instruction manual en - testo · 7 preface dear testo customer, your decision to buy the testo 300...
TRANSCRIPT
testo 300 XXL
Instruction manual en
3
Contents
PageContents..........................................................................................................................3Preface ............................................................................................................................7Principal safety information ..............................................................................................8System description 9
Control unit ..............................................................................................................9Device operation ....................................................................................................10Touchscreen (option) ..............................................................................................11Overview of the measuring system..........................................................................12Menu overview 14
Control unit with analyser box..........................................................................14Control unit......................................................................................................15
Commissioning ..............................................................................................................16Function menu and function bar..............................................................................16Status display of the LED (analyser box)..................................................................17Power supply ..........................................................................................................18
Power supply, analyser box/control unit ..........................................................18Mains operation ......................................................................................18Rechargeable battery operation ..............................................................18
Power supply, control unit................................................................................18Mains operation ......................................................................................18Rechargeable battery operation with battery pack ..................................18Battery operation ....................................................................................18Rechargeable battery operation with standard batteries..........................18
Charging batteries ..................................................................................................19Charging batteries, analyser box/control unit ..................................................19Charging battery pack, control unit..................................................................19
Example measurement ..................................................................................................20Correct use of the flue gas probe............................................................................20Preparing for measurements with the control unit and analyser box........................21Core current search with the control unit and analyser box ....................................22Flue gas measurement at burners with control unit and analyser box ....................23Conducting a CO measurement in gas burners with the control unit and the analyser box ..............................................................24Draught measurement with the control unit and analyser box ................................25
In compliance with the conformity certificate, the units fulfil the guidelines ofDIN EN 61010-1 safety regulations for electrical devices, control, regulating
and laboratory equipment of guideline 73/23/EEC.
In compliance with the conformity certificate, the unit fulfils the guidelines of2004/108/EEC.
PageDifferential pressure measurement with the control unit ..........................................26Differential pressure measurement with the control unit ..........................................27CO measurement with the control unit ....................................................................28Gas leak probe with control unit..............................................................................29Humidity measurement with the control unit............................................................30Velocity measurement with the control unit..............................................................31
Volume measurement in a duct ......................................................................32Mean value ......................................................................................................33Volume flow measurement ..............................................................................34Velocity measurement with the pitot tube ........................................................35
CO2 measurement with the control unit ..................................................................36Current/voltage measurement with the control unit ................................................37
Printing ..........................................................................................................................38Printing the measurement results ............................................................................38
Main menu Memory (analyser box and control unit) ......................................................39Overview ................................................................................................................39Saving without selecting a location..........................................................................40Creating a new location ..........................................................................................41Selecting saved location..........................................................................................42Using an existing location as format for a new location ..........................................43Changing existing location ......................................................................................44Deleting existing location ........................................................................................45Barcode pen overview ............................................................................................46Reading a location with the barcode pen ................................................................47Reading a location from the measurement menu ....................................................47Reading several locations with barcode pen from the measurement menu ............47Displaying saved readings ......................................................................................48Mean calculation ....................................................................................................49Printing mean calculation result ..............................................................................50Deleting entire memory............................................................................................51Printing saved readings of a location ......................................................................52Free memory? ........................................................................................................53
Main menu Probe for control unit ..................................................................................54Scaling ....................................................................................................................54Resetting ................................................................................................................55Info..........................................................................................................................55
Main menu Input for control unit ....................................................................................56Parameters..............................................................................................................56
Temperature ....................................................................................................56Humidity ........................................................................................................57Pressure ..........................................................................................................58Density ............................................................................................................59
Contents
4
5
Contents
Page
Pitot tube factor ..............................................................................................60Cross section ..................................................................................................61Offset factor ....................................................................................................62Info ..................................................................................................................62
Alarm limits..............................................................................................................63Timer ......................................................................................................................63
Main menu Device for control unit ................................................................................64Changing date/time ................................................................................................64
Setting the date ..............................................................................................64Setting the time ..............................................................................................66
Auto off ..................................................................................................................66Printer ....................................................................................................................67
Setting the contrast ........................................................................................67Entering print text ............................................................................................67
Light........................................................................................................................68Diagnostic ..............................................................................................................68Units ......................................................................................................................69
Main menu Service for control unit ................................................................................70Operating values ....................................................................................................70Resetting factory settings ........................................................................................70Address ..................................................................................................................70Device data ............................................................................................................70Language ................................................................................................................71
Main menu Periphery (with touchscreen only) ................................................................71Solid fuel measurement option for analyser box ............................................................72
CO measuring ranges in different dilution stages ....................................................73Main menu Sensors for analyser box and control unit ..................................................73
Recalibration ..........................................................................................................74NO2 addition ..........................................................................................................75Print cal. data..........................................................................................................75
Main menu Input for analyser box and control unit ........................................................76Spot number index / WTT / oil derivatives ..............................................................76Fuel ........................................................................................................................77O2rel/CO2max/coefficient ........................................................................................77Parameters..............................................................................................................79
Pressure ..........................................................................................................79Pitot tube factor ..............................................................................................80Cross section ..................................................................................................81Offset factor ....................................................................................................82Info ..................................................................................................................83
Dew point AT ..........................................................................................................84Main menu Device for analyser box and control unit......................................................85
View (display sequence) ..........................................................................................85Diagnostic ..............................................................................................................86
6
Contents
PageMain menu Service for analyser box and control unit ....................................................87
Operating values ....................................................................................................87Switching off CO/NO ..............................................................................................88Device data ............................................................................................................89Bus address............................................................................................................89
Maintenance ..................................................................................................................90Inserting new paper roll in printer ............................................................................90Changing the battery pack ......................................................................................91
Control unit......................................................................................................91Analyser box....................................................................................................91
Filter change ..........................................................................................................92Condensation trap ..................................................................................................92Flue gas probe ........................................................................................................93
Cleaning the flue gas probe ............................................................................93Cleaning the flue gas probe with gas path closed............................................93Changing a defective thermocouple ................................................................93
Changing O2, CO and NO measuring cells..............................................................94Installation of CO/NO2 measuring cell ....................................................................94Installation of NO measuring cell ............................................................................94Maintenance of the gas pumps ..............................................................................95
Opening the analyser box ................................................................................95Cleaning the main gas pump and the purging pump ......................................95Cleaning the main gas pump with optional gas preparation ............................96
Changing the pump cassette of the condensation pumpwith optional gas preparation ..................................................................................97Upgrades ................................................................................................................98
Calculation information ..................................................................................................99Error messages ..........................................................................................................100Technical data..............................................................................................................101Ordering data ..............................................................................................................106Notes
7
Preface
Dear Testo customer,
Your decision to buy the testo 300 XXL was an excellent choice.
The testo 300 XXL is designed for flue gas analysis and all measurements inheating systems.
Thousands of customers buy our high-quality products every year. There are atleast 7 good reasons for this:
1) Our price/performance ratio is good. Reliable quality at a fair price.
2) Substantially extended warranty periods up to 2 years – depending on the model.
3) We provide an optimum solution to your measuring tasks with over 40 years ofprofessional experience.
4) Our high quality standards are confirmed by the ISO 9001 certificate.
5) Of course, our devices also bear the CE label required by the EU.
6) Calibration certificates for all relevant measurements. Seminars, consulting andcalibration on site.
7) We don’t leave you on your own after your purchase.
Use the device only in the operating andmeasuring ranges specified in “Technical data”.
Use the device only under the conditions and forthe purpose for which it was designed. For this,particularly observe the safety information and thetechnical data.
To prevent damage to the instrumentation anditems under test, suitably trained persons shouldonly use the equipment.
Observe all local Health and Safety regulationsbefore operating the test equipment.
Do not make measurements on live components.
Observe the warnings on the control unit andanalyser box!
Magnets of the control unitThe magnets of the control unit can damage ECcards, telephone cards etc.
Operation with mains adapter Use only the original mains adapter to operate thedevice. Replace damaged mains cable only withan original mains cable of the manufacturer oragent (i.e. by authorised service).
Check for leaksThe entire measuring system (probe, condensationtrap, filter, hoses and connections) must bechecked for leaks before each measurement, e.g.by pushing on a collapsed rubber bladder.Measuring errors can be caused if external airinfiltrates.
Expose the gas outletWhen measuring, ensure that the gas outlets areexposed to allow the gas to escape unobstructed.If this is not the case, the measurement resultscan be falsified.
Empty the condensation trapThe condensation trap must only be emptied whenthe pump is switched off (the measuring cells areotherwise in danger of damage)!
Disposing of batteriesThe end user is legally required to dispose of allused batteries in accordance with their country’sregulations.
Disposing of measuring cellsStoring the analyser in rooms containing solvents,vapours or gases will damage the measurementcells e.g. cleaning fluids, polishes, paints, cookingoperations. Protect the device from direct sunlightand store at temperatures <45°C.
Storing the deviceStoring the analyser in rooms containing solvents,vapours or gases will damage the measurementcells e.g. cleaning fluids, polishes, paints, cookingoperations. Protect the device from direct sunlightand store at temperatures <45°C.
Charge the battery before useBefore making the first measurement or afterseveral days of disuse, fully charge the internalbattery (see menu Service Operating values). Charge the battery at leastevery 4 weeks during longer periods of disuse toprevent damagingly low discharging.
Changing the configurationSwitch off the device to change the configuration(e.g. probe change), as the device reads theprobe-specific values when it is switched on.
PC cable connectedWhen the PC cable is connected, it is impossibleto issue a print command.
Using probes with plug headEnsure the good contact of probes with plugheads. Fully tighten the union nut on the probehandle.
Opening the deviceIf the device is opened, misused or subjected toforce, the warranty will be voided.
Because the pressure sensor heats up, it isrecommended to wait on the “Warm up” for 15minutes to achieve the best possible accuracy.
8
Principal safety information
System descriptionControl unit
9
On/off button
Function keys
Function bar
Touchscreen (option)
on the rear:Contacts to analyser box
Reading display
Device active display
Touchscreen penholder
Printer
Reading window
Display lighting
Mains adapter connection
Pressure connection
Probe connection
Cancel/back
PC connection
Main menu display
Status display
Location
Acknowledgement Cursor keys
Connection of the spiral BUScable to the analyser box
10
System descriptionDevice operation
The analyser measurement box is controlled, readand programmed using the control unit.
The operation of the testo 300 XXL is designed sothat the key functions are visible at all times onthe display.
The keypadThe function keys allow rapid changes of thefunctions of the device and the overall control ofthe measurement. The significance of the functionkeys is indicated in the function bar. Not allfunction keys are used in every menu. Arrows inthe function bar indicate further accessiblefunctions in the menu. When the right orleft cursor key is pressed, these additionalfunctions are displayed and can be activated withthe function key.
With the up /down cursor keys, youcan browse between the reading windows (in themeasurement menu) or select a menu item in alist. When you have reached the lastwindow/menu item, the device automatically loopsback to the first window or the first menu item.
With the Browse key, you can branch fromthe measurement menu to the main menu or froma submenu to the measurement menu.
The I/O button switches the device on and off.
With the Cancel key, you can terminateselected procedures or a chosen selection and exitsubmenus. When you exit a submenu, you willalways move one menu window backwards untilyou reach the main menu.
With the Acknowledge key, the selection orthe execution of the selected function is accepted.If the key is pressed in the measurementmenu, the menu “Device selection list” appears.The menu provides information on which device isin use. 001=Control unit active, 002=Analyser boxactive etc. If an analyser box is connected, this isindicated automatically in the window when thedevice is started.
With the Lighting key , the display lightingcan be switched on and off or the automatic timercan be started.
The displayThe display constantly shows all importantinformation.
In the Reading/Menu display area, 6 readingsper window are indicated in the measurementmenu. Only three readings are indicated in largerlettering when the zoom function is used (byfunction key).
The function bar indicates the significance of theindividual function keys . Their effects canchange according to the menu. The description ofthe function keys changes accordingly on thedisplay.
The function bar can be assigned as desired toall functions of the function menu. When the key
is pressed together with a function key, thefunction menu opens. The entries in the functionmenu change according to the connectedcomponents of the measuring system.
The Operation display provides a graphic de-scription of the current mode of the device, e.g.whether the pump of the device is running and ifthe device is running in mains operation.
The Location box constantly shows whether andwhich location hasn’t been selected and is active.If no location has been selected, “NONAME”appears on the display.
In the measurement menu, the Reading windowon the right indicates the measured value windowin which you are currently working, e.g. 2/5indicates that you are working in the second of 5possible pages.
Errors are indicated on the display as soon as theyoccur. The “!” symbol in the status display oftenindicates the occurrence of errors.
The touchscreen function (option) allowsdirect access to the functions indicated on thedisplay using a special pen. A brieftouch is sufficient to activate the function.
11
System descriptionTouchscreen (option)
Touch a position on the display with the specialpen. The main menu appears. Each menu itemcan be accessed and activated with the specialpen. The example shows the individual steps toset the lighting status of the control unit.
Measurement menu
Measurement menu
Main menu
Change “Device” settings
Change “Device” settings
or
Adjustment menu Adjust “Light”
Adjust “Light”
Set “On/Off” Set automatic
PPeerriipphheerryy
PPeerriipphheerryy
12
System descriptionOverview of the measuring system
Measuring cell heaterprotects againstdestruction bycondensation
Gas connection
DP measurement
Spiral BUS cable connection
Flue gas probeconnectionProbe
connection
110/230 V50/60 Hz
Carryinghandle
Control unitcatch
Contact strips to thecontrol unit
LED status display
Gas path filter
Gas outlet 2
Fresh air inlet filterFresh air inlet
Diluting air filterGas outlet 1
Condensation tank
13
System descriptionOverview of the measuring system
0
12.1 1.2792.4
180.1 22.2
BrSt
l
AT °C VT °C
h%
00204/06
Noname
HoldMax
MinZoom
Volume flowratemeasurementwith funnel
Pressure measurement
Temperaturemeasurement
CO safetymeasurement
Gas leak measurement
CO2/humiditymeasurement
Current measurementVoltagemeasurement
Cable connection controlunit/spiral analyser box cable
Control unit
4 user-definedfunction keys
Mains connection/fast battery charge
Integrated differentialpressure measurement
Integrated printer
Multi-probeinput
RS 232interface
Display
14
MEMORY
SENSORS
INPUT
DEVICE
SERVICE
MMaaiinn mmeennuu
Op. values Switch-off Device data Bus address
Display sequenceUnitsView
Diagnostic
Spot number/HCTFuelO2rel/CO2maxParameter Dewpoint/ambient
Smoke number 1Smoke number 2Smoke number 3Heat carrier temp.Oil derivative Info
O2 rel. CO2max. coeff.
Temperature VLHumidity VL Dewpoint VL
Pres. Pitot tube factorCross-sectionOffset factor Info
Recal. NO2 additionPrint cal. data
CONO
CO NO
Read outMean valueFuel meas.Delete memoryFree memory?
