INSTRUCTION MANUAL

DIGITAL THEODOLITE

DT-101

Page 2: Manual Teodolito DT 100 En

Foreword

Thank you for purchasing the TOPCON DIGITAL THEODOLITE, DT-101. For the best

performance of the instruments, please carefully read these instructions and keep them

in a convenient location for future reference.

1. Do not aim the instrument directly into the sun .Aiming the instrument directly into the sun can result in serious damage to the eyes. Damage tothe instrument could also result from exposing the instrument's objective lens to direct sunlight.The use of a solar filter is suggested to alleviate this problem.

2. Setting the instrument on a tripodWhen mounting the instrument on a tripod, use a wooden tripod when possible. The vibrationsthat may occur when using a metallic tripod can effect the measuring precision.

3. Installing the tribrachIf the tribrach is installed incorrectly , the measuring precision could be effected. Occasionallycheck the adjusting screws on the tribrach. Make sure the base fixing lever is locked and thebase fixing screws are tightened.

4. Guarding the instrument against shocksWhen transporting the instrument, provide some protection to minimize risk of shocks. Heavyshocks may cause the measurement to be faulty.

5. Carrying the instrumentAlways carry the instrument by its handgrip.

6. Do not expose the instrument to extreme heat.Do not leave the instrument in extreme heat (+122 degrees F) longer than necessary. It couldeffect the performance of the instrument. Do not expose the instrument's objective lens todirect sunlight without a filter. Damage could occur inside the instrument.

7. Battery level checkConfirm battery level remaining before operating.

GENERAL HANDLING PRECAUTIONS

Page 3: Manual Teodolito DT 100 En

CONTENTS1 NOMENCLATURE AND FUNCTIONS ..................................................................1-1

1.1 Nomenclature ........................................................................................................................ 1-11.2 Display................................................................................................................................... 1-31.3 Operating Key........................................................................................................................ 1-31.4 Serial Signal RS-232C Connector ......................................................................................... 1-4

2 PREPARATION FOR MEASUREMENT................................................................2-12.1 Setting Instrument up for Measurement ................................................................................ 2-12.2 POWER Switch Key ON........................................................................................................ 2-22.3 Battery Power Remaining Display ......................................................................................... 2-22.4 Vertical Angle Tilt Correction ................................................................................................. 2-3

3 ANGLE MEASUREMENT......................................................................................3-13.1 Measuring Horizontal Angle Right and Vertical Angle .......................................................... 3-13.2 Switching Horizontal Angle Right/Left.................................................................................... 3-23.3 Measuring from the Required Horizontal Angle..................................................................... 3-23.4 Vertical Angle Percent Grade(%) Mode ................................................................................ 3-33.5 Repetition Angle Measurement ............................................................................................. 3-3

4 DISTANCE/COORDINATE MEASUREMENT.......................................................4-14.1 Combined Use with EDM ...................................................................................................... 4-14.2 How to connect DT-101 to EDM............................................................................................ 4-14.3 Power supply ......................................................................................................................... 4-14.4 Distance Measurement (Continuous Measurement)............................................................. 4-24.5 Distance Measurement (N-time Measurement) .................................................................... 4-34.6 Coordinate Measurement ...................................................................................................... 4-44.7 Tracking Mode....................................................................................................................... 4-5

5 THE OTHER FUNCTIONS ....................................................................................5-15.1 Stadia Surveying ................................................................................................................... 5-15.2 Auto Cut Off........................................................................................................................... 5-15.3 Setting Minimum Angle Reading ........................................................................................... 5-1

6 COMBINED USE OF INSTRUMENT WITH THE DATA COLLECTOR.................6-16.1 Connecting DT-101 to the data collector ............................................................................... 6-16.2 Mode Available with Data Collector ....................................................................................... 6-16.3 Stake-out ............................................................................................................................... 6-16.4 Coordinate Measurement with Data Collector....................................................................... 6-2

7 SELECTING MODE...............................................................................................7-17.1 Items of the Selecting Mode.................................................................................................. 7-17.2 How to Set Selecting Mode ................................................................................................... 7-1

8 HANDLING POWER SOURCE .............................................................................8-19 CHECK AND ADJUSTMENT ................................................................................9-1

• Pointers on the Adjustment ................................................................................................... 9-1• Notes on the Tribrach............................................................................................................ 9-2

9.1 Checking /Adjusting the Plate Level ...................................................................................... 9-29.2 Checking /Adjusting the Circular Level.................................................................................. 9-29.3 Adjustment of the Vertical Cross-hair.................................................................................... 9-39.4 Collimation of the Instrument................................................................................................. 9-49.5 Checking / Adjusting the Optical Plummet Telescope........................................................... 9-59.6 Adjustment of Vertical Angle 0 Datum................................................................................... 9-6

Page 4: Manual Teodolito DT 100 En

10 CORRECTION FOR REFRACTION AND EARTH CURVATURE.......................10-110.1 Distance Calculation Formula.............................................................................................. 10-1

11 CORRECTION FOR H.I. VALUE.........................................................................11-112 DETACH/ATTACH OF TRIBRACH .....................................................................12-113 PRECAUTIONS...................................................................................................13-114 OPTIONAL ACCESSORIES................................................................................14-115 TARGET (PRISM) SYSTEM................................................................................15-116 ERROR DISPLAYS .............................................................................................16-117 SPECIFICATIONS...............................................................................................17-1

STANDARD SET COMPOSITION!Instrument (with lens cap) ...................................................................................... 1pc."Carrying case .......................................................................................................... 1pc.#Plumb bob set ......................................................................................................... 1set$Tool kit [Cleaning brush, Screw driver, Rod pin, Plumb bob hook] ......................... 1set%Sunshade................................................................................................................ 1pc.&AA Batteries ........................................................................................................ 4pcs.'Cable PC-5S ........................................................................................................... 1pc.(Plastic rain cover..................................................................................................... 1pc.)Silicon cloth ............................................................................................................. 1pc.*Instruction manual................................................................................................... 1pc.(Make sure that all of the above items are with the instrument when purchased.)

Page 5: Manual Teodolito DT 100 En

1 NOMENCLATURE AND FUNCTIONS

1-1

1 NOMENCLATURE AND FUNCTIONS1.1 Nomenclature

Objective lens

Connector (to beconnected to EDM)

Instrumentcenter mark

Sighting collimator

Horizontal motionclamp

Horizontaltangent screw

Optical plummettelescope

Display

Lower horizontalmotion clamp Serial signal RS-232C

connector

Page 6: Manual Teodolito DT 100 En

1 NOMENCLATURE AND FUNCTIONS

1-2

Hand grip

Telescopeeyepiece

Battery

Base

Tribrach fixing lever

Leveling screw

Circular level

Operating keys

Vertical tangentscrew

Plate level

Vertical motionclamp

Telescope focusing knob

Hand grip fixing screw

Page 7: Manual Teodolito DT 100 En

1 NOMENCLATURE AND FUNCTIONS

1-3

1.2 Display

! Display marksDisplay Contents Display Contents

V V-angle Remaining Battery Capacity

HR H-angle right EDM working

HL H-angle left Horizontal distance

Ht Repetition angle measurement Relative elevation

NAVG The number of repetition/ Average of angle

Slope distance

REP Repetition angle measurement N coordinate

TILT Tilt correction mode E coordinate

F Function key selection mode Z coordinate

% Percent grade SO Stake-out measurement

G Unit display GON

1.3 Operating Key

Keys FUNCTION Keys FUNCTION

R/L Selection for horizontal angle right / leftmeasurement

REC Pressing once : Data holdPressing again : Data transmission

HOLD Holding the horizontal angle REP Repetition angle measurement

FUNC Upper function selection Illumination of display and reticleON/OFF

ANG% V-angle displaySelection for vertical angle/percentdisplay

Coordinate measurement modeSwitching to display N, E/Z coordinate

0 SET Horizontal angle 0set Distance measurement modeSwitching HD( ) / SD( ) / RE( )

POWER Power switch , Moving the blinking digit to the left orright

Increasing the blinking numeral

Page 8: Manual Teodolito DT 100 En

1 NOMENCLATURE AND FUNCTIONS

1-4

1.4 Serial signal RS-232C connector

The serial signal connector is used for connecting the DT-101 with a computer orTOPCON Data Collector, which enables the computer to receive measured data from theDT-101 or to send preset data of horizontal angle, etc. to it.

