manual isa in situ spectral analyser

62
Version of the Manual: 5.1 en www.go-sys.de Manual ISA In Situ Spectral Analyser

Upload: others

Post on 13-Nov-2021

15 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Manual ISA In Situ Spectral Analyser

Version of the Manual: 5.1 en

www.go-sys.de

Manual ISA

In Situ Spectral Analyser

Page 2: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 2 / 62 www.go-sys.de [email protected]

Copyright This manual contains information which is the intellectual property of GO Systemelektronik GmbH. The user is obliged to use this information exclusively to run the instrument. It is not permitted to pass this information to third parties. Reproducing, copying, editing or extracting the manual contents is only allowed with the express permission of GO Systemelektronik GmbH. Changes GO Systemelektronik GmbH retains the right to modify the contents of the manual without prior notice. Liability exclusion GO Systemelektronik GmbH takes no responsibility for correct system operation under all possible operating conditions. It is not possible to guarantee that the software will function completely without error under all possible circumstances. GO Systemelektronik GmbH cannot therefore accept liability for direct or indirect damage resulting from system operation or the contents of this manual. Product observance Within the scope of our obligation for product observance GO Systemelektronik GmbH will endeavour to warn third parties about all identified dangers which could arise from the interaction between hardware and soft-ware and from the use of other components. Effective product observance is only possible with adequate information from the end user about the planned field of application and the hardware and software used. If the conditions of use change or if the hardware or software are changed, due to the complex relationships between hardware and software it is no longer possible to describe all possible dangers and their effects on the total system, in particular on our system. This manual does not describe every possible property and combination of the system. For further information, please contact GO Systemelektronik GmbH. Manufacturer’s declaration When installing the system it is necessary to ensure correct electrical connection, protection against moisture and foreign bodies and excessive condensation, and system heating which can arise from both correct and incorrect use. It is the responsibility of the installer to ensure that the correct installation conditions are pro-vided.

© GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Allemagne Tel.: +49 431 58080-0 Fax: +49 431 58080-11 www.go-sys.de [email protected]

Creation date: 14.7.2015 Described firmware version of the BlueBox: 2.78.36

Described software versions: AMS: 3.3.3.0 ISA Data: 3.3.3.0

Spectrum Visual: 3.3.4.0 Version of the manual: 5.1 en

Article number of the manual: DOC 486 6002-E-5.1-BDA File name: Manual ISA V5p1 en.pdf

Page 3: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 3 / 62 www.go-sys.de [email protected]

blank page

Page 4: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 4 / 62 www.go-sys.de [email protected]

Table of contents

1 Introduction .................................................................................................................................................................................................... 6 2 Overview .......................................................................................................................................................................................................... 6

2.1 Specifics DualBeam ................................................................................................................................................................................ 8 3 Scope of Delivery ............................................................................................................................................................................................. 9

3.1 Scope of Delivery ISA TS, ISA T4, ISA T4 DualBeam ............................................................................................................................... 93.2 Additional Scope of Delivery ISA mobil ................................................................................................................................................ 11

4 Technical Data .............................................................................................................................................................................................. 12 5 Tips before Starting ...................................................................................................................................................................................... 13

5.1 Safety Notices and Warnings ................................................................................................................................................................ 135.2 ATEX Notes ............................................................................................................................................................................................. 135.3 Setup Schematic ................................................................................................................................................................................... 15

6 Commissioning ............................................................................................................................................................................................. 15 7 Measurement Cycle ...................................................................................................................................................................................... 16 8 Menu Operation ............................................................................................................................................................................................ 17

8.1 The Spectrometer Menu ........................................................................................................................................................................ 178.1.1 Interval and Average ..................................................................................................................................................................... 188.1.2 Table and Diagram ....................................................................................................................................................................... 18

8.1.2.1 Table View .............................................................................................................................................................................. 198.1.2.2 Diagram View ......................................................................................................................................................................... 20

8.1.3 Sensor Information ........................................................................................................................................................................ 218.2 The Maintenance Menu ......................................................................................................................................................................... 228.3 Specifics DualBeam .............................................................................................................................................................................. 23

9 Operation with AMS ...................................................................................................................................................................................... 24

9.1 The AMS Start Window .......................................................................................................................................................................... 249.2 The Sensor Setup Window .................................................................................................................................................................... 25

9.2.1 The Configuration Window ........................................................................................................................................................... 279.2.2 The Spectra Window (Calibration Window) ................................................................................................................................. 29

9.2.2.1 The Views of the Spectra Window (Calibration Window) .................................................................................................... 319.3 Specifics DualBeam .............................................................................................................................................................................. 32

10 Saving the Spectra Data in the BlueBox ................................................................................................................................................... 34 11 Transmitting the Data to the PC ................................................................................................................................................................ 34

11.1 Transmitting the BlueBox Data with BlueBox SQL Software ........................................................................................................... 3511.2 Transmitting the BlueBox Data with ISA Data .................................................................................................................................. 36

12 Spectrum Visual .......................................................................................................................................................................................... 37

12.1 Call up and Display the Spectra ......................................................................................................................................................... 3712.2 Input of Reference Values and Probe Numbers ................................................................................................................................ 3912.3 Printing of the Displayed Spectra ...................................................................................................................................................... 4012.4 Converting of the Displayed Spectra into Other Data Formats ....................................................................................................... 4012.5 Exporting Spectra Data ...................................................................................................................................................................... 4112.6 Importing Spectra Data ...................................................................................................................................................................... 42

Page 5: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 5 / 62 www.go-sys.de [email protected]

13 Language Options ...................................................................................................................................................................................... 42 14 Applications ................................................................................................................................................................................................. 43

14.1 Example: Spectral Absorption Coefficient (SAC or SAK) at 254 nm .................................................................................................. 4314.2 Example: Chemical Oxygen Demand (COD or CSB) .......................................................................................................................... 4514.3 ISA Formula Examples ........................................................................................................................................................................ 47

15 Parameter Accuracy ................................................................................................................................................................................... 49 16 CAN-bus Connection and Termination ...................................................................................................................................................... 50 17 Connections of the External Sensor Module ............................................................................................................................................. 51 18 Connection of the Compressed Air Line ..................................................................................................................................................... 51 Appendix A – The Configuration Data Sheet .................................................................................................................................................. 52 Appendix B – Connector Pin Assignment at the BlueBox in the Version TS ................................................................................................. 55 Appendix C – ISA mobile .................................................................................................................................................................................. 58

1 Overview ................................................................................................................................................................................................... 582 Technical data .......................................................................................................................................................................................... 593 Startup ...................................................................................................................................................................................................... 604 Removal from service ............................................................................................................................................................................... 625 Label with access data ............................................................................................................................................................................ 62

Page 6: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 6 / 62 www.go-sys.de [email protected]

1 Introduction

This manual describes the operation at work of the In Situ Spectral Analyzer (called ISA) from GO Systeme-lektronik. The spectral analyzer is available in four versions:

• ISA TS BlueBox with integrated spectrometer sensor module

• ISA T4 BlueBox with external spectrometer sensor module

• ISA T4 DualBeam BlueBox with an external spectrometer sensor module with an integrated additional independent reference optical measurement path

• ISA mobile ISA TS with two accumulator batteries and a power up control module, integrated in a suitcase

The operation performs at the display of the BlueBox (menu operation) and with the BlueBox PC Software∗ incl. the Spectrum Visual program for the graphical representation of the spectra. This manual describes only the spectral analyzer specific operation. The general operation of the BlueBox and the BlueBox PC Software is described in the attached manuals: Manual BlueBox T4 und Manual BlueBox PC Software This manual does not describe the commissioning, maintenance and service. This is described in the at-tached Manual ISA and Process Spectrometer Commissioning – Maintenance –Service.