System descriptionMenu overviewControl unit with analyser box
from height Absolute
Pres. Metres abovem.s.l.Diff. pressure
DurationNo. valuesInfoSaveDelete Duration
DilutionInfoSaveDeleteYes
No
15
MEMORY
SENSORS
INPUT
DEVICE
SERVICE
PERIPHERY*
MMaaiinn mmeennuu
Op. valuesReset factoryAddressDevice dataLang.
Change dateAuto OffPrinterLightDiagnosticUnits
Touchscreen cal.
Set date
°C, °F
m/s, fpm
ppm, %
Off, td°C, g/m3,g/kg, J/g
m3/h, cfm, m3/m,m3/s, l/s, M3/h
hPa, inW, mbar,Pa, bar, psi
ParameterAlarm limitsTimer
Temp.Humid.Pres.DensityPitot tube factorCross-sectionOffset factorInfo
CO limitCO2 limitdP limit 1dP limit 2Acoustic alarm
DurationNo./valuesInfoSaveDelete
CircleSquareRectangleArea
from heightabsolute
PressureMetres above m.s.l.Diff. pressure
contrastPrint text
On/offAutomatic
Temp.Humid.VelocityVol.-FlowPressureGas
yesno
Probe socket 1
min inputmax inputmin outputmax outputUnitDecimal placeInfoSaveDelete
ScalingResetInfo
Read outMeanDelete memoryFree memory?
System descriptionMenu overview
Control unit
Off5 Min.
10 Min.....
*with touchscreen option only
Function menu and function bar
Commissioning
16
The function bar can be assigned as desired to all functionsof the function menu. Pressing the key with a functionkey opens the function menu. The entries in the functionmenu change according to the connected components ofthe measuring system.
FUNCTION MENU ANALYSER BOX/CONTROL UNITPump starts. Flue gas measurement runs.
Doubles the character size.
Saves readings.
Differential temperature measurement with 2 probe inputs ofthe analyser box.
Graphic display of the temperature during core currentsearch.
Fine draught with 0.001 hPa resolution.
Starts solid fuel measurement.
Prints readings.
Printer paper line feed.
Re-zeros the measuring cell.
Fresh air purging of the measuring cells.
Starts CO undiluted measurement.
FUNCTION MENU, CONTROL UNIT
Doubles the character size.
Freezes the current readings.
Shows the highest value during a measurement.
Shows the lowest value during a measurement.
Calculates mean value.
Additional display for volume flow rate measurement.
Displays in metres per second.
Zeros external pressure probe.
Zeros CO probe.
Starts solid fuel measurement.
Analyser box is detected without restarting the device.
PPSSttaarrtt
uuCCOO oonn
SSttaarrtt
MMeemm..
DDeellttaa TT
CCoorrccuurr
DDrraagghhtt
PPrriinntt
ZZeerroo
GGaass
ZZoooomm
ZZoooomm
HHoolldd
MMiinn
MMaaxx
VVooll
MMeeaann
mm//ss
hhPPaa==00
ppppmm==00
SSeeaarrcchh
SSttaarrtt
LLFF PPrr
17
Commissioning
Function menu and function bar
LED 1
LED 2
LED 3
DDeellttaa TT
MMeemm..
ddPP 11
ddPP 22
PPrriinntt
LLFF PPrr
DDiiaagg..
EESSCC
OOKK
TTeesstt
CChhaannggee
EEnndd
IInnffoo
aa AA
<<
Status display of the LED
Saves readings.
T1 + T2, displays differential temperature.
Differential pressure 0 - 80 mbar
Differential pressure 0 - 1000 mbar
Prints readings.
Printer paper line feed
Direct display of all error messages
FUNCTION BAR
Cursor up (optional with touchscreen)
Cursor down (optional with touchscreen)
Menu exit (optional with touchscreen)
Confirms input (optional with touchscreen).
Set values come into effect
Change
Prints test page.
Accepts settings and closes window.
Upper/lower case switch
Back + deletes 1 character.
Space
Overview of settings
Sets standard values.
Current value comes into effect
CCuurrrr..
LED1 (Power):
Mains operation Green/Permanent
Battery operation (batt. full) Green/Flashing
Battery operation (batt. empty) Red/Flashing
Battery recharging, Off mode Off
LED2 (Status):
Measuring Green/Permanent
Fresh air/Zeroing Green/Flashing
Defect Red/Flashing
LED3 (Battery recharging):
Battery recharging (fast charge) Green, flashing
Batt. full, compensation charge Green, permanent
18
CommissioningPower supply
115/230 V
Mainsadapter0554.1084
or
or
Plug mains cable into analyser box. The power supply forthe control unit is ensured when- control unit is connected to the analyser box by the
contact strips or- control unit is connected to the analyser box by the spiral
BUS cable
The power supply for the control unit in rechargeablebattery operation is ensured when:- control unit is connected to the analyser box by the
contact strips or- control unit is connected to the analyser box by the spiral
BUS cable
Observe battery capacity.Ensure correct polarity.
There should always be rechargeable batteries/batteries inthe control unit.
Power supply, analyser box/control unit
Control unit power supply
MAINS OPERATION
Connect mains adapter to the control unit.
MAINS OPERATION
RECHARGEABLE BATTERY OPERATION
RECHARGEABLE BATTERY OPERATION with battery pack(art. no. 0554.0097)
RECHARGEABLE BATTERY OPERATION WITH STANDARD
BATTERIES
BATTERY OPERATION
19
Battery charging
Commissioning
Charging batteries, analyser box/control unit
Charging battery pack, control unit
The battery packs must be installed in the analyser box andthe control unit for charging.
Plug mains cable into analyser box.
NoteThe units must be switched off. Charging is impossibleduring operation.
The charging of the battery pack for the control unit isensured when:- control unit is connected to the analyser box by the
contact strips or- control unit is connected to the analyser box by the
spiral BUS cable.
Charge level displays:- Analysis box:
Recharging is in progress: Recharging battery LEDflashes green, Recharging is complete: Recharging battery LED lights up green
- Control unit: Shows charge level in display.
The battery pack must be installed in the control unit forcharging.
Standard batteries cannot be charged.
Plug mains adapter into the control unit.
NoteThe control unit must be switched off. Charging isimpossible during operation.
Charge level display:- Charge level is shown in display.
115/230 V
Mainsadapter0554.1084
The end user is legally required to dispose of all used batteries inin accordance with their country’s regulations.
CChhaarrggiinngg CCoonnttrrooll UUnniitt
CChhaarrggiinngg CCoonnttrrooll UUnniitt
20
Example measurementCorrect use of the flue gas probe
The measurement of the flue gas temperature is made bythe thermocouple at the tip of the flue gas probe. The probetube protects the thermocouple, but has openings in toallow the stream of flue gas to envelop the thermocouple.
Thetemperature
measurement is made during the zeroing phase. The fluegas probe can be adjusted in the flue at this time (Note: theAT probe must be connected). This makes it possible toconduct a core current search during the zeroing phase. Thetemperature measured by the flue gas probe is interpretedby the testo 300 XXL as the combustion air temperatureunless a AT probe is connected (Note: the probe must beremoved from the flue). When the zeroing phase has beencompleted, this value is saved as the combustion airtemperature value.
All dependent parameters are calculated by this value.
incorrect
Flue gasstream
Thermocoupleincorrect
correct
correct
The thermocouple must always be exposed to the streamof flue gas to allow an exact measurement of the fluegas temperature and thereby a precise determination of
the flue gas losses. It must not be screened by a web ofthe probe pipe.
The tip of the thermocouple must not touch the guard. Ifnecessary, bend the tip of the thermocouple accordingly.
Example measurement
21
Preparing for measurementwith the control unit and analyser box
Control unit + analyser box + flue gas probe
1 Connect flue gas probe to analyser box.
2 Switch on .
3 Zeroing phaseAny connected CO probe and the gas measuring cells ofthe analyser box are zeroed during the zeroing phase.
4 Fuel selectionThe menu “Fuel selection” appears during the zeroingphase. Select the desired fuel with the keys or andconfirm with .
5. Back to the zeroing phase.
NoteAfter zeroing phase back to the measurement menu withthe key.
10 sec
Flue gas probe
Analyser box
Initialisation
Zeroing phase
Select the fuel
Connect the flue gas probe before switching on the controlunit and the analyser box.
Note
The analyser box does not need to be activated in thedevice selection menu. It is detected automatically whenthe measuring unit starts.
LLooccaattiioonn 33
22
Example measurement
Starting core current search:
1 Connect flue gas probe to analyser box.
2 Place flue gas probe in flue pipe.
3 Activate the function menu with the key and one of
the function keys. Select function key with and confirm with .
4 Start core current search with .
5 Determine core current temperature.
6 Terminate core current search with .
Note
A core current search is also possible during the zeroingphase.
Corcur
Corcur
Core current searchwith the control unit and analyser box
Measurement menu
Measurement menu
Core current menu
Select core current
10 sec
Initialisation
Device zeroing
Automatic fuel display
Zeroing phase
3 x
1 sec
Function menu
CCoorrccuurr
LLooccaattiioonn 33
LLooccaattiioonn 33
Preparing for measurement1 Specify new location and select (see page 41):
2 The four function keys of the control unit are freelyconfigurable with the functions of the function menu.
3 Activate the function menu with the key and one of
the function keys. Select function key with and confirm with .
4 Start measurement with .
5 Position the flue gas probe in the core current:Seek flue gas temperature AT max, then fasten the probewith the cone in this position.
6 When the readings no longer change, stop the pump.Place flue gas probe in fresh air. The readings are retained until the pump is next started.
PStart
PStart
Example measurementFlue gas measurement at burners
with the control unit and the analyser box
23
Back to the measurement menu
If the flue gas probe is positioned vertically or at an angleof up to 45°, falling drops of condensation can cause a
sudden drop in the temperature.
Remedy: Readjust flue gas probe.
Measurement menu
Location
Measurement menu
Function menu
Measurement menu
Measurement menu
1 sec
Conducting a CO measurement in gas burners (flue gas path test) with the control unit and analyser box
1 Activate the function menu with with the key andone of the function keys. Select function key with and confirm with .
2 Start measurement with .
3 Measure undiluted CO with .Display of the function key changes to .
4 Finish undiluted CO measurement with .Undiluted CO is saved.
5 Finish measurement with . PStop
uCO off
uCO off
uCO on
PStart
uCO on
uCO off
uCO on
24
Example measurement
Measurement menu
Function menu
Measurement menu
CO limit for the rejection of a system (in relation toundiluted flue gas): > 1000 ppm uCO
Draught measurementwith the control unit and the analyser box
25
Preparing for measurement
Use the flue gas probe for the draught measurement. Thedraught measurement can be conducted before or after aflue gas measurement.
Start measurement
1 Activate the function menu with the key and one of
the function keys. Select function key with and confirm with .
2 Start measurement with .
3 Place probe in flue.
4 Display of the current reading.
Draught
Draught
Example measurement
A draught measurement is only possible with the pumpswitched off. The + pressure inlet of the device must beopen!
Do not change between battery and mains operationduring the draught measurement (voltage fluctuationsinfluence the result of the measurement)!
If you conduct the draught measurement after a flue gasmeasurement, a head of pressure exists in the hose afterthe pump has stopped, which must be dispelled. Thisoccurs within approx. 30 seconds.
Remove any traces of condensation in the flue gas probe(shake out the probe with the tip towards the floor).
Sensor zeros automatically, probe must be outside the flue.
Negative sign indicates vacuum, positive sign indicates pressure in the flue.
Select “Draught”
Measurement menu
Flue gas probeoutside flue
Place probe in flue
Stable reading?
Autom. zeroing
Result
Measurement menu
Function menu
1 sec
DDrrgghhtt
Preparing for measurement
Selection of a suitable measuring range. Two measuring ranges P1 (-80 to +80 hPa) and P2 (-1000to +1000 hPa) are available in the function menu:
Start measurement1 Device selection: control unit
2 Activate the function menu with the key and one ofthe function keys. Select the function key (lowrange) or (high range) with andconfirm with .
3 Zero sensor with function key .
4 Start measurement with function key , stopmeasurement with function key or the measure-ment is terminated automatically when the time hasexpired (Input menu - Timer).
5 Display and printing of the current reading.
TStop
TStart
dP1
dP2
dP1
26
Differential pressure measurement with the control unit
Example measurement
Do not use in the main duct!
Ensure that there are no leaks between the samplingpoint and the device (escaping gas and atmosphericoxygen can create an explosive mixture)!
Do not smoke or use naked flames during themeasurement!
Observe the rated pressure of the connections!
Do not exceed the measuring range!
No gas enters the device because an air cushion formsbetween the sensor and the gas inlet. The measuring unitis therefore not protected against explosions.
Measurement menu
Measurement menu
Measurement menu
Function menu
Select “dP1” and/or“dP2”
Automatic sensorzeroing
Start measurementStop measurement Print
1 sec
10 sec
dP1
TTSSttaarrtt
TTSSttoopp
01:00 min
00:27 min
27
Measurement with control unit + temperature probe
For pure temperature measurement, e.g. supply / returntemperature
The testo 300 XXL control unit has a multi-probe input.
1 Activate the function menu with the key and one ofthe function keys. Select the function key with and confirm with .
2 Measure T1: Press function key .
3 Measure T2: Press function key .
4 Display of the measured differential temperature value
Measurement with control unit + analyser box + 2temperature probes
Device selection: flue gas measuring box
1 Configure function key with (see above)
2 display of the current value of Temp.1 (probeconnection), Temp.2 (flue gas probe connection) and thedifferential temperature.
DeltaT
DeltaT
T2
T1
DeltaT
Differential temperature measurementwith the control unit
DisplayTemp. 1
DisplayTemp. 2
Before switching on the testo 300 XXL, connect thetemperature probe!
Example measurement
T 1 T 2
Measurement menu
Differential temperature
Measurement menu
Measurement menu
Function menu
Select “Delta T”
1 sec
DDeellttaa TT
TT 11
TT 22
4499..88
1155..44
3344..44
Measurement with: control unit + CO probe
1 - Initialisation - Zeroing of the CO probe (60 sec) After zeroing, the CO probe is ready for measuring.
2 Open cap.
Note The direction of flow of the gas to the probe influencesthe measuring accuracy! Optimum measurement resultsare achieved when the probe is moved gently backwardsand forwards. Frontal flow onto the probe leads toexcessive readings!
3 Close cap after the measurement.
28
CO measurementwith the control unit
Example measurement
Connect probe before switching on the control unit!
Cap must be closed during zeroing (measuring errorotherwise possible).
Open the cap only during the measurement, then replace.
Warning message when adjusted CO value is exceeded(change limit in Main menu “Input - Alarm limits - CO
limit”).
