preparation of poly(thionine) modified screen-printed carbon electrode and its application to...
TRANSCRIPT
Biosensors and Bioelectronics 19 (2003) 277�/282
www.elsevier.com/locate/bios
Preparation of poly(thionine) modified screen-printed carbonelectrode and its application to determine NADH in flow injection
analysis system
Qiang Gao, Xiaoqiang Cui, Fan Yang, Ying Ma, Xiurong Yang *
State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Renmin Street 159,
Changchun, Jilin 130022, China
Received 16 November 2002; received in revised form 2 April 2003; accepted 15 May 2003
Abstract
A poly(thionine) modified screen-printed carbon electrode has been prepared by an electrooxidative polymerization of thionine in
neutral phosphate buffer. The modified electrodes are found to give stable and reproducible electrocatlytic responses to NADH and
exhibit good stability. Several techniques, including cyclic voltammetry, X-ray photoelectron spectroscopy (XPS) and scanning
electron microscopy (SEM), have been employed to characterize the poly(thionine) film. Further, the modified screen-printed
carbon electrode was found to be promising as an amperometric detector for the flow injection analysis (FIA) of NADH, typically
with a dynamic range of 5�/100 mM.
# 2003 Elsevier B.V. All rights reserved.
Keywords: Electropolymerization; Thionine; NADH; Screen-printed carbon electrode; FIA
1. Introduction
The determination of reduced b-dihydronicotinamide
adenine dinucleotide (NADH) is very important in
enzyme assays, due to its participation in the enzymatic
catalysis of more than 250 dehydrogenases useful both
in bioprocesses and analytical applications (Katakis and
Dominguez, 1997; Lobo et al., 1997). The electrochemi-
cal oxidation of NADH has been studied (Gorton, 1986)
and attempts have been made to use surface-modified
electrodes in order to speed up the rate of heterogeneous
electron transfer between an electrode and NADH in
solution (Gorton and Dominguez, 2002).The use of electropolymerization to prepare modified
electrodes is a practical approach to the direct formation
of films on small or irregularly shaped electrodes,
especially microelectrodes or microarray electrodes
(Emr and Yacynych, 1995). The method has also been
used to prepare modified electrodes with electrocatalytic
* Corresponding author. Tel.: �/86-431-568-9278; fax: �/86-431-
568-9711.
E-mail address: [email protected] (X. Yang).
0956-5663/03/$ - see front matter # 2003 Elsevier B.V. All rights reserved.
doi:10.1016/S0956-5663(03)00212-4
activity toward NADH oxidation (Pariente et al., 1995;
Lobo et al., 1996; Bartlett et al., 1997; Zhou et al., 1998;
Cai and Xue, 1997; Silber et al., 1996; Chen and Lin,
2001). Thionine, as a redox dye, has been studied
extensively due to its potential utility in photogalvanic
cell (Albery et al., 1979) and sensor (Ruan et al., 1998;
Xiao et al., 1999) applications. Albery et al. (1979)
reported on the modification of platinum and tin
dioxide electrodes with thionine. They applied a positive
voltage for 15 min (1�/1.5 V vs. SCE) on electrodes
immersed in a thionine solution containing 0.05 M
H2SO4, and estimated the organic coverage to be of the
order of 20 monolayers. The thionine film was, however,
immediately disordered when the pH value of the
solution was increased to higher than pH 4. Another
way for the preparation of poly(thionine) modified
electrodes is potential-sweep electrolysis in the potential
range of �/0.2 to �/1.4 (vs. Ag/AgCl) in an acetonitrile
solution containing NaClO4 and thionine. Using this
method, modified electrodes with catalytic activity for
oxidation of NADH have been prepared on the basal-
plane pyrolytic graphite electrode (Ohsaka et al., 1993)
and microband gold electrode (Cai et al., 1995). In
Q. Gao et al. / Biosensors and Bioelectronics 19 (2003) 277�/282278
addition, Yang et al. (1999) have presented a two-step
method for the electropolymerization of thionine in
neutral phosphate buffer on glassy carbon electrodes
and constructed a H2O2 biosensor.Screen-printed electrode has combined advantages of
being simple, inexpensive and versatile (Alvarez-Icaza
and Bilitewski, 1993). Reports about the use of electro-
polymerization of redox dyes on screen-printed carbon
electrodes are very rare. Much effort has been directed
towards the goal of preparing new electrode materials
that can exhibit a facile electrochemical oxidation of
NADH (Wang et al., 1998; Sprules et al., 1995;Aviamescu et al., 2002). We found that the poly(thio-
nine) modified screen-printed carbon electrodes, which
can be prepared by the electropolymerization of thio-
nine in neutral phosphate buffer, considerably electro-
catalyzes the oxidation of NADH and have a good
stability.