2 Overview

The ISA (In Situ Spectral Analyzer) is a spectrometer which works in the wavelength range between 200 and 708 nm (UVVIS). The result of a single measurement is an absorption spectra over the whole wavelength range. A spectrometer is a very multi-purpose measuring instrument because it can be implemented in very diverse applications through a relevant calibration. Multi-parameter measurements are also possible. This is a great advantage in comparison to electro-chemical sensors. Moreover, ISA offers an adjustable measuring-path-length whereby the number of potential applications is increased even more. A great advantage of ISA is the special coating of the glass plates in the measuring-path. Combined with the possibility to automatically clean the measuring section with compressed air, ISA achieves very long in- ser vice times and maintenance intervals. Only the optics and the compressed air cleaning are integrated into the high grade steel (material number 1.4404 or Titanium) sensor head. The entire control- and analysis-electronics are mounted in the accompany-ing sensor module. ISA is thereby usable in a high temperature range (up to 110°C). I SA is thus suitable for use in medicine or in the food industry, since the measuring head can be sterilized at high temperatures. ISA can be used in potentially explosive areas, for details, see 5.2 ATEX Notes.

Note: A separate manual for commissioning, maintenance and service is attached. Manual ISA and Process Spectrometer Commissioning – Maintenance –Service

∗ especially with the software AMS

Page 7: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 7 / 62 www.go-sys.de [email protected]

The properties of ISA in summary:

• optical sensor linked by fibre optic to the analysis unit

• recording across the complete absorption spectrum 200 nm – 708 nm

• optional independent reference optical measurement path (DualBeam)

• suitable for a temperature range from 0 °C to +110 °C

• sensor adaptable to different media with an easy to adjust measuring-path-length (0.5 – 20 mm)

• saving of raw data and calibrated data

• calibration adjustment by telemetry

• software for calibration and service

• cleaning of measuring-path with compressed air

• easy installation

• high cost-efficiency

Device scope:

• purely optical sensor head with sensor head cable (integrates two fibre optic cables and a com-pressed air line), available lengths of the sensor head cable : 2.5 m; 6 m; 10 m

• freely adjustable measuring path from 0.5 to 20 mm

• capture of absorption spectra in the range 200 nm – 708 nm

• telemetrical input of calibration algorithms for parameter calculation

• comprehensive software to carry out calibrations

• multi-parameter calculation

• automatic cleaning via compressed air

• process control

• waste water monitoring

• water quality monitoring

Example parameters:

• Nitrate: for NO3/NO3-N measurements Measurement range 0.1 - 100 mg/l NOx_eq in water bodies (other measurement ranges are possible)

• Carbon compounds TOC/COD: Calibration by comparison analysis. Resolution and accuracy depend on the stability of the water-matrix and its ability to be analysed.

Page 8: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 8 / 62 www.go-sys.de [email protected]

2.1 Specifics DualBeam Via a beam splitter, a reference beam with 5% of the intensity is generated and directed to a photometer. The measured intensity can be queried via the AMS∗ software.

• Allows cross calculation with a reference parameter.

• Allows control of the system stability during spectrometer operation.

∗ AMS is part of the BlueBox PC Software.

Reference beam 5 %

Probe Spectrometer

Calculationwith AMS

Xenon lamp

Beam splitter

Photometer

Page 9: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 9 / 62 www.go-sys.de [email protected]

3 Scope of Delivery The spectral analyzer is available in four versions:

• ISA TS BlueBox with integrated spectrometer sensor module

• ISA T4 BlueBox with external spectrometer sensor module

• ISA T4 DualBeam BlueBox with an external spectrometer sensor module with an integrated additional independent reference optical measurement path

• ISA mobile ISA TS with two accumulator batteries and a power up control module, integrated in a suitcase

3.1 Scope of Delivery ISA TS, ISA T4, ISA T4 DualBeam 1. BlueBox

If the spectrometer will not be integrated into an existing BlueBox system, the BlueBox is part of the deliv-ery. The BlueBox is the central control unit of ISA. In the BlueBox measured data are saved and for-warded, the values for the desired parameters are calculated and calibration parameters are stored. Via the PC interface all ISA data and settings can be remotely controlled and read out for further processing. The BlueBox is delivered with all the required accessories. Information about the properties, installation and use of the BlueBox can be found in its accompanying manual.

2. Sensor head The ISA sensor head is manufactured from high grade steel (optional titanium). Only the optics and the compressed air cleaning are integrated in the measuring head. The sensor head can thereby be deployed in high temperature ranges (up to +110 °C).

3. Sensor head cable The sensor head is connected with the spectrometer sensor module via the special covered sensor head cable. The spectrometer sensor module contains all the electronics. In the measuring head cable there are two fibre optic cables and a compressed air line. The probe cable must not be bent or kinked in a tighter radius than 40 mm. At installation the sensor head shall not be hung up at the measuring head cable; therefore use the appropriate lugs.

4. Sensor module with compressed air connector (only ISA T4 and ISA T4 DualBeam). The entire ISA control- and analysis-electronics are fitted in the Sensor module.

5. Software USB-Stick with the system software and the manuals

6. USB-Dongle Protection against unauthorized access

7. Configuration data sheets BlueBox/Spectrometer sensor module

8. Spectrometer data sheet from Zeiss

9. CAN-bus-cable (only ISA T4 and ISA T4 DualBeam)

Page 10: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 10 / 62 www.go-sys.de [email protected]

Scope of Delivery ISA TS Article-No. 486 0060

Scope of Delivery ISA T4 Article-No. 486 0004 + 486 6002/6004

Page 11: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 11 / 62 www.go-sys.de [email protected]

Scope of Delivery ISA T4 DualBeam Article-No. 486 0004 + 486 6002/6004

3.2 Additional Scope of Delivery ISA mobile

1. Case with a integrated combination of ISA TS with a BlueBox Power Up Control inclusive two 12 V ac-cumulator batteries

2. 12 V DC battery charger for the accumulator batteries with manual

3. 2 spare fuses 4 A, medium time lag

See Appendix C – ISA mobile

Article-No. 486 0M60

Page 12: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 12 / 62 www.go-sys.de [email protected]

4 Technical Data In Situ Spectral Analyser (ISA) ISA is an in situ submersible spectrum analyser. Wavelength range 200 to 708 nm, resolution 2 nm Measurand UV VIS spectra in the range 200 – 708 nm Measurement principle spectrum analysis Measurement interval adjustable, min. 60 s Light source Xenon flash lamp Sensor module Power supply 24 VDC (18 – 36 VDC) via CAN-bus-cable Compressed air connector quick connector for 4 mm PU-tube, 4 – 6 bar Temperature range 0 °C to +40 °C Weight approx. 2.6 kg Housing material aluminium Dimensions 280 mm x 170 mm x 90 mm Protection class IP65 Sensor head Material high grade steel (material number 1.4404) – Titanium optional Temperature range 0 °C to 110 °C Weight 1.5 kg

102 mm 102 mm

Ø 4

4 m

m

measuring-path, adjustable between 1 und 20 mm

1. glass plate in the measuring-path / 2. glass plate on the opposite

lug lug

connection cable

approx. 25 mm

Page 13: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 13 / 62 www.go-sys.de [email protected]

5 Tips before Starting 5.1 Safety Notices and Warnings Please read carefully the operating manuals before starting the BlueBox and ISA. Keep these manuals at hand. Do not pass the BlueBox or ISA on to other persons without the user manuals. The manufacturer does not take responsibility for incorrect handling or operation. The instrument is constructed according to the low voltage code and to the safety regulations for electronic measuring devices. Correct function and safety can only be ensured when both general and system-specific safety measures are followed. Before connecting the power supply, make sure that the power supply voltage is suitable. Correct and safe function can only be maintained under the climatic conditions specified in chapter 4 Techni-cal data. If the instrument is moved from a cold to a warm environment, condensation might form which could influ-ence its function. In this case, wait for the instrument to reach equilibrium with the new surroundings before use. Maintenance and repairs may only be carried out by GO-authorised technicians. If it is suspected that the instrument cannot be used without danger, it must be shut down and prevented from further use. The user’s safety might be affected if the instrument shows signs of damage, does not function properly, has undergone long storage under unsuitable conditions or was subject to extreme transport conditions. If in doubt, contact the manufacturer "GO Systemelektronik GmbH” and send the instrument for repair or maintenance if necessary.

