lethabo power station units no. 1 to 6 particulate ......no. 1 to 6 at lethabo power station during...
TRANSCRIPT
-
Lethabo Power Station
Units No. 1 to 6 Particulate
Emission Monitor
Correlations,
Unit No. 4 Start-up &
Unit No. 6 Shutdown Report
1st July 2016
Report No. RSL222
-
P.O. Box 2459 Noordheuwel ext 4 Krugersdorp 1756 Vat No. 4670254459
10 Chisel Street Boltonia
Krugersdorp 1739
Tel: +27 (011) 660 9432 Tel: +27 (011) 954 6652 Fax:+27 (086) 689 4572
www.stacklabs.co.za
Eskom Holdings SOC Limited
Lethabo Power Station
Private Bag X415
Vereeniging
1930
Attention: Yaseen Waja
Tel: 0164575062
Cell: 0837864988
Date
Order No. Report No. Enquiries
1st July 2016 PO: 4501942164
RSL222 Tel: 011 954 6652
Cell: 082 458 7438
e-mail: [email protected]
REPORT No.: RSL222
LETHABO POWER STATION UNITS NO. 1 TO 6 PARTICULATE EMISSION MONITOR
CORRELATIONS, UNIT NO. 4 START-UP & UNIT NO. 6 SHUTDOWN REPORT
Sir
Herewith the finalised report for the particulate emission monitor correlations conducted on Units No. 1 to 6 at Lethabo Power Station during the period from December 2015 to July 2016, Unit No. 4
Start-up Measurements and Unit No. 6 Shutdown Measurements.
We thank you for the opportunity to be of service. We trust that your requirements were interpreted correctly. Should you however have any queries, please contact us at the above numbers, we will
gladly assist.
PH Pretorius
Stacklabs
ISO 9096, 12141 & 10155
STACKLABS ISO #:825268/:2007-05-23 © ISO 2003
-
II
Source Information
Source Location: Lethabo Power Station
Private Bag X415
Vereeniging 1930
South Africa
Permit No. TBA Contact: Yaseen Waja
Title: System Engineer
Telephone: (016) 457 5062
Listed Activity Category: 1.1 Solid Combustion Installation
Survey Date: December 2015 to July 2016
Pollutants: Total Particulate Matter (TPM) Negotiated Limit: 100 mg/Nm
3 dry 10% O2
Method: ISO 9096, 12141 & 10155
Plant: Unit No. 1
Survey Date: 9th to 18
th March 2016
Measured Concentration: 12.6 to 58.1 mg/Nm3 dry 10 % O2
Compliance: Yes – All Measurements No. 1 to 11
Plant: Unit No. 2
Survey Date: 19th to 23
rd December 2015
Measured Concentration: 11.6 to 91.2 mg/Nm3 dry 10 % O2
Compliance: Yes – Measurements No. 1 to 12
Measurement No. 9- plant malfunction
Plant: Unit No. 4
Survey Date: 13th to 17
th June 2016
Measured Concentration: 37.9 to 126.3 mg/Nm3 dry 10 % O2
Compliance: Yes – Measurements No. 3 to 12
No – Measurements No. 1 & 2 (618.37 MW SB)
Plant: Unit No. 5
Survey Date: 12th to 17
th December 2015
Measured Concentration: 65.8 to 144 mg/Nm3 dry 10 % O2
Compliance: Yes – Measurements No. 7, 8, 9, 10, 12, 14 & 15 No – Measurements No. 1 to 6, 11 & 13
Plant: Unit No. 6 Survey Date: 12
th to 7
th December 2015
Measured Concentration: 18.7 to 172.3 mg/Nm3 dry 10 % O2
Compliance: Yes – Measurements No. 7 to 12, 14 & 15 No – Measurements No. 1 to 6 & 13
Testing Laboratory
Laboratory: Stacklabs 10 Chisel Street Boltonia Krugersdorp 1739
Contact: Mr PH Pretorius Title: Managing Member
Telephone: (011) 954 6652
E Mail: [email protected]
-
III
Report No. RSL222 Reference
Revision
SLOP019
0
Title: Units No. 1 to 6 Particulate Emission Monitor
Correlations, Unit No. 4 Start-up & Unit No. 6 Shutdown
Date 2012/05/25
Revision Date No Revision
Compiled by Functional representative Authorised by
………………… …………………... …………………….. WJ Bronkhorst
Team Leader
P Buell
Quality Representative
PH Pretorius
Managing Member
Review and Certification
All work, activities and tasks performed for this report were carried out under my direction and
supervision. All work was conducted in compliance with listed standards
I have reviewed the details, calculations, results, conclusions and all written material contained within
this report, and hereby certify that the presented material is authentic and accurate.
Name: PH Pretorius
Title: Managing Member
2016/07/01
Sign:___________________________ Date:______________________
-
IV
Table of Contents
Section Page
Source Information............................................................................................................ II
Review and Certification...................................................................................................... III Report Summary.................................................................................................................. V
1. Monitoring Objective........................................................................................................... 1 2. Plant Description and General Operating Procedures.......................................................... 1 3. Methods and procedures....................................................................................................... 3 4. Results.................................................................................................................................. 5 5. Discussion............................................................................................................................ 17 6. Monitor Deviations............................................................................................................ 47 7. Recommendations.............................................................................................................. 47 8. Acknowledgements............................................................................................................. 47 9. References............................................................................................................................ 47 10. Distribution.......................................................................................................................... 47
Appendix No. 1 .................................................................Detailed Measurement Results 48
Appendix No. 2 .......................................................................................Plant parameters 67
Appendix No. 3 ...................................................................................Correlations Graphs 95 Appendix No. 4.................................................Water Vapour Concentration Calculation 112
Appendix No. 5............................................................................... Outlines of Procedures 114
Appendix No. 6............................... Particulate Emission Monitor Calibration Certificate 117
Appendix No. 7.................................................................................Particle Size Analysis 120 Appendix No. 8.................................................................................... Filter Mass Sheets 124
Appendix No. 9............................................................................... Calibration Certificate 130
-
V
REPORT SUMMARY
In order to meet the requirements of the South African Air Quality Act No. 39 of 2004
(Reference 9.1), all Solid Fuel Combustion Installations (Category 1.1) such as Lethabo Power Station, are obligated to monitor several listed pollutants during operation. These pollutants
include particulate matter, Sulphur Dioxide and Oxides of Nitrogen. The Act further specifies
that all particulate monitoring programs are required to meet international standards such as ISO 9096 & 12141 (Reference 9.2 & 9.3) and therewith ISO 10155 (Reference 9.4). To this
end and in keeping with the ESKOM Standard for Emission Monitoring and Reporting (420-
56242363) (Reference 9.5) Eskom’s scheduled particulate emission monitor correlations on
five Units at its Lethabo Power Station situated near Vereeniging in the Northern Free State. The risks areas included Units No. 1, 2, 4, 5 & 6’s electrostatic precipitators and the only
pollutant considered was particulate matter (particulate emission monitors correlations).
Continuous particulate emission monitors generally operate on a principle of passing a light
beam through a dust-laden gas stream. The dust concentration is then determined by measuring
the degree of light attenuation resulting from the dust within the gas stream. However the
degree of light attenuation is dependent on more than just the quantity of dust. Dust characteristics such as colour, size, shape and distribution of the dust within a duct, will all
impact on the amount of light that is attenuated. In addition, these characteristics will change
over time as the plant’s electrostatic precipitator or bag filters performances change. In order to compensate for these characteristic changes and to facilitate the determination of mass
emissions from the light attenuation, a correlation of light attenuation and measured mass
emissions must be conducted at regular intervals. The correlation of the light attenuation and mass emission requires the determination of the actual mass emission emitted from a particular
plant by means of isokinetic or Gravimetric dust sampling. This procedure must be conducted
over a range of plant operating conditions representative of the plant’s general operating
conditions and is referred to as a particulate emission monitor correlation or a dynamic calibration.
Stacklabs, an environmental testing laboratory, was contracted by Eskom to complete the required particulate emission monitor correlations on Units No. 1, 2, 4, 5 & 6. In addition to
the five particulate emissions monitor correlations, Stacklabs was also contracted to conduct
particulate emission measurements during a light up period on Unit No. 4 and a shutdown period on Unit No. 6.