In the present paper, we reported on the preparation
of the poly(thionine) modified screen-printed carbonelectrodes and its electrocatalysis for NADH oxidation.
X-ray photoelectron spectroscopy (XPS) and scanning
electron microscopy (SEM) were employed to charac-
terize the modified electrodes. The determination of
NADH using the modified electrodes was also per-
formed in a flow injection analysis (FIA) system.
2. Experimental
2.1. Chemicals and reagents
Reduced b-dihydronicotinamide adenine dinucleotide
(NADH, AMRESCO) was used as received. Thionine
(dye content 85%) was obtained from Aldrich. Ascorbic
acid, uric acid and dopamine were purchased from
Sigma. Carbon ink (Electrodag 423SS) was obtainedfrom Acheson Colloids (Japan).
Unless otherwise stated all chemicals and reagents
used are of analytical grade. Buffers were prepared using
water from a Milli-Q ultra-pure water system. 0.2 M
phosphate buffer (pH 6.9) was prepared by mixing the
stock solution of potassium dihydrophosphate and
sodium hydroxide.
2.2. Apparatus
Amperometric and cyclic voltammetric experiments
were performed with a CHI 832 (Shanghai, China). All
experiments were carried out with a conventional three-
electrode system with the screen-printed carbon elec-
trode as working electrode, a platinum foil as counter
electrode, and an Ag/AgCl (saturated potassium chlo-ride) as reference electrode. All solutions tested were
thoroughly deoxygenated by bubbling pure nitrogen,
and a continuous flow of nitrogen was maintained over
the solution during experiments at room temperature
(239/2 8C).
SEM was performed on a JXA-840 (JEOX, Japan)
apparatus. The acceleration voltage was 20 kV. Sampleswere gold sputtered prior to SEM measurements.
XPS measurements were conducted with an ESCA-
LAB MK II spectrometer (VG Co., UK) with a Mg Karadiation (hg�/1253.6 eV) as the X-ray source and a
pass energy of 20 eV. The pressure inside the analyzer
was maintained at 6.5�/10�7 Pa. Data analysis was
based on deconvolution of a high-resolution composite
XPS peak into peaks of the individual species indifferent states. The deconvolution was carried out by
a nonlinear regression analysis provided by the instru-
ment, with position, height, and width of each individual
Gaussian peak as variable parameters. The criterion for
the best fit was the good agreement between the
experimental points and fitted curves.
The flow injection system comprised of an ISIF-C
brainpower flow injection analyzer (Ruimai Company,Xi’an, China) equipped with two three-channel peristal-
tic pumps, a rotary injection valve with a 50 ml of sample
loop and a home-made flow cell.
2.3. Preparation of base electrodes and poly(thionine)-
modified screen-printed carbon electrodes
The screen-printed carbon electrodes used as a base
electrodes were prepared by the method described in thepaper (Zhang et al., 2001). The poly(thionine) modified
electrodes were prepared by a potential-sweep electro-
lysis at 50 mV s�1 in the potential range of �/0.5 to 1.1
V (vs. Ag/AgCl) in phosphate buffer containing 0.1 mM
thionine. The electrodes thus prepared were treated in
an ultrasonic bath for a few minutes to dissolve thionine
monomer adsorbed on the electrode surface or trapped
in the polymer matrix. The surface concentration of theelectroactive moieties within the poly(thionine) films
could be controlled by appropriately choosing the
number of the cyclic scans. Unless otherwise stated,
the poly(thionine) modified electrodes with cyclic scans
of 25 times were used.
