low-flow ground water sampling: an update on proper application and use gary girolimon, p.e. ted d....

60
Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental Systems Inc. Copyright © QED Environmental Systems, Inc. 2008; all rights reserved.

Upload: eleanore-casey

Post on 15-Jan-2016

225 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Low-Flow Ground Water Sampling: An Update on

Proper Application and Use

Gary Girolimon, P.E.

Ted D. Miller Associates, Inc.

Representatives for

QED Environmental Systems Inc.

Copyright © QED Environmental Systems, Inc. 2008; all rights reserved.

Page 2: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Early research into purging resulted in guidelines to purge wells to remove “stagnant” water

• “Rule of thumb“ is to remove 3 to 5 well volumes” prior to sampling.

• Wells with insufficient water; evacuate and sampled upon recovery, typically within 24 hours.

• Little concern was given to how purging methods and devices (e.g., bailers) affected the chemistry of ground water samples.

Page 3: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

What does the sample represent with traditional purging methods?

Water from other vertical zones

Normally Immobile NAPL Microglobules

Normally Immobile Colloids and Sediment Elevate Turbidity

Water Chemically Altered by Gas Exchange

Page 4: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Problems with Traditional Well Purging Practices

• High purge volumes can cause underestimation due to dilution or overestimation due to induced plume migration, contaminant mobilization and increased sample turbidity.

• Dewatering lower-yield wells causes losses of VOCs, affects DO and CO2 levels, and increases sample turbidity.

• Excessive drawdown can cause overestimation (“false positives”) from landfill gas infiltration or from mobilization of soil-bound contaminants in the overlying formation.

• Purge water generated can be tens or hundreds of gallons from each well, with the associated handling and disposal costs.

Page 5: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Purging with Bailers and Inertial Lift Pumps

• Stagnant water mixes into the sampling zone, requiring extensive purging to obtain formation water.

• Surging action increases turbidity, which can affect both metals and some organic compounds.

• Insertion of the bailer and decanting samples causes aeration that can reduce VOCs and dissolved metals

• Inability to reproduce the same purging rate and sampling position leads to significant data imprecision and inaccuracy.

Page 6: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

High-Rate Purging with Electric Submersible Pumps

• Excessive drawdown pulls stagnant water into sampling zone and mobilizes smear zone contaminants.

• Water from chemically distinct zones above and below the screened zone are pulled into the borehole.

• Less accurate samples result from effects on volatiles, turbidity, mixing and dilution.

Page 7: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Hand Bailing and High-Rate Pumping Elevate Turbidity

• sample filtration adds cost and time in field or laboratory

• If filtration is not permitted, turbidity can greatly elevate concentrations of metals and some organics (e.g., PAHs) where they are bound to soils

• the chemistry of turbid samples that have been filtered to remove excess turbidity is not the same as that from low turbidity samples

Page 8: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Limitations in traditional purging methods led to the evolution of low-flow purging

• Low-flow purging and sampling is a methodology that reduces disturbance to the well and aquifer typically caused by bailing or over-pumping.

• Low-flow pumping from the well screen controls turbidity that could affect analytical results.

• Because turbidity is low, unfiltered samples can be used to get a better estimation of the true mobile contaminant load.

• Samples represent a flow-weighted average of the screen zone.

• Sampling accuracy and precision are greatly improved.

Page 9: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Low-Flow Sampling

• Low pumping rate minimizes drawdown, mixing and formation stress, isolates stagnant water.

• Low stress = low turbidity, improved sample accuracy, reduced purge volumes.

• Samples represent formation water and naturally mobile contaminants, not stagnant water in the well or mobilized contaminants.

Page 10: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Lower flow = better samples

Low-flow purging and sampling controls turbidity and delivers higher quality samples - a clear advantage.