! The following data will be output at each mode.Mode Output

Angle mode V, HR (or HL)

Horizontal distance, Relative elevation mode HD, V, HR (or HL), VD

Slope distance mode SD, V, HR (or HL), HD

Coordinate mode N, E, Z, HR (or HL)

Repetition angle mode Hm(Average), Hn(Total)

! Output at the tracking mode is displayed as distance data only.

The details necessary for the connection with the DT-101 are obtained from its InterfaceManual which is optionally available. Please refer to the manual.

Page 9: Manual Teodolito DT 100 En

2 PREPARATION FOR MEASUREMENT

2-1

2 PREPARATION FOR MEASUREMENT2.1 Setting Instrument Up For Measurement

Mount the instrument to the tripod. Level and center the instrument precisely to insure thebest performance. Use tripods with a tripod screw of 5/8 in. diameter and 11 threads perinch, such as the Type E TOPCON wide- frame wooden tripod.

Reference: Leveling and Centering the Instrument1.Setting up the TripodFirst, extend the extension legs to suitablelengths and tighten the screws on theirmidsections.

2.Attaching the Instrument on the TripodHead

Place the instrument carefully on the tripod headand slide the instrument by loosening the tripodscrew. If the plumb bob is positioned right overthe center of the point, slightly tighten the tripodscrew.

3.Roughly Leveling the Instrument by Usingthe Circular Level

! Turn the leveling screws A and B to movethe bubble in the circular level. The bubbleis now located on a line perpendicular to aline running through the centers of the twoleveling screws being adjusted.

" Turn the leveling screw C to bring thebubble to the center of the circular level.

4. Centering by Using the Plate Level! Rotate the instrument horizontally by using

the Horizontal motion/clamp screw andplace the plate level parallel with the lineconnecting leveling screws A and B, andthen bring the bubble to the center of theplate level by turning leveling screws A andB.

" Rotate the instrument 90° (100g) around itsvertical axis and turn the remaining levelingscrew or C to center the bubbleonce more.

# Repeat the procedures ! and " for each 90°(100g) rotation of the instrument and checkwhether the bubble is correctly centered for allfour points.

5.Centering by Using the Optical PlummetTelescope

Adjust the eyepiece of the optical plummettelescope to your eyesight.Slide the instrument by loosening the tripod screw,place the point on the center mark, and then tightenthe tripod screw. Sliding the instrument carefullynot to rotate that allows you to get the leastdislocation of the bubble.

6.Completely Leveling the InstrumentLeveling the instrument precisely in a similar way to4. Rotate the instrument and check to see that thebubble is in the center of the plate level regardlessof telescope direction, then tighten the tripod screwhard.

Levelingscrew B

Leveling screw C

90°

PointCenter mark

Leveling screw C

Leveling screw A

Levelingscrew B

Leveling screw A

Page 10: Manual Teodolito DT 100 En

2 PREPARATION FOR MEASUREMENT

2-2

2.2 POWER Switch Key ON

! Turn the power switch ON.Display initializes for two seconds and showsZERO SET.

" Rotate the telescope to set the instrument atvertical angle reading of 0°

$ Confirm the battery power remaining display. Confirm the battery power remaining display.Replace with battery when battery level is low. Refer to Chapter 2.3 “Battery PowerRemaining Display" .

Note : For setting the vertical angle at 0°, an electronic datum0 is provided on the vertical angle circle. If the telescopeis turned and the sensor passes the datum 0, anglemeasurement begins.The datum 0 is placed near the level position of thetelescope, the vertical angle setting of 0 can easily beset by rotating the telescope.

2.3 Battery Power Remaining DisplayBattery power remaining display indicates the power condition.

Battery Power Remaining Display

Measurement is possible.

The power is poor. The batteryshould be recharged orreplaced.

Measurement is impossible.Need to recharge or replace thebattery.

Note : 1) The battery operating time will vary depending on the environmental conditions suchas ambient temperature etc. It is recommended for safety to charge the batterybeforehand or to prepare spare full charged batteries.

2) For general usage of the battery, see chapter 8" Power source and charging".

POWER switch key ON.

Battery Power Remaining Display

Page 11: Manual Teodolito DT 100 En

2 PREPARATION FOR MEASUREMENT

2-3

2.4 Vertical Angle Tilt Correction

When the tilt sensors are activated, automatic correction of vertical and horizontal anglefor mislevelment is displayed.To ensure a precise angle measurement, tilt sensors must be turned on. The display canalso be used to fine level the instrument. If the tilt over display “b" appears theinstrument is out of automatic compensation range and must be leveled manually.

When the instrument tilted over correction range.

V bTILT

HR 120°30'40"

Standing Axis in the X directiontilted over

$ To set auto tilt correction from the moment that power is on, refer to chapter 7" VerticalAngle Tilt correction ( Tilt ON/OFF)".

$ The display of Vertical or Horizontal angle is unstable when instrument is on an unstablestage or a windy day. You can turn off the auto tilt correction function of V angle in thiscase.

Standing axis Zenith

$ DT-101 compensate the vertical anglereading due to inclination of the standingaxis in the X direction.

Inclination of the standingaxis in the X direction

Page 12: Manual Teodolito DT 100 En

3 ANGLE MEASUREMENT

3-1

3 ANGLE MEASUREMENT3.1 Measuring Horizontal Angle Right and Vertical Angle

Make sure the mode is in Angle measurement.Operating procedure Operation Display

! Collimate the 1st target (A). Collimate A V 90°30'40"

HR 20°30'40"

" Set horizontal angle of target A at 0° 00' 00".Press the [0 set] key twice.

[0 SET]Twice V 90°30'40"

HR 0°00'00"

# Collimate the 2nd target (B).The required V/H angle to target B will bedisplayed.

Collimate B V 98°40'40"

HR 123°45'20"

How to Collimate (Reference)! Point the telescope toward the light. Turn the diopter ring and adjust the diopter so that the cross

hairs are clearly observed.(Turn the diopter ring toward you first and then backward to focus.)

" Aim the target at the peak of the triangle mark of the sighting collimator. Allow a certain spacebetween the sighting collimator and yourself for collimating.

# Focus the target with the focusing knob.

* If parallax is created between the cross hairs and thetarget when viewing vertically or horizontally whilelooking into the telescope, focusing is incorrect ordiopter adjustment is poor. This adversely affectsprecision in measurement or survey.Eliminate the parallax by carefully focusing and usingdiopter adjustment.