2.4. Flow injection analysis
A home-made three-electrode flow cell (20 ml volume),
equipped with a poly(thionine) modified screen-printed
carbon electrode, a stainless steel counter electrode and
an Ag/AgCl reference electrode, was used. The
poly(thionine) modified screen-printed carbon electrode
was inserted into the flow-through cell and connected
finely with a CHI 832 Potentiostat. Then, a carrier
electrolyte (0.2 M phosphate buffer pH 6.9) wasdelivered and a potential (0.1 V vs. Ag/AgCl) was
applied to obtain a stable baseline. Amperometric
signals were recorded using a PC 586 when NADH
Q. Gao et al. / Biosensors and Bioelectronics 19 (2003) 277�/282 279
solution was injected with a flow rate of 0.5 ml min�1.
Peak height of the oxidation current was used as a signal
and three injections were made for each concentration of
NADH standard.
3. Result and discussion
3.1. Properties of the poly(thionine) modified screen-
printed carbon electrode
Cyclic voltammograms (Fig. 1) showed the growth
process of the poly(thionine) film with consecutivecyclic. The first cyclic voltammogram (initial potential
of 0.4 V) showed a pair of sharp reversible peaks at the
region of the monomer redox peak, which could be due
to thionine monomer redox on the screen-printed
carbon electrode. On further potential scanning, a
well-defined shoulder peak was observed at a peak
potential of 0.9 V (vs. Ag/AgCl). The irreversible anodic
reaction corresponded to the oxidation of the �/NH2
groups of the thionine molecule. During the thionine
electropolymerization, a pair of new reversible peaks
with a peak potential of 0.1 V grew up, which increased
gradually and shifted to a cathodic peak potential (Epc)
of �/0.1 V and an anodic peak potential (Epa) of 0.27 V
with increasing scan number. This demonstrated that
the polymer film was formed on the screen-printed
carbon electrode.The potential sweep range, especially the upper
potential limit, was most important factor for preparing
poly(thionine) modified electrode. If the potential sweep
was confined to the range that upper limit was below 0.8
V, then a simple CV corresponding to the two-electron
reduction of thionine to leucothionine was observed
while polymerized thionine was not obtained. The
Fig. 1. Cyclic voltammograms recorded during the electropolymeriza-
tion of thionine in phosphate buffer (pH 6.9) containing 0.1 mM
thionine. Scan rate, 50 mV s�1. Initial potential, 0.4 V (vs. Ag/AgCl).
mechanism of thionine electropolymerization has been
proposed and confirmed (Bauldreay and Archer, 1983;
Hutchinson and Hester, 1984). Schlereth and Karyakin
(1995) have reported that the cation-radical species were
formed at about 0.8 V for the electropolymerization of
phenoxazines or phenothiazines if the parent monomer
has primary (or secondary) amino groups as ring
substituents. On the bases of these discussions, it was
suggested that the oxidation potential not less than 0.9 V
was mainly based on the same reason. In order to
achieve the formation of the poly(thionine) film, the
electrode potential must be larger than the potential
where the oxidation of NH2 groups of the thionine
molecule (probably to the cation radicals, Bauldreay
and Archer, 1983) occur, as the usual case for most
NH2-group containing aromatic compounds (Ohsaka et
al., 1984, 1991; Kunimura et al., 1988).
Fig. 2 showed the cyclic voltammograms of the
poly(thionine) modified screen-printed carbon electrode
at various scan rates. The relationship between peak
current and scan rate was found to be linearly with v1/2,
which indicates that the charge-transport process within
the film was diffusion-controlled. Further successive
potential scanning at 100 mV s�1 for 1 h caused a
decrease in current response less than 5%.
