Calibration certificate No. : Ges/VAD-130402-06/LG-01/13-14 Issue Date: 04/04/2013
Client : O.P. Goyal
Site Name : -
Equipment under calibration : Load Gauge
Make : Masta
Serial No : 100/579
Range : 0-1000 KN
Least Count : 5 KN
Ram Dia. : 165 mm
Condition on Receipt : Good
Date of Calibration : 04/04/2013
Recommended Date of Next calibration : 03/04/2014
CHECKED BY APPROVED BY
Note (1) This Certificate / Report shall not be reproduced except in full.
Page 1 / 2
CALIBRATION CERTIFICATE
Summary of Measurements: The Load Gauge was calibrated against Proving Ring upto full Scale by direct
comparison method. The observed results along with the estimated accuracy in the measurements are given on
page 2.
(2) The measurement results quoted in this Certificate / report refers only to the items submitted andapply at the time of measurement.
(3) This Calibration / Test has been carried out using measurement standards and istruments whose
calibration is traceable to National / international standards
-
7/30/2019 LG-01-VAD-130401-06- op goyal
2/8
Measuring Equipment used :
Calibration certificate No. : Ges/VAD-130402-06/LG-01/13-14
0'0 180'0 360'0 AVG.
0.0 0.0 0.0 0.0 0.0 --
100.0 104.2 104.0 104.6 104.3 0.62
200.0 199.2 198.7 199.4 199.1 0.32
300.0 299.6 298.0 297.3 298.3 0.77
400.0 398.7 398.4 397.8 398.3 0.23
500.0 498.2 497.8 498.5 498.2 0.14
600.0 598.7 598.1 597.5 598.1 0.20
700.0 697.5 697.2 697.8 697.5 0.09
800.0 797.3 798.1 797.5 797.6 0.10
900.0 897.1 897.5 898.1 897.6 0.11
CHECKED BY APPROVED BY
Page 2 / 2
CALIBRATION CERTIFICATE
The FMD Proving Ring No. 08242 of full scale Range 1000 KN Externally calibrated by NATIONAL
COUNCIL FOR CEMENT AND BUILDING MATERIALS, BALLABGARH. vide certificate no.
11090407/D/5/C-178, Dated- 14/11/2013 and Valid up to 26 Months from the date of calibration.
Measurements Results: The calculated reading generated by FMD Proving Ring and the indicated Load by the
Gauge under calibration is given below.
Load Gauge Calibration
Proving Ring
Reading
Load Gauge Reading, kN Relative
Repeatability
Error %
-
7/30/2019 LG-01-VAD-130401-06- op goyal
3/8
-
7/30/2019 LG-01-VAD-130401-06- op goyal
4/8
100 48.9 213.91 104.60199 4.60199
200 93.2 213.91 199.36412 -0.63588
0.0 0.0 -- 300 136.8 213.91 292.62888 -7.37112
100.0 104.3 4.25 400 184.2 213.91 394.02222 -5.97778
200.0 199.1 -0.46 500 229.2 213.91 490.28172 -9.71828
300.0 292.6 -2.48 600 274.4 213.91 586.96904 -13.03096
400.0 393.5 -1.62 700 320.2 213.91 684.93982 -15.06018
500.0 489.4 -2.13 800 368.6 213.91 788.47226 -11.52774
600.0 586.0 -2.33 900 413.2 213.91 883.87612 -16.12388
700.0 685.8 -2.03
800.0 788.0 -1.49
900.0 884.3 -1.74
Proving
Ring
Reading
Load
Gauge
Reading,
kN
Error %
Load Gauge Calibration
-
7/30/2019 LG-01-VAD-130401-06- op goyal
5/8
MSMED Regd. No. 24-019-22-01281 &VUDA REGD. NO. UDA/PLAN-05/233/2005
Calibration certificate No. : Ges/VAD-130418-05/LG-01/13-14 Issue Date: 18/04/2013
Client : O.P. Goyal
Site Name : -
Equipment under calibration : Load Gauge
Make : Masta
Serial No : 100/579
Range : 0-1000 KN
Least Count : 5 KN
Ram Dia. : 165 mm
Condition on Receipt : Good
Date of Calibration : 18/04/2013
Recommended Date of Next calibration : 17/04/2014
CHECKED BY APPROVED BY
Note (1) This Certificate / Report shall not be reproduced except in full.
Page 1 / 2
CALIBRATION CERTIFICATE
Summary of Measurements: The Load Gauge was calibrated against Proving Ring upto full Scale by direct
comparison method. The observed results along with the estimated accuracy in the measurements are given on
page 2.
(2) The measurement results quoted in this Certificate / report refers only to the items submitted andapply at the time of measurement.
(3) This Calibration / Test has been carried out using measurement standards and istruments whose
calibration is traceable to National / international standards
-
7/30/2019 LG-01-VAD-130401-06- op goyal
6/8
Measuring Equipment used :
Calibration certificate No. : Ges/VAD-130418-05/LG-01/13-14
0'0 180'0 360'0 AVG.
0.0 0.0 0.0 0.0 0.0 --
100.0 103.2 103.5 103.8 103.5 0.58
200.0 199.8 198.5 199.4 199.2 0.65
300.0 299.6 298.6 297.5 298.6 0.70
400.0 398.9 398.6 397.8 398.4 0.28
500.0 499.1 497.9 498.8 498.6 0.24
600.0 598.8 598.7 598.2 598.6 0.10
700.0 698.1 697.2 698.8 698.0 0.23
800.0 799.3 798.9 797.9 798.7 0.18
900.0 897.7 897.1 897.5 897.4 0.07
CHECKED BY APPROVED BY
Page 2 / 2
The FMD Proving Ring No. 08242 of full scale Range 1000 KN Externally calibrated by NATIONAL
COUNCIL FOR CEMENT AND BUILDING MATERIALS, BALLABGARH. vide certificate no.
11090407/D/5/C-178, Dated- 14/11/2013 and Valid up to 26 Months from the date of calibration.
Measurements Results: The calculated reading generated by FMD Proving Ring and the indicated Load by the
Gauge under calibration is given below.
Load Gauge Calibration
Proving Ring
Reading
Load Gauge Reading, kN Relative
Repeatability
Error %
CALIBRATION CERTIFICATE
-
7/30/2019 LG-01-VAD-130401-06- op goyal
7/8
-
7/30/2019 LG-01-VAD-130401-06- op goyal
8/8
100 48.9 213.91 104.60199 4.60199
200 93.2 213.91 199.36412 -0.63588
300 136.8 213.91 292.62888 -7.37112
400 184.2 213.91 394.02222 -5.97778
500 229.2 213.91 490.28172 -9.71828
600 274.4 213.91 586.96904 -13.03096
700 320.2 213.91 684.93982 -15.06018
800 368.6 213.91 788.47226 -11.52774
900 413.2 213.91 883.87612 -16.12388
avg 1 2 3
50.1 49.4 49.6 49.7 213.9 107.2 105.7 106.1
93.4 92.8 93.2 93.1 213.9 199.8 198.5 199.4
136.6 136.8 136.8 136.7 213.9 292.2 292.6 292.6
183.9 183.8 183.6 183.8 213.9 393.4 393.2 392.7
228.60 228.60 228.90 228.7 213.9 489.0 489.0 489.6
273.9 273.8 373.6 307.1 213.9 585.9 585.7 799.2
320.6 320.8 320.8 320.7 213.9 685.8 686.2 686.2
368.2 368.4 368.4 368.3 213.9 787.6 788.0 788.0
412.8 413.8 413.6 413.4 213.9 883.0 885.2 884.7