filter:target(marker)andanalogdataoftenneedtobefilteredtoremove noise.TherearemanyfilteroptionsinV3Dpipelines(seethe“SignalFilter”sectioninthelefthandpaneofthepipelinewindow).Ingeneral,rawdatashouldbefilteredbeforeperformingcalculations.Whichfiltersshouldbeusedandatwhatstepoftheanalysistheyshouldbeimplementeddependsonthetaskbeingperformedandtheanalysisrequired.Inthe“R_LE_walk_pipeline”example,alowpassButterworthfilterisappliedtothemarkerdatatoremovesignalabove8Hz.The“NUM_REFLECTED”lineinthecommandprovides100reflectedpointsoneithersideofthedatatominimizetimeshiftsinthedataduetothefilter.Notethatwhenthetargetdataarefiltered,thefiltereddataareplacedinanewfolder:TARGET>PROCESSED,whiletheoriginal,unfiltereddataremainintheTARGET>ORIGINALfolder.

E.Interpolate:thiscommandwillfilldatagapsofaspecifiedsize.Thisshouldnotbe usedinlieuoftrackingdatainQTMbutratherasafinalsweeptocorrectanysmallgaps thatweremissedintracking.

F.Select_Active_File:thiscommandcanbeusedtoselectallfilesorgroupsoffilesfor furtherpipelinecommands.Whenpairedwithanearlier“Assign_Tags_To_Files,”the Select_Active_Filecommandcanbeusedtoperformpipelinecommandsononlyasubsetoffiles.

G.Eventcommands(e.g.,Event_Minimum,Event_Maximum,Event_Threshold):canbeusedtoautomaticallycreateeventssuchasheel-strikes,toe-offs,maximumflexion,etc. Thesecommandscanbefoundinthe“EventCreation”and“EventManagement” sectionsofthelefthandpaneofthepipelinewindow.Notethatitisoftenmostusefulto editthesecommandsusingthe“Text”optionratherthanthedefaultGUIwindow.

Updated:16October2017 RoomCalorimeterCore 128

H.Forloops:the“For_Each”and“End_For_Each”commandscanbeusedtobeginand endforloops.Inthe“R_LE_walk_pipeline”example,theseareusedtorecursively renamegaitcycleevents.

I.Signalmathcommands:thesecommandscanbeusedtoperformsimplearithmeticonsignals.Forexample,inthe“R_LE_walk_pipeline,”thenumberofframesbetweenconsecutiveheelstrikesandbetweenheelstrikesandtoe-offsareusedtodeterminegaitcycletimes(usingSubtract_Signals)and,subsequently,thepercentofthegaitcyclespentinthestancephase(usingDivid_Signals).

J.Metriccommands:thesecommandscreatenewsignalsinthe“METRIC”folder.

K.Compute_Model_Based_Data:thiscommandisperhapsthemostimportantcommandinV3Dpipelines.Withit,ausercancomputeahugerangeofvariables.Wheneditingthiscommandthefollowingwindowappears.Init,younamethedatavariableyouarecreating,identifywhattypeofvariableyouarecalculating,anddefinethedatabeing usedtocalculatethisnewvariable.TherearemanydifferenttypesofvariablesthatcanbecreatedandthespecificsofthealgorithmsusedorcomputationalordercanbefoundintheV3Donlinedocumentation.The“UseNegative”optionsallowyoutomultiplyasignalby-1asyoucalculateit,whichiscommonlyusedtoaligndatadirectionswithcommonreportingpractices(e.g.,theright-handrulewillgenerallyresultinkneeextensionbeingapositivevalue,whilereportingpracticesgenerallyfollowthatkneeflexionshouldbepositive).CardanSequenceshouldgenerallyfollowfromaxisoflargestexpectedexcursiontoaxisofsmallestexpectedexcursionforthemovementofinterest.Inthebelowexamplewindow,fortherighthipduringwalkingwewouldexpectthelargestexcursiontobeinthesagittal,orX,axis,followedbyfrontal(Yaxis)andtransverse(Z)motion.

Datavariableyou’recreating(musttypeinname)

Whattypeofvariableyou’recreating

Definitionsforvariablecalculation

Updated:16October2017 RoomCalorimeterCore 129

L.Export_Data_To_Ascii_File:thiscommandisusedtoexportdatatoatab-delimitedtextfilewhichcanbereadinexcelorimportedintoafurtherdataanalysisprogram,suchasMatLab.IntheGUIwindowwhichpopsupwhenyoueditthiscommand,youcandesignatethepathforyourexportedAsciifile,selectoptionssuchasexportingthemeanandstandarddeviationofyourdata(ifyouhavemultipletrials),normalizeyourdata(tomakealltrialsthesamerelativelength),amongotheroptions.Whenyouselectthesignaltype,folder,name,andcomponentyou’dliketoexport,youmustclick“AddaNewSignal”tosavetheseselectionsintothecommand.Likewise,ifyouwishtoeditanexistingcommand,youmustclickthe“ModifySelected”buttonaftermakingyoureditstosavethem.Intheexamplefrom“R_LE_walk_pipeline,”notethateachstrideofdatamusthaveaseparatelydesignatedeventsequenceandthusaseparatelineintheGUI.Ifyourunthiscommand(withanappropriatelyrenamedfilenamepath),youcanopentheexportedfileinexceltoviewtheformatofV3DAsciifiles.If“ExportMeanandStandardDeviation”isselected,themeanandstandarddeviationofalllinesintheGUIwillbecomputed,andthesewillappearascolumns2-7intheAsciiexport.Notethatifyoure-runthe“Export_Data_To_Ascii_File”commandwithoutchangingthe“ExportedAsciiFile”name,youwilloverwriteanyexistingfileatthatfileaddress.V3Dwillnotgivea“areyousureyouwishtooverwrite…”warningdialogbeforeoverwritingexistingdata.