All required site measurements for the five correlations, the light up and shutdown
measurements were carried out during the period from December 2015 to July 2016. The relevant results and correlation graphs have been presented in this report and may be
summarised as follows:
-
VI
Report Summary
Units No. 1, 2, 4, 5 & 6 Particulate Emission Monitor Correlations
Unit No. 1
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 1.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
March 2016
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 10.3 5.0 32.8
2 10.3 4.9 38.5
3 13.8 5.5 45.5
4 14.5 5.6 58.1
5 12.5 5.3 39.7
6 6.5 4.4 12.9
7 10.9 5.0 43.8
8 8.3 4.6 25.7
9 6.5 4.4 12.6
10 7.9 4.6 29.6
11 7.3 4.5 22.9
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
Unit No. 1 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 1
March 2016 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/04/26 5.1455 * mA – 18.5028 0.97 58.1 0 to 0.3
2 2016/04/26 34.3035 * mA – 135.1348 0.97 58.1 0 to 2.0
-
VII
Unit No. 1 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s)
as well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 611.5 719.3
2 621.9 725.3
3 619.2 697.8
4 622.5 702.0
5 622.5 705.9
6 532.6 567.3
7 521.4 566.0
8 525.4 585.6
9 427.5 472.3
10 440.4 478.8
11 438.0 476.2
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Unit No. 1 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 1
March 2016 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2014/04/26 1.1331 * Air Flow (kg/s)
– 5.9209 0.998 725.3 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
VIII
Unit No. 2
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 2
except where indicated.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
Dec 2015
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 18.9 6.2 83.3
2 >20.0* 6.6 77.7
3 14.5 5.6 63.0
4 18.4 6.2 79.0
5 >20.0* 6.6 89.8
6 16.3 5.8 67.3
7 19.3 6.3 79.7
8 >20.0* 6.7 91.2
10 6.4 4.4 15.5
11 6.3 4.3 13.7
12 7.7 4.6 15.2
13 6.0 4.3 11.6
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
*Measurements No. 5 & 8 were excluded from the Correlation on Output No. 1 due to signal
saturation. Measurement No. 9 was excluded from Output No. 1 & 2 Correlations due to burst filter.
Unit No. 2 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 2
December 2015 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/02/23 5.4767 * mA – 21.7274 0.99 83.3 0 to 0.3
2 2016/02/23 34.3566 * mA – 136.2854 0.99 91.2 0 to 2.0
-
IX
Unit No. 2 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s)
as well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 651.4 736.4
2 657.8 766.9
3 661.7 759.9
4 652.7 773.2
5 649.7 784.1
6 649.3 735.1
7 653.7 774.7
8 650.4 772.1
9 541.3 592.1
10 538.1 536.1
11 543.4 540.4
12 540.3 545.2
13 546.7 547.4
14 536.6 548.1
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 2
December 2015 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2016/02/23 1.1457 * Air Flow (kg/s)
– 17.4328 0.99 784.1 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
X
Unit No. 4
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 4.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
June 2016
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 29.0* 7.8 126.3
2 28.2* 7.6 110.9
3 18.7 6.2 67.1
4 21.1* 6.6 61.5
5 20.8* 6.5 90.1
6 18.4 6.2 58.4
7 20.9* 6.5 72.9
8 13.7 5.4 37.9
9 15.0 5.7 53.3
10 14.2 5.5 52.3
11 15.1 5.7 54.8
12 22.3* 6.7 96.8
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
*mA on Output 1 for Measurements No. 1, 2, 4, 5, 7 & 12 were extrapolated from Output 2’s results on request of customer.
Unit No. 4 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 4
June 2016 Correlations
Issued Date Function Correlation *MME Monitor
Output No.
yyyy/mm/dd mg/Nm
3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2 Range
Extinction
1 2016/07/01 4.4949 * mA – 17.4959 0.99 67.1 0 to 0.3
2 2016/07/01 31.7272 * mA – 128.2120 0.98 126.3 0 to 2.0
-
XI
Unit No. 4 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s)
as well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 616.5 747.9
2 618.8 766.4
3 618.7 804.8
4 601.0 799.6
5 605.9 817.8
6 608.2 822.6
7 608.8 783.6
8 514.7 652.0
9 532.5 640.0
10 452.7 494.2
11 435.1 492.1
12 434.2 491.7
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 4
June 2016 Correlation
Issued Date Function Correlation *MMGF Monitor
Output No.
yyyy/mm/dd Nm
3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2 Range
n/a 2016/07/01 1.2879 * Air Flow (kg/s)
– 16.5418 0.99 822.6 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
XII
Unit No. 5
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 5
except where indicated.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
Dec 2015
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 >20.0* 8.3 113.4
2 >20.0* 8.3 114.7
3 >20.0* 8.4 120.4
4 >20.0* 8.0 110.3
5 >20.0* 9.1 144.0
6 >20.0* 8.9 135.7
7 18.2 6.1 77.3
8 >20.0* 6.9 88.2
9 >20.0* 7.0 97.4
10 >20.0* 6.5 80.4
11 >20.0* 8.0 109.6
12 19.1 6.3 65.8
13 >20.0* 8.0 123.1
14 >20.0* 6.7 80.9
15 19.8 6.4 71.6
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155. *Measurements No. 1 to 6, 8 to 11, 13 & 14 were excluded from the Correlation on Output No. 1 due
to signal saturation.
Unit No. 5 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 5
Dec 2015 Correlations
Issued Date Function Correlation *MME Monitor
Output No.
yyyy/mm/dd mg/Nm
3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2 Range
Extinction
1 2016/02/23 4.7217 * mA – 18.6368 0.90 77.3 0 to 0.3
2 2016/02/23 27.1495 * mA – 103.0318 0.99 144 0 to 2.0
-
XIII
Unit No. 5 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s)
as well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 609.1 638.8
2 610.4 637.9
3 603.8 650.2
4 605.4 668.0
5 606.6 644.7
6 599.1 673.6
7 532.3 581.9
8 520.7 593.2
9 516.8 571.8
10 498.0 626.8
11 474.9 569.2
12 500.5 617.0
13 450.9 548.1
14 447.1 552.6
15 491.6 593.2
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 5
December 2015 Correlation
Issued Date Function Correlation *MMGF Monitor
Output No.
yyyy/mm/dd Nm
3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2 Range
n/a 2016/02/23 1.0984 * Air Flow (kg/s)
– 17.0016 0.99 673.6 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
XIV
Unit No. 6
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 6
except as indicated.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
Dec 2015
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 >20.0* 9.4 155.6
2 >20.0* 9.8 165.7
3 >20.0* 9.5 141.3
4 >20.0* 8.5 120.4
5 >20.0* 9.2 137.1
6 >20.0* 10.5 172.3
7 10.6 5.0 23.0
8 11.6 5.1 25.1
9 9.9 4.9 18.7
10 16.6 5.9 44.8
11 13.8 5.5 30.6
12 15.8 5.8 42.0
13 >20.0* 8.0 121.4
14 10.6 5.0 23.4
15 12.3 5.2 25.8
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155. *Measurements No. 1 to 6 & 13 were excluded from the Correlation on Output No. 1 due to signal
saturation.
Unit No. 6 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 6
December 2015 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/02/23 3.4385 * mA – 14.1750 0.99 44.8 0 to 0.3
2 2016/02/23 28.0913 * mA – 116.7179 0.99 172.3 0 to 2.0
-
XV
Unit No. 6 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s)
as well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 609.1 638.8
2 610.4 637.9
3 603.8 650.2
4 605.4 668.0
5 606.6 644.7
6 599.1 673.6
7 532.3 581.9
8 520.7 593.2
9 516.8 571.8
10 498.0 626.8
11 474.9 569.2
12 500.5 617.0
13 450.9 548.1
14 447.1 552.6
15 491.6 593.2
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 6
December 2015 Correlation
Issued Date Function Correlation *MMGF Monitor
Output No.
yyyy/mm/dd Nm
3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2 Range
n/a 2016/02/23 1.0984 * Air Flow (kg/s)
– 17.0016 0.99 673.6 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
XVI
Unit No. 4 Start-up Measurements
The following table presents the conditions as captured during the boiler Light up on Unit No.
