Page 1: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

Read carefully the instructions published in this manual before the first use of the level meter. Keep the manual at a safe place. The manufacturer reserves the right to implement changes without prior notice.

INSTRUCTION MANUALprůmyslová elektronika

Ultrasonic level meters Ulm – 70

Firmware: v.2.0

Page 2: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface
Page 3: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

Table of conTenTs

1 .  Measuring principle .................................................................................................................3

2 .  Range of applications ..............................................................................................................3

3 .  Variants of sensors ..................................................................................................................3

4 .  Dimensional drawings .............................................................................................................4

5 .   Installation and putting into operation ..................................................................................4

6.    Installation instructions ..........................................................................................................5

7.  Electrical connection ...............................................................................................................9

8.  Examples of ULM-70 connection ............................................................................................10

8.1 . Wiring diagram of the level meter with current output and PDU unit .................................10

8.2 . WiringdiagramofthelevelmeterwithcurrentoutputandPDUunit .............................. in versions for explosive areas ..........................................................................................11 8 .3 . WiringdiagramofthelevelmeterwithcurrentoutputandMGUunit ................................13

8 .4 . WiringdiagramofthelevelmeterwithModbus/RS485andMGUunit ...........................13

8 .5 . WiringdiagramofthelevelmeterwithModbus/RS485andPDUunit ............................13

9 .  Set-up elements .......................................................................................................................14

10.  Status signalization ................................................................................................................14

11 .  Operation and setting ............................................................................................................15

10.1 . Basicconfiguration ..........................................................................................................15

10.2 . Servicesettings ...............................................................................................................18

10 .3 . Additionalfunctions .........................................................................................................20

12.  HART® communication protocol ...........................................................................................22

13.  Modbus® communication protocol ........................................................................................23

14 .  Order code ..............................................................................................................................23

15 .  Accessories ............................................................................................................................23

16 .  Safety, protection, compatibility and explosion proof .......................................................24

17 .  Use, manipulation and maintenance ....................................................................................24

18 .  Marking of labels ....................................................................................................................25

19 .  Menu structure .......................................................................................................................27

20 .  Specifications .........................................................................................................................28

21 .  Packing, shipping and storage .............................................................................................30

Page 4: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.3

1 .  Measuring principle

UltrasoniclevelmeterULM® is a compact measuring device consisting of two parts - main level me-ter(thebodywithmeasuringelectronics)anddisplaymodule.Usingtheelectroacousticconverter,the level meters transmit the sequence of ultrasonic pulses that spread towards the surface level. Theconverterrecuperatesreflectedacousticwavesthataresubsequentlyprocessedintheelec-tronicmodule.Theintelligentevaluationblockfiltersoutinterferingsignals,comparesthecleanedreceivedsignalwith the false reflectionmap (e.g. frommixers, ladders, reinforcementetc.)andselectsasuitablereflection(echo).Basedontheperiodduringwhichtheindividualpulsesspreadtowardsthesurfacelevelandbackandbasedonthemeasuredtemperatureinthetank,theinstantdistancetothesurfaceleveliscalculated.Accordingtothelevelheight,thelevelmeteroutputissetandthemeasuredvalueisdisplayedonthedisplay.

2 .  range of applicaTions

Forcontinuousnon-contactlevelmeasurementofliquids(watersolutions,seweragewater,etc.),mashandpastematerials(sediments,sticks,resinsetc.)inclosedoropenvessels,sumps,reser-voirsandopenchannels.Incasethelevelofbulk-solidmaterialsismeasured,themeasurementrange is reduced.The levelmeterscancontinuouslymeasure levelsofbulk-solidmaterialswitha low concentration of dust particles. Consult the manufacturer on recommended use of the level meterforbulk-solidmaterials.

3 .  VarianTs of sensors

All operations described in this instruction manual have to be carried out only by trained personnel or an accredited person. Warranty and post warranty service must be exclusively carried out by the manufacturer.

Improper use, installation or set-up of the level meter can result in crashes in the application (overfilling of the tank or damage of system components).

The manufacturer is not responsible for improper use, losses of work caused by either direct or indirect damage, and for expenses incurred during installation or use of the level meter.

safeTy

ULM  –70_ –02 Measuring range from 0.15m to 2m, plasticPVDFtransmitter,mechanicalconnectionwiththreadG1".

ULM –70_–06 Measuring range from 0.25m to 6m, plasticPVDFtransmitter,mechanicalconnectionwiththreadG1½".

ULM –70_–10 Measuring range from 0.4m to 10m, plasticPVDFtransmitter,mechanical connection with thread G2¼".

ULM–70_–20 Measuring range from 0.5m to 20m, plasticPVDFtransmitter,mechanicalconnectionwithaluminiumalloyflange.

Page 5: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

4© Dinel, s.r.o. ULM–70

4 .  DiMensional Drawings

ULM–70_–02

Pg11 cable gland

Ground terminal

variant ULM-70 with protective conductor

ULM–70_–06

ULM–70_–20

ULM–70_–10

This procedure includes the following three steps.

•  insTallaTion

•  elecTric connecTion

•  seTTing

5 .  insTallaTion anD puTTing inTo operaTion

Page 6: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.5

Fig. 2: Level meter dead zone

m – Dead zone (blind zone, blocking distance)

6 .    InstallatIon InstructIons• Install the level meter in the vertical positionintotheupperlidofthetankorreservoirusingaweld-

ingflange,afasteningnutoraflangesothatthelevelmeteraxiscanbeperpendiculartothesurfacelevelofthemeasuredliquid(Fig.1).

•   Themin.dimensional parameters to install the level meterintoalidoraceilingofatankaregiveninFig.3.

• When installing in an open channel(reservoir,drainetc.),installthelevelmeterontoabracketascloseaspossibletotheexpectedmax.level.

• Inconnectionwiththemeasurementprinciple,nosig-nals reflectedintheareaimmediatelyunderthelevelmetercanbeevaluated.The zone(Fig.2)determinesthe min. distance possible between the level meterand the highest surface level. The min. distances to themediumaregiveninthechapter"Specifications".

• It is necessary to install the levelmeter so that thebin level cannot interfere with the dead zone when filleduptothemaximum.Ifthemeasuredlevelinter-fereswiththedeadzone,thelevelmeterwillnotworkproperly.

Fig. 1: Recommended installation in the tank

Fig. 3: Installation distance from the tank wall

• If themaximumsurface level in the tank interfereswith thedeadzone, the levelmeterhastobemountedintoahigherinstallation neck.Inthisway,thetankcanbefillednearlyuptothemaximumvolume.Theinnernecksurfacehastobeevenandsmooth(withoutedgesandweldedjoints);theinneredgeshouldberoundedwheretheultrasonicwaveleavesthepipe.Theneckdiametershouldbeaslargeaspossiblebuttheneckheightshouldbeaslowaspossible.RecommendeddimensionsoftheinputneckaregiveninFig.4.

ULM–70–02 ; 10 d > 1/12 c (min. 200 mm)

ULM–70–06 d > 1/8 c (min. 200 mm)

ULM–70–20 d > 1/10 c(min. 200 mm)

Fig. 4: Possible installation of the installation neck

a – Neck heightb – Neck widthm – Dead zone (blind zone, blocking dist.)

ULM–70–02 ; 06 a < 3 bb > 100 mm

ULM–70–10 a < 1,5 bb > 100 mm

ULM–70–20 a < 1,5 bb > 150 mm

d – Distance from the tank wall

c – Maximum reach of the level meter

Page 7: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

6© Dinel, s.r.o. ULM–70

• Iftheemittedacousticsignalofthelevelmeterisaffectedbynearobjects(roughnessonwallsofthetank,variouspartitions,mixersetc.),itisnecessarytomapfalsereflectionsbyactivatingthemode"TEACHING".Incaseofinstalledmixers,itisnecessarytoputthemixerstopositionunderthelevelmeter(directthemixerpaddletotheultrasonicsignalbeam)(Fig.5and6).

• Donotinstallthelevelmeterinorabovethefillingpoint(Fig.7).

