  • IRT670USER’S MANUAL

    High Performance 30:1, Wide-rangeInfrared Thermometer with Memory

    INTRODUCTION:Thank you for purchasing this Infrared Thermometer. This is a non-contact infrared temperature measuring instrument. Tomeasure a temperature, point the unit at the desired object, pull the measuring trigger and hold. Make sure the target area islarger than the unit’s spot size. For larger target objects, you must be within the target distance.

    IRT670 Manual-FINAL3-050608:Layout 1 5/6/08 9:40 AM Page 1

    99 Washington Street Melrose, MA 02176 Phone 781-665-1400Toll Free 1-800-517-8431

    Visit us at www.TestEquipmentDepot.com

    http://www.testequipmentdepot.com

  • Features include broad temperature and high DS ratio. Thisallows the user to monitor the target from a long distance;far away from potential risk.

    FEATURES:• High DS ratio• Adjustable emissivity from 0.1 to 1.00 in 0.01 steps• Ultra low power consumption in shutdown mode• Extended long time measuring reliability• Laser sighting On/Off is switchable• Backlit LCD display• °F or °C selectable• Electronic trigger lock• Temperature data storage• Audible alarms

    APPLICATIONS:• Manufacturing processes of semiconductor technology• Automotive repair and maintenance• Food safety and processing• HVAC energy audits• Electrical troubleshooting• Testing terminals on circuits• Scientific experiment• Air conditioning

    SAFETY INFORMATION:Read the following safety information carefully beforeattempting to operate or service the meter. Only qualifiedpersonnel should perform repairs or servicesnot covered in this manual.

    LASER WARNING NOTE!Do not point laser directly at eyes.Use caution around reflective surfaces.

    Keep out of reach of children!

    CAUTIONS!DO NOT submerge the unit in the water.This product is not designed for use in medical evaluations.The product can only be used to measure body temperaturefor reference. It is meant for industrial and scientificpurposes.

    SAFETY SYMBOLSDangerous, refer to this manual before using themeter.CE Certification

    This instrument conforms to the following standards:EN61326: Electrical equipment for measurement, control and

    laboratory use.IEC61000-4-2: Electrostatic discharge immunity test.IEC61000-4-3: Radiated, radio-frequency, Electromagnetic

    field immunity test.IEC61000-4-8: Power frequency magnetic field immunity

    test.Tests were conducted using a frequency range of 80-1000MHz with the instrument in three orientations.The average error for the three orientations is ±1.0 °F(±0.5 °C) at 3V/m throughout the spectrum. However,between 781-1000MHz at 3V/m, the instrument may notmeet its stated accuracy.

    The device may not be disposed in the trash. Itpromotes the re-use recycling and other forms of

    recovery of used materials and components, and toimprove the environmental performance of all operators(manufacturers, traders, treatment facilities) involved in thelife cycle of products. Dispose of the product appropriatelyin accordance with relevant regulations.

    IRT670 Manual-FINAL3-050608:Layout 1 5/6/08 9:40 AM Page 2

  • DisplayScreen

    Control Panel

    Laser Sighting

    MeasuringWindow

    MeasuringTrigger

    Battery andCover

    SPECIFICATIONS:Distance/Spot Ratio: 30:1Temperature Range: -25.6° to 1400°F (-32° to 760°C)Accuracy: ±5°F (±3°C) within -25.6° to -4°F (-32° to -20°C)

    ±3°F (±2°C) within -4° to 212°F (-20° to 100°C)±2% within 212° to 1400°F (100° to 760°C)

    Thermopile: 8 ~ 14µmRepeatability: ±2°F (±1°C)Resolution: 0.1°F (0.1°C)Response Time: 500 ms.Operating Temperature: 32° to 122°F (0° to 50°C),

    10 to 90%RHAuto Power Off: Automatically after approx. 6 secondsEmissivity: Adjustable 0.1 ~ 1.0°F/°C Switchable: YESBacklight: YESLaser Sight Switchable: YESAudible alarm: YESMax/Min/Avg/ΔT: YESAuto Measuring: YES10 Point Memory: YESDual Display: YESTripod Mount: YESDimensions: 7.8" x 6.5" x 2" (200 x 166 x 51mm)Battery Type: “9V” batteryWeight: 10.6 oz. (300 g) Approx.Accessory: “9V” battery, Instruction Manual,

    Soft Carrying Case

    OPERATIONS OF INSTRUMENTQUICK START:To measure a temperature, point the unit at the target thatyou want to measure, pull themeasuring trigger and hold.Be sure to consider the target area that is inside the angle ofvision of this instrument. The single spot of laser is used foraiming only

    UNIT DIAGRAM:

    IRT670 Manual-FINAL3-050608:Layout 1 5/6/08 9:40 AM Page 3

  • 1. Laser/Buzzer On/Off2. Low Battery3. Data Log4. ΔT/Emissivity5. High/Low Alarm6. Memory Key7. Down Button8. SCAN/HOLD/AUTO9. °F/°C Indication

    10. Main TemperatureDisplay

    11. Max/Min/Avg12. Set/Record Value13. Lock Key14. Mode Key15. Up Button

    Parameter Setup Mode

    • Scan and display thetemperature duringmeasurement

    • The maximum temperaturewithin one shot measuring

    • The minimum temperaturewithin one shot measuring

    • The average temperaturewithin one shot measuring

    • The highest and lowesttemperature difference withinone shot measuring

    • Recall the stored data byusing � or � button

    • Emissivity for more accuracyon different materials

    • Highest alarm temperaturesetting

    • Lowest alarm temperaturesetting(*The parameters areadjusted by using� or � buttons)

    LCD AND CONTROL PANEL: OPERATION FUNCTIONS:To operate the advanced functions of the unit, press theMODE button to change the functions. The sequentialoperations and the corresponding explanations are shownin the following chart.