4 on the 16th of March 2016. The table presents the boiler load in MW, particulate emission
monitor’s indications from Output No. 2 and the measured particulate emissions in mg/Nm3
(dry) @ 10% O2.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015
Correlation
Measured Dust
Mass
2015
Correlation
Measured Dust
Mass
Test No. MW % mA mg/Nm3 (dry)
@ 10% O2 mg/Nm3 (dry) @ Stack O2
1 0.0 12.2 >20.0* 828.1 638.0
2 0.0 11.7 >20.0* 3841.9 3242.9
3 370.0 7.9 >20.0* 2557.2 3043.5
Unit No. 6 Shutdown Measurement Results
The following table presents the conditions as captured during the boiler shutdown on Unit No. 6 on the 27
th & 28
th of June 2016. The table presents the boiler load in MW, stack gas
oxygen concentration in % v/v, particulate emission monitor’s indications from Output No. 2
and the measured particulate emissions in mg/Nm3 (dry) @ 10% O2 & mg/Nm
3 (dry) @ the
actual stack gas oxygen concentrations.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015
Correlation
Measured Dust
Mass
2015
Correlation
Measured Dust
Mass
Test No. MW % mA mg/Nm3 (dry)
@ 10% O2
mg/Nm3 (dry)
@ Stack O2
1 0.0 13.6 >20.0* 11353.1* 7608.3
2 0.0 20.6 >20.0* 173 559.6* 6311.3
The particulate emissions reported as mg/Nm3 Dry corrected to 10% O2 are not a true representation of
the concentrations due to the high stack gas oxygen concentrations.
Report Recommendations
It is recommended that:
The particulate emissions from Units No. 1 to 6 reported to the authorities, are according to the correlation function presented in Figures No. 1 & 2 of this report.
The next compliant correlation is conducted on Unit No 3 during the first quarter of 2016.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 1 of 131
1. Monitoring Objective
The primary objective of the particulate emission monitor correlations presented in this report
is to quantify the concentrations of all particulate matter released into the atmosphere during the
operations of Units No. 1, 2, 4, 5 & 6 at Lethabo Power Station.
The requirement for conducting the particulate emission monitor correlations is to demonstrate
compliance with the requirements of the National Environmental Management: Air Quality Act
(39/2004), Listed Activities and Associated Minimum Emission Standards in terms of Section 21, Category 1.1. The substance applicable under Category 1.1 is Particulate Matter which is
capped at 100 mg/Nm3 dry @ 10% O2. This cap limit however, in the case of Lethabo Power
Station, was negotiated to 100 mg/Nm3 dry @ 10% O2.
The objective of the particulate emission measurements surveys that were conducted during the
unit light up and shutdown periods on Units No. 4 & 6 respectively, was to quantify the concentrations of all particulate matter released into the atmosphere during the unit light up and
shutdown periods.
The requirement for conducting the particulate emission survey during the light up and shutdown periods has been set in the Stations Atmospheric Licence.
2. Plant Description & General Operating Procedure
2.1. The pulverised fuel steam generating boilers on which the particulate emission monitor correlations, light up and shutdown measurements were conducted, were originally designed
for a maximum steam flow of 600 MW. These units were originally equipped with four
parallel Brandt / Babcock electrostatic precipitators with a design capacity of 997 Am3/s gas
flow and a performance collecting efficiency of 99.88%. The electrostatic precipitators were later retro-fitted with sulphur trioxide flue gas conditioning and recently upgraded with high
frequency transformers on the first five fields.
Electrostatic Precipitator Design Data for Unit under Test (original Design)
Description @ 97 % MCR
Number of ESPs in use 4
Gas Volume Flow Rate Worst Coal 997 m3/s
Gas Temperature 132°C
Inlet Dust Burden 34.9 g/Nm3
Outlet Dust Burden n/a mg/Nm3
Carbon in Dust 2%
Sulphur in Coal n/a%
Pressure Drop 0.25Kpa
Temperature Drop 11°C
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 2 of 131
2.2. During the emission surveys, the particulate emission measurements were conducted through purpose built ports situated on the unit’s stacks at a level approximately 150m above ground
level. For a diagrammatic representation of the test positions, see the following sketch.
Measurements were conducted along sections Q—Q.
2.3. The operating regime followed during the survey periods included all significant activities that arise during normal production as well as some fault conditions. The correlation included measurements at the following Boiler Loads:
Measurement
No. Unit No. 1 Unit No. 2 Unit No. 4 Unit No. 5 Unit No. 6
Load (MW) Load (MW) Load (MW) Load (MW) Load (MW)
1 609.4 561.4 618.7 601.5 614.6
2 609.3 561.7 618.8 601.4 614.9
3 609.7 561.5 618.7 601.5 614.8
4 609.8 561.7 618.1 602.3 615.1
5 609.5 562.0 618.1 601.2 615.1
6 500.3 562.4 618.6 601.0 614.9
7 500.3 562.5 605.0 502.7 421.2
8 500.1 562.5 500.2 502.7 421.3
9 400.6 401.9 500.0 490.3 373.6
10 400.5 401.9 375.1 494.7 491.8
11 400.7 401.8 375.2 452.0 491.9
12 - 398.7 375.1 495.4 491.9
13 - 398.6 - 450.3 448.0
14 - 398.5 - 450.4 448.0
15 - - - 487.0 485.1
S/S 1 - - 0.0 - 330.0 to 0.0
S/S 2 - - 0.0 - 0
S/S 3 - - 0 to 378.0 - -
S/S Start-up & Shutdown Measurements
For additional information of the plant parameters please see Appendix No. 2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 3 of 131
3. Methods & Procedures The method selected for the determination of the particulate emission concentration on Units
No. 1, 2, 4, 5 & 6 was in compliance with ISO 12141 Stationary source emissions -
Determination of concentration of particulate matter (dust) at low concentrations – Manual gravimetric method.
This general method can be briefly described as follows:
A sharp-edged nozzle is positioned in the duct facing into the moving gas stream and a sample
flow of the gas is extracted isokinetically for a measured period of time. To allow for non-
uniformity of the distribution of particulate concentration in the duct, samples are taken at a pre-selected number of stated positions in the duct cross-section. The particulate matter
entrapped in the gas sample is separated by a filter medium, then dried and weighed. The
particulate concentration is calculated from the weighed particulate mass and the gas sample volume. The particulate mass flow rate is calculated from the particulate concentration and
the duct gas volumetric flow rate. The particulate mass flow rate can also be calculated from
the weighed particulate matter, the sample time, the area of the sample plane and the nozzle
opening.
The degree to which this sample represents the total gas flow depends on
Homogeneity of the gas velocity within the sampling plane;
A sufficient number of sampling points in the sampling plane;
Isokinetic withdrawal of the plane
As stated above, the gas has to be sampled at more than one sampling point in the sampling
plane, dependant on the sampling plane area. This plane is normally divided into equal areas, at the centres of which gas is withdrawn. To determine the particle concentration in the plane,
the nozzle is moved from one sampling point to the other, extracting gas isokinetically at each
point. Sampling periods should be equal for each sampling point, resulting in a composite
sample. If equal sampling areas cannot be chosen, the sampling period shall be proportional to the sampling area. The sample is extracted through a sampling train, which principally
consists of:
A sampling probe tube with entry nozzle;
A particle separator;
A gas metering system, in-stack or external;
A suction system
The particle separator and / or the gas metering system may be located either in the duct, or
placed outside the duct.
It is necessary to avoid condensation of vapour in the sampling train during gas sampling,
because it will interfere with particle separation, particulate condition and flow measurements.
To this end, the probe tube, the particle separator and the gas flow-measuring device may be heated above the relevant dew point where necessary. The water vapour may intentionally be
removed downstream of the particle separator, to make use of a dry gas meter for the
measurement of sampled gas volume.
For isokinetic sampling, the gas velocity at the sampling point in the duct has to be measured,
and the corresponding sample gas flow has to be calculated and adjusted.
Normally a Pitot static tube is used for the measurement of duct gas velocity. If the sample gas
flow-measuring device is used within the duct, the relationship between the measured pressure
drop and the pitot static tube differential pressure is simple, facilitating the adjustment to
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 4 of 131
isokinetic conditions. If the gas meter device is situated outside the duct, the calculation of the
isokinetic sample gas flow rate is more complicated. The calculation may also include the duct gas density under standard conditions (which may be derived from the dry gas composition and
the moisture content), the temperature, static pressure of the gas in the duct, the gas meter
device, and the water vapour content of the duct gas, if the sample gas flow is measured after water removal.
After sampling, the collected particulate matter is completely recovered, dried and weighed and
the concentrations are determined.
All Stacklabs isokinetic sampling is carried out employing procedures and equipment that
comply with the requirements of ISO 9096 12141 1992 (Reference 1). All Stacklabs sampling equipment is calibrated by SANAS accredited laboratories.
Lethabo personnel were responsible for the setting of the plant prior to the test period. Stacklabs was contracted to provide the service of particulate emission monitors correlation
through isokinetic dust sampling only.
The broad outlines of filter weighing, pre-test preparations, sampling system integrity checks and sampling procedures are discussed in Appendix No 3.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 5 of 131
4. Results 4.1. Unit No. 1
The following table presents the results of the particulate emission monitor’s indications from Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm
3 (dry) @ 10% O2.
