raith150 two ebl writing manual
TRANSCRIPT
-
7/29/2019 Raith150 Two EBL Writing MANUAL
1/5
EBL writing:
1) 1st select a particle2) Go to file and click new positionlist.
While selecting there New positionlist a new window will open like this:
3) From Control bar choose Write Field Manager. Automatically there will be a window like this containingdifferent magnification and field size and select one of your desired and to confirm that your chosen write field
is selected click the set writefield botton().
-
7/29/2019 Raith150 Two EBL Writing MANUAL
2/5
4) Once your mask size is selected opening Write field control you will have another window named ScannedManager. There u shou select 1
stWritefield Alignment Procedures and then Manual and selecting * um WF
Manual ALWF * um marks u need to drag into the New positionlist.
As a whole you can see up to this.
5) Then scanning the dragged item a window will open to align writefield. Do itfor each selected item like say(200X40,200X20,200X5). After dragging one
item do writefield alignment then delete it from position list and drag another
one if you have
Opened window is like this
This ends the basic writefield alignment we do.
-
7/29/2019 Raith150 Two EBL Writing MANUAL
3/5
Exposure:
Now you need to measure beam current.
1) Click stage Control from control bar2) Go to its Position and select Faraday Cup on holder
there. Then stage will move such a way gun will reach
near Faraday cup.
3) Then you should manually place Gun on Faraday cup &zoom so that ur sreen is totally dark (It ensures all
electrons are captured by Faraday Cup and hence proper
measurement).
4) Next measure current going Exposoure option fromcontrol as shown.
(It is good practice to save the position before moving stage to faraday
cup,so that the position where I did write field alignment is not lost.
5) Now you can set exposure parameters. Like beam current windowsimultaneously other windows are created like exposure parameter where
just clicking the calculator exposure parameter calculation will appear.
6) opens the Exposure Parameter Calculation as said earlier just now.7) selects which layers in the pattern that is going to be written (exposed).8) opens a window (called Exposure parameter calculation) with
parameters that is sometimes useful to change, e.g. the beam settling time,
the FBMS beam width, etc.
9) Entre required Step size & Dose and calculate the dwell time and press ok.Note:
1. Step size - the length the beam is moved between each dwell time, when the beam is still, making theexposure. Usually 5 - 15 nm for W.F. size of 60 - 100 m and 100 - 200 nm for W.F. size of 800 - 1500 m
(bonding pads).
2. Dwell time - the time the beam is still at each location.3. Beam current - the e--current reaching the sample. This should already be correct since the earlier
measurement should have put the measured value here.
4. Area dose - the dose required for an area to fully develop that area, i.e. fully remove all resist but not more.Sometimes wrongly referred to as "Sensitivity". It is rather the inverse of sensitivity.
-
7/29/2019 Raith150 Two EBL Writing MANUAL
4/5
These four parameters are interdependent according to this
formula:Writing:
1) Again open New position list.2) Open your GDS database from where you can drop the GDS file that you wanna write into the new positionlist.3) Select properties from Drop down list. Select layers to be written and define boundaries..You can see these.
If you click edit button as shown here you can find
Then go to property.A new windo will be popped up where you can select the working area and position list.
-
7/29/2019 Raith150 Two EBL Writing MANUAL
5/5
Learn more to finsh
Ref:
http://www.nanophys.kth.se/nanophys/facilities/nfl/manual/
http://www.nanophys.kth.se/nanophys/facilities/nfl/manual/http://www.nanophys.kth.se/nanophys/facilities/nfl/manual/http://www.nanophys.kth.se/nanophys/facilities/nfl/manual/