4 glass tubes for toning
4 rubber stoppers nº5
4 stirrer magnets 6x20 mm
1 nut clip for burette
1 inox rack for 12 tubes
1 power voltage feeder
1 titration liquid 500 mL
1 starch indicator 250 mL
1 acid 500 mL
1 alkali 500 mL
1 beaker 100 ml p.p. for furring
1 instructions manual
ref. 3026056
ref. 47185050
ref. 137115
ref. 63204115
ref. R292
ref. 0002065
ref. 1006009
ref. 1005001
ref. 1006008
ref. 1006010
ref. 409100
Included material
ref. 1010012
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
·
·
·
·
1 pipette 25 mL double line A class
2 graduated pipettes 5 mL
1 graduated pipette 10 mL
1 burette 25 mL
ref. 3019026
ref. 3019013
ref. 3019017
ref. 3019068
Required material
·
·
·
·
·
·
·
·
Ripper method
simple and easy to use
very practical
without maintenance
no need of calibration
autotune
economic reagents
compact design
power voltage: 12V·
Characteristics
175m
m
15
0m
m
165m
m
50
0m
m
Peso2,15 Kg
1011
12
13
With the help of pipettes add in one of the tubes
to analyse, and
Place the tube inside the Toning.
till the limit of the upper side of the instrument.
Place the tip of the burette in the center of the tube.
Press the key.A led will shine counterclockwise, the sample will
stir and will autotune. Once finished the autotune a double “pip” will
sound indicating the begining of the analysis. , whites, will
remain illuminated indicating . Otherwise and
indicating in adjustment. In this case
.
Add the reagent observing the luminous variation from the circle of leds.
When all leds are nearly iluminated add the reagent more carefully till
maintain all the . In this case 3 long beeps
will sound indicating end of analysis.
Take note of the spended volume in the burette.
and will get the content of free sulphurous in
milligrams per liter [ ].
For further analysis follow .
25 mL of the wine
sample 1-2 mL of Starch Indicator 5 mL of Acid.
Add distilled water
run/stop
Led #1 and #2
adjustment OK led #1 led
#2 will blink error go to
observations
circle lighted during 10 seconds
The read volume is
multiplied by 10 factor
mg/L SO
steps from 05 to 122
06
070809
Operative Techinque
Free sulphurous:
05
InstructionsAssembly
Set the rod in the corresponding hole from the back and turn it
clockwise, later set the nut clip.
Place a 25 mL burette in the burette support. with
.
Place the glass tubes in the rack and put one stirrer magnet inside them.
Connect the power voltage feeder and press . The leds will shine
on indicating current in the instrument. will remain .
Fill it Titration Liquid
(Potassium Iodate N/128)
power
Led #1 illuminated
01
02
0304
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
Using a fill, with the help of a pipette,
, and .
Fill with water till the mark and place inside the instrument.
Fill the burette and center it in the middle of the tube.
Press the .You will hear an acoustic
signal and will begin the special mode for modena vinegars.
Analyse like in wines. Follow .
Once the analysis is finish the spent volume in the burette by
(see next page table) and will obtain the mg SO /L.2
tube of 25 cm. long 10 mL of
modena vinegar sample 2 mL of Starch Indicator 5 mL of Acid
run/stop button during 2 seconds
steps 10 and 11
multipy 25
factor
20
212223
24
With the help of pipettes add in one of the tubes
to analyse and .
Cover it with the rubber stopper, mix it and
in the rack.
Once time elapsed uncover it and add and
.
Follow .
For further analysis follow .
25 mL of the sample of
wine 10 mL of Alkali
let it rest during 15 minutes
1-2 mL of Starch Indicator 5
mL of Acid
steps from 06 to 12 without adding distilled water
steps from 14 to 17
14
15
16
1718
Introduction and Operative Technique for balsamic vinegars
Fill and
Cover it, mix it, and let it during .
Past this time add and .
Follow former steps .
To repeat an analysis follow steps .
10 mL of modena vinegar sample 15 mL of Alcali.
rest 5 minutes
2 mL of Starch 8 mL of Acid
from 21 to 25
from 20 to 25 and from 26 to 29
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
2627282930
25
By default, balsamic vinegars don’t have (or just a few which is
contemptible) free SO but, by effect of sugar and the intense colour, an
error is titrated.This error can be quantified and is generated the same
one when titrating total SO . Once analysed, this error is substracted
from total SO result obtaining the mg/L of real total SO .
2
2
2 2
19
Total sulphurous:
Step_1: Free sulphurous:
Step_2:Total sulphurous:
Observations
It is recommended to add the sample and reagents with the glass tube
inside the rack to avoid spill liquids over the instrument.
The must be before each analysis.
having added the Acid reagent.
If and means that the sample of wine have not been able to
pass the autotune because it may be a very intensive in colour wine (bigger
than ) or very muddy (bigger than
).In such case repeat for the free SO for total
SO with and where accor-
ding to attached table from next page.The graphics show an illustrative sum-
mary:
It is important not to touch the tube once is inserted inside the
instrument as glass walls from the tube are not completely regular so a
turn or movement will alter the light beam.
The addition of reagent must be done in continuous way.When the
is achieved it is essential the addition by drops in order not to pass
over.
For the totals analysis it is not necessary to add water.
To the instrument press button during .
Burette not included.Any burette can be used from the laboratory.
2
2
tube clean and dry
Analyse immediately after
led #1 led #2 blink
15 units of intensity 3000 units of turbidity
NTU steps 05 to 12 or 14 to 17
10 mL of sample the multiplier factor will be 25
final
zone
disconnect power 3 seconds
Try to maintain always the same working conditions.
To enter in press the button during
till 3 short beeps are heard.
To recover normal working mode press once button.
After 15 minutes of not working (no analysis done) the instrument is auto
shutdown.
modena vinegars mode run/stop 2
seconds
run/stop
NTU
0 3000
25 mL 10 mL
I
0 15
25 mL 10 mL
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
04
06
01
07
04
09
08
02
01
03
10 20 25
IODINE N/50 64 32 25.6
IODINE N/128 25 12.5 10
TABLE OF MULTIPLIER COEFFICIENTS
SAMPLE VOLUME [mL]
Codes guide
Led #1
Current
run/stop
power
, white, lighted:
in the instrument.
Connect: press once.
Shut down: press during 3 seconds.
Led #1 Led #2 led #3
OK
and , white, lighted and ,
blue, blinking:
Adjustment . Beggining of analysis.
Led #1 Led #2
not OK 10
mL of sample multiplier factor 25
and , white, blinking:
Adjustment . Repeat analysis with
and .
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
All blue leds lighted:
.Will keep like this till
or is pressed.
Read the burette spent volume and multiply by
the corresponding factor according to table.
End of analysis run/stop
power
Codes guide
It covers all the manufacture defects, provided that no manipulation has
been done by not authorised personnel, during 2 years, except defects
induced for bad use.
Warranty
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
All leds lighted:
If it keeps like this
during 10 seconds the analysis will be finished,
otherwise 1 or 2 more drops of Titration
Liquid will be needed.
Testing end of analysis:
GAB Sistemática Analítica S.L.
Sant Jordi, 30
08734 Moja-Olèrdola
BARCELONA [Spain]
Tel. +34 938 171 842
Fax. +34 938 171 436
www.gabsystem.com