The measurements listed below were all included in the final particulate emission monitor
correlations for Unit No. 1.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
March 2016
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 10.3 5.0 32.8
2 10.3 4.9 38.5
3 13.8 5.5 45.5
4 14.5 5.6 58.1
5 12.5 5.3 39.7
6 6.5 4.4 12.9
7 10.9 5.0 43.8
8 8.3 4.6 25.7
9 6.5 4.4 12.6
10 7.9 4.6 29.6
11 7.3 4.5 22.9
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
Unit No. 1 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 1
March 2016 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2014/04/26 1.1331 * Air Flow (kg/s)
– 5.9209 0.998 725.3 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 6 of 131
Unit No. 1 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s) as
well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 611.5 719.3
2 621.9 725.3
3 619.2 697.8
4 622.5 702.0
5 622.5 705.9
6 532.6 567.3
7 521.4 566.0
8 525.4 585.6
9 427.5 472.3
10 440.4 478.8
11 438.0 476.2
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 1
March 2016 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2014/04/26 1.1331 * Air Flow (kg/s)
– 5.9209 0.998 725.3 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 7 of 131
4.2. Unit No. 2
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 2 except where indicated.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
Dec 2015
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 18.9 6.2 83.3
2 >20.0* 6.6 77.7
3 14.5 5.6 63.0
4 18.4 6.2 79.0
5 >20.0* 6.6 89.8
6 16.3 5.8 67.3
7 19.3 6.3 79.7
8 >20.0* 6.7 91.2
10 6.4 4.4 15.5
11 6.3 4.3 13.7
12 7.7 4.6 15.2
13 6.0 4.3 11.6
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
*Measurements No. 5 & 8 were excluded from the Correlation on Output No. 1 due to signal
saturation. Measurement No. 9 was excluded from Output No. 1 & 2 Correlations due to burst filter.
Unit No. 2 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 2
December 2015 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/02/23 5.4767 * mA – 21.7274 0.99 83.3 0 to 0.3
2 2016/02/23 34.3566 * mA – 136.2854 0.99 91.2 0 to 2.0
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 8 of 131
Unit No. 2 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s) as
well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 651.4 736.4
2 657.8 766.9
3 661.7 759.9
4 652.7 773.2
5 649.7 784.1
6 649.3 735.1
7 653.7 774.7
8 650.4 772.1
9 541.3 592.1
10 538.1 536.1
11 543.4 540.4
12 540.3 545.2
13 546.7 547.4
14 536.6 548.1
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 2
December 2015 Correlation
Issued Date Function Correlation *MMGF Monitor
Output No.
yyyy/mm/dd Nm
3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2 Range
n/a 2016/02/23 1.1457 * Air Flow (kg/s)
– 17.4328 0.99 784.1 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 9 of 131
4.3. Unit No. 4
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 4.
Monitor Output
(0 to 0.3 Ext)
Output No. 1
Monitor Output
(0 to 2 Ext)
Output No.2
June 2016
Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 29.0* 7.8 126.3
2 28.2* 7.6 110.9
3 18.7 6.2 67.1
4 21.1* 6.6 61.5
5 20.8* 6.5 90.1
6 18.4 6.2 58.4
7 20.9* 6.5 72.9
8 13.7 5.4 37.9
9 15.0 5.7 53.3
10 14.2 5.5 52.3
11 15.1 5.7 54.8
12 22.3* 6.7 96.8
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
*mA on Output 1 for Measurements No. 1, 2, 4, 5, 7 & 12 were extrapolated from Output 2’s results on
request of customer.
Unit No. 4 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 4
June 2016 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/07/01 4.4949 * mA – 17.4959 0.99 67.1 0 to 0.3
2 2016/07/01 31.7272 * mA – 128.2120 0.98 126.3 0 to 2.0
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 10 of 131
Unit No. 4 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (kg/s) as
well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 616.5 747.9
2 618.8 766.4
3 618.7 804.8
4 601.0 799.6
5 605.9 817.8
6 608.2 822.6
7 608.8 783.6
8 514.7 652.0
9 532.5 640.0
10 452.7 494.2
11 435.1 492.1
12 434.2 491.7
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 4
June 2016 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2016/07/01 1.2879 * Air Flow (kg/s)
– 16.5418 0.99 822.6 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 11 of 131
4.4. Unit No. 5
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 5 except
where indicated.
Monitor Output (0 to 0.3 Ext)
Output No. 1
Monitor Output (0 to 2 Ext)
Output No.2
Dec 2015 Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 >20.0* 8.3 113.4
2 >20.0* 8.3 114.7
3 >20.0* 8.4 120.4
4 >20.0* 8.0 110.3
5 >20.0* 9.1 144.0
6 >20.0* 8.9 135.7
7 18.2 6.1 77.3
8 >20.0* 6.9 88.2
9 >20.0* 7.0 97.4
10 >20.0* 6.5 80.4
11 >20.0* 8.0 109.6
12 19.1 6.3 65.8
13 >20.0* 8.0 123.1
14 >20.0* 6.7 80.9
15 19.8 6.4 71.6
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
*Measurements No. 1 to 6, 8 to 11, 13 & 14 were excluded from the Correlation on Output No. 1 due
to signal saturation.
Unit No. 5 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 5
Dec 2015 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/02/23 4.7217 * mA – 18.6368 0.90 77.3 0 to 0.3
2 2016/02/23 27.1495 * mA – 103.0318 0.99 144 0 to 2.0
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 12 of 131
Unit No. 5 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (k/s) as
well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 609.1 638.8
2 610.4 637.9
3 603.8 650.2
4 605.4 668.0
5 606.6 644.7
6 599.1 673.6
7 532.3 581.9
8 520.7 593.2
9 516.8 571.8
10 498.0 626.8
11 474.9 569.2
12 500.5 617.0
13 450.9 548.1
14 447.1 552.6
15 491.6 593.2
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 5
December 2015 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2016/02/23 1.0984 * Air Flow (kg/s)
– 17.0016 0.99 673.6 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 13 of 131
4.5. Unit No. 6
The following table presents the results of the particulate emission monitor’s indications from
Outputs No. 1 & 2 as well as the measured particulate emissions in mg/Nm3 (dry) @ 10% O2.
The measurements listed below were all included in the final correlations for Unit No. 6 except
as indicated.
Monitor Output (0 to 0.3 Ext)
Output No. 1
Monitor Output (0 to 2 Ext)
Output No.2
Dec 2015 Correlation
Measured Dust Mass
Test No. mA mA mg/Nm3 (dry) @ 10% O2
1 >20.0* 9.4 155.6
2 >20.0* 9.8 165.7
3 >20.0* 9.5 141.3
4 >20.0* 8.5 120.4
5 >20.0* 9.2 137.1
6 >20.0* 10.5 172.3
7 10.6 5.0 23.0
8 11.6 5.1 25.1
9 9.9 4.9 18.7
10 16.6 5.9 44.8
11 13.8 5.5 30.6
12 15.8 5.8 42.0
13 >20.0* 8.0 121.4
14 10.6 5.0 23.4
15 12.3 5.2 25.8
All particulate emissions reported as mg/Nm3 Dry corrected to 10% O2. The final correlation includes three zero points as described in ISO 10155.
*Measurements No. 1 to 6 & 13 were excluded from the Correlation on Output No. 1 due to signal
saturation.
Unit No. 6 Particulate Emission Monitor Correlation Equations
Lethabo Power Station Unit No. 6
December 2015 Correlations
Issued Date Function Correlation *MME Monitor
Output
No. yyyy/mm/dd
mg/Nm3(d) @ 10% O2
= m * mA + c Coefficient
mg/Nm3(d)
@ 10% O2
Range
Extinction
1 2016/02/23 3.4385 * mA – 14.1750 0.99 44.8 0 to 0.3
2 2016/02/23 28.0913 * mA – 116.7179 0.99 172.3 0 to 2.0
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 14 of 131
Unit No. 6 Air flow to gas flow correlation
The following table presents the results of the plant recorded total Air Flow indication (k/s) as
well as the measured Gas Flow results (Nm3 (dry) @ 10% O2).