The fibre optic cable must not be bent or flexed with a radius less than 40 mm.

The sensor head must not be hung on the sensor head cable, please use the lugs for installation of the Sensor head.

Page 14: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 14 / 62 www.go-sys.de [email protected]

5.2 ATEX Notes Guideline 94/9/EC, known as the ATEX directive, requires in Annex II to the fulfilment of basic safety require-ments for devices that are provided within the EU for use in explosive areas.

The sensor head of the spectrometer has the following ATEX-characterisation∗ :

II 3/- G Ex op is IIA T4 Gb/-

The BlueBox respectively the sensor module must necessarily be located outside the explosive area.

The power supply may not fit against the housing of the BlueBox respectively of the sensor module. The housing of the BlueBox respectively of the sensor module must always be grounded.

The electrical resistance between the lower lug of the spectrometer sensor head and the ground screw of the BlueBox respectively of the sensor module must be less than 50 Ω.

Parameter: Electrical data: maximal input voltage of the sensor module: 36 V DC Range of ambient temperature: Sensor head: 0 °C to +110 °C Sensor module: 0 °C to +40 °C Special conditions for safe use: The ambient temperature of the sensor is 0 °C to +110 °C. The sensor module must be installed outside of hazardous areas. The ambient temperature of the sensor module is 0 °C to +40 °C. The basic safety and health requirements are fulfilled by compliance with: EN 60079-0:2009 General requirements EN 60079-28:2007 Optical Radiation 'op is' Identification of the measuring head: Laser engraving Identification of the BlueBox/sensor module: abrasion resistant label on the outside right

∗ If a separate ATEX certificate is attached, this certificate is valid. II 2/- G Ex op is IIB T4 Gb/- is available on request.

Page 15: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 15 / 62 www.go-sys.de [email protected]

5.3 Setup Schematic In the diagram below is a schematical representation of module interconnections. The Sensor module should be mounted close to the place of measurement. The maximum distance between Sensor module and Sensor head is determined by the length of fibre optic cable (depending on version up to 10 m). The Sensor module and the BlueBox are in version ISA T4 connected by CAN-bus cable (type of cable: LiYCY 2x2x0.5 mm² with an impedance of approx. 100 Ω). Cable lengths up to 200 m are possible∗, so that the BlueBox as central control unit of the measuring system can be mounted in a control station. A PC (or Notebook) with BlueBox system-software installed is necessary for the configuration of the system. For direct connection with a PC/Notebook use a standard or crosslink network-cable.

6 Commissioning The commissioning of the ISA spectrometer is described in detail in the attached Manual ISA and Process Spectrometer Commissioning - Maintenance - Service.

∗ Consider the voltage drop.

Page 16: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 16 / 62 www.go-sys.de [email protected]

7 Measurement Cycle

1. Cleaning (only when compressed air cleaning is active) Compressed air cleaning of the sensor head optics (Duration: Parameter Cleaning time)

2. Wait (only when compressed air cleaning is active) The wait time after flushing serves to ensure that air bubbles can dissipate. (Duration: Parameter Waiting time)

3. Dark measuring A dark measuring quantifies the dark current and is used to compensate individual system properties. The dark measurement is factory preset and cannot be changed.

4. Heating The Xenon lamp is heated up through a few light flashes. (Number of flashes: Parameter Heating).

5. Measuring The period in which an actual measurement occurs. It consists of an individual measurement (Number of light flashes per single measurement: Parameter Intensity) repeated a number of times (Number of repeti-tions: Parameter Average). The final measurement result is the arithmetic mean of the individual meas-urements.

6. Idle time The time until the beginning of the next measurement cycle.

The input of settings during the measuring process time is not possible.∗

∗ Except in extreme cases there is enough time for the input of settings after the end of the measuring process time to the begin ing of the next measurement cycle. The measurement process period is calculated from the Heating period, Dark measuring period and Measuring period. The Heating period is generally negligible. The dark measurement is negligible too, because it takes place only once. Heating + Dark measuring + (Intensity x Average) 100 light flashes have a duration of approx. 1.6 s.

Measuring timeInitialisation

time

Idle time

Measuring process time

Process time

Interval

Measurement cycle without and with compressed air cleaning

newInterval

newInterval

Measuring timeInitialisation

time

Idle time

Measuring process time

Heating

Dark measuring

Measuring

Heating

Dark measuring

Measuring

CleaningWait

Page 17: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 17 / 62 www.go-sys.de [email protected]

8 Menu Operation 8.1 The Spectrometer Menu

Switches to the input of the measurement interval. The minimum interval is 60.

Switches to the input of the measurement average: Number of measurements whose arith-metic average is the measured value.

Switches to the maintenance menu. See 8.2 The Maintenance Menu

Switches to the table display.

Switches to the diagram display.

Switches to the sensor information.

Selects the sensor for the multiple parameter display or not.

Switches back to the previous menu.

Menu0

30000

Spectrometer

21677

04:19:5221.11.14

<--

System

Sensors

Actuators

User

Ser vice Help

09:19:3927.08.14

Main menu

<--

Inter val

Average

Calib ration

Diagram

Info

Table selected

<--

1/2Spectrometer

Calculated

BSA00123-A

SAK254

BSA00123-B

Spectrometer

Inter val

Average

Calib ration

Table

Diagram

Info

Selected

Selected

<–

Spec trometer menu

Page 18: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 18 / 62 www.go-sys.de [email protected]

8.1.1 Interval and Average

Spectrometer menu

Spectrometer menu

Input interval in Seconds Input average Minimum interval is 60 s

The current value is displayed.

Deletes the last entered character.

Saves the value and switches to the previous menu.

Switches to the previous menu without saving the value.

8.1.2 Table and Diagram The spectrometer value displayed at the parameter display is the MVR, the Maximum digital Value of a Raw spectrum.

These values are displayed in a table and in a chart.

Inter val

Average

Inter valAktual value 60

1 2 3

4 5 6

7

0

8 9

<

<–Ok

AverageAktual value 6

1 2 3

4 5 6

7

0

8 9

<

<–Ok

<

Ok

<–

Menu0

32000

Spectrometer

21677

04:19:5221.11.14

Page 19: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 19 / 62 www.go-sys.de [email protected]

8.1.2.1 Table View

Spectrometer menu

Displays the previous day's values.

Displays the following day's values.

Scrolls the list.

Switches to the spectrometer menu.

∗ maximum digital Value of a Raw spectrum

Table

Table

Statistics List

Statistics

min. max. mittl .

1h 18230 18340 18785

6h 18234 18340 18787

12h 18236 18340 18788

24h 18228 18340 18784

7d 18230 18340 18785

Spectromter [ ]

<--

List

10:06:12 18755

10:05:12 18760

10:04:12 18754

10:03:12 18748

10:02:12 18753

10:01:12 18752

Spectrometer [ ]

<--< >

14.03.15

<

>

<–

Statistics for the MVR* over the last 24 hours and the last 7 days

Daily maximum values of the MVR*

Page 20: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 20 / 62 www.go-sys.de [email protected]

8.1.2.2 Diagram View

Spectrometer menu

These views are only overviews. Not all maximum values are displayed.