Fig. 5: False echo from obstacles in the tank Fig. 6: False echo from the mixer paddle

Fig. 7: Level meter installation outside the influence of filling

• Incasethelevelofbulksolidsismeasured,themeasurementrangeisreducedDuetoabsorp-tionofacousticwavesbyabulkmedium,short-eningofthemeasuringrangeoccursbyupto50% depending on the grain size. We therefore recommend selecting a level meter with greater range than the maximum range of measuring the medium. It is also appropriate to use a di-rectionalhorn(seeimage8),whichreducestheshorteningofthemeasuringrange,becauseitbetterconcentratesacousticenergywhilepre-serving the same beam angle, and improvesthe sensitivity when receiving the reflectedecho. We recommend to consult the use with the manufacturer.

Fig. 8: Level meter installation in silo or hopper

hornadapter

Page 8: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.7

!

!

• Scattering or attenuation of theultrasonic signal can result if the surfacelevelhasbeenmoderate-ly stirred or rippled (byamixer,coming liquid etc.). It can resultin reduction of the measurement rangeorunreliablefunctionofthelevelmeter(Fig.10).

• Rotatingmixer blades can cause thatthesurfaceisstirred,whichre-sultsinfalsereflectionsoftheultra-sonic signal from the surface level and unreliable operation of thelevelmeter(Fig.13).(obr.11).For a rippled or swirling level, you can use the directional horn to elimi-nate scattering of the ultrasonic signal.

• During filling, mixing and other processes, foam can arise on the surface level of the measured liq-uid. The foam considerably absorbs the ultrasonic sig-nal which might cause malfunction of the level meter (Fig.9).Forsuchcases, it isnecessary tosetup"SENSI-TIVITY"modeto"high"orcontactthemanufacturerifneed. Incaseofathinlayeroffoam,itisalsopossibletousethedirectionalhornforimprovingreceiptofthereflectedecho.

Fig. 9: Foam on the surface

Fig. 10: Moderately stirred surface

Fig. 11: Intensely stirred surface

• If the level sensor is mounted to bottlenecks and places with ba-rriers,ornearunevenwallsor thefilling area, where the transmissi-on signal could be distorted, werecommend using a guide tube(acoustichorn).The tubemustbemade from a single material with a smooth inner surface (see image 12a,12b).Theminimumtubedia-metermusthavethedimension"b"according to image 4 on page 5. After installing, you must performthe procedure "LEARNING". Werecommend consulting with the manufacture on the construction of theguidetube.

Obr. 12a: Short guide tube installation

ventilation holes

sleeves for mounting

Obr. 12b: Total guide tube installation

Page 9: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

8© Dinel, s.r.o. ULM–70

• Thelevelmetermustnotbeinstalledinplaceswithdirectso-lar radiationandmustbeprotectedagainstweathereffects.

• If the installation in places with direct solar radiation is inevi-table,itisnecessarytomountashielding coverabovethelevelmeter(Fig.13).

• Itissuitabletorunthecableunderacablebushing(obliquelydown inslack)according toFig.14 topreventpenetration of humidity.Thentherainandcondensingwatercanflowofffreely.

• The cable bushing and connector have to be sufficiently tightenedtopreventpenetrationofhumidity.

• To lower theminimum distance to themeasuredmedium,a reflection boardmadefromsolid,evenandsmoothmate-rialcanbeinstalledtothelevelmeter.Thenthetankcanbefillednearlyuptothemaximumheight.Thesolutionissuit-ableforopentanksandreservoirs(Fig.15).

H + L = mrtvá zóna

Fig. 13: Solar radiation shielding cover

Fig. 14: Prevention to avoid intrusion of humidity

Fig. 15: Reflection board

Page 10: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.9

7 .  elecTrical connecTionThelevelmeter isconnectedtoconsequential (evaluating)devicewithasuitablecablewith theouterdiameterof6to8mmusingscrewterminalslocatedunderthedisplaymodule.Therecom-mendedcrosssectionofcoresforthecurrentversion2x0,5÷0,75mm2 and for the version with Modbuscommunication2x2x0,25mm2 (twistedpair,shielded).Pluspole(+U)isconnectedtotheterminal(+),minuspole(0V)totheterminal(-)andtheshielding(onlyforshieldedcables)totheterminal( ).Commu-nicationwiresAandBofthelineRS-485(forversion"M"-Modbus)areconnectedtotheterminalsAandB.Procedure to connect the cable to the level meter: 

1. Unscrew the nut of the upper transparent lid.

2. Take the upper edge of the display module andtakeitoutcarefullybymildswingingup.

3. If you cannot grasp the module, you can usea small screwdriver. Insert it as far as the seam and usefromseveralsidestoslightlyliftthemodule.

4. Release the cableoutlet and thread the strippedsupplycablein.

5. Connectthecabletothescrewterminalsaccord-ingtothediagraminFig.17or18.Firmlytightentheterminalsandthecableoutlet.

6. If the level meter with Modbus is involved as aterminalforRS-485,werecommend(toavoidre-flectionsontheline)toconnect120Ωterminationresistor.This is donebymovinga small lever oftheswitchmarked120ΩtotheONposition.Onthelevelmetersconnected to the lineRS-485asanintermediatedevice, the termination resistorsarenotconnected(switchremainsoff).

7. Insert the displaymodule back into the head sothattheconnectorisproperlyconnected.

8. Slidesiliconesealonthethreadofthelevelmeterbody,thentightenthenutoftheupperlid.Connectthecabletoconsequentialdevice.

display unit connector

Terminal block

metal clip

Electrical connection must be done in de-energized state!

With regard to possible occurrence of electrostatic charge on non-conductive parts of the level meter, all level meters for explosive spaces (ULM–70Xi type) must be grounded. It will be done using a screw placed on the head of the level meter under the cable outlet.

Fig. 17: Inside view of screw terminals of the level meter with current output ULM-70_-_-_-I

Fig. 16: Wiring diagram of the level meter with current output ULM-70 _-_-_- I

+ A B ISGND

ON

120

display unit connector

Terminal block

metal clip

120Ω switch

Fig. 18: Wiring diagram of the level meter with Modbus ULM-70_-_-_-M

Fig. 19: Inside view of screw terminals of the level meter with Modbus ULM-70_-_-_-M

Page 11: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

10© Dinel, s.r.o. ULM–70

The supply voltage source should be preferably realized as a stabilized power supply unit with safe voltage from 18 to 36 V DC (18 ÷ 30 V DC for Xi version), which can be a part of the evaluation or display device.

In case of strong electromagnetic interferences (EMI), parallel cable ducting with power lines, or when cable length exceeds 30 m we recommended to use shielded cable.

It is also necessary to design and take measures to reduce the effects of static electricity to a safe level in the wiring.

Installation in explosive atmospheres needs to be carried out in compliance with CSN EN 60079-14 (Electrical installations for explosive gaseous atmospheres – Part 14: Electrical installations in dangerous areas other than mining) and possibly also in compliance with other standards relating to the area concerned.

8 .  exaMples of ulM-70 connecTion

PDU-420-W (input 4 ... 20 mA )

(0V) -

ULM-70N-_ _-_-I 

User manual - METER PDU-4xx-P

Figure 4.7. Terminals description (relay outputs, device with current output)

Figure 4.8. Terminals description (OC-type outputs)

Figure 4.9. Connection of 2-wire current converters

11

1 2 5 6 7 8 9

GND DATA+DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

10 11 12 13 14 15

R1

3 4

optional ACTIVEcurrent output

- +

R2

n.c. n.c. n.c.n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++

(optional) (optional)

Powersupply

(depending on version)

1631

1732

1833

1934

2035

1 2 5 6 7 8 9

GND DATA+DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

10 11 12 13 14 15

3 4

OC1 OC2 - + + -

OC1 ÷ OC4: Umax = 30V DC,

Imax = 30mA, Pmax = 100mW

+ -

- +

OC3

OC4(optional)

(optional)

n.c.n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++(optional)(optional)

Powersupply

(depending on version)

1631

1732

1833

1934

2035

5 6 24V DC

- +

+ -

+-

+ -

internally connected

1631

1732

1833

1934

2035

1631

1732

1833

1934

2035

ULM-70N-_ _-_-I 

connectedinternally

PDU-4xx-P (input 4 ... 20 mA )

8 .1 .  Wiring diagram of the level meter With current output and pdu unit

(+U) + (+U) + (0V) -

connectedinternally

User manual - METER SRP-N118

a) b)

Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)

Figure 4.3. Method of connecting cables to the clamping connectors

All connections must be made while power supply is disconnected !