Recorded Total Air Flow
Measured Gas Flow
Test No. kg/s Nm3 (dry) @ 10% O2
1 609.1 638.8
2 610.4 637.9
3 603.8 650.2
4 605.4 668.0
5 606.6 644.7
6 599.1 673.6
7 532.3 581.9
8 520.7 593.2
9 516.8 571.8
10 498.0 626.8
11 474.9 569.2
12 500.5 617.0
13 450.9 548.1
14 447.1 552.6
15 491.6 593.2
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
Lethabo Power Station Unit No. 6
December 2015 Correlation
Issued Date Function Correlation *MMGF Monitor
Output
No. yyyy/mm/dd
Nm3/s (d) @ 10% O2
= m * kg/s Air Flow + c Coefficient
Nm3/s (d)
@ 10% O2
Range
n/a 2016/02/23 1.0984 * Air Flow (kg/s)
– 17.0016 0.99 673.6 n/a
Gas Flow reported as Nm3/ s Dry corrected to 10% O2
*MMGF: Maximum Measured Gas Flow during correlation period
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 15 of 131
4.6. Unit No. 4 Light up Measurement Results
The following table presents the conditions as captured during the boiler Light up on Unit No. 4
on the 16th
of March 2016. The table presents the boiler load in MW, particulate emission monitor’s indications from Output No. 2 and the measured particulate emissions in mg/Nm
3
(dry) @ 10% O2.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015 Correlation
Measured Dust
Mass
2015 Correlation
Measured Dust
Mass
Test No. MW % mA mg/Nm3 (dry)
@ 10% O2
mg/Nm3 (dry)
@ Stack O2
1 0.0 12.2 >20.0* 828.1 638.0
2 0.0 11.7 >20.0* 3841.9 3242.9
3 370.0 7.9 >20.0* 2557.2 3043.5
4.7. Unit No. 6 Shutdown Measurement Results
The following table presents the conditions as captured during the boiler shutdown on Unit No.
6 on the 27th & 28
th of June 2016. The table presents the boiler load in MW, stack gas oxygen
concentration in % v/v, particulate emission monitor’s indications from Output No. 2 and the measured particulate emissions in mg/Nm
3 (dry) @ 10% O2 & mg/Nm
3 (dry) @ the actual stack
gas oxygen concentrations.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015 Correlation
Measured Dust
Mass
2015 Correlation
Measured Dust
Mass
Test No. MW % mA mg/Nm3 (dry)
@ 10% O2
mg/Nm3 (dry)
@ Stack O2
1 0.0 13.6 >20.0* 11353.1* 7608.3
2 0.0 20.6 >20.0* 173 559.6* 6311.3
The particulate emissions reported as mg/Nm3 Dry corrected to 10% O2 are not a true representation of
the concentrations due to the high stack gas oxygen concentrations.
The detailed results of the correlation measurements for Units No 1, 2, 4, 5 & 6 have been
presented in the following Tables:
Appendix No. 1 Detailed Measurement Results
Appendix No. 2 Plant parameters
Appendix No. 3 Correlations Graphs
Appendix No. 7 Particle Size Analysis Appendix No. 8 Filter Mass Sheets
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 16 of 131
The following abbreviations were used in the text, tables and figures:
MCR Maximum Continuous Rating °C Degrees Celsius
% v/v Percentage on a Volume-by-Volume basis
Am3 Actual Cubic Metres
Nm3 Normal Cubic Metres
g/s Grams per second
mg/s Milligrams per second
Fo Fields out ESP Electrostatic Precipitator
FFP Fabric Filter Plant
ext. Extinction mg/Nm
3 (d) @ 10% O2 Milligrams per Normal cubic meters dry corrected to 10% Oxygen
• ‘Actual’ refers to the measured temperature and pressure conditions of the gases in the duct
• ‘Normal’ refers to the actual conditions being normalised to 0 °C and 101,325 kPa.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 17 of 131
5. Discussion
5.1. Unit No. 1 Particulate Emission Correlation
The correlation on Unit No 1 was completed form the 9th to the 19
th of March 2016. A total of
eleven measurements were conducted of which all eleven measurements were included in the
final particulate emission monitor correlations. The following trends indicate the particulate
emission monitor signals captured during Measurement No. 1 on the 9th of March with 609
MW boiler loading and the Measurement No. 10 on the 19th
of March with 400 MW boiler loading.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 18 of 131
The outputs settings of all the particulate emission monitors included in this report have been
set for 0 to 0.03 extinction on Output No. 1 and 0 to 2.0 extinction on Output No. 2.
The measured particulate emission concentrations on Unit No. 1 ranged from 12.6 to 58.1 mg/Nm
3 Dry @ 10% O2, well below the negotiated cap value of 100 mg/Nm
3 Dry @ 10% O2.
The corresponding particulate emission monitor signals, recorded from Output No. 1, ranged
from 6.5 to 14.5 mA. The resulting co-ordinates produced the following correlation with a correlation coefficient of 0.974, which falls within the standards tolerance of 0.95 to 1.05.
4
6
8
10
12
14
16
18
20
Mo
nit
or
Ou
tpu
t (m
A)
Lethabo Unit No. 1 TPM Monitor Output Data (mA) Measurement No. 10 Output No. 2
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 19 of 131
A correlation derived for Output No. 2 based on the monitor settings of 0 to 2.0 extinction is
presented in the following Figure. This correlation also complies with the requirements of the ISO 10155, and includes the cap limit of 100 mg/Nm
3 Dry @ 10% O2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 20 of 131
5.2. Unit No. 1 Airflow to Gas Flow Correlation In addition to the measurement of the particulate emissions, gas volume flow rates were also
determined during this correlation period. The measured gas volume flow rates, which ranged
from 472.3 to 725.3 Nm3/s Dry @ 10% O2, were correlated against the plants corresponding
Total Air flow Rates to produce the following correlation with a coefficient of 0.998.
The following trends indicate Boiler load and Total Air Flow captured during Measurement No. 1 and demonstrates the typical load and air flow stability experienced during the correlation
period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 21 of 131
Plant parameters and trends for each measurement have been presented in Appendix No. 3.
The corresponding stack gas velocity data captured during Measurement No. 1 produced the
following profile.
The minimum, maximum and average gas velocity for this measurement was recorded as 27.8,
34.6 and 31.8 m/s respectively. The minimum and maximum deviation from the Average
velocity is therefore 12.7 & 8.8 % respectively.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 22 of 131
5.3. Unit No. 2 Particulate Emission Correlation
The correlation on Unit No 2 was completed from the 19th to the 23
rd of December 2015. A
total of thirteen measurements were conducted of which only eight measurements were included in the final particulate emission monitor correlation for Output No. 1 and twelve for
the correlation for Output No. 2. Measurements No. 2, 5, 8 & 9 were excluded from the
correlation for Output No. 1 as a result of monitor signal saturation on the first output.
Measurement No. 9 was excluded from the correlations for Outputs No. 1 & 2 as a result of a burst filter.
The following trends indicate the particulate emission monitor signals captured during Measurement No. 5 on the 20
th of December with 562 MW boiler loading. The trends show the
saturated signal on Output No. 1 and the normal trend from Output No. 2.
The outputs settings have been set to 0 to 0.03 extinction on Output No. 1 and 0 to 2.0
extinction on Output No. 2.
4
6
8
10
12
14
16
18
20
Par
ticu
late
Em
issi
on
(mA
)
Lethabo Power Station Units No. 2 Measurement No. 5 Stack Emission Dust Emission mA Output 1
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 23 of 131
The measured particulate emission concentrations on Unit No. 2, included in the correlation for
Output No. 1, ranged from 11.6 to 83.3 mg/Nm3 Dry @ 10% O2 and for output No. 2 ranged
from 11.6 to 91.2 mg/Nm3 Dry @ 10% O2. The maximum measured emission of 91.2 mg/Nm
3
Dry @ 10% O2 was below the negotiated cap value of 100 mg/Nm3 Dry @ 10% O2. The
corresponding particulate emission monitor signals, recorded from Output No. 1, ranged from 6.0 to 19.3 mA. The resulting co-ordinates produced the following correlation with a
correlation coefficient of 0.997, which falls within the standards tolerance of 0.95 to 1.05.
A correlation derived for Output No. 2 based on the monitor settings of 0 to 2.0 extinction is
presented in the following Figure. This correlation also complies with the requirements of the ISO 10155, and includes the cap limit of 100 mg/Nm
3 Dry @ 10% O2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 24 of 131
5.4. Unit No. 2 Airflow to Gas Flow Correlation
In addition to the measurement of the particulate emissions, gas volume flow rates were also
determined during this correlation period. The measured gas volume flow rates, which ranged from 536.1 to 784.1 Nm
3/s Dry @ 10% O2, were correlated against the plants corresponding
Total Air flow Rates to produce the following correlation with a coefficient of 0.989.
The following trends indicate Boiler load and Total Air Flow captured during Measurement No.
1 and demonstrates the typical load and air flow stability experienced during the correlation
period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 25 of 131
Plant parameters and trends for each measurement have been presented in Appendix No. 3.