The electrochemical behavior of poly(thionine) as
other polymer dyes was affected by pH (Torstensson
and Gorton, 1981). The anodic and cathodic peak
potentials shifted negatively with increasing pH in
phosphate buffer. The E8 value decreased by 55 mV/
pH between pH 6 and 8 for the immobilized poly(thio-
nine), which was close to the expected Nernstian value
of 59 mV for a two-electron, two-proton process.
It was reported that the screen-printed electrode and
conventional electrode differed in their electrochemical
behavior (Yoon and Kim, 1996). They were suggested
Fig. 2. Cycle voltammograms of the poly(thionine) modified screen-
printed carbon electrode at various potential scan rates. Potential scan
rates; 10, 20, 40, 60, 80, 100 mV s�1.
Fig. 3. Scanning electron micrograph of poly(thionine) modified
screen-printed carbon electrode. The modified electrode was prepared
by repeating 100 times of potential sweep in phosphate buffer (pH 6.9)
containing 0.1 mM thionine. Magnification was 1000.
Q. Gao et al. / Biosensors and Bioelectronics 19 (2003) 277�/282280
that the difference might be attributed to the increased
surface area of the screen-printed electrode due to rough
and porous structure. Fig. 3 showed the typical electronmicrograph of the screen-printed electrode at low
magnification (1000-fold), a rough and jagged structure
with randomly distributed carbon particles was ob-
served, which supports the presumption that the thick
film electrode acted like a huge array of microelectrodes.
Fig. 4. (A) XPS spectra of N1s for electropolyzerated thionine (a) and
adsorbed thionine (b). (B) Resolved XPS peaks of electropolymer-
izated thionine. The poly(thionine) modified screen-printed electrode
was prepared by repeating 100 times of potential sweep in phosphate
buffer (pH 6.9) containing 0.1 mM thionine.
3.2. X-ray photoelectron spectroscopy (XPS) analysis
High-resolution XPS spectra of N1s were recorded
and the binding energy of the nitrogen of poly(thionine)
has been investigated. Fig. 4 showed the XPS peaks of
N1s for thionine molecules in the adsorbed film (Fig.
4A.b) and the polymer film (Fig. 4A.a). For the
adsorbed thionine molecules on the electrodes, a
screen-printed carbon electrode was immersed in thethionine-containing phosphate buffer for an hour.
Adsorption process could be followed by cyclic voltam-
metry with potential range from �/0.5 to 0.6 V (vs. Ag/
AgCl), which was characterized by the fact that the
cyclic voltammograms reached a steady-state. The
detailed analysis of the N1s peak of poly(thionine)
showed three different peaks, as shown in Fig. 4B,
indicating that three chemical states of nitrogen atomsexisted in the polymer film. The one at 400.1 eV (42.9%)
was corresponding to the amino nitrogen atoms, while
another at 398.1 eV (29.5%) was ascribed to the
heterocyclic nitrogen atoms. The third one at 399.3 eV
(27.6%) came from a new state of nitrogen atoms.
According to Kessel and Schultze (1990) and Camalli et
al. (1990), this kind of nitrogen atoms might be
originated from the bonding via an amino nitrogenatom in one thionine molecule to a carbon atom in
another one. Thus, the electropolymerization of thio-
nine took place under our experimental conditions.
3.3. Electrocatalysis of poly(thionine)-modified
electrodes for oxidation of NADH
For further electrochemical measurements, the elec-
trode was then transferred into the phosphate buffer
with or without NADH. Fig. 5 showed typical voltam-
mograms demonstrating the electrocatalytic activity of
the poly(thionine) modified screen-printed carbon elec-
trodes for the oxidation of NADH. The reversible redox
response of the poly(thionine) film can be observed over
a wide range of potentials ca. �/0.2 to 0.2 V (vs. Ag/
AgCl) and its center was located ca. 0.05 V (vs. Ag/
AgCl) as shown in voltammogram a. Voltammogram b
observed in the presence of NADH showed an enhanced
oxidation current and a large negative shift in the anodic
peak potential of about 300 mV, compared with that
obtained at the bare screen-printed carbon electrode
(voltammogram c). These results demonstrated the
Fig. 5. Cyclic voltammograms of (a, b) poly(thionine) modified screen-
printed carbon electrode in the absence (a) and presence (b, c) of 1 mM
NADH. Voltammogram (c) was obtained at a bare screen-printed
carbon electrode. Scan rate, 10 mV s�1.