Page 11: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Island County Landfill - Unfiltered Metals Concentrations - Well E2S

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

Dec-88 May-90 Sep-91 Jan-93 Jun-94 Oct-95 Mar-97 Jul-98 Dec-99 Apr-01 Sep-02 Jan-04

Date

Co

ncen

trati

on

in

mg

/L

Chromium

Nickel

Changed to low-flow sampling and dedicated bladder pump Dec. 1999

Effect of low-flow sampling data accuracy and precision

Page 12: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental
Page 13: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Cost Savings with Low-Flow Sampling(From Schilling, 1995)

Low-flowPurging

Three WellVolumes

Purging Analysis:Total Purge Volume (15 wells) 61 gallons 743 gallonsAverage Volume Purged 3.3 gallons 50 gallonsAverage Pumping Rate 0.3 GPM 2-5 GPMAverage Purging Time per Well 13 minutes 50 minutesTotal Purging Time (15 wells) 3.25 hours 12.5 hours

Economic Analysis (in US Dollars):Time for Purging Wells (a) $500 $1,875Disposal costs (b) $1,300 $3,750Cost per Sampling Event $1,800 $5,625Annual Sampling Costs (quarterly

sampling)

$7,200 $22,500

Sampling costs for 30 years $216,000 $675,000

(a) Two-person crew at $150/hr.USD(b) First drum = $1,000; additional drums = $300 (drum = 55 US gallons/208 liters).

Page 14: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Advantages of Low-Flow Sampling• Low-flow is a consistent, performance based standard for

purging, rather than an arbitrary rule of thumb.• It documents completeness of purging for every sample,

overcoming potential temporal variability in factors that can affect required purge volume.

• Low-flow sampling can reduce sampling costs:– Sample Quality - reduced turbidity, more accurate dissolved

concentrations, and a better estimate of the true mobile contaminant load

– Direct cost savings - reduced purge water handling & disposal, reduced purging time (in some wells).

– Indirect cost savings - improved data accuracy and precision (no false statistical “hits”), and better data = better decisions.

Page 15: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

What Does a Low-Flow Sample Represent?

• While low-flow sampling is recognized as a solution for improving sample quality and reducing purge water volume, defining what the sample represents, in terms of the zone within the well screen that is sampled, is still often questioned.

• Pump placement can be an issue when changing to LFS from traditional purging; the concern is that samples will come from a narrow zone within the well screen, possibly “missing” contamination above or below the pump intake.

• Field experiments, laboratory simulations and numerical modeling support the position that samples are derived from the entire screen zone under low-flow pumping conditions.

Page 16: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Vertical Concentration Profiles (Puls and Paul, 1998)

• Chromium concentrations were averaged throughout the well screen zone under very low flow pumping conditions (0.25 lpm), whereas concentrations were known to be measurably stratified within the surrounding formation based on direct-push samples taken next to the screen.

• The concentrations from low-flow samples were virtually identical to the mean concentration of the multi-level and direct-push samples taken.

• Bailed sample concentrations were biased lower than the low-flow pumped sample results.

Device DMLS Geoprobe Low-Flow Bailer

Cr (mg/l) 1.69 1.86 1.76 1.05

Page 17: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

MODFLOW Simulations

• MODFLOW computer simulations with high vertical definition by Varljen, et al. (GWMR, Winter 2006) show that flow across a well screen, regardless of pump placement, captured water from the entire well screen zone.

• The objectives of the modeling experiment were to demonstrate the actual monitoring zone achieved during low flow purging and sampling and assess the effects of:– screen length– pumping rate– pump position in screen– sand pack configuration– formation heterogeneity and anisotropy– full vs. partial screen penetration

Page 18: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

E n d C a p

S a n d P a c k

G r o u t

P u m p I n t a k eS c r e e n Z o n e

Varljen, et al. 2006

Horizontal View of MODFLOW Simulation

Page 19: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Varljen, et al. 2006

Model Output at 0.25 L/min Steady State Flow

Page 20: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Varljen, et al. 2006

Vertical Distribution of Flux into a 10-foot Well Screen and Effect of Changes in Pumping Rate

Page 21: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Vertical Distribution of Flux into 20-foot Well Screen

Varljen, et al. 2006

Page 22: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Varljen, et al. 2006

Effect of Pump Placement on Vertical Flux Distribution

Page 23: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Varljen, et al. 2006

Effect of Pump Placement on Vertical Flux Distribution

Page 24: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Varljen, et al. 2006

Effect of Heterogeneities on Flux Distribution Pattern

Page 25: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Varljen, et al. 2006

Effect of Heterogeneities on Flux Distribution Pattern

Page 26: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

• Water is drawn from the entire length of the well screen during low-flow purging and sampling.

• The flow pattern remains the same in well screens from 5 - 20 feet long.