∗ maximum digital Value of a Raw spectrum

Diagram

Diagram

Auto scale Range

Auto scale

Spectrometer 24h

13 16 19 22 01 04 07 10 13

22460

22440

22430

22420

22410

22400

22390

22380

22370

22360

22350

Range

Spectrometer 24h24000

19200

14400

9600

4800

0

13 16 19 22 01 04 07 10 13

Maximum values of the MVR* over the last 21 – 24 hours (de-pending on the value range), scaled to the min/max of all the measured values

Pressing on the Display switches to the spectrometer menu.

Maximum values of the MVR* over the last 21 – 24 hours (depending on the value range)

Pressing on the Display switches to the spectrometer menu.

Page 21: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 21 / 62 www.go-sys.de [email protected]

8.1.3 Sensor Information

Spectrometer menu

Comment general comment Sensor serial no. CAN-ID∗ of the spectrometer (8-digit designation of the Data Acquisition

Module) + consecutive number of the sensor (0-9) Interval measurement interval of the spectrometer Average measurement average Resolution default value is 1 no decimal places Parameter name of the parameter, here nothing Unit unit of the parameter, here nothing Last update date of the last measurement

Switches back to the spectrometer menu.

∗ In other contexts also called DAM-ID.

Info

Spectrometer

Comment UV-VISSensor se rial no. isa000011Inter val 60Average 5Resolut ion 1ParameterUnitLast update 15:51:34 18.01.15

<–

Page 22: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 22 / 62 www.go-sys.de [email protected]

8.2 The Maintenance Menu

Switches to the menus of the clear water calibration. See Manual ISA and Process Spectrometer Commissioning – Mainte-nance –Service there 4.2.6 Clear Water Calibration

Switches the compressed air cleaning of the spectrometer sensor head on and off. The button is also a status indicator.

Switches to the input menus of the interval of the compressed air cleaning: 1 compressed air cleaning before each measurement, 2 compressed air cleaning before each second measurement and so on.

Opens the input menu of the intensity.

Switches to the input menu of the measurement path length of the sensor head in mm. Only visible if the path length is used in an AMS formula (variable name ISA.Pathlength). The first input is performed in the AMS configuration window (see 9.2.1 The Configura-tion Window).

Opens the display of the current absorption spectrum.

Menu0

30000

Spectrometer

21677

04:19:5221.11.14

<--

System

Sensors

Actuators

User

Ser vice Help

09:19:3927.08.14

Main menu

<--

Spektrom ter

Inter val

Average

Calib ration

Diagram

Info

Table selected

<--

1/2Spectrometer

Calculated

BSA00123-A

SAK254

BSA00123-B

Password

****

1 2 3

4 5 6

7

0

8 9

<

<–

Spectrometer

ISA DI-water calib ration

ISA Purging inter val

Path lengthIntensity spec

ISA Purging on

<–Abs spectrum Raw spectrum

e

ISA DI-water calib ration

ISA Purging on

ISA Purging off

ISA Purging inter val

Intensity spec

Path length

Abs spectrum

Maintenance menu

Page 23: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 23 / 62 www.go-sys.de [email protected]

Opens the display of the current raw spectrum.

Switches back to the spectrometer menu.

8.3 Specifics DualBeam Not any

Raw spectrum

<–

Page 24: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 24 / 62 www.go-sys.de [email protected]

9 Operation with AMS 9.1 The AMS Start Window Start the program AMS, e.g. via the program BlueBox SQL like here. The AMS start window opens:

Depending on the sensor status the sensor icon has a different colour.

green – The sensor works.

grey – The name of the sensor has been set with AMS to "n/c", therefore the sensor is not active.

red – sensor error

blue with magnifying glass – waiting for the first measured value

1. Selection of the BlueBox

3. Double-click on "Spectrometer", or similar

2. Password input. The password can be found in the configuration data sheet BlueBox under "2. Network". Press the RETURN key.

• (green circle) The password is valid.

• (grey circle) The password is invalid.

Page 25: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 25 / 62 www.go-sys.de [email protected]

9.2 The Sensor Setup Window This opens the sensor setup window. The factory default values are specified for the spectrometer. Usually, the user has only to change the interval and the average.

Opens the configuration window of the spectrometer. See 9.2.1 The Configuration Window

Switches to the spectra window (calibration window). See 9.2.2 The Spectra Window (Calibration Window)

Name1 Sensor name, also retrieved by the other BlueBox programs.

Comment1

An entry in the comment field • determines how a measured value will be saved in the database

and • is an user defined comment text for AMS and the BlueBox SQL Software.

Parameter1 Name of the measured parameter

Unit2 Unit of the output value

1 maximum 20 characters 2 More than 5 characters can’t be displayed at the BlueBox display.

formula input field (see Manual AMS)

MVR* of the last recorded raw spectrum, used for function control

formula entry buttons (see Manual AMS)

Page 26: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 26 / 62 www.go-sys.de [email protected]

Digits before Factory preset, 5 digits before the decimal point are displayed.

Digits after

Factory preset, 0 digits after the decimal point are displayed. Since here integer counts are measured, there are no digits after the decimal point.

Interval

Time period between the measurement results Minimum 60 s

Min. Value Lowest expected measuring value, factory preset 0

Max. Value Highest expected measuring value, factory preset 32000

Average The number of single measurements from which the arithmetic mean is derived. This arithmetic mean is the recorded measured value.

Wait until the end of a measurement before making an input!

Page 27: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 27 / 62 www.go-sys.de [email protected]

9.2.1 The Configuration Window

Sensor setup window

The input of settings during the measuring process time is not possible. See 7 Measurement Cycle

Cleaning ON/OFF Turns the automatic compressed air cleaning on and off. The button is also status indicator.

Cleaning time Duration of compressed air cleaning in seconds

Cleaning interval Interval of the compressed air cleaning: 1 Compressed air cleaning before each measurement, 2 Compressed air cleaning before each second measurement and so on.

Wait time To ensure that air bubbles from a compressed air cleaning do not interfere with a meas-urement, a wait time in seconds can set here. This is the time between the end of the compressed air purging and the measurement.

Note regarding the compressed air cleaning: In most applications, it will be advisable to use the com-pressed air cleaning of the spectrometer. In this case it is necessary to connect a compressed air line to the allocated plug-in connection of the BlueBox respectively of the sensor module.

Intensity Number of light flashes before the measurement

Path length [mm] Input of the measurement path length of the sensor head in mm. The path length is measured e.g. with a vernier calliper; take care not to damage the optics. If the path length is used in an AMS formula (variable name ISA.Pathlength), you must enter the path length mandatory here. Otherwise, the software uses a default value (10 mm).

Heating Number of light flashes before the measurement

Page 28: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 28 / 62 www.go-sys.de [email protected]

Zeiss-Coefficients Base settings of the optics. Do not change.

Saves the Zeiss-coefficients. In the case that the Zeiss-coefficients need to be changed, the Checksum field serves to ver-ify the inputted values. The Zeiss-coefficients and their associated Checksum can be found on the Zeiss datasheet.

Checksum

Page 29: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 29 / 62 www.go-sys.de [email protected]

9.2.2 The Spectra Window (Calibration Window)

Sensor setup window

View raw spectra:

View of the raw spectra: The last five raw spectra are displayed. Spectrum ① is a raw spectrum before, spectrum ② is a raw spectrum after a intensity calibration. View absorption spectra:

View of the absorption spectra: The last five absorption spectra are displayed.