Updated:16October2017 RoomCalorimeterCore 130

IV.Reports

V3Dcancreatereportstoassistwithdatavisualization.Thesereportscanplotvariablesthathavebeencalculatedandcangenerallybeusefulforcheckingcalculations.Areportisincludedinthe“Reports”tabofthe“R_LE_walk.cmo”fileandthecorrespondingreporttemplatefileisincludedindemonstrationfilesas“R_LE_report_template.rgt.”Thisreportconsistsentirelyof2Dgraphsofjointkinematics.Reportsarealsoabletogeneratetables,bargraphs,andothergraphics.Belowaresomeguidelinesforcreatingareport.

1.Whencreatinganewreportfromscratch,firstdesignatethelocationoftheitem(graph/graphic)youwishtoadd,selectthetypeofitemyouwishtoadd,andclickthe“Add”button.

Locationofitemtoadd

Itemtoadd

Updated:16October2017 RoomCalorimeterCore 131

2.An“Add/ModifyGraph”windowwillpopup.Hereyoucandesignatewhatdatatographinwhichposition.Belowisanexampleofthiswindowpopulatedforplottingonestrideofsagittalhipangledata.

3.Inareportwithmultiplemovementcyclesandmultiplesvariablesofinterest,therewillbemanylinesinthereportgenerationdialog.Alongwithmultiplecolumnsandrowsofplotsonasinglepage,reportscancontainmultiplepages,allowingforeasyvisualizationoflotsofdata.Ifyouareinterestedingeneratingreports,itisrecommendedthatyouexplorethedocumentationonreportsintheV3Dwiki(https://www.c-motion.com/v3dwiki/index.php/Tutorial:_Creating_a_Report)forfurtherinformationandexamples.

Updated:16October2017 RoomCalorimeterCore 132

UNIVERSITYOFMASSACHUSETTSAMHERST

INSTITUTEFORAPPLIEDLIFESCIENCE

HumanTestingCenter

HumanMotionCore

Chapter2:DelsysTrignoEMG

1. GeneralInformationaboutTrignoSensors1.1. Sensors1-8haveEMG+TriaxialAccelerometersensors.

1.1.1. WhenusedinconjunctionwithQTMthesensorshavecollectionfrequenciesofEMG1925.25Hz(upsampledto2000Hz)and148.48Hz(upsampledto??).

1.2. Sensors9-16haveEMG+TriaxialAccelerometer,GyroscopesandMagnetometers.1.2.1. WhenusedinconjunctionwithQTMthesensorshavecollectionfrequenciesofEMG1925.25Hz

(upsampledto2000Hz)and148.48Hz(upsampledto??).

2. ConfiguringtheTrignoSensors

2.1. Openthe‘DelsysTrignoContorlUtility’2.2. Inthelowerrightcornerofthescreenpress‘Configure’

Addtherestoftheconfigurationscreens–whattheyarehowtheywork.

Updated:16October2017 RoomCalorimeterCore 133

3. PairingSensors.

3.1. Pairthesensorsyouwouldliketouse3.1.1. Pushthepairbuttonin“DelsysTrignoControlUtility’software.

3.1.2. Pushthebuttononthesensoruntilyouseeaflashinggreenlight.Youshouldalsoreceivea

confirmationonthecomputerscreenthatthesensorhassuccessfullypaired

4. Placementandorientationofsensors5. EnablingSensorsinQTM

5.1. InputDevice

6. LabelingSensorsinQTM

Updated:16October2017 RoomCalorimeterCore 134

7. AccessingSensordatainQTM

7.1. Sensordatacanbeviewedinrealtime

Updated:16October2017 LivingScienceCore 1

MANUALOFPROCEDURES

LivingScienceCore

InstituteofAppliedLifeSciences

LifeScienceLaboratoriesS360240ThatcherRoad,

Amherst,MA01003

atthe

UniversityofMassachusetts,Amherst

Updated:16October2017 RoomCalorimeterCore 2

UNIVERSITYOFMASSACHUSETTSAMHERST

INSTITUTEFORAPPLIEDLIFESCIENCE

HumanTestingCenter

LivingScienceCore

Chapter1:OxyconMobile

1.1 Introduction:TheOxyconmobileisaportableindirectcalorimetryunitthatmeasuresmetabolicresponsestoexercise,workorotheractivitiesthatstationarymetabolicunitscannotmeasure.Thewirelesscapabilitiesofthisunitoperateviatelemetry,upto1,000minlineofsightrange.TheOxyconMobileaffordsuserstheopportunityforon-the-spotrecordingofcardiopulmonaryanddeducedparameterssuchas:VO2,VCO2,RER,VEO2,VECO2,PETO2.VE,VT,BF,BR,andSpO2,etc.Researcherscanchoosetomonitorparametersineither,breath-by-breathorintrabreathepochs.Additionally,researcherscanmonitorthesedatainreal-time(viatelemetry),orposthoc,asdataarestoredtoaninternalSDmemorycard.

