partial discharge measurement, acquisition and...
TRANSCRIPT
CHAPTER 2
Partial Discharge Measurement,
Acquisition and Experimental Studies
28
CHAPTER 2
Partial Discharge Measurement, Acquisition and Experimental Studies
2.1 Partial Discharge- Characteristics and Types of Sources
IEC 60270 Standard on Partial Discharge Measurements defines Partial Discharge [7]
as ‘a localized electrical discharge that only partially bridges the insulation between
conductors and which may or may not occur adjacent to a conductor’. Stated otherwise,
the electrical breakdown phenomenon confined to the localized regions of the insulating
medium between two conductors at different potentials is called Partial Discharge (PD).
PD phenomena are inherently self-quenching mechanisms which may be ascribed to
the fact that the electric field intensity below a certain point is extremely low to sustain
continued growth of the discharge. This aspect may also be attributed to the space charge
(memory propagation effect) [63] formed during the discharge process which initiates
reduction of the local electric field intensity that is insufficient to sustain the discharge.
PD is also inherently a stochastic process which exhibits substantial statistical variability
in its major characteristics such as pulse amplitude, pulse shape and time of occurrence.
A few important factors that may influence the stochastic nature of PD include the
availability of the initiatory electron in the gap, memory propagation due to the effect of
residual charge (space charge) from previous pulses, role of irradiation, growth of
discharge-induced cavities etc wherein each factor may be interrelated. Further, the
29
presence of memory propagation effects makes PD phenomenon a complex non-
Markovian process imposing restrictions on the capability to predict and diagnose PD.
PD may be broadly classified based on the mechanism of discharge and site of
location of discharge [4] as internal PD, which occurs inside the dielectric and external
PD, which takes place at the interface or on the periphery of the dielectric. In general
there are five types of PD namely, (1) Internal discharges occurring in gas filled cavities
and oil filled cavities leading to breakdown (2) Surface discharges occurring in gases or
in oil in the presence of electrical stress component parallel to the surface of the dielectric
(3) Corona discharges occurring at sharp points protruding from electrodes in gases and
in liquids (4) Electrical trees originating from conducting particles, electrodes or from
cavity in solid insulation and (5) floating part discharges occurring in badly grounded
components in a High Voltage (HV) test circuit.
2.2 Methods for Partial Discharge Measurement and Their Relevance to Pattern
Recognition
Though several attributes of PD such as magnitude, rise time, recurrence rate, phase-
relationship, time interval between successive pulses, discharge inception and extinction
voltage etc characterize its occurrence, the more appropriate features from the viewpoint
of pattern recognition and also as represented in a majority of modern digital PD
measurement and acquisition systems are the phase angle of occurrence of PD pulses (φ),
magnitude of the apparent charge during discharge (q) and the discharge rate (n).
30
IEC 60270 offers essentially three modes of performing studies on PD testing and
measurement namely straight (direct) detection, polarity discrimination and balanced
bridge method. The first method is usually appropriate when the testing is carried out
under controlled laboratory testing condition wherein substantially lesser influence of
electromagnetic interference and background noise is ensured. The latter two techniques
are usually resorted to when appropriate mechanism for reduction and mitigation of noise
has to be ensured for improved signal to noise ratio. Though these schemes offer better
mechanisms for countering noise during measurements the first method is more
commonly used in practice and is more appropriate in the context of this research since
detailed study and analysis of PD signatures is carried out in appropriately shielded
controlled laboratory environment. Figure 2.1 shows the general test arrangement using
IEC 60270- Method 1.
Figure 2.1: Typical PD Measurement and Acquisition Test Setup (Ct - Equipment Under Test, Cc- Coupling Capacitor, Cp- Reference Pulse Calibrator, Z and Cq- components of the Quadrupole Arrangement)
31
From the context of obtaining the quasi-integrated PD pulses, coupling devices
(commonly referred to as quadrupole) comprise band pass filters of either wide band or
narrow band characteristics. Since identification of the pulse polarity is closely related to
the identification of the source of PD, it is evident that the appropriate choice of the
quadrupole becomes vital. However, in this research, detailed studies and analysis carried
out on the role of the polarity of the pulses in classification capability of the proposed
hybrid system of PNN clearly indicates that the classification capability is not seriously
compromised. This aspect is evident and may be attributed to the fact that the Gaussian
kernel utilized in the PNN modular versions is compact, positive and serves as an
excellent optimal basis function in the least square sense and hence normally distributes
the noise input during fitting of the data.
