scecal fermilab beam test analysis scecal group meeting kyungpook national university, daegu,...
TRANSCRIPT
ScECAL Fermilab Beam Test analysis
ScECAL Group Meeting
Kyungpook National University, Daegu, Republic of Korea,August 6th , 2010
Adil Khan, Satoru Uozumi, DongHee Kim
MIP Calibration
MIP Calibration Constant
Data Entries Mean RMS
Low Temp Runs09 2160 143.7 27.87
High Temp Runs09 2160 178.2 32.69
2008 Mip Const 2160 160.8 31.39
Mean value of each Energy Point
Run# Energy Point Mean Sigma
530408 3 GeV 435.287 41.0071
530460 6 GeV 920.475 57.365
530506 12 GeV 1829.93 92.746
530513 16 GeV 2427.81 97.0098
530532 25 GeV 3674.23 134.242
530538 32 GeV 4722.46 155.915
Xprofile of Correlated Temp vs Energy
3Gev 6Gev
12Gev 16Gev
25Gev 32Gev
P0:1766.79P1:-64.860
P0: 745.460P1: 49.3527
P0: 3033.53P1:-93.634
P0: 3240.64P1:-43.9452
P0: 2495.76P1: 37.6158 P0:-10544.5
P1:641.561
Run560140Run560155
Run560160
Run560164
Run560180
Run560181
Run560182
Run560185Run560205Run560206Run560215
Run560223Run560224Run560225Run560229Run560232Run560234Run560378 Run560529 Run560562Run560584 Run560586Run560588
Run560254Run560260Run560269Run560271
Run560532Run560533
Run list for Moun 32 Gev 2009 data
Only the circled oneare good runs
Bad Runs
LowTemp Runs
HighTemp Runs
9 Fitting all the channels• Using Gaussian Convoluted-Landau Function
MuonRun 32GeV, Layer1Mip Distribution of High Temp Runs Mip Distribution of Low Temp Runs
MIP Selection
Muon Event Recorded in Online Monitor
Strip =j
Layerx !=i
Event selectionExample: For X-Layer i, Strip j→Check a hit of the strip j on otherX-layers excepting i .(total 14 layers)MIP event Hit Definition:
ADC >ADCped + 3σped
Fitting MIP Distribution• Gaussian convoluted landau
distribution function