nanomanufacturing metrology 1 modeling scanning microscopy for industrial metrology j. s....
TRANSCRIPT
Nanomanufacturing Metrology 1
Modeling Scanning Microscopy for Industrial Metrology
J. S. VillarrubiaNational Institute of Standards & Technology
March 9, 2010
Nanomanufacturing Metrology2
I. What goes on in an SEM?
II. How is this relevant to PED’s mission and to industry?
III. Impact of our work
This talk is about
Nanomanufacturing Metrology3
I. What goes on in an SEM?
Nanomanufacturing Metrology4
Electron gun
Electron optical column
SE Detector
Fast (E>50 eV) backscattered electrons
Electrons scatter within a finite “interaction volume.”
The size of this volume is related to —signal level —spatial res. —edge bloom size —topographic contrast
Sample
An SEM animation:
We simulate all of these phenomena using Monte Carlo methods.
Nanomanufacturing Metrology5
Our simulator (JMONSEL) allows samples to have complicated 3-dimensional shapes.
Nanomanufacturing Metrology6
Within each region we simulate the main scattering phenomena: 1. Elastic scattering
Nucleus
electron
+Z
e
Nanomanufacturing Metrology7
Within each region we simulate the main scattering phenomena: 2. Secondary electron generation
SE generation via a plasmon intermediarySE generation by impact scattering
+ nucleii
electrons
Nanomanufacturing Metrology8
Within each region we simulate the main scattering phenomena: 3. Interface scattering
2 2 4
10
8
6
4
2
Outside
Inside
An electron must “go uphill” to get out of the sample.
Sometimes it bounces. (Transmission probability < 1.)
Nanomanufacturing Metrology9
To write our simulator, we take advantage of NIST strengths.
Our division, PED, has a long history of involvement in modeling the SEM, AFM and Optics. JMONSEL, is an outgrowth of our earlier MONSEL program.
We collaborate with NIST’s Microanalysis Division (who have vast experience in modeling x-ray scattering . There is a little overlap in the physics, but a lot in 3-D sample shapes.
For our secondary electron generation model, we use methods proposed and developed by Cedric Powell and David Penn (Physics Lab). Much of our input data is the same input data used for NIST Standard Reference Database 71 (Electron Inelastic Mean Free Paths).
Our elastic scattering model is based on NIST Standard Reference Database 64 (Electron Elastic Scattering Cross Section Database).
Nanomanufacturing Metrology10
II. How is this relevant to PED’s mission and to industry?
In generalExample 1: Critical dimension metrologyExample 2: Defect metrologyExample 3: Contour metrology
Nanomanufacturing Metrology11
Generally: At least 3 kinds of questions that require modeling frequently arise in industrial microscopy:
1. What am I looking at?—Does this shape or this one go with
2. How do I optimize my measurement? — What instrument settings for best topographic contrast?— How about best material contrast?— What about highest resolution for given throughput?
3. What are the limits? —If an electrical contact were broken at the bottom (sub-surface)
would its image look any different?—If I could make a transistor gate this small, would I be able to
measure it?
Measurement 40 2002040z
40 20 0 20 40
x
40
20
0
20
40
y
50050
z
50 0 50
x
50
0
50
y
Nanomanufacturing Metrology
12
To answer these questions, the tool’s image is not enough.
For optimization, we want to know how the image would change if we changed instrument settings.
Questions about limits often arise when the image shows nothing—and we want to know: would we see X if it were there?
Questions about image interpretation (especially quantification) require us to decide whether the image we have is consistent with this or that instrument state (e.g. beam size & shape) & sample state (e.g., feature size & shape).
We need the image AND a model.
Nanomanufacturing Metrology13
Example 1: Modeling assigns edge positions for nanometer-scale critical dimension metrology
Where are the edges?With what uncertainty?
-?- -?-
w
w – 2
Nanomanufacturing Metrology14
Example 2: JMONSEL is simulating industrial defect test samples to assist defect detection
Measurement JMONSEL Simulation
Intentional defect array sample modeled at SEMATECH request.
Uses: (1) Answer questions about limits of detection. (2) Help design and optimize detection algorithms.
Nanomanufacturing Metrology15
Example 3: JMONSEL has identified a measurement artifact in contour metrology
1
1 .1
1 .2
1 .3
1 .4
1220 1218 1216 1214 1212 1210
100
80
60
40
20
0
1
1
1
1 .1
1 .1
1 .2
1 .21 .2 1 .2
1 .3
1 .3
1 .3
1 .3
1 .41 .4
1 .4
1 .4
1260 1250 1240 1230 1220 1210 1200
100
80
60
40
20
0
These brightness contours shift ~1.5 nm when they move from isolated to dense neighborhood. These are 30 nm lines and spaces.
Contours of constant brightness
Shift
This effect has since been observed experimentally.
Nanomanufacturing Metrology16
III. Impact of our work
Nanomanufacturing Metrology17
Examples from AFM and Statistics work:
AFM: We developed, published, and posted software for modeling Atomic Force Microscopy.
1. The methods are now in widespread use.2. One of our papers on the subject has been cited 182 times and counting.3. Citing papers document uses in
– CD metrology– Nanoscale mechanical characterization by nanoindentation– Characterization of biological materials– Surface roughness metrology– Studies of crystallization and melting in polymers– Tribology– …
Statistics: We alerted the industry to a bias error in linewidth roughness measurement and proposed a solution.
• The original paper was “Best metrology paper of 2005” (SPIE)• All major CD-SEM suppliers now use unbiased metrics.
Nanomanufacturing Metrology18
Examples from SEM simulation work of this project:
Our earlier SEM code (MONSEL) is being used by: • — at least 5 semiconductor electronics manufacturers, • — 5 instrument suppliers, and • — 7 universities or research institutes.
One CD-SEM vendor sent a guest researcher to work with us. We continue to interact.
A (Model-Based Library) method we developed to enable model results to be applied to metrology was recipient of Nanotech Briefs Nano50 award in 2005.
Our ongoing work is partially funded by International SEMATECH (a semiconductor industry consortium). New code to be released December 2010.
Questions?