2.3 Fabrication of Benchmark Laboratory Models- Single and Multiple Sources
The laboratory models are fabricated to replicate the PD patterns that are
representative of the source of discharge displayed in the oscilloscope in line with the
recommendations of [64, 65]. This aspect is a basic yet vital requirement since
repeatability and reliability of the PD signatures obtained which are characteristic
features for a particular PD source would in turn serve as a mechanism to correlate and
validate the effectiveness of the test methodology in addition to providing a high
confidence level of the results interpreted during measurements and discrimination using
AI tools. Hence, the focus of this research is on appropriate validation of the
32
methodology with divergent types of benchmark models rather than providing inferences
for complex real-time insulation.
The dielectric material used for the analysis of discharge signature patterns in this
research pertaining to internal PD including that of treeing mechanism is a synthetic
polymer ‘poly methyl methacrylate- PMMA’ (commonly referred to as perspex® or
plexiglas®). Though this material has been primarily used in this research due to its visual
(transparent) characteristic, it is worth mentioning that PMMA is characterized by
excellent optical and weather resistant properties. In addition it has been observed by
several researchers [66] that the time to breakdown characteristics of PMMA is relatively
good. Additional major characteristics of PMMA includes high arc resistance (hence
providing plausible application in high voltage circuit breakers), decreasing permittivity
with increasing frequency etc. During the course of the experiments, the test object was
placed in transformer oil bath to prevent surface discharges and external flashovers.
2.3.1 Experimental Study 1: Laboratory Models simulating Multiple Source PD Patterns
12mm thick, 80 mm diameter perspex with electrode-bounded cavity of 2 mm depth
simulates internal PD while a similar setup with gliding discharges is replicated to
replicate surface discharges. Figure 2.2 (a) and Figure 2.2 (b) depicts the arrangement of
the models.
33
Figure 2.2: Laboratory Benchmark Model representing (a) Electrode Bounded Cavity (b) Surface Discharge
Corona discharge in air is replicated with a point electrode initiated from a stainless
steel rod with an approximate angle of 85° and a plane electrode of 10mm thickness and
60mm diameter. A similar arrangement immersed in transformer oil replicates oil-corona
discharges. Figure 2.3 (a) and Figure 2.3 (b) depicts the models.
Figure 2.2: Laboratory Benchmark Model simulating (a) Air Corona and (b) Oil Corona
HV
10mm Diameter, Polished Sharp Needle
Earth Electrode
Tip Radius: 100 μm
Perspex Non- Metallic Chamber
Perspex Non- Metallic Chamber
HV
10mm Diameter, Polished Sharp Needle
Tip Radius: 100 μm
Earth electrode
Transformer Oil
34
Electrode bounded cavity with air-corona is produced by inserting a needle
configuration (2 mm) from the high voltage (HV) electrode in addition to a 2 mm
electrode-bounded cavity on perspex in the HV electrode. Multiple electrode bounded
cavities replicate with nine numbers of 2mm and 4mm holes. Figure 2.4 represents the
replicated multiple source PD models. Figure 2.5 portrays a snapshot of the reference
models utilized for analysis.
Figure 2.4: Laboratory Models simulating Multiple Source PD (a) Electrode bounded cavity with air- corona (b) Multiple Cavity discharges
Figure 2.5: Laboratory models replicating Corona (Air and Oil), Electrode-bounded cavity, Multiple Cavity and Electrode Bounded Cavity with Corona
2.3.2 Experimental Study 2: Laboratory Models simulating PD initiated Electrical Treeing
PD initiated electrical treeing studies have been carried out by several researchers
since this category is one of the most important phenomena that describe degradation in
35
electrical insulation system. Since patterns of PD before and after treeing initiation are
complex and appear visibly similar, the focus of this study is on ascertaining the ability of
the proposed dynamic modular PNN version in discriminating the stochastic PD patterns
before and during treeing.