Figure 4.4. Terminals description (relay outputs)

9

!

+ Uo -n.c

.

R1 R2

GND

0/1 - 5V0/2 - 10V

0/4 - 20mA

+

+

GND

GN

D

DA

TA

+

151617 12 11 10 9 8 71314 123456

DA

TA

-

RS - 485

(option) (option)

Po

wer

su

pp

ly(d

ep

end

ing

on

ve

rsio

n)

24V (+5%, -10%)

Imax = 100 mA

PE

n.c

.

5 mm

ma

x.

1.6

mm6 mm

ma

x.

2 m

m

Connection of PDU-420-W is valid for firmware version 6.00 or higher. The older versions (up to version 5.99), the level meter output +U is connected to the terminal 7 and the output 0V to the terminal 10.

Page 12: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.11

8 .2 .  Wiring diagram of the level meter With current output and pdu unit in versions for explosive areas

(+U) +

(0 V) –

ULM-70Xi-_ _-_-I 

IRU-420-I

explosive area

User manual - METER SRP-N118

a) b)

Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)

Figure 4.3. Method of connecting cables to the clamping connectors

All connections must be made while power supply is disconnected !

Figure 4.4. Terminals description (relay outputs)

9

!

+ Uo -n.c

.

R1 R2

GND

0/1 - 5V0/2 - 10V

0/4 - 20mA

+

+

GND

GN

D

DA

TA

+

151617 12 11 10 9 8 71314 123456

DA

TA

-

RS - 485

(option) (option)

Po

wer

su

pp

ly(d

ep

end

ing

on

ve

rsio

n)

24V (+5%, -10%)

Imax = 100 mA

PE

n.c

.

5 mm

max.

1.6

mm6 mm

ma

x. 2

mm

PDU-420-W-230V (input 4 ... 20 mA )

Connection of PDU-420-W is valid for firmware version 6.00 or higher. The older versions (up to version 5.99), the terminal 9 of the IRU unit is connected to the terminal 10 of the PDU unit and the terminal 10 of the IRU unit is connected to the terminal 11 of the PDU unit.

non-explosive area

User manual - METER SRP-N118

a) b)

Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)

Figure 4.3. Method of connecting cables to the clamping connectors

All connections must be made while power supply is disconnected !

Figure 4.4. Terminals description (relay outputs)

9

!

+ Uo -n.c

.

R1 R2

GND

0/1 - 5V0/2 - 10V

0/4 - 20mA

+

+

GNDG

ND

DA

TA

+151617 12 11 10 9 8 71314 123456

DA

TA

-

RS - 485

(option) (option)

Po

wer

su

pp

ly(d

ep

end

ing

on

ve

rsio

n)

24V (+5%, -10%)

Imax = 100 mA

PE

n.c

.

5 mm

ma

x.

1.6

mm6 mm

ma

x.

2 m

m

LN

230 V AC

Page 13: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

12© Dinel, s.r.o. ULM–70

User manual - METER SRP-94

Double numeration means, that depending on device version, particular terminal can be marked according to the top or bottom number.

Figure 4.6. Terminals description (relay outputs)

Figure 4.7. Terminals description (OC-type outputs)

11

GND DATA+

DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

1 2 5 6 7 8 9

10 11 12 13 14 15

3 4

31 32 33 34 35

OC1 OC2 - +

OC1 ÷ OC4: Umax = 30V DC,

Imax = 30mA, Pmax = 100mW

OC3

OC4

+ - + -

+ -

(optional) (optional) (optional)

(optional)

Power

supply

(depending on version)

n.c. n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++

Power

supply

(depending on version)

GND DATA+

DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

1 2 5 6 7 8 9

10 11 12 13 14 15

3 4

31 32 33 34 35

R1 R2 R3

R4

(optional) (optional) (optional)

(optional)

n.c. n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++

1631

(+U) +

(0 V) –

ULM-70Xi-_ _-_-I 

IRU-420-I

PDU-4xx-P-230V (input 4 ... 20 mA )

User manual - METER SRP-94

Double numeration means, that depending on device version, particular terminal can be marked according to the top or bottom number.

Figure 4.6. Terminals description (relay outputs)

Figure 4.7. Terminals description (OC-type outputs)

11

GND DATA+

DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

1 2 5 6 7 8 9

10 11 12 13 14 15

3 4

31 32 33 34 35

OC1 OC2 - +

OC1 ÷ OC4: Umax = 30V DC,

Imax = 30mA, Pmax = 100mW

OC3

OC4

+ - + -

+ -

(optional) (optional) (optional)

(optional)

Power

supply

(depending on version)

n.c. n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++

Power

supply

(depending on version)

GND DATA+

DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

1 2 5 6 7 8 9

10 11 12 13 14 15

3 4

31 32 33 34 35

R1 R2 R3

R4

(optional) (optional) (optional)

(optional)

n.c. n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++

1631

explosive area

non-explosive area

User manual - METER PDU-4xx-P

Figure 4.7. Terminals description (relay outputs, device with current output)

Figure 4.8. Terminals description (OC-type outputs)

Figure 4.9. Connection of 2-wire current converters

11

1 2 5 6 7 8 9

GND DATA+DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

10 11 12 13 14 15

R1

3 4

optional ACTIVEcurrent output

- +

R2

n.c. n.c. n.c.n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++

(optional) (optional)

Powersupply

(depending on version)

1631

1732

1833

1934

2035

1 2 5 6 7 8 9

GND DATA+DATA-

RS - 485

+ Uo -

+24V +5%, -10% Imax = 100mA

10 11 12 13 14 15

3 4

OC1 OC2 - + + -

OC1 ÷ OC4: Umax = 30V DC,

Imax = 30mA, Pmax = 100mW

+ -

- +

OC3

OC4(optional)

(optional)

n.c.n.c.

GND

0/1 - 5V, 0/2 - 10V

0/4-20mA

++(optional)(optional)

Powersupply

(depending on version)

1631

1732

1833

1934

2035

5 6 24V DC

- +

+ -

+-

+ -

internally connected

1631

1732

1833

1934

2035

1631

1732

1833

1934

2035

User manual - METER SRP-N118

a) b)

Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)

Figure 4.3. Method of connecting cables to the clamping connectors

All connections must be made while power supply is disconnected !

Figure 4.4. Terminals description (relay outputs)

9

!

+ Uo -n.c

.

R1 R2

GND

0/1 - 5V0/2 - 10V

0/4 - 20mA

+

+

GND

GN

D

DA

TA

+

151617 12 11 10 9 8 71314 123456

DA

TA

-

RS - 485

(option) (option)

Po

wer

su

pp

ly(d

ep

end

ing

on

ve

rsio

n)

24V (+5%, -10%)

Imax = 100 mA

PE

n.c

.

5 mm

max.

1.6

mm6 mm

ma

x. 2

mm

LN

230 V AC

Page 14: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.13

(+U) +

(0 V) -

ULM-70N-_ _-_-I 

MGU-800 (with module IUI4)

8 .3 .  Wiring diagram of the level meter With current output and mgu unit

12 13 14 18 19

A B +_

S

RS 485 - Master+24V DC

Imax=120 mA

(A) Y

E

(B) G

N

(0V)

WH

(+U)

BN

Cab.

Sh.

ULM-70N-_ _-_-M 

(A) YE(B) GN

(0V) WH(+U) BN

Cab.Sh.

MGU-800 (without additional modules)

8 .4 .  Wiring diagram of the level meter With modbus / rs485 and mgu unit

8 .5 .  Wiring diagram of the level meter With modbus / rs485 and pdu unit

ULM-70N-_ _-_-M 

Page 15: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

14© Dinel, s.r.o. ULM–70

ESC OK

www.dinel.cz

Ultrasonic Level Meter ULM-70

ESC OK

1745T E mm

10 .  sTaTus signalizaTion

1) symbol appears in the lower left corner of the display

display function

"NO ECHO" Lighting intermittently – the level meter is not able to receive echo for a long time. Incorrect installation of the level meter

"DEAD ZONE" Lighting intermittently – the measured level is in the "dead zone" of the level meter or the ultrasonic converter is dirty.