10 sec
Initialisation
CO probe zeroing
60 sec
Measurement menu
110000
Explosive and flammable gases, particularly natural gas andpropane, are detected by the gas leak probe even in lowconcentrations in the air. For example, suitable for findingleaks in gas pipes, vessels and equipment.1 Connect the probe before switching on the
testo 300 XXL.
2 After the device has been switched on, the warming upphase of the leak detection probe is automatically started(approx. 30 sec). Green, yellow and red LED, steadytone.
3 Ready for use Green LED flashes steadily, steady tone stops.
4 Conduct leak detection:• When gas escapes, acoustic signal
(ticking) is audible, becoming faster with increasingconcentration.
• When the 1st switching limit (> 200 ppm) is exceeded,yellow LED flashes.
• When the 2nd switching limit (> 1%) is exceeded, redLED flashes, steady tone. No readings of the gas leakprobe are indicated in the display.
The power supply of the gas leak probe is provided by thecontrol unit.
29
Service tip:The trigger limit is factoryadjusted to 200 ppmCH4 gas.
The 1st alarm limit canbe finely adjusted in arange from 0...1000 ppm(perform in fresh air only).
Example measurementGas leak probe
with control unit
Do not use the gas leak probe and testo 300 XXL inenclosed rooms or facilities in which gases have
accumulated into an explosive mixture.
The use of other electrical equipment is also prohibited.
It must be ensured that the gas concentration does notexceed 20% LEL.
Duration of measurement with the gas leak probe with fullycharged battery, maximum 2 hours.
Remove the gas leak probe from the deviceimmediately after the measurement to prevent heavybattery discharge.
Never leave the gas leak probe connected when chargingthe batteries.
red LED
yellow LED
green LED
Trimmer potentiometerfor fine adjustment ofthe 1st alarm limit.
30
Example measurementHumidity measurementwith the control unit
Measurement with: combination humidity probe, e.g.Part no. 0636.9740
1 Connect combination humidity probe to the probe inputof the control unit.
2 Switch on the control unit.
3 Current readings of relative humidity, temperature anddew point are displayed.
Changing units:
- Press key- In Main menu - Device - Units – Change unit
A detailed description is contained in chapter Main menu,Device for control unit - Units.
10 sec
Initialisation
Zeroing
Measurement menu
31
Example measurementVelocity measurement
with the control unit
Special notes on the thermal velocity probe withdirection probe 0635.1041
Fully push down the cap of the probe.
During the measurement, observe the markings of theprobe head and on the telescope. The markings must forma line.
For velocity measurements with a knowndirection of flow, the arrow mark on the probehead must point in the direction of flow.
For velocity measurements with an unknowndirection of flow, turn the probe 360° around itsaxis. Observe the reading. The maximum valueindicates the direction of flow.
NoteMeasurement against the direction of the arrow falsifies thereadings by up to 40%.
32
Example measurementVelocity measurement (hot wire or vane)with the control unit
Volume flow measurement in a duct with 500 mmdiameter with velocity probe.
1 Connect the velocity probe to the probe input of thecontrol unit.
2 Switch on control unit.
3 Activate the function menu with the key and one ofthe function keys. Select the function key with and confirm with .
4 Enter duct diameter:- Press key.- From Main menu – Input – Parameter – Cross section -Circle – enter value
5 Accept the value with function key .
6 Return to the measurement menu with .
7 Start measurement with . VOL
EndZeroing
10 sec
Initialisation
Measurement menu
Measurement menu
Measurement menu
Measurement menu
VOL
1 sec
20 sec
Select “Vol”
Main menu
Select “Input”
Select “Parameter”
Function menu
Select “Cross section”
Select “Circle”
Value is automaticallysaved
Enter value
End
Vol
33
Example measurement
Velocity measurement (hot wire or vane)with the control unit
Mean value
1 Connect the velocity probe to the probe input of thecontrol unit.
2 Switch on control unit.
3 Activate the function menu with the key and one ofthe function keys. Select the function key with
and confirm with .
4 Enter averaging:- Press function key . - Select menu and enter value.
5 Accept value with function key and return to themeasurement menu.
6 Start averaging with .
7 End averaging with .
8. Return to the measurement menu with .ESC
End
Start
End
Mean
Zeroing
10 sec
Initialisation
Measurement menu
Measurement menu
Measurement menu
1 sec
20 sec
Select “Mean”
Select “Timed/Multi-point”
Start averaging
Calculating meanvalue
Mean calculationresult
Function menu
Enter value
EndEnd
ESCStart22..2255
Mean
34
Example measurementVelocity measurement (volume velocity measurement with funnel)with the control unit
Volume flow measurement at an inducing opening
A volume flow rate funnel is required for volume flow ratemeasurement at an inducing opening (grid or cap withannular gap).
The opening of the funnel must completely cover the grid(max. 200 x 200 mm with 0554.0400 or max. 350 x350 mm with 0554.0410).
1 Connect velocity probe, preferably 0635.1041, to theprobe input of the control unit.
2 Switch on control unit.
3 Activate the function menu with the key and one of
the function keys. Select function key with and confirm with .
4 Enter duct diameter:- Press key- From Main menu - Input - Parameter - Cross section -
Circle - enter value 8.82 cm.
5 Accept the value with function key
6 Return to the measurement menu with .
7 Start measurement with . VOL
End
VOL
Zeroing
10 sec
Initialisation
Measurement menu
Measurement menu
Measurement menu
Measurement menu
1 sec
20 sec
Select “Vol”
Main menu
Select “Input”
Select “Parameter”
Function menu
Select “Cross section”
Select “Circle” Value is automaticallysaved
Enter value
End
Vol
35
Example measurementVelocity measurement (pitot tube)
with the control unit
Velocity measurement with pitot tubeMethod 1 (limited): Connect the pitot tube directly to the pressure inlet of thecontrol unit. Limitation: Only the pressure value is shownduring the measurement. The values for velocity speed andvolume flow are shown when the measurement is complete.Method 2: Use the pitot tube preferably with the pressure probe orderno. 0638.1445 due to the optimum accuracy. Measuringrange up to approx. 40 m/s. An additional channel is alsodisplayed (hPa, m/s, m3/h). The velocity is calculated automatically by the device fromthe pressure difference at the pitot tube.
NoteThe pitot tube factor of the standard Testo pitot tubes is1.00 and need not be changed. Enter the factor 0.67 forthe straight testo pitot tubes. For other pitot tubes, ask thesupplier for the pitot tube factor and store. 1 Connect the pressure probe to the probe input or the
pitot tube to the pressure outlet of the control unit.2 Switch on control unit.3 Enter the pitot tube factor:
- Press key- From Main menu – Input – Parameter – Pitot factor –
Set value5 Accept the value with function key .6 Return to the measurement menu with .
End
Select “Parameter”
Main menu
Select “Input”
Select “Pitot factor”
Measurement menu
Value is automaticallysaved
Enter value
End
10 sec
5 sec
Initialisation
Zeroing
Measurement menu
36
Example measurementCO2 measurement with the control unit
The probe 0632.1240 measures concentrations of 0...9999ppm.1 Connect probe, switch on control unit.
2 Warming up phase of the CO2 sensor
3 Display of a reading
4 Enter absolute pressure:- Press key.- From Main menu – Input – Parameter – Pressure –Absolute pressure – Set value
5 Accept the value with function key .
6 Return to measurement menu with .
7 Display of the current reading.
End
The measured CO2 value depends on the absoluteatmospheric pressure. This effect is compensated by thedevice internally. For this, enter the absolute pressure of
the location in the main menu. The displayed CO2 valueis automatically compensated to the entered absolutepressure.
To avoid the effects of the CO2 content of respired air, holdthe probe as far away from your body as possible.
The measuring principle is based on infrared absorption.This type of probe has a relatively high powerconsumption. For long-term measurements, use themains adapter and rechargeable batteries.
If sudden changes occur in the concentration, the probewill require 30 – 60 sec. to adapt to the environment.Gently waving the probe reduces the adaptation time.
10 sec
30 sec
Initialisation
Measurement menu
Measurement menu
EEnndd
Warming up phase
Main menu
Select “Parameter”
Select “Pres.”
Select “Absolute”Enter value
Select “Input”
Measurement with control unit + CO2 probe
37
Example measurementCurrent/voltage measurement
with the control unit
Measurement with control unit + cable 0554.0007
Currents and voltages are measured with the probe0554.0007. These signals can be assigned to a different physical unit.
Example: 0...20 mA are to represent 0...100 %RH whendisplayed later.
1 Connect current and voltage cable, switch on controlunit.
2 Enter the physical unit:- Press key.- From Main menu - Probe - Scaling- Probe socket 1 -Menu min. input, max. input... Set values
- Min. input - - enter 0 - - Max. input - - enter 20 - - Min. output - - enter 0 - - Max. output - - enter 100 -- Enter unit- %RH - - Enter decimal places - .
3 Scaling info
4 Return to the measurement menu with .
5 Display of the current reading.
Main menu
Select “Probe”
Select “Scaling”Select “min input” Set value Display info
Set further values(see above)
Select “Probe” socket 1
Measurement menu
10 sec
Initialisation
Measurement menu
EEnndd
EEnndd
OOKK
EEnndd
EEnndd
EEnndd
EEnndd
38
PrintingPrinting the measurement results
Printout of the currently displayed readings:
1 Activate the function menu with the key andfunction key.
2 Configure function key with .
3 With key back to the measurement menu.
4 Press function key .
5 Print.
If no location is active, the printout is made with“NONAME”.
testo 300 XXLSN: 000321/D
Smith Ltd.Manchester
Location: furnace 5
15:01:00 08:20:15
Fuel: Natural gas
10.9 % CO20.45 % O22348 ppm CO
320 ppm NO15 ppm NO2
1050 °C Flue gas temp.32.5 °C Comb. temp.15.2 m/s Velocity
1.2 % Flue gas loss
Batch 25/2
Date/time
Memory identifier
Fuel
Readings
1 sec
Measurement menu
Function menu
Select print
Measurement menu
39
Main menu MemoryOverview
The memory organisation of the testo 300 XXL is designedsuch that a so-called “Location” must exist to identify asaved reading. This location can consist of up to 20alphanumeric characters.
For example, a location may be:- the name of the customer- a customer number or- the designation of a location.
Several measurements can be saved for each location. Thisis possible by pressing the function key “Mem.” severaltimes.
It is also possible to place locations in directories(see Measurement data processing).
The maximum possible number of measurements dependson whether each individual measurement is saved with itsown location or all measurements are saved exclusively withone location.
40
Main menu MemorySaving without selecting a location
Measurement menu
Function menu
Select “Mem.”
Measurement menu
Save under“NONAME”
Depending on the selected devices, the readings aresaved in the activated component (control unit or
analyser box).
If no location has been chosen after the device wasswitched on, data can still be saved. The location“NONAME” is used for this. This location is the defaultsetting.
1 sec
Save currently displayed readings
1 Activate the function menu with the key andfunction key.
2 Configure function key with .
3 With key to get back to the measurement menu.
4 Press function key .
5 Save.
If no location is active, the data is saved under “NONAME”.
MMeemm..
MMeemm..
MMeemmoorryy
1. Change to menu “Location” with key and .
2. Press key in window “Location”.
3. With keys , select “New loc”and confirm with key .
4. Enter the location.
5. With function key , the location is accepted in the“Location” menu.
End
Change
41
Main menu MemoryCreating a new location
Measurement menu
Location
“New loc” selected Name location
Displays new location
Measurement menu
Create new location
change
OK
Confirm selec-ted characterwith “OK” key.
End
42
Main menu MemorySelecting saved location
1. Change to menu “Location” with key and .
2. Select desired location with keys .
3. Accept with key . Returns to the measuring menu.
4. With the function key “Mem.”, the current readings aresaved under the selected location.
This menu item is accessible during the zeroing phase of thetesto 300 XXL.
Measurement menu
Measurement menu
Location
saved locationselected
43
Main menu MemoryUsing an existing location as format for a new location
Location selected
Select “copy”
Location
1. Change to menu “Location” with key and .
2. Select the location to serve as the source for the copy.
3. Press .
4. Activate the “copy” function. Press . The cursor isbehind the last character of the location.
5. Press function key “ ” to the desired character. The lastcharacter is deleted each time.
6. With keys , , , , select the desiredcharacter (text input) and accept with key .
7. When the location is complete, press function key, returning to the location list. The newly created
location is displayed. Branch to the measurement manu with key .
End
Change
Measurement menu
change
“copy” is activated
Renaming a location
Location renamed
Displays new location
Measurement menu
End
Confirmselectedcharacterwith “OK”
key
44
Main menu MemoryChanging existing location
Location selected
Select “Change”
Location
1. Change to menu “Location” with key and .
2. Select the location to be changed.
3. Press .
4. Activate the “Change” function. Press function key “OK”.
5. Press function key “ ” to the desired character. The lastcharacter is deleted each time.
6. With keys , , , , select the desiredcharacter (text input) and accept with key .
7. When the location is complete, press function key ,returning to the location list. The renamed location isdisplayed. The changed location is deleted.Branch to the measurement menu with key .
End
Change
Measurement menu
change
“Change” activated
Changes the location name
Location renamed
Displays new location
Measurement menu
End
Confirmselected
letter with“OK”
key
45
Main menu MemoryDeleting existing location
1. Change to menu “Location” with key and .
2. Select the location to be deleted.
3. Press .
4. Activate “Delete” function. Press function key .
5. Location is deleted.
6. With function key or key back to the mea-surement menu.
ESC
OK
change
Measurement menu
Select “Delete”
Entry deleted
“Delete” activated
Location selected
Location
Measurement menu
change
Note
s OOKK
46
Main menu MemoryBarcode pen overview
The barcode pen automates the allocation of measurementdata to a location.
If a barcode label is attached on site, this can be read withthe barcode pen.
The barcode pen is connected to the RS-232 socket on thebottom of the device and can be connected anddisconnected whilst the device is switched on.
Activate the barcode pen by holding the tip of the pen brieflyagainst a white surface, e.g. the barcode label. Then movethe pen evenly and quickly in direct contact across thebarcode pattern. A correct reading is indicated by a signaltone.
The location is automatically accepted in the location list ofthe testo 300 XXL and is selectzed as the active location. Itis unnecessary to read or manually enter the customernumber of the customer where you are measuringbeforehand with a PC or by hand into the testo 300 XXL.When you visit a customer whose customer number is notsaved in the device, you can read the number directly onsite by the barcode into the testo 300 XXL. If the locationread by the barcode pen already exists in the testo 300XXL, this is activated.
Note
Barcodes can be created and decoded with the evaluationsoftware.
47
Main menu MemoryReading a location with the barcode pen (during the zeroing phase)
You can also enter the location after the expiry of the zeroingphase.