    OPERATION MODE

    IRT670 Manual-FINAL3-050608:Layout 1 5/6/08 9:40 AM Page 4

  • OPERATION REMARKS:• Memory: Activate the record function by pushing the “M”button. To delete all the records, press the up or downbutton to DATA0 and then press the “M” button.

    • LOCK: Push the button to continuously measure anddisplay the temperature without pulling the measuringtrigger.

    • The above functions can always be activated by followingany of the steps of the Operations Mode in the flow chart.

    • In SCAN mode, the LCD display both the currenttemperature in Celsius or Fahrenheit. The unit will HOLDthe last reading for 30 seconds after the trigger isreleased. When the battery is low, the battery icon willdisplay, but the unit will continue to function.

    • While DATA# flashes on the left bottom, the value on theMain Temperature Display can be recorded in “#” log.Simply press the “M” button.

    °F/°C AND BATTERY CHANGE:The unit is powered by“9V” battery and displaystemperature in either °F/°C.The user has to replace thebattery when the batteryvoltage drops below thevoltage for reliableoperation and at the sametime the low batterysymbol will appear. Tochange the “9V” battery,pull and open the unit’shandle by using the finger.Change the “9V” batterywith a new one and pushthe battery cover back inplace.

    TECHNIQUES OF INFRARED THERMOMETER:FIELD OF VIEW (FOV) RATIO =DISTANCE TO DIAMETER (DS) RATIOThe field of view is the angle of vision at which theinstrument operates and it is determined by the optics of theunit. The FOV is the ratio of the distance from the target tothe target diameter. The smaller the target, the closer youshould be to it. When the target diameter is small, it isimportant to bring the thermometer closer to the target toinsure that only the target is measured, excluding thesurroundings.

    EMISSIVITY:Emissivity is the ability of an object to omit or absorb energy.Perfect emitters have an emissivity of 1, emitting 100% ofincident energy.An object with an emissivity of 0.8 will absorb 80% andreflect 20% of the incident energy. Emissivity is defined asthe ratio of the energy radiated by an object at a giventemperature to the energy emitted by a perfect radiator atthe same temperature. All values of emissivity fall between0.0 and 1.0.Non-contact temperature sensors measure IR energy emittedby the target, have fast response, and are commonly used tomeasure moving and intermittent targets, targets in avacuum, and targets that inaccessible due to hostileenvironments, geometry limitations, or safety hazard.The cost is relatively higher, although in some cases it iscomparable to contact devises.

    °F/°C switchable

    Laser On/Off

    “9V” battery

    IRT670 Manual-FINAL3-050608:Layout 1 5/6/08 9:40 AM Page 5

  • MAINTENANCE:Cleaning the lens: Blow off loose particles using cleancompressed air. Gently brush remaining debris away with acamel’s hair brush. Carefully wipe the surface with a moistcotton swab. The swab may be moistened with water.

    NOTE:DO NOT use solvents to clean the lens.CLEANING THE HOUSING: Use soap and water on a damp

    sponge or soft cloth.

    EMISSIVITY TABLEMaterial Temp °F/°C EmissivityGold (pure highly polished) 227/440 0.02Aluminum Foil 27/81 0.04Aluminum Disc 27/81 0.18Aluminum Household (flat) 23/73 0.01Aluminum (polished prate 98.3%) 227/400 0.04

    577/1070 0.06Aluminum (rough plate) 26/78 0.06Aluminum (oxidized @ 599°C) 199/390 0.11

    599/1110 0.19Aluminum surfaced roofing 38/100 0.22Tin (bright tinned iron sheet) 25/77 0.04Nickel wire 187/368 0.1Lead (pure 99.95-unoxidized) 127/260 0.06Copper 199/390 0.18

    599/1110 0.19Steel 199/390 0.52

    599/1110 0.57Zinc galvanized sheet iron (bright) 28/82 0.23Brass (highly polished) 247/476 0.03Brass (hard rolled-polished w/lines) 21/70 0.04Iron galvanized (bright) - 0.13Iron plate (completely) 20/68 0.69Rolled sheet steel 21/71 0.66Oxidized iron 100/212 0.74Wrought iron 21/70 0.94Molten iron 1299-1399/

    3270-2550 0.29

    Copper(polished) 21-117/70-242 0.02Copper (scraped shiny not mirrored) 22/72 0.07Cooper (plate heavily oxidized) 25/77 0.78Enamel (white fused on iron) 19/66 0.9Formica 27/81 0.94Frozen oil - 0.93Brick (red-rough) 21/70 0.93Brick (silica-unglazed rough) 1000/1832 0.8Carbon (T-carbon 0.9% ash) 127/260 0.81Concrete - 0.94Glass (smooth) 22/72 0.94Granite (polished) 21/70 0.85Ice 0/32 0.97Marble (light gray polished) 22/72 0.93Asbestos board 23/74 0.96Asbestos paper 38/100 0.93

    371/700 0.95Asphalt (paving) 4/39 0.97

    IRT670 Manual-FINAL3-050608:Layout 1 5/6/08 9:40 AM Page 6

    99 Washington Street Melrose, MA 02176 Phone 781-665-1400Toll Free 1-800-517-8431

    Visit us at www.TestEquipmentDepot.com

    http://www.testequipmentdepot.com