pressure measurement and calibration instrumentssika-usa.com/sturteverant/wp-content/uploads/... ·...

16
Quality by tradition Pressure Measurement and Calibration Instruments

Upload: others

Post on 03-Jun-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Quality by tradition

Pressure Measurementand Calibration Instruments

Page 2: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

TemperatureStandard Signals

Pressure

% r.F.

°C°F

bar

mbarpsi

mA

Pressure Measurement and Calibration Instruments

Our Pressure instruments and calibration instruments are designed and built with long-term reliability, innovative calibration

technology, and superior quality in mind. A comprehensive range of pressure products and services, with standardized

electrical signals are offered. Our products are engineered and manufactured to DKD standards (Germany's NIST equivalent).

Our pressure calibrators, simulators, controllers and indicators, transmitters, and sensors help our customers to ensure the

consistent long-term quality of their products, and improve the effiency of their production processes. Our strength is our fl

in developing customized solutions, based on our welth of technical and application experience.

Page 3: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

3 in 1

Full version / P1000.1

Hydraulic or pneumatic

application areas in which wetting of the test sample is not allowed or the use of aggressive or ionizing substances must be avoided, air is the ideal test medium.

many cases can only be covered by several pumps from other suppliers.

Manual pressure generation in the medium pressure range up to 60 bar (870psi) using a handle is unique – no other pump can do this.

Integrated negative pressure capability enables operating (-14psi)

enables switching from positive pressure to negative pressure with no need for special tools.

pitch is used for precise pressure adjustment in the low mbar range, enabling accurate settings in the low pressure region.

The easily operated hydraulic test pumps and pressure

pressure range. They have a built-in reservoir for the hydraulic , (5075, 10135, or

14500 psi) can be generated, depending on the model.

A single pump for a variety of requirements

Standard adapter kit

G¼ G½ M12 x 1.5 M20 x 1.5

Adapter kits

OEM

Depending on the model a matching pressure hose is part of

BSP, NPT, and metric adapters for all commonly used

replacement full-setequipment is held in a carrying case with a foam-rubber insert.

Calibration Certificates

OEM version and full version

Depending on the model a matching pressure hose is part of

Calibration Certificates are supplied with each calibrator, anddocument the traceability of your instruments to the nationalmeasurement system. SIKA calibrators come standard with an Exclusive 3 year Calibration Certificate, or 350 working hours.

3

Page 4: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Pressure is one of the most commonly measured unit ofmeasurement, which is why precise and reliable pressure measurement is especially important.

However, the characteristics of even the best sensor or transducer can be altered by a wide variety of factors. This drift cannot be prevented, and it leads to incorrect readings.

Calibration allows these deviations to be measured,stored and measuring devices or activities should be calibrated before being used.

Good reasons for proper and reliable calibration:

Maintaining consistently high product quality

Process optimization

Increasing productivity

Employee and customer safety

Environmental requirements / ecological aspects

z

costfor performing the necessary test and calibration tasks quickly.

Requirements for pressure sources

The essential requirements for manual pressure generation are:

Easy and quick connection to test samples

Simple and easy pressure generation

Maintenance free operation

These aspects have been taken into account and implemented in the design of our test pumps and pressure generators.

Test pumps or pressure generators

stationary use. Unlike complicated and sensitive plate scales, there is no need to take gravity into account or perform complicated alignment, since measurements are based on

is one-hand operation, which makes repetitive test runs more ergonomic.

Manual pressure pumps

Mobile and simple

Test sample

Digital pressure gauge(Reference)

Easy pressure generation

Handling

Pressure sources

4

Page 5: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Practical

The test pumps and pressure generators are designed to enable the direct connection of all pressure systems to be tested using adapters. The test sample is easily connected using the rugged industrial hose with integrated quick coupling and supplied

using a positioning adapter.

The required test pressure is initially generated using the handles

result, the pressure on both instruments is the same.

The pressure relief valve allows continuous pressure reduction and ensures accurate and easy testing, even with decreasing pressure.

In the simplest case, the pressure is indicated by an analogue

held instrument can also be used. The accuracy or adjustment of the pressure measuring device being tested can be checked by comparing the indicated reference value with the measured value for the device under test.

