RP-HPLC, EFFICIENCY PARAMETERS, RESOLUTION,
COMPARISON OF SENSITIVITY, SELECTIVITY
AND FIELD OF APPLICATIONS OF
DETECTORS
Mr. Shrishail Pattadakal
Monday, April 10, 2023 Mr. Shrishail Pattadkal 2
Contents
Introduction
Types of HPLC
RP-HPLC
HPLC Parameters
Resolution
Conclusion
HPLC Detectors
Monday, April 10, 2023 Mr. Shrishail Pattadkal 3
Introduction
H PLC
= High
= Pressure
= Liquid
= Chromatography
(Performance)
Monday, April 10, 2023 Mr. Shrishail Pattadkal 4
What is HPLC ?
HPLC is a form of column chromatography used frequently in biochemistry and analytical chemistry. To separate, identify and quantify compounds.
Separation technique involving mass-transfer between stationary and mobile phase.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 5
Types of HPLC
HPLC
NP-HPLC RP-HPLC Aq. NP-HPLC
Monday, April 10, 2023 Mr. Shrishail Pattadkal 6
Normal Phase HPLC
Using normal phase solvents :
Most retained compounds are polar
Least retained compounds are less polar
More non-polar solvents make compounds retain longer
Monday, April 10, 2023 Mr. Shrishail Pattadkal 7
Aqueous NP Chromatography
Aqueous normal-phase chromatography (ANP) is a chromatographic technique which encompasses the mobile phase region between reversed-phase chromatography (RP) and organic normal phase chromatography (ONP). This technique is used to achieve unique selectivity for hydrophilic compounds, showing normal phase elution using reverse-phase solvents.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 8
Reverse Phase HPLC
Solvents using :
Most retained compounds are least polarLeast retained compounds are most polarMore polar solvents make compounds retain longer.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 9
Principle• Nonpolar (nonspecific) interactions of analyte with hydrophobic
adsorbent surface (-C18, C8, Phenyl, C4)
• Difference in analyte sorption affinities results in their separation
• More polar analytes retained less
• Analytes with larger hydrophobic part are retained longer
• Almost no separation of structural isomers
Monday, April 10, 2023 Mr. Shrishail Pattadkal 10
Parameters Used In HPLC
1) Retention parameter
2) Column efficiency parameter
3) Peak symmetric parameter
Monday, April 10, 2023 Mr. Shrishail Pattadkal 1111
Retention Parameters
Monday, April 10, 2023 Mr. Shrishail Pattadkal 1212
Efficiency Parameters
Monday, April 10, 2023 Mr. Shrishail Pattadkal 13
Peak symmetry
S : Symmetry factor ( T : Tailing factor )
S = 2f
W0.05
fW0.05
h x 0.05h
S = 1 : The peak is completely symmetric.S > 1 : TailingS < 1 : Leading
Monday, April 10, 2023 Mr. Shrishail Pattadkal 1414
Resolution
Monday, April 10, 2023 Mr. Shrishail Pattadkal 15
Selectivity
• Selectivity (a) is equivalent to the relative retention of the solute peaks and, unlike efficiency, depends strongly on the chemical properties of the chromatography medium.
• The selectivity, a, for two peaks is given by α = k2´ /k1´ = V2 - V0/V1 – V0 = V2/V1
where V1 and V2 are the retention volumes, and k2´ /k1´ are the capacity factors, for peaks 1 and 2 respectively, and V0 is the void volume of the column.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 1616
Eluent Composition Effect on Selectivity
0
2
4
6
8
10
12
14
16
18
0 1 2 3 4 5 6 7
0
5
10
15
20
25
0 1 2 3 4 5 6 7
0
2
4
6
8
10
12
14
16
0 1 2 3 4 5 6 7 8 9
0
2
4
6
8
10
12
0 2 4 6 8 10 12
90% MeCN
80% MeCN
70% MeCN
60% MeCN
Sensitivity sensitivity was considered to define how small a mass or
concentration of a solute could be unambiguously identified by a detector. This concept of sensitivity still persists, although it has been replaced in some cases by minimum detectable mass (MDM) or, alternatively, minimum detectable concentration (MDC).