CoreCompetencies:

1. Beabletofollowcalibrationinstructions2. Understandhowtooutfitparticipantwithequipotent3. Reviewstepsforsettingup,collecting,andreviewingsessiondatathroughascreenreport4. Familiarizeyourselfwithhowtogeneratea.CSVreport

1.2 HowtoCalibrate:Belowisastep-by-stepoutlineforcalibratingtheunit.

PreliminaryCalibrationSteps1.) RetrieveComputerfromOxyconcase

a. Username:labsup Password:lab2.) Collectreceiverunit(picturedbelow)

Updated:16October2017 RoomCalorimeterCore 3

3.) PluginpowerandconnectcomputertoreceiverunitusinggreyUSBcable

Updated:16October2017 RoomCalorimeterCore 4

4.) Grabthetwopacks(SBX,DEX)andamemorycarda. SBX:Sensorbox(O2&CO2sensor)b. DEX:Dataexchangeunit

Powercord

Updated:16October2017 RoomCalorimeterCore 5

5.) WarmupSBXbyconnectingittothereceiverwithyellowEthernetcable

6.) Openlabmanager

a. Allowwarm-upfor15minutes

Updated:16October2017 RoomCalorimeterCore 6

7.) Assemblenecessarycomponentsofmaska. Attachturbinetoturbineholder(metal(silver)sideinfirst)

b. Makesureturbineisnotwetc. Connectinfraredsensortoturbine

d. Besuresamplingtubeisattachedtotheturbineholder(oppositeendofinfrared)

Updated:16October2017 RoomCalorimeterCore 7

8.) AttachmaskcomponentsandtightensamplingtubeandinfraredsensortoSBX

Updated:16October2017 RoomCalorimeterCore 8

Updated:16October2017 RoomCalorimeterCore 9

9.) PlaceappropriatestudymemorycardinDEXunitandaddbattery(makesureitischarged)

a. IfDEXdoesn’tturnoffrightaway,pressoffbutton

10.) After15minutes,labmanagershouldopenuptoahomescreen

Updated:16October2017 RoomCalorimeterCore 10

11.) ConnectSBXtoDEXusingyellowEthernetcable

a. AfterafewsecondsofSBXvibrating,alightshouldflashonSBXb. Ifthebatteryislow,SBXwillbeepandanotewillpopuponthescreenc. BesuregreenLED’sonbothunitsarelitpriortocalibration

1.3OxyconMobileCalibration

1. ClickAmbientconditionsonthelabmanager

Updated:16October2017 RoomCalorimeterCore 11

a. DEXsensesi. Usuallyaccuratetemperatureandbarometricpressureii. Manuallyenterhumidity

b. CheckgreyboxinthelaboratoryontheParvometaboliccartforenvironmentalconditions

i. PressF1tochangetemperature,barometricpressureandhumidityii. Leavebarometricpressureasis

c. PressF12tosave

2. ConnectSBXbacktoreceiverusingyellowEthernetcablea. WaitforSBXtogetpower(willvibrateandLEDlightsupgreenwhenreadytocontinue)

3. SetupforAutoVolumeCalibration

a. ClickAutovolumecalibrationsfromcalibrationsgroup

Updated:16October2017 RoomCalorimeterCore 12

b. Waitforgraphtocomeup

4. Connectwhitesamplingtube,intoreceiverandSBX

Updated:16October2017 RoomCalorimeterCore 13

5. ConnectmouthpiecewithturbineintoPCA(blackhole)withsampletubeholecovered

6. PressF1tobeginAutoVolumeCalibration

Updated:16October2017 RoomCalorimeterCore 14

a. If%differenceislessthan3%pressF12tosaveb. If%differenceisgreaterthan3%,checklinesandconnectionsandretry

7. GasAnalyzerCalibration

a. Connectgascylinderintounit,pushclipuntilitislocked

Updated:16October2017 RoomCalorimeterCore 15

b. Opengascylinderbytwistingthevalvefromthecalibrationgastwocompleteturns

i. Cylindercontains15.98%O2and3.9%CO2concentrations

c. ClickonGasAnalyzerCalibrationinthelabmanager

Updated:16October2017 RoomCalorimeterCore 16

d. Clickonthesettingstabatthetopandenterinanalysisresultsfromthe“CertificateofAnalysis”providedbyAirgas,thissheetshouldbewiththegastanks

e. Then,pressF1tobegin,Graphwillappearandcalibrationwillbeginf. Calibrationwillsampleroomairalongwithcylinderair

g. O2peaksat21andvalley;bluelineh. CO2willvalleyat0andthenpeak;redlinei. Ifpercentdifferencelessthan3%pressF12andsavej. Turnofaircylinderk. Ifpercenterrorgreaterthan3%,repeatprocess