Mobile and independent

use. The lightweight and compact design make our productseasy to transport directly to the measurement site. The instruments can be used immediately and do not require anadditional power supply. There is no need to take along nitrogenbottles or connect the equipment to a compressed air system.Wear-free manual pressure generation is simple and easy, regardless ofambient temperature and orientation.

Possible areas of application

anywhere, including on site, in workshops, test and measurement rooms, as well as laboratories. They cover a broad spectrum of industries with diverse applications.

Manufacturing and production

Maintenance and service

Quality assurance and test equipment monitoring

Repair

testing, adjusting and calibrating pressure sensors, pressure gauges, pressure switches, safety valves and all types of pressure devices. They are optimized in their function and use

Practical and independent

5

Practice orientated

Reference connection

Fine adjustment

Test connection

Force applied by pressing the handles together

Discharge

Mobile use

Page 6: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Type P 4

Pneumatic pressure pumps

Type P 4

OEM version

Pressure medium

Dimensions

Weight (2.16 Lbs.)

Pressure ranges

Negative pressure -0.3 bar (-4.35 psi) [depending on test sample / reference]

Positive pressure 4 bar (58 psi)

Connections

Reference

Test sample G¼ with quick coupling and pressure hose (1m)

Full version

Adapter kit Chrome-plated brass

Gasket kit

Dimensions

Weight (9.26 Lbs.)

6

Page 7: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Type P 40.2 and P 60

Type P 40.2 P 60

OEM version

Pressure medium

Dimensions

Weight (2.43 Lbs.)

Pressure ranges

Negative pressure -0.95 bar -0.95 bar

Positive pressure 40 bar 60 bar

Connections

Reference G¼

Test sample G¼ with quick coupling and pressure hose (1m)

Full version

Adapter kit Chrome-plated brass

Gasket kit

Dimensions

Weight (9.26) Lbs.

7

Page 8: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Type P 350.1

Hydraulic pressure pumps

Type P 350.1

OEM version

Pressure medium

Dimensions Approx. 240 x 170 x 50 mm

Weight Approx. 1.1 kg (2.43 Lbs.)

Pressure ranges

With destilled water 0...350 bar (5,076 psi)

0...350 bar (5,076 psi)

Connections

Reference G½

Test sample G¼ with quick coupling and pressure hose (1m)

Full version

Adapter kit Chrome-plated brass

Gasket kit

Dimensions Approx. 450 x 370 x 110 mm

Weight Approx. 3.5 kg (7.72 Lbs.)

8

Page 9: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Type P 700.2 - 1000

Type P 700.2 - 1000

OEM version

Pressure medium

Dimensions

Weight (4.19 Lbs.)

Pressure ranges

With destilled water bar (10,150 psi) (14,500 psi)

(10,150 psi) (14,500 psi)

Connections

Reference G¼

Test sample G¼ with quick coupling and high pressure hose (1 m), 1000 bar

Full version

Adapter kit Stainless steel

Gasket kit

Dimensions

Weight (11 Lbs.)

9

Page 10: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Hydraulic comparison test pump

Type P 1400.GSIKA test pumps and pressure generators can be usedeverywhere, including workshops, calibration, test andmeasurement rooms as well as laboratories.

They cover a broad spectrum of industries with diverseapplications. • Assembly and commissioning • Manufacturing and production • Maintenance and service • Quality assurance and test equipment monitoring • Repair and Calibration

The latest generation of comparison test pumps provideprecision throughout the entire pressure range. All wettedparts are made of stainless steel, ensuring trouble-free,wear-free operation.The sophisticated design of these comparison test pumpsfacilitates the easy installation of test samples, even withlarge nominal diameters, without sacrificing the functionality.The built-in adapters on the pressure connections also allowfor free alignment of the reference pressure gauge and“test sample” instrument under test.

Type P 700.G2 P 700. GW P 1000.GW P1400.G

Pressure ranges

With distilled water 0...700 bar 0...10 000 psi 0...1000 bar 0...15 000 psi

0...700 bar 0...10 000 psi 0...1400 bar 0...12 000 psi

OEM version

Pressure medium Distilled water Distilled water

Dimensions Approx. 340 x 225 x 130 mm / Approx. 13.39 x 8.86 x 5.12 in.

