magnetic plasma

Upload: riky-ikhwan

Post on 14-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/29/2019 Magnetic Plasma

    1/6

    Influence of magnetic fields on negative corona discharge currents

    Junfeng Mi, Dexuan Xu*, Yinghao Sun, Shengnan Du, Yu Chen

    Laboratory of Discharge Plasma and Pollution Control Engineering, Department of Environmental Science and Engineering,

    Northeast Normal University, Changchun 130024, China

    a r t i c l e i n f o

    Article history:

    Received 1 July 2007

    Received in revised form 31 March 2008

    Accepted 16 April 2008

    Available online 28 May 2008

    Keywords:

    Magnetically enhanced

    Negative corona discharge

    Plasma

    a b s t r a c t

    The mechanism of magnetically enhanced negative corona discharges was studied by comparing the

    influences of two different magnetic fields on the negative corona discharge current. In the magneticallyenhanced corona discharge, a local magnetic field is formed near the discharge electrode by using the

    small permanent magnets, and the corona discharge currents are enhanced because the Larmor

    movements of free electrons enhance the ionizations of the gas molecules in the ionization region. It is

    assumed that the increase of the discharge currents attributes to only the enhanced ionization process in

    the small ionization region, and is not relevant to the lengthening trajectory of free electrons in the wide

    drift inter-electrode region. In the enhanced ionization-region magnetic field, the mechanism of mag-

    netically enhanced corona discharges was explained and the relative increase of the discharge current

    could exceed 150%. However, in a weakened ionization-region magnetic field, the mechanism of mag-

    netically enhanced corona discharges was inconspicuous and the relative increase of the discharge

    current could only reach about 25%. It is predicted that after the magnetic field of magnetically enhanced

    negative corona discharges had been fixed, the relative increases of the discharge current for varied

    mean electric fields in the inter-electrode region could have a maximum value, according to the

    mechanism of magnetically enhanced negative corona discharges. The above predication was completely

    validated by the current experiment data. In addition, the optimum combinations between the electric

    field and the magnetic field were obtained. In order to reach the largest relative increase of the discharge

    current under 4 kV/cm mean electric field intensity in the inter-electrode region for a practical elec-trostatic precipitator, the optimum magnetic field with a magnetic flux density of about 0.43 T at the

    edges of magnetic st rings should be selected.

    2008 Elsevier B.V. All rights reserved.

    1. Introduction

    Recently, more and more countries are beginning to limit the

    emissions of micron and sub-micron aerosol particles. Thus the

    corona discharges have progressively employed for electrostatic

    precipitators (ESP) and pre-chargers as well as some new applica-

    tions, such as sterilization [1], the enhancement of chemical vapor

    deposition [2], separation [3], and the production of ozone [4].The theory for charging particles is well developed for the better

    application of ESP. Pauthenier and Moreau-Hanot [5] developed an

    expression for field charging of large aerosol particles. Fuchs and

    Bricard developed the same statement for diffusion charging of

    smaller aerosol particles, and the combined field and diffusion

    charging theory was discussed by Liu and Kapadia [6]. A few

    experimental and theoretical studies [710] indicated that Fuchs

    theory had successfully predicted the charging probability of fine

    particles. The experimental and mathematical study carried out by

    Reischl et al. [11] demonstratedthat Fuchs theory was also valid for

    bipolar diffusion charging of fine particles, while the charging

    probability of positive ions was less than that of negative ions. In

    addition, there was no difference between varied gases and particle

    materials. The charges on fine aerosol particles also vary in agree-

    ment with diffusion charging theory, which tended to promote the

    charging of fine aerosol particles.Based on the above charging theories, the diffusion charging

    should be intensified for removal of fine aerosol particles in ESP. In

    addition, the diffusion charging will be a determined factor when

    the fine aerosol particles pass through ESP with a weak intensity of

    the applied electric field as well as a higher gas temperature.

    However, the field charging will be a determined factor when the

    large aerosol particles pass through ESP with a strong intensity of

    the applied electric field and a lower gas temperature [12].

    According to the diffusion charging theory of ions, it is advan-

    tageous for diffusion charging to supply a higher concentration of

    ions when the intensity of the electric field is weaker. However, the

    weak electric field could not induce a high concentration of ions in* Corresponding author.

    E-mail address: [email protected] (D. Xu).

