monitor radio flu oro equipment
TRANSCRIPT
-
8/17/2019 Monitor Radio Flu Oro Equipment
1/6
TECHNICAL STANDARD Radiographic Fluoroscopic Equipment / 1
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical
medical physicists in the United States. The College is a nonprofit professional society whose primary purposes are to advance the science of radiology,
improve radiologic services to the patient, study the socioecono mic aspects of the pr actice of r adiology, and encour age continuing education for
radiologists, radiation oncologists, medical physicists, and persons practicing in allied professional fields.
The American College of Radiology will periodically define new practice parameters and technical standards for radiologic pract ice to help adva nce the
science of radiology and to improve the quality of service to patients throughout the United States. Existing practice parameters and technical standards
will be reviewed for revision or renewal, as appropriate, on their fifth anniversary or sooner, if indicated.
Each practice parameter and technical standard, representing a policy statement by the College, has undergone a thorough consensus process in which it has
been subjected to extensive review and approval. The practice parameters and technical standar ds recognize that the safe and ef fective use of diagnostic
and therapeutic radiology requires specific training, skills, and techniques, as desc ribed in each document. Reproduction or modification of the published
practice parameter and technical standard by those entities not providing these services is not authorized.
Revised 2011 (Resolution 4)*
ACR TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS
PERFORMANCE MONITORING OF RADIOGRAPHIC AND FLUOROSCOPIC
EQUIPMENT
PREAMBLE
This document is an educational tool designed to assist practitioners in pr oviding appropriate radiologic care for
patients. Practice Parameters and Technical Standards are not inflexible rules or requirements of practice and are
not intended, nor should they be used, to establish a legal standard of care1. For these reasons and those set forth
below, the American College of Radiol ogy and our collaborating medical specialty societies caution against the
use of these documents in litigation in which the clinical decisions of a practitioner are called into question.
The ultimate judgment regarding the propriety of any specific procedure or course of action must be made by the
physician or medical physicist in light of all the circumstances presented. Thus, an approach that differs from the
guidelines, standing alone, does not nec essarily imply that the approach was below the st andard of care. To the
contrary, a conscientious practitioner may responsibly adopt a course of action different from that set forth in theguidelines when, in the reasonable judgment of the practitioner, such course of action is indicated by the condition
of the patient, lim itations of available resources, or advances in knowledge or technology subsequent to
publication of the guidelines. However, a practitioner who employs an approach substantially different from these
guidelines is advised to document in the patient record information sufficient to explain the approach taken.
The practice of medicine involves not only the science, but also the art of dealing with the prevention, diagnosis,
alleviation, and treat ment of disease. The variety and complexity of hum an conditions make it impossible to
always reach the most appropriate diagnosis or to pred ict with certainty a particular response to treat ment.
Therefore, it should be recognized that adherence to these guidelines will not assure an accurate diagnosis or a
successful outcome. All that should be expected is t hat the practitioner will follow a reasonable course o f action
based on current knowledge, available resources, and the needs of the patient to deliver effective and safe medical
care. The sole purpose of these guidelines is to assist practitioners in achieving this objective.
1 Iowa Medical Society and Iowa Society of Anesthesiologists v. Iowa Board of Nursing, ___ N.W.2d ___ (Iowa 2013) Iowa Supreme Court refuses to find
that the ACR Technical Standard for Management of the Use of Radiation in Fluoroscopic Procedures (Revised 2008) sets a national standard for who may
perform fluoroscopic procedures in light of the standard’s stated purpose that ACR stan dards are educational tools and not intended to establish a legal
standard of care. See also, Stanley v. McCarver, 63 P.3d 1076 (Ariz. App. 2003) where in a concurring opinion the Court stated that “published standards or
guidelines of specialty medical organizations are useful in determining the duty owed or th e standard of care applicable in a g iven situation” even though
ACR standards themselves do not establish the standard of care.
-
8/17/2019 Monitor Radio Flu Oro Equipment
2/6
2 / Radiographic Fluoroscopic Equipment TECHNICAL STANDARD
I. INTRODUCTION
This technical standard w as revised by the Ameri can College of Radiology (ACR) with assistance from the
American Association of Physicists in Medicine (AAPM).