8. ReassembleSBXandDEXwithyellowEthernet

Updated:16October2017 RoomCalorimeterCore 17

9. Reassemblemouthpiecewithsamplingtubea. PlacenewbatteryinDEX

Updated:16October2017 RoomCalorimeterCore 18

1.4TestingwiththeOxyconMobile

1. Inputpatientdataa. PressF2tocreatenewsubjectb. BesuretofilloutLastName,FirstName,Identification,DOB,Gender,HeightandWeight

(pressentertonavigatethroughpatientinputs)

c. PressF12tosavecurrentsubjectinformation

2. Sizesubjectforheartratebelta. Wetsensorsb. Attachsensorstobeltc. Subjectputsonheartratemonitorensuringthesensorisatthesternum

3. Sizesubjectformask

a. Havesubjectputmaskonandcoverbreathingholeb. Havethesubjectblowtomakesuretherearenoleaksc. Attachthefourclipstothemaskd. Havethesubjectholdthemaskinplacewhiletheresearcherclipstheblackheadgear

nettothemask(taggoesinfront)

4. Sizesubjectforbackpacka. AttachtheSBXandDEXtothebackpack(wiresfacingup)b. Subjectputsonbackpackc. Attachmaskwithwhitesamplingtube(whitesamplingtubeshouldbeclosesttothe

subjectsface)

Updated:16October2017 RoomCalorimeterCore 19

5. Tobegindatacollection,clickbreathbybreath

a. MakesureDEXandSBXhavegreenlights

i. Ifyellow,flipSBXandDEX,checkHRbeltposition,checkbattery

6. EnsureallPatientDataiscorrect

7. CheckTestsettings

Updated:16October2017 RoomCalorimeterCore 20

a. Ergometer-Modalitymustbebicycleb. DeadSpace-clickappropriatesizecorrespondingtosubject’smasksizec. Averaging-Method,Breaths

8. PressF1orplaytoanalyzebaselineinformation(backgroundzeroing,notedinbottomrightcorner)

9. PressF1orplayagaintostartrecordingdata10. Afterdatacollectioniscompletebesuretheupperleftbutton(play)isnowa“stop”buttonto

endrecording11. Press12toexitandsave

1.5ReviewingtheData

1. ClickonScreenReporttoviewcollecteddata

2. Fromthereportlist,choosethereportviewyouwouldlike(studyspecific,formoreinformationon

creatingnewdatavisualizationtemplatesseebelow)

Updated:16October2017 RoomCalorimeterCore 21

3. PressOKtogeneratethereport

4. ToexportusabledatapressF9andsaveonalocaldrive

5. ToimportfromstoragedeviceselectFlashdiskreader

Updated:16October2017 RoomCalorimeterCore 22

6. Importthefileyouwouldliketoreview7. MakesuretheCurrentpatientmatchesthefileyouaretryingtoimport8. PressF1toreadfile9. AfterfileisloadedpressF12tosave10. RepeatSteps1-411. ToviewsavedfileinExcel,openExcelandgotoFile-Open12. Findthesavedfile13. WhenpromptedbytheImportWizard,MakesureImportRowisat1andFixedWidthisselected

14. ClickNext

Updated:16October2017 RoomCalorimeterCore 23

15. Whenpromptedwithtextimport,alsoclicknext

16. Onthefinalstep,Clickadvancedandmakesureseparatorsarecorrect

17. Afterimportiscomplete,selectSaveasandbesuretocheckthecorrectfileextension

Updated:16October2017 RoomCalorimeterCore 24

1.6CleaningEquipment1. Disassembleallpiecestobecleaned2. Wipeturbineholderandinfraredsensorwithalcoholwipes.3. Headgearstrapsandheartratemonitorsshouldberinsedandhunguptodry4. Masksshouldbesoakedina25:1cleaningsolutionfor30secondsandrinsedthoroughly5. Theturbinemustbesoakedinaspecialsolution

a. Inthemarkedcontainer,fillthetothebrimwithSterilizingandDisinfectingsolutionb. Intheclearplasticcuplocatednearby,fillthecuptothefirstindent(blackline)withthe

Activatorsolutionc. PourtheActivatorsolutioninthesterilizingsolutionandmixthoroughlyd. Placetheplasticgrateintothesolutionwiththeturbineinsidee. Allowtheturbinetositfor15minutesf. Removefromsolutionandrinsewithwaterg. Letairdry

1.7CreatingaNewDataVisualizationTemplate1. Clickontheutilitiestab2. Enterpassword:abcd3. Clickonreportdesigner4. Namenewreport,outputunit:defaultprinter(JLABpdfprinter)5. Clickok6. Clickonthetableattheright(hasF3onbutton)

Comment[MB3]:Isthisactuallythepassword?

Updated:16October2017 RoomCalorimeterCore 25

7. ChooseOxycontable,thenclicksummary8. Hereyoucanchoosemeasurementprogram(breath-by-breath,endtidal,cardiacoutput,intrabreath,

etc.)9. Dependingonthemeasurementprogramselectedyoucanaddparameterstoyoursummary,dothis

byhighlightingaparameteryouareinterestedinandclick“Insert”.Thiswilladdittoyourscreenreportsummary

10. Byfurthernavigatingthroughthetabsatthetopofthedialoguebox,youcanselectaveragingtype(intervals),layout,marker,andsetupsettings

11. PressF12tosavethereport,andnamethereport.Note:thenameyougivethereportwillbehowit

ispresentedinthescreenreportsection(seeReviewingtheDatatore-familiarizeyourselfwiththisstep)

Updated:16October2017 RoomCalorimeterCore 26

1.8Contacts

1. KimatCarefusion:18002312466Operatorwillpromptyouthroughascreener,click1atfirstquestion,click2atsecondquestion,click3atthirdquestion.