    Contents lists available at ScienceDirect

    Journal of Electrostatics

    j o u r n a l h o m e p a g e : w w w . e l s e v i e r . c o m / l o c a t e / e l s t a t

    0304-3886/$ see front matter 2008 Elsevier B.V. All rights reserved.doi:10.1016/j.elstat.2008.04.010

    Journal of Electrostatics 66 (2008) 457462

    mailto:[email protected]://www.sciencedirect.com/science/journal/03043886http://www.elsevier.com/locate/elstathttp://www.elsevier.com/locate/elstathttp://www.sciencedirect.com/science/journal/03043886mailto:[email protected]
  • 7/29/2019 Magnetic Plasma

    2/6

    the general corona discharges. Therefore, the fine aerosol particles

    could be effectively captured when a higher concentration of ions

    and weaker electric field are supplied by ESP. In order to charge fine

    aerosol particles, it is necessary to enhance the corona discharge

    currents in ESP with a weak intensity of electric field.

    Researches on enhancing corona discharge current using

    a magnetic field have been carried out in recent years. Initially, the

    study on a small cylinder corona nozzle posited in the inner-hole of

    a cylindrical permanent magnet where a weakened ionization-

    region magnetic field was formed, was reported [13]. Although the

    propagation of an ionization region was noticed in the experiment,

    the influence of the magnetic field on ionization was not given

    enough attention. We then conducted research on a magnetically

    enhanced corona pre-charger [14]. An enhanced ionization-region

    magnetic field was formed by installing a small permanent magnet

    near the discharge electrode, which causes the magnetic flux

    densityto be higher in the ionization region than in the drift region.

    When the magnetic flux density was 0.43 T at the edges of mag-

    netic strings, the discharge current was increased by 1.3 times

    under a high voltage of 15 kV. The concentration of negative ions in

    the charging region of the pre-charger was greatly increased as

    well. This result is of great benefit to the charging of fine aerosol

    particles. It was supposed that the dominant mechanism of mag-netic enhancement in the corona discharges involves the Larmor

    movements of free electrons which enhance ionizations of the gas

    molecules near the discharge electrode, and the lengthening

    trajectory of free electrons also induces a little increase of corona

    discharge current.

    In the current study, we conductedthe research on the influence

    of the magnetic field on negative corona discharge current to fur-

    ther explore the mechanism of magnetically enhanced corona

    discharge. Only the magnetically enhanced negative corona dis-

    charges were discussed in this paper for its wide use.

    2. Experimental apparatus

    The magnetic field was applied near the discharge electrode;

    herein called the ionization-region magnetic field. The experi-

    mental apparatus is shown in Fig. 1. The effective length of the

    stainless-steel wire electrode (4) was 10 mm and 0.5 mm in di-

    ameter. Two cylindrical, permanent magnet strings (5) of 6 mm in

    diameter were made from permanent magnet disks and assembled

    at opposite ends of the wire electrode. The magnetic flux density

    near the discharge electrode could be changed by increasing or

    decreasing the numberof magnet disks in the magnetic strings. The

    spacing between the two magnetic strings was kept at 12 mm. The

    wire electrode was positioned exactly in the center of stainless

    cylinder electrodes (3). The length of cylinder electrode was 70 mm

    and its diameter 70 mm. A high-voltage power supply (1) (Beijing

    Electrostatic Instrument Factory, China, GJ F 100), capable of de-

    livering negative voltage, was measured using a high voltage

    divider (2) (Shanghai Huisha Instrument Co., Ltd., China, FRC-50).

    Between the cylinder electrode and grounding wire, two ammeters

    with different measurement ranges were connected. One ammeter

    (Cany Precision Instruments Co., Ltd., China, BS15/6) was used to

    measure the corona onset voltage, and the other (Cany Precision

    Instruments Co., Ltd., China, BS15/16) was used to measure the

    discharge current.

    Fig. 2 shows the magnetic and the electric lines of force between

    discharge and cylinder electrode when the strongest magnetic flux

    density (near magnetic string edge) is 0.43 T. The magnetic lines of

    force were approximately perpendicular to the electric lines of

    force near the wire electrode. The magnetic flux density of mag-

    netic string edges was measured using a gaussmeter (Shanghai NO.

    4 Multimeter Manufacturing Co., Ltd., China, CT5 A). It is obvious

    that the magnetic flux density gradually decreases from the wireelectrode to the cylinder electrode when a permanent magnet is

    applied near the wire electrode, i.e., the influence of the magnetic

    field on the ionization region is stronger than the drift region.

