quality control lecture cpath master 2014 ain shams
DESCRIPTION
Basics of quality management or assurance program detailing values of internal quality control material analysis and interpretation and external quality control or proficiency testing programs in medical laboratoriesTRANSCRIPT
![Page 1: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/1.jpg)
![Page 2: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/2.jpg)
![Page 3: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/3.jpg)
All management systems are focused on getting the job done.
More efficient and effective management guarantees the job to be done in the proper way.
Larger profit Better serviceVs
![Page 4: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/4.jpg)
STAGE ACTIVITIES PERFORMED1. TOTAL QUALITY MANAGEMENT Management approach centered around
“Customer Satisfaction”2. QUALITY MANAGEMENT All of the below plus the economic
aspects of “Cost of Quality”3. QUALITY SYSTEM “Comprehensive and Coordinated”
efforts to meet quality objectives4. QUALITY ASSURANCE Systematic activities to provide
“Confidence” that the organizationmeets requirements for quality
5. QUALITY CONTROL Operational techniques applied to“Specific Tasks” for quality andregulatory compliance.
![Page 5: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/5.jpg)
![Page 6: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/6.jpg)
Quality : degree to which a set of inherent characteristics fulfills requirements
Quality control: part of quality management focused on fulfilling quality requirements
Quality assurance: part of quality management focused on providing confidence that quality requirements will be fulfilled
Quality management: coordinated activities to direct and control an organization with regard to quality
![Page 7: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/7.jpg)
1942-521942-52 US Military develop requirements for contractors for shell, aircraft, missile US Military develop requirements for contractors for shell, aircraft, missile suppliers suppliers (Quality without 100% inspection)(Quality without 100% inspection)
19471947 International Organization for Standardization (ISO) International Organization for Standardization (ISO) created to adopt industrial created to adopt industrial standardsstandards
19591959 US Department of Defense established US Department of Defense established MIL-Q-9858MIL-Q-9858 quality management quality management
19631963 MIL-Q-9858 is internationalized as anMIL-Q-9858 is internationalized as an ABCAABCA standardstandard
19681968 NATO adopts MIL-Q-9858A as Allied Quality Assurance Publication 1 (NATO adopts MIL-Q-9858A as Allied Quality Assurance Publication 1 (AQAP-1AQAP-1))
19791979 British Standards Institute (BSI) developed AQAP-1 for civilian use British Standards Institute (BSI) developed AQAP-1 for civilian use BS 5750BS 5750
![Page 8: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/8.jpg)
Calibration: the process of testing and adjustment of an instrument, kit, or test system to provide a known relationship between the measurement response and the value of the substance measured by the test procedure
Effectiveness: the extent to which planned activities are realized and planned results achieved
Efficiency: the relationship between the results achieved and the resources used
![Page 9: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/9.jpg)
Non-QMS processes – processes that are not contained in either the QSEs or in the laboratory’s path of workflow, such as those within the finance, sales, or marketing functions.
Path of workflow: the sequential processes in a laboratory’s activities that transform a request for examination into the laboratory information that is captured in the report of results.
Plan: written account of intended future course of actions aimed at achieving specific goal(s) or objective(s) within a specific timeframe and explains in detail what needs to be done, when, how, and by whom.
![Page 10: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/10.jpg)
![Page 11: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/11.jpg)
OrganizationCustomer FocusFacilities and SafetyPersonnelPurchasing and InventoryEquipmentProcess Management
Pre-examinationExaminationPost-examination
Documents and RecordsInformation ManagementNonconforming Event ManagementAssessmentsContinual Improvement
Pree-xaminationExaminationPost-examinationInformation
Management
Quality System Essentials - those elements which must be applied to all operations in the service’s path of workflow
Path of Workflow - as it relates to the processes involved in the pre-, during and post- phases of delivery of services with the laboratory as an example
![Page 12: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/12.jpg)
![Page 13: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/13.jpg)
![Page 14: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/14.jpg)
![Page 15: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/15.jpg)
![Page 16: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/16.jpg)
! Internal Quality Control (IQC) Procedures
! External Quality Assessment (EQA)
! Quality Management
![Page 17: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/17.jpg)
Internal Quality Control
![Page 18: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/18.jpg)
⇒ Done during daily routine work⇒ Provides an immediate control⇒ Errors are corrected immediately⇒ Routinely collect and analyze data
from every test run or procedure
Why do we need Internal Quality Control?