MDM or MDC are classically defined as the mass or concentration of solute passing through the detector that will give a signal equivalent to twice that of the noise. Originally the signal given by a detector for unit mass or unit concentration change was termed the detector response.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 17
HPLC Detectors
Monday, April 10, 2023 Mr. Shrishail Pattadkal 19
Requirements for an HPLC detector
• Good sensitivity (high signal, low noise)
• No interference from mobile phase
• Must be able to work in a liquid phase environment
Monday, April 10, 2023 Mr. Shrishail Pattadkal 20
Types of Detectors
• Refractive Index• UV-Visible• Fluorescence• Conductivity (for ion chromatography)• Mass Spectrometry
Monday, April 10, 2023 Mr. Shrishail Pattadkal 21
Types of HPLC Detectors
Solute Property
UV-Vis
Electrochemical
Fluorescence
FixedVariable
Photodiode ArrayAmperometry
Pulse Amperometry
Voltammetry
Coulometry
Bulk Property
Refractive Index
ConductivitySuppressed
Non-suppressed
Deflectance Type
Reflectance Type
Monday, April 10, 2023 Mr. Shrishail Pattadkal 22
UV-Vis Absorbance Detector
UV Detectors measure the ability of a sample to absorb light. This can be accomplished at one or several wavelength.
Fixed wavelength measures at one wavelength, usually 254 nm.
Variable wavelength measures at one wavelength at a time, but can detect over a range of wavelengths.
Diode Array measures a spectrum of wavelengths simultaneously.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 23
UV-Vis Absorbance Detector
Monday, April 10, 2023 Mr. Shrishail Pattadkal 24
Fluorescent Detectors
• Fluorescent detectors measures the ability of a compound to absorb then re-emit light at given wavelength.
• Each compound has a characteristic fluorescence.
• The excitation source passes through the flow-cell to a photo detector while a monochromator measures the emission wavelengths.
• Has sensitivity limit of 10-9 to 10-11 gm/ml.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 25
Monday, April 10, 2023 Mr. Shrishail Pattadkal 26
Fluorosence
Monday, April 10, 2023 Mr. Shrishail Pattadkal 27
Electrochemical Detector• It measure compounds that undergo oxidation
or reduction reactions.
• Usually accomplished by measuring gain or loss of electrons from migrating samples as they pass between electrodes at a given difference in electrical potential.
• Has sensitivity of 10-12 to 10-13 gm/ml.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 28
Optical Detectors
There are mainly 4 types of detectors,
1) Circular Dichroism Chiral Detector.
2) Chiral Detector.
3) Chemiluminiscence Detector.
4) Refractive Index Detector.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 29
Circular Dichroism Chiral Detector
• Simultaneous acquisition of UV factor signals.
• Direct determination of optical isomer separation & purity.
• Spectral scanning capability.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 30
Chiral Detectors• These are specially designed for optically active
compounds.
Chemiluminescence Detector
• Detection limits as demand increases for detection of trace amounts of lipids, nucleotides, nitrogen oxides & catachol amines.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 31
Refractive Index
• Monitors are the closest to being universal HPLC detectors, as nearly all dissolved solutes alter the refractive index of the mobile phase.
• They are differential detectors, generating a signal that depends on the difference between the RI of the pure mobile phase and the modified value caused by the dissolved solute.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 32
Instrumentation of Refractive Index
ConclusionHPLC is probably the most universal type of analytical procedure;
its application areas include, quality control, process control, forensic analysis, environmental monitoring and clinical testing. In addition HPLC also ranks as one of the most sensitive analytical procedures and is unique in that it easily copes with multi-component mixtures. It has achieved this position as a result of the constant evolution of the equipment used in LC to provide higher and higher efficiencies at faster and faster analysis times with a constant incorporation of new highly selective column packings.
Monday, April 10, 2023 Mr. Shrishail Pattadkal 33
Monday, April 10, 2023 Mr. Shrishail Pattadkal 34
Bibliography
Separation Chemistry – R.P.Budhiraja, New age international(p) Limited.
Analytical Chromatography – Gurudeep R. Chatwal Edited By Madhu Arora, Himalaya Publishing House