The corresponding stack gas velocity data captured during Measurement No. 1 produced the
following profile.
The minimum, maximum and average gas velocity for this measurement was recorded as 29.0
33.6 and 31.8 m/s respectively. The minimum and maximum deviation from the Average
velocity is therefore 8.7 & 5.6 % respectively.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 26 of 131
5.5. Unit No. 4 Particulate Emission Correlation
The correlation on Unit No 4 was completed form the 13th to the 17
th of June 2016. A total of
twelve measurements were conducted of which six measurements were included in the final particulate emission monitor correlation for Output No. 1 and twelve for the correlation for
Output No. 2. Measurements No. 1, 2, 4, 5, 7 &12 were excluded from the correlation for
Output No. 1 as a result of monitor signal saturation on the first output.
The following trends indicate the particulate emission monitor signals captured during
Measurement No. 7 on the 15th of June 2016 with 619 MW boiler loading. The trends show the
saturated signal on Output No. 1 and the normal trend from Output No. 2.
The outputs settings have been set to 0 to 0.03 extinction on Output No. 1 and 0 to 2.0
extinction on Output No. 2.
The measured particulate emission concentrations on Unit No. 4, included in the correlation for
Output No. 1, ranged from 37.9 to 67.1 mg/Nm3 Dry @ 10% O2 and for output No. 2 ranged
from 37.9 to 126.3 mg/Nm3 Dry @ 10% O2. The maximum measured emission of 126.3
mg/Nm3 Dry @ 10% O2 was above the negotiated cap value of 100 mg/Nm
3 Dry @ 10% O2.
The corresponding particulate emission monitor signals, recorded from Output No. 1, ranged
from 13.7 to 18.7 mA. The resulting co-ordinates produced the following correlation with a
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 27 of 131
correlation coefficient of 0.988, which falls within the standards tolerance of 0.95 to 1.05.
A correlation derived for Output No. 2 based on the monitor settings of 0 to 2.0 extinction is presented in the following Figure. This correlation also complies with the requirements of the
ISO 10155, and includes the cap limit of 100 mg/Nm3 Dry @ 10% O2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 28 of 131
5.6. Unit No. 4 Airflow to Gas Flow Correlation
In addition to the measurement of the particulate emissions, gas volume flow rates were also
determined during this correlation period. The measured gas volume flow rates, which ranged from 491.7 to 822.6 Nm
3/s Dry @ 10% O2, were correlated against the plants corresponding
Total Air flow Rates to produce the following correlation with a coefficient of 0.991.
The following trends indicate Boiler load and Total Air Flow captured during Measurement No.
6 and demonstrates the typical load and air flow stability experienced during the correlation
period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 29 of 131
Plant parameters and trends for each measurement have been presented in Appendix No. 3.
The corresponding stack gas velocity data captured during Measurement No. 1 produced the
following profile.
The minimum, maximum and average gas velocity for this measurement was recorded as 25.9,
34.8 & 31.8 m/s respectively. The minimum and maximum deviation from the Average
velocity is therefore 18.6 & 9.7 % respectively.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 30 of 131
5.7. Unit No. 4 Particulate Matter Size Distribution In addition to the correlations for the particulate emissions and the gas flows, a set of extended
measurements were conducted on Unit No. 4 during the 23rd
and 24th
of March of 2016. The
purpose of these extended measurements was to provide sufficient particulate matter, from the stack gas, for a particle size analysis. The collated particulates from all the measurements
conducted on the 23rd
and 24th of March were combined into one sample. The following figure
and tables, take from the lab report, presents the particle size distribution determined from this
sample.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 31 of 131
The results show that approximately 20% of the particles fall below 1 µm, 63% fall below 2.5
µm and 95% fall below 10µm. The remaining 5% of particles fall between 10 and 100 µm.
The full lab sheet has been presented in Appendix No. 8.
5.8. Unit No. 5 Particulate Emission Correlation
The correlation on Unit No 5 was completed form the 12th
to the 17th of December 2015. A
total of fifteen measurements were conducted of which only three measurements were included
in the final particulate emission monitor correlation for Output No. 1. All fifteen measurements were however included in the correlation for Output No. 2. Only data from Measurements No.
7, 12 & 15 could be included in the correlation for Output No. 1 as these were the only three
measurements where the monitor signal did not saturate the first output.
The following trends indicate the particulate emission monitor signals captured during
Measurement No. 1 on the 12th of December 2015 with 602 MW boiler loading. The trends
show the saturated signal on Output No. 1 and the normal trend from Output No. 2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 32 of 131
The outputs settings have been set to 0 to 0.03 extinction on Output No. 1 and 0 to 2.0 extinction on Output No. 2.
The measured particulate emission concentrations on Unit No. 5, included in the correlation for
Output No. 1, ranged from 65.8 to 77.3 mg/Nm3 Dry @ 10% O2 and for output No. 2 ranged
from 65.8 to 144.0 mg/Nm3 Dry @ 10% O2. The maximum measured emission of 144.0
mg/Nm3 Dry @ 10% O2 was above the negotiated cap value of 100 mg/Nm
3 Dry @ 10% O2.
The corresponding particulate emission monitor signals, recorded from Output No. 1, ranged
from 18.2 to 19.8 mA. The resulting co-ordinates produced the following correlation with a
correlation coefficient of 0.988, which falls within the standards tolerance of 0.95 to 1.05.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 33 of 131
A correlation derived for Output No. 2 based on the monitor settings of 0 to 2.0 extinction is
presented in the following Figure. This correlation also complies with the requirements of the
ISO 10155, and includes the cap limit of 100 mg/Nm3 Dry @ 10% O2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 34 of 131
5.9. Unit No. 5 Airflow to Gas Flow Correlation
In addition to the measurement of the particulate emissions, gas volume flow rates were also
determined during this correlation period. The measured gas volume flow rates, which ranged from 548.1 to 673.6 Nm
3/s Dry @ 10% O2, were correlated against the plants corresponding
Total Air flow Rates to produce the following correlation with a coefficient of 0.989.
The following trends indicate Boiler load and Total Air Flow captured during Measurement No.
1 and demonstrates the typical load and air flow stability experienced during the correlation
period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 35 of 131
Plant parameters and trends for each measurement have been presented in Appendix No. 3.
The corresponding stack gas velocity data captured during Measurement No. 5 produced the following profile.
The minimum, maximum and average gas velocity for this measurement was recorded as 22.3, 28.6 & 25.6 m/s respectively. The minimum and maximum deviation from the Average
velocity is therefore 12.8 & 11.6 % respectively.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 36 of 131
5.10. Unit No. 6 Particulate Emission Correlation
The correlation on Unit No 6 was completed form the 12th
to the 17th of December 2015. A
total of fifteen measurements were conducted of which eight measurements were included in
the final particulate emission monitor correlation for Output No. 1. All fifteen measurements were however included in the correlation for Output No. 2. Measurements No. 1 to 6 & 13
were excluded from the correlation for Output No. 1 as a result of monitor signal saturation on
the first output.
The following trends indicate the particulate emission monitor signals captured during
Measurement No. 2 on the 12th of December 2015 with 602 MW boiler loading. The trends
show the saturated signal on Output No. 1 and the normal trend from Output No. 2.
The outputs settings have been set to 0 to 0.03 extinction on Output No. 1 and 0 to 2.0
extinction on Output No. 2.
The measured particulate emission concentrations on Unit No.6, included in the correlation for
Output No. 1, ranged from 18.7 to 44.8 mg/Nm3 Dry @ 10% O2 and for output No. 2 ranged
from 18.7 to 172.3 mg/Nm3 Dry @ 10% O2. The maximum measured emission of 172.3
mg/Nm3 Dry @ 10% O2 was above the negotiated cap value of 100 mg/Nm
3 Dry @ 10% O2.
The corresponding particulate emission monitor signals, recorded from Output No. 1, ranged
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 37 of 131
from 9.9 to 16.6 mA. The resulting co-ordinates produced the following correlation with a
correlation coefficient of 0.995, which falls within the standards tolerance of 0.95 to 1.05.
A correlation derived for Output No. 2 based on the monitor settings of 0 to 2.0 extinction is
presented in the following Figure. This correlation also complies with the requirements of the ISO 10155, and includes the cap limit of 100 mg/Nm
3 Dry @ 10% O2.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 38 of 131
5.11. Unit No. 6 Airflow to Gas Flow Correlation
In addition to the measurement of the particulate emissions, gas volume flow rates were also
determined during this correlation period. The measured gas volume flow rates, which ranged from 539.3 to 805.8 Nm
3/s Dry @ 10% O2, were correlated against the plants corresponding
Total Air flow Rates to produce the following correlation with a coefficient of 0.981.