Telescope eyepiece(Diopter ring)

Function knob

Reference

Page 13: Manual Teodolito DT 100 En

3 ANGLE MEASUREMENT

3-2

3.2 Switching Horizontal Angle Right/Left

Make sure the mode is Angle measurementOperating procedure Operation Display

! Collimate the 1st target (A). Collimate A V 90°10'20"

HR 120°30'40"

" Press the [R/L] key.The mode Horizontal angle Right (HR)switches to (HL) mode.

[R/L] V 90°10'20"

H L 239°29'20"

# Measure as HL mode.

$ Every time pressing the [R/L] key, HR/HL mode switches.

3.3 Measuring from the Required Horizontal Angle(Holding the Angle)

Make sure the mode is angle measurementOperating procedure Operation Display

! Set the required horizontal angle, usingHorizontal tangent screw

Display angle V 90°10'20"

HR 130°40'20"

" Press the [HOLD] key. *1) [HOLD] V 90°10'20"

HR 130°40'20"

# Collimate the target. Collimate Blinks

% Press the [HOLD] key to finish holding thehorizontal angle.

[HOLD] V 90°10'20"

HR 130°40'20"

*1) To return to the previous mode, press any key except the [HOLD] and the [FUNC+FUNC]key.

Page 14: Manual Teodolito DT 100 En

3 ANGLE MEASUREMENT

3-3

3.4 Vertical Angle Percent Grade(%) ModeMake sure the mode is Angle measurement

Operating procedure Operation Display

! Press [F4]() key to get the function as onpage 2. V 90°10'20"

HR 120°30'40"

" Press the [ANG%] key. *1) [ANG/%] % V -0.30

HR 120°30'40"

*1) Every time pressing the [ANG/%] key, the display mode switches.$ When the measurement is carried out over than±45° (±100%) from the horizontal, the

display shows [- - - - -] .

3.5 Repetition Angle MeasurementMake sure the mode is Angle measurement

Operating procedure Operation Display

! Press the [FUNC] key. [FUNC] F V 90°10'20"

HR 120°30'40"

" Press the [REP] key. [REP] REP Ht 0°00'00"

0 H

# Collimate the target A. and press the [0SET]key twice.

Collimate A[0SET][0SET]

REP Ht 0°00'00"

0 H

% Collimate the target B using the horizontalclamp and tangent screw.Press the [HOLD] key.

Collimate B[HOLD]

REP Ht 130°25'20"

1 AVG H 130°25'20"

& Recollimate target A using the horizontalclamp and tangent screw, and press the[R/L] key.

RecollimateA

[R/L]

' Recollimate target B using the horizontalclamp and tangent screw, and press the[HOLD] key.

RecollimateB

[HOLD]

REP Ht 260°50'40"

2 AVG H 130°25'20"

2 measurements

Page 15: Manual Teodolito DT 100 En

3 ANGLE MEASUREMENT

3-4

( Repeat & to ' to measure the desirednumber of repetitions. REP

Ht 521°41'20"4 AVG

H 130°25'20"

) To return to the normal angle mode, pressthe [FUNC] key and [HOLD] key. [FUNC]

[HOLD]

[Example] 4 measurement

$ Horizontal angle can be accumulated up to(2000°00'00" - minimum reading)(horizontal angle right) or -(2000°00'00" - minimumreading)(horizontal angle left) . In case of 5 second reading, horizontal angle can beaccumulated up to ±1999°59'55".

Page 16: Manual Teodolito DT 100 En

4 DISTANCE/COORDINATE MEASUREMENT

4-1

4 DISTANCE/COORDINATE MEASUREMENT4.1 Combined Use with EDM

Before measuring distance, set the atmospheric correction and the prism constant ofEDM.For details, see the EDM Instruction manual.

4.2 How to connect DT-101 to EDM

Cable PC-5S is used

4.3 Power supply

Each on-board battery is used .NOTE: It is not possible to supply the power to DT-101 and EDM from external battery .

PC-5S

Page 17: Manual Teodolito DT 100 En

4 DISTANCE/COORDINATE MEASUREMENT

4-2

4.4 Distance Measurement (Continuous Measurement)

Make sure the mode displays angle measurement.Operating procedure Operation Display

! Collimate the target by DT-101 and thecenter of the prism by EDM respectively.

Collimate

" Set the instrument for distancemeasurement mode (HR, ) pressing the[FUNC] key and the [0SET] key to acceptslope distance data from EDM.

[FUNC][0SET] m

HR 120°30'40"

# When EDM switch ON slope distancecontinuous mode in turn and measurementstarts. *1)

EDMPower ON

* 123.456 m

HR 120°30'40"

The data from EDM to DT-101 displayshorizontal distance. *2),3)

*1) When EDM is working, the " * " mark appears in the display.*2) The result is shown with buzzer sound.*3) Pressing the [FUNC] key and the [0 set] key again, the display changes to RE ( ), SD ( )

in turn.$ It is possible to choose the display order ( → → ) or ( → → ) for initial

measuring distance mode. Refer to Chapter 7 “ SELECTING MODE”.$ To return to the normal angle measuring mode, press the [ANG%] key.

Prism

Target

DT-101

166mmEDM

166mm

Page 18: Manual Teodolito DT 100 En

4 DISTANCE/COORDINATE MEASUREMENT

4-3

4.5 Distance Measurement (N-time Measurement)This is to get average of total number of distance measurement. Set the EDM for numberof distance measurement mode and enter frequency number of measurement.(To set EDM for number of distance measurement mode, refer to the EDM instructionmanual.)

Make sure the mode displays angle measurement.Operating procedure Operation Display

! Collimate the target by DT-101 and thecenter of the prism by EDM respectively.

Collimate

" Set the instrument for distancemeasurement mode (HR, ) pressing the[FUNC] key and the [0SET] key to acceptslope distance data from EDM. *1)

[FUNC][0SET] m

HR 120°30'40"

# EDM switch ON.Press the [MEAS] key of EDM, the number setwill be measured and displayed each time,then the average value is displayedfollowingly.

EDMPower ON

EDM[MEAS]

* 123.456 m

1 HR 120°30'40"

123.456 m3

HR 120°30'40"

3 measurements

*1) In the number of distance measurement, the measured value will not display on the DT-101 combined with other EDMs except DM-A5. Press [REC] key of EDM to display the dataEDM at that time.

Page 19: Manual Teodolito DT 100 En

4 DISTANCE/COORDINATE MEASUREMENT

4-4

4.6 Coordinate MeasurementThe following description is for the coordinate measurement with the origin (0,0,0) at theinstrument point. The coordinate measurement includes measurements for latitude,departure and relative elevation. If any other point than the instrument point is set as thecoordinate point, use of DATA COLLECTOR is required to set the coordinate of theinstrument point. Refer to chapter 6.4 “Coordinate measurement"

Make sure the mode displays angle measurement.Operating procedure Operation Display

! Use the horizontal clamp and tangentscrew to set the shown horizontal angle inthe direction angle of point A and lock.

Displayangle V 90°10'20"

HR 130°40'20"

" Press the [HOLD] key. 'HR' blinks. [HOLD]

# Collimate the target A, and press the[HOLD] key to finish holding. Collimate A

[HOLD] V 90°10'20"

HR 130°40'20"

% Collimate the target point B (prism) using thehorizontal clamp and tangent screw.

