Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD Director of IonFlux Product Development Fluxion Biosciences

Upload: others

Post on 23-Aug-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

Liquid Perfusion without the Overhead:Fast, Serial Liquid Replacement for PAM Assay Development

Ali Yehia, PhD!Director of IonFlux Product Development Fluxion Biosciences

Page 2: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Background

• Founded in 2005• Headquarters in California• Worldwide distribution• Winner of 5 NIH grants• Two time R&D 100 Award winner

• BioFlux (2008)• IonFlux (2010)

IonFlux BioFlux IsoFlux

Page 3: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

IonFlux System TM

Built by Fluxion Biosciences, distributed and supported by Molecular Devices www.moldev.com

Page 4: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

IonFlux 16 & HT

• Automated Patch Clamp System based on a unique in-plate liquid handling strategy

• No liquid handler so footprint akin to a simple plate reader

• Comes in two models: • IonFlux 16:

• 16 Amplifiers • IonFlux HT:

• 64 Amplifiers

Page 5: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Recording Plates• Standard SBS plate format

• 96 Wells for IF 16 • 384 Wells for IF HT

• Bottom of plate is replaced by intricate laminar flow system → providing an integrated

compound delivery and recording

Page 6: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

• Two Traps: → Two independent recordings

• 8 different compounds or concentrations

• Two different IC per plate • Independent conduit:

• No mixing • No cross contamination • No pipetting delays • No electrode/compound

contact

Recording Plates

Page 7: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Cover of Assay for August 2012

Page 8: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

System Integration

Page 9: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Mechanism of Action• Plates are primed with all solutions

get close to the action • Cells introduced and trapped • Seals are formed • Extra suction forms Whole Cell

Patch

Page 10: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Recording• Fast & successive compound

release • Continuous compound washing

within the gaps • No pipetting for compound

delivery allowing simultaneous application and recording across the plate

• 32 experiments completed in matter of minutes

• 500 compounds with 8 point dose response completed in few hours

Page 11: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Ensemble & Single Cell recording

• Ensemble recordings designed with 20 cells (High success rate) • Currents summed not averaged

• Single cell recording designed for highest quality (Giga-seals)

20-cell ensemble Single cell

Page 12: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Trapping Cells

• Transparent plates • Cells trapped and patched to

the side → Imaging possible

• Cell flow: • Increases access of cells

to patch holes • Can be interrupted to

trapping

Page 13: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2012 Fluxion Biosciences. All rights reserved.

PneumaticPressure

InterfaceGasket

Electrode

Heating element

Temperature Control• Heating elements right under the

plate • Accurate temperature control with

positive feedback • Rapid changing of temperature

(~1 min)

Page 14: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Temperature Control

Page 15: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Flexibility in Solution Replacement

• IonFlux software provides extensive control over the delivery of compounds and solutions

• Wash steps are as simple as separating two compound additions

• Unique successive compound additions with no break in between

• Rapid exchange of solutions via parallel compound delivery system

Page 16: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Allosteric Modulators

• Both PAM and Agonist are needed to initiate response.

• The effect is not seen without agonist.

• In order to study the effect we need:

• Response consistency from agonist activation → Repeatability

• Flexibility and speed in solution displacement

Dawson GR et al 2005

Page 17: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

GABAA Receptors

• Superimposed GABAA receptors current sweeps (HEK cells expressing α1β3γ2 - Eurofins PrecisION Cat# CYL3053)

• EC50 from 40 recording wells with normalized currents showing a value of 3.32 ± 0.1 µM

Page 18: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

GABAA Receptors: Agonists

• Different agonists providing different responses

• Dose response showing consistent results for all three agonists

muscimol > GABA > isoguvacine

Page 19: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

Repeatability

• Superimposed GABA responses to 10 µM show minimal sweep to sweep variation, providing perfect substrate for the study of modulators effects

Page 20: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

Repeatability

Consistent and repeatable response to successive additions of GABA is vital for the study of PAM effects

Page 21: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

Positive Allosteric Modulation

• Successive addition of diazepam shows a marked increase in the response to GABA.

Page 22: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

PAM Dose Response

• Successive addition of increasing concentrations of diazepam, showing marked increase in to the base1µM GABA response.

• Successive addition of increasing concentrations of triazolam, showing similar increase in base1µM GABA response.

Page 23: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

PAM Dose Response

Page 24: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Rapid Liquid Displacement AssaysRapid Liquid Displacement Assays

Page 25: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.© 2014 Fluxion Biosciences Inc.

Rapid Liquid Displacement Assays

Increasing GABA concentration recorded within the same sweep.

Page 26: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Rapid Liquid Displacement AssaysOpen channel modulation assays: 1 µM GABA → 1 µM Diazepam

Page 27: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Rapid Liquid Displacement AssaysOpen channel modulation assays + blocker

Page 28: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Rapid Liquid Displacement AssaysNMDA double agonist: 30 µM Glycine

Page 29: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

Development

Page 30: Liquid Perfusion without the Overhead · Liquid Perfusion without the Overhead: Fast, Serial Liquid Replacement for PAM Assay Development Ali Yehia, PhD! Director of IonFlux Product

© 2014 Fluxion Biosciences Inc.

IonFlux System - Summary• The IonFlux system provides a suitable

and extensive platform for PAM assay development with benefits including: • Continuous flow of all solutions • Integrated or in-plate liquid

handling system • Rapid and serial displacement of

compounds • Extensive control in solution

exchange • Fast assay completion with

parallel execution of compound additions.