Reading a location from the measurement menu
Reading several locations with the barcode pen frommeasurement menu.Procedure as described above (simply repeat theprocedure). The location last read is active.
10 sec
Initialisation Device zeroing
Measurement menu
Measurement menu
Measurement menu
48
Main menu Memory
Displaying saved readings
1.Change to menu “Location” with key and .
2.Select location and confirm with key .
3.Press key in the measurement menu.
4.Select “Memory” menu and confirm with key .
5.Activate “Read out” menu with key .
6.Display of the saved readings.
Main menu
Memory menu
Memory menu
Measurement menu
Displays savedreadings
Displays savedreadings
Measurement menu
Location selected
Location
2 x
49
Main menu MemoryMean calculation
Mean value menu
Mean value menuActivated
Specify length of time
Main menu
Measurement menu
Start MeanCalculation
1.Press in the measurement menu.
2.Select “Memory” menu and confirm via .
3.Select mean value menu and press .
4.Enter duration of mean calculation. Confirm reading byclicking on .
5.Enter number of mean calculations. Confirm reading byclicking on .
6. Information on input values. Mean calculation can bestarted by clicking on .
Note:Mean calculation stops automatically once the specifiedlength of time has elapsed.
EEnndd
EEnndd
EEnndd
Enter number
EEnndd
OOKK
OOKK
Information Automatic Stop Mean calculation
SS tt aarr tt
50
Main memory MemoryPrinting mean calculation result
Main menu
Read out menu
Displaying and selectingsaved readings
1. Start mean calculation
2. To measurement menu after automatic stop.
3.Change to “Main menu” menu by clicking on .
4.Activate “Memory” menu.
3.Select “Read out” sub-menu.
4. Confirm by clicking on .
5.Displays saved readings. Select value which is to beprinted and start printing by clicking on
.
6.Return to measurement menu by clicking on .
Start Mean calculation
Measurement menu
Automatic stop
Display of selectedvalues
Prints out data
Return tomeasurement menu
PPrr ii nn tt
PPrr ii nn tt
Example of printout
51
Main menu MemoryDeleting entire memory
Main menu
Measurement menu
Measurement menu
Memory menu
Select “Deletememory”
Select yes
1. Change to menu “Main menu” with key .
2. Activate the “Memory” menu.
3. Select submenu “Delete memory”.
4. Delete memory “Yes/No”
5. Confirm with key .
6. Return to measurement menu via
52
Main menu Memory
Printing saved readings of a location
Measurement menu
Measurement menu
Location selected
Select “Print location”
Print locationLocation
cchhaannggee Location
1. Change to menu “Location” with key and .
2. Select location and activate function key .
3. Select menu “Print location”.
4. Start printing with key .
5. Then return to the measurement menu with key .
CChhaannggee
53
Main menu Memoryfree memory?
Measurement menu
Measurement menu
Main menu
Select “Memory”
Select “free memory?”
Displays free memory
1. Change to the main menu with key .
2. Activate the “Memory” menu with .
3. Select “free memory?”
4. Display free memory with key .
5. Return to the measurement menu with key .
The menu provides information on the available memorycapacity to save readings.
54
Scaling
Select “Probe”
Select “Scaling”
Select “min input”
Set value
Display info
Set further values(see above)
Select “Probesocket 1”
Measurement menu
Measurement menu
Currents and voltages are measured with the probe0554.0007. These signals can be assigned to a different physical unit.
Example: 0...20 mA are to represent 0...100 %RH whendisplayed later.
1 Connect current and voltage cable, switch on controlunit.
2 Enter the physical unit:- Press key.- From Main menu - Probe - Scaling- Probe socket 1 -Menu “min input.”, “max input”... “Set values”
- Min input - - enter 0 -- Max input - - enter 20 - - Min output - - enter 0 -- Max output - - enter 100 -- Enter unit
- %RH - - Enter decimal places -
3 Scaling info
4 Return to the measurement menu with .
EEnndd
EEnndd
OOKK
EEnndd
EEnndd
EEnndd
EEnndd
Main menu Probefor control unit
55
Main menu Probefor control unit
Info
Information on the entered values for calibration andscaling.
Resetting
Entered scaling values are reset to the factory settings.
Select “Probe”
Select “Reset”
Probe socket 1
Reset
Measurement menu
Measurement menu
1. Change to the main menu with key .
2. Activate “Probe” menu with .
3. Select “Reset”.
4. Display the probe socket with key . .
5. Press key to reset.
5. Return to the measurement menu with key .
56
Current readingcomes into effect
or
Menu: Set humidity value
Parameters
NoteThe temperature, relative humidity and absolute pressureinfluence the air density at the measuring point. When thesevalues have been entered, the density is calculatedautomatically. The result is composed as follows:
TemperatureHumidity Density
Metres above sea levelBarometric pressure Absolute pressureDifferential pressure
TEMPERATURE
Enter the temperature as a parameter:
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select the“Temp.” menu.
4 Accept the current reading with key or select“Input temperature” with the cursor keys.
5 With function key , an automatic branch is madeto the menu “Input humidity” or with key back to themeasurement menu.
Main menu Inputfor control unit
Measurement menu
Select “Input”
Select “Parameter”
Select “Temp.”
Enter temperature
EEnndd
EEnndd
57
HUMIDITY
Enter the humidity as a parameter:
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select Humiditymenu.
4 Accept the current reading with key or select“Input humidity” with the cursor keys.
5 With function key , an automatic branch is madeto the menu “Input barometric pressure” or with key ,back to the measurement menu.
Main menu Inputfor control unit
Select “Humid.”
EEnndd
Current readingcomes into effect
Menu: Setpressure value
Enter humidity
EEnndd
Measurement menu
Select “Input”
Activate “Parameter”
or
58
PRESSURE
NoteThe absolute pressure is determined in the measuringprocedure depicted to the side. The absolute pressureresults from• Altitude pressure (metres above sea level)
This has an annual average at sea level of 1013 mbar. Thehigher the location is above sea level, the lower thepressure.
• Barometric pressure This has an annual average independent of the altitude1013 mbar. This pressure can fluctuate from the annualaverage by approx. ±20 mbar depending on the weather(see local barometer display).
• Differential pressure This is the higher or lower pressure in the duct.
The entry of the absolute pressure (in hPa only; it isimpossible to switch to other units) has an effect on thepressure-dependent measurement values. These areautomatically compensated by the pressure: for humidity(g/kg, J/g), CO2 and in all thermal probes!
Enter the pressure as a parameter:
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select Absolutepressure menu.
4 Accept the current reading with key or select“Input absolute pressure” with the cursor keys.
5 With function key , an automatic branch is madeto the menu “Input density” or with key , back to themeasurement menu.
Main menu Input
for control unit
Select “Pres.”
Enter the absolutepressure
Select “Absolute”
Menu: Setdensity value
EEnndd
Current readingcomes into effect
or
EEnndd
Measurement menu
Select “Input”
Activate “Parameter”
59
DENSITY
NoteFactory setting: 1293 g/m3
If this value is not changed, this value is used for thecalculation. The individual values are not taken into account.
The temperature, relative humidity and absolute pressureinfluence the air density at the measuring point. Thesevalues can be entered.
When these values have been entered, the density iscalculated automatically.
Enter density as a parameter:
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select the“Density” menu.
4 Accept the current reading with key or select“Input density” with the cursor keys.
5 With function key , an automatic branch is made
to the menu “Input pitot tube factor” or with key ,back to the measurement menu.
EEnndd
Main menu Inputfor control unit
Select “Density”
Enter the density
Menu: “Set pitottube factor”
Measurement menu
Select “Input”
Activate “Parameter”
Current readingcomes into effect
EEnndd
or
60
PITOT TUBE FACTOR
NoteThe pitot tube factor of the standard Testo pitot tubes is1.00 and need not be changed. Enter the factor 0.67 for thestraight testo pitot tubes. For other pitot tubes, ask thesupplier for the pitot tube factor and store.
Enter the pitot tube factor as a parameter:
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select “Pitotfactor” menu.
4 Accept the current reading or select “Set pitot tubefactor” with the cursor keys.
5 With function key , an automatic branch is madeto the menu “Cross section - Circle” or with key ,back to the measurement menu.
Main menu Input
for control unit
EEnndd
Measurement menu
Select “Input”
Activate “Parameter”
Select “Pitot factor”
Set the pitot tubefactor
Menu: “Crosssection”
EEnndd
61
Main menu Inputfor control unit
CROSS SECTION
Enter the cross section as a parameter:
1 Connect probe, switch on control unit.
2 Activate the function menu with the key and one ofthe function keys. Select the function key with
and confirm with .
3 Enter duct diameter:- Press key.- From Main menu – Input – Parameter – Cross section
- Circle – input value (8.82 cm)- Square – input value- Rectangle – input value
- Length a – input value- Length b – input value
- Area – Input value
4 Accept the respective entered value with function key.
5 Go automatically to the menu “Offset factor” or pressto go back to measurement menu.
6 Start measurement with . VOL
End
VOL
Measurement menu
Select “Input”
Activate “Parameter”
Select ”Cross section”
Respective setvalue comes into
effect
Enter offset factor
or
EEnndd
Select “Circle”,“Square”,
“Rectangle”, “Area”
Enter value for”Circle”, “Square”,
“Rectangle”, “Area”.
62
Main menu Input
for control unit
Measurement menu
Select “Input”
Activate “Parameter”
Select “Offset factor”
Set the offset factor
Set value comesinto effect
EEnndd
Measurement menu
Select “Input”
Activate “Parameter”
Select “Info”
Parameter info
Measurement menu
INFO
Overview of all parameters
1 Change to the main menu with key .
2 From Main menu – Input – Parameter – select the “Info”menu.
3 Return to the measurement menu with key .
OFFSET FACTOR
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select “Offsetfactor” menu.
4 Accept the current value with key or select “Setoffset factor” with the cursor keys.
5 Go automatically to the menu “Offset factor” or pressto go back to measurement menu.
EEnndd
Offset factor calculation:
v = k s
k = Offset factor s = Pitot tube factor p = Pressure (Pa) rho = Density (kg/m3)
2prho
Parameter info
63
Alarm limits
Note:The alarm for CO is factory adjusted and activated to 30ppm and for CO2 to 5000 ppm. The alarm limits for dP areset to zero and deactivated. The acoustic alarm is switchedon.
If alarm limits are set and activated, a message appears onthe display if these are exceeded during measurements. If the acoustic alarm is switched on, a signal tone indicatesthat values have been exceeded.
NotedP measurementsWhen the timer is switched on, the alarm limit automaticallyactivates.
Enter alarm limits:
1 Switch on the control unit.
2 Change to the main menu with key .
3 From Main menu - Input – Alarm limits: - CO limit - enter value- Enter CO2 value- dP limit 1 - enter value- dP limit 2 - enter value- Switch acoustic alarm on or off
4 Enter characters with the cursor keys.
5 Accept the respectively entered value with function key.
6 Return to the measurement menu with key .
EEnndd
Timer
In addition to the alarm limit, a timer can be set for the dPmeasurement (primary and main test to TRGI). When thetimer is switched on, the alarm limit for dP is automaticallyactivated.
Main menu Inputfor control unit
Activate “Alarmlimits”
Select “Audible alarm”
Set “CO”, “CO2limit”, “dP limit 1”,
“dP limit 2”
Switch on theacoustic alarm
Accept the setvalue
EEnndd
OOKK
Accept the setting
Measurement menu
64
Main menu Devicefor control unit
Measurement menu
Main menu
Select “Changedate”
Set the date
Current date
Select number
Confirm selection,automatic branch tonext item
EEnndd
cchhaannggee
Measurement menu
2 x
EEnndd
cchhaannggee
Changing the date
SETTING THE DATE
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Device – Change date – Set date/timemenu
4 Activate menu “Set date” with key .
5 With the cursor keys select numbersin the nuumeric block and accept with key .
6 The date box automatically jumps to the next number.
NoteWith the function keys or with the icons left and right of the numeric block, it is possible tobypass numbers.
7 Terminate input with function key .
Date box
Numeric block
65
SETTING THE TIME
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Device – Change date – SetDate/time menu.
4 With key , select menu “Change date” and activatewith key .
5 With the cursor keys select numbersin the numeric block and accept with key .
6 The date box automatically jumps to the next number.
NoteWith the function keys or with the icons leftand right of the numeric block, it is possible to bypassnumbers.
7 Terminate input with function key .
Main menu Devicefor control unit
EEnndd
cchhaannggee
Set the time
Current time
Select number
Time box
Numeric block
Confirm selection,automatic branch tonext item
EEnndd
CChhaannggee
Measurement menu
2 x
Measurement menu
Main menu
Select “Changedate”
66
Main menu Devicefor control unit
Measurement menu
Measurement menu
Main menu
Select “Device”
Select “Auto off”
Set time
The entry isaccepted
automatically.
Auto off
The period after which the control unit switches off when notin use is specified in the menu “Auto off”.
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Device – Auto off – Set time.
4 This is accepted automatically with key . .
5 Return to the measurement menu with key .
67
Main menu Devicefor control unit
Measurementmenu
Measurement menu
Main menu
Select “Device”
Select “Printer”
Select “Contrast” Set the contrast
Printer
SETTING THE CONTRAST
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Device - Printer - Contrast-Set contrast
4 This is accepted automatically with key .
5 Return to the measurement menu with key .
TTeess tt
EEnndd
EEnndd
Measurementmenu
Measurement menu
Main menu
Select “Device”
Select “Printer”
Select “Print text” Select “Line”
Label the line
ENTERING PRINT TEXT
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu - Device - Printer – Print text – Selectlines 1 - 3 or footnote – enter text (e.g. company name,address)
4 The entered line is accepted automatically with key .
5 Return to the measurement menu with key .
EEnndd
EEnndd
Test printout
Light
Note
“On/Off” Pressing key switches the lighting on,pressing again back off.
“Automatic” Pressing key switches the lighting on,after approx. 3 min. the lighting is automaticallyswitched off.
1 Switch on control unit.
2 Change to main menu with key .
3 From Main menu - Device - Light – select “Light type”.
4 The selection is accepted automatically with key .
5 Return to the measurement menu with key .
68
Diagnostic
Displays all unrectified error messages.
Main menu Devicefor control unit
Measurement menu
Main menu
Select “Device”
Select “Light”
Select “On/Off/automatic”
OOKK
OOKK
Measurement menu
69
Units
Example: Set °C
1 Switch on the control unit.
2 Change to the main menu with key .
3 From Main menu – Device - Units - Temperature - enter °C.
4 Accept the entered value with function key .
5 Return to the measurement menu with key .
Available units:
Temperature: °C, °F
Humidity: Off, td°C, g/m3, G/kg, J/g
Velocity: m/s, fpm
Volume flow: m3/h, cfm, m3/m, m3/s, l/s, m3/h
Pressure: hPa, inW, mbar, Pa, bar, psi
Gas: ppm, %
Main menu Devicefor control unit
Measurement menu
Main menu
Select “Device”
Select “Units”
Select theparameter
Specify unit
OOKK
OOKK
Measurement menu
70
Operating values
Resetting factory settings
Main menu Servicefor control unit
Display operating values
Display device data
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu - Service – select “Op. values”.