    In the second experiment, the magnetic field was applied near

    the collecting electrode; herein called the drift region magnetic

    field. The experimental apparatus is shown in Fig. 3. The effective

    length of the stainless-steel wire electrode (4) was 70 mm and

    0.5 mm in diameter. The length of the cylinder electrode (3) was

    100 mm, with an insidediameterof 56 mmand an outside diameter

    of 60 mm. The cylinder magnet (5) was made from permanent

    magnet rings and was installed outside the cylinder electrode, and

    the cylinder magnet was 50 mm long with an inside diameter of

    60 mm and an outside diameter of 72 mm. The magnetic flux

    density in theentire inter-electrode regioncouldalso be changed byincreasing or decreasing the numberof magnet rings of the cylinder

    magnet. The wire electrode was positioned exactly in the center of

    the cylinder electrode (3). The functions of the high-voltage power

    Fig.1. Schematic of experimental apparatus with the ionization-region magnetic field.

    1. HV power supply, 2. HV divider, 3. cylinder electrode, 4. wire electrode, and 5.magnet strings.

    Fig. 2. Schematic of the magnetic lines of force and the electric lines of force in theionization-region magnetic field.

    J. Mi et al. / Journal of Electrostatics 66 (2008) 457462458

  • 7/29/2019 Magnetic Plasma

    3/6

    supply (1) and the high voltage divider (2) in this experiment wereaccordant with those in the experiment of the ionization-region

    magnetic field.

    Fig. 4 shows the magnetic and the electric lines of force between

    the wire and cylinder electrodes when the mean magnetic flux

    density is 0.015 T near the wire electrode in the drift region mag-

    netic field experiment. The strongest magnetic flux density is

    0.39 T. In order to determine the mean magnetic flux density near

    the wire electrode, four equidistant points on the axis in the ioni-

    zation region were considered. The average of the four values was

    taken as the mean magnetic flux density in the ionization region.

    It is obvious that this permanent magnet could induce a stron-

    ger magnetic field in the drift region than that in the ionization-

    region magnetic field. The magnetic lines of force were also

    approximately perpendicular to the electric lines of force near thewire electrode. Clearly, the magnetic flux density gradually in-

    creased from the wire electrode to the cylinder electrode.

    3. Influence of magnetic fields on the currents of negative

    corona discharges

    As seen in Fig. 5, the corona discharge current was greatly en-

    hanced by increasing the magnetic flux density in the ionization-

    regionmagnetic field. When the magnetic fluxdensity was0.43 Tat

    the edges of magnetic strings, the current increased by about 1.5

    times under the mean electric field intensity of 4 kV/cm, which isthe ratio of the inter-electrode voltage to the spacing between the

    discharge electrode and the collecting electrode. The current in-

    creased by about 1.2 times under the mean electric field intensity of

    4 kV/cm, when the magnetic flux density was 0.38 T at the edges of

    magnetic strings.

    However, as seen in Fig. 6, it was inconspicuous of the increase

    of corona discharge current in the drift region magnetic field. As an

    external cylinder magnet was employed, the discharge current

    increased about 25% when the mean magnetic flux density was

    0.015 T near the wire electrode.

    In order to decrease the magnetic flux density near the dis-

    charge electrode, the magnet rings were decreased. The discharge

    current was almost unchanged when the mean magnetic flux

    density was about 0 T in the ionization region. However, themagnetic flux density was stronger in the drift region under this

    condition. It could be concluded that the corona discharge current

    could not be enhanced by the Larmor movements of free electrons

    in the drift region, i.e., the increase of the discharge current is not

    relevant to the lengthening trajectory of free electrons in the wide

    drift region.

    The inter-electrode region can be divided into two parts as we

    know, namely, the ionization region and the drift region. The ion-

    ization region occupies only about 0.5% of the whole inter-electrode

    Fig. 3. Schematic of experimental apparatus with drift region magnetic field. 1. HV

    power supply, 2. HV divider, 3. cylinder electrode, 4. wire electrode, and 5. cylinder

    magnet.

    Fig. 4. Schematic of the magnetic lines of force and the electric lines of force in thedrift region magnetic field.

    Fig. 5. Characteristic curves showing the discharge current as function of inter-elec-

    trode mean electric field intensity for the negative corona discharges under different

    magnetic field intensities in the ionization-region magnetic field experiment.