Q. Gao et al. / Biosensors and Bioelectronics 19 (2003) 277�/282 281
electrocatalytic oxidation of NADH by the poly(thio-
nine) film.
3.4. Detection of NADH in flow injection system
The flow-rate dependence of the current response at aconstant concentration of NADH was examined by
recording the peak currents at different flow rates. The
current response was found to decrease with increasing
flow rate. At a flow rate of 0.5 ml min�1, the
determination of NADH can be performed in 3 min
including the sampling and washing, giving a through-
out of about 20 samples per hour with a relative
standard deviation (R.S.D.) of 4.5%.
Fig. 6. Flow-injection responses to NADH injection at poly(thionine)
modified screen-printed carbon electrode. The concentration of
NADH solution was increased from 5�/10�6 to 1.5�/10�3 M.
Applied potential, 0.1 V (vs. Ag/AgCl). Flow rate, 0.5 ml min�1.
Electrolyte and carrier, phosphate buffer (pH 6.9).
Fig. 6 displayed the flow-injection response of the
poly(thionine) modified screen-printed carbon electro-
des for NADH solution of increasing concentration
from 5 mM to 1.5 mM. Well-defined and sharp peakswere observed using a detection potential of 0.1 V (vs.
Ag/AgCl). The flow injection peak currents were
proportional to the NADH concentration. At higher
concentrations, there appeared to be a leveling off of the
response, likely due to kinetic limitation. At the lower
concentrations, a linear correlation was observed. The
present modified electrodes possessed a dynamic range
of 5�/100 mM. The resulting calibration plot had a slopeof 1.14 mA mM�1 and correlation coefficient, 0.999. A
detection limit of 3�/10�6 M can be estimated on the
basis of the signal-to-noise (S/N�/3). In all cases the
response was rapid and reproducible. After an initial
loss activity, the electrodes exhibited a very stable
response during hours of continuous flow injection.
The current response of the modified electrodes in the
presence of some interferent compounds was alsoexamined. The concentration of ascorbic acid, uric
acid and dopamine were 0.05, 0.32 and 0.02 mM,
respectively, which are their individual values in human
serum samples. Dopamine and uric acid exhibited no
response. On the contrary, a large current response was
observed for ascorbic acid. Thus, a preseparation or
preoxidation of ascorbic acid (Shin et al., 2001) was
required for use of poly(thionine) modified electrodes asan electrochemical detector for NADH.
The sensing elements were found to be stable for at
least 1 week and maintained 90% of their activity during
that period while being stored at 4 8C between measure-
ments.
4. Conclusion
In this paper, a poly(thionine) film on the screen-
printed carbon electrode was prepared by the oxidative
electropolymerization of thionine in neutral phosphate
buffer, and the resulting poly(thionine) modified elec-
trodes exhibited a good electrocatalytic activity for
NADH oxidation and good stability. In a FIA, the
limit of detection of NADH was estimated to be of theorder of 3�/10�6 M. The modified screen-printed
carbon electrodes can be base electrodes for dehydro-
genase enzyme biosensor.
5. Supporting information available
XPS characterization of N1s. This material is avail-
able free of charge via the Internet at http://pubs.ac-
s.org.
Q. Gao et al. / Biosensors and Bioelectronics 19 (2003) 277�/282282
Acknowledgements
Support of this study by the National Natural Science
Foundation of China (No. 20299030) is gratefullyacknowledged.
References
Albery, W.J., Foulds, A.W., Hall, K.J., Hillman, A.R., Egdell, R.G.,
Orchard, A.F., 1979. Thionine coated electrode for photogalvanic
cells. Nature 282, 793�/797.