! Ensure that test results are reliable (trustable)
! Ensure that test results are reproducible (Closeness of the results of assays of the same test carried under changed conditions)
! Control quality of daily routine work
![Page 19: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/19.jpg)
Monitoring quality of laboratory testing, accuracy and precision of laboratory
results
![Page 20: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/20.jpg)
Precise and Inaccurate
Precise and Accurate
![Page 21: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/21.jpg)
Imprecise and Inaccurate
Imprecise and ?? Accurate
![Page 22: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/22.jpg)
It ensures continual check that the laboratory’s work does not fluctuate and that reports are validated before they are released.
It is based on monitoring the procedures which are actually used for the tests in the laboratory. It includes:
· Control charts with tests on control materials· Duplicate tests on all specimens or on a proportion
of the specimens· Delta check, comparing current test results with
previous results· Consistency of mean values of patient data
![Page 23: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/23.jpg)
Use 2-3 levels of controls covering the medical decision points
At least 20 control values over a period of 20-30 days for each level of control
Perform statistical analysis Develop Levey-Jenning chart Run -regularly- with patient samples Monitor control values on chart using
Westgard rules Take immediate corrective action, if
needed Record actions taken (if any)
![Page 24: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/24.jpg)
QC material must be available in large quantities stored in small aliquots (Ideally should last for at least 1 year)
Always consider as Biohazardous Requires very accurate reconstitution if
needed Always deal with and store as
recommended by manufacturer
![Page 25: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/25.jpg)
Need data set of at least 20 points obtained over 20-30 days better by different operators in different times of day
Calculate mean, standard deviation, Coefficient of variation and determine target ranges
Develop Levey-Jenning charts Plot –on the chart- control values each
run/ day Make decisions regarding acceptability
of run using Westgard rules Monitor over time and at defined
intervals
![Page 26: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/26.jpg)
1. 192 mg/dL2. 194 mg/dL3. 196 mg/dL4. 196 mg/dL5. 160 mg/dL6. 196 mg/dL
7. 200 mg/dL8. 200 mg/dL9. 202 mg/dL10. 255 mg/dL11. 204 mg/dL12. 208 mg/dL13. 212 mg/dL
![Page 27: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/27.jpg)
80
85
90
95
100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Day
Mean
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
![Page 28: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/28.jpg)
80
85
90
95
100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Day
Mean
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
![Page 29: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/29.jpg)
![Page 30: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/30.jpg)
Ideally should have control values clustered around the mean (+/-2 SD) with little variation in the upward or downward direction
Imprecision = large amount of scatter around the mean. Usually caused by errors in technique
Inaccuracy = may see as a trend or a shift, usually caused by change in the testing process
Random error = no pattern. Usually poor technique, malfunctioning equipment
![Page 31: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/31.jpg)
Random Error (RE): Imprecision
Causes:1. Pipetting error2. Temperature error3. Mixing defect4. Machine need
troubleshooting
Systematic Error (SE): Inaccuracy
Causes:1. Deterioration of
control material2. Deterioration of
calibrator3. Deterioration of
reagents
![Page 32: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/32.jpg)
Allows determination of whether an analytical run/s (in control) or (out of control)
![Page 33: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/33.jpg)
![Page 34: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/34.jpg)
![Page 35: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/35.jpg)
![Page 36: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/36.jpg)
![Page 37: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/37.jpg)
![Page 38: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/38.jpg)
![Page 39: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/39.jpg)
![Page 40: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/40.jpg)
![Page 41: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/41.jpg)
![Page 42: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/42.jpg)
![Page 43: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/43.jpg)
![Page 44: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/44.jpg)
![Page 45: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/45.jpg)
![Page 46: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/46.jpg)
October 1, 2008
![Page 47: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/47.jpg)
October 1, 2008
![Page 48: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/48.jpg)
![Page 49: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/49.jpg)
When you're running 2 or 4 control levels, use the rules: 13s/22s/R4s/41s/10x
When you're running 3 control levels, use a set that works for multiples of threes:13s/2of32s/R4s/ 31s/12x
![Page 50: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/50.jpg)
Rejection rule = Out of control=1. Stop testing2. Identify and correct problem3. Repeat testing on pt samples and
control4. Don’t report pt results until problem
is solved and controls indicate proper performance
![Page 51: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/51.jpg)
Change one variable at a time to troubleshoot an out of control test.