The following trends indicate Boiler load and Total Air Flow captured during Measurement No.
2 and demonstrates the typical load and air flow stability experienced during the correlation
period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 39 of 131
Plant parameters and trends for each measurement have been presented in Appendix No. 3.
The corresponding stack gas velocity data captured during Measurement No. 2 produced the
following profile.
The minimum, maximum and average gas velocity for this measurement was recorded as 27.4,
37.7 & 33.8 m/s respectively. The minimum and maximum deviation from the Average
velocity is therefore 19.0 & 11.5 % respectively.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 40 of 131
5.12. Unit No. 4 Start Up or Light Up Measurements
On the 16th of March 2016 a set of three measurements were conducted on Unit No. 4 during a
Unit Light Up/Start Up. The first two measurements were conducted prior to synchronisation
and the third during the load ramping from 130 to 350 MW. The following trends and tables show the Boiler Load, Ait flow & Mill PF flow variation during the measurement period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 41 of 131
Measurement No. 1
The following table presents the conditions as captured during Measurement No. 1.
Boiler Load
Stack Oxygen
Monitor Output
(0 to 2 Ext) Output No.2
2015
Correlation Measured Dust Mass
Test No. MW % mA mg/Nm3 (dry) @
Stack O2
1 0.0 12.2 >20.0 638.0
LoadMill A PF
flow
Mill B PF
flow
Mill C PF
flow
Mill D PF
flow
Mill E PF
flow
Mill F PF
flow
Total air
flow
Total air
flow
MW
US-
04NM10U
603-C153
US-
04NM20U
603-C157
US-
04NM30U
603-C161
US-
04NM40U
603-C165
US-
04NM50U
603-C169
US-
04NM60U
603-C173
US-
04NG00F6
00-C90
US-
04NG00F6
00-C90
Average 0.0 4.6 1.0 -0.4 3.4 -2.6 -0.2 272.9 272.9
Max 0.0 5.5 10.4 3.8 3.6 11.9 3.4 537.1 537.1
Min 0.0 3.2 -11.2 -34.3 1.2 -34.3 -30.8 118.8 118.8
11:45:00 0.0 4.8 -0.6 -34.3 1.2 -0.1 -30.8 340.8 340.8
11:50:00 0.0 3.3 3.8 -34.3 3.5 -5.0 -30.4 366.8 366.8
11:55:00 0.0 3.2 3.8 3.6 3.6 3.2 3.4 537.1 537.1
12:00:00 0.0 3.2 3.8 3.6 3.5 3.2 3.4 209.8 209.8
12:05:00 0.0 3.2 3.8 3.8 3.6 3.2 3.4 280.3 280.3
12:10:00 0.0 3.2 3.8 3.6 3.6 3.2 3.4 227.9 227.9
12:15:00 0.0 3.2 3.8 3.6 3.5 3.2 3.4 226.2 226.2
12:20:00 0.0 3.2 3.8 3.6 3.6 3.2 3.4 118.8 118.8
12:25:00 0.0 5.5 -10.6 3.6 3.6 -34.3 3.4 167.2 167.2
12:30:00 0.0 5.5 -11.2 3.6 3.5 -34.3 3.4 174.8 174.8
12:35:00 0.0 5.4 -7.2 3.6 3.5 -14.2 3.4 245.5 245.5
12:40:00 0.0 5.4 10.4 3.6 3.6 11.9 3.4 303.9 303.9
12:45:00 0.0 5.3 3.9 3.6 3.6 3.9 3.4 288.3 288.3
12:50:00 0.0 5.4 -7.0 3.6 3.6 -7.3 3.4 272.3 272.3
12:55:00 0.0 5.3 5.3 3.6 3.5 1.3 3.4 289.3 289.3
13:00:00 0.0 5.3 5.2 3.6 3.5 9.8 3.4 295.5 295.5
13:05:00 0.0 5.4 -2.8 3.6 3.5 -1.5 3.4 275.8 275.8
13:10:00 0.0 5.5 -0.3 3.6 3.5 -1.9 3.4 274.8 274.8
13:15:00 0.0 5.4 8.0 3.6 3.5 3.3 3.4 290.8 290.8
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 42 of 131
Measurement No. 2
The following table presents the conditions as captured during Measurement No. 2.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015
Correlation
Measured Dust Mass
Test No. MW % mA mg/Nm3 (dry) @
Stack O2
2 0.0 11.7 >20.0 3242.9
LoadMill A PF
flow
Mill B PF
flow
Mill C PF
flow
Mill D PF
flow
Mill E PF
flow
Mill F PF
flow
Total air
flow
MW
US-
04NM10U
603-C153
US-
04NM20U
603-C157
US-
04NM30U
603-C161
US-
04NM40U
603-C165
US-
04NM50U
603-C169
US-
04NM60U
603-C173
US-
04NG00F6
00-C90
Average 0.0 5.5 8.9 3.6 3.5 10.6 3.4 311.7
Max 0.0 5.6 28.7 3.6 3.6 28.1 3.4 332.0
Min 0.0 5.4 -0.8 3.6 3.5 -1.3 3.4 296.1
14:00:00 0.0 5.4 8.8 3.6 3.6 11.6 3.4 315.6
14:05:00 0.0 5.4 10.8 3.6 3.6 7.0 3.4 313.3
14:10:00 0.0 5.5 2.9 3.6 3.5 0.5 3.4 300.4
14:15:00 0.0 5.5 3.3 3.6 3.5 -1.3 3.4 296.1
14:20:00 0.0 5.5 4.6 3.6 3.5 2.4 3.4 301.2
14:25:00 0.0 5.5 3.3 3.6 3.6 11.8 3.4 315.2
14:30:00 0.0 5.5 0.8 3.6 3.5 6.3 3.4 300.0
14:35:00 0.0 5.5 -0.8 3.6 3.6 5.2 3.4 302.1
14:40:00 0.0 5.5 -0.5 3.6 3.6 4.0 3.4 302.5
14:45:00 0.0 5.5 9.1 3.6 3.5 9.8 3.4 310.5
14:50:00 0.0 5.4 20.2 3.6 3.6 19.4 3.4 319.3
14:55:00 0.0 5.5 28.7 3.6 3.6 28.1 3.4 332.0
15:00:00 0.0 5.5 16.6 3.6 3.5 19.7 3.4 328.1
15:05:00 0.0 5.6 16.7 3.6 3.6 20.8 3.4 314.8
15:10:00 0.0 5.5 7.3 3.6 3.6 10.8 3.4 309.2
15:15:00 0.0 5.5 10.2 3.6 3.5 13.9 3.4 327.0
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 43 of 131
Measurement No. 3
The following table presents the conditions as captured during Measurement No. 3.
.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015
Correlation
Measured Dust Mass
Test No. MW % mA mg/Nm3 (dry) @
Stack O2
3 370.0 7.9 >20.0 2557.2
LoadMill A PF
flow
Mill B PF
flow
Mill C PF
flow
Mill D PF
flow
Mill E PF
flow
Mill F PF
flow
Total air
flow
MW
US-
04NM10U
603-C153
US-
04NM20U
603-C157
US-
04NM30U
603-C161
US-
04NM40U
603-C165
US-
04NM50U
603-C169
US-
04NM60U
603-C173
US-
04NG00F6
00-C90
Average 378.0 7.3 31.2 3.6 18.6 31.8 15.9 440.5
Max 610.1 25.4 38.2 3.6 38.8 37.9 37.4 604.9
Min 131.3 3.2 15.3 3.6 -34.3 14.7 -10.5 338.9
15:45:00 131.3 5.6 30.4 3.6 3.6 32.6 3.4 346.1
15:50:00 160.6 5.5 30.0 3.6 -34.3 32.9 3.4 338.9
15:55:00 201.7 5.5 31.6 3.6 -34.3 34.0 3.4 348.0
16:00:00 248.5 5.6 31.0 3.6 -10.5 34.3 3.4 367.6
16:05:00 294.6 5.6 15.3 3.6 5.8 14.7 3.4 367.0
16:10:00 337.0 5.5 32.6 3.6 28.3 32.8 3.4 418.0
16:15:00 348.6 5.5 33.7 3.6 33.0 33.4 3.4 410.4
16:20:00 357.5 5.8 34.6 3.6 25.6 34.9 3.4 414.6
16:25:00 356.5 5.6 33.2 3.6 34.4 33.5 -10.5 436.9
16:30:00 363.0 3.2 21.5 3.6 25.2 22.4 21.0 433.0
16:35:00 409.8 4.7 28.2 3.6 29.0 28.2 27.9 441.2
16:40:00 452.2 3.2 31.5 3.6 31.8 31.9 31.4 457.0
16:45:00 493.9 3.2 35.1 3.6 35.5 35.1 34.4 477.3
16:50:00 517.5 4.4 32.6 3.6 33.2 32.8 32.6 504.5
16:55:00 551.0 8.5 35.4 3.6 36.1 35.1 35.2 519.5
17:00:00 593.0 22.1 38.2 3.6 38.8 37.9 37.4 604.1
17:05:00 610.1 25.4 34.8 3.6 35.6 34.8 34.5 604.9
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 44 of 131
5.13. Unit No. 6 Shutdown Measurements
On the 27th
of June 2016 two measurements were conducted on Unit No. 6 during a Unit
Shutdown. The first measurement was conducted during the load reduction from 270 to 0 MW.