4 Return to the measurement menu with key .
Reset to initial settings.
Address
Input via PC software only.
Measurement menu
Main menu
Select “Service”
Select “Op. values”
Measurement menu
Device data
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Service – select “Device data”.
4 Return to the measurement menu with key .
Control Unit
Control Unit
71
Main menu Service
Main menu Periphery
for control unit
Set “Language”
Measurement menu
Main menu
Select “Service”
Select “Language”
Calibrate thetouchscreen
Measurement menu
Main menu
Select “Periphery”
Activate “Touchscreen”
Measurement menu
Language
1 Switch on control unit.
2 Change to the main menu with key .
3 From the main menu - Service - Lang. - Set language.
4 Return to the measurement menu with key .
(with touchscreen only)
1 Switch on control unit.
2 Change to the main menu with key .
3 From Main menu - Periphery - Touchscreen – Calibratetouchscreen.
4 Return to the measurement menu with key .
Measurement menuCCaall..
The use of a probe preliminary filter is obligatory formeasuring solid fuels:For 335mm probe: 0440 7435For 700mm probe: 0440 7436
1 Connect analyser box and control unit; switch on controlunit.
2 Go to main menu via .3 Click on Main menu - Memory - Solid fuel measurement
and confirm with .4 Specify “Length of time” using the arrow buttons. Confirm
value via and proceed to “CO range” menu.
5 Set required measuring range and confirm via button.
6 Information on values entered.7 Start measurement by clicking on .
Mean calculation stops once the specified length of time haselapsed.Note:If you wish to use the default values (duration: 15min.,measuring range: CO High), you can also select “Info”directly in the “Solid fuel measurement” menu and start themeasurement.
SS tt aarr tt
EEnndd
72
Solid fuel measurement optionfor analyser box
Measurement menu
Main menu
Info
Selecting solid fuelmeasurement
Measurement menu
Solid fuel measurementis activated
Specify length of time
Measurement stopsautomatically
Start measurement
Fresh air
Measurement
2 min.
Set CO range
EEnndd
SS tt aarr tt
73
Solid fuel measurement optionfor analyser box
Notes:
The given measuring ranges are recommendations toachieve the highest possible accuracy. Higher deviation is tobe expected outside these measuring ranges.
Note: The toxic sensors may be damaged if COconcentrations are greater than 40 000 ppm. Remove thetoxic sensors during applications of this kind in order toprevent the sensors from being damaged. Measurementscan be carried out in this range using the O2 sensor.
CO`s lower explosion limit is reached if CO concentrations ofmore than 10.8 % (108 000 ppm) come into contact with air.
The lower explosion limit is significantly lower in combusionprocesses such as wood combustion due to the nonburnedhydrocarbons and hydrogen.
CO measuring ranges with CO Low or CO High
CO meas. CO concentration Measuring range CO Accuracyrange in sensor (ppm) (ppm)
from........to from........to
Low 1 0 1000 00 1000 See Technical data
High 2 50 1500 1000 30000 ± 7.5 % of read.
1 corresponds to: undiluted2 corresponds to: 20xdilution
74
Main menu Sensorsfor analyser box
Sensors
RECALIBRATION
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu - Sensors – to menu “Recalibration”
4 With key , select the parameter and activatewith key .
5 With the cursor keys select numbersin the numeric block and confirm with key .
6 Accept entered value with .
7 The newly entered CO or NO value is displayed.
8 Return to the measurement menu with key .
Measurement menu
Select “Sensors”
Select “Recal.”
Enter parametervalue
Recalibration resultSelect number
Confirm selection
SS tt aarr tt
SS tt aarr tt
Measurement menu
Select parameter
75
Main menu Sensorsfor analyser box
Measurement menu
Measurement menu
NO2 ADDITION
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu – Sensors – to menu “NO2 addition”
4 With the cursor keys select numbersin the numeric block and confirm with key .
5 Accept entered value with .
6 Return to the measurement menu with key .
Measurement menu
Select “Sensors”
Select “NO2addition”
Enter NO2 additionSelect number
Confirm selection
EEnndd
PRINT CALIBRATION DATA
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu – Sensors – to menu “Print cal. data”
4 Start print with key .
5 Return to the measurement menu with key .
NoteAusdruck abbrechen mit Taste .
EEnndd
Measurement menu
Select “Sensors”
Select ”Print cal.data
76
Main menu Inputfor analyser box
Spot number / Oil derivative / Heat carrier temperature
Note:Enter spot number and oil drivatives only for oil fuels andpropane.
1 Connect analyse box to control unit; switch on controlunit.
2 Change to main menu with key .
3 From Main menu - Input – Smoke number/WTT (HCT)- Smoke number 1 – enter value- Smoke number 2 – enter value- Smoke number 3 – enter value- Heat carrier temp. - enter value
With the cursor keys select numbersin the numeric block and confirm with key .
4 Accept the respectively entered value with function key.
5 When the heat carrier temperature has been entered andconfirmed, an automatic branch is made to the “Oilderivative” menu.
6 Select “Oil derivative?”, “No / Yes”. Branch to the “Info”menu with functional key .
7 Return to the measurement menu with key .
Select the menu
Enter “Smoke number 1” “Smoke number 2” “Smoke number 3” “Heat carrier temp.”
Oil derivative no/yes
Set comes intoeffect
EEnndd
Accept the settings
OOKK
Measurement menu
Select “Input”
Select “Spot (smoke) number/ WTT”
Select number
Confirm selection
EEnndd
Display info
Measurement menu
77
EXAMPLE:Set “Light oil”
1 Connect analyser box to control unit; switch on controlunit.
2 During the zeroing phase, the menu“Select fuel” is automatically activated.
3 Select “Light oil” with keys .
4 With key return to the zeroing phase or themeasurement menu.
Fuel
O2 reference, CO2max and coefficient
Main menu Inputfor analyser box
Measurement menu
Zeroing phase
Zeroing phase
or
“Select Light oil”
3 sec
Available fuels and their factors
* These factory adjusted values can be freely selected.
A2, B fuel-specific factors FBr conversion factor mg/m3 in g/GJ
The O2 reference index, CO2 max. and all factors for freelydefined fuels.
The coefficients A2, B are only adjustable for the freelydefinable fuels “Fuel 1 and 2”. “Coeff” appears in the menuwhen “Fuel 1 or 2” is selected.
Fuel A2 B f CO2max* O2 ref.* FBr
Fuel oil, light 0.68 0.007 0.68 15.4* 3* 0.2637Fuel oil, heavy 0.61 - 0.61 15.9* 3* 0.2616Natural gas 0.66 0.009 0.66 11.9* 3* 0.2665Propane 0.63 0.008 0.63 13.7* 3* 0.2613Anthracite 0,74 - 0.74 20.5* 8* 0.2640Coal briquettes 0,75 - 0.75 18.9* 8* 0.2748Wood fuels, coke 0.74 - 0.74 20.3* 13* 0.1885Brown coal, peat 0.90 - 0.90 19.8* 8* 0.2647Coking gas 0.60 0.011 0.60 10.3* 3* 0.2461Urban gas 0.63 0.011 0.63 13.6* 3* 0.2418Fuel 1 0* 0* 0* 0* 0*Fuel 2 0* 0* 0* 0* 0*
78
Main menu Inputfor analyser box
Select “Light oil”
Select“O2ref./CO2max”
Select “O2ref.”
Enter “O2ref.” forLight oil
Enter CO2max for Light oil
Zeroing phase
3 sec
Measurement menu
EEnnddEEnndd
EXAMPLE
“Light oil” is selected.
1 Connect analyser box to control unit; switch on controlunit.
2 During the zeroing phase, the menu “Select fuel” isautomatically activated.
3 Activate “Light oil” with keys .
4 With key , change to menu “Select fuel”and with key , select menu “O2ref/CO2max”.
5 With , select “Input O2ref.” and confirm with .
6 With the cursor keys select numbersin the numeric block.
7 With , accept value and change to menu“CO2max”.
8 With the cursor keys select numbersin the numeric block.
9 With and , back to the measurementmenu.
EEnndd
EEnndd
79
Main menu Inputfor analyser box
Zeroing phase
Select “Input”
3 sec
2 x
Parameters
PRESSURE
NoteThe absolute pressure is determined in the measuringprocedure depicted to the side. The absolute pressureresults from• Altitude pressure (metres above sea level)
This has an annual average at sea level of 1013 mbar. Thehigher the location is above sea level, the lower thepressure.
• Barometric pressure This has an annual average independent of the altitude1013 mbar. This pressure can fluctuate from the annualaverage by approx. ±20 mbar depending on the weather(see local barometer display).
• Differential pressure This is the higher or lower pressure in the duct.
The entry of the absolute pressure (in hPa only; it isimpossible to switch to other units) has an effect on thepressure-dependent measurement values. These areautomatically compensated by the pressure: for humidity(g/kg, J/g), CO2 and in all thermal probes!
Enter the pressure as a parameter:
1 Connect the probe. Connect analyser box to control unit;switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select Abs.pressure menu.
4 Accept the current reading with key or enter“Absolute pressure” with the cursor keys.
5 With function key , an automatic branch is madeto the menu “Set pitot tube factor” or with key backto the measurement menu.
Enter the absolute pressure
Select “Absolute”
or
EEnndd
Current readingcomes into effect
Set menu: ”Pitottube factor”
EEnndd
Select “Parameter”
Select “Pres.”
80
PITOT TUBE FACTOR
NoteThe pitot tube factor of the standard Testo pitot tubesequals 1.00 and need not be changed. Enter the factor 0.67for the straight testo pitot tubes. For other pitot tubes, askthe supplier for the pitot tube factor and store.
Enter the pitot tube factor as a parameter:
1 Connect the probe. Connect analyser box to control unit;switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select “Pitotfactor” menu.
4 Accept the current reading with key or select“Set pitot tube factor” with the cursor keys.
5 With function key , an automatic branch is madeto the menu “Circle - Cross section” or with key ,back to the measurement menu.
EEnndd
EEnndd
Select “Pitot factor”
Enter the pitot tubefactor
Main menu Inputfor analyser box
Zeroing phase
Select “Input”
3 sec
2 x
Select “Parameter”
Menu “Set cross-section”
EEnndd
81
Main menu Inputfor analyser box
Zeroing phase
Select input
3 sec
2 x
Select “Parameter”
CROSS SECTION
Enter the cross section as a parameter:
1 Connect the probe. Connect analyser box to control unit;switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select “Crosssection” menu.
4 Enter duct diameter:- Press key.
- Circle – input value - Square – input value- Rectangle – input value
- Length a – input value- Length b – input value
- Area – input value
5 With function key , accept the respectivelyentered value and automatically branch to menu “Setoffset factor” or back to the measurement menu with key .
Select “Crosssection”
Accept the respectivelyset value and branch tomenu “Set offset factor”
EEnnddSelect “Circle”,
“Square”,“Rectangle”, “Area”
Enter value for“Circle”, “Square”,
“Rectangle”,“Area”
End
82
Main menu Inputfor analyser box
Measurement menu
Select input
Activate “Parameter”
Select “Offset factor”
Set “Offset factor”
Accept the setvalue
EEnndd
OFFSET FACTOR
1 Connect probe, switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Parameter – select “Offset.factor” menu.
4 Accept the current value with key or select “Set Offset factor” with the cursor keys.
5 Return to the measurement menu with key .
Offset factor calculation
v = k s
k = Offset factor s = Pitot tube factor p = Pressure (Pa) rho = Density (kg/m3)
2prho
EEnndd
83
Main menu Inputfor analyser box
Info display
Select “Info”
Measurement menu
INFO
Overview of all parameters.
1 Change to the main menu with key .
2 From Main menu – Input – Parameter – select the “Info”menu.
3 Return to the measurement menu with key .
Zeroing phase
Main menu
3 sec
2 x
Select “Parameter”
Dew point/ambient air (AT)
1 Connect humidity probe to control unit. Connect analyserbox to control unit; switch on control unit.
2 Change to the main menu with key .
3 From Main menu – Input – Dew point/ambient air – selectthe “Temperature/ambient air” menu.
4 With the cursor keys , enter thetemperature reading or accept the current reading withfunction key .
5 Function key branches automatically to menu “Humidity/ambient air”.
6 Enter the numbers as described in item 4.
7 Function key branches automatically to menu “Dew point/ambient air”.
8 Automatic display of the dew point in the input box.With function key , the value is accepted.
9 Return to the measurement menu with key .
84
Main menu Inputfor analyser box
Zeroing phase
Select Input
3 sec
2 x
Select “Dew point/ambient air”
Activate “Temperature/ambient air”
Enter AT temperature
EEnndd
EEnndd
EEnndd
Enter AT humidity
Display AT dew point
Measurement menu
Select number
Confirm selection
EEnndd
EEnndd
EEnndd
85
View
The display sequence of the readings is defined in thismenu. The sequence can be freely selected.
Example: The first empty display box is occupied by the reading “Fluegas temperature”.
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu – Device – View – Display seq. menu
4 With the function keys , find an emptyreading space.
5 With the cursor keys , assign thereading space.
6 With , open the window PARAMETER, UNIT, INSERT,DELETE and activate menu “Parameter”.
Note“Insert” menuParameters can be inserted at any desired reading space.The existing values are shifted by one space.
“Delete” menuDeletes selected parameters.
7 Select AT parameter with keys and confirm with.
Available parameters:O2 CO NO H2NOx FT AT dTT1 T2 qA CO2l h uCo SootOILD HC.T O2rf CO2MDraught Accu DT F/hPump DP Fuel
Main menu Devicefor analyser box
Zeroing phase
Main menu
3 sec
2 x
Select “Device”
Activate “View”
Display sequence
8 Automatic branch to menu “Unit”.
NoteIf a parameter is selected, a matching unit is automaticallypresented. This suggestion can be changed.
9 With key , the parameter is accepted.
10 Return to the parameter menu with function key .
EEnndd
86
Main menu Devicefor analyser box
Assign empty readingspace
Select “Parameter”
Select “AT”
Select “Unit”
Accept the newparameter
EEnnddDiagnostic
Displays all unrectified error messages.
Find empty readingspace
Measurement menu
87
Main menu Servicefor analyser box
Operating values
Display operating values
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu - Service – “Op. values” menu.
4 Return to the measurement menu with key .Select “Service”
Select “Op. values”
Measurement menu
Zeroing phase
Main menu
3 sec
2 x
Switching off CO/NO/NO2
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu – Service – Switch-off.