Collimate B

& Set the instrument for coordinatemeasurement mode pressing the [FUNC]key and the [ANG%] key to accept slopedistance data from EDM.

[FUNC][ANG%] m

' EDM switch ON, the coordinatemeasurement starts.N coordinate ( ) and E coordinate ( ) isdisplayed.

EDMPower ON

* 123.456 m

23.456 m

( Press the [FUNC] key and the [ANG%] key todisplay Z coordinate ( )*1)

[FUNC][ANG%]

* 3.456 m

HR 95°30'40"

*1) Pressing the [FUNC] key and the [ANG%] key, N ( ) , E ( ) and Z ( ) data is displayedalternately.

$ Z coordinate measured by DT-101 is the difference in height between the horizontalposition of the instrument telescope and target.

E

Prism (n,e,z)

z

N

n

Instrument point(0,0,0) e

Page 20: Manual Teodolito DT 100 En

4 DISTANCE/COORDINATE MEASUREMENT

4-5

4.7 Tracking Mode

This mode measures in shorter time than in fine mode.It is very useful when tailing the moving object or carrying out stake-out work.

Tracking Mode :The unit to be displayed : 10mmOperating procedure Operation Display

! Collimate the target by DT-101 and thecenter of the prism by EDM respectively.

Collimate

" Set the instrument for distancemeasurement mode (HR, ) pressing the[FUNC] key and the [0SET] key to acceptslope distance data from EDM.*1)

[FUNC][0SET] m

HR 120°30'40"

# EDM switch ON.Set the EDM to the slope distance mode intracking.*2)

EDMPower ON

EDM[MEAS]

* 123.456 m

HR 120°30'40"

The data from EDM to DT-101 displayshorizontal distance in tracking mode.

The display unit is 1cm.

*1) To measure the relative elevation ( ) and slope distance ( ), press the [FUNC] and the[0SET] key.

*2) At the condition of more than 2,000 m of measuring distance, the displayed value by thetracking mode is :

Displayed value = Measured distance -2,000mIn this case measurement must be done without atmospheric correction and correctionfor refraction calculation and correction for refraction and earth curvature way.

Page 21: Manual Teodolito DT 100 En

5 THE OTHER FUNCTIONS

5-1

5 THE OTHER FUNCTIONS5.1 Stadia Surveying

DT-101 can be used for stadia surveying, which is simply a convenient method formeasuring distance with the stadia hairs of the instrument, in combination with agraduated rod, such as leveling rod or stadia rod (which is preferable for longdistances).The distance from the center of the instrument to the rod is found by sightingthrough the instrument on to the rod and multiplying the stadia interval or reading (i.e.,the interval between the apparent locations of the bottom and top stadia hairs on the rod)by 100.

! First, set up the rod on the point." Then, sight through the telescope of the properly leveled instrument and determine

the distance or interval “l " on the rod between the bottom and top stadia hairs.# The horizontal distance “L" from the plumb line center of the instrument to the rod is

equal to 100 times the stadia interval or stadia reading or rod reading or “l ".L = 100 l

5.2 Auto Cut OffIf no key operation is given for more than 30 minutes, the power turns off automatically.To set this function, refer to Chapter 7 ”SELECTING MODE".

5.3 Setting Minimum Angle ReadingSelect minimum display unit for angle measurement. It is possible to select it as shownbelow.To set this function, refer to Chapter 7”SELECTING MODE".

1" / 5" ( 0.2mgon / 1mgon)

L=100× l

Stadia hairsl l

Page 22: Manual Teodolito DT 100 En

6 COMBINED USE OF INSTRUMENT WITH THE DATA COLLECTOR

6-1

6 COMBINED USE OF INSTRUMENT WITH THEDATA COLLECTORA wider variety of applications is a available if the instrument is used with the datacollector.For further information concerning the data collector, refer to data collector instructionmanual.

6.1 Connecting DT-101 to the data collectorConnect serial signal connector of DT-101 with Data collector using P-1.

6.2 Mode Available with Data Collector(1) Data input mode

The Data collector gives a prompt message guide the operation in the display window.The operator can enter the data by keyboard operation according to the message.For further information concerning the Data collector, refer to Data collectorinstruction manual.

(2) DT-101 control modeIn this mode , setting a standard distance ( for stake out), coordinate of instrumentpoint and horizontal angle is possible.

(3) Data StorageTo store the data obtained by DT-101 to Data collector, use the [REC] key of DT-101.Press the key twice and the data will be output to the Data collector.

6.3 Stake-outThe difference between the measured distance and the distance preset by data collectoroperation is displayed.

Displayed value=Measured distance - Standard ( Preset) distance! Stake-out operation can be performed for horizontal distance ( ), relative elevation

( ) or slope distance ( ).Procedure" Set the standard distance from Data collector.# Press the [FUNC] and the [0SET] key of the DT-101.$ Press the [FUNC] and [0SET] key, to select one of the distance measurement mode

( , , )$ Set the EDM to the distance measurement mode in tracking.

The difference between the measured distance and the standard distance isdisplayed.

! To return to normal distance measurement mode, reset the standard distance to “0"or turn the power switch off once.

! Stake-out operation can be carried out in the continuous measurement modeaccording to the similar operation procedure.

Page 23: Manual Teodolito DT 100 En

6 COMBINED USE OF INSTRUMENT WITH THE DATA COLLECTOR

6-2

6.4 Coordinate Measurement with Data Collector

The coordinate measurement from a coordinate origine can be carried out after settingthe coordinate of instrument point using Data collector.

Procedure" Input the coordinate of instrument point from Data collector.# Collimate target A.$ Set the direction angle from Data collector.% Collimate target B using horizontal tangent screw and horizontal motion clamp.& Set the instrument for coordinate measurement mode pressing the [FUNC] key and

the [ANG%] key to accept slope distance data from EDM.' Set the EDM to the slope distance mode.

N coordinate ( ) and E coordinate ( ) from the set origin are measured repeatedly.Pressing the [FUNC] key and [ANG%] key , Z coordinate ( ) key is measuredrepeatedly.

! Pressing the [FUNC] key and the [ANG%] key, N ( ) , E ( ) and Z ( ) data isdisplayed alternately.

! To return to the coordinate measurement with the origin at the instrument point, resetthe coordinate of instrument point at (0,0,0).

N

E

Instrument point C

Prism(n,e,z)

zn

Origin(0,0,0)e

Page 24: Manual Teodolito DT 100 En

7 SELECTING MODE

7-1

7 SELECTING MODEBy operating the keys, the following modes are available.

7.1 Items of the Selecting Mode

Selecting Mode 1

DigitNo.

Items Contents Setting value0

Setting value1

1 Angle unitDEG/GON

Choose angle unit degree(DEG) or /gon(GON). DEG GON

2 Angle unitMIL

Choose angle unit MIL. DEG/GON MIL

3 V angleZ0 / H0

Choose the vertical angle reading from zenithor from horizontal.

Zenith 0 Horizontal 0

4 Minimumangle unit

Choose the minimum angle unit. 5" 1"

5 Tilt correctionON/OFF

Setting the function of tilt correction. OFF ON

6 Auto cut offON/OFF

Setting the function of power off automaticallywhen it is continued more than 30 minutes.

OFF ON

7 REC TYPE Select REC-A or REC-B for data output.REC-A : The measurement is started and new

data is output.REC-B : The data being displayed is output.

