Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2 , Garrett M. Chinn 3 , Richard R. Jahan-Tigh 1,2 1 Department of Dermatology, University of Texas, Houston Medical School, Houston, TX, USA 2 Division of Dermatology, Department of Medicine, University of Texas MD Anderson Cancer Center, Houston, TX, USA 3 Division of General Internal Medicine, Department of Medicine, Harvard Medical School, Massachusetts General Hospital, Boston, Massachusetts

Upload: colin-pierce

Post on 27-Dec-2015

221 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

Research Techniques Made SimpleFlow Cytometry II: Mass and

Imaging Cytometry

Hung Doan1,2, Garrett M. Chinn3, Richard R. Jahan-Tigh1,2

1 Department of Dermatology, University of Texas, Houston Medical School, Houston, TX, USA

2 Division of Dermatology, Department of Medicine, University of Texas MD Anderson Cancer Center, Houston, TX, USA

3 Division of General Internal Medicine, Department of Medicine, Harvard Medical School, Massachusetts General Hospital, Boston, Massachusetts

Page 2: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

1. Single-cell suspensions are prepared from tissue/blood

2. Rare earth metal–labeled antibodies incubated with cells (35 or more antibodies labeling each cell)

Mass Cytometry (CyTOF)

3. Cells are injected into argon gas stream where they are exposed to a plasma torch reaching temperatures of 10,000 K

4. Cells and their metal labels are then vaporized, atomized, and ionized, and the times-of-flight (TOFs) of the ions from the plasma to a mass spectrometer detector are measured, creating a profile for each cell.

Page 3: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

Red outlines present in the two upper panels and the blue outlines of the bottom two panels represent the cells that are CD8+ and specific for VP3 of all the CD8+ cells (represented by black dots).

Figure 2(a)

Cells that are positive for both markers (in this

case, CD103 and integrin-β7) are present inthe upper right quadrant.

Those that are negative for the two markers are present in the bottom left.

This group of cytometry fluorescence plots compares CD8+ T-cells from the intestinal intraepithelial lymphocytes (IEL) that are specific for a rotavirus epitope called VP3 to the same VP3 specific T-cells from the peripheral blood (i.e., they are gated on CD8+).

Newell EW, et al. (2013) Nat Biotechnol 31:623-9

Page 4: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

This graph displays the first two principal components (i.e., PC1 and PC2) or the first two groups of variables (e.g., cell markers) that best predicts the variance in the data.

Specifically, it shows that peripheral blood lymphocytes reactive for the rotavirus VP3 epitope (Blood-Rota-VP3) appear to have a similar cellular surface marker expression profile to intestinal intraepithelial lymphocytes (IEL) reactive to the VP3 epitope which is distinct from rotavirus-reactive lymphocytes in peripheral blood not reactive to VP3 (Blood-Rota-nonVP3). Based on these and other data, the authors propose that VP3 exposure may happen in the intestine and the lymphocytes

circulate from there to the peripheral blood.

Each data point represents a donor’s cell marker expression data from either peripheral blood or IEL specific for the VP3 epitope.

Figure 2(b)

Newell EW, et al. (2013) Nat Biotechnol 31:623-9

Page 5: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

• Each row in this clustergram is the expression data for a single donor. Only the variables that were calculated to be most important for this set of data as derived from the principal component analysis are listed along the X-axis (i.e., CD69, CD103, etc).

Figure 2(c)

• Each square correlates with the cumulative expression data from millions of that specific donor’s T cells that were vaporized after passing through the plasma torch.

• A separate calculation involving the Euclidian distance between values allows similar samples to be clustered together.

Newell EW, et al. (2013) Nat Biotechnol 31:623-9

Page 6: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

Imaging Cytometry (ImageStream)

1. Cells of interest are labeled with fluorescent antibodies and injected into a liquid stream.

2. The flow of the liquid causes the cells and/or particles to line up until they are struck by an excitatory light source causing them to fluoresce.

3. In contrast to classic flow, images of each detected particle are captured using a microscope objective (20x, 40x, or 60x) along with their respective fluorescence data. In classic flow, no image is captured - only the fluorescence and scatter data are present.

Page 7: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

Imaging Cytometry (ImageStream)

1. Cells of interest are labeled with fluorescent antibodies and injected into a liquid stream.

2. The flow of the liquid causes the cells and/or particles to line up until they are struck by an excitatory light source causing them to fluoresce. 3. The most common imaging cytometry platform, ImageStream,

captures image data for each of these events that pass through the detector system using one or more microscope objectives.

Page 8: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

Figure 3• Each row on the left

represents a single event, in this case, either a dermal dendritic cell (DDC) or a Langerhans cell (LC) that had its image captured as it passed through the detector. Each image is multispectral (composed of multiple separate wavelengths of light), which allows only certain channels to be viewed simultaneously.

Columns show cell images in gray from bright-field (traditional light microscopy), and the green spots are

green-fluorescent protein–producing Staphylococcus aureus. Overlay shows the bacteria inside each cell.

Columns with blue images represent early endosomes due to fluorescence of the Alexafluor 405 (AF405), and the reddish-orange images mark lysosomes dyed with phycoerythin (PE).

van der Aar AM, et al. (2013) J Invest Dermatol 133:1240-9

Page 9: Research Techniques Made Simple Flow Cytometry II: Mass and Imaging Cytometry Hung Doan 1,2, Garrett M. Chinn 3, Richard R. Jahan-Tigh 1,2 1 Department

SUMMARY• Mass cytometry (CyTOF) and imaging cytometry (ImageStream) are

technological offshoots of traditional flow cytometry that allow characterization of multiple simultaneous cellular and subcellular parameters.

• Mass cytometry uses rare earth metal isotopes to label cells that are then read by a mass spectrometer, allowing for more than 35 labels on each cell.

• Imaging cytometry (ImageStream) combines high-resolution light and fluorescent microscopy to capture an image for each event passing through the detector.

• During the mass cytometry (CyTOF) process the cells are destroyed, making subsequent cell sorting and analysis impossible.

• Imaging cytometry (ImageStream) suffers from fluorescent dye “spillover,” which continues to limit its multiplexing capabilities.

• The data analysis for both techniques can be complicated given the highly multiplexed nature of these techniques.