4 With key , select menu “CO”, ”NO” or “NO2”.
5 With the cursor keys select numbersin the numeric block and confirm with key .
6 Accept entered value with .
7 Return to the measurement menu with key .
88
Main menu Servicefor analyser box
Select number
Confirm selection
EEnndd
EEnndd
Zeroing phase
Main menu
3 sec
2 x
Select “Service”
Select “Switch-off”
Activate “CO”, “NO”or ”NO2”
Enter CO, NO orNO2 value
Measurement menu
89
Main menu Servicefor analyser box
Select number
Measurement menuEEnnddConfirm selection
Device data
Display device data
Enter bus address
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu - Service – Device data menu
4 Return to the measurement menu with key .Select “Service”
Select “Device data”
Measurement menu
Zeroing phase
Main menu
3 sec
2 x
Select “Service”
Select “Bus address”
Zeroing phase
Main menu
3 sec
2 x
Bus address
1 Connect analyser box to control unit; switch on controlunit.
2 Change to the main menu with key .
3 From Main menu - Service – Bus Address – Change BusAddress
4 With the cursor keys , select anumber on the numeric block and confirm with key .Possible numeric combinations: 01 - 99.
5 Accept entered value with . Automaticbranch to the main menu.
The control unit and analyser box are factory adjusted todifferent bus address numbers.
If these are equivalent, the bus address number of thecontrol unit or the analyser box must be changed to preventconnection problems.
EEnndd
90
MaintenanceInserting new paper roll in printer
Function menu
Select paper feed (LFPr).
Measurement menu
LLFF PPrr
1 sec
Measurement menu
Paper roll feed
1 Activate the function menu with the key and one ofthe function keys. Select the function key with and confirm with .
2 Open printer cover and remove plastic roller.
3 Insert a suitable paper roll. Observe paper guide: thecoated side of the thermal paper must point towards theprinter head after threading in.
4 Press function key . The paper is drawn in.
5 Close cover.
LF Pr
LF Pr
91
MaintenanceChanging the battery pack
Analyser box
Rear
1 Disconnect mains unit
2 Disengage catch.
3 Remove battery pack and pull plug from socket afterdisengaging.
4 Insert new battery pack (ensure that the plug engages).
5 Put on and close cover.
Control unit
Rear
1 Pull off mains adapter.
2 Disengage catch.
3 Remove battery pack and pull plug from socket.
4 Insert new battery pack (adhere to marking on plug wheninserting)
5 Put on and close cover.
92
Maintenance
Filter change
1. If the filter is visibly contaminated, it must be replacedunder all circumstances.
2. Replace the filter if the pump performance drops(audibly).
In most cases, it is sufficient to replace the flue gas filter.
- To replace a filter, remove the filter cover by twisting to theleft.
- Take out the used filter and insert the new filter. - Screw on the filter cover.- The cross strut of the filter housing must be aligned with
the markings on the box housing.
Spare filters with Part no. 0554.3381 available.
Condensation trap
To be observed- Only empty the condensation trap when the pump is
switched off. - Do not damage the sealing rings when assembling the
condensation trap.- Pull out the condensation trap horizontally to empty the
condensation.- Open the drain plug.
Filter, flue gas
Filter, freshair inlet
Filter, diluting air
Condensationtank
Drain plug
Filter housing Cross strut
Remove condensationtank
93
Maintenance
Flue gas probe
Cleaning the flue gas probeIf the flue gas is heavily laden with dust, it is possible thatsections of the gas path preceding the hose filter willbecome contaminated or blocked.
Cleaning the flue gas probe with gas path closedPull off the probe shaft, place and move about in hot water.Then blow out with air or clean with a round brush (e.g.brass).
Changing a defective thermocouplePull the bend protection spring of the rear outlet from theguide by twisting to the left and remove the left hose. Pullout the thermocouple by the cable.
Remove the bend protection spring and remove the cablefrom the slotted hose.
When inserting a new thermocouple, do not crush thethermocouple cable; push in fully using a small screwdriver.
Remove the thermocouple only if defective.
O2, CO, NO2 and NO measuring cell replacement
- Switch off the device and isolate from the mains.- Open the large cover on the rear of the measuring box.- Remove measuring cell heater.- Pull the hose connections from the used cell.- Remove the measuring cell from the device.- Insert and connect a new measuring cell.
Caution! Remove the auxiliary circuit board before installing
the NO cell.
Caution! The O2 measuring cell requires a compensationperiod of approx. 60 min. after replacement in the
testo 300 XXL. Only then can accurate measure-ments be made. The device must not be switched on
while doing this.
Installation of an NO measuring cell
Installation of CO / NO2 measuring cell
NO measur-ing cell
Measuringcell
Auxiliarycircuit board
Short-circuit bridgeCaution!
Remove the short-circuit bridge wheninstalling a new measuring cell.
Measuring cells
Measuring cell heater
94
Maintenance
95
Disconnect the mains plug prior to maintenance.
MaintenanceMaintenance of the gas pumps
Opening the analyser box
1. Switch off the device and pull the mains plug.. 2. Remove the condensation trap. 3. Unlock the three filter housings. 4. Loosen the 8 Philips screws in the lower part of the
device housing. 5. Turn the device back over and remove the upper part of
the device housing.
Cleaning the main gas pump and the purging pump
1. Remove Philips screw from the plastic pump bracket. 2. Push the plastic bracket gently aside. 3. Pull out the pump carefully. 4. Take the pump tool from the holder in the handle. 5. Push the pump tool into the guides of the pump head. 6. Remove the pump tool from the pump head. 7. Remove the diaphragm holder from the pump head
and remove the diaphragm. 8. Clean the pump diaphragm, pump cup and pump head
with spirit or water. 9. Insert the diaphragm in the diaphragm holder and place
in the pump head.10. Place pump head on the pump. 11. Remove pump tool. 12. Insert pump in the installation block. 13. Replace the pump in the gas measuring block and
fasten the pump with Philips screws to the plasticpump bracket.
14. Put on the upper part of the housing. Ensure that nocables are trapped.
15. Turn the device over and tighten the 8 Philips screws. 16. Install the filter housing and condensation trap.
Screws Screws
Main gas pump Purging pump
Pumptool
Pump(Part. no.0239.0014)
Pump cup
Pump head
Pump tool
Pump tool(art. no.0192.0468)
Diaphragms(Part. no.0193.0072)
Diaphragmholder
96
MaintenanceMaintenance of the gas pumps
Cleaining the main gas pump with the gas preparationoption
1. Remove Philips screw from the plastic pump bracket. 2. Push the plastic bracket gently aside. 3. Pull the gas pump upwards out of the gas measuring
block. 4. Loosen the 4 fastening screws in the pump head of the
main gas pump. 5. Pull off the pump head. 6. Remove the two circlips from the depressions of the
pump head (front and rear). 7. Remove and clean the pump diaphragms (water or
spirit). 8. If necessary, blow through the inlet and outlet pipes
with compressed air. 9. Reattach the diaphragms with the circlips.10. Attach the pump head to the main gas pump and
fasten with the screws. 11. Replace the pump in the gas measuring block and
fasten the pump with Philips screws to the plasticpump bracket.
12. Put on the upper part of the housing. Ensure that nocables are trapped.
13. Turn the device over and tighten the 8 Philips screws. 14. Install the filter housing and condensation trap.
Main gas pump Purging pump
Main gaspump
Fasteningscrews
Pump head
97
MaintenanceChanging the pump cassette of the condensation pump with optional gas
preparation
1. Empty the condensation trap. 2. Remove the cover. 3. Unlock and pull off the pump cassette. 4. Remove the bend protection spring and push onto the
hose of the inlet of the new pump cassette. 5. Push on hoses.
NoteEnsure that the hoses are not trapped or constrained.
6. Push the replacement cassette onto the motor shaftuntil it engages.
7. Attach cover.
unlock
InletOutlet
98
MaintenanceUpgrades
1. Switch off the device. 2. Open the measuring cell cover.
Upgrade modules are connected at one of the emptyslots. The measuring modules NO and HC must onlybe inserted at the slots marked “NO” and “Type A/HC”
– observe switch settings. Connect the measuringmodules NO2, SO2, H2S at any desired slot marked
“Type A” (see label in the measuring cell cover).
3. Remove the hose connector and plastic adapter of anempty slot.
4. Push on the extension module. 5. Connect the hose connector to the gas hose nipple of
the extension module. 6. Install the cell heater on the extension module. 7. Close the measuring cell cover. 8. Switch on the device with the control unit or using the
control software. 9. Insert the added parameter to the display sequence.
Heater measurement cell
Hoseconnector
Plastic adapter
99a
German calculation information
The following equations were used to calculate the following values:
CO2max x (21% - O2%) CO2max : fuel-specific maximumCO2 value: CO2 = CO2 value
21% 21 % : oxygen content of the air in %O2 % : measured oxygen content
in %
A2 FT : flue gas temperatureFlue gas loss: qA = (FT-AT) + B - KK AT : ambient (combustion) temp.
(21 - O2) A2/B : fuel-specific factors
21 : oxygen content of the airO2 : measured oxygen contentKK : is a factor which qA produces
as a negative value when the temperature is not reached.Required for measurement of fuelvalue systems.
If the fuel-specific factors A2 and B are equal to zero, the Siegert equation is used, applying the factor f.
(FT - AT) FT : flue gas temperatureqA = f x AT : ambient (combustion) temp.
CO2 CO2 : calculated CO2 valuef : fuel-specific factor
Technicalcombustionefficiency: Eta = 100 - qA If qA is negative, Eta becomes greater than
100%.
Excess air CO2max CO2max : fuel-specificindex λ: λ = maximum CO2 value
CO2 CO2 : calculated CO2 valuef : fuel-specific factor
NOx: NOx = NO + NOadd x NO NOadd : NO2 addition factor
Please refer to page 77
UK calculation information
Conversion of ppm to mg/m3 referred to the O2 reference value(freely selectable according to fuel)
CO (mg/m3) CO = x CO (ppm) x 1.25
02 set : Oxygen content in the airO2: Measured oxygen content
NOx (mg/m3) NOx = x NO (ppm) x 1 + x 2.05
02 set : Oxygen content in the airO2: Measured oxygen content
Conversion of ppm to g/GJ
CO (g/GJ) CO= x CO (ppm) x FBr x 1.25
NOx (g/GJ) NOx= x NOx (ppm) x FBr x 2.05
Conversion of (ppm) to mg / kWh
CO (mg/kWh) CO= x CO (ppm) x FBr x 3.6 x 1.25
NOx (mg/kWh) NOx= x NOx (ppm) x FBr x 3.6 x 2.05
FBr See “Fuel selection”
02 set - O2ref
(02 set -O2)
02 set - O2ref
(02 set -O2)
NO-add
100
Fitted onlywith NO:
02 set
02 set -02 meas.
02 set
02 set -02 meas.
02 set
02 set -02 meas.
02 set
02 set -02 meas.
99b
UK calculation information (continued)
Kgr x (FT - AT) X x (2488 + 2.1 x FT - 4.2 x AT)
CO2
K1 x CO
CO2 + COQgr x 1000
Knet x (FT - AT) X x (210 + 2.1 x FT - 4.2 x AT)
CO2
K1 x Qgr x CO
Qnet x CO2 + COQgr x 1000
Efficiencies
CO / CO2 ratio
Effg=100-
Effn=100-
+ +
++
CO (ppm)
CO2 (%) x 10000rat =
99c
X = MH2O+9H
Error messages Cause Remedy
NO-value drifts Loss of aux. voltage Measure only after 2hfor NO
Double module Measuring module already installedDilution Gas flow rate in the dilution
path too high / too lowO2 signal too high O2 value above 20.9% Switch device off / on
O2 cell spent Replace O2 cellCO signal unstable CO cell drifts excessively Replace cell if necessaryCO signal too high CO signal not at zero Wait until regenerated
CO swiched-off CO value above selectedswitch-off limit
NO signal unstable NO cell drifts excessively Replace cell if necessaryNO signal too high NO-signal not at zero Wait until regenerated
NO swiched-off NO-value above selectedswitch-off limit
NO2 signal unstable NO2 cell drifts excessively Replace cell if necessaryNO2 signal too high NO2 signal not at zero Wait until regenerated
NO2 swiched-off NO2 value above selectedswitch-off limit
Battery low Connect device tomains
Device Device temperaturetemperature is outside the
operating temperature rangePump flow rate Too low / too high Check pump / gas path
gas flow rateGas cooling system Gas cooler not working
Cell temperature Cell temperature outsidetoo high specifications
If we were unable to answer your question, please contact your distributor or Testo Cus-tomer Service. For contact data, see back of this document or web pagewww.testo.com/service-contact
100
Error messages
101
Technical data
General technical data
Storage and transport temperature ..................-20...+50 °CAmbient temperature ..........................................-5...+45 °CBattery operation time ........................................approx. 7h(with pump running, lighting off)
Analyser box testo 300 XXL Data memory: ........................................................1 MByteHousing:........................................................................ABSProtection class:............................................................IP20Max. flue gas pressure: ............................................50 hPaFlow rate of the internal pump: ................approx. 0.8 l/min.Weight:......................................................................3200 gDimensions in mm (L x W x H): ....................395 x 275 x 95Power supply: ..............integrated mains adapter 115/230 V
or Testo battery packNumber of measuring cells: ......................................max. 4 Battery operating time (depending on equipping): ................................ approx. 7 hO2Measuring range: ....................................................0...25%Accuracy: ..............................................±0.2 vol.% absolutet90:................................................................approx. 30 secResolution:..............................................................0.1 vol%CO2Measuring range: ..............................................0...CO2 maxAccuracy: ............................................................±0.2 vol.% t90:................................................................approx. 30 secResolution:............................................................0.01 vol%Note: ......................................................calculated from O2CO (H2 compensated)Measuring range: ..........................................0...10000 ppm Accuracy: ..........................................±10 ppm (<100 ppm)
±5% of value (100...2000 ppm)±10% of value >2000 ppm
t90:................................................................approx. 40 secResolution: ................................................................1 ppm
NO2Measuring range: ..............................................0...500 ppmAccuracy*: ............................................±5 ppm (<100 ppm)
±5% of value (³100 ppm)t90:................................................................approx. 40 secResolution: ..............................................................0.1 ppmNOMeasuring range: ............................................0..3000 ppmAccuracy: ............................................±5 ppm (<100 ppm)
±5% of value (100... 2000 ppm)±10% of value (>2000 ppm)
t90:................................................................approx. 30 secResolution: ................................................................1 ppmFine draught Measuring range: ....................................................±40 hPaAccuracy: ..............................................±0.03 hPa (<3 hPa)
±1.5% of value (³3 hPa)Resolution: ............................................................0.01 hPaNote: ....................................................max. overload 1 barPrecision draught analyser unitMeasuring range: ....................................................±10 hPaAccuracy: ..............................................±0.03 hPa (<3 hPa)
±1.5% of value (>3 hPa)Resolution: ..........................................................0.001 hPaNote: ....................................................max. overload 1 barTemperatureMeasuring range:..........................................-40...+1200 °CAccuracy: ......................................±0.5 °C (-40...+99.9 °C)
±0.5% of value (+100...1200 °C)Resolution: ..........................................0.1 °C (-40...99.9 °C)
1 °C (+100...+1200 °CNote: ..................................................thermocouple type KVelocity measurementMeasuring range: ................................................0...40 m/sAccuracy: ..............................................................±0.2 m/sResolution: ..............................................................0.1 m/sNote: ................................................calculated from dP, °C,
gas density, absolute pressureFlue gas dew pointMeasuring range: ............................................0...99.9 °C tdt90: ..........................................................................<20 secResolution: ............................................................0.1 °C tdNote: ........................................calculated from O2, °C, fuel
102
Technical data
* Accuracy only in conjunction with the installed gas preparation and a flue gas probe with NO2/SO2 hose.