Alvarez-Icaza, M., Bilitewski, U., 1993. Mass production of biosen-
sors. Anal. Chem. 65, 525A�/533A.
Aviamescu, A., Noguer, T., Aviamescu, M., Marty, J.L., 2002. Screen-
printed biosensors for the control of wine quality based on lactate
and acetaldehyde determination. Anal. Chim. Acta 458, 203�/213.
Bartlett, P.N., Birkin, P.R., Wallace, E.N.K., 1997. Oxidation of b-
dihydronicotinamide adenine dinucleotide at poly(aniline)-coated
electrodes. J. Chem. Soc., Faraday Trans. 93, 1951�/1960.
Bauldreay, J.M., Archer, M.D., 1983. Dye-modified electrodes for
photogalvanic cells. Electrochim. Acta 28, 1515�/1522.
Cai, C.X., Xue, K.H., 1997. Electrocatalysis of NADH oxidation with
electropolymerized films of azure I. J. Electroanal. Chem. 427,
147�/153.
Cai, C.X., Ju, H.X., Chen, H.Y., 1995. Electrocatalysis of poly-
thionine modified microband gold electrode for oxidation of
reduced nicotinamide adenine dinucleotide. Chem. J. Chin. Univ.
16, 368�/372.
Camalli, M., Caruso, F., Mattogno, G., Rivarola, E., 1990. Adducts of
tin(IV) and organotin(IV) derivatives with 2,2?-azopyridine II.
Crystal and molecular structure of SnMe2Br2AZP and further
mossbauer and photoelectronic spectroscopic studies. Inorg. Chim.
Acta 170, 225�/231.
Chen, S.M., Lin, K.C., 2001. The electrocatalytic properties of
polymerized neutral red film modified electrodes. J. Electroanal.
Chem. 511, 101�/114.
Emr, S.A., Yacynych, A.M., 1995. Use of polymer films in ampero-
metric biosensors. Electroanalysis 7, 913�/923.
Gorton, L., 1986. Chemically modified electrodes for the electrocata-
lytic oxidation of nicotinamide coenzymes. J. Chem. Soc. Faraday
Trans. 1 82, 1245�/1258.
Gorton, L., Dominguez, E., 2002. Electrocatalytic NAD(P)H oxida-
tion at mediator modified electrode. Rev. Mol. Biotechnol. 82,
371�/392.
Hutchinson, K.W., Hester, R.E., 1984. Raman spectroscopic studies of
a thionine-modified electrode. J. Chem. Soc. Faraday Trans. I 80,
2053�/2071.
Katakis, I., Dominguez, E., 1997. Catalytic electrooxidation of NADH
for dehydrogenase amperometric biosensors. Mikrochim. Acta
126, 11�/32.
Kessel, R., Schultze, J.W., 1990. Surface analytical and photoelec-
trochemical investigations of conducting polymers. Surf. Interf.
Anal. 16, 401�/406.
Kunimura, S., Ohsaka, T., Oyama, N., 1988. Preparation of thin
polymeric films on electrode surfaces by electropolymerization of
o -aminophenol. Macromolecules 21, 894�/900.
Lobo, M.J., Miranda, A.J., Lopez-Fonseca, J., Tunon, P., 1996.
Electrocatalytic detection of nicotinamide coenzymes by poly(o -
aminophend) and poly(o -phenylenediamine)-modifed carbon paste
electrodes. Anal. Chim. Acta 325, 33�/42.
Lobo, M.J., Miranda, A.J., Tunon, P., 1997. Amperometric biosensors
based on NAD(P)-dependent dehydrogenase enzymes. Electroana-
lysis 9, 191�/202.
Ohsaka, T., Ohnuki, Y., Oyama, N., Katagiri, G., Kamisako, K.,
1984. IR absorption spectroscopic identification of electroactive
and electroinactive polyaniline films prepared by the electro-
chemical polymerization of aniline. J. Electroanal. Chem. 161,
399�/405.