2

1

Page 30: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 30 / 62 www.go-sys.de [email protected]

Deletes the spectra display.

Switches between the view of the raw spectra and the absorption spectra back and forth.

Prints the graph.

Saves the graph as a pixel image in jpg-format (795 x 427).

Performs an intensity calibration.

Switches the display of the fingerprint on and off. only in absorption spectra view

Performs a zero calibration. only in absorption spectra view

The spectrometer measures raw spectra, from the raw spectra, the system calculates the absorbance spectra. In an absorbance spectrum of the degree of absorption of the individual wavelengths is represented with a logarithmic scale. The maximum value is 4.51.

𝑒𝑖 = −𝑙𝑜𝑔10 𝑠𝑝𝑒𝑐𝑡𝑟𝑢𝑚𝑖

𝑐𝑙𝑒𝑎𝑟 𝑤𝑎𝑡𝑒𝑟 𝑠𝑝𝑒𝑐𝑡𝑟𝑢𝑚𝑖

𝑖=255

𝑖=0

Using the Zeiss coefficients (see before the previous page) to each even wavelength of 200 nm to 708 nm there is assigned a value. Uneven wavelengths get the value of the preceding even wavelength.

Page 31: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 31 / 62 www.go-sys.de [email protected]

9.2.2.1 The Views of the Spectra Window (Calibration Window) View of the raw spectra Display of the last five raw spectra

Button up: view of the raw spectra Button down: view of the extinction spectra

View of the extinction spectra Display of the last five extinction spectra

Button up: view of the extinction spectra Button down: view of the extinction spectra and of the fingerprint

View of the fingerprint Additional to the extinction spectra the last stored fingerprint is shown. The fingerprint is an additionally stored spectrum. It is saved via the drop-down menu under Application. Click on <Save fingerprint> stores the extinction spectrum of the last measurement under the term fingerprint. The values of this fingerprint can be can be accessed with AMS-formula∗, i.e. as reference values.

∗ see Manual BlueBox PC Software and 14.3 ISA Formula Examples there ISA fingerprint difference

Page 32: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 32 / 62 www.go-sys.de [email protected]

9.3 Specifics DualBeam The values of the DualBeam photometer can be accessed via the AMS formula entry "ISA0(1);"1. In the following example, the query is performed in a virtual (calculated) sensor with the name "Dual Beam". The corresponding sensor setup window is opened by double-clicking on the icon. See 9.1 The AMS Start Window Example: Sensor setup window DualBeam query in a virtual sensor

The measured values of the DualBeam photometer are the summed digital raw (Counts) of the xenon lamp flashes of at the A/D converter of the photometer. Range = 0-50000 With the (active) formula entry "ISA0(1);"1 the measured values are transmitted to the virtual sensor.

Opens the calibration window. See Manual AMS

Name2 Sensor name, also retrieved by the other BlueBox programs.

Comment2

An entry in the comment field • determines how a measured value will be saved in the database

and • is a user defined comment text for AMS and the BlueBox SQL Software.

Parameter3 Name of the measured parameter

1 Please note, that in systems without DualBeam the entry "ISA0(1);" queries the raw value of the first pixel of a current spectrum. 2 maximum 20 characters 3 More than 5 characters can’t be displayed at the BlueBox display.

Formula input field (see Manual AMS)

Display of the current digital raw value (count) of the photometer

Buttons for the formula input (see Manual AMS)

Formula input is active

Page 33: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 33 / 62 www.go-sys.de [email protected]

Unit4 Unit of the output value, empty since the value is dimensionless.

Digits before Factory preset, 5 digits before the decimal point are displayed, results from the value range.

Digits after

Factory preset, 0 digits after the decimal point are displayed. Since here integer counts are measured, there are no digits after the decimal point.

Interval

The interval is already predetermined from the photometer setting, therefore is here the standard entry 1.

Min. Value Lowest expected measuring value, results from the value range.

Max. Value Highest expected measuring value, results from the value range.

Average The average is already predetermined from the photometer setting, therefore is here the standard entry 1.

4 More than 5 characters can’t be displayed at the BlueBox display.

Page 34: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 34 / 62 www.go-sys.de [email protected]

10 Saving the Spectra Data in the BlueBox What spectra data will be saved in the BlueBox is determined by the adjustment Save Mode in the drop-down menu under Applications in the calibration window (see 11.2).

These settings are the same as in the Save Mode option of ISA Data (see 11.2 Transmitting the BlueBox Data with ISA Data). 11 Transmitting the Data to the PC There are two ways of transmitting the spectra data from the BlueBox to the PC.

• Transmitting with BlueBox SQL software Here the whole data of the BlueBox are transmitted to the PC.

• Transmitting with ISA Data Here only the raw spectra and the extinction spectra are transmitted. It is possible to transmit only se-lected spectra.

Both ways save the data in a SQL data base. The memory location was determined at the installation of the SQL-server.

Off (no spectrum)

No spectra will be saved.

Extinction spectrum

Only extinction spectra will be saved in the BlueBox.

Raw and extinction spectrum

Raw spectra and extinction spectra will be saved in the BlueBox.

Page 35: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 35 / 62 www.go-sys.de [email protected]

11.1 Transmitting the BlueBox Data with BlueBox SQL Software For to transmit the data from the BlueBox to the PC you have to call up <Setup>< Get Data…> in the program BlueBox SQL.

When checking these items in the sensor selection field the following data will be transmitted to the PC:

configuration data of the BlueBox

the measurement data of all connected sensors (in case of ISA for each the MVR∗ of the raw spectrum)

the raw spectra and the extinction spectra

the measurement data of the selected sensors (in case of ISA for each the MVR* of the raw spectrum)

∗ MVR = Maximum digital Value of a Raw spectrum

If this check box is activated, all those values are transmitted automatically, which had been recorded since the point in time of the last data transmission. This point in time is determined by the topical data set of the data base.

Only for measurement data. If this check box is activated, only those values are transmitted, which differs from the previous value for a certain degree. The value of this difference is determined in: BlueBox SQL <Options><Sensor De-tails…> "Data transmission level" (see Manual BlueBox SQL).

Only active, if the connection is established via modem or gateway.

Transmits the data to the PC.

BlueBox selection

Sensor selection field example

to

only

at d

isab

led

Auto

mat

ic

"All extended data" must be checked, checked, otherwise no spectra are transferred.

from

Page 36: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 36 / 62 www.go-sys.de [email protected]

11.2 Transmitting the BlueBox Data with ISA Data To transfer the spectra from the BlueBox to the PC/Notebook call up the program ISA Data.

A list then appears containing all the spectra in the BlueBox for the selected time period. Individual spectra, or all of them, can be selected for the data transfer to the PC/Notebook. Red marked data is already stored, black marked data not. The listing will be actualized with click on .

Inverts the selection.

Selects all.

Transfers the selected data to the data base.

Select the BlueBox.

Select the period.

from to

Page 37: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 37 / 62 www.go-sys.de [email protected]

Additional you can choose under <Setup><Save Mode>, which spectral data is stored on the BlueBox.

These settings are the same as in AMS (see 10 Saving the Spectra Data in the BlueBox). 12 Spectrum Visual 12.1 Call up and Display the Spectra The tool "Spectrum Visual" is available for the graphical display of spectra.

Select the spectra in the list that should be graphically displayed.

Extinction Spectrum

Only extinction spectra will be saved in the BlueBox.

Raw- and Extinction Spectrum

Raw spectra and extinction spectra will be saved in the BlueBox.

Off No spectra will be saved.