Note:Havereceiverunitonhandwhenyoucall,ithasserialnumbertheywillneedwhenassistingyou.

2. MichaelBusa–CoreDirector413-577-0574

[email protected]

Updated:16October2017 RoomCalorimeterCore 27

UNIVERSITYOFMASSACHUSETTSAMHERST

INSTITUTEFORAPPLIEDLIFESCIENCE

HumanTestingCenter

LivingScienceCore

Chapter2:ActigraphLink(GT9X)

2.1Supplies

1) ActiGraphGTXaccelerometerchargingcable2) ActiGraphdock3) ActiLife6software4) ComputertorunActiLife6software

2.2DeviceCharging

TheActiGraphGT9XaccelerometercontainsarechargeableLithiumPolymerbatterywhichiscapableofprovidingpowerforcontinuousdatacollectionperiodsof10ormoredayswithoutrecharging.Ifthebatteryhasbeenfullydepletedatthetimeofcharging,itwilltakeapproximately3-4hoursfortheActiGraphGT9Xaccelerometertoreachfullcharge.ChargingProcedures

1) ConnectActiGraphdocktocomputerorwalloutletusingminiUSBcable2) PlugtheActiGraphLinkintothedockwiththeActiGraphlogofacingup.Once

connected,theredLEDlightontherightsideofthedockwillturnyellow,thedevicescreenwilldisplaytheserialnumber,andthebattericonwillblinktoindicatecharging.

3) Oncethedeviceisfullycharged,theyellowlightwillturngreenandthebatteryicononthedevicewillshowasfullandstayonsteady

2.3DeviceInitializationTheActiGraphGT9XaccelerometersareinitializedwiththeActiLife6softwarepackage.TheActiGraphGT9XaccelerometerwillnotcollectdataunlessitisinitializedusingtheActiLifesoftware.

Updated:16October2017 RoomCalorimeterCore 28

InitializationProcedures

1) OpentheActiLife6softwarebydouble-clickingonthedesktopiconorbysingle-clickingonthestartmenushortcutfortheprogram.

2) PlugthechargingcableintoastandardUSBportonthecomputerrunningtheActiLife6

software.3) Selecttheaccelerometerthatwillbeinitializedbycheckingtheboxbesidethe

accelerometerdevicenameandserialnumber.

4) Fromthe“Initialize”buttonchoose“RegularInitialization”.Thiswillopentheprimary

initializationwindow.

1)

3)

Updated:16October2017 RoomCalorimeterCore 29

5) Selectasamplingrate6) Selectidlesleepmode.7) Note:SelectWirelessOption,heartrate,sleepmode,andIMU8) Clickthe“EnterSubjectInfo”button.Thiswillopenthesecondaryinitializationwindow.

6)

7)

8)

4)

5)

Updated:16October2017 RoomCalorimeterCore 30

9) Selectastartdateandtimeatthetimesubjectorientationisscheduledtobegin10) Selectastopdateandtimetostopdatacollectionataspecifiedtime(canbeleftblank)

11) Inputanappropriateidentifier(e.g.,“S#”)inthe“SubjectName”field.12) Inputtheappropriate“Limb”and“Side”information(e.g.righthiporleftwrist)13) Clickthe“Initialize1Device”button.Thiswillclosethewindowandreturntothemain

softwarewindow.14) Ifproperlyinitialized,thewords“finishedinitializing”shouldappearunderneaththe

“Status”column.

15) TheActiGraphGT9Xaccelerometerisnowreadyfordatacollection.Unplugaccelerometerfromthechargingcable.

9)

11)

12)

13)

Updated:16October2017 RoomCalorimeterCore 31

2.4AccelerometerPlacement

TheActiGraphGT9Xcanbewornatthehip(aroundthewaist)andonthewrist.Elasticbeltsandwriststrapswillbeusedtosecuretheaccelerometerstotheparticipant’sbody(seepicturesfordetails).

Securedaroundthewaistattherighthipandattachedtoanelasticbelt.MakesurethattheGT9Xispositioneddirectlyonthebody’sside(mid-axillaryline)

2.5DataRetrievalandDownload

ThepreviouslydescribedActiLife6softwareutilityisrequiredtodownloaddatafromtheActiGraphGT9Xaccelerometer.

Updated:16October2017 RoomCalorimeterCore 32

DownloadProcedures1) OpentheActiLife6softwarebydouble-clickingonthedesktopiconorbysingle-clickingonthestart

menushortcutfortheprogram.

2) PlugtheActiGraphLinkintothedockwiththeActiGraphgoldplatescontactingthegoldpinsofthedock.3) PlugthechargingcableintoastandardUSBportonthecomputerrunningtheActiLife6software.4) Selecttheaccelerometerthatwillbeinitializedbycheckingtheboxbesidetheaccelerometerdevice

nameandserialnumber.