Ohsaka, T., Ohba, M., Sato, M., Oyama, N., Tanaka, S., Nakamura,
S., 1991. Formation of a noval electroactive film by electropoly-
merization of 5-amion-1-naphthol. J. Electroanal. Chem. 300, 51�/
66.
Ohsaka, T., Tanaka, K., Tokuda, K., 1993. Electrocatalysis of
poly(thionine)-modified electrodes for oxidation of reduced nico-
tinamida adenine dinucleotide. J. Chem. Soc. Chem. Commun.,
222�/224.
Pariente, F., Lorenzo, E., Tobalina, F., Abruna, H.D., 1995. Aldehyde
biosensor based on the determination of NADH enzymatically
generated by aldehyde dehydrogenase. Anal. Chem. 67, 3936�/
3944.
Ruan, C., Yang, F., Lei, C.H., Deng, J.Q., 1998. Thionine covalently
tethered to multilayer horseradish peroxidase in a self-assembled
monolayer as an electron-transfer mediator. Anal. Chem. 70,
1721�/1725.
Schlereth, D.D., Karyakin, A.A., 1995. Electropolymerization of
phenothiazine, phenoxazine and phenazine derivatives: character-
ization of the polymers by UV�/visible difference spectroelectro-
chemistry and fourier transform IR spectroscopy. J. Electroanal.
Chem. 395, 221�/232.
Shin, J.H., Choi, Y.S., Lee, H.J., Choi, S.H., Ha, J.H., Yoon, I.J.,
Nam, H., Cha, G.S., 2001. A planar amperometric creatinine
biosensor employing an insoluble oxidizing agent for removing
redox-active interferences. Anal. Chem. 73, 5965�/5971.
Silber, A., Hampp, N., Schuhmann, W., 1996. Poly(methylene blue)-
modified thick-film gold electrodes for the electrocatalytic oxida-
tion of NADH and their application in glucose biosensor. Biosens.
Bioelectron. 11, 215�/223.
Sprules, S.D., Hart, J.P., Wring, S.A., Pittson, R.A., 1995. Reagent-
less, disposable biosensor for lactic acid on a screen-printed carbon
electrode containing Meldola’s Blue and coated with lactate
dehydrogenase, NAD� and cellulose acetate. Anal. Chim. Acta
304, 17�/24.
Torstensson, A., Gorton, L., 1981. Catalytic oxidation of NADH by
surface-modified graphite electrodes. J. Electroanal. Chem. 130,
199�/207.
Wang, J., Pamidi, P.V.A., Jiang, M., 1998. Low-potential stable
detection of b-NADH at sol�/gel derived carbon composite
electrodes. Anal. Chim. Acta 360, 171�/178.
Xiao, Y., Ju, H.X., Chen, H.Y., 1999. A reagentless hydrogen peroxide
sensor based on incorporation of horseradish peroxidase in
poly(thionine) film on a monolayer modified electrode. Anal.
Chim. Acta 391, 299�/306.
Yang, R., Ruan, C.M., Dai, W.L., Deng, J.Q., Kong, J.L., 1999.
Electropolymerization of thionine in neutral aqueous media and
H2O2 biosensor based on poly(thionine). Electrochim. Acta 44,
1585�/1596.
Yoon, H.C., Kim, H.S., 1996. Electrochemical characteristics of a
carbon-based thick-film L-lactate biosensor using L-lactate dehy-
drogenase. Anal. Chim. Acta 336, 57�/65.
Zhang, C.X., Gao, Q., Aizawa, M., 2001. Flow injection analytical
system for glucose with screen-printed enzyme biosensor incorpor-
ating Os-complex mediator. Anal. Chim. Acta 426, 33�/41.
Zhou, D.M., Sun, J.J., Chen, H.Y., Fang, H.Q., 1998. Electrochemcial
polymerization of toluidine blue and its application for the
amperometric determination of b-D-glucose. Electrochim. Acta
43, 1803�/1809.