    Fig. 6. Characteristic curves showing the discharge current as a function of inter-

    electrode mean electric field intensity for negative corona discharges under differentmagnetic field intensities in the drift region magnetic field experiment.

    J. Mi et al. / Journal of Electrostatics 66 (2008) 457462 459

  • 7/29/2019 Magnetic Plasma

    4/6

    volume, while the drift region occupies about 95.5%. In the current

    study, two different applications of the magnetic field have been

    designed to produce different influences, respectively, on the ion-

    ization region and drift region.

    It is obvious that the increase of the corona discharge current is

    mainly determined by the magnetic flux density in the ionization

    region and is independent of the magnetic flux density in the drift

    region. The increase of the discharge current is only due to the

    enhanced ionization process in the small ionization region near

    discharge electrode, and is not relevant to the lengthening trajec-

    tory of free electrons in the drift region.

    3.1. Influence of magnetic fields on the free electrons in the

    ionization region

    In the conventional negative corona discharges, free electrons

    move along the electric lines of force due to the Coulomb forces of

    the electric field. In the ionization magnetic field, however, the

    magnetic lines of force lie perpendicular to the electric lines of force

    in the vicinity of the wire electrode. The free electrons are acted

    upon both Lorentz and Coulomb forces, hence Larmor movements

    are formed.In order to elucidate this process, some correlative numerical

    values of Larmor movements were estimated as follows.

    The Larmor frequency of free electrons

    f Bq2pm

    (1)

    where B is the magnetic flux density, q is an electronic charge, and

    m is electronic mass. In the ionization-region magnetic field, f is

    1.21010 rad/s when the magnetic flux density is 0.43 T at theedges of magnetic strings. The period of gyration T thus is about

    0.8 1010 s.According to Ref. [15], we can also get the radius of ionization

    region and the electric field intensity.

    The radius of ionization region

    a r0 0:03ffiffiffiffiffi

    r0p

    (2)

    where r0 is the radius of the curvature on the discharge electrode.

    So the radius of ionization region is about 7.1104 m.The electric field intensity can be estimated by using the Peek

    formula in the ionization region [16]

    E 3 106f

    T0p

    Tp0 0:03

    ffiffiffiffiffiffiffiffiffiffiT0p

    Tp0a

    s !(3)

    wherefis the coarseness of surface of the wire electrode,fz0.6; T0is the standard temperature, T0 273 K; p0 is the thermodynamicstandard state pressure,1.013

    105 Pa; Tis the real temperature, K;

    p is the real pressure, Pa.

    Therefore, the electric field intensity is about 1.7 106 V/m inthe ionization region when the magnetic flux intensity is 0.43 T at

    the edges of magnetic strings. Besides, when the collision is not

    taken into account, the mean velocity can be calculated when a free

    electron moves from the surface of wire electrode tothe edge of the

    ionization region by the following two Equations:

    qE$a 12

    mV2 (4)

    V 12

    V (5)

    where Vis the velocity of free electron at the edge of the ionizationregion, and V is its mean velocity, which is about 1.0 107 m/s.

    Thus, it will take the free electron about 7.11011 s moving fromthe surface of the wire electrodeto the edge of the ionization region

    when the magnetic flux density is 0 T. The moving time of free

    electrons will increase in the ionization region when the ioniza-

    tion-region magnetic field is applied, and the gyration time is at

    least about 0.88 period.

    And the radius of gyration

    R mVBq

    1:3 104 m (6)

    According to the above analyses, the moving distance of free

    electrons will increase at least by 8.2104 m in the ionizationregion when the free electrons move from the surface of the dis-

    charge electrode. Therefore, the number of collisions will increase

    to 215.2% and the mean energy of free electrons will decrease by

    115.2% between two collisions (an electron free path) for these free

    electrons under this condition. In conclusion, the discharge cur-

    rents will be enhanced due to the exponential increase of collisions

    for all the free electrons in the ionization region.

    However, the trajectory of free electrons is very complicated

    because of the collisions. In any case, it is certain that the trajectory

    of free electrons substantially lengthened and the mean energy offree electrons decreases in the ionization region when the ioniza-

    tion-region magnetic field is applied.