5) Clickonthe“Download”button.Thiswillopentheprimarydownloadwindow.Click“ChangeLocation”toselectdestinationfile

6) Withinthe“GT9XDownloadOptions”box,makesurethe“CreateAGDFile”boxischeckedandspecifythe“Epoch”.

7) Checktheappropriateboxesforyourprotocol.

1)

4)

5)

Updated:16October2017 RoomCalorimeterCore 33

8) Clickthe“DownloadAllDevices”button.Theprimarydownloadwindowwillcloseandreturntothemain

softwarewindowanddialogboxwillappearaskingyoutonamethefile.Thedownloadprocessmaytakeseveralminutes.

5)

5))

8)

7)

6)

9)

Updated:16October2017 RoomCalorimeterCore 34

9) Ifproperlydatawasproperlydownloaded,thewords“finisheddownloading”shouldappearunderneath

the“Status”column.

10) Nowitisnecessarytoprocessthedefaultfilter.agdfile.Select“Scoring”>Checkallitem>click“RemoveSelected”>Click“AddDataset”>SelectthepreviouslydownloadedAGDFile

11) Click“Calculate”12) Goto“Details”onthecalculatedfileandclick“ExportAllEpochs”13) Goto“OpenContainingFolder”,renamethefile(S#_ActiGraph_Location.csv)andputitinthe

appropriatefolder

10)

Updated:16October2017 RoomCalorimeterCore 1

MANUALOFPROCEDURES

RoomCalorimeterCore

InstituteofAppliedLifeSciences

LifeScienceLaboratoriesS360240ThatcherRoad,

Amherst,MA01003

atthe

UniversityofMassachusetts,Amherst

Updated:16October2017 RoomCalorimeterCore 2

UNIVERSITYOFMASSACHUSETTSAMHERST

INSTITUTEFORAPPLIEDLIFESCIENCE

HumanTestingCenter

RoomCalorimeterCore

Chapter1:QuickGuideforCalibration/Validation

Supplies

QuickGuide:CalorimeterRoomOperation

GasTankCheckListPriortocalibrationtestsandvalidationstudies,checktankpressures.

Tank MinimumPressure Tests

CO2 100psi Infusion,ShortCircuit,BlenderCalibration

O2 800psiforblendercal,NAforinfusion

Blendercalibration

StartCalRQ Login

SelectMode

HumanorValidationStudy Calibration

CalCheck

Complet

NullTest

No

Yes

RunStudy NullTest

Figure 1. Study Setup Flow

Updated:16October2017 RoomCalorimeterCore 3

N2 1500forblendercal,800psiforInfusion,500forShortCircuit

Infusion,ShortCircuit,BlenderCalibration

ZeroTank 400psi TankCalibration,CalCheck

SpanTank 400psi TankCalibration,CalCheck

RefTank 400psi Alltests–alwaysopen

Panelinstrumentsettingsforalltests StdSetting

1 Chamber1O2Rotameter 1.25L/min

2 Chamber1CO2Rotameter 1.25L/min

3 Chamber2O2Rotameter 1.25L/min

4 Chamber2CO2Rotameter 1.25L/min

5 InflowO2Rotameter 1.25L/min

6 InflowCO2Rotameter 1.25L/min

7 Chamber1SamplePressureGauge 2psi

8 Chamber1SamplePressureRegulator 2psi

9 Chamber2SamplePressureGauge 2psi

10 Chamber2SamplePressureRegulator 2psi

11 InflowSamplePressureGauge 2psi

12 InflowSamplePressureRegulator 2psi

Updated:16October2017 RoomCalorimeterCore 4

Updated:16October2017 RoomCalorimeterCore 5

AnalyzerCalibration

CalCheck:flowzeroandspangasestocheckvalues

AnalyzerGain:Subtracttheapparentconcentrationonthecomputerofthespanvaluefromtheapparentconcentrationonthecomputerofthezerovalue.Thisdeltashouldmatchthedeltaofthespanfromtheblenderortanks(Span(20.95)–Zero(20.05)=0.9delta).

AnalyzerCalibration:Flowzeroandspangases,changeslopeandoffsetvaluesintheanalyzer.Ifanalyzervaluesareoutofrange(<19.8%or>21.2%forO2;>1.1%forCO2)performcalibrationonanalyzers,whenanalyzersarestable,performcalibrationonsoftware,ifneeded.

SoftwareCalibration:Flowzeroandspangases,changeslopeandoffsetvaluesinthesoftware

Delta:Thedifferencebetweenthespanvalueandthezerovalue.Subtracttheapparentconcentrationofthespanvaluedisplayedonthecomputerfromtheapparentconcentrationonthecomputerofthezerovalue.

𝐷𝑒𝑙𝑡𝑎 = 𝑧𝑒𝑟𝑜 − 𝑠𝑝𝑎𝑛 eq(1)0.9 = 20.95 − 20.05 eq(2)

Datafileisrecordingwhentheloggingbutton(J)displays“LoggingOn”andEnterNotebox(K)appears.Filerecordingwillstopwhenloggingbuttondisplays“LoggingOff”or“Done”(L)ispressed.“Done”alsoclosesthewindow.