• Pump placement has no significant effect on the flow pattern into the well; geology is the controlling factor. More water comes from the most permeable zones near the well screen, regardless of pump position.

Modeling Conclusions

Page 27: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Low-Flow Sampling Application Guidelines – The Basics

• Flow rates must be controlled to pump without continuous drawdown (water level must stabilize) and not increase turbidity.

• Drawdown is based on well performance, not arbitrary guidance.

• Indicator parameters are monitored for stabilization to indicate formation water and purging completeness.

• Dedicated sampling equipment is preferred. Portable pumps require larger purge volumes, can increase turbidity and require decontamination between wells, but are still better than bailing or high-rate pumping.

Page 28: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Purging Flow Rates and Drawdown

• From US EPA, 1996: “Typically, flow rates on the order of 0.1 - 0.5 L/min are used, however this is dependent on site-specific hydrogeology. Some extremely coarse-textured formations have been successfully sampled in this manner at flow rates to 1 L/min.”

• The goal is to achieve a stabilized pumping water level in the well as quickly as possible. This reduces mixing within the borehole, and draws water only from the sampling zone.

• Flow rates are established for each well based on drawdown and turbidity values measured during purging.

Page 29: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Drawdown• From USEPA 1996: “The goal is minimal drawdown

(0.1m) during purging. This goal may be difficult to achieve under some circumstances… and may require adjustment based on site-specific conditions and personal experience.”

• This drawdown example has been interpreted as a maximum drawdown limit by some agencies without any consideration for site-specific conditions. There is no data to support this, or any other drawdown limit.

• A study by Vandenberg and Varljen shows that the true goal is to establish a stable pumping water level during purging, with indicator parameter stabilization following this.

Page 30: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Correlation of Drawdown and Indicator Parameter Stabilization

Drawdown and Specific Conductance During PurgingSt. John's Landfill Well D-2A

0

2

4

6

8

10

12

14

16

14:00 14:30 15:00 15:30 16:00

Time

Dra

wd

ow

n (

ft)

475

480

485

490

495

500

505

510

Sp

. C

on

d.

(uS

/cm

)

Pump OffDrawdown

Specific Conductance(10-minute moving average)

Casiing 0.5 1 1.5 2 2.5 3

At the point where the water level stabilized, the indicator parameters measured (conductivity shown above) also stabilized.

(Vandenberg and Varljen, 2000)

Page 31: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Drawdown Recommendations

• Drawdown should be observed and recorded for each well based on a stabilized pumping water level

• Extended purging at low flow rates, even where drawdown exceeds several feet, is preferable to high-rate well volume pumping, bailing or well evacuation.

• Drawdown within the screen should be avoided, but is still preferable to evacuation and recovery, and may be unavoidable in water table wells and low yield wells.

• Very-low-yield wells that will not stabilize at flow rates of 0.1 - 1.0 L/min should be candidates alternative sampling methods (such as minimal purge sampling) in lieu of well evacuation.

Page 32: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Purging Very-Low-Yield Wells

• In very low yield formations, purging to remove stagnant water will dewater the well, resulting in large changes in water chemistry and highly altered samples.

• The preferred approach is sampling the water in the screen zone without dewatering the well, purging 2x the equipment volume prior to sampling and measure field parameters prior to sampling.

• Applies to wells that cannot be pumped at low rates (100 ml/min) without continuous drawdown.

• Minimum purge sampling is less disruptive than well evacuation and minimizes sample chemistry alteration (the “lesser of two evils” approach).

Page 33: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Designing the Sampling System to Minimize Purge Volumes

• The sampling system can be designed to minimize displaced water and reduce the equipment volume– short/smaller diameter pump design – 100 to150 mL

– smaller tubing ID (less volume/foot) – 5 to 10 ml/foot

– Flow cell closely surrounds probes - 150 to 200 mL

• Work with your lab to determine lowest possible sample volume acceptable for analysis to minimize drawdown of water column during sampling.

Page 34: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Sampling Flow Rates• Sampling flow rates “less than 0.5

L/min are appropriate.” (US EPA 1996).

• Use rates at or below the purging flow rate for metals and other inorganic parameters, lower rates (100 ml/min.) for VOCs and filtered samples.

• Fill larger sample bottles first, then reduce the flow rate (if needed) for VOCs and any filtered parameters.

• Sampling at 100 ml/minute for all parameters can extend sampling times unnecessarily.