    3.2. Influence of magnetic fields on the free electrons in

    the drift region

    In the drift region, the collisions could not promote ionization

    because of the lower kinetic energy of free electrons. In addition,

    the neutralizations could notoccur because only the negative space

    charges (free electrons and negative ions) existed in the drift region

    in the magnetically enhanced negative discharges. In the above two

    experiments, the number of collisions between free electrons and

    gas molecules also increased in the drift inter-electrode region forthe influence of the magnetic field. However, more free electrons

    were attached by the gas molecules to form the negative ions in the

    drift inter-electrode region. In a word, the charge concentration

    increased and the drift velocity along the electric line of force for

    both negative ions and free electrons decreased in these two

    experiments.

    It can also be explained why the increase of the discharge cur-

    rents is not relevant to the lengthening trajectory of free electrons

    in the wide drift region. We suppose that the output negative

    charges from the ionization region are Q within per unit time (t).

    Since both the neutralizations of charges and the ionizations are

    inexistent in the drift region, all the charges Qshould come into the

    collecting electrode in the same unit time according to the principle

    of electric current continuity. Therefore, the discharge currentIQ/t, is a fixed value. It is obvious that the increase of the dis-charge currents is not relevant to the lengthening trajectory of free

    electrons in the wide drift region, which is in accordance with the

    experimental data. The discharge currents cannot be enhanced by

    the lengthening trajectories of negative ions as well as that of the

    free electrons in the magnetically enhanced negative corona

    discharges.

    In summary, the mechanism of the magnetic enhancement is

    that the magnetic field could affect the ionization region. Therefore,

    the corona discharge current was enhanced by the Larmor move-

    ments of free electrons in the ionization region, and the space

    charge concentration is enhanced in the whole inter-electrode

    region. Moreover, the lengthening trajectories of free electrons and

    negative ions cannot increase the corona discharge current in thedrift region. Remarkably, to increase the corona discharge current,

    J. Mi et al. / Journal of Electrostatics 66 (2008) 457462460

  • 7/29/2019 Magnetic Plasma

    5/6

    it is important to install the permanent magnet near the discharge

    electrode to induce a stronger magnetic flux density.

    3.3. Optimum combination between electric field and

    magnetic field

    In the ionization-region magnetic field, the radius of gyration for

    the Larmor movements increases with the mean electric field in-

    tensity when the magnetic field remains unchanged. The moving

    distance of a free electron consists of both the distance along the

    electric line of force and the distance caused by Larmor movement

    in an electron mean free path. When the magnetic field was fixed,

    the moving distance of a free electron along the electric line of force

    will increase with the inter-electrode electric field intensity in ev-

    ery electron mean free path. That means the moving distances of

    free electrons caused by Larmor movements will decrease with the

    increasing inter-electrode electric field intensity in every electron

    mean free path when the magnetic field was fixed. Therefore, the

    relative increase (magnetically enhanced corona discharge to con-

    ventional one) of the collision number between free electrons and

    gas molecules should decrease with the enhancing electric field

    intensity when the magnetic field remains unchanged. This tendsto restrain the enhancement of ionizations. However, the mean

    energy of free electrons will increase with the electric field in-

    tensity which tends to enhance ionization.

    According to above analyses, it is predicted that the relative

    increase of corona discharge current could have a maximum value

    at a special point because the change of the relative increase of

    collision number is in the opposite direction of that of the mean

    energy of free electrons.

    The characteristic curves (Fig. 7) were generated by analyzing

    the data of the ionization-region magnetic field experiment,

    showing the relative increase of discharge current as a function of

    the inter-electrode mean electric field intensity for negative corona

    discharges under different magnetic flux densities of 0.38 T and

    0.43 T. Fig. 7 shows that the relative increase of the discharge cur-rent has a maximum value, which is accordant with our prediction.

    The optimum combination between the electric field and the

    magnetic field was obtained when the relative increase of the

    discharge current is the maximum value. It greatly saves the energy

    at this time.

    The relative increase of discharge current would exceed 150%,

    when the mean electric field intensity is about 4.2 kV/cm and the

    magnetic flux density is 0.43 T. The characteristic curve (Fig. 8) was

    generated by analyzing the data of the ionization-region magnetic

    field experiment, showing the relative increase of current asa function of the magnetic flux density for negative corona dis-

    charges under mean electric field intensity of 4 kV/cm, which is the

    value for the practical operation of the ESP. According to Fig. 8, the

    relative increase of the discharge current could attain the maxi-

    mum value by using a proper magnetic field when the mean

    electric field intensity is 4 kV/cm. The efficiency of the ESP will be

    enhanced at the maximum value in the ionization-region magnetic

    field because of the increased concentration of negative ions. The

    optimum combination between the mean electric field of 4 kV/cm

    and the magnetic field in the ionization region of about 0.43 T is

    especially significant in the industrial applications of corona

    discharges. Moreover, the relative increase of discharge current

    could reach about 150%.