"NO PASSWORD" It will appear in the item "MENU" – the level meter is protected using a pass-word against unauthorised setting. Enter the correct password (see p. 19).

Symbol "T" 1) Lighting permanently – "TEACHING" mode activation.

Symbol "E" 1) Lighting intermittently – correct echo receiving (of the reflected signal) from the measured surface level.

Symbol

ESC OK

www.dinel.cz

Ultrasonic Level Meter ULM-70

ESC OK

1745T E mm 1)

Lighting permanently – level meter is locked against unauthorized settings by a password. You must enter the correct password to unlock it (see page 19).

Level meter locked

Ultrasonic transmitter

Upper lid nutCable gland

Control buttons

OLED display Label

9 .  seT-up eleMenTs

Button   

• Set-upmodeaccess• Confirmationofselectediteminthemenu• Movethecursorintheline• Savingofset-updata

Button 

• Moveinthemenu• Change of values

Button   

• Cancelling of carried out changes• Shiftonelevelup

Echo receiving

UnitsTeachingmode activation

Displayof measured values

Set-up elements

Fig. 20: Full view of the level meter

ESC OK

www.dinel.cz

Ultrasonic Level Meter ULM-70

ESC OK

1745T E mm

Page 16: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.15

START ING

Dinel®

11 .  operaTion anD seTTing

Setthelevelmeterusing3buttonsplacedonthedisplaymodule(seeChapterSet-upelements).After5min.ofinactivity,thelevelmeterautomaticallyreturnsbacktothemeasurementmode.Ifthepassword is active, the levelmeterwill be also locked.The val-uesthathavenotbeenconfirmedusingthebutton willnotbesaved!After themeter is locked, youcannot change the setting!Whenyouattempttoedit,thewords"NOPASSWORD"willappearonthedisplay.Howtounlockthelevelmeterisgivenonpage16.Afterconnectionofthesupplyvoltagetothelevelmeterthedisplayshowsthelogo"Dinel"andthetext"Starting"(approx.15s).Then,thelevelmetergoestothemeasuringmodeandthedisplayshowsthe current measured value.

MIN LEVEL and MAX LEVEL

Youcan freelydefine theminimum / maximum distance  from the  front surface of  the  level meter(item"LEVEL"forcurrents4/20mA).The"DISPLAY"isintendedtosetthevaluedisplayedonthedisplay.Settingtheunitsisdoneinthe"UNITS".

Ifinthebottomofthedisplayappears(whenenteringthevalues)theinscription"OUTOFLIMITS",thevaluespecifiedfortheitem"LEVEL"isoutsidethemeasuringrangeofthelevelmeter.Iftheinscription"SPANTOOSMALL"isshown,itmustbespecifiedalargerspanbetweenMinandMaxvalues.Formoreinformation,seechapter"Specifications".Thedecimalpointpositionoftheitem'LEVEL'isfirmlyset(accordingtotheselectedunits),intheitem"DISPLAY"itisfreelyadjustable

11 .1 .  basic configurationAfterthefirststartofthelevelmeteritisnecessarytoperformthebasicconfiguration(settingofthemeasuringrange,choiceofunitsandpossibly damping).The settingsareaccessible in thebasicmenubypressing the"BASICSETTINGS".

UN ITS

LEVEL : mm

DISPLAY: %

TEMPERATURE °C

MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING

MIN ACTUAL LEVEL0 2 8 7 0 mm

OUTPUT : 0 4 . 0 0 mA

LEVEL : 0 6 0 0 0 mm

DISPLAY: 0 0 0 0 0 %

MAX ACTUAL LEVEL0 2 8 7 0 mm

OUTPUT : 2 0 . 0 0 mA

LEVEL : 0 0 2 5 0 mm

DISPLAY: 0 0 1 0 0 %

ACTUAL LEVEL: Actual distance to levelOUTPUT: current 4 mA / 20 mALEVEL: Definition of the min / max levelDISPLAY: The value showed on the displayMIN LEVEL

MAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO

Page 17: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

16© Dinel, s.r.o. ULM–70

UNITS

LEVEL : mm

DISPLAY: %

TEMPERATURE °C

DAMP ING

05

UNITS

Levelmetercanprocessandconvertalargenumberofdifferentphysical values. The setting is doneintheitem"UNITS".

LEVEL: Unit selection (mm, cm, m, in, ft)DISPLAY: The unit showed on the display (%, mm, cm, m, in, ft, l, hl, m3, gal, bbl, mA)TEMPERATURE: Temperature unit (°C, °F)

DAMPING

Settingthe response time of the measurements. Thefunctionisusefulforsuppressinglevelfluc-tuations,wavesandrapidchangesofthelevel.The reaction time will depend on the exponential function.Dampingwithadefineddelayinsecondsrepresentsthetimewhenexponentialreaches2/3ofitsmaximumvalue.

The damping time can be set in the interval from 0 to 99 s.

1. To enter to the menu press thesamebuttontoselect"BASICSETTINGS".Then,usingand select"UNITS".

2. Nowthemenuitem"UNITS"isshown. Bypressingthe and buttonmakethesettingsofindividual items.

3. Bypressing buttonsave thedata.Bynextpressesof thebutton leave the menu. The level meter returns to measurement mode.

1. To enter to the menu press thesamebuttontoselect"BASICSETTINGS".Then,usingand select"DAMPING".

2. Nowthemenuitem"DAMPING"isshown. Bypressingthe and buttonmakethesettingsof individual items.

3. Bypressing buttonsave thedata.Bynextpressesof thebutton leave the menu. The level meter returns to measurement mode.

1. To enter to the menu press thesamebuttontoselect"BASICSETTINGS". Then,usingand select "MINLEVEL"or"MAXLEVEL".

2. Nowitisshowntheitem"MINLEVEL"("MAXLEVEL").Bypressing and set the output cur-rent"OUTPUT",thedistanceforthedefinedcurrent"LEVEL"thevalueonthedisplay "DISPLAY".

3. Bypressing buttonsave thedata.Bynextpressesof thebutton leave the menu. The level meter returns to measurement mode.

MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING

MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING

Page 18: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.17

You can set the sensitivity in three degrees: LOW – MEDIUM – HIGH.

SENSITIVITY

Thesettingisdefinedinthreestepsofthelevelmetersensitivity.

"LOW" – Low sensitivity in case of surrounding interferences affecting the measurement. "MEDIUM" – Medium sensitivity (suitable for most applications)."HIGH" – Enhanced sensitivity for measured mediums partly absorbing the ultrasonic signal (bulk solids, foams)

If there are no above obstacles in the tank, it is not necessary to start this mode.

TEACHING

The mode serves for suppressing false reflections resulting fromreflectionof theultrasonicsignalfromroughnessesonwallsofthetank,variouspartitions,mixersorotherobstacles.Thesensorstartingthismodedetectsfalsereflectionsandsavetheminthememory.Thenthesefalsereflectionswillnotaffectthesubsequentmeasurement(theyaremasked).

Before starting the mode it is necessary to empty the tank as much as possible (prefer-ably completely).

1. To enter to the menu press thesamebuttontoselect"SERVICE".Then,using andselect"TEACHING".

2. Nowitisshowntheitem"TEACHING".Bypressing setthevalue"LEVELDISTANCE"(dis-tanceto the level)–supposeddistancefromthefaceof thesensor to themediumlevel. If thedistancetothelevelisnotpreciselyknown,enteravalueratherlower(inthetolerancefieldasshowninFig.19).

3. Afterenteringthe"SETLEVELDISTANCE"bypressing buttonthesystemstarts"teaching"(falsereflectionmapping).Duringthemapping,thedisplayshowsflashingsign"RUNNING".

4. Themappingoffalseechoescanbecompletedwhenyouseetheinscription"PressOKtostop"andyoupress .

5. Theprocedureiscompletelyfinishedwhenyoucanseetheinscription"DONE".Itisthenpos-sibletoexitthemenubyrepeatedpressingthebutton .

SENS IT IV ITY

MED IUM

TEACH ING

SET LEVEL D ISTANCE :

0 6 0 0 0 m m

START

MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING

MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING

TEACH ING

SET LEVEL D ISTANCE :

0 6 0 0 0 m m

RUNN ING

TEACH ING

SET LEVEL D ISTANCE :

0 6 0 0 0 m m

DONEPRESS ESC TO EX IT

The mode "TEACHING" will stop automati-cally after ca. 1000 measurements.