The performance of all radiographic and fluoroscopic equipment must be evaluated upon installation and
monitored at least annually by a Qualified Medical Physicist to ensure that the equipment is functioning properly
and that patients are not exposed to unneces sary doses of radiation. Additional or more frequent monitoring may
be necessary after r epairs that might change the radiation exposure to pa tients or perso nnel or the imaging
performance of the eq uipment. Although it is n ot possible to c onsider all p ossible variations of equipment performance to be monitored, adherence to this technical standard will assi st in maximizing image quality and in
reducing patient radiation doses.
II. GOAL
The goals are to produce the highest quality diagnostic image at the lowest reasonable radiation dose consistent
with the clinical use of the equipment and the information requirement of the examination and to establish and
maintain performance standards.
III. QUALIFICATIONS AND RESPONSIBILITIES OF PERSONNEL
A Qualified Medical Physicist is an individual who is co mpetent to practice independentl y one or m ore of thesubfields in medical physics. The A merican College of Radiology (ACR) considers certification, cont inuing
education, and experience in the appropriate subfield(s) to demonstrate that an individual is competent to practice
one or more of the subfields in medical physics, and to be a Qu alified Medical Ph ysicist. The ACR strongl y
recommends that the indi vidual be certified in the ap propriate subfield(s) by the American Board of Radiology
(ABR), the Canadian College of Physics in Medicine, or by the American Board of Medical Physics (ABMP).
A Qualified Medical Physicist should meet the ACR Practice Guideline for Continuing Medical Education
(CME). (ACR Resolution 17, 1996 – revised in 2012, Resolution 42)
The appropriate subfield of medical physics for this technical standard is Diagnostic Medical Phy sics. (Previous
medical physics certification including Radiological Phy sics, Diagnostic Radiological Ph ysics, and Diag nostic
Imaging Physics are also acceptable.)
Understanding of the relationship between i mage quality and patient radiation dose is essential for proper
monitoring of equipment performance. The medical physicist must be fam iliar with the principles of i maging
physics and radiation protection; the current guidelines of the National Council on Radiation Protection and
Measurements (NCRP); federal and local laws and regula tions pertaining to the performanc e of the equipment
being tested; the function, clinical uses, and performance specifications of the imaging equipment; and calibration
processes and limitations of the instruments used for testing performance.
The medical physicist may be assisted by other properly trained individuals in obtaining test data for performance
monitoring. These individuals must be properly trained and approved by the medical physicist in the techniques of
performing the tests, the function and li mitations of the imaging equipment and test instruments, the reasons for
the tests, and the importance of the test results. The tests will be performed by or under the general supervision ofthe medical physicist, who is responsible for and must review, interpret, and approve all data and provide a signed
report.
IV. PERFORMANCE CHARACTERISTICS TO BE MONITORED
A. Performance Evaluation
The performance of each r adiographic and fluoroscopic unit must be evaluated at least annually. This evaluation
should include, but not be limited to, the following tests (as applicable):
http://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdfhttp://www.acr.org/~/media/FBCDC94E0E25448DAD5EE9147370A8D1.pdf
-
8/17/2019 Monitor Radio Flu Oro Equipment
3/6
TECHNICAL STANDARD Radiographic Fluoroscopic Equipment / 3
1. Integrity of unit assembly.
2. Collimation and radiation beam alignment.
3. Fluoroscopic system resolution.
4. Automatic exposure control system performance.
5.
Fluoroscopic automatic b rightness control performan ce (high-dose-rate, pulsed modes, field-of-view
[FOV] variation).
6. Image artifacts.
7. Fluoroscopic phantom image quality.
8.
kVp accuracy and reproducibility.9.
Linearity of exposure versus mA or mAs.
10. Exposure reproducibility.
11. Timer accuracy.
12. Beam quality assessment (half-value layer).
13. Fluoroscopic entrance exposure rates.
14.
Image receptor entrance exposure.
15. Equipment radiation safety functions.
16. Patient dose monitoring system calibration
17. Video and digital monitor performance.
18. Digital image receptor performance.
For further inform ation on com puted radiography [CR] and digital radiogra phy [DR] systems please see th eACR–AAPM–SIIM Practice Guideline for Digital Radiography [1].
B. Quality Control Program
A continuous quality control (QC) program must be implemented for all radiographic and fluoroscopic units. The
program should be established with t he assistance of the medical physicist. The medical physicist should identify
the person responsible for performing the tests and may choose to modify the frequency of testing based on t he
system’s usage and performance. The QC program should include, but not be lim ited to, the following tests (asapplicable):
1. Appropriateness of technique factors.
2.
Visual equipment checklists.3. Phantom images.
4. Repeat analysis.
5. Viewboxes, image monitors, and viewing conditions.
6. Laser film printer quality control.
7.