2.4 Preparation of Industrial Objects for PD Testing and Pattern Recognition
2.4.1 PD Pattern Recognition approach for Pollution Severity Initiated Flashover in
Ceramic Insulators
Contamination flashover has become a vital aspect in the design of high-voltage
outdoor insulation. Hence, this research involves conducting studies pertaining to severity
associated flashover prediction in ceramic insulators using PD signature patterns as a
technique for diagnosis. Since this research focuses on conducting predictive tests to
determine the performance of polluted insulator due to dynamic changes in the PD
pattern for varying salinity, an artificial pollution test is conducted with equivalent salt as
the pollutant to analyze the performance of the insulators. Since the clean fog test [42]
reflects the contamination mechanism occurring in industrial locations, testing is carried
out by dipping the insulator into slurry consisting of 40g of kaolin with varying levels of
salinity (39 gm/l, 57 gm/l, 67 gm/l and 91 gm/l). It is observed during studies that except
for the clean and dry case of leakage current waveforms, the remaining cases are similar
in spite of the varying polluting conditions. On the contrary, in the case of PD patterns, it
is revealed that there exist significant differences between patterns that either affect or do
36
not affect insulator flashover performance and arcing due to polluted surfaces. It is
observed that PD on polluted surfaces clearly exhibits a larger distribution with much
scattered number of PD pulses as compared to dry and wet insulator before the initiation
of scintillation led flashover. A few researchers have also reported on similar study [72-
74] that clearly delineates the importance of similar studies.
2.4.2 PD Pattern Recognition in Cross Linked Polyethylene (XPLE) Cables and
Distribution Transformer Bushings
Extruded power cables are formed by polyethylene (PE), cross linked polyethylene
(XPLE), ethylene propylene rubber (EPR) etc. Since, the advantage utilizing the PE
polymer to form cross-linked structure for better strength and increased operating
temperature limits have enabled power utilities to prefer XPLE power cables in most
electrical installations, detailed PD measurement and recognition studies have been
carried out on 33kV extruded XLPE power cable that simulate some of the significant
and typical categories of installation defects. PD related failure in power cables are
generally categorized into internal cavity discharges, surface (gliding) discharges along
interfaces due to missing semi-conductive screen, corona discharges as sharp protrusions
at end terminations and electrical treeing. The presence of protrusions on the semi-
conductive screen which are pointed in the radial direction to the cable conductor leads to
a major critical situation due to high intensity electrical stresses. Surface discharges at the
interface between air and XLPE insulation due to missing semi-conductive screen may
lead erosion culminating in degradation.
37
Hence, for the purpose of investigations four major types of artificial sources that
represent the occurrence of PD in 33kV, single core XPLE power cable samples namely
electrode bounded cavity (between the semi-conductive screen), surface discharges
(along the interface between air-XLPE insulation), corona discharges (due to sharp
protrusion at the end termination) and multiple source cavity with surface discharges is
taken up for analysis.
Electrode bounded cavity defect is replicated by forming a flat cavity of 5mm
diameter and 2mm depth formed on the surface between the XLPE dielectric and the
semi-conducting layer. The cavity is formed by removing a small part of the outer semi-
conductive screen and by making the cavity of the appropriate dimensions and then
covering the area with an aluminium tape. In addition, the measuring electrode is covered
with a grounded aluminium strip to ensure good screening during measurement. Figure
2.6 shows the schematic arrangement of the cable taken up for PD measurement.
Figure 2.6: Typical Sample of Cable simulating Electrode-bounded Cavity Defect
The artificial defect simulating surface discharges due to missing semi-conductive
screen is created by removing a small segment of the outer semi-conductive screen of the
38
cable as shown in Figure 2.7. This type of defect is most prevalent during installation of
cable and cable jointing since the discharges are produced due to electric field
enhancement at the edge of the semi-conductive layer leading to gliding discharges.