    According to the mechanism of the magnetic enhancement, therelative increase of the discharge current is determined by two

    factors: (1) The collision number between the free electrons and

    the gas molecules in the ionization region; and (2) The mean

    energy of free electrons. If the mean energy of free electrons is

    beneath the minimum ionization energy of the gas molecules, most

    of the gas molecules will not ionize. Under that condition, the

    discharge current could not increase even if the collisions between

    the free electrons and the gas molecules increase. In contrast, when

    the number of the collisions decreases, the discharge current may

    not increase even if the mean energy of free electrons exceeds the

    minimum ionization energy of the gas molecules.

    In the ionization region, besides elastic collisions between the

    free electrons and the gas molecules, ionizations, excitations and

    attachments are present at the same time. If the mean electric fieldintensity is fixed, the mean energy of the free electrons will

    decrease when the magnetic field is used. Therefore, when the

    collisions between the free electronsand gas molecules occurin the

    ionization region, the probability of excitations and attachments

    increases, whereas the probability of ionization collisions decreases

    in the magnetically enhanced corona discharges. However, the

    number of collisions between the free electrons and gas molecules

    increases when the magnetic field is applied.

    According to above analyses, the mean energy of the free elec-

    trons decreases in the ionization region when the magnetic field is

    applied, while the number of the collisions between free electrons

    and gas molecules increases. Therefore, the optimum combinations

    between the electric fields and magnetic fields may be as follows.

    If the mean electric field intensity is fixed, the number ofcollisions between the free electrons and gas molecules will

    Fig. 7. Characteristic curves showing the relative increase of current as a function of

    inter-electrode mean electric field intensity for negative corona discharges underdifferent magnetic field intensities.

    Fig. 8. Characteristic curves showing the relative increase of current as a function

    of magnetic flux density for negative corona discharges under mean electric field of

    4 kV/cm.

    J. Mi et al. / Journal of Electrostatics 66 (2008) 457462 461

  • 7/29/2019 Magnetic Plasma

    6/6

    increase with the magnetic flux density increasing, while the

    mean energy of free electrons will decrease with the magnetic

    flux density increasing. Initially, due to the increased number of

    collisions (a decisive factor), the relative increase of the discharge

    current grows with the magnetic flux density. Subsequently, the

    relative increase of the discharge current attains a maximum

    value when the magnetic flux density is a particular value. Then

    the relative increase of the discharge current decreases with the

    increase of the magnetic flux density because of the obvious

    decreased mean energy of free electrons, which, in turn, becomes

    the decisive factor. Thus the relative increase of the discharge

    current would have a maximum value with the magnetic flux

    density increasing.

    Furthermore, if the magnetic flux density is fixed, the mean

    energy of free electrons increases with the mean electric field in-

    tensity in the ionization region, while the number of the collisions

    between the free electrons and gas molecules decreases with the

    mean electric field increasing. At first, due to the increased mean

    energy of free electrons, which is a decisive factor, the relative in-

    crease of the discharge current grows with the mean electric field

    intensity. After that, the relative increase of the discharge current

    attains a maximum value when the mean electric field intensity is

    a particular value. Then the relative increase of the discharge cur-rent decreases with the increasing of the mean electric field

    intensity because of the obvious decreased number of collisions,

    which becomes decisive factor. Thus, there would be a maximum

    value of relative increase of the discharge current with the mean

    electric field intensity increasing.

    4. Conclusions

    In this paper, according to the research on the corona discharges

    in the magnetic field, the following conclusions can be established:

    (1) The mechanism of the magnetic enhancement is that the

    magnetic field could affect the ionization region. Therefore, thecorona discharge current was enhanced by the Larmor move-

    ments of free electrons in the ionization region, and the space

    charge concentration is enhanced in the whole inter-electrode

    region. Moreover, the lengthening trajectories of free electrons

    and negative ions cannot increase the discharge current in the

    drift region.