The variables are usually common to all instruments or procedure systems and include the following:
1. Try a fresh vial of QC 2. Try a new reagent (same lot) 3. Try a new consumable (water, part, solution, tubing)
if applicable. 4. Try a new reagent (new lot) 5. Recalibrate (current calibrator) 6. Recalibrate (new lot calibrator) If a system cannot be brought into control,
notify senior staff. Do not process patient samples for the test affected.
![Page 52: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/52.jpg)
When changing to a new lot number of control material, ideally there should be an overlap period while the new material is being analyzed to establish the new control limits.
In cases where the overlap period is not sufficient, it is possible to establish the mean value for the new control material in a short time, over say a five-day period, or to start with the manufacturer’s values.
Then apply the previous estimate of SD to establish the control limits.
These control limits should be temporary, until sufficient data is collected to provide good estimates of both the mean and SD of the new material.
![Page 53: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/53.jpg)
New lots of a quality control material should be analyzed for each analyte in parallel with the lot of control material in current use.
Ideally, a minimum of at least 20 measurements should be made on separate days when the measurement system is known to be stable, based on QC results from existing lots.
If the desired 20 data points from 20 days are not available, provisional values may have to be established from data collected over fewer than 20 days.
Possible approaches include making no more than four control measurements per day for five different days
![Page 54: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/54.jpg)
Retained patients samples◦ Original samples must be assayed in the lab under accepted
control run conditions◦ Stored aliquited immediately after finishing 1st analysis◦ Storage conditions must meet the requirements for analytes to
be measured◦ Avoid repeated warming/ thawing◦ Stability of the measurand must be carefully respected◦ Better to be alternating with QC material◦ Result of the 1st assay is used as the base for comparison and
calculation (as if gold value)◦ Allowable analytical imprecision is used for judgment of
acceptance or rejection◦ Data calculated from biological variations/ CLIA limits are used
for the issue
![Page 55: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/55.jpg)
Result comparison with an in-control method/ instrument/ procedure for the measurand◦ The in-control instrument/ method is considered
the reference one◦ Not more than one hour time gap between the
parallel assays◦ Principles of assays must be considered if
different◦ Inaccuracy limits is used for comparison◦ Data calculated from biological variations/ CLIA
limits are used for the issue
![Page 56: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/56.jpg)
Individual Patient Results:◦Clinical Correlations◦Correlation with other laboratory tests◦Intralaboratory duplicates◦Delta check with previous test results◦Limit Check
Multiple Patients:◦Test distribution statistics◦Monitoring patients means
![Page 57: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/57.jpg)
QUESTIONS
![Page 58: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/58.jpg)
External Quality Control
![Page 59: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/59.jpg)
Inter-laboratory comparisons and other performance evaluations that may extend throughout all phases of the testing cycle, including interpretation of results; determination of individual and collective laboratory performance characteristics of examination procedures by means of interlaboratory comparison
![Page 60: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/60.jpg)
A program in which multiple samples are periodically sent to members of a group of laboratories for analysis and/or identification, in which each laboratory’s results are compared with those of other laboratories in the group and/or with an assigned value, and reported to the participating laboratory and others
![Page 61: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/61.jpg)
Introduced into laboratory medicine more than 60 years ago to address that results for aliquots of the same sample were different when measured by different laboratories.