TEACH ING

SET LEVEL D ISTANCE :

0 6 0 0 0 m m

RUNN INGPRESS OK TO STOP

If during the scanning of the tank in the bottom of the display appears the dialog "press OK to stop" (see figure) the level meter already found no further obstacles and "TEACHING" mode may be terminated. If it is not terminated, the level meter is still ready for the possible presence of obstacles (e.g. paddles of the agitator). Once it registers a further obstacle, the dialogue disappears and the obstacle is erased. This process may be repeated up to 1000 cycles. After this the "TEACHING" mode is automatically stopped.

Page 19: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

18© Dinel, s.r.o. ULM–70

Incaseofinstalledmixers,itisnecessary to posi-tion the mixers under the level meter (direct the mixerbladetotheultrasonicsignalbeam).

Note:Iftherearesignificantobstaclesintheupperhalf of the tank,multiple  false  reflections can occur especially in closed tanks. In such casesit isnecessary to reduce the level in the tankasmuchsepossibletocorrectlymaskthesepossiblemultiplefalsereflections.

1/2 of Tank

Obstacles

Tolerance zone for "LEVEL DISTANCE" value

MEDIUM TEMPERATURE

The level meter is equipped with automatic  temperature  compensation. If for instance in the tank there is a difference of 10°Cbetween the temperature of themeasuredmaterial (me-dium) and the temperature at themounting site of the levelmeter (see themode "DIAGNOS-TICS, page 20), the measuring accuracy will be reduced by around 1% of the set range. Ifthis function is activated, this temperature difference can be compensated. If in the tank (openchannel) isabigdifferencebetweenthetemperatureof themeasuredmedium(liquid)andtem-perature in theplaceof installationof theULM(seemode, "DIAGNOSTICS"page20), it isad-vised to improve the precision of the measurement by the zone temperature compensation. Otherwise,thismodeisnot necessary to run.

MEDIUM TEMPERATUREFA ILURE MODEHARTFACTORY DEFAULTRESET

MED IUM TEMPERATURE

YES

+20°C

Inactive compensation (initial state), the word "NO" appears on the display.See the "UNITS" menu for temperature unit selection (°C or °F).

Afterstartofthezone temperature compensationmodeit isnecessarytosetthetemperatureof the surface of the medium. The level meter then calculates the average value from the medium temperature and the temperature at the installation place of the level meter. With such an average temperatureitcountsincalculatingofthevelocityofacousticwavespropagationandfordetermina-tion of the level position.

Leve

l Dist

ance

NO ECHO: Current in case of echo lossDEAD ZONE: Dead zone currentThe values can be set in three steps: 3.75 mA, 22 mA and LAST (last measured data).

FAILURE MODE

It defines the output current of the level meter when the measured medium level is in the dead zone ("DEADZONE")oroutsidethemeasurementrangeincaseofecholoss("NOECHO").

FA ILURE MODE

NO ECHO: 3 .75 mA

DEAD ZONE : 22 mA

11 .2 .  services settingsIn the supplemented configuration, you can set parameters ofsensitivity, mapping of false reflections, temperature differencecompensation, behaviour in case of fault conditions or HART® communication.Here,youcansetthesensorintotheinitialstateorresetitaswell.Thesettingsareaccessibleinthebasicmenuundertheitem"SERVICE".

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDINFO

MED IUM TEMPERATUREFA ILURE MODEHARTFACTORY DEFAULTRESET

Fig. 21: Level distance zone

Page 20: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.19

HART

HART® mode(pointtopoint,multidrop)andmultidropmodeaddress setting.Upto15unitscanbeconnectedtoonetwo-wiredcableinthemultidropmode.

In case of the address "00", the point to point mode is enabled. The range from "01" to "15" is reserved for addresses in the multidrop mode.

After you press the button "RUNNING" will be displayed for about 3 sec. After the initial values are set, "DONE" will be appear on the display.

FACTORY DEFAULT

To reset the initial values ofthelevelmetersetbythemanufacturer,pressthebutton (see the Factorydefaulttable,p.28).

RESET

Complete restartofthelevelmeter.Thesameeffecthasalsoashort-timeinterruptionofthesupplyvoltage.Toenabletheresetting,pressthebutton .

FACTORY DEFAULT

ARE YOU SURE?

FACTORY DEFAULT

RUNN ING

FACTORY DEFAULT

DONEPRESS ESC TO EX IT

RESET

ARE YOU SURE?During the restart process, "RUNNING" will be displayed. Then the level meter will be automatically turned off and on.

MODBUS

This itemispartofamenuwithModbusoutput levelmeterULM-70_-_-_-M.Modbusmode isintendedforthesettingsofthelevelModbusaddresses,baudrateandparitysettings.

MODBUS

ADDRESS : 001

BAUD RATE : 9600

FORMAT: 8N1

ADDRESS: 1 to 247 (default 1) BAUD RATE: 4800, 9600, 19200 (default 9600) FORMAT : 8N1, 8O1, 8E1, 8N2 (default 8N1)

data:

parita: N – non parityO – odd parityE – even parity

8 – number of bits

FORMAT: number of stop bits: 12

MEDIUM TEMPERATUREFA ILURE MODEHARTFACTORY DEFAULTRESET

HART

00

POLL ING ADDRESS

MED IUM TEMPERATUREFA ILURE MODEMODBUSFACTORY DEFAULTRESET

MED IUM TEMPERATUREFA ILURE MODEMODBUSFACTORY DEFAULTRESET

MED IUM TEMPERATUREFA ILURE MODEMODBUSFACTORY DEFAULTRESET

Page 21: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

20© Dinel, s.r.o. ULM–70

DIAGNOSTICS

Itcontainsinformationabouttheactualtemperatureinsidethetank(or about the compensated temperature) "TEMPERATURE" andcurrent flowing through the loop "CURRENT". If the temperaturecompensation("MEDIUMTEMPERATURE")isactivated,thecor-rectedtemperatureisdisplayed.

DISTANCE TO LEVELTEMPERATURECURRENT

TEMPERATURE

+23°C

CURRENT

17 .30 mA

The temperature is measured inside the tank where the level meter is installed.If the temperature of the measured medium is different, we recommend you to carry out the temperature compensation "MEDIUM TEM-PERATURE" because of accuracy (see p. 15). Then the displayed temperature is an average value from the temperature set in the "MEDIUM TEMPERATURE" and the actual temperature measured by the sensor.

DISTANCE TO LEVELTEMPERATURECURRENT

CLONE SETTINGS

This mode is intended for copyingofthelevelmeter(ULM–70body) configuration  into  the  display module (DM–70) andback.Thedisplaymodulecanthenberemovedfromthelevelmeterbodyandputintoanotherlevelmeterandmaketherethesettingstransfer(cloning).The"CLONESETTINGS"modetransfersalldata,excludingsettingofthe"Teaching"andHART®.

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDINFO

CLONE SETT INGS

NOW CLON ING

CLONE SETT INGS

SENSOR D ISPLAYMODULE

D ISPLAYMODULE

SENSOR

→→

1. Press toenterthemenuandselecttheitem"CLONESETTINGS".Copyingofthesettingsfromthebodyofthelevelmetertodisplaymoduleisdonebyselecting"SENSOR→ DISPLAYMODULE".TotransferthesettingsfromthedisplaymoduletoanotherlevelmeterselecttheitemDISPLAYMODULE→ SENSOR.

2. Theselectedmodestartsbypressingbutton Duringtransmissionthedisplayshows"NOWCLONING".

3. Aftercompletingtheprocessinthemiddleofthescreendisplays"DONE".Itisthenpossibletoleavethemenuandthemodebypressingthebutton .

CLONE SETT INGS

SENSOR D ISPLAYMODULE

D ISPLAYMODULE

SENSOR

→→

ARE YOU SURE?

11 .3 .  additional functionsAdditionalfunctionsincludemodestodisplaytemperatureinthetankortofindouttheactualflowingcurrentintheloop.Besides,tolockmodificationsusingapasswordandinformationaboutthelevelmeterversion.Allofthefunctionsareaccessiblefromthemainmenu.