Figure 2.7: Cable Sample replicating Gliding Discharges due to Missing Semi-
Conductive Layer
Multiple sources pertaining to one major form of the complex fully overlapped PD
signature i.e., cavity with surface discharge is replicated as shown in Figure 2.8.
Figure 2.8: Cable Sample depicting Multi-Source (Electrode-bounded Cavity with
Surface Discharge) Defect
39
Detailed studies have also been carried out on various sources of PD on 11kV high
voltage bushings of a 315 kVA, 11kV/ 433V, Dyn11 distribution transformer at varying
applied voltages that replicates real-time external PD sources such as gliding discharges
due to pollution, corona discharges and partially overlapped multiple sources defects
(gliding discharges with air-corona).
2.5 Laboratory Experimental Setup for PD Pattern Recognition Studies
Investigations have been carried out using a 10kVA, 100kV, 50Hz test transformer
(MWB make) with a W.S Test Systems make (Model No.: DTM-D®) Digital PD
Measurement and Acquisition System measurable in the range 2-5000pC. A Tektronix
built-in oscilloscope (TDS 2002B) alongside a tunable adjustable filter-insert (Model:
DFT-1) with a variable center frequency of range 600 kHz- 2400 kHz at a bandwidth of
9 kHz is utilized for acquiring the appropriate PD pulses. Facility is provided to display
the measured PD pulses either in pico-coulomb (pC) or in milli-volt (mV) in accordance
with stipulations laid down by IEC 60270. The direct detection and measurement test
setup as recommended in IEC 60270 is utilized in carrying out the studies in this research
work since the tests have been carried out in a controlled laboratory environment thus
obviating the need for alternative strategies for noise suppression. Notwithstanding, the
PD measurement and acquisition system comprises window gating facility to mask and
suppress unwanted background noise during measurement. Further, a 1nF coupling
capacitor is also included in the test setup to improve the transfer characteristics of the
test circuit. Calibration of the test setup is carried out using a reference calibrator (Model:
40
PDG®) and in line with the requirements of IEC 60270. Figure 2.9 shows the PD
measurement and acquisition system used in this research.
Figure 2.9: Digital PD Measurement and Acquisition System and a typical display of PD patterns on a sinusoidal time base for air-corona discharges 2.5.1 Experimental Test setup for Laboratory Models simulating Multiple Source PD
Patterns
Figure 2.10 and Figure 2.11 depicts the typical test arrangement utilized for carrying
out PD pattern recognition studies on laboratory benchmark models. It is worth
mentioning that the tests were performed under controlled laboratory conditions with
appropriate shielding of the laboratory test setup including that of the control panel along
with the measurement system.
41
Figure 2.10: Typical Test Arrangement depicting PD pattern recognition system for Laboratory Models
Figure 2.11: Snapshot depicting the typical experimental layout for PD studies on Laboratory Benchmark Models
Quadrople Calibrator
Coupling Capacitor
Test Transformer
Test Object
42
2.5.2 Experimental Test setup for Laboratory Models replicating studies for PD initiated
Electrical Treeing
This research envisages studies based on samples comprising virgin moulds of
perspex blocks of 80mm diameter and 12mm thickness. The high voltage electrode
comprises a stainless steel needle of 1mm thickness with a tip radius 500μm inserted to a
depth of 4mm by hard pressing. The sample is placed in a test cell containing transformer
oil to prevent flashover. Figure 2.12 and Figure 2.13 replicates the test setup utilized for
carrying out electrical treeing studies.
Figure 2.12: Typical Test Arrangement depicting PD pattern recognition system for Electrical Treeing Studies
43
Figure 2.13: Snapshot of the Typical Test Arrangement utilized for Electrical Treeing Studies
2.5.3 Experimental Test setup for PD Pattern Recognition for Pollution Severity initiated
Flashover in Ceramic Insulators
A series of experiments are performed on four insulators of varying dimensions and
salinities. Measurement of correlated PD and leakage current were studied for the
following cases: 1. pin of the insulator, 2. cap of the insulator and 3. pin and cap of
insulator. The generic test setup for acquiring PD signatures for analysis is shown in
Figure 2.14.