Measurement No. 2 was conducted after with zero MW load and with just fans running. The following trends and tables show the Boiler Load, Mill PF flow variation during the
measurement period.
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 45 of 131
Measurement No. 1
The following table presents the conditions as captured during Measurement No. 2.
LoadMill A PF
flow
Mill B PF
flow
Mill C PF
flow
Mill D PF
flow
Mill E PF
flow
Mill F PF
flow
MW
US-
06NM10U
603-C153
US-
06NM20U
603-C157
US-
06NM30U
603-C161
US-
06NM40U
603-C165
US-
06NM50U
603-C169
US-
06NM60U
603-C173
Average 55.5 5.1 4.4 3.8 8.6 7.2 7.5
Max 272.1 5.2 28.7 3.8 37.0 34.0 8.9
Min 0.0 5.0 2.9 3.8 -7.0 -12.0 3.5
20:40:00 272.1 5.2 28.7 3.8 4.1 0.4 8.8
20:45:00 263.9 5.2 6.1 3.8 37.0 34.0 8.8
20:50:00 243.4 5.2 2.9 3.8 14.5 -3.3 8.8
20:55:00 186.3 5.1 2.9 3.8 29.4 12.9 8.8
21:00:00 88.9 5.2 2.9 3.8 30.0 9.5 8.8
21:05:00 0.0 5.2 2.9 3.8 6.0 -8.8 8.8
21:10:00 0.0 5.2 2.9 3.8 -7.0 -12.0 8.9
21:15:00 0.0 5.2 2.9 3.8 1.2 -9.2 8.8
21:20:00 0.0 5.2 2.9 3.8 2.3 -6.5 8.9
21:25:00 0.0 5.2 2.9 3.8 3.9 29.1 8.9
21:30:00 0.0 5.2 2.9 3.8 5.4 32.0 8.9
21:35:00 0.0 5.2 2.9 3.8 6.6 33.1 8.9
21:40:00 0.0 5.2 2.9 3.8 6.3 3.8 8.9
21:45:00 0.0 5.2 2.9 3.8 5.8 3.8 8.1
21:50:00 0.0 5.1 2.9 3.8 3.6 3.8 6.4
21:55:00 0.0 5.0 2.9 3.8 3.6 3.8 3.5
22:00:00 0.0 5.0 2.9 3.8 3.6 3.8 3.5
22:05:00 0.0 5.0 2.9 3.8 3.6 3.8 3.5
22:10:00 0.0 5.0 2.9 3.8 3.6 3.8 3.5
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 46 of 131
Measurement No. 2
The following table presents the conditions as captured during Measurement No. 1 & 2 with the
particulate concentration in mg/Nm3 dry @ stack O2 conditions.
Boiler
Load
Stack
Oxygen
Monitor Output
(0 to 2 Ext)
Output No.2
2015
Correlation
Measured Dust Mass
Test No. MW % mA mg/Nm3 (dry) @
Stack O2
1 55.5 13.6 >20.0 7608.3
2 0.0 20.6 >20.0 6311.3
LoadMill A PF
flow
Mill B PF
flow
Mill C PF
flow
Mill D PF
flow
Mill E PF
flow
Mill F PF
flow
Total air
flow
MW
US-
04NM10U
603-C153
US-
04NM20U
603-C157
US-
04NM30U
603-C161
US-
04NM40U
603-C165
US-
04NM50U
603-C169
US-
04NM60U
603-C173
US-
04NG00F6
00-C90
Average 0.0 5.5 8.9 3.6 3.5 10.6 3.4 311.7
Max 0.0 5.6 28.7 3.6 3.6 28.1 3.4 332.0
Min 0.0 5.4 -0.8 3.6 3.5 -1.3 3.4 296.1
14:00:00 0.0 5.4 8.8 3.6 3.6 11.6 3.4 315.6
14:05:00 0.0 5.4 10.8 3.6 3.6 7.0 3.4 313.3
14:10:00 0.0 5.5 2.9 3.6 3.5 0.5 3.4 300.4
14:15:00 0.0 5.5 3.3 3.6 3.5 -1.3 3.4 296.1
14:20:00 0.0 5.5 4.6 3.6 3.5 2.4 3.4 301.2
14:25:00 0.0 5.5 3.3 3.6 3.6 11.8 3.4 315.2
14:30:00 0.0 5.5 0.8 3.6 3.5 6.3 3.4 300.0
14:35:00 0.0 5.5 -0.8 3.6 3.6 5.2 3.4 302.1
14:40:00 0.0 5.5 -0.5 3.6 3.6 4.0 3.4 302.5
14:45:00 0.0 5.5 9.1 3.6 3.5 9.8 3.4 310.5
14:50:00 0.0 5.4 20.2 3.6 3.6 19.4 3.4 319.3
14:55:00 0.0 5.5 28.7 3.6 3.6 28.1 3.4 332.0
15:00:00 0.0 5.5 16.6 3.6 3.5 19.7 3.4 328.1
15:05:00 0.0 5.6 16.7 3.6 3.6 20.8 3.4 314.8
15:10:00 0.0 5.5 7.3 3.6 3.6 10.8 3.4 309.2
15:15:00 0.0 5.5 10.2 3.6 3.5 13.9 3.4 327.0
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 47 of 131
6. Monitoring Deviations
6.1. The numbers of measurements included in some of the correlations are inconsistent with the requirements of the ESKOM Standard for Emission Monitoring and Reporting (420-
56242363).
7. Recommendations
It is recommended that:
7.1. The particulate emissions from Units No. 1, 2, 4, 5 & 6, reported to the authorities, are according to the correlation functions presented in Figures No. 1 to 10 of this report.
7.2. New compliant correlations are conducted on Units No. 1, 2, 4, 5 & 6 during the first quarter of 2018.
7.3. The first Output range on Unit No. 1 to 6 be selected to a higher extinction range in order to
include the cap limit of 100 mg/Nm3 dry @ 10% O2.
8. Acknowledgement
The author expresses sincere appreciation for the co-operation and assistance of the Lethabo
personnel during the correlation.
9. References
9.1. South African Air Quality Act No. 39 of 2004 9.2. ISO 9096 Stationary source emission – Manual Determination of mass concentration of
particulate matter.
9.3. ISO 12141 Stationary source emission – Determination of concentration of particulate matter (dust) at low concentrations – Manual gravimetric method.
9.4. ISO 10155 Stationary source emissions – Automated monitoring of mass concentrations of particles – Performance characteristics, test methods and specifications
9.5. 420-56242363 ESKOM Standard for Emission Monitoring and Reporting
10. Distribution 10.1. Yaseen Waja Eskom Lethabo Power Station 10.2. Ebrahim Patel Eskom Enterprises Park 10.3. R Rampiar Eskom Enterprises Park
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 48 of 131
Appendix No. 1
Detailed Measurement Results
-
Lethabo Power Station Units No 1, 2, 4, 5 & 6 Emissions Dec 2015 to June 2016 Report No. RSL222 Page 49 of 131
Units No. 1
Table No. 1
Measurements No. 1 to 5
Date 9th
to 16th
March 2016
Dust Concentration [mg/Nm
3] is the measured dust concentration Normalised to gas conditions at 0°C and 101,325 kPa.
Customer Name Lethabo Lethabo Lethabo Lethabo Lethabo
Unit No. 1 1 1 1 1
Location Stack Stack Stack Stack Stack
Measurement No. 1 2 3 4 5
Date yyyy/mm/dd 2016/03/09 2016/03/09 2016/03/16 2016/03/16 2016/03/16
Start Time 00H00 11H16 15H06 12H07 13H53 15H35
End Time 00H00 13H38 16H20 13H25 15H07 16H47
Plant Load MW 610 610 610 610 610
Monitor