103
Technical data
Differential pressureMeasuring range ............................................-40...+40 hPaAccuracy: ....................................1% of value (-3...-40 hPa)
0.03 hPa (-2.99...+2.99 hPa)1.5% of value (+3...+40 hPa)
Resolution......................................0.01 hPa (-40...+40 hPa)EfficiencyMeasuring range: ..................................................0...120%Resolution: ..................................................................0.1%Note: ........................................calculated from O2, °C, fuelFlue gas lossMeasuring range: ..............................................-20...99.9%Resolution: ..................................................................0.1%Note: ........................................calculated from O2, °C, fuelExcess airMeasuring range: ..................................................1...20.00 Resolution: ....................................................................0.01 Note: ........................................calculated from O2, °C, fuelPump flow rate displayMeasuring range: ..............................................0...1.5 l/minAccuracy: ......................10% of measuring range end valueDevice temperatureMeasuring range:................................................-5...+50 °CAccuracy: ................................................................±0.5 °CResolution: ................................................................0.1 °C
Control unitStorage and transport temperature ..................-20...+50 °COperating temperature........................................-5...+45 °CPower supply: ..........................................8V mains adapter
Testo rechargeable batteryBattery type AA
Housing:........................................................................ABSProtection class:............................................................IP20Weight:........................................................................850 gDimensions in mm (L x W x H): ....................252 x 115 x 58Display: ................................................LCD graphic displayPrinter: ..................................................................integratedData memory: ........................................................1 MByteautomatic detection of all connected probesCO probeMeasuring range ..............................................0...500 ppmAccuracy ........................................................±5% of valueCO2 probeMeasuring range ..........................................0...10000 ppmAccuracy ........................................................±5% of value
Pressure range 1Meas. range ..................................................-80...+80 hPaAccuracy..................................±1% of value (-50...-80 hPa)
±0.5 hPa (-49.9...+49.9 hPa)±1% of value (+50...+80 hPa)
Resolution ..................................±0.01 hPa (-80...+80 hPa)Max. overload ........................................................1200hPaPressure range 2Meas. range ..........................................-1000...+1000 hPaAccuracy............................±2% of value (-200...-1000 hPa)
±4 hPa (-199,9...+199,9 hPa)±2% of value (+200...+1000 hPa)
Resolution ............................±0.1 hPa (-1000...+1000 hPa)Max. overload ........................................................1200hPaTemperatureProbe type: ................................................Type K (NiCr-Ni)Measuring range:........................................-200...+1370 °CAccuracy: ........................................±1% (-200...-100.1 °C)
±0.4 °C (-100.2...+200 °C)±1 °C (-200.1...+1370 °C)
Resolution....................................±0.1 °C (-200...+1370 °C)Probe type: ..................................................................NTCMeasuring range:............................................-40...+150 °CAccuracy: ........................................±0.4 °C (-40...-10.0 °C)
±0.2 °C (-10.1...+50 °C)±0.4 °C (+50.1...+150 °C)
Resolution........................................±0.1 °C (-40...+150 °C)Probe type: ................................................................PT100Measuring range:..........................................-200...+800 °CAccuracy: ..........................................±1 °C (-200...-100 °C)
±0.4 °C (-99.9...-50 °C)±0.1 °C (-49.9...+99.9 °C)
±0.4 °C (+100...+199.9 °C)±1 °C (+200...+800 °C)
Resolution ......................................±0.1 °C (-200...-100 °C)±0.01 °C (-99.99...+300 °C)±0.1 °C (+300.1...+800 °C)
Probe type................................................Type J (Fe-CuNi)Measuring range:........................................-200...+1000 °CAccuracy: ........................................±1% (-200...-150.1 °C)
±0.4 °C (-150.2...+150 °C)±1 °C (+150.1...+1000 °C)
Resolution....................................±0.1 °C (-200...+1000 °C)Probe type............................................Type S (Pt10Rh-Pt)Measuring range: ............................................0...+1760 °CAccuracy: ..............................................±1% (0...+1760 °C)Resolution............................................±1 °C (0...+1760 °C)
104
Technical data
105
Technical data
Current/voltage measurementMeasuring range ..................................................0...20 mAAccuracy ............................................±0.04 mA (0...20 mA)Resolution ..........................................±0.01 mA (0...20 mA)HumidityResolution ................................±0.01 % RH (0...100 % RH)Velocity (thermal)Measuring range: ..............................................0...+20 m/sAccuracy:......................................±0.01 m/s (0...-1.99 m/s)
±0.02 m/s (+2...+4.99 m/s)±0.04 m/s (5...+20 m/s)
Resolution ......................................±0.01 m/s (0...+20 m/s)
WarrantyControl unit: 24 months (excluding wear parts
and printer mechanism)Analyser box: 24 months (excluding wear parts,
dirt filter and measuring cells)Measuring cells: CO/NO/NO2 12 months
O2 18 monthsProbes: 12 months (excluding filter)Battery: 12 monthsAccessories: 6 months
106
Ordering data
Designation Order no.
Control unitTrade version 0563.0354
Options for control unitIRDA interface to replace connectingcable control unit/PC 0440.0560Touchscreen with panel 0440.0559
Accessories for control unitVelocity probe 0635.1041Voltage/current probe 0554.0007Spiral connecting cable 0449.0052Mains adapter for control unit 0554.1084Powerbox 0554.1045Spare pen for touchpanel 0554.0563Printer paper 0554.0569Barcode pen 0554.0460CO probe 0632.1247CO2 probe 0632.1240Standard rechargeable batteries with charger 0554.0110Connecting cable control unit/PC 0409.0178TradeLeak detection probe 0632.1246Hose connection set, pressure 0554.0315
Analyser boxAdditional basic equipment, testo 300 XXL 0563.0352
Options for measuring boxPrecision draught measurement, 30 mbar 0440.3921NO2 measuring cell 0440.3926Gas preparation 0440.0355NO measuring cell 0440.3935
AccessoriesLockable bracket 0554.0203Connecting cable, spiral 0449.0052Connecting cable, 5m 0449.0043Connecting cable, 20m 0449.0044Carrying strap for measuring box or control unit 0554.0434PC evaluating software 0554.0508
107
Ordering data
Designation Order no.
AccessoriesPC configuration software 0554.0507Delta T probe set 0600.4593AT probe 0600.9797Flue gas probe, 700 mm, +1000 °C, special hose 2.2 m 0600.8551Flue gas probe, 335 mm, +1000 °C, special hose 2.2 m 0600.8552Flue gas probe, 335 mm, +500 °C, special hose 2.2 m 0600.8550Probe 500 °C, 180 mm 0600.9556Probe 500 °C, 300 mm 0600.9557Leather case 0516.0307Aluminium case 0516.0351
UpgradesTouchscreen (by testo service) 0554.0559Gas preparation (by testo service) 0554.3955NO measuring cell 0554.3922CO dilution 0554.0555NO2 measuring cell 0554.3926Fresh air valve 0554.0557Trigger input 0554.3932IRDA upgrade (by testo service) 0554.0560Solid fuel measurement CO (via testo service) 0554.0567
Spare parts / ConsumablesO2 measuring module 0390.0070CO/H2 measuring module 0390.0072NO measuring module 0390.0074NO2 measuring module 0390.0075Spare filter (20) 0554.3381Printer paper (6 rolls) 0554.0569Pump diaphragm for main gas/rinsing gas pump 0193.0072Pump diaphragm for main gas pump (with gas preparation option) 0193.0049Pump cassette of condensate transport pump (with gas preparation option) 0440.0013Rechargeable battery pack for analyser box 0554.1097Rechargeable battery pack for control unit 0554.1098Mains unit for control unit 0554.1084
Control unit
108
Ordering data
NiCr-Ni surface probe
Very fast reacting surface probe with flexiblethermocouple band, rated length = 150 mm, probe tip Ø 10 mm
Very fast reacting surface probe with flexiblethermocouple band, probe tip inclined 90°,rated length = 49 mm, probe tip Ø 10 mm
Rugged surface probe, straightrated length = 150 mm, probe tip Ø4 mm
Rugged surface probe, inclined by 90°, for constrictedconditions; rated length = 130 mm, probe tip Ø 4 mm
Rugged probe with flexible TE bandrated length = 200 mm, probe tip Ø 15 mm
Pipe contact probe for pipes up to 2ý diameterSpare measuring head
Magnetic probe for measurement on metal surfacesa) Adhesion force approx. 20 Nb) For higher temperatures, adhesion force approx. 10 N
Flat-head probe with telescope (215 ... 660 mm), for rapidmeasurements at poorly accessible points, measuring head Ø 25mm, height 12 mm
Miniature probe for measurements at electronic components,small motors…;rated length = 100 mm, probe tip Ø 5 mm
Roller probe for surface measurements on rollers and rotating drums, perm. peripheral speed..18…400 m/min, rated length = 274 mm, probe tip Ø 33 mm
Cu-CuNi probe
NTC probeHigh-precision probe for measurement of air and gastemperatures with exposed, mechanically protected detector
Pt100 surface probeRugged surface probe,rated length = 150 mm, Ø probe tip 9 mm
Velcro probe for pipes with Ø max. 100 mm
-200…+300 °Cbriefly up to +500 °C
Class 23
-200…+600 °CClass 1
25
-200…+600 °CClass 1
25
-200…+700 °CClass 2
3 Spiral cable PUR
3
-60…+130 °CClass 2
5 1.5 mPVC
Class 2-50…+170 °C-50…+400 °C
L =1.5 msilicone glass fibre
-50…+250 °CClass 2
<3 1.8 mPVC
-50…+400 °C (briefly upto +500 °C), Class 2 3 1.5 m
PVC
on request
-50…+240 °CClass 2
-40…+350 °CClass 1
Spiral cable PUR
0604.01940614.0194*
0604.09940614.0994*
0604.99930614.9993*
0604.98930614.9893*
0600.0394
0600.45930602.0092
0600.47930600.4893
0600.2394
0600.1494
0600.5093
Measuringrange/accuracy
t99sec.
Connecting cable Order no.
-200…+300 °Cbriefly up to +500 °C
Class 2
-40…+130 °Cto UNI curve 60
Spiral cablePUR
0610.9714
-50…+400 °C, Class B 40Plug-on headOrder connecting
cable (p. 56).
0604.99730628.0018*
-50…+150 °C, Class B 40 1.6 m, PTEEribbon cable
0628.0019
Plug-on head
Order connecting cable
(page 56).
* with EEPROM: • Measuring range restriction saved in probe • Adjustment to 1 measuring point possible • Surface addition for surface probe adaptable to measuring task
109
Ordering data
-200…+600 °CClass A
75
75
0604.9773
0628.0017*
* with EEPROM: • Measuring range restriction saved in probe • Adjustment to 1 measuring point possible • Surface addition for surface probe adaptable to measuring task
-200...+1250 °CClass 1
601.5 m PVC
0600.5993
0363.1712
all Class 1
-200…+900 °C-200…+900 °C-200 +1100 °C-200 +1100 °C
3.53.53.53.5
0600.53930600.54930600.57930600.5893
Spiral cable PUR 0600.5593
0604.97940614.9794*
-200…+400 °CClass 1 3 0604.0293
0614.0293*-200…+600 °C
Class 1 1
0604.05930614.0593*
1-200…+1100 °CClass 1
-200…+600 °CClass 1
<1
-200…+400 °CClass 1
1.5 msilicone
0600.25933
0604.04930614.0493*
Fast-response immersion/penetration proberated length = 150 mm, probe tip Ø 3 mm
Very fast-repsonse immersion/penetration probe for meas. inliquids; rated length = 150 mm, probe tip Ø 1.5 mm
Very fast-response immersion/penetration probe for hightemperatures, rated length = 470 mm, probe tip Ø 1.5 mm
Very fast-response immersion/penetration probe formeasurements in gases and liquids with thin, lightweight tip; rated length = 150 mm, probe tip Ø 0.5 mm
Rugged immersion/penetration probe in V4A stainless steel,watertight and heat resistant, e. g. for foodindustry; rated length = 150 mm, probe tip Ø 3 mm
Smelt probe for measurements in non-ferrous metal smelts with exchangeable, inclined tip. Service life of the tip: up to500 measurements in aluminium smelts
Spare tip
Plug-on tips for high temperatures, flexible. Please also order handle (see below).a) Outer jacket, stainless steel 1.4541b) Outer jacket, stainless steel 1.4541c) Outer jacket, Inconel 2.4816d) Outer jacket, Inconel 2.4816
Handle for plug-on measuring tips
Pt100 air probeStandard air proberated length = 150 mm, Ø probe tip 9 mm
Precision air proberated length = 150 mm, Ø probe tip 9 mm
NiCr-Ni immersion/penetration probe Measuringrange/accuracy
t99sec.
Connectingcable
Order no.
Plug-on head
Order connecting cable
(page 56).
** 1/10 Class B (0…+100 °C)1/5 Class B (rest of range)
to DIN IEC 751
-100…+400 °C 1/10 Class B**
Plug-on headOrder connecting cable(page 56).
Plug-on headOrder connecting cable(page 56).