A

B

C

ED

C

ED

F

IHG

KJ

LFigure2.CalibrationScreen

Updated:16October2017 RoomCalorimeterCore 6

BlenderCalibration:

1. OpenN2,CO2andO2tanks2. FlipInstrumentationPanelvalvestomatchthoseshowninFigure3.Flipblendervalvesfor

calibration.3. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer4. Oncethegasisflowing,watchtheO2(A)andCO2(B)graphsuntilthevaluesaresteady.Turn

Averagingon,andletblenderrunforatleastanhour.5. Afterhouriscomplete,pressthe“SetZero”button(D).6. Afterthevaluesarestored,enterthecalibrationtank’sgasconcentrationsintheirrespective

entryboxesinthe“CalGas”column(C)(i.e.ifacalibrationgasconsistingofablendof21.02%O2and0%CO2isflowingintotheanalyzers,21.02wouldbeenteredinthe“CalGas”columnfortheO2analyzersand0wouldbeenteredfortheCO2analyzers).DefaulttankvaluescanbechangedonthePreferencesscreen.

7. Repeatsteps3-6usingtheSpanBlend.8. Selectradiobuttons(F)foreachanalyzeryouwouldliketocalibrate.Buttonswillturnblueif

calculationswillbeperformedforthatanalyzer.9. Afterzeroandspanvalueshavebeenstored,press“Calculate”(G).Newslopesandoffsets

arecalculated,butnotappliedtoanalyzergasvalues.10. Ifslopesandoffsetsareacceptable,press“Apply”(H).Newslopesandoffsetsaresavedto

theconfigurationfile.11. Recordthenewslopeandoffsetvaluesinlabnotebook.12. FlowZerogasblendthroughtheanalyzerstoconfirmthecalibrationvalues.Repeatwiththe

Spangasblend.Ifeitherblenddoesnotmatchthesetvalues,repeatthecalibration.

Figure3.PanelLayoutforBlenderCalibration

Comment[MB4]:Whatareacceptablevalues?

Updated:16October2017 RoomCalorimeterCore 7

GasTankCalibration

1. Openzeroandspantanks2. FlipvalvesasthoseshowninFigure43. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer4. Oncethegasisflowing,watchtheO2(A)andCO2(B)graphsuntilthevaluesaresteady.Once

steady,pressthe“SetZero”button(D).5. Afterthevaluesarestored,enterthecalibrationtank’sgasconcentrationsintheirrespective

entryboxesinthe“CalGas”column(C)(i.e.ifacalibrationgasconsistingofablendof21.02%O2and0%CO2isflowingintotheanalyzers,21.02wouldbeenteredinthe“CalGas”columnfortheO2analyzersand0wouldbeenteredfortheCO2analyzers).DefaulttankvaluescanbechangedonthePreferencesscreen.

6. Repeatsteps3-6usingtheSpanCalibrationGas.7. Selectradiobuttons(F)foreachanalyzeryouwouldliketocalibrate.Buttonswillturnblueif

calculationswillbeperformedforthatanalyzer.8. Afterzeroandspanvalueshavebeenstored,press“Calculate”(G).Newslopesandoffsets

arecalculated,butnotappliedtoanalyzergasvalues.9. Ifslopesandoffsetsareacceptable,press“Apply”(H).Newslopesandoffsetsaresavedto

theconfigurationfile.10. Recordthenewslopeandoffsetvaluesinlabnotebook.11. FlowZerogasblendthroughtheanalyzerstoconfirmthecalibrationvalues.Repeatwiththe

Spangasblend.Ifeitherblenddoesnotmatchthesetvalues,repeatthecalibration.

Figure4.PanellayoutforSpanGasCalibration(flipthehighlightedCalibrationSourcevalvefromSpanGastoZeroGasforspancalibration)

Updated:16October2017 RoomCalorimeterCore 8

TankCalCheck

Purpose:Checkanalyzercalibrationvaluesbeforeastudy(humanandvalidation).

1. OpenZeroandSpancalibrationtanks2. SetpanelupasshowninFigure4.3. Flowzerogastotheanalyzersbyturningthe“CalibrationSource”valveto“ZeroGas”.4. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer5. Oncethegasisflowing,watchtheO2(A)andCO2(B)graphsuntilthevaluesaresteady.Once

steady,recordthezerovalue.6. Repeatsteps3-4usingthespangas.7. Calculatethegasdeltabysubtractingthezerogasmeasuredontheanalyzerfromthespan

valuemeasuredonthevalue.8.

𝐷𝑒𝑙𝑡𝑎 = 𝑍𝑒𝑟𝑜𝑇𝑎𝑛𝑘𝐺𝑎𝑠𝑃𝑒𝑟𝑐𝑒𝑛𝑡𝑎𝑔𝑒 − 𝑆𝑝𝑎𝑛𝑇𝑎𝑛𝑘𝐺𝑎𝑠𝑃𝑒𝑟𝑐𝑒𝑛𝑡𝑎𝑔𝑒 ex.O2:0.9 = 20.95 − 20.05ex.CO2:0.99 = 0 − 0.99

9. Verifythedeltaiswithin+/-3%oftheexpecteddelta.10. Ifdeltaisnotwithin+/-3%oftheexpecteddelta,performaTanksoftwarecalibration.

