pore size of hplc reversed phase materials and insulin quantification

Upload: bujie84

Post on 04-Apr-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/30/2019 PORE SIZE OF HPLC REVERSED PHASE MATERIALS AND INSULIN QUANTIFICATION

    1/1

    The different pore size brought about little or no effect on

    the retention time (Figure 2) and in agreement with the

    previous report. 10 The absence of effect on retention

    time suggests that this analytical procedure is robust

    with respect to different pore size column. The

    calibration curve constructed was evaluated on its

    correlation coefficient. The peak area of the insulin

    appeared linear in the range of 0.1-0.5 mg/ml for both

    columns (Table 1). The correlation coefficients exceeded

    the proposed recommended value of 0.999.9 Thus, bothof the columns signify a good linearity of analytical

    method (Table 1). Accuracy and precision of the samples

    were calculated for between-day and within-day. The

    results were calculated using the equation of observed

    mean concentration over estimated concentration for

    accuracy and standard deviation over mean for

    precision. Both columns demonstrate a full recovery of

    insulin over the range of 0.1 -0.5mg/ml of insulin (Table

    2). The results depicted the close proximity between the

    obtained and estimated results hence, it can be

    concluded that the analytical method is accurate. The

    column with pore size 300 is more sensitive byexhibiting low LOD and LOQ values 80 pore sizecolumn (Table 3). It shows that 300 pore size columncan detect lowest amount of insulin and a reliable lower

    amount of insulin for quantification. The system

    suitability was studied by using 0.3mg/ml of insulin

    concentration. The mean, SD and %RSD of retention

    time and peak area were calculated as shown in Table 5.

    The retention time and peak area %RSD of both 80and 300 pore size columns are all under the 2%acceptance criterion therefore both are suitable for

    insulin quantification. The stability of insulin was

    investigated by measuring the concentration changes in

    the standard samples over time. It was assessed by

    subjecting the insulin samples at room temperature,

    25oC for 0 and 24h. The results showed that analytical

    system with the 300 pore size column is more stablewith the mean 98.97% recovery compared to 97.11%

    recovery for 80 pore size column is significantlydifferent (Table 4, p