44
Figure 2.14: Generic layout of the test Arrangement depicting PD pattern recognition system for Prediction of Flashover due to Pollution Severity
2.6 Partial Discharge Test Measurement and Data Acquisition
2.6.1 Experimental Study 1: Measurement and Data Acquisition for Laboratory Models
simulating Multiple Source PD Patterns
The signature analysis was carried out after due verification of patterns acquired for
reference benchmark models. These patterns were verified for their appropriateness with
the reference patterns given in IEC Guidelines [8] and [64]. Figure 2.15 Figure 2.16,
igure 2.17and Figure 2.18 learly demonstrate the non-stationary behaviour of PD
signatures due to various sources of discharges for varying applied voltages.
45
Figure 2.15: Typical Single Source PD Signature of Air Corona Discharges
Figure 2.16: Single Source PD Signature of Single Electrode Bounded Cavity
Discharges
Figure 2.17: Multiple Source PD Signature-Electrode bounded Cavity with Air Corona
46
Figure 2.18: Single Source PD Signature of Surface Discharges in Air (gliding
discharges)
Acquisition of PD data is carried out initially for small sets of patterns due to both
single and multiple PD sources. The fist data comprises a total of two sets of training
database i.e. 20 and 25 sets. A total of 52 PD fingerprints samples are indicated in Table
2.1. The second study comprises large database PD patterns of defects for varying
applied voltages as indicated in Table 2.2. It is pertinent to note that the patterns acquired
exhibit the statistical variations in the pulse patterns for each cycle of the sinusoidal
voltage that display the inherent non–Markovian nature of PD making the task of
classification difficult. The task becomes even more challenging due to varying applied
voltages. In order to ascertain the capability of the modular PNN versions in classifying
PD signatures that exhibit non-stationary and complex stochastic behaviour, preliminary
studies in each case has been conducted by digitally superimposing PD signatures
obtained for single source defects on similar lines with [40], since these patterns may
result in even more onerous challenges during recognition of partially and fully
overlapped patterns. Detailed analyses have clearly indicated the robustness of the
scheme in discriminating digitally superimposed signatures also. This aspect is reported
based on this research study in [118] and deliberated in detail in Chapter 5, Section 5.2.
47
Table 2.1: Small Dataset PD Signatures for Benchmark Laboratory Models
PD Category
Type of PD Label for Classification
No. of PD Patterns
1 Electrode Bounded Cavity EC 8 2 Surface Discharge S 10 3 Oil-Corona OC 10 4 Air-Corona C 6 5 Electrode Bounded Cavity with
Corona ECC 8
6 Bounded Cavity with Surface Discharge
VS 10
Table 2.2: Database pertaining to different PD Sources for Various Applied Voltages
PD Category Type of PD Applied Voltage
(kV) Total No. of
Patterns
1
Electrode Bounded Cavity
7.28 120 9.1
9.6
2
Air-Corona 13.65 120 20.93
22.75
3 Oil-Corona 20.93
120 29.12 31.85
4
Electrode Bounded Cavity with Air-Corona
9.1 120 9.6
13.6
5 Multiple Electrode Bounded Cavity
9.1 120 9.6
13.6
2.6.2 Experimental Study 2: Measurement and Data Acquisition for Laboratory Models
simulating PD initiated treeing patterns
As discussed in the previous sections, the fabricated sample is placed in a test cell
containing transformer oil in order to prevent surface discharges and flashover during
48
testing and analysis. In addition, since it has been observed by a few researchers that the
signature patterns of PD before and during treeing inception are not quite distinctive, the
patterns are selected such that only after a threshold tree length [5] the patterns provide
distinct details regarding ‘no tree’ and ‘tree’ patterns thus enabling appropriate
identification. Figure 2.19 show photographs of the test arrangement and the nature of
treeing growth during studies.
(a) Branch-type dendrite treeing initiation (b) Branch-type tree growth Figure 2.19: Photographs of dendrite morphology propagation and growth in Electrical Treeing Studies
Studies have been carried out at an applied voltage of 23.5kV and a total of 200
datasets pertaining to PD and tree initiated patterns have been acquired. Since it has been
observed by researchers that the PD patterns and those pertaining to inception of tree are
not quite distinct, the patterns utilized for identification are so chosen that data are
obtained for ‘no tree’ and ‘tree’ patterns.
49
2.6.3 Experimental Study 3: Measurement and Data Acquisition of PD Signature
Patterns for Pollution Severity Initiated Flashover Studies in Ceramic Insulators
Measurement of correlated PD and leakage current were studied for the following
cases: 1. pin of the ceramic insulator, 2. cap of the ceramic insulator and 3. both pin and
cap of ceramic insulator. Table 2.3 indicates the characteristics and details regarding the
studies conducted on individual disc insulators for varying salinities.
Table 2.3: Characteristics and Dimensions of Disc Insulator Samples
Sample No.
Characteristics and Dimensions Creepage Distance
(mm)
Diameter (mm)
Cantilever Strength
(kN)
Glazing Height of Cap
Diameter of Cap
Salinity (g/l)
Sample 1 340 265 45 Dark Brown
85 828 39
Sample 2 327 264 45 Light Brown
90 878 67
Sample 3 320 270 45 Dark Brown
90 873 57
Sample 4 295 265 45 Light Brown
104 968 91
Table 2.4 indicates study carried out for two sets of samples conducted on a single 12kV
insulator indicating the various stages of dry-band formation leading to pollution severity
initiated flashover studies.
50
Table 2.4: Database of PD patterns during to pollution performance studies in ceramic insulators (Sample 2 and Sample 3)
Category of PD
Type of PD Applied Voltage (kV)
No. of Patterns
Sample 2 Sample 3
1
PD (wetted without Dry band arcing)
6.1 5.4 40
2
PD during Dry band (at pin end)
2.8 4.8 40 3.6 5.6
3 PD during Dry band (at cap end)
6.4 5.9 80 10.7 6.2
4 PD during Dry band (at pin and cap)
12.4 7.4 40 16.3 8.1
A few researchers have also reported on similar study [73] that clearly delineates the
importance of similar studies. Figure 2.20 clearly illustrates this aspect. Typical snapshots
of the studies carried out during pollution performance studies in insulators are depicted
in Figure 2.21.
Figure 2.20: PD and leakage current waveforms during Pollution Performance Test on Insulators under wet-condition with 67g/ l: (a) Applied voltage 16.6kV; (b) Applied voltage of 24.1 kV
51
(a) Arcing at Cap (b) Arcing at Cap (c) Flashover Initiation (d) Flashover and Pin Figure 2.21: Typical Images of Tests on Insulators Prediction of Flashover due to Pollution Severity
2.6.4 Experimental Study 4: Measurement and Data Acquisition of PD Signature
Patterns for Multiple Source Defects in XPLE Power Cables and Transformer
Bushing
PD signature datasets obtained from various samples of cables and 11kV transformer
bushing is indicated in Table 2.5 and Table 2.6 respectively. Incidentally, it is significant
to note that the PD signatures have been obtained for two major forms of multi-source
defects namely the partially overlapped signatures (cavity with air corona) and fully
overlapped signatures (cavity with surface discharge) in order to ascertain the capability
of the proposed hybrid PNN variant (S-Transform versions) in providing solutions to
recognition of non-stationary PD pulse patterns.
52
Table 2.5: Dataset of PD Signatures obtained from Samples of 33kV XPLE Power
Cable
Table 2.6: Dataset of PD Signatures obtained from 11kV Bushing of Distribution
Transformer
PD data acquisition is carried out using PD Gold® software developed by HV
Solution Inc, U.K. which is interfaced with the PD detection and measurement system.