Table1.DeltaExample

ExpectedDelta

AllowableDifference

0.9 0.0271 0.03

Updated:16October2017 RoomCalorimeterCore 9

NullTest

Anulltestverifiestheoffsetofthegasanalyzers.Thetestisperformedbyflowingtheinflowair(medair)toboththeinflowandoutflowanalyzers.AnoffsetcorrectionismadeontheCalibrateAnalyzersscreeninCalRQTMorcanbemadeduringpostprocessing.

1. Openafiletologdata(HumanStudy,ValidationStudy,orCalibration)2. TurntheSampleSourceValvetoNull(Figure5).3. Letthesystemrununtilthehumidityreading,plusVO2andVCO2onthesoftwarestabilize.This

indicatesthatthedryersareconditioned.4. Oncethereadingsarestable,recordthenulldataforatleast10min.5. Ifapplyingthenullcorrectioninrealtime,calculateoffsetcorrectionbysubtractingInflowO2

concentrationfromOutflowO2concentration.RepeatforCO2.ApplyoffsetcorrectiontotheinflowanalyzeroffsetforbothO2andCO2.

Figure5.Panellayoutfornulltest

Updated:16October2017 RoomCalorimeterCore 10

ValidationTestingShortCircuit

Figure6.ShortCircuit(SC)RackPanelConfiguration.Redboxeshighlightthevalvesthatmustbeturnedforthecorrectgasstreamtoreachtheanalyzers.**ShadedredboxhighlightsthevalvethatdetermineswheretheSCinfusionissent.Turnthevalvetoselectthedesiredchamber**

Updated:16October2017 RoomCalorimeterCore 11

Figure7.ShortCircuitMFCPanelConfiguration

1. SetuprackpanelasshowninFigure4.ChangeMFCpanelasshowninFigure7

2. Figure9.3. StartaValidationStudy(File->NewValidationStudy)4. ChooseInfusionRadioButton5. EnterValidationID–FilenamewillbeValidationID–Chamber#-Infusion-Date6. SetdesiredInflowandOutflowrates(Edit->ManualMFCSet)

a. InflowRate=60lpmb. OutflowRate=0lpmc. UsethehighflowMFCforchamber2.Setto60lpm

7. EnterflowrateinInflowSettingpress“ApplyInflow”.CheckCurrentInflowRateisasexpected.

Updated:16October2017 RoomCalorimeterCore 12

8. Runanull–samegasshouldbegoingtoboththeinflowandoutflowanalyzers(Valvesonthepanelshouldbepointingtonull)

9. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer10. Setchambervolumeto0onthePreferencesScreen.11. OpenBlenderScreen(Edit->AdjustBlender).ChooseFlowRateradiobutton12. SelectAppropriateMFCandsetflowrateorselectdesiredprofile.Press“Apply”tostartflow

throughblender.13. FloweachN2/CO2blendforatleast10minafterVO2andVCO2havestabilized

a. Quickcheck:VCO2shouldbeclosetoCO2flowrate;VO2shouldbecloseto

(N2flow+CO2flow)*[O2inflow]

Updated:16October2017 RoomCalorimeterCore 13

Infusion

Figure8.InfusionPanelConfiguration-Redboxeshighlightthevalvesthatmustbeturnedforthecorrectgasstreamtoreachtheanalyzers.**Shadedredboxhighlightsthevalvethatdetermineswheretheinfusionissent.Turnthevalvetoselectthedesiredchamber**

Updated:16October2017 RoomCalorimeterCore 14

Figure9.ChamberStudyMFCPanelSetup

1. SetupRackpanelasshowninFigure6.Selectthecorrectchambertosendgasinfusion.ChangeMFCpanelasshownin

2. Figure93. StartaValidationStudy(File->NewValidationStudy)4. ChooseInfusionRadioButton5. EnterValidationID–FilenamewillbeValidationID–Chamber#-Infusion-Date6. SetdesiredInflowandOutflowrates(Edit->ManualMFCSet)

a. InflowRate=60lpmb. OutflowRate=0lpmc. UsethehighflowMFCforchamber2.Setto60lpm

7. EnterflowrateinInflowSettingpress“ApplyInflow”.CheckCurrentInflowRateisasexpected.

Updated:16October2017 RoomCalorimeterCore 15

8. Runanull(~15mins)–samegasshouldbegoingtoboththeinflowandoutflowanalyzers(Valvesonthepanelshouldbepointingtonull)

9. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer10. SetChamberVolumetoappropriatevolume(Chambers=32500liters,Flex=5000liters)11. Closechamberdoor.12. Addanotetodatafilewithchambersetup(whatfurnitureistheroom,platform,bikeor

emptyinthesmallroom)13. SwitchnullvalvetoStudy.Letrunforabout15minutes14. OpenBlenderScreen(Edit->AdjustBlender).ChooseFlowRateradiobutton15. SelectAppropriateMFCandsetflowrateorselectdesiredprofile.Press“Apply”tostartflow

throughblender.16. Setflowratesandselect“Apply”17. VerifyOutflowO2andOutflowCO2valuesarechangingasexpected.