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO

D ISTANCE TO LEVEL

01220 mm

DISTANCE TO LEVELTEMPERATURECURRENT

Page 22: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.21

CLONE SETT INGS

WARN ING–CLON ING IS NOT POSS IBLE

WRONG SENSOR TYPE

PRESS ESC TO EX IT

CLONE SETT INGS

WARN ING–CLON ING IS NOT POSS IBLE

NO SETT INGS SAVED

PRESS ESC TO EX IT

Incompatible type of level meter. Transfer of the settings can be realized only with the same type of level meter (e.g. ULM–70–02 → ULM–70-02, ULM–70–10 → ULM–70–10) and with the firmware version 2.0 and later.

PASSWORD

You can lock the level meter data against unauthorized editing. Afteractivatingthepasswordthedatamayberead,butcannotbeedited.Ifyoutrytoeditthesettings(withouttruepassword)thedisplayshows"NOPASSWORD".Thepasswordcanbeany5-digitnumericcombination.Thecom-bination of numbers 00000 is reserved for disabling the pass-word.

1. Usethebuttons and inthemenu"PASSWORD"toselectthemode"ENTER"forenteringthepasswordorthemode"CHANGE"forchangingthepassword(whenactivated,thewordsaredisplayedinversely).Pressthebutton onceagaintoconfirmtheselection.Youcanchangethepasswordonlywhenthelevelmeterisunlocked.Otherwise,thewords"NOPASSWORD"willbedisplayed.

2. Nowyoucaneditthepassword.Theactualediteditemisdisplayedinversely.Pressthebuttontomovetothenextposition(clockwisedirection),button servestochangethevalues(0...9).

3. Aftertheoperationiscompleted,confirmtheediteddatabypressingthebutton .

Display of status information to confirm data:"YES" – correctly edited password"NO" – incorrectly edited password"OK" – the password saved (only in case of "CHANGE")The password is automatically hidden after it is edited or changed ("00000" will appear). To deactivate the password, edit the nu-merical combination "00000" in the mode "CHANGE".

TEACH ING

SET LEVEL D ISTANCE :

0 6 0 0 0 m m

START

NO PASSWORD

PASSWORD

ENTER

0 0 0 0 0

PASSWORD

CHANGE

0 0 0 0 0

The data set is not stored into the display module (DM–70). The transfer can not be done. It is necessary to repeat the procedure of the copying the settings in the mode "CLONE SETTINGS".

The level meter with activated password will be automatically locked after 5 minutes of inactivity or after 5 min. from switching to measuring mode. Locking of level meter is indicated in the lower left corner of the screen by the letter "L".

If the password is lost, contact the manufacturer.

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO

Page 23: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

22© Dinel, s.r.o. ULM–70

INFO

Informationaboutthetype,serialnumberandproductiondateofthelevelmeter(type,serialnumber–SNandfirmwareversion–SW).

SENSOR

ULM-70N -06 - I HART

SN: 140025

SW: 2 .0

D ISPLAY MODULE

DM-70

SN: 140041

SW: 4 .1

SENSORD ISPLAY MODULE

SENSORD ISPLAY MODULE

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO

SENSOR

ULM-70N -06 -M MODBUS

SN: 140025

SW: 2 .0

LANGUAGE

Settingthelanguageofdisplaymenu.

BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO

LANGUAGE

ENGL ISH

You can set three kinds of language:ČESKY – ENGLISH – по русски

12 .  proTokol HarT®

Universal communication interface for data communication of peripheral devices with the level meter.Datatransmissionrunsthroughthesamelineasthe4÷20mAcurrentloopwithoutimpact on analog communication. Forsettingthelevelmeterandcollectionofmeasureddata,itisnecessarytohaveavailableaHARTcommunicator,bywhichitispossibletocommunicatedirectlywiththelevelmeter,orusingit,tomediatecommunicationwithaperipheraldevice,seeimage 22.

≈ 250 R

Obr. 22: Typical hardware configuration with HART

ULM–70_-_ _-_-I

RS232 / USB

HART®

communicator

Power supply unit

Page 24: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.23

13 .  proTokol MoDbus®

14 .  orDer coDe

Standard – incl. in the price of the level sensor Optional – for extra charge

• 1pcofSeal(forULM–70_–02–I,06–I)• free-to-download programme Basic Scada

Level(fortheModbusversion)

• FixingnutsG1"andG1½"andG2 ¼• HornadapterST–G1,

STG1,5andST–G2,25• for versionModbusconvertorURC-485

02 – 0,15 ... 2 m06 – 0,25 ... 6 m10 – 0,4 ... 10 m20 – 0,5 ... 20 m

ULM – 70 –

Output type: I – current outputM – RS-485 (Modbus RTU)

– – –

Process connection: G – threadF – flange

Electrical connection: B – short cable outletH – outlet for protective conductor

– Set-up elements: D – basic version with displayL – without display, metal lid

Performance: N – Normal – usable in non-explosive areas onlyXi – Ex. proof – suitable for explosive areas

Maximum reach:

15 .  accessories

Obr. 23: Typical hardware configuration with Modbus®

DatacommunicationtakesplacealongaserieslineofastandardRS-485withprotocolModbusRTU.Alistofrelevantvariablesisprovidedinaseparateannex.

Tosetupthelevelmeterandcollectmeasureddata,youcanusethesoftwareapplication"BasicSCADAlevel",whichisfreelyavailableattheWebsitewww.dinel.cz.ConnectingthelevelmetertoaperipheraldevicecanbeperformedusingaconverterURC-485,seeimage23.

RS485 / USB

convertor URC-485

ULM–70_-_ _-_-M

Power supply unit

Page 25: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

24© Dinel, s.r.o. ULM–70

It is forbidden to perform any modifications or interventions into the ULM–70 level meter without manufacturer's approval. Potential repairs must be carried out by the manufacturer or by a manufacturer authorized service organization only.

Installation, commissioning, operation and maintenance of the ULM–70 level meter has to be carried out in accordance with this instruction manual; the provisions of regulations in force regarding the installation of electrical equipment have to be adhered to.

Installation in areas with potentially explosive atmospheres must be carried out in accordance with standard EN 60079-14 (Electrical apparatus for explosive gas atmospheres - Part 14: Electrical installations in hazardous areas other than mines) and respectively in accordance with other standards that apply to a given area.

16 .  safeTy, proTecTion, coMpaTibiliTy anD explosion proof

The levelmeterULM–70 isequippedwithprotectionagainst reversepolarityandoutputcurrentoverload.ProtectionagainstdangerouscontactissecuredbylowsafetyvoltagethatcomplieswithEN332000-4-41.ElectromagneticcompatibilityaccordingtoEN55022/B,EN61326/Z1andEN61000-4-2to6.ExplosionproofofULM–70Xitypecomplieswiththefollowingstandards:EN60079-0:2007;EN60079-11:2007;EN60079-26:2007andexaminedbyFTZÚ-AO210Ostrava-RadvanicecertificateNo.:FTZÚ09ATEX0277X.

Special conditions for safe use ULM–70Xi:

ThedeviceisdesignedforconnectiontotheisolatingrepeaterIRU-420.Whentheotherapprovedsupplyunitisused,whoseoutputparameterssatisfyabovementionedoutputparameters,itisnec-essary tohaveagalvanicseparationor, ifsupplyunitwithoutgalvanicseparation isused (Zenerbarriers),itisnecessaryprovidepotentialequalizationbetweensensorandpointofbarrierearthing.Forapplicationinzone0thepresentexplosiveatmospheres-mixtureofairwithflammablegases,vapourormistsmustcomply:0,8bar<p<1,1bar.Thedevicemustbeinstalledinsuchaway,topreventmechanicaldamageofsensorface.Itisnecessarycarriedoutearthingbyscrewwhichisplaced on head of level meter.

17 .  use, ManipulaTion anD MainTenance

The levelmeterdoesnot requireanypersonnel for itsoperation.Follow-updisplayingdevice isusedtoinformthetechnologicalentityoperatingpersonnelonthemeasuredsubstancelevelheightduring the operation.