PT/EQA programs are now an essential component of a laboratory’s quality management system.
PT/EQA is a component of laboratory accreditation requirements
![Page 62: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/62.jpg)
![Page 63: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/63.jpg)
![Page 64: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/64.jpg)
Ideal samples for a PT/EQA program would fulfill a range of criteria: ◦ Stable for the conditions under which they will be
transported and stored◦ Homogeneous across all the aliquots produced◦ Have analyte concentrations that include the expected
clinical range◦ Include appropriate sample types (e.g., urine, whole
blood, serum)◦ Available in sufficient volume◦ Inexpensive enough for cost not to be an impediment◦ Behave in clinical laboratory measurement procedures in◦ the same manner as patient samples◦ Samples from a single donor or pooled samples from
multiple donors can be used
![Page 65: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/65.jpg)
Samples have traceable reference values (when reference values are used)
Behave like patient samples (commutability) The laboratory may also consider cost Similarity of PT samples to patient samples Method compatibility with peer groups Size of peer groups Frequency of challenges Timeliness and usefulness of reports Educational content Customer service
![Page 66: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/66.jpg)
![Page 67: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/67.jpg)
category 1 programs are limited because of:• Technical aspects such as a lack of reference measurement
procedures, absence of certified reference materials, inability to prepare commutable samples;
• Practical considerations such as the difficulty of preparing samples covering the full measuring interval and the complicated logistics of preparation and distribution of fresh or frozen samples;
• Psychological limitations such as lack of awareness of the quality factors important in PT/EQA or unwillingness to adopt these;
• Economic concerns because distributing commutable samples in sufficient quantity and providing target values with reference measurement procedures is expensive
![Page 68: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/68.jpg)
PT samples should be tested in the same manner as patient samples, to the extent possible
Some laboratories may improperly test PT samples differently from patient samples, by repeat testing of PT samples when patient samples are tested only once, or by having a specific analyst test PT samples rather than rotating PT testing among all the personnel who perform patient testing.
There should be no attempt to produce “best” results by replicate analysis or testing immediately following internal QC or recalibration
![Page 69: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/69.jpg)
The core content of the result report should resemble as closely as possible the content of a routine clinical result report
If the usual report is deemed inappropriate for a PT report, it may equally be inappropriate for a clinical report
A copy of all PT reports should be retained within the laboratory in order to verify the information handling by the PT provider.
![Page 70: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/70.jpg)
Sending a set of samples from an organizing body to a group of participating laboratories for measurement of 1 or more analytes present in the samples
Samples are intended to simulate the clinical samples usually measured
Laboratories are not informed of the analyte concentration or activity in a particular sample
Timely schedules for running and reporting results are included
Laboratory perform measurements in the same manner as for patient samples
Results for the samples are returned to the PT/EQA organizer for evaluation of conformance to the expected results
![Page 71: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/71.jpg)
The organizer prepares a report that includes:◦ the results reported by a laboratory◦ the method used for the measurements◦ the target values expected for each analyte◦ evaluation of whether the individual laboratory’s results
met the performance requirements◦ Reports may also include evaluation of the performance
of the various measurement procedures used by the participants
The laboratory evaluates its performance according to the provider report
![Page 72: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/72.jpg)
Limits or quality standards around the target value are established against which performance can be assessed by:◦ Regulatory: wider like US CLIA, German Rili-BAeK◦ Statistical: ± 2-3 SD◦ Clinically-based: on a difference that may affect clinical decisions or
on biological variation
Total error limits including bias, imprecision, and analytical nonspecificity can contribute to the variation in a single result
Have different limits to separately assess bias and imprecision when replicate samples are included
PT/EQA limits are set as a minimum standard to identify results that indicate poor performance. Thus, meeting these standards may not indicate that performance is optimal nor that performance meets all clinical needs