REC-A REC-B

8 CRLF It is possible to output the data with carriagereturn and line feed.

OFF ON

Selecting Mode 2

DigitNo.

Items Contents Setting value0

Setting value1

1 Distance unit Choose measuring unit for distance meter orfeet.

meter feet

2 Distancedisplay order

Choose the distance display order. → → → →

3 H.I. errorcorrection

Setting the function of the H.I. value ( difference in instrument heights)

OFF ON

4 W-CorrectionON/OFF

Setting the function of the correction forrefraction and earth curvature.

OFF ON

5 W-CorrectionCoefficient

Set correction for refraction and earthcurvature, coefficient of refraction as ; K=0.14

K 0 20

K=0.14 K=0.20

6 NEZMEMORY

It is possible to retain the coordinate ofinstrument point after turning switch off.

OFF ON

7 NEZ / ENZ Select a coordinate displaying order either NEZor ENZ.

NEZ ENZ

8 Unused

Page 25: Manual Teodolito DT 100 En

7 SELECTING MODE

7-2

7.2 How to Set Selecting Mode

! Selecting Mode 1

<Example Setting>:Minimum angle unit “ 5" ", Tilt correction :OFFOperating procedure Operation Display

" While pressing the [R/L] key, power ON.The current setting is shown with the digitNo.1 blinking.

[R/L]+

Power ONS E L E C T 1

0 0 1 1 1 0 0 0

Digit No.8 Digit No.1

# Let the digit No.4 to be set blink by pressingthe [ ] key. [ ]

S E L E C T 1

0 0 1 1 1 0 0 0

Blinking

$ Press the [ ] key to set 0 for the digit. [ ]S E L E C T 1

0 0 1 1 0 0 0 0

% Let the digit No.5 to be set blink by pressingthe [ ] key. [ ]

S E L E C T 1

0 0 1 1 0 0 0 0

Blinking

& Press the [ ] key to set 0 for the digit. [ ]S E L E C T 1

0 0 1 0 0 0 0 0

' Press the [0 SET] key. [0 SET]S E L E C T 1

S E T

( Turn the power OFF. Power OFF

! Pressing [ ] key, blinking digit moves to the right. Pressing the [ ] key while the digit No.1 is blinking, blinking digit moves to the digit No. 8. Pressing the [ ] key while the digit No.8 is blinking, blinking digit moves to the digit No.1.! Every time pressing [ ] key, the blinking digit value 0/ 1 switches.

Page 26: Manual Teodolito DT 100 En

7 SELECTING MODE

7-3

! Selecting Mode 2

<Example Setting>:H.I. correction : ON , NEZ memory :ONOperating procedure Operation Display

" While pressing the [ANG%] key, power ON.The current setting is shown with the digitNo.1 blinking.

[ANG%]+

Power ONS E L E C T 2

0 0 0 0 0 0 1 0

Digit No.8 Digit No.1

# Let the digit No.3 to be set blink by pressingthe [ ] key. [ ]

S E L E C T 2

0 0 0 0 0 0 1 0

Blinking

$ Press the [ ] key to set 1 for the digit. [ ]S E L E C T 2

0 0 0 0 0 1 1 0

% Let the digit No.6 to be set blink by pressingthe [ ] key. [ ]

S E L E C T 2

0 0 0 0 0 1 1 0

Blinking

& Press the [ ] key to set 1 for the digit. [ ]S E L E C T 2

0 0 1 0 0 1 1 0

' Press the [0 SET] key. [0 SET]S E L E C T 2

S E T

( Turn the power OFF. Power OFF

! Pressing [ ] key, blinking digit moves to the right. Pressing the [ ] key while the digit No.1 is blinking, blinking digit moves to the digit No. 8. Pressing the [ ] key while the digit No.8 is blinking, blinking digit moves to the digit No.1.! Every time pressing [ ] key, the blinking digit value 0/ 1 switches.

Page 27: Manual Teodolito DT 100 En

8 HANDLING POWER SOURCE

8-1

8 HANDLING POWER SOURCE! For removingPush the lock lever downward and pull out the battery

! Replace of battery(DB-35)

" Push the hook downward and take the lid out.# Take out the old batteries and put new batteries as illustration shows in direction of

⊕ .$ Insert a convex in a upper hole. Click to close the lid by pressing it.

Replace all 4 batteries to new ones at the sametime.Don't mix the old batteries to the new ones.

! For installing

Place the base of the on-board battery into the main body, push the on-board batterytoward the instrument side till the battery clicks into position.

Lock lever

Lid

ConvexHole

Hook

Lid

Page 28: Manual Teodolito DT 100 En

9 CHECK AND ADJUSTMENT

9-1

9. CHECK AND ADJUSTMENT! Pointers on the Adjustment

" Adjust the eyepiece of the telescope properly prior to any checking operation whichinvolves sighting through the telescope.Remember to focus properly, with parallax completely eliminated.

# Carry out the adjustments in the order of item numbers, as the adjustments aredependent one upon another. Adjustments carried out in the wrong sequence mayeven nullify previous adjustment.

$ Always conclude adjustments by tightening the adjustment screws securely (but donot tighten them more than necessary, as you may strip the threads, twist off thescrew or place undue stress on the parts).Furthermore, always tighten by revolving in the direction of tightening tension.

% The attachment screws must also be tightened sufficiently, upon completion ofadjustments.

& Always repeat checking operations after adjustments are made, in order to confirmresults.

! Notes on the Tribrach

Note that the angle measuring precision may be effected directly if the tribrach has notbeen installed firmly." If any leveling screw becomes loose and slack or if collimation is unstable due to the

looseness of leveling screws, adjust by tightening the adjusting screws (in 2 places)installed over each leveling screw with a screwdriver

# If there is any slack between the leveling screws and the base, loosen the set screwof the holding ring and tighten the holding ring with adjusting pin, until it is properlyadjusted. Re-tighten the set screw on completing the adjustment.

Adjustment screw

Set screw

Adjustment screw

Leveling screwHolding ring

Page 29: Manual Teodolito DT 100 En

9 CHECK AND ADJUSTMENT

9-2

9.1 Checking /Adjusting the Plate Level

! Check" Place the plate level parallel to a line

running through the centers of twoleveling screws, say, A and B. Usethese two leveling screws only andplace the bubble in the center of theplate level.

# Rotate the instrument 180° or 200garound the vertical axis and checkbubble movement of the plate level. Ifthe bubble has been displaced, thenproceed with the following adjustment.

! Adjustment" Adjust the level adjustment capstan

screw, with the accessory adjustingpin and return the bubble towards thecenter of the plate level. Correct onlyone-half of the displacement by thismethod.

# Correct the remaining amount of thebubble displacement with the levelingscrews.

$ Rotate the instrument 180° or 200garound the vertical axis once moreand check bubble movement. If thebubble is still displaced, then repeatthe adjustment.

9.2 Checking /Adjusting the Circular Level! Check" Carefully level the instrument with the plate level only. If the bubble of the circular

level is centered properly, adjustment is not required. Otherwise, proceed with thefollowing adjustment.

! Adjustment" Shift the bubble to the center of the circular level, by adjusting three capstan

adjustment screws on the bottom surface of the circular level, with the accessoryadjusting pin.