Maintenanceofthisequipmentconsistsinverificationofintegrityofthelevelmeterandofthesup-plycable.Dependingonthecharacterofthesubstancemeasured,werecommendtoverifyatleastonceperyeartheclarityoftheultrasoundtransduceremittingfieldandtocleanit,respectively.Incaseanyvisibledefectsarediscovered, themanufactureror resellerof thisequipmentmustbecontactedimmediately.

The device must be installed to prevent tensile overload rope electrode level meter.

The  device  must  be  installed  in  such  a  way,  to  prevent  mechanical  damage  of sensor face.

Page 26: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.25

18 .  Marking of labels

Symbolofproducer:logoDinel®

Internetaddress:www.dinel.czLevelmetertype:ULM–70N–__–_–I–_–_Serialnumber:Ser.No.:xxxxx–(fromtheleft:productionyear,serialprodutionnumber)Supplyvoltage:Ui = 18...36V=Outputcurrentrange:I =4÷20mAAmbienttemperaturerange: ta=-30...+__°C(see.Temperature range according to type)Protectionclass:IP67Compliancemark: Electro-wastetake-backsystemmark:

ULM-70N-20-F-I-B-D

U = 18 36 V...

I = 4 20 mA...t = -30 ... +60 °C

a

ULM-70N-10-G-I-B-D

U = 18 36 V...

I = 4 20 mA...t = -30 ... +60 °C

a

ULM-70N-06-G-I-B-D

U = 18 36 V...

I = 4 20 mA...t = -30 ... +70 °Ca

ULM-70N-02-G-I-B-D

U = 18 36 V...

I = 4 20 mA...t = -30 ... +70 °Ca

IP67

IP67

IP67

IP67

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

Ser. No.: ______

Ser. No.: ______

Ser. No.: ______

Ser. No.: ______

Varianty značení:

ExampleoflabelfortypeofULM-70N-02-G-B-D

Temperature range according to type:

02 ta = -30 ... +70°C06 ta = -30 ... +70°C10 ta = -30 ... +60°C20 ta = -30 ... +60°C

Labels for type of ULM-70N-_ _-_-I-_-_:

Clasification non-explosive performance:

02 II1/2GExiaIIBT5Ga/Gb

06 II1/2GExiaIIBT5Ga/Gb

10 II1/2GExiaIIAT5Ga/Gb20 II2GExiaIIAT5Gb

Labels for type of ULM-70Xi-_ _-_-I-_-_:

ULM-70Xi-02-G-I-B-D

IP67

U = 30 Vi

t = -30 ... + 70 °Ca

II 1/2 G Ex ia IIB T5 Ga/Gb

P = 0,99 W, I = 132 mAi i

C = 370 nF, L = 0,9 mHi i

FTZÚ 09 ATEX 0277X

1026

I = 4 20 mA...

ULM-70Xi-06-G-I-B-D

IP67

1026

ULM-70Xi-10-G-I-B-D

IP67

1026

ULM-70Xi-20-F-I-B-D

IP67

1026

U = 30 Vi

t = -30 ... + 70 °Ca

P = 0,99 W, I = 132 mAi i

C = 370 nF, L = 0,9 mHi i

I = 4 20 mA...

U = 30 Vi

t = -30 ... + 60 °Ca

P = 0,99 W, I = 132 mAi i

C = 370 nF, L = 0,9 mHi i

I = 4 20 mA...

U = 30 Vi

t = -30 ... + 60 °Ca

P = 0,99 W, I = 132 mAi i

C = 370 nF, L = 0,9 mHi i

I = 4 20 mA...

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

II 1/2 G Ex ia IIB T5 Ga/Gb

FTZÚ 09 ATEX 0277X

II 1/2 G Ex ia II T5A Ga/Gb

FTZÚ 09 ATEX 0277X

II G Ex ia II T52 A Gb

FTZÚ 09 ATEX 0277X

Varianty značení:

Ser. No.: ______

Ser. No.: ______

Ser. No.: ______

Ser. No.: ______

ExampleoflabelfortypeofULM-70Xi-02-G-I-B-D

Symbolofproducer:logoDinel®

Internetaddress:www.dinel.czLevelmetertype:ULM–70Xi–__–_–I–_–_Serialnumber:Ser.No.:xxxxx– (fromtheleft:productionyear,serialprodutionnumber)Outputcurrentrange: I =4...20mAMax.internalvalues:Ui = 30V=,Ii=132mA;Pi =0,99W;Ci=370nF;Li=0,9mHAmbienttemperaturerange: ta=-30...+__°C(viz.Teplotnírozsahdletypu)Labelofnon-explosivedevice: ,Performance:II_GExiaII_T5_/_Numberofcertificateofintrinsicallysafety:FTZÚ09ATEX0277XProtectionclass:IP67Compliancemark: ,No.ofauthorizedpersonexaminingcontrolofsystemquality:1026Electro-wastetake-backsystemmark:

Temperature range according to type:

02 ta = -30 ... +70°C06 ta = -30 ... +70°C10 ta = -30 ... +60°C20 ta = -30 ... +60°C

Page 27: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

26© Dinel, s.r.o. ULM–70

ULM-70N-20-F-M-B-D

ULM-70N-10-G-M-B-D

ULM-70N-06-G-M-B-D

ULM-70N-02-G-M-B-D

IP67

IP67

IP67

IP67

U = 18 36 V...

RS-485 (Modbus RTU)

t = -30 ... +70 °Ca

U = 18 36 V...

RS-485 (Modbus RTU)

t = -30 ... +70 °Ca

U = 18 36 V...

RS-485 (Modbus RTU)

t = -30 ... +60 °Ca

U = 18 36 V...

RS-485 (Modbus RTU)

t = -30 ... +60 °Ca

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

www.dinel.cz

DinelMade in Czech Republic

Ser. No.: ______

Ser. No.: ______

Ser. No.: ______

Ser. No.: ______

Varianty značení:

Labels for type of ULM-70N-_ _-_-M-_-_:

ExampleoflabelfortypeofULM-70N-10-G-M-B-D

Symbolofproducer:logoDinel®

Internetaddress:www.dinel.czLevelmetertype:ULM–70N–__–_–I–_–_Serialnumber:Ser.No.:xxxxx–(fromtheleft:productionyear,serialprodutionnumber)Supplyvoltage: U= 18...36V=Dataoutput:RS-485(ModbusRTU)Ambienttemperaturerange: ta=-30...+__°C(see.Temperature range according to type)Protectionclass:IP67Compliancemark: Electro-wastetake-backsystemmark:

Real label size is 70x20mm.

Temperature range according to type:

02 ta = -30 ... +70°C06 ta = -30 ... +70°C10 ta = -30 ... +60°C20 ta = -30 ... +60°C

Page 28: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.27

19 . 

Men

u s

Tru

cTu

reM

INM

AX

05

00

SE

NS

OR

GR

LM

-70

N-1

0-I

HA

RT

EL

EC

TR

OD

E:

10

00

mm

SN

: 1

40

01

5

SW

: 2

.1

DIS

PL

AY

MO

DU

LE

DM

-70

SN

: 1

40

08

5

SW

: 4

.1

SE

NS

OR

DIS

PL

AY

MO

DU

LE

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

BA

SIC

SE

TT

ING

SS

ER

VIC

ED

IAG

NO

ST

ICC

LO

NE

SE

TT

ING

SP

AS

SW

OR

DL

AN

GU

AG

EIN

FO

MIN

LE

VE

LM

AX

LE

VE

LU

NIT

SD

AM

PIN

GS

EN

SIT

IVIT

YT

EA

CH

ING

AC

TU

AL

02

87

0m

m

OU

TP

UT

:0

4.0

0m

A

LE

VE

L:

06

00

0m

m

DIS

PL

AY

:0

00

00

%

AC

TU

AL

02

87

0m

m

OU

TP

UT

:2

0.0

0m

A

LE

VE

L:

00

25

0m

m

DIS

PL

AY

:0

01

00

%

UN

ITS

LE

VE

L:

mm

DIS

PL

AY

:%

DA

MP

ING

ME

DIU

M T

EM

PE

RA

TU

RE

FA

ILU

RE

MO

DE

HA

RT

FA

CT

OR

Y D

EF

AU

LT

RE

SE

T

TE

AC

HIN

G

TA

NK

MU

ST

BE

EM

PT

Y

ST

AR

T

FA

ILU

RE

MO

DE

NO

EC

HO

: 4

.00

mA

HA

RT

RE

SE

T AR

E Y

OU

SU

RE

?