Raw Spectrum View of the extinction spectra

Raw Spectrum View of the raw spectra

1th Derivation Displays the graphs of the first derivation. Has no influence on the exported spectra data.

Selection BlueBox

Selection Spectrometer

Selection time period

Selection spectra

Selection Timezone

Selection View

Page 38: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 38 / 62 www.go-sys.de [email protected]

Reload: If in the meantime new spectra were stored in the database, click on Reload loads the new spectra. Spectra that are already selected remain activated.

Resets the zoom view to not zoomed.

Moves the selected time period by one day ahead.

Moves the selected time period by one day return.

Selects all entries that have reference values and/or probe names (see 13.2).

You can zoom the spectra diagram in and out by drawing a rectangle to the right or to the left with the mouse while pushing the shift button and the left mouse button. Click on resets the zoom view to not zoomed.

Page 39: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 39 / 62 www.go-sys.de [email protected]

12.2 Input of Reference Values and Probe Numbers

Double-click on an item in the list of spectra opens an input window:

Here you can enter the reference value of a reference measurement. The entered decimal point (comma or period) depends on the language of the operating system, but there is always saved a point, false entries are ignored. The reference value is needed in the application-specific calibration (see 10) in connection with a one-parameter calibration, and is stored together with the spectral data at an export in the JCAMP-DX format (see 13.5). Pressing Ctrl + double-clicking on an item in the list of spectra opens an input window:

Here you can enter the Probe number∗ of a reference measurement, commas entered are stored as a point. This probe number is required in the application-specific calibration in connection with a multi-parameter calibration, and is stored together with the spectral data at an export in the JCAMP-DX format (see 12.5).

∗ Also characterised as Probe name. Character set: ASCII standard

Page 40: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 40 / 62 www.go-sys.de [email protected]

12.3 Printing of the Displayed Spectra Click on <Print> prints the displayed spectra on the selected printer.

12.4 Converting of the Displayed Spectra into Other Data Formats The saved spectra can be converted into the following formats: Picture .bmp (Bitmap) .emf (Metafile) .jpg (JPEG)

Native .tee

Data .txt XML html (Table) Excel

Page 41: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 41 / 62 www.go-sys.de [email protected]

12.5 Exporting Spectra Data You can export the selected spectral data of all spectra as a file. Click on <File><Export> opens a selection menu. JCAMP-DX (with the extension. jdx) is a standardized file format for exchanging spectra and related chemical and physical information between systems from different manufactures. In connection with the ISA spec-trometer, the JCAMP-DX files are needed for calculating the calibration coefficients in the application-specific calibration (see chapter 10). After clicking on <JCAMP-DX> an input window appears. Here you can enter the unit of reference values5. The unit of reference values is needed in the ap-plication-specific calibration (see Chapter 10) in connection with a one-parameter calibration, in all other cases, press Enter without <OK>. In the next window you specify name and memory location of the file. The file will then be saved with the file extension .jdx. The spectral data of the selected spectra are then summarized in a .jdx file.

Text Format The spectral data of each selected spectrum will be stored in a single .txt file. B+L Format The spectral data of each selected spectrum will be stored in a single .asc file. The modification date6 of the file is the recording time point of the spectrum. BlueBox Database The spectral data of all selected spectra are stored in one .isa file. With this .isa file the spectral data could be imported into another BlueBox database (see 13.6 Importing spec-tra data into the data base). 5 Character set: ASCII standard 6 not the creation date

Page 42: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 42 / 62 www.go-sys.de [email protected]

12.6 Importing Spectra Data Files in the format of the BlueBox database (extension .isa, see previous page) allow to transfer spectral data from other BlueBox databases into the current database.

13 Language Options

Page 43: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 43 / 62 www.go-sys.de [email protected]

14 Applications The ISA Spectrometer offers by means of AMS (Advanced Managing Software) the possibility to program so called virtual Sensors. In combination with ISA numerous possibilities can be realized. With the help of two examples the analytical possibilities of ISA will be shown:

• Spectral Absorption Coefficient (SAC) at 254 nm

• Chemical Oxygen Demand (COD)

14.1 Example: Spectral Absorption Coefficient (SAC or SAK) at 254 nm The Spectral Absorption Coefficient (SAC) allows assertions to be made concerning the contamination level of water bodies by organic materials. The basis for this example is a measurement according to DIN 38404-3 (2006). The light attenuation per meter of water (unit 1/m) at 254nm is measured. Light attenuation at 254nm is a measure of the organic material contamination level of water bodies because many organic materials absorb ultraviolet light. A virtual sensor for the measurement of SAC is already pre-programmed.

Select SAK254 with double-click.

Page 44: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 44 / 62 www.go-sys.de [email protected]

The following window appears:

Explanation of text in the formula input field: Value =ISA(254)*(1000/ISA.Pathlength); ⇒ determines the measured wavelength (here 254 nm) and converts the measured value into the unit 1/m

and writes this in the variable “Value“ „ISA.Pathlength“ is a system variable (value of the variable is the length of the measurement path in

mm), see 9.2.1 The Configuration Window and 8.2 The Maintenance Menu if (Value<0) Value = 0; ⇒ suppresses the output of negative values Value; ⇒ outputs the value The result is the Spectral Absorption Coefficient (SAC) at 254 nm.

formula input field

actual value of the SAK254

Page 45: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 45 / 62 www.go-sys.de [email protected]

14.2 Example: Chemical Oxygen Demand (COD or CSB) The Chemical Oxygen Demand (COD) is a measurement of how much oxygen is required by the chemical proc-esses in waste water. A high value points to a high percentage of organic material in the waste water. The algorithm to calculate COD is determined by a comparison of values from test samples. Not less than 25 test samples are required.

Select CSB with double-click.

Page 46: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 46 / 62 www.go-sys.de [email protected]

The following window appears:

The COD value is the sum of the weighted values of the absorption coefficient of four wavelengths (388 nm, 390 nm, 532 nm, 542 nm). Explanation of text in the formula input field: Value = 13.2757; ⇒ sets the output value to 13.2757 Value += ISA(388)* -275944.1177; ⇒ adds to the output value the with 13.2757 multiplied absorption coefficient at a wavelength of 388 nm etc. if ( Value < 0 ) Value = 0; ⇒ suppresses the output of negative values Value; ⇒ outputs the value

formula input field

current CSB value

This example is not generally applicable.

Page 47: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 47 / 62 www.go-sys.de [email protected]

14.3 ISA Formula Examples What follows is a list of examples to show how a virtual sensor can be programmed to display ISA calculations.

In this formula the absorption values at frequencies 284nm, 332nm and 628nm are used to generate a CSB value. A lower limit is also implemented. The exact coefficients are derived from ISA Soft software calcula-tions.

ISA NO3 example

# IS A N O3 ( exa m pl e) Valu e = - 0 .06 347 ; Valu e += I SA (28 4) * 28 .5 47863 ; Valu e += I SA (33 2) * - 51 . 92771 1; Valu e += I SA (62 8) * 30 .1 10743 ; i f ( Valu e < 0 ) Va lu e = 0 ; Valu e;

A saved spectrum fingerprint can be used as a reference to look for medium contamination. In this formula, the sum of the absolute value of the differences of the present spectrum to the saved fingerprint at all fre-quencies in the range 200 to 710nm is calculated. The frequency range can of course be adjusted.