Page 35: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Indicator Parameters for Purging• Indicator parameters often include pH, temperature,

conductivity, DO, ORP (redox) and turbidity.• DO and C are the most reliable indicators, based on

published research and field experience.– pH stabilizes readily, often shows little change.– Temperature measured at the well head is affected by sunlight,

ambient temperature, and some electric pumps– Turbidity cannot indicate when purging is completed. It should be

measured primarily to support sample data and prevent excessive pumping/formation stress.

• Stabilization criteria are typically + 5-10% of readings, or a range of units (e.g., + 0.2 mg/L DO) where percentages are not appropriate. Stabilization occurs when three consecutive readings fall within the criteria.

Page 36: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Measuring indicator parameters

• Traditional approaches use hand-held or bench-top instruments that expose samples to air and make precise measurement intervals difficult.

• Readings may not appear stable even though water chemistry has stabilized.

Page 37: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

An in-line flow cell isolates water from air, maintaining water chemistry and allowing automated measurement. Open-top “flow containers” can’t achieve accurate values for dissolved oxygen or redox due to rapid gas exchange.

Page 38: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Typical flow-cell output provides simultaneous display of parameters while storing readings for future recall

Page 39: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Typical Indicator Parameter Stabilization Curves

0

2

4

6

8

10

12

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32

Time, minutes

Te

mp

, p

H,

DO

, T

urb

0

100

200

300

400

500

600

EC

, IS

E

Temp oC

pH

DO mg/l

Turb ntu

EC uS

ISE mv

Page 40: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Regulatory and Industry Guidance on Low-Flow Sampling Procedures

• US EPA Low-Flow (Minimal Drawdown) Ground-Water Sampling Procedures (April 1996)

• US EPA Ground-Water Sampling Guidelines for Superfund and RCRA Project Managers (May 2002)

• Various EPA Regional SOPs or guidance• State guidance document and SOPs

– Florida DEP FS2200 (June 2004)

• ASTM D6771 (2002), Standard Practice for Low-Flow Sampling

Page 41: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

• Some regulatory guidance documents and SOPs have requirements for comparison testing, vertical profiling, or arbitrary operational limits (e.g., maximum drawdown values or screen lengths) that don’t consider site-specific conditions or recognize the published scientific literature.

• These requirements and restrictions are likely to cause potential users to stick with bailing and high-rate/high volume purging. In the end, no one benefits - users spend more money and time to get less meaningful data.

Regulatory Guidance on Low-Flow Sampling Procedures

Page 42: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Passive and No-Purge Sampling: The next evolutionary step?

• Over the past decade, researchers and manufacturers have studied sampling wells without purging.

• The methodology is based on the concept that ambient groundwater flow through the well screen would maintain an approximation of the groundwater chemistry within the screen zone of the well.

• Samples taken represent the discrete interval within the screen where the samplers are placed or activated, rather than a flow-weighted average of the screen zone.

• Several passive or no-purge samplers can be deployed at various depths in the well screen in order to identify any contaminant stratification within the water column.

Page 43: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Polyethylene Diffusion Samplers

DiffusionSamplers

Weight

Cross Sectional View

Typical sampler is 1-2 feet in length

Page 44: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Certain VOCs, primarily chlorinated and aromatic hydrocarbons, will diffuse through the polyethylene bag

TCE

BTEX

VC

Page 45: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

MTBE

METALS

SVOCs

COLLOIDS

PESTICIDES

RADIONUCLIDES

Most other common analytes, such as metals, pesticides, semi-volatiles and some volatiles can’t diffuse into the bag

Page 46: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

No-Purge Sampling can be performed with various grab samplers

HydraSleeve ™ Snap Sampler™

Kemmerer Sampler

Page 47: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

How do low-flow sampling and no-purge/passive sampling compare?

• Studies comparing LFS and no-purge/passive sampling show good correlation in some wells but significant differences in others. This may be the effects of contaminant stratification In the well, ambient vertical flow and mixing and the operational limitations of some passive sampling devices.

• Cost comparisons are typically based on a single sampling point. If multiple passive/no-purge samplers are used, sampling cost savings are quickly offset by increased analytical costs of multiple samples.