    (2) The relative increase of the discharge current is much larger

    when the permanent magnet is applied near the discharge

    electrode than near the collecting electrode.

    (3) When the magnetic field is fixed, the relative increases of the

    discharge current for varied inter-electrode mean electric fields

    have a maximum value in the magnetically enhanced negative

    corona discharges.

    (4) In order to reach the largest relative increase of the discharge

    current under the inter-electrode mean electric field intensity

    of 4 kV/cm for a practical ESP, the optimum magnetic field with

    a magnetic flux density of about 0.43 T at the edges of magnet

    strings should be selected.

    (5) The relative increase of the discharge current is determined by

    two factors. One is the collision number between the free

    electrons and the gas molecules in the ionization region, and

    the other is the mean energy of the free electrons. Moreover,

    the changecharacteristic of the relative increase of the collision

    number is in contrast to that of the mean energy of free elec-

    trons, leading to optimum combinations between the electric

    fields and the magnetic fields.

    References

    [1] M. Yamamoto, M. Nishioka, M. Sadakata, Sterilization by H2O2 droplets undercorona discharge, J. Electrostat. 56 (2002) 173187.

    [2] K. Yu, J. Zou, Y. Ben, Y. Zhang, C. Liu, Synthesis of NiO-embedded carbonnanotubes using corona discharge enhanced chemical vapor deposition, Di-

    amond Relat. Mater. 15 (2006) 12171222.[3] J. Li, Z. Xu, Y. Zhou, Application of corona discharge and electrostatic force to

    separate metals and nonmetals from crushed particles of waste printed circuitboards, J. Electrostat. 65 (2007) 233238.

    [4] K. Yanallah, S. Hadj Ziane, A. Belasri, Y. Meslem, Numerical modeling of ozoneproduction in direct current corona discharge, J. Mol. Struct. 777 (13) (2006)125129.

    [5] M.M. Pauthenier, Moreau-Hanot, Charging of spherical particles in an ionizingfield, J. Phys. Radium 7 (1932) 590613.

    [6] B.Y.H. Liu, A. Kapadia, Combined field and diffusion charging of aerosol par-ticles in the continuum regime, J. Aerosol Sci. 9 (1978) 227242.

    [7] B.Y.H. Liu, D.Y.H. Pui, On unipolar diffusion charging of aerosols in the con-tinuum regime, J. Colloid Interface Sci. 58 (1977) 142149.

    [8] A. Hussin, H.G. Scheibel, K.H. Becker, J. Porstendorfer, Bipolar diffusioncharging of aerosol particles I: experimental results within the diameterrange 430 nm, J. Aerosol Sci. 14 (1983) 671677.

    [9] M. Adachi, Y. Kousaka, K. Okuyama, Unipolar and bipolar diffusion charging ofultra-fine aerosol particles, J. Aerosol Sci. 16 (1985) 109123.

    [10] A. Wiedensohler, H.J. Fissan, Bipolar charge distributions of aerosol particles inhigh-purity argon and nitrogen, Aerosol Sci. Technol. 14 (1991) 358364.[11] G.P. Reischl, J.M. Makela, R. Karch, J. Necid, Bipolar charging of ultrafine par-

    ticles in the size range below 10 nm, J. Aerosol Sci. 27 (1996) 931949.[12] A.V. Filippov, Combined field-diffusion charging of aerosol particles, J. Aerosol

    Sci. 27 (1996) 197198.[13] D. Xu, J. Li, Y. Wu, L. Wang, D. Sun, Z. Liu, Y. Zhang, Discharge characteristics

    and applications for electrostatic precipitation of DC corona with sprayingdischarge electrodes, J. Electrostat. 57 (2003) 217224.

    [14] D. Xu, L. Sheng, H. Wang, Y. Sun, X. Zhang, J. Mi, Study of magnetically en-hanced corona pre-charger, J. Electrostat. 65 (2007) 101106.

    [15] H.J. White, Industrial Electrostatic Precipitation, Addison-Wesley, Massachu-setts, 1963, pp. 319323.

    [16] X. Chen, L. Zou, Y. Huang, L. Ke, W. Chen, S. Huang, Assumption and verifi-cation of normal distribution model of electric field in wire-plate ESP, J. Wu-han Univ. Sci. Technol. 29 (4) (2006) 356359.

    J. Mi et al. / Journal of Electrostatics 66 (2008) 457462462