Weight Approx. 9.9 kg / Approx. 21.8 lbs.

Connections

Reference G⅜ A left, G¼, G½

Test sample G⅜ A left, G¼, G½, G⅛

Full version

Adapter kit Stainless steel

Gasket kit ® Seals and O-rings

Dimensions Approx. 450 x 370 x 150 mm / Approx. 17.71 x 14.57 x 5.91 in.

Weight Approx. 12.6 kg / Approx. 27.7 lbs.

Depending on the model distilled water and hydraulic oil can be used as test medium. Distilled water is an excellentcalibration medium, since it can be removed without a problem and without oil residue. For test pressures above1000 bar, hydraulic oil is recommended.

The filling of the pressure body and test pressure generation are easily done via a rotatable spindle. Large volumescan be easily loaded via the pressure medium reservoir which is built-in on the test pump. Pressure fluctuations causedby air and gas components in the filling liquid, compensation of thermodynamic effects and precise adjustment of therequired test pressure are done via the fine adjustment.

10

Page 11: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

11

Digital pressure gauges are particularly suitable for both stationary and mobile measurement and display of pressure. They can be used as reference pressure gauges to simplify the checking, adjustment and calibration of other pressure measurement devices directly on site.High accuracy in signal acquisition is achieved by using high-performance measuring cells with electronic linearization of the characteristic curve. Suitable instruments are available for a wide variety of measurement tasks.Ease of use is assured by innovative design and advanced

selected conveniently at the press of a button. Excellent protection against dust and moisture is provided by a membrane keypad or rubber buttons. Integrated supplementary functions make our digital pressure gauges true all-rounders.

Advantages at a glance

Exact and reliable measurement

High operational readiness

Easy and clear readout

for

Easy to assemble and use

Supplementary functions for extra value

EX-version availabe on request

Negative / Positive / and Differential Pressure

Measuring ranges from -1 bar (-14.5 psi) upto 2500 bar (36250 psi), with high overpressure protection are available. Very small differential pressures in the millibar range can also be measured. Differential pressure measuring cells or two independent measuring inputs are used for this purpose.

Resolution / Accuracy

It is often necessary to use several mechanical pressure gauges when measurements must be made over a wide pressure

high resolution and precision can handle this task with just one instrument.

spectrum of requirements. This precision is often found only in

gauges are designed for use in harsh industrial environments.

Tare / Zero

offset adjustment easy and eliminates the need for tedious mechanical adjustment. Single-point adjustment allows the linear characteristic curve to be shifted in positive or negative direction over the entire measuring range.

Linearization

Multi-point adjustment can be performed if it is necessary to adjust the indicated values at different test points. Two-point

the measuring cell curve. Some digital pressure gauges allow up to six offset values to be programmed in order to shift the characteristic curve to meet the most stringent customer expectations.

Battery operation / Auto-Off

Power is supplied by long-life batteries (ordinary or

function switches off the power to the instrument after prolonged inactivity. The electronics are designed for extremely low power

than 1000 hours.

Digital pressure gauges

Exact and reliable

Type E2 / D2 and P

Page 12: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Data memoryThe logger function for local data storage can be used to record pressure curves automatically and perform leak tests. The integrated data memory in digital pressure gauges allows a variety of data sets to be recorded directly. The time interval between samples is programmable and the maximum

displayed on a PC. Data import at the press of a button is also possible. In this case the data is shown directly on the display. In this process the values are automatically annotated with the date and time of day using an integrated real-time clock.

Supplementary functions

Analog output

a control console or in a control room as well as the connection of external recorders and indicating instruments.

Relay output / Alarm signalling

Digital pressure gauges allow limit contacts to be closed even at low pressures. There is no need for high actuation forces for magnetic spring or inductive contacts, which makes it easier to signal critical equipment conditions and perform supplementary

pressure exceeds the range of the programmed minimum and maximum pressure levels.Temperature display

Temperature measurement is often required in addition to pressure measurement. For this purpose, a temperature sensor is integrated in the measuring cell to detect the temperature of the medium. The process temperature can be displayed at the press of a button. This allows two quantities to be measured at a single measuring point, which saves time and costs.