![Page 73: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/73.jpg)
Determines the accuracy by comparing PT/EQA results to those from a reference measurement procedure or from a designated comparison method or to an all-participant (or all-method) mean/ median. This arrangement is now referred to as accuracy-based evaluation
Assess agreement with other measurement procedures and imprecision among all methods as well as within a method group
Peer group evaluation provides valuable information to assess quality, verifying that a laboratory is using a measurement procedure in conformance to the manufacturer’s specifications and to other laboratories using the same technology
![Page 74: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/74.jpg)
![Page 75: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/75.jpg)
![Page 76: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/76.jpg)
![Page 77: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/77.jpg)
![Page 78: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/78.jpg)
![Page 79: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/79.jpg)
![Page 80: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/80.jpg)
![Page 81: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/81.jpg)
![Page 82: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/82.jpg)
![Page 83: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/83.jpg)
![Page 84: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/84.jpg)
PT/EQA result represents 1 point in time and will occasionally be a random error
Repeat the measurement using a stored aliquot of the PT/EQA sample (assuming the measurand was stable on storage) to confirm if the problem has persisted or to conclude that the problem no longer exists and the original unacceptable result was a random event, and therefore no corrective action is indicated. If the repeated result is still unacceptable, the laboratory conducts further investigation to identify the root cause, and then initiates corrective action
![Page 85: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/85.jpg)
• Gather data related to the testing event to include records of calibration, reagent use, QC results, and maintenance procedures;
• Obtain other data on assay performance, e.g., previous PT/EQA results and relevant patient data;
• Identify the root cause of the error;• Take corrective action and preventive action ifindicated;• Monitor the success of the corrective action;• Document the investigation and the corrective
action.
![Page 86: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/86.jpg)
• Was the testing material received in satisfactory condition?
• Was the appropriate sample tested?• Were procedures for sample preparation followed?• Was the appropriate method used for analysis?• Was the method performed according to
documented procedures?• Were appropriate reagents and controls used?• Was equipment operated according to
documented procedures?
![Page 87: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/87.jpg)
• Was equipment appropriately maintained?• Was QC acceptable at the time of testing PT
samples?• Were results interpreted appropriately?• Has this problem occurred previously with PT
samples? Are data consistent with previous PT distributions? Is there a trend leading to failure or is the current set completely unexpected?
• Did repeat testing on the properly stored residual sample produce similar results?
• Were patient results acceptable at the time of PT testing?
![Page 88: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/88.jpg)
1. Clerical error;2. Methodologic problem;3. Equipment problem;4. Technical problem;5. Problem with proficiency testing materials;6. Problem with evaluation of results; and7. No explanation after investigation: An
investigation fails to reveal an explanation for an unacceptable PT result 19 to 24% of the time
![Page 89: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/89.jpg)
![Page 90: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/90.jpg)
![Page 91: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/91.jpg)
![Page 92: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/92.jpg)
![Page 93: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/93.jpg)
Split-Sample With Another Laboratory Internal Split-Sample Procedures Audit-Sample Procedure Analysis of Manufacturer’s Product Calibrator or
Trueness Control Material Analysis of Interlaboratory Quality Control Data Averages of Patient Data
![Page 94: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/94.jpg)
Reference Intervals Reevaluation of Interpreted Results Direct Observation of Technique-Dependent Tests Clinical Correlation Studies Government and University Interlaboratory
Comparison Programs Analysis of Data From Qualitative Alternative
Assessment Procedures
![Page 95: Quality control lecture CPath master 2014 Ain Shams](https://reader036.vdocuments.net/reader036/viewer/2022062512/554b23f2b4c905a2058b460b/html5/thumbnails/95.jpg)
YOUR VALUABLE QUESTIONS