180°

Plate level

Levelingscrew B

Levelingscrew A

Half amountdisplacement

Level adjustmentcapstan screw

Capstan adjustment screws

Bottom of tribrach

Page 30: Manual Teodolito DT 100 En

9 CHECK AND ADJUSTMENT

9-3

9.3 Adjustment of the Vertical Cross-hair

Adjustment is required if the vertical cross-hair is not in a place perpendicular to thehorizontal axis of the telescope ( since it must be possible to use any point on the hair formeasuring horizontal angles or running lines).

! Check

" Set the instrument up the tripod and carefully level it.# Sight the cross-hairs on a well defined Point A at a distance of, at least, 50 meters (

160ft.) .$ Next swing the telescope vertically using the vertical jog and shuttle, and check

whether the point travels along the length of the vertical cross-hair.% If the point appears to move continuously on the hair, the vertical cross-hair lies in a

plane perpendicular to the horizontal axis ( and adjustment is not required ).& However, if the point appears to be displaced from the vertical cross-hair, as the

telescope is swung vertically, adjustment is required in the reticle plate.

! Adjustment" Unscrew the cross-hair adjustment section cover, by revolving it in the

counterclockwise direction, and take it off. This will expose four eyepiece sectionattachment screws.

# Loosen all four attachment screws slightly with the accessory screw-driver (whiletaking note of the number of revolutions).Then revolve the eyepiece section so that the vertical cross-hair coincides to Point A'.Finally, re-tighten the four screws by the amount that they were loosened.

$ Check once more and if the point travels the entire length of the vertical cross-hair,further adjustment is not required.

Note : Perform following adjustment after complete above adjustment .Chapter 9.4 "Collimation of the Instrument", Chapter 9.6 "Adjustment of Vertical AngleZero Datum" .

Eyepiece

Eyepiece sectionattachment screws

Eyepiece sectionattachment screws

Page 31: Manual Teodolito DT 100 En

9 CHECK AND ADJUSTMENT

9-4

9.4 Collimation of the InstrumentCollimation is required to make the line of sight of the telescope perpendicular to thehorizontal axis of the instrument, as otherwise, it will not be possible to extend a straightline by direct means.

! Check" Set the instrument up with clear sights of

about 50 to 60meters (160 to 200 ft.) onboth sides of the instrument.

# Sight Point A at approximately 50meters (160 t.) distance.

$ Use the vertical motion clamp andvertical tangent screw only and rotatethe telescope 180° or 200g around thehorizontal axis, so that the telescope ispointed in the opposite direction.

% Sight Point B, at equal distance as PointA.

& Use the horizontal motion clamp andhorizontal tangent screw, and rotate theinstrument 180° or 200g around thevertical axis. Fix a sight on Point A oncemore.

' Use the vertical motion clamp andvertical tangent screw only and rotatethe telescope 180° or 200g around thehorizontal axis once more and fix a sighton Point C, which should coincide withprevious Point B.

( If Points B and C do not coincide, adjustin the following manner.

! Adjustment" Unscrew the cross-hair adjustment section cover.# Find Point D at a point between Points C and B,

which should be equal to 1/4th the distancebetween Points B and C and measured from PointC. This is because the apparent error betweenPoints B and C is four times the actual error sincethe telescope has been reversed twice during thechecking operation.

$ Shift the vertical cross-hair line andcoincide it with Point D, by revolving theleft and right capstan adjustment screwswith the adjusting pin.Upon completingthe adjustment, repeat the checkingoperation once more. If Points B and Ccoincide, further adjustment is notrequired. Otherwise , repeat theadjustment.

50mA

50m

50m

50m

BC

D

CC

A A

BB

A

Eyepiece

Eyepiece

Telescope

Capstanadjustmentscrews Capstan

adjustmentscrews

Page 32: Manual Teodolito DT 100 En

9 CHECK AND ADJUSTMENT

9-5

Note1) : First, loosen the capstan adjustment screw on the side to which the vertical cross-hair line must be moved. Then tighten the adjustment screw on the opposite side byan equal amount which will leave the tension of the adjustment screws unchanged.Revolve in the counterclockwise direction to loosen and in the clockwise direction totighten, but revolve as little as possible.

Note2) : Perform following adjustment after complete above adjustment . Chapter 9.6"Adjustment of Vertical Angle 0 Datum".

9.5 Checking / Adjusting the Optical Plummet TelescopeAdjustment is required to make the line of sight of the optical plummet telescope coincidewith the vertical axis ( as otherwise the vertical axis will not be in the true vertical whenthe instrument is optically plumbed).

! Check

" Coincide the center mark and the point.# Rotate the instrument 180° or 200g around the vertical axis and check the center

mark.If the point is properly centered in the center mark, adjustment is not required.Otherwise, adjust in the following manner.

! Adjustment

" Take off the adjustment section robber cap of the optical plummet telescopeeyepiece, by revolving it in the counterclockwise direction, and take it off. This willexpose four capstan adjustment screws which should be adjusted with the accessoryadjusting pin to shift the center mark to the point. However, correct only one-half ofthe displacement in this manner.

# Use the leveling screws and coincide the point and center mark.$ Rotate the instrument 180° or 200g around the vertical axis once more and check the

center mark. If it is coincided to the point, then further adjustment is not required.Otherwise, repeat the adjustment.

Note : First, loosen the capstan adjustment screw on the side to which the center mark mustbe moved. Then tighten the adjustment screw on the opposite side by an equal amountwhich will leave the tension of the adjustment screws unchanged.Revolve in the counterclockwise direction to loosen and in the clockwise direction totighten, but revolve as little as possible.

Capstanadjustmentscrews

Capstan adjustment screws

Eyepiece 1/2 of displacement

Plumment telescope

Page 33: Manual Teodolito DT 100 En

9 CHECK AND ADJUSTMENT

9-6

9.6 Adjustment of Vertical Angle 0 Datum

If when measuring the vertical angle of target A at telescope position normal (direct) andreverse settings, the amount of normal and reverse measurements combined is otherthan 360° (ZENITH-0), half of the difference from 360° is the error amount from corrected0 setting. Carry out adjustment. As adjustment for vertical angle 0 setting is the criteriafor determining instrument coordinate origin, use special care for adjustment.

Operating procedure Operation Display" Level the instrument properly with the plate

level.

# While pressing the [0SET] key, turn the powerswitch ON. [0SET]

+Power ON

V 0 SET

$ Turn the telescope to do 0 set. TurnTelescope V SETP 1

TILT

% Collimate target A from the telescopeproperly in normal setting. Collimate A

(Normal)

& Press the [0SET] key. [0SET] V SETP 2

TILT

' Collimate target A in reverse telescopesetting. Collimate A

(Reverse)

( Press the [0SET] key.Measured value is set and carry out normalangle measurement.

[0SET]SET

) Turn the power switch OFF. PowerOFF

! If there is any misoperation during checking, display for error appears. Then repeat theabove procedure from the start.

! Check that the instrument works properly by sighting a target at normal and reversetelescope settings and check if the amount of normal and reverse setting is 360°.

Page 34: Manual Teodolito DT 100 En

10 CORRECTION FOR REFRACTION AND EARTH CURVATURE

10-1

10 CORRECTION FOR REFRACTION AND EARTHCURVATUREThe instrument measures distance, taking into account correction for refraction and earthcurvature.

Note : If the telescope is positioned within ±9° from the nadir or zenith, no measurement willresult even if the correction function for refraction and earth curvature works.The display shows "E51".