Explosion protection

Explosion-proof versions are also available for use in potentially such as plants

and drilling platforms.

Example applications

Continuous or temporary checking of a wide variety of system pressures

and elimination of problem areas and avoiding structural damage

or air conditioning systems

Recording pressure drops for the determination of leakage rates in leak tests

Measurement of barometric air pressure for the determination of weather conditions

Reference pressure gauge for calibration tasks

Technical data

Accuracy (full scale)

Pressure range (bar) Resolution

-1...3 bar-1...40 bar-1...60 bar0...400 bar

0...1000 bar

1 mbar10 mbar10 mbar100 mbar100 mbar100 mbar

10 mbar10 mbar100 mbar100 mbar100 mbar

Pressure range (psi) Resolution

-14.5...43.51 psi-14.5...580.15 psi-14.5...870.23 psi0...5,801 psi

0...14,503 psi

0.0150.1450.1451.4501.4501.450

0.1450.1451.4501.4501.450

0...10,152 psi

Type E2, D2

Type R

Technical data

Accuracy (full scale)

Pressure range Resolution

-1...1 bar (-14.5...14.50 psi)-1...2.5 bar (-14.5...36.26 psi)-1...5 bar (-14.5...72.52 psi)-1...10 bar (-14.5...145 psi)-1...20 bar (-14.5...290 psi)-1...40 bar (-14.5...580 psi)-1...60 bar (-14.5...870 psi)0...100 bar (0...1450 psi)0...250 bar (0...3,626 psi)0...350 bar (0...5,076 psi)0...500 bar (0...7,252 psi)

(0...10,152 psi)0...1000 bar (0...14,503 psi)0...1500 bar (0...21,755 psi)0...2000 bar (0...29,007 psi

1 mbar (0.015 psi)1 mbar (0.015 psi)1 mbar (0.015 psi)10 mbar (0.145 psi)10 mbar (0.145 psi)10 mbar (0.145 psi)10 mbar (0.145 psi)100 mbar (1.450 psi)100 mbar (1.450 psi)100 mbar (1.450 psi)100 mbar (1.450 psi)100 mbar (1.450 psi)1 bar (14.50 psi)1 bar (14.50 psi)1 bar (14.50 psi)

12

Page 13: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

13

Functions

Type E2 / D2 R

Adjustment options

LinearisationTare / Zero

6 points

Selectable units

PressureTemperature

bar, mbar, kPa, MPa, PSI, kg/cm² bar°C

Features

Measuring inputsPC connection Analogue outputRelay outputBuilt-in versionExplosion protection

1 x direct 1 x directRS232 (optional)

Data memory

Number of memorysRecording intervalRecording durationData sets

60.000 values (auto)1 sec...10 h1 min...1000 hPressure / Temperature

Display / Representation

Multi-functional LCDBargraphIllumination

Min / max value

4 ½ digit 4 digit

Measuring rate

StandardPeak / Fast

10 msec.10 msec.

100 msec.

Process connection

Connection optionsMaterialMedium temperatureFor aggressive media

G¼1.4404

(-4...176 °F)

G½1.4542

(14...158 °F)

Housing

Degree of protectionDimensionMaterialOperating temperatureWeight

Zinc casting0...50 °C (32...122 °F)540 g (1.2 Lbs.)

IP65 (front) / IP40

(14...158 °F)450 g (1.0 Lbs)

Power

Auto-off functionBattery typeExt. powerBattery operation 1500 h

Page 14: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

Type LType P

Technical data

Accuracy (bar)

Pressure range (bar) Resolution

-1...1 bar-1...2.5 bar-1...5 bar-1...10 bar-1...20 bar-1...40 bar-1...60 bar0...100 bar0...250 bar0...350 bar0...500 bar

0...1000 bar0...1500 bar0...2000 bar0...2500 bar

1 mbar1 mbar1 mbar10 mbar10 mbar10 mbar10 mbar100 mbar100 mbar100 mbar100 mbar100 mbar1 bar1 bar1 bar1 bar