Darkroom and screen cleanliness.
8. Processor quality control.
9. Screen film speed matching.
10. Analysis of fixer retention.
11. Darkroom fog.
12.
Screen-film contact.
13. CR and DR system performance.
C. Acceptance Testing
Initial performance testing of imaging equipm ent must be performed upon installation and before clinical use.
This testing must be more comprehensive than pe riodic performance and m ust be consistent with curren t
acceptance testing practices. Electrical safety of the equipment must also be tested by appropriate personnel prior
to its initial clinical use.
http://www.acr.org/~/media/3E08C87AD6E6498D9E19769E5E5E390D.pdfhttp://www.acr.org/~/media/3E08C87AD6E6498D9E19769E5E5E390D.pdf
-
8/17/2019 Monitor Radio Flu Oro Equipment
4/6
4 / Radiographic Fluoroscopic Equipment TECHNICAL STANDARD
D. Written Survey Reports and Follow-Up Procedures
The medical physicist must provide a written report of the findings of acceptance test ing and performance
evaluation to the responsi ble physician(s), if appropriate, a nd to the professional(s) in charge of obtaining or
providing necessary service to the equipment. If appropriate, the medical physicist should initiate the required
service. Written reports must be provided in a timely manner consistent with the im portance of any adverse
findings.
If use of the equipm ent would pose imm inent danger to patients or staff, the m edical physicist must take
immediate action to prevent its use.
V. RADIATION SAFETY IN IMAGING
Radiologists, medical physicists, regist ered radiologist assistants, radiologic technologists, and all supervisin g
physicians have a responsibility for safety in the workplace by keeping radiation exposure to staff, and to society
as a whole, “as low as reasonably achievable” (ALARA) and to assure that radiation doses to individual patients
are appropriate, taking into account the possible risk from radiation exposure and the diagnostic image quality
necessary to achieve the clinical objec tive. All personnel that wor k with ionizing radiation must understand the
key principles of occupat ional and public radiation protection (justification, optimization of protection and
application of dose limits) and the pri nciples of proper management of radiation d ose to patients (justification,
optimization and the use of dose refere nce levels) http://www-
pub.iaea.org/MTCD/Publications/PDF/p1531interim_web.pdf
Nationally developed guidelines, such as the ACR’s Appropriateness Criteria®, should be used to help choose the
most appropriate imaging procedures to prevent unwarranted radiation exposure.
Facilities should have and adhere to po licies and procedures that require varying ionizing radiation exam ination
protocols (plain radiography, fluoroscopy, interventional radiology, CT) to take into account patient body habitus
(such as patient dimensions, weight, or body mass index) to optimize the relationship between minimal radiation
dose and adequate image quality. Automated dose reduction technologies available on imaging equipment should be used whenever appropriate. If such technology is not available, appropriate manual techniques should be used.
Additional information regarding patie nt radiation safety in i maging is avai lable at the Image Gently® for
children (www.imagegently.org) and I mage Wisely® for adults ( www.imagewisely.org) websites. Theseadvocacy and awareness campaigns provide free educational materials for all stakeholders involved i n imaging
(patients, technologists, referring providers, medical physicists, and radiologists).
Radiation exposures or other dose i ndices should be measured and patient radiation dose estimated for
representative examinations and types of patients by a Qualified Medical Phy sicist in accordance with the
applicable ACR technical standards. Regular auditing of patient dose indices should be performed by comparing
the facility’s dose information with national benchmarks, such as the ACR Dose Index Registry, the NCRP
Report No. 172, Reference Levels and Achievable Doses in Medical and Dental Imaging: Reco mmendations for
the United S tates or the Conference of Radiation Control Program Director’s National Evaluation of X-ray
Trends. (ACR Resolution 17 adopted in 2006 – revised in 2009, 2013, Resolution 52).
Patient radiation dose must be estimated for radiographic and fluoroscopic equipment at least annually. Tables of patient radiation exposure for representative examinations must be prepared and supplied to the facility. These
tables must be prepared using m easured radiation output data and imaging techniques provided by the facility.
These results must be compared with appropriate guide lines or recommendations when they are available [2-3 ].
The medical physicist should assist facilities in understanding and developing policies and procedures to evaluate
risks to patients, personnel, and physicians from studies and interventions requiring prolonged radiation exposure
[3-13].