10.1 Distance Calculation FormulaDistance Calculation Formula; with correction for refraction and earth curvature taken intoaccount. Follow the Formula below for converting horizontal and vertical distances.

Horizontal distance D=AC(α) or BE(β)Vertical distance Z =BC(α) or EA(β)D=L{cosα-(2 θ-γ)sinα}Z=L{sinα-( θ-γ)cosα}θ=L·cosα/2R Earth curvature

correcting itemγ=K·Lcosα/2R Atmospheric

refractioncorrecting item

K=0.14 or 0.2 Coefficient ofrefraction

R=6372km Radius of earthα ( or β) Altitude angleL Slope distance

! The conversion formula for horizontal and vertical distances is as follows whencorrection for refraction and earth curvature is not applied.D=L·cosαZ=L·sinα

Note : The coefficient of the instrument has been set at 0.14 before shipment (K=0.14).if the “K" value is to be changed, refer to "13. Selecting mode".

Radius of earthR=6372Km

θ

βD′

B

Z

C

D

A

-Z

E

Page 35: Manual Teodolito DT 100 En

11 CORRECTION FOR H.I. VALUE

11-1

11 CORRECTION FOR H.I. VALUE

If a measurement is performed as shown below without using a target, a slope distancemeasured with an EDM will be accompanied by an error corresponding to H.I. value (difference in instrument height).DT-101 is available automatically taking into account correction for an error. Refer tochapter 7 "SELECTING MODE"

! Calculation for H.I. Correction (Reference)

H.I. error correction is obtained in the following manner.

S' = 0.166 cosθ2 + S - ( 0.166 sin 2 2θ)

S' : Slope distance from DT-101S : Measured distance from EDMθ : Zenith angle from DT-101

Prism center

DT-101

θ

S′

S

H.I.=166mm

EDM

Page 36: Manual Teodolito DT 100 En

12 DETACH/ATTACH OF TRIBRACH

12-1

12 DETACH/ATTACH OF TRIBRACHThe instrument is easily detached or attached to the tribrach, with a tribrach fixing leverloosened or tightened for this purpose.

! Detachment" Loosen the tribrach fixing lever, by revolving it 180° or 200g in the

counterclockwise direction (which will point the triangle mark upwards).# Grip the carrying handle firmly with one hand while holding the tribrach with the

other. Then lift the instrument straight upwards and off.

! Attachment" Hold the instrument by the carrying handle, with one hand, and carefully lower it

on top of the tribrach while, at the same time, coinciding the alignment piece withthe tribrach alignment groove on the instrument and tribrach respectively.

# When fully seated, revolve the tribrach fixing lever 180° or 200g clockwise ( whichwill point the triangle mark downwards again).

! Locking the Tribrach Fixing LeverThe tribrach fixing lever can be locked, to prevent it be accidentally removed,especially if the upper instrument section is not being detached very often. Simplytighten the securing screw on the fixing lever with the accessory screwdriver, foundin the case.

Alignment piece

Alignment groove

Securing screwTribrach locking lever

Page 37: Manual Teodolito DT 100 En

13 PRECAUTIONS

13-1

13 PRECAUTIONS1. For transportation, hold by the handle or yoke of the instrument. Never hold by the

lens barrel as it can affect the fixing bracket inside and reduce the accuracy of theinstrument.

2. Never expose the instrument without a filter to direct sunlight. It may damage thecomponents inside the instrument.

3. Never leave the instrument unprotected in high temperature. The temperatureinside instrument may easily reach up to 70°C or above and will reduce the servicelife.

4. When a high degree of precision is required for measurement, provide shadeagainst direct sunlight for the instrument and tripod.

5. Any sudden change of temperature to the instrument or prism may result in areduction of measuring distance range, i.e. when taking the instrument out from aheated vehicle.

6. When opening the carrying case and taking out the instrument, place the casehorizontally, then open the case.

7. When storing the instrument in the case, always align the white indication marks onthe instrument, and set the telescope at the vertical position.

8. For transportation, provide dampening or a cushion appropriately to avoid suddenshock or vibration.

9. For cleaning the instrument after use, remove dust using a cleaning brush, thenwipe off with a cloth.

10. For cleaning the lens surface, use a cleaning brush to remove the dust, then use aclean lintless cotton cloth. Moisten it with alcohol ( or mixture with ether ) to wipegently in a rotational motion from the center out.

11. Even if any abnormality occurs, never attempt to disassemble or lubricate theinstrument yourself. Always consult with TOPCON or your dealer.

12. To remove the dust on the case, never use thinner or benzine. Use a clean clothmoistened with neutral detergent.

13. Check each part of the tripod after extended use. Parts (screws or clamps) maywork themselves free.

Page 38: Manual Teodolito DT 100 En

14 OPTIONAL ACCESSORIES

14-1

14 OPTIONAL ACCESSORIES

Page 39: Manual Teodolito DT 100 En

14 OPTIONAL ACCESSORIES

14-2

Data collectorSuitable for systemization of measuring instrument. Measuring data will be automatically stored andtransferred to a computer system, making measuring operations more efficient and saving time andeffort in such operation.

Prism unit case, Model 3This is the plastic case to store and carry various sets of prisms.The case covers one of the following prism sets:! Tilt single prism set! Tilt single prism set with a target plate! Fixed triple prism unit! Fixed triple prism unit with a target plate! External dimensions:

427(L)×254(W)×242(H) mm! Weight:3.1kg

Prism unit case, Model 6Fixed 9 prisms unit or tilting 3 prisms unit can be stored in this case. Especially, this is a very easycase to carry. Soft material is used.! External dimensions:

250(L)×120(W)×400(H) mm! Weight:0.5kg

Prism unit case, Model 51 prisms unit or fixed 3 prisms unit can be stored in this case. Especially, this is a very easy case tocarry. Soft material is used.! External dimensions:

200(L)×200(W)×350(H) mm! Weight:0.5kg

Page 40: Manual Teodolito DT 100 En

14 OPTIONAL ACCESSORIES

14-3

Trough Compass-5The trough compass is simply mounted on top of the carrying handle

Diagonal EyepieceObservation in an easy posture will be provided up to the zenith position

Optical plummet tribrachThis is detachable tribrach having built-in optical plummet telescope( Compatible with Wild)

Gadget case, Model 1A case to store and carry accessories.! External dimensions:

300(L)×145(W)×220(H) mm! Weight:1.4kg

Page 41: Manual Teodolito DT 100 En

14 OPTIONAL ACCESSORIES

14-4

Wide-frame extension leg tripod, Type E (Wood)! Flat head 5/8" × 11 threads with adjustable legs.

Back pack, Model 2Convenient for use in mountainous terrain.

Prism setsSee the description on Chapter 15 "TARGET (PRISM) SYSTEM".

Eyepiece prismThe eyepiece prism is supplied with a single dark filter glass for use on the telescope and is veryconvenient when tilting the telescope up to 40° from the zenith. The images are observed upsidedown, but not reversed laterally.symbol 129 \f "Wingdings" \s 10"# Simply screw the eyepiece prism directly into the female

screw mounts on the eyepiece of the telescope.symbol 130 \f "Wingdings" \s 10$# Use the dark filter glass ( with spring clips ) on the eyepiece

prism fixed to the telescope eyepiece, when observing the sun directly, otherwise take it off fornormal use.

Page 42: Manual Teodolito DT 100 En

14 OPTIONAL ACCESSORIES

14-5

Aluminum extension leg tripod,Type E! Flat head 5/8" symbol 180 \f "Symbol" \s 10× 11 threads with adjustable legs.

Eyepiece (26symbol 180 \f "Symbol" \s 11××××)

Solar reticle, Model 5A reticle designed for collimation of the sun.Can be used together with Solar Filter.

Solar filterA filter designed exclusively for direct collimation of the sun.

Page 43: Manual Teodolito DT 100 En

15 TARGET (PRISM) SYSTEM

15-1

15 TARGET (PRISM) SYSTEM

Pole adaptor-F2 Tribrach Optical plummettribrach

Prismadaptor-S2

Prismadaptor-2

• Tilting prismholder-2

• Triple prismholder-2

• Adjustable target-2

• Nine prism holder-2

Prism-2Prism-2

Target pole-2

Page 44: Manual Teodolito DT 100 En

16 ERROR DISPLAYS

16-1

16 ERROR DISPLAYS

Display Contents Countermeasure

b Instrument tilts over than 3 minutes. Level the instrument properly.

E01 Displayed when the instrument rotatedabnormally.

Pressing [0SET] key, returns tomeasuring mode.

E02 Displayed when the telescope rotatedabnormally.

Press [0SET] key, and after displayof "0 SET", set vertical angle zero setby rotating telescope.

E03 Displayed when a internal problemexists with the angle measuringsystem.

Switch OFF the power , then ONagain. Sometimes error occurs whenvibrates, clear the vibration.

E04 Displayed when the discrepancy valueof each measuring is more than 30"while repetition angle measurement isoperated.

Press [0SET] key, measure again fromthe beginning.

E51 Displayed when measurement carriedout within ±9° from zenith or nadir atthe Earth curvature and refractioncorrection mode is ON.

Set correction for refraction and earthcurvature mode OFF or measure outof ±9° from the zenith or Nadir.

E70 When "Adjustment of Vertical Angle 0Datum" is adjusted in wrongprocedure.When "Adjustment of Vertical Angle 0Datum" is carried out to the range outof ±45° from horizontal .

Switch OFF the power , then ONagain. Confirm the procedure andadjust again.

E80’s Mainly at the time data transmissionbetween DT-101 and externalinstrument.

Confirm the connection cables and theoperation procedure are correct.

E99 Abnormality in internal memory systemwhile "Adjustment of Vertical Angle 0Datum" is operated.

Switch OFF the power , then ONagain. Confirm the procedure andadjust again.

! If errors still persist after attempting to clear them, contact your local TOPCON dealer orTOPCON Head office.

Page 45: Manual Teodolito DT 100 En

17 SPECIFICATIONS

17-1

17 SPECIFICATIONSTelescope Length 149mm

Objective lens 45mm

Magnification 30×

Image Erect

Field of view 1°30'Resolving power 2.5"

Minimum focus 90cm

Stadia ratio 100

Stadia constant 0

Electronic Angle Method Incremental reading

Measurement Detecting Horizontal: 2side

Vertical: 2side

Minimum reading 1"/5"(0.2mgon/1mgon)

Accuracy *1) 2"

Diameter circle 71mm

Display Unit 2side

Illumination Display !

Reticle !

Compensator Tilt sensor Automatic verticalcompensator

Compensating range ±3'

Optical Plummet Magnification 3×

Telescope Field of view 5°

Focusing 0.5m~∞

Level Sensitivity Plate level 30" / 2mm

Circular level 10' / 2mm

Power Supply Battery 4 AA batteries

Operating Time Alkaline manganese drybattery

12 hours

Size Instrument height 176mm (6.93in)

DimensionD×W×H(mm)

149×180×313(5.87×7.09×12.3 in)

Weight(Including batteries)

4.4 kg(9.7 lb)

*1) Standard deviation based on DIN 18723

Page 46: Manual Teodolito DT 100 En

TOPCON CORPORATION75-1 Hasunuma-cho, Itabashi-ku, Tokyo, 174 Japan.Phone: 3-3967-1101 Fax: 3-3960-4214 Telex:0727-2384

TOPCON AMERICA CORPORATIONCORPORATE OFFICE65, West Century Road, Paramus, New Jersey 07652, U.S.A.Phone: 201-261-9450 Fax: 201-387-2710 Telex: 134338TOPCON CALIFORNIA3380 Industrial BLVD, Suite 105, West Sacramento, CA. 95691, U.S.A.Phone: 916-374-8575 Fax: 916-374-8329TOPCON MIDWEST1728 West Algonquin Road, Arlington Heights, Illinois 60005, U.S.A.Phone: 708-818-9188 Fax: 708-818-9342TOPCON DENVER4895 Joliet Street ,Suite E Denver Co 80239 U.S.APhone: 303-373-0990 Fax: 303-373-0998

TOPCON LASER SYSTEMS, INC.396 Earhart Way, Livemore, CA 94550,U.S.A.Phone: 510-443-8161 Fax: 510-443-7302

TOPCON EUROPE B.V.ESSE Baan 11, 2908 LJ Capelle a/d IJssel, The Netherlands.Phone: 10-4585077 Fax: 10-4585045 Telex: 23783

TOPCON DEUTSCHLAND G.m.b.H.Halskestr. 7,47877 Willich, GERMANYPhone: 02154-9290 Fax: 02154-929-111 Telex: 8531981 TOPC D

TOPCON S.A.R.L.104/106, Rue Rivay, 92300 Levallois-Perret, FRANCEPhone: 1-4106 9494 (MEDICAL) 1-4106 9490 (TOPOGRAPHIE)Fax: 1-47390251 Telex: 620287

TOPCON ESPANA S.A.HEAD OFFICEFrederic Mompou, 5 08960, Sant Just Desvern Barcelona, SPAINPhone: 3-4734057 Fax: 3-4733932MADRID OFFICEAvenida Ciudad de Barcelona 81,1 Planta 28007, Madrid, SPAINPhone: 1-552-4160 Fax: 1-552-4161

TOPCON OPTICAL SVENSKA A.B.Industrivä Nagen 4 P.O. Box 2140 43302 S!!!!vedalen SWEDENPhone: 031-261250 Fax: 031-268607 Telex: 21414

TOPCON SINGAPORE PTE. LTD.Alexandra Distripark, Block 4, #05-15, Pasir Panjang Road, SINGAPORE 0511Phone: 2780222 Fax: 2733540 Telex: RS 26622

TOPCON AUSTRALIA PTY. LTD.408 Victoria Rode, Gladesville, NSW 2111 AUSTRALIAPhone: 02-817-4666 Fax: 02-817-4654

TOPCON INSTRUMENTS (THAILAND) CO.,LTD.147/229-230, New Southern Bus Station, Pinklao-Nakarnchaisri Rd.Boromrashinee Road, Bangplad Bangkok, Noi Bangkok 10700 THAILANDPhone: 662-435-4002 Fax: 662-435-4005

TOPCON INSTRUMENTS (MALAYSIA) SDN.BHD.Lot 226 Jalan Negara Dua, Pusat Bandar Taman Melawati, Taman Melawat, 53100Kuala Lumpur, MALAYSIAPhone: 03-4079801 Fax: 03-4079796

BLOOMFIELD COMPUTING SERVICES PTY. LTD.408 Victria Road, Gladesville, NSW 2111, Australia.Phone: 02-817-4666 Fax: 02-817-4654