ISA fingerprint difference

# IS A F i n ge r pr in t D i f f e re nc e exa m ple F P = 0 ; fo r ( i=2 30 ; i<= 70 0 ; i+ =2 ) F P += ab s ( I SAF P ( i ) ) ; F P ;

The formula calculates the Spectral Absorption Coefficient (SAC) at 254 nm. ISA SAK254

# IS A SA C25 4 SAC = I SA( 254 )* (100 0 / I SA .P at hle n gth ) ; i f ( SA C < 0 ) S AC = 0 ; SAC ;

Page 48: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 48 / 62 www.go-sys.de [email protected]

ISA SAK254 with turbidity compensation # IS A SA C25 4 wi th t u rbi d ity c o m pe ns a t i o n of f s et = 0 ; fo r ( i=6 00 ; i<7 00 ; i+= 2) of fs et += I SA ( i ) ; of f s et / = 50 ; # cal c ulat es t he a ve ra g e of the abs o r pt i o n s pe ct ru m d ri f t # in th e r an g e 60 0n m to 700n m SAC = ( I SA (25 4) -o f fs et )* ( 1000 .0 / I SA .P at hle n gt h ) ; i f ( SA C < 0 ) S A C = 0 ; SAC ; ISA Modbus export This formula makes spectral data retrievable for a Modbus master device. # IS A S p ec tr u m M o dbu s - Exp o rt fo r ( i=0 ; i <=25 5; i++ ) M O DR (i ) = ISA 0( i ) ; For more information see Service documentation Modbus configuration with MODConfig.exe Article-No. DOC 486 0003-E-1.2-SD-3

Page 49: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 49 / 62 www.go-sys.de [email protected]

15 Parameter Accuracy Parameters, calculated with ISA spectral data can achieve the accuracy of 5% (typical 5 % - 10 %). In detail the accuracy can vary by the change of the water matrix. If the water matrix has higher variability, for example day / night or seasonal changes, this has to be analyzed and a special calibration must be performed. Changes in the water matrix can be detected by other parameters (conductivity, pH value, temperature etc). 1. The accuracy of spectral data calculated parameters always is influenced by the quality of calibration.

A higher number of calibration points will have the result of a more accurate calibration!

2. The range of the reference value pairs∗ must cover the entire measuring range evenly distributed as possi-ble. The minimum number of reference value pairs is 25, a smaller number reduces the quality of the cali-bration and may subsequently lead to erroneous measurements.

3. The right choice of the analytical method and the care in the procedure are the most important require-ments for the accuracy of the calculation. The accuracy of the ISA parameter calibration is depending upon the specific accuracy of the chemical method for the parameter.

4. After a calibration it must be checked and, if necessary, adjusted over a longer period (e.g. one week). This increases the stability of the measurement.

Measurements with high accuracy over a longer period are only possible with correspondingly adapted main-tenance, a maintenance schedule is highly recommended here.

In general periodic cleaning and recalibration increases the measurement quality. The interval of the mainte-nance is determined by the measurement conditions and can last from several weeks to several months. See Manual ISA and Process Spectrometer Commissioning – Maintenance – Service

∗ For the calculation and calibration to the desired parameter of a specific application, it is necessary that for each parameter

reference values from chemical laboratory analysis and the corresponding spectral values are provided. In case of a one-parameter calibration a spectrum is assigned to one reference value, in a multi-parameter calibration there is more than one reference value assigned. The spectral data of a spectrum plus one or more corresponding measurement values are called reference value pair.

Page 50: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 50 / 62 www.go-sys.de [email protected]

16 CAN-bus Connection and Termination

External sensor module

DIP switch at On ⇒ Sensor module is terminated

DIP switch at Off ⇒ Sensor module is not terminated

Termination of the BlueBox, see Manual BlueBox

CAN-bus socket with adjacent

termination switch

Page 51: Manual ISA In Situ Spectral Analyser

ISA

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 51 / 62 www.go-sys.de [email protected]

17 Connections of the External Sensor Module

18 Connection of the Compressed Air Line

Push the PU-tube into the plug connector until it stops (approx. 5 mm).

Press with a suitable tool on the outer ring of the plug connector and remove the PU-tube.

plug connector

4 mm PU-tube

GO Systemelektronik GmbH

CAN-Bus-Connection

Sensor head connection

Compressed air connector (4 – 6 bar)

Page 52: Manual ISA In Situ Spectral Analyser

ISA – The Configuration Data Sheet

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 52 / 62 www.go-sys.de [email protected]

Appendix A – The Configuration Data Sheet The configuration data sheet contains the necessary settings to run the BlueBox. Example BlueBox TS:

Page 53: Manual ISA In Situ Spectral Analyser

ISA – The Configuration Data Sheet

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 53 / 62 www.go-sys.de [email protected]

1. BlueBox TS: Serial Number A1234 BlueBox Password (PIN) xxxxx Storage Device CF-256

Serial Number Serial number of the BlueBox With this serial number the BlueBox is identified by the BlueBox PC Software. set at the factory, not changeable BlueBox Password (PIN) Password of the BlueBox Is required to change the BlueBox system settings. set at the factory, not changeable Storage Device Model and size of the internal memory of the BlueBox, here CF-256

(CF = CompactFlash, 256 = 256 MB) set at the factory, changeable by replacing

2. Network:

IP Address 192.168.1.167 Netmask 255.255.255.0 Gateway 0.0.0.0 Port 14111 Login Name bluebox Password xxxxx

IP Address IP address of the BlueBox

At this address, the BlueBox is addressed on the network. set at the factory, changeable

Netmask Netmask of the BlueBox set at the factory, changeable Gateway Default gateway of the Blue Box set at the factory, changeable Port Network Port of the BlueBox set at the factory∗, not changeable Login Name User name for a modem connection set at the factory, not changeable Password Network password of the BlueBox Is needed to access the BlueBox via the AMS software. set at the factory, not changeable

∗ 14111 / or when encryption is enabled 14110

Page 54: Manual ISA In Situ Spectral Analyser

ISA – The Configuration Data Sheet

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 54 / 62 www.go-sys.de [email protected]

3. BlueGate Settings: IP Address 91.221.182.1411 Password BlueGate xxxxx

IP Address IP address of an Internet Gateway can be configured at the factory, changeable Password BlueGate Password of an Internet Gateway can be configured at the factory, changeable2

4. BlueBox PC Software - BlueGate Settings: Host datagateway.go-sys.de Username bluebox Password Windows xxxxx

If the BlueBox is accessed via a gateway (e.g. a UMTS connection), you have to enter these access data in the BlueBox SQL Software.

Only for a BlueBox with build-in spectrometer

5. Spectrometer Components: Spectrometer Board Serial Number ISA12345 ISA Sensor Head Serial Number 1234 Fibre Optic Cable Serial Number 12/3456 Zeiss Module Number 123456

ISA Spectrometer Board Serial Number Serial number of the spectrometer board set at the factory, not changeable ISA Sensor Head Serial Number Serial number of the sensor head set at the factory, not changeable Fibre Optic Cable Serial Number Serial number of the sensor head cable set at the factory, not changeable Zeiss Module Number Serial number of the Zeiss spectrometer module set at the factory, not changeable

1 IP-Address of the GO webserver (default address) 2 changeable only at the default address

Page 55: Manual ISA In Situ Spectral Analyser

ISA – Connector Pin Assignment BlueBox TS

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 55 / 62 www.go-sys.de [email protected]

Appendix B – Connector Pin Assignment at the BlueBox in the Version TS The BlueBox has the following connection options:

1. Power supply (24 VDC) via a 4-pin M12 panel plug (male)

2. CAN-bus connection via a 5-pin M12 panel socket (female) Connection of BlueBox sensor-/actuator modules via a 4-wire CAN-bus connection cable

3. Relay 1 and 2 output connection via a 8-pin M12 panel socket (female), contact load 48 V/0,5 A

4. Connection of the compressed air line. See 18 Connection of the Compressed Air Line

5. Socket of the sensor head cable

6. LAN connection for LAN cables in standard or cross-link design

7. Serial interface connection via a 3-pin M8 panel plug (male), RS-232 or RS-485 (optional with 4-pin M8 panel plug and 12 VDC power supply)