FA

CT

OR

Y D

EF

AU

LT

AR

E Y

OU

SU

RE

?

DIS

TA

NC

E T

O L

EV

EL

01

22

0 m

m

DIS

TA

NC

E T

O L

EV

EL

CU

RR

EN

TT

EM

PE

RA

TU

RE

+2

C

CL

ON

E S

ET

TIN

G

SE

NS

OR

DIS

PL

AY

MO

DU

LE

DIS

PL

AY

MO

DU

LE

SE

NS

OR

→ →

PA

SS

WO

RD

EN

TE

R

00

00

0

LA

NG

UA

GE

EN

GL

ISH

PO

LL

ING

AD

DR

ES

S

SE

NS

ITIV

ITY

SE

NS

ITIV

ITY

: M

ED

IUM

1 2

3

4 5

6!

7!

8!

MA

TE

RIA

L:

LIQ

UID

ME

DIU

M T

EM

PE

RA

TU

RE

YE

S

+2

C

CU

RR

EN

T 17

.30

mA

Page 29: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

28© Dinel, s.r.o. ULM–70

TecHnical specificaTions – leVel MeTer

Measuringrange1)ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20

0.15 ... 2 m0.25...6m0.4 ... 10 m0.5 ... 20 m

Adjustablemeasuringrange(SPAN) Min.200mm

Supplyvoltage ULM–70N–__ULM–70Xi–__

18...36VDC18 ... 30 V DC

Output ULM–70_–__-_-IULM–70_–__-_-M

4...20mA(Limitvalues3.9...20.5mA),HART®

RS-485withprotocolModbusRTU

Current consumption ULM–70_–__-_-IULM–70_–__-_-M

4...20mA/Max.22mAMax.20mA

ResolutionULM–70_–02;10ULM–70_–06ULM–70_–20

<1mm<2mm<2.5mm

Accuracy(withinthetotalrange) 0.15 %

Temperature error Max.0.04%/K

Operatingfrequency

ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20

120kHz75kHz50kHz30kHz

Beamwidth(-3dB)ULM–70_–02;10ULM–70_–06ULM–70_–20

10°14°12°

Ambienttemperaturerange ULM–70_–02;06ULM–70_–10;20

-30...+70°C-30...+60°C

Short-timetemperaturestressresistance +90°C/1hour

Max.operationoverpressure(ontransmissionsurface) 0.1MPa

Sensitivity 3steps(low–medium–high)

Damping 0 ... 99 s

Measuringperiod 1 ... 4 s

Risetime ca. 30 s

AdditionaltechnicaldataforExproof 2)–Max.internalvalues Ui=30VDC;Ii=132mA;Pi=0.99W;Ci=370nF;Li=0.9mH

Failure indication(echoloss,levelindeadzone3),internalfailure)

Adjustableinmodes:3.75mA;22mA;Lastmeasuredvalue

Maximalresistanceofcurrentoutputload(U=24VDC) Rmax=270Ω4)

Mechanicalconnection

ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20

ScrewingwiththreadG1''ScrewingwiththreadG1½''ScrewingwiththreadG2¼''Aluminiumalloyflange

1) In case the level of bulk-solid materials is measured, the measurement range is reduced.2) Allowed pressure range in the zone 0: 80 ... 110 kPa.3) Dead zone = Blind zone = Blocking distance4) Including 250Ω resistor in case of HART® connection.

20 .  specificaTions

Page 30: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

ULM–70 © Dinel, s.r.o.29

facTory DefaulT

ULM–70_–02 ULM–70_–06 ULM–70_–10 ULM–70_–20

MINLEVEL(Distancetomin.level) 2 000 6000 10 000 20 000

MAXLEVEL(Distancetomax.level) 150 250 400 500

UNITS mm;%;°C mm;%;°C mm;%;°C mm;%;°C

DAMPING 2 5 10 10

SENSITIVITY MEDIUM MEDIUM MEDIUM MEDIUM

MEDIUMTEMPERATURE NO NO NO NO

FAILUREMODE–NOECHO 3.75mA 3.75mA 3.75mA 3.75mA

FAILUREMODE–DEADZONE1) 22mA 22mA 22mA 22mA

POOLINGADDRESS(HART®) 00 00 00 00

PASSWORD Nopassword Nopassword Nopassword Nopassword1) Dead zone = Blind zone = Blocking distance

area classificaTion (according to EN 60079-10 and EN 60079-14)

ULM–70N–__ Performance for non-explosive areas

ULM–70Xi–02ULM–70Xi–06

Explosiveproof–suitableforexplosiveareas(combustiblegasesorvapours)II1/2GExiaIIBT5Ga/GbwithIsolatingrepeater(IRU–420),thewholelevelmeter

–zone1,frontheadpart–zone0

ULM–70Xi–10Explosiveproof–suitableforexplosiveareas(combustiblegasesorvapours)

II1/2GExiaIIAT5Ga/GbwithIsolatingrepeater(IRU–420),thewholelevelmeter –zone1,frontheadpart–zone0

ULM–70Xi–20 Explosiveproof–suitableforexplosiveareas(combustiblegasesorvapours)II2GExiaIIAT5GbwithIsolatingrepeater(IRU–420),thewholelevelmeter–zone1

TecHnical specificaTions – Display MoDule

Displaytype MatrixOLED

Resolution 128x64pixel

Characterheight/Numberofdigitsmeasuredvalue 9mm/5Digits

Displaycolour Yellow

Buttons Membraneswitchpanel

Ambienttemperaturerange -30...+70°C

Weight 46g

TecHnical specificaTions – leVel MeTer

Protection class IP67

Weight

ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20

0.3kg0.4kg0.6kg3.1kg

Page 31: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

30© Dinel, s.r.o. ULM–70

useD syMbols

Toensuremaximumsafetyofcontrolprocesses,wehavedefinedthefollowingsafetyinstructionsandinformation.Eachinstructionislabeledwiththeappropriatepictogram.

Alert, warning, danger

Thissymbol informsyouaboutparticularly important instructionsfor installationandop-erationofequipmentordangeroussituations thatmayoccurduring the installationandoperation.Notobservingtheseinstructionsmaycausedisturbance,damageordestructionofequipmentormaycauseinjury

Information

Thissymbolindicatesparticularlyimportantcharacteristicsofthedevice.

Note

Thissymbolindicateshelpfuladditionalinformation.

21 .  packaging, sHipping anD sTorage

ThedeviceDLM–35ispackagedinapolyethylenebag,andtheentireconsignmentisplacedintoacardboardbox.Asuitablefillermaterialisusedinthecardboardboxtopreventmechanicaldamageduring transport.

Removethedevicefromthepackagingonlyjustbeforeusing,therebyprotectingitfrompotentialdamage.

Aforwardingcompanywillbeusedtoshipgoodstothecustomer.Uponprioragreement,orderedgoodscanbepickedupinpersonatcompanyheadquarters.Whenreceiving,pleasechecktoseethattheconsignmentiscompleteandmatchestheorder,ortoseeifanydamagehasoccurredtothepackaginganddeviceduringtransport.Donotuseadeviceclearlydamagedduringtransport,butrathercontactthemanufacturerinordertoresolvethesituation.

If thedeviceistobefurthershipped, itmustbewrappedin itsoriginalpackagingandprotectedagainst impact and weather conditions.

Storethedeviceinitsoriginalpackagingindryareascoveredfromweatherconditions,withhumid-ityofupto85%withouteffectsofchemicallyactivesubstances.Thestoragetemperaturerangeis-20°Cto+60°C.

Level meters of variants ULM–53_– 02, 06, 10 are fitted with protective caps to prevent damage to the ultrasonic transducer. Remove the cover prior to putting into operation.

Page 32: INSTRUCTION MANUAL - Dinel, s.r.o. - Měření hladiny a … the electroacoustic converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the surface

Dinel, s.r.o.U Tescomy 249

760 01 ZlínCzech Republic

Tel.: +420 577 002 003Fax: +420 577 002 007E-mail: [email protected]

www.dinel.cz

The lastest version of this instruction manual can be found at www.dinel.czVersion: