MEDICAL GAS CABINET Chart’s Medical Gas Cabinet is the final piece of equipment for the gasified liquid oxygen to pass through before entering the house oxygen line in the hospital. It is the command center and ensures that the oxygen storage and distribution system is functioning properly. The Medical Gas Cabinet monitors the liquid levels and pressures of the primary and secondary liquid supply tanks, and interfaces with the hospital’s alarm system. Cabinets Available Liquid cylinder with high pressure reserve Bulk liquid with mini-bulk reserve High pressure cylinder with high pressure cylinder reserve Rear View of Cabinet Showing Points of Connection Required On-Site Plumbing and Gauge Layout

Upload: trinhduong

Post on 27-Apr-2018

220 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: MEDICAL GAS CABINET - Chart Industriesfiles.chartindustries.com/11661251_MedicalGasCabinet.pdf · MEDICAL GAS CABINET Chart’s Medical Gas Cabinet is the final piece of equipment

MEDICAL GAS CABINET

Chart’s Medical Gas Cabinet is the final piece of equipment for the gasified liquid oxygen to pass through before entering the house oxygen line in the hospital. It is the command center and ensures that the oxygen storage and distribution system is functioning properly. The Medical Gas Cabinet monitors the liquid levels and pressures of the primary and secondary liquid supply tanks, and interfaces with the hospital’s alarm system.

Cabinets Available

• Liquid cylinder with high pressure reserve

• Bulk liquid with mini-bulk reserve

• High pressure cylinder with high pressure cylinder reserve

Rear View of Cabinet Showing Points of Connection Required On-Site

Plumbing and Gauge Layout

Page 2: MEDICAL GAS CABINET - Chart Industriesfiles.chartindustries.com/11661251_MedicalGasCabinet.pdf · MEDICAL GAS CABINET Chart’s Medical Gas Cabinet is the final piece of equipment

V2

PS1

V4

V1 R1 V1

G1

G1

G2

V2R2 CV

RV2

R3 CVRV1

RV1

CV

PS2

OUTLET

FLOW

FLOWFLOW

INLETINLET

MAIN RESERVE

RV2 DV

DV

V3

RV1

MEDICAL GAS CABINET

Chart Industries, Inc.U.S. : 1-800-877-3093Worldwide: 1-952-758-4484

Copyright © 2006 Chart IndustriesPN 11661251

www.chart-ind.com

R1-MAIN FINAL LINE REGULATOR R2-BACK-UP FINAL LINE REGULATORR3-RESERVE SYSTEM REGULATORV1-MAIN FINAL LINE REGULATOR ISOLATION VALVESV2-BACK-UP FINAL LINE REGULATOR ISOLATION VALVESV3-RESERVE SYSTEM REGULATOR ISOLATION VALVEV4-RESERVE IN-USE PRESSURE SWITCH ISOLATION VALVERV1-RELIEF VALVE (350 PSI)RV2-RELIEF VALVE (75 PSI)G1-FINAL LINE PRESSURE GAUGEG2-RESERVE SYSTEM PRESSURE GAUGECV-CHECK VALVESPS1-RESERVE IN-USE PRESSURE SWITCHPS2-RESERVE PRESSURE LOW PRESSURE SWITCHDV-DIVERTER VALVE

NFPA APPROVED AND CERTIFIED

Flow SchematicHospital Final Line For Liquid Reserve

APPROXIMATE DIMENSIONS: 2' x 3' x 4' (L x W x H)

Typical Wiring Schematic

PRESSURE SWITCH 1“RESERVE IN-USE”MAIN TANK LOW

PRESSURE SWITCH 2“RESERVE PRESSURE LOW”

RESERVE TANK LOW

PRESSURE SWITCH 1“LOW LEVEL ALARM”

MAIN TANK

PRESSURE SWITCH 2“LOW LEVEL ALARM-RESERVE”

RESERVE TANK

LOCATED INCONTROL PANEL

LOCATED ON TANK

PS1

PS2

LLS1

LLS2

TO CUSTOMER “RESERVE IN-USE” ALARM

TO CUSTOMER “RESERVE PRESSURE LOW” ALARM

TO CUSTOMER “LOW LEVEL ALARM-MAIN” ALARM

TO CUSTOMER “LOW LEVEL ALARM-RESERVE” ALARM

TERMINAL STRIP IN CONTROL PANEL

JUNCTION BOX IN CONTROL PANEL

INTERNAL WIRING

FIELD WIRING BY OTHERS