pore size of hplc reversed phase materials and insulin quantification
TRANSCRIPT
-
7/30/2019 PORE SIZE OF HPLC REVERSED PHASE MATERIALS AND INSULIN QUANTIFICATION
1/1
The different pore size brought about little or no effect on
the retention time (Figure 2) and in agreement with the
previous report. 10 The absence of effect on retention
time suggests that this analytical procedure is robust
with respect to different pore size column. The
calibration curve constructed was evaluated on its
correlation coefficient. The peak area of the insulin
appeared linear in the range of 0.1-0.5 mg/ml for both
columns (Table 1). The correlation coefficients exceeded
the proposed recommended value of 0.999.9 Thus, bothof the columns signify a good linearity of analytical
method (Table 1). Accuracy and precision of the samples
were calculated for between-day and within-day. The
results were calculated using the equation of observed
mean concentration over estimated concentration for
accuracy and standard deviation over mean for
precision. Both columns demonstrate a full recovery of
insulin over the range of 0.1 -0.5mg/ml of insulin (Table
2). The results depicted the close proximity between the
obtained and estimated results hence, it can be
concluded that the analytical method is accurate. The
column with pore size 300 is more sensitive byexhibiting low LOD and LOQ values 80 pore sizecolumn (Table 3). It shows that 300 pore size columncan detect lowest amount of insulin and a reliable lower
amount of insulin for quantification. The system
suitability was studied by using 0.3mg/ml of insulin
concentration. The mean, SD and %RSD of retention
time and peak area were calculated as shown in Table 5.
The retention time and peak area %RSD of both 80and 300 pore size columns are all under the 2%acceptance criterion therefore both are suitable for
insulin quantification. The stability of insulin was
investigated by measuring the concentration changes in
the standard samples over time. It was assessed by
subjecting the insulin samples at room temperature,
25oC for 0 and 24h. The results showed that analytical
system with the 300 pore size column is more stablewith the mean 98.97% recovery compared to 97.11%
recovery for 80 pore size column is significantlydifferent (Table 4, p