-
8/17/2019 Monitor Radio Flu Oro Equipment
5/6
TECHNICAL STANDARD Radiographic Fluoroscopic Equipment / 5
ACKNOWLEDGEMENTS
This technical standard was revised according to the pro cess described under the hea ding The Process for
Developing ACR Practice Parameters and Technical Standards on t he ACR website
(http://www.acr.org/guidelines) by the Committee on Practice Parameters and Technical Standards of the
Commission on Medical Physics with assistance from the AAPM.
Principal Reviewers
Mahadevappa Mahesh, MS, PhD, FACR
John M. Boone, PhD, FACR
Committee on Practice Parameters and Technical Standards– Medical Physics
(ACR Committee responsible for sponsoring the draft through the process)
Richard A. Geise, PhD, FACR, Chair
Tariq A. Mian, PhD, FACR, Vice Chair
William K. Breeden, III, MS
Laurence E. Court, PhD
Martin W. Fraser, MS
Nicholas J. Hangiandreou, PhD
Bruce E. Hasselquist, PhD
Ralph P. Lieto, MSMahadevappa Mahesh, MS, PhD, FACR
James T. Norweck, MS
Janelle L. Park, MDDoug Pfeiffer, MS
Gerald A. White, Jr., MS, FACR
James M. Hevezi, PhD, FACR, Chair, Commission
REFERENCES
1. American College of Radiology . Practice guideline for digital radiography . http://www.acr. org/
SecondaryMainMenuCategories/quality_safety/guidelines/dx/digital_radiography.aspx. Accessed May 24,
2011.2. American College of Radiology. Practice guideline for diagnostic reference levels in medical x-ray imaging.
http://www.acr.org/SecondaryMainMenu
Categories/quality_safety/guidelines/med_phys/reference_levels.aspx. Accessed May 24, 2011.
3. Public Health Advisory: Avoidance of Serious X-Ray Induced Skin Injuries to Patients During
Fluoroscopically-Guided Procedures. Rockville, Md: Food and Drug Administration; 1994.
4. Specification and Acceptance Testing and Quality Assurance of Diagnostic X-Ray Imaging Equipment.
College Park, Md: American Association of Physicists in Medicine; 1994. AAPM Monograph 20.
5. Managing the Use of Fluoroscopy in Medical Institutions. College Park, Md: Americ an Association of
Physicists in Medicine; 1998. AAPM Report 58.
6. Cardiac Catheterization Equipment Performance. College Park, Md: American Association of Physicists in
Medicine; 2001. AAPM Report 70.
7. Structural Shielding Design for Medical X-Ray Imaging Facilities. Bethesda, Md: National Counci l onRadiation Protection and Measurements; 2004. NCRP Report 147.
8. Performance Standards for Diagnostic X-Ray Systems and their Major Components: Federal Register; June
10, 2005. Final Rule 21 CFR Part 1020.30-1020.32.
9. Balter S, Ho pewell JW, Miller DL, Wagner LK, Ze lefsky MJ. Fluoroscopically guided interventional
procedures: a review of radiation effects on patients' skin and hair. Radiology 2010;254:326-341.
10. Mahesh M. Fluoroscopy: patient radiation exposure issues. Radiographics 2001;21:1033-1045.
11. Miller DL, Balter S, Wagner LK, et al. Quality improvement guidelines for recording patient radiation dose in
the medical record. J Vasc Interv Radiol 2009;20:S200-207.
-
8/17/2019 Monitor Radio Flu Oro Equipment
6/6
6 / Radiographic Fluoroscopic Equipment TECHNICAL STANDARD
12. Seibert JA, Filipow L, Andriole K, ed. Practical Digital Imaging and PACS . Madison, Wisc: Medical Physics
Publishing; AAPM Monograph 25; 1999.
13. Stecker MS, Balter S, Towbin RB, et al. Gu idelines for patient radiation dos e management. J Vasc Interv
Radiol 2009;20:S263-273.
*Practice parameters and technical standards are publishe d annually with an effective date of October 1 in the
year in which am ended, revised or approved b y the ACR Council. For practice parameters and technical
standards published before 1999, the e ffective date was January 1 foll owing the y ear in which the practice
parameters or technical standard was amended, revised, or approved by the ACR Council.
Development Chronology for this Technical Standard
1992 (Resolution 11)
Amended (Resolution 13)
Revised 1997 (Resolution 17)
Revised 2001 (Resolution 18)
Revised 2006 (Resolution 29, 16g, 17)
Amended 2009 (Resolution 11)
Revised 2011 (Resolution 4)