8. Current output 2 connection via a 4-pin M8 panel socket (female), 4 to 20 mA

9. Current output 2 connection via a 4-pin M8 panel socket (female), 4 to 20 mA

10. Pulse input 2 connection via a 4-pin M8 panel socket (female) PNP/NPN selectable (optional static), switching current approx. 6 mA, measurement range up to 20 kHz, debounced up to 100 Hz

11. Pulse input 2 connection via a 4-pin M8 panel socket (female) PNP/NPN selectable (optional static), switching current approx. 6 mA, measurement range up to 20 kHz, debounced up to 100 Hz

12. Ground connection

13. Borehole (with cap) for an antenna or cable modem

Modems (optional):

• UMTS modem with antenna

• Analogue modem

• ISDN modem

2 3 5 6 7 1 8 9

10 11 12

13

4

Page 56: Manual ISA In Situ Spectral Analyser

ISA – Connector Pin Assignment BlueBox TS

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 56 / 62 www.go-sys.de [email protected]

Connector Pin Assignment at the BlueBox T4

Power supply 24 VDC panel plug (M12, male) 3 +24 VDC 4 GND 24 V

CAN-bus panel socket (M12, female) 1 CAN-H 2 CAN-L 3 +24 VDC 4 GND 24 V

Relay outputs panel socket (M12, female) 1 NC1 normally closed 1 2 NO1 normally open 1 3 P1 switch 1 4 NC2 normally closed 2 5 NO2 normally open 2 6 P2 switch 2

43

1 2

34

3

4

1 2

56

Page 57: Manual ISA In Situ Spectral Analyser

ISA – Connector Pin Assignment BlueBox TS

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 57 / 62 www.go-sys.de [email protected]

RS-232 or RS-485 panel plug (M8, male) RS-232 RS-485 1 GND GND 3 RX RX/TX- 4 TX RX/TX+

RS-232 or RS-485 with optional power supply panel plug (M8, male) RS-232 RS-485 1 GND GND 2 12 VDC 12 VDC 3 RX RX/TX- 4 TX RX/TX+

Current output 4 – 20 mA panel socket (M8, female) 1 +24 VDC 2 I-Return

Pulse input panel plug (M8, male) 1 24 V 2 PNP 3 NPN 4 GND

1

4

3

1

2 4

3

2

1

1

2 4

3

Page 58: Manual ISA In Situ Spectral Analyser

ISA – ISA mobil

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 58 / 62 www.go-sys.de [email protected]

Appendix C – ISA mobile 1 Overview ..................................................................................................................................................................................................... 58

2 Technical data ............................................................................................................................................................................................ 59 3 Startup ......................................................................................................................................................................................................... 60 4 Removal from service............................................................................................................................................................................... 62 5 Label with access data ............................................................................................................................................................................. 62

1 Overview

ISA mobile is an integrated combination of ISA TS with a BlueBox Power Up Control inclusive two 12 V accu-mulator batteries in a case. The Power Up Control

• Converts the 12 V voltage of a battery into a 24 V power supply for a BlueBox-System. • Starts up and shuts down the BlueBox at user specified intervals. • Protects the accumulator batteries from deep discharge.

The Power Up Control is described in the manual BlueBox - Power Up Control. The use of ISA mobile differs from it only in the choice between automatic mode and manual mode (see page 61, there item 7).

Before prolonged storage and transportation for a long period, the fuses of the accumulator batteries have to be removed.

The fibre optic cable must not be bent or flexed with a radius less than 40 mm.

The sensor head must not be hung on the sensor head cable, please use the lugs for installation of the sensor head.

Page 59: Manual ISA In Situ Spectral Analyser

ISA – ISA mobil

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 59 / 62 www.go-sys.de [email protected]

2 Technical data

case housing polycarbonate; 59 cm x 43 cm x 21 cm

protection class IP65 with closed case and attached inside nut protection cap and attached protection caps to the external electrical connections

environmental conditions: ambient temperature: storage: -10 °C to +50 °C operation: 0 °C to +40 °C

weight 19 kg

accumulator batteries Pb accumulator, 2 x 7 Ah

length sensor head cable 6 m

connections outside∗

compressed air connector quick connector 4 mm PU-tube, 4 - 6 bar

12 VDC 4-pin mounting socket with pins (male) to connect the charger device

CAN-bus 4-pin mounting socket (female)

compressed air connector 12 VDC CAN-bus

fuses (a) for accumulator battery 1 left side under the mounting plate 4 A medium time lag

(b) for accumulator battery 2 right side under the mounting plate 4 A medium time lag

∗ connections BlueBox see: 20 Connections of the BlueBox / Version TS

a b

Page 60: Manual ISA In Situ Spectral Analyser

ISA – ISA mobil

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 60 / 62 www.go-sys.de [email protected]

3 Startup

1. Open the case.

2. Mount the fuse (if needed).

3. Charge the accumulator batteries with the supplied charge device (via the 12 V DC connection at the outside right of the box). The charging time is 8 hours in total discharge

4. Pull the tab on the holding clamps to the top.

5. Disconnect the sensor head from the holding clamps and insert the sensor head with the sensor head cable through the screw hole in the foam sleeve to the outside. The sensor head cable may not be in a tighter radius than 40 mm bent or kinked. It also requires the sensor head will not be suspended from the measuring head cable, use instead the lugs at the sensor head.

6. Close the inside of the screw hole with the union nut of the sensor head cable.

Page 61: Manual ISA In Situ Spectral Analyser

ISA – ISA mobil

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 61 / 62 www.go-sys.de [email protected]

7. Choose the operation mode of the Power Up Control.

• Off: Transport and storage mode, the accumulators are electrically disconnected. delivery status • automatic mode: The Power Up Control is activated. • manual mode: The Power Up Control is deactivated. The BlueBox is continuously supplied with power. For a complete description of the Power Up Control, refer to the Manual BlueBox - Power Up Control.

7.1. If the Power Up Control is in the measuring time∗, the BlueBox starts up automatically

7.2. If the BlueBox do not starts up in the automatic mode, switch briefly back to manual mode and back. The BlueBox starts up.

7.3. Proceeding for changes of interval and measuring time: If the BlueBox do not starts up in the automatic mode, switch to manual mode. The BlueBox starts up.

8. Connect the compressed air line with the compressed air connector at the right outer side of the case (see 19 Connection of the compressed air line).

∗ see Manual BlueBox-Power Up Control

Page 62: Manual ISA In Situ Spectral Analyser

ISA – ISA mobil

GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 Page 62 / 62 www.go-sys.de [email protected]

4 Removal from service

1. If the Power Up Module is in operation, shutdown the BlueBox with the menu operation.∗

2. Switch setting on operation mode Off. (see page above)

3. Before prolonged storage and transportation for a long period, the fuses of the accumulator batteries have to be removed.

5 Label with access data

On the upper right side of the mounting plate there is a label with the access data of the BlueBox.

Serial Number serial number of BlueBox With this serial number the BlueBox is identified by the BlueBox PC Software.

set at the factory, not changeable Display password of the BlueBox Is required to change the BlueBox system settings. set at the factory, not changeable Password network password of BlueBox Is needed to access the BlueBox via the AMS software. set at the factory, not changeable

Login Name user name for a modem connection set at the factory, not changeable

These and more access data can be found in the attached configuration sheet.

∗ see Manual BlueBox there 8.2.1.3 Shutdown

Central Unit Serial Number: AB1234 Display: 12345 Network Password: ab123c Login Name: abcdefg