PDGold©
data acquisition software provides facility to acquire high resolution PD signals
PD Category
Type of PD Discharge Inception Voltage
Applied Voltage (kV)
Total No. of Patterns
1
Electrode Bounded Cavity 2.9
3.2 150 3.6
4
2
Air-Corona 3.4
3.8 150 4.2
4.4
3 Surface Discharge 3.6
4 150 4.5
4.8
4
Electrode Bounded Cavity with Surface Discharge
2.8
4 150 4.4
4.8
PD Category
Type of PD Discharge Inception Voltage
Applied Voltage (kV)
Total No. of Patterns
1
Air-Corona 5.8
6.2 150 7.5
8.1
2 Surface Discharge 4.4
4.8 150 5.6
7.1
3 Air Corona with Surface Discharge
6 6.8
150 7.3 7.8
53
at a sample rate of 1 sample per 2.5 nanoseconds (ns). The unit also detects PD for 50Hz
power cycle which allows the user to observe the shape of the PD pulses detected in
addition to acquiring the phase resolved PD (PRPD) patterns in real-time. The software
also provides a PD threshold level for recording the number of PD pulses (count) per
cycle [74]. The set-up of the unit for an on-line PD testing requires filling a user-friendly
‘Data Input Form’ with dropdown menus. The pulses are displayed in both sinusoidal and
elliptical forms selectable in auto or manual mode. In manual mode, the user may record
the data for a considerable period though the data acquired in this study is limited to
duration of 5-10 minutes. The set-up of the unit for an on-line PD testing requires filling
a user-friendly “Data Input Form” with dropdown menus as indicated in Figure 2.22. The
user fills the data input form with appropriate information and an automatic test report is
generated by the software for ease of recall.
Figure 2.22: Screenshot of Data Input Form during PD Measurement and Acquisition
54
The software acquires the PD pulses synchronously on a 50 Hz power cycle so as to
display the pulses in both sinusoidal and elliptical forms selectable in auto or manual
mode. In manual mode, the user may record the data for a period of 5-10 minutes which
is acquired from a minimum of 240 to a maximum of 750 waveforms per channel. The
recording can be stopped whenever necessary by clicking on the “Stop” button. Figure
2.23 shows typical PD pulses acquired during the testing, measurement and acquisition
process.
Figure 2.23: Typical snapshot of the output from Digital PD Measurement and Acquisition System of PD patterns on an elliptical-time base of electrode bounded cavity with air-corona Figure 2.24 and Figure 2.25 show the oscilloscopic display and the screenshot of the PD
plot respectively (φ -q and q-n) for the test carried out for a period of 5 minutes.
55
Figure 2.24: Typical waveform representation on Oscilloscope depicting (a) Single Electrode Bounded Cavity with Air- Corona (b) Air-Corona on Sinusoidal Base
Figure 2.25: Typical Plot of PD patterns of φ -q and q-n exhibited by Single Electrode Bounded Cavity Benchmark Model
2.7 Summary
The salient characteristics, properties and terminologies pertaining to PD are
deliberated in detail for the context of PD pattern recognition. The role of the
appropriateness of the type of testing method in pattern recognition and its limitation is
explained. The need for validation of PD signatures acquired during testing and
56
measurement, rationale behind the procedure of utilizing PD signature database based on
laboratory benchmark models for pattern recognition studies and its impact on
discrimination is summarized. Detailed specifications on the formulation of the test
circuit arrangement, experimental setup and procedure for data acquisition is discussed
from the point of view of pattern recognition and discrimination.
A few major aspects with regard to PD testing, measurement and acquisition are
summarized below:
1. Need for appropriate choice of PD measurement circuit among the various methods
recommended by IEC 60270 such as direct detection method, bridge detection method,
pulse discrimination method to suit laboratory or site conditions become vital during
measurement. In all cases appropriate PD calibration has to be carried out ensuring that in
each case the signal to noise ratio is high based on appropriate choice of the bandwidth of
tunable filter during acquisition.
2. Need to carry out proper cross- validation of the acquired PD pulse signatures for
various benchmark models in line with the recommendations given in [8, 64] since
patterns are observed to be highly non-Markovian and non-stationary in nature making
the task of discrimination difficult.