1 mbar1 mbar1 mbar10 mbar10 mbar10 mbar10 mbar100 mbar100 mbar100 mbar100 mbar100 mbar1 bar1 bar1 bar1 bar

0.1 mbar0.5 mbar0.5 mbar1 mbar2 mbar5 mbar5 mbar10 mbar20 mbar50 mbar50 mbar50 mbar100 mbar200 mbar500 mbar

Pressure range (psi) Resolution

-14.50...14.50 psi-14.50...36.26 psi-14.50...72.52 psi-14.50...145 psi-14.50...290 psi-14.50...580 psi-14.50...870 psi0...1,450 psi0... 3,626 psi0... 5,076 psi0... 7,251 psi

0... 14,503 psi0... 21,755 psi0... 29,007 psi0... 36,259 psi

0.015 psi0.015 psi0.015 psi0.145 psi0.145 psi0.145 psi0.145 psi1.1450 psi1.1450 psi1.1450 psi1.1450 psi1.1450 psi14.50 psi14.50 psi14.50 psi14.50 psi

0.015 psi0.015 psi0.015 psi0.145 psi0.145 psi0.145 psi0.145 psi1.1450 psi1.1450 psi1.1450 psi1.1450 psi1.1450 psi14.50 psi14.50 psi14.50 psi14.50 psi

0.001 psi0.007 psi0.007 psi0.015 psi0.029 psi0.073 psi0.073 psi0.145 psi0.290 psi0.725 psi0.725 psi0.725 psi1.450 psi2.900 psi7.252 psi

0... 10,152 psi

Technical data

Accuracy (full scale)

Pressure range (bar)

-1...2 bar-1...10 bar-1...20 bar0...200 bar0...400 bar

0...1000 bar

0.1 mbar1 mbar1 mbar10 mbar50 mbar100 mbar100 mbar

1 mbar10 mbar50 mbar100 mbar100 mbar

1 mbar10 mbar50 mbar100 mbar100 mbar

Pressure range (psi)

-14.50...29.01 psi-14.50...145 psi-14.50...290 psi0...2,900 psi0...5,801 psi

0...14,503 psi

0.001 psi0.015 psi0.015 psi0.145 psi0.725 psi1.450 psi1.450 psi

0.015 psi0.145 psi0.725 psi1.450 psi1.450 psi

0.725 psi0.725 psi0.725 psi1.450 psi1.450 psi

Resolution

Resolution

0...10,152 psi

14

Page 15: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

15

Functions

Type P L

Adjustment options

LinearisationTare / Zero

6 points

Selectable units

Pressure bar, mbar, kPa, MPa, PSI bar, mbar, hPa, kPa, MPa, PSI, mmHg, inHg,

Features

Measuring inputsPC connectionAnalogue outputRelay outputBuilt-in versionExplosion protection (optional)

1 x directRS232 (optional)

(optional)

1 x direct

Ex II 2G Ex ia II C T6

Data memory

Number of memoriesRecording intervalRecording durationData sets

Display / Representation

Multi-functional LCDBargraphIllumination

Min/max value

5 digit

Measuring rate

StandardPeak / Fast

100 msec. 500 msec.

Process connection

Connection optionsMaterialMedium temperatureFor aggressive media

G½1.45420...50 °C

G¼1.44350...50 °C

Housing

Degree of protectionDimensionMaterialOperating temperatureWeight

IP65 (front) / IP40

0...50 °C (32...122 °F)900 g (2.0 Lbs.)

IP65 (front) / IP54

0...50 °C (32...122 °F)210 g (0.46 Lbs)

Power

Auto-off functionBattery typeExt. powerBattery operation

24 VDC1x 3 VCR

2000 h

Page 16: Pressure Measurement and Calibration Instrumentssika-usa.com/sturteverant/wp-content/uploads/... · Digital pressure gauges are particularly suitable for both stationary and mobile

2016 Edition

PRESSCAT0216

Mechanical Measuring Instruments

Flow Measuring Instruments

Electronic Measuring and Calibration Instruments

SIKA USA Inc.1500 S. Sylvania Ave. Suite 111Sturtevant, WI 53177

Phone (414) 886-2695 Fax (414) 898-0101

Email: [email protected]