110
Ordering data
Please alsoorder adapter
0600.1693
0644.1109Pack of 5
max. temp. +400 °C
max. temp. +200 °C0644.1607Pack of 2
0.3 m PVC 0600.1693
0…+120 °C±0.5 °C (0...50 °C)±1 °C (50...120 °C)
1.5 m 0554.0670
-18…+260 °C*** 2 0600.0750Spiral cable
PUR
0554.0004
Standard immersion/penetration proberated length = 200 mm, probe tip Ø 3 mm
Precision immersion/penetration proberated length = 200 mm, probe tip 3 mm
Glass sheath for immersion/penetration probes 0604.0273 and0628.0015
Flexible precision immersion probe rated length = 1000 mm,
Rugged probe with ground tip, watertight and heat resistantrated length = 150 mm, Ø probe tip 3 mm
NiCr-Ni thermocouple pairsThermocouple pair in glass fibre insul. thermoconductors Insulation:Double conductor flat oval, individually countering with glass fibre braiding, both conductorswith common glass fibre braiding and impregnated with lacquer; wire Ø 0.2 mm; outside Ø1 x 0.8 mm
Thermocouple pair for adhesive att., alumin. foil substrateAttach to measuring points with common adhesives or silicone thermal conduction paste, orderno. 0554.0004, Ø extension 2 x 0.2 mm, thickness 0.1 mm
Adapter for connection of NiCr-Ni thermocouple pairs and probeswith exposed wire ends
Misc. temperature probesGlobe thermometer to measure radiated heat, NTC probe, globe Øapprox. 150 mmComplies with requiremements of ISO 7243, ISO 7726, DIN EN 27726, DIN 33403.
Infrared probe for contact-free temperature measurement at live, poorly accessible and rotating parts
Accessories for temperature probesSilicone thermal conduction paste (14 g), Tmax = +260 °C
* with EEPROM: • Measuring range restriction saved in probe • Adjustment to 1 measuring point possible • Surface addition for surface probe adaptable to measuring task
***fulfills EN 61326 - 1 : 1997
-200…+400 °CClass A
20 0604.0273
-100…+400 °C1/10 Class B**
30 0628.0015*
-200…+400 °CClass A 30
1.5 msilicone
0604.2573
0554.7072
-100…+400 °C1/10 Class B**
80in water
0628.0016*
** 1/10 Class B (0…+100 °C)1/5 Class B (rest of range)
to DIN IEC 751
Plug-on headOrder connectingcable (page 56).
Plug-on headOrder connectingcable (page 56).
Plug-on headOrder connectingcable (page 56).
Pt100 immersion/penetration probe Measurementrange/accuracy
t99sec.
Connecting cable
Order no.
Please alsoorder adapter
0600.1693
111
Ordering data
Standard room air probe up to +70 °CØ 12 mm (probe tip)
Duct humidity/temperature probe,telescopic extension attachable, Ø12 mm
Telescope, length 340 - 800 mm
High-precision reference humidity/temperature probe, Ø 21 mm
Flexible humidity probe with minimodule formeasurements e. g. on material test benches.Module cable length 1500 mm
0…100 % RH(probe tip)
-20 +70 °C
±2 % RH (2...98 % RH)±0.4 °C(0…50 °C)±0.5 °C (rest ofrange)
<12at
2 m/s
0636.9740
0…100 %RH(probe tip)
-20…+70 °C
±2 % RH (2..98% RH)±0.4 °C (0…50 °C)±0.5 °C (rest of range)
****
<12at
2 m/s
3 mPUR
0636.9715
0430.9715
0…100 % RH(probe tip)
-20…+70 °C
±1 % (10...90 % RH)
(15…30 °C)
±2 % (rest of range)****
<12at
2 m/s
0636.9741
0…100 % RH-20…+125 °C
±2 %RH (2...98 %RH)±0.4 °C (0…50 °C)±0.5 °C (rest of range)
20 0628.0013
* Specified accuracy at rated temperature +25 °C, temperature coefficient ±0.03 % / °C
Other probes for testo 950, 650, 400
CO probe
CO2 probe
Leak detection probeMechanical rpm probe with plug-on head.
Current/voltage cable (±1 V, ±10 V, 20 mA)
0…500 ppm ±5 ppm (0…100 ppm)±5 % of value (r.o.r.)
0632.1247
0…1 vol. % CO20…10000 ppm CO2
50 ppm±2 % ofvalue (0..5000ppm)100 ppm±3 % ofvalue (rest of range)
0632.1240
0554.0007
20…20000 rpm ±1 digit 0640.0340
0 20mA/4 20mA 0±1V/0 ±10V
±0.04 mA±0.001 V / ±0.01 V
Meas. range Accuracy * Order no.t99sec.
Connectingcable
Order connectingcable
(page 56).
Humidity/temperature probe for air conditioning/ventilation
Plug-on headOrder connecting
cable (page 56).
Plug-on headOrder connecting
cable (page 56).
Plug-on headOrder connecting
cable (page 56).
***fulfills EN 61326 - 1 : 1997****fulfills EN 61326 - 1 : 1997 in connection with control unit
112
Ordering data
0…100 % RH-20…+180 °C
±2 %RH (2...98 %RH)±0.4 °C (0…50 °C)±0.5 °C (rest of range)***
<30at
2 m/s0628.0022
0…100 % RH-50…+50 °C tpd
Pressure dew pointmeasurement at rated temp.+25°C:±0.9…>4 °C tpd(5…40 °C) *****
1 5min
typically
2 min
0636.9840
0…100 % RH-50…+50 °C tpd
Pressure dew pointmeasurement at rated temp.+25°C:±0.8…4 °C tpd(5…40 °C) *****
1 5min
typically
2 min
0636.9841
0…100 % RH-20…+140 °C
±2 %RH (2...98 %RH)
±0.4 °C (0…50 °C)±0.5 °C (rest of range)***
<30at
2 m/s
0628.0014
0 100 % RH-20 +120 °C
±2 %RH (2...98 %RH)±0.4 °C (0…50 °C)±0.5 °C (rest of range)***
<30at
2 m/s
0636.2140
0…100 % RH-20…+180 °C
±2 %RH (2...98 %RH)±0.4 °C (0…50 °C)±0.5 °C (rest of range)***
<30at
2 m/s
0628.0021
* Specified accuracy at rated temperature +25 °C, temperature coefficient ±0.03 % / °C
0…100 % RH(probe tip)-20…+70 °C
±2 %RH (2...98 %RH)±0.4 °C (0…50 °C)±0.5 °C (rest of range)***
<12at
2 m/s
Plug-on headOrder connecting cable
(page 56).
Plug-on headOrder connecting cable
(page 56).
Plug-on headOrder connecting cable
(page 56).
Plug-on headOrder connecting cable
(page 56).
Plug-on headOrder connecting cable
(page 56).
Plug-on headOrder connecting cable
(page 56).
Plug-on headOrder connecting cable
(page 56).
0636.0340
Humidity/temperature probe for heavy industrial use
Bar probe forhumidity/temperature measurements in stacked goods, rated length 320 mm
Rugged humidity probe, e. g. for measurement ofmaterial compensating humidity or formeasurements in air exhaust ducts, rated length300 mm, Ø 12 mm
Rugged, high-temperature probe up to+180 °C, rated length 300 mm, Ø 12 mm
Flexible, non-rigid humidityprobe for measurements at poorlyaccessible points
Standard pressure dew point probe formeasurements in compressed air systems
Precision pressure dew point probe formeasurements in compressed air systems
Flexible, non-rigid humidityprobe for measurements at poorly accessiblepoints.
Meas. range Accuracy * Order no.t99sec.
Connectingcable
100...66 wet
0...1 very dry0636.0565
freely scalable, comparative mea-surements, no water content
0...100kΩ= 100...0%
0636.0365Material humidity probe
Material/building humidity cable
^
***fulfills EN 61326 - 1 : 1997
*****does not fulfill: EN 61326 - 1 : 1997; does not fulfill EN 61326 - 1 : 1997 / A1:1998
113
Ordering data
Adapter for humidity probe 0636.XX60 and Pt100 probe 0605.XX73/XX72
Connecting cable for probes with plug-on head, length 1.5 m, jack material PUR
Connecting cable for probes with plug-on head, length 5 m, jack material PUR
Extension lead between plug-on head cable and device, length 5 m, jack material PUR
Telescope for probes with plug-on head, max. extension length 1 m, cable length 2.5 m, jack material PUR
Inspection and calibration set for humidity probes and triple probe
Adapter for humidity calibration, triple probe 0635.1545, order with calibration set.
Inspection and storage humidity (33 %RH) for humidity probes and triple probe
Ø 21 mm
Ø 12 mm
0554.0640
0554.0647
0628.0012
Ø 21 mmØ 12 mm
0554.06650554.0755
Ø 21 mm 0554.0667
Cover caps for all humidity probes Ø 12 and 21 mm
Metal guard basket, material stainless steel V4A. Rapid compensation time, rugged andheat resistant. Application: for flow velocities <10 m/s.
Wire mesh filter, material stainless steel V4A. Rapid compensation time, protects againstdirt and destruction. Application: meteorology, water spray, condensation.Installation filter for metal guard basket and plastic cap.
Cap with wire mesh filter
PTFE sintered filter, material PTFE. Favourable behaviour in condensation, waterrepellent, high resistance to aggressive media. Applications: compressed air measurements, wet areas (long-term measurements), highflow velocities.
Stainless steel sintered cap, material stainless steel V2A. Very rugged, suitable forpenetration, can be cleaned with compressed air, mechanical probe protection.Applications: High mechanical loads, high flow velocities.
Surface adapter for humidity probes Ø 12 mm
Ø 12 mm 0554.0757
Ø 21 mm
Ø 12 mm
0554.0666
0554.0756
Order no.
Order no.
Connecting cable
0554.0454
0430.0143
0430.0145
0409.0063
0430.0144
0554.0660
0554.0661
0554.0636
Accessories for humidity/temperature probes, triple probe + rpm probe
114
Ordering data
0.6…20 m/s-30…+140 °C *
±(0.2 m/s ±1 %of value) 0635.9443
0.4…60 m/s-30…+140 °C *
±(0.2 m/s±1 % of value****
0635.9540
0.4…40 m/s-30…+140 °C *
±(0.2 m/s±1 % of value****
0635.9640
0.25…20 m/s ±(0.1 m/s+1.5 % of value)
0635.9440
0.2…15 m/s ±(0.1 m/s+1.5 % of value)
0635.9340
0.7…30 m/s ±0.3 m/s, ±5 % of value 0635.9045
Differential pressureDifferential pressureAbsolute pressure
±10 hPa*****±100 hPa***2000 hPa***
±0.03 hPa±0.1 hPa (0…20 hPa)±0.5 % of value (rest)±5 hPa
0638.14450638.15450638.1645
Pressure cellPressure cell
-1…10 bar0…30 bar
±1% of high range,overload protected upto 32 bar±1% of high range,overload protected upto 70 bar
0638.17400638.1840
Telescope for plug-on vane probes/ length max. 1 m/ extension on request/ connecting cable for pressure probes 0638.1740 and 0638.1840
Handle for plug-on vane probes
Gooseneck, flexible connection between measuring probe and connector
Magnetic probe holder for vane probes
2.3 m silicone 0430.0941
DIN round plug1.5 m silicone
0430.3545
0430.0001
0554.0430
Pressure probes Meas. range
Meas. range
Measuring method Accuracy
Accuracy
Connecting cable
Order no.
Order no.
Order no.
Order no.Accessories for plug-on vane probes
Pressure probe for measurement of flowvelocities and differential pressure a)
b) or absolute pressure c)Pressure probe in refrigerant-resistant stainlesssteel, without cable, screw-in thread 7/16”UNFa) Low-pressure probeb) High-pressure probe
Accessories for pressure probes0554.02250409.1740
Magnetic holder for pressure probes 0638.1445 / 0638.1545 / 0638.1645 Connecting cable for pressure probes 0638.1740 and 0638.1840
Plug-on velocity probes
Vane measuring probe, plugs onto handle or telescopeØ 12 mm
Vane/temperature measuring probe, plugs onto handle ortelescope, Ø 16 mm
Vane/temperature measuring probe, plugs onto handle ortelescope, Ø 25 mm
Plug-on vane for integrating velocity measurement. Operating range -20…+60 °C, 90° angling, Ø 60 mm
Plug-on vane for integrating velocity measurement. Operating range -20…+60 °C, 90° angling, Ø 100 mm
Cup anemometer for meteorological wind measurement.Cable 3 m.
* in short-term measurments
***fulfills EN 61326 - 1 : 1997****fulfills EN 61326 - 1 : 1997 in connection with control unit
*****does not fulfill: EN 61326 - 1 : 1997; does not fulfill EN 61326 - 1 : 1997 / A1:1998
115
Ordering data
3-function probe for simultaneous measurement of temperature,humidity and velocity. With plug-on head – please order connectingcable (see p. 56).
0…10 m/s0…100 %RH(probe tip)-20…+70 °C
as 0635.1549****±2 %RH (2…98 %RH)±0.4 °C (0…+50 °C)±0.5 °C (rest of range)
0635.1540(see note)
Comfort probe for turbulence measurements, with telescope and tripod. Fulfils the requirements ofDIN 1946 Part 2 and VDI 2080
0…5 m/s0…+50 °C
0…5 m/s: ±(0.03 m/s ±4% of value)
±0.3 °C
0628.0009(see note)
Note: Thermal probes are calibrated to the average annual atmospheric pressure of 1013 hPa atsea level when calibrated in m/s. In still air (v < 0.5 m/s), the displayed temperature value will be in error due to local heatingeffects from the thermal probe.
0…10 m/s-20…+70 °C
0…10 m/s: ± (0.03 m/s +5 % of value)
0…10 m/s:± (0.03 m/s + 5 % of value)
0…20 m/s: ± (0.03 m/s +4 % of value)
0635.1549(see note)
0…10 m/s-20…+70 °C
0635.1049(see note)
0…20 m/s-20…+70 °C
0635.1041(see note)
0.6…20 m/s-40…+350 °C
±0.2 m/s ±2 % of value**** 0635.6045
Meas. range Accuracy Order no.
Economical, rugged hot ball probe for measurements in the low flow raterange. With handle.
Rugged hot ball probe with handle and telecope (190 - 850 mm)for measurements in the low flow rate range.
Rapid-response hot wire probe with telescope (160 - 760 mm)for measurements in the low flow rate range with direction detection.
High-temperature probe with handle for long-term measurements up to+350 °C / Ø 25 mm
Velocity probes with handle/telescope for testo 400
Meas. range Accuracy Order no.Other probes
Accessories for velocity probes Order no.
Stainless steel pitot tubes(longer designs on request)
Silicone hose to connect pitot tubes with pressure probes
Volume flow rate measuring funnel to measure the suction rate of ventilationsystemsa) Measuring range 20…400 m3/h; funnel factor 22b) Measuring range 20…400 m3/h; funnel factor 22- In conjunction with 0635.1041 or 0635.9540. -
+600 °C+600 °C+600 °C+600 °C+1000 °C+1000 °C
Temp. max. Ø
7 mm7 mm8 mm8 mm8 mm8 mm
Length
350 mm1000 mm500 mm1000 mm350 mm750 mm
0635.21450635.23450635.21400635.22400635.20410635.2042
0554.0440
0554.04000554.0410
5 m
****fulfills EN 61326 - 1 : 1997 in connection with control unit
Notes
Notes
0970 3003 en 02 en-GB