• The biggest difference is in what the sample represents. LFS represents a flow-weighted average of the well screen, while a no-purge/passive sample may represent either a depth-discrete sample of the well screen (but not necessarily the surrounding formation), or may represent some undefined average value of the screen zone due to ambient vertical flow and mixing within the well.

Page 48: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Potential Bias/Error in Samples from Sampling Equipment

• Temperature and pressure changes from heat generated by the device and lift principle

• Loss or gain of dissolved gasses due to rotating impellers and air/water interface in the device

• VOC loss due to aeration, agitation or pressure drop• Turbidity increase from surging and mixing in well• Material leaching, sorption/desorption

Page 49: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Sample Temperature Increase from Sampling Pump

21

21.5

22

22.5

23

23.5

24

0 20 40 60 80 100 120

Time, minutes

Te

mp

era

ture

, C

Bladder pump

RediFlo 2

(From Oneacre and Figueras, 1996)

Page 50: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Effect of sampling devices on VOC concentrations

(From Pohlmann, et al., 1991)

0

200

400

600

800

1000

1200

1400

1600

Co

ncen

trati

on

, u

g/L

Benzene Chlorobenzene

Multi-level

Bailer

Bladder pump

Page 51: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Drive Air

Bladder

Page 52: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Bladder pumps are available in a wide variety of materials (SS, PVC, PTFE) and sizes to fit wells down to 1” diameter and depths to 1,000 feet or greater.

Page 53: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Bladder pumps designed specifically for portable use feature inert polyethylene bladders and quick disassembly without tools for easy decontamination.

Page 54: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Bladder Pumps Advantages• Highest degree of sample accuracy and precision under

widest range of field conditions; not subject to operator influence like bailers and high-flow pumps.

• Systems can be installed in minutes and are easy to operate with minimal training.

• Simple design with only three moving parts has been proven to be virtually maintenance-free for the life of most applications.

• Ability to operate at low flow rates without damage from heating or running dry makes bladder pumps the best choice for low-flow purging and sampling applications.

Page 55: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Dedicated System Portable System

Pumps Remain in Each WellPumps Remain in Each Well Moves From Well To WellMoves From Well To Well

Should your sampling system be dedicated or portable?

Page 56: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Portable Sampling Systems• Lower equipment cost (purchase/rent/lease)

• Higher operating costs (operation/maintenance)– assembly/disassembly at each well– cleaning labor (average 1 labor-hour/well)– cleaning supplies (buckets, brushes, detergent & water)– reagents (isopropanol, methanol, hexane, acid wash)– cleaning blank sample analysis– re-sampling & added analysis if any “suspect” samples

• Potential for cross-contamination of samples & wells

• Greater exposure to contaminants & cleaning agents

Page 57: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Dedicated Sampling Systems• Higher initial equipment cost (purchase/lease)

• Lower operating costs (operation/maintenance)– no assembly/disassembly at each well– no cleaning labor– no cleaning supplies or reagents– no cleaning blank samples– labor typically 30% - 50% lower overall

• Eliminates cross-contamination from sampling equipment

• Reduces potential for operator exposure to contaminants

Page 58: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Cost Savings / Ease of Use• End-user design is simpler than bailers or portable pumps• Dedicated systems reduce sampling labor and expenses,

resulting in a break-even in two years or less for most sites doing quarterly sampling

Page 59: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Summary• Sampling methods are like tools in a toolbox – each has an

appropriate use based on the task at hand.• Traditional purging and sampling methods often bias sample results

and can generate significant waste.• Passive and no-purge sampling methods can identify contaminant

stratification within a well screen, but this information may not reflect the actual concentration in the formation due to ambient vertical flow and redistribution of contaminants within the borehole.

• Low-flow sampling can improve sample accuracy and precision and avoid biases associated with turbid samples. It provides flow-weighted average concentrations from typical monitoring wells, and is the best approach for determining the true mobile contaminant load in most monitoring scenarios.

• Bladder pump systems offer the best combination of reliability and sample accuracy. Dedicated systems can reduce sampling costs by eliminating decontamination between wells and reducing field labor.

Page 60: Low-Flow Ground Water Sampling: An Update on Proper Application and Use Gary Girolimon, P.E. Ted D. Miller Associates, Inc. Representatives for QED Environmental

Thank you!For more information, contact:

Gary GirolimonTed D. Miller Associates

Tel: [email protected]

www.qedenv.comwww.micropurge.com