scope of accreditation · 2020. 2. 28. · based on jis b 7515 (using calibrations tester &...
TRANSCRIPT
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
All types of Caliper
Digital/
Dial/
Vernier-0.02 mm
GTS-WP-22 Based on
BS EN ISO 13385-1&
BS 887 (only for limits of
error reference)
0 - 600mm 20μm Laboratory/
Customer
Premises
(Std. room/
Metrology)
> 600mm to 1500mm 26μm
1500mm < to 2000mm 35μm
Vernier Caliper
(0.05 mm)
GTS-WP-22
Based on BS EN ISO
13385-1 & BS 887 (only
for limits of error
reference)
0 - 600mm (already
available)
± 36μm
600mm < to 1500mm ± 40μm
1500mm < to 2000mm ± 43μm
External Micrometer
(Digital)
LC:0.001 mm)
GTS-WP-23
Based on BS EN ISO 3611
&
BS 870 (only for limits of
error reference)
0-25mm
(already available)
1.7μm Laboratory/
Customer
Premises
(Std. room/
Metrology)
>25mm up to 100mm 2.2μm
>100mm up to 500mm 8.6μm
>500mm up to 925mm 15 μm
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 2 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
External Micrometer
(Analogue
LC:0.01 mm)
GTS-WP-23
Based on BS EN ISO 3611
&
BS 870 (only for limits of
error reference)
0 to 25mm
(already available)
± 2.9μm Laboratory/
Customer
Premises
(Std. room/
Metrology)
>25mm up to 100mm ± 3.8μm
>100mm up to 500mm ± 8.7μm
>500mm up to 925mm 15μm
Micrometer Setting
Standard
GTS-WP-23
Based on BS EN ISO 3611
(using ULMS)
Up to 100 mm 1.1μm Laboratory
>100 Up to 600mm 3.1μm
GTS-WP-23
Based on BS EN ISO 3611
(using HMS)
Up to 100mm 1.8μm
>100mm Up to 600mm 5.2μm
>600mm Up to 900mm 9.2μm
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 3 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Depth Micrometer
(Digital/analogue)
GTS-WP-28
Based on BS EN ISO 6468
Up to 25mm 2.0μm Laboratory >25mm up to 100mm 2.2μm
>100mm Up to 300mm 4.1μm
Up to 100mm /0.01mm 6.3μm
>100mm Up to 300mm
/0.01 mm
12μm
Tubular Micrometer
/Inside Micrometer
(Digital/ Analogue)
GTS-WP-24
Based on BS EN ISO 959
Up to 300mm 4.3μm Laboratory >300mm Up to 500mm 5.3μm
>500mm up to 1000mm 15μm
0.001mm to
300mm/0.01mm
7.2μm
300mm < to
2000mm/0.01mm
21μm
Inside Micrometer
(Caliper Type)
GTS-WP-24
Based on BS EN ISO 959
0mm to 50mm 2.4μm Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 4 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Dial/Digital
Indicators
GTS-WP-26
Based on BS EN ISO 463&
BS 907(only for limits of
error reference) (using
ULMS)
0.01mm to 100mm
/0.01mm
7.7μm Laboratory
0.001mm to 50mm
/0.001mm
2.3μm
Using Dial calibration
Tester
0.01mm to 25mm
/0.01mm
(Existing Scope )
25mm < to
100mm/0.01mm
7μm
Laboratory/
Customer
Premises
(Std. room)
Bore Gauge
(Ordinary/Digital)
GTS-WP-27
Based on JIS B 7515
(using Calibrations tester
& ULMS)
Up to 400mm/0.001mm 3.2μm Laboratory
Up to 400mm/0.01mm 6.5μm
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 5 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
LVDT
(Ordinary/Digital)
GTS-WP-168
Based on ASTM F2537
(using ULMS/Slip Guges)
UP to 200mm 0.2+(5*L/1000)
μm Laboratory/
Customer
Premises
(Std. room)
Dial Test Indicator /
Lever Type Dial
Gauges
GTS-WP-172
Based on BS EN ISO 463,
BS 2795 & IS 11498
(using ULMS)
Up to 1mm/0.001mm 1.7μm Laboratory
Up to 1mm/0.01mm 5.8μm
Using Dial Calibration
Tester
Up to 1mm/0.01mm 6.9μm Laboratory/
Customer
Premises
Dial / Digital
Thickness Gauges
GTS-WP-36
Based on JIS B7503; JIS
B7524
(Using ULMS/slip gauge)
Up to 25mm /0.001mm 2.8μm Laboratory
Up to 25mm /0.01mm 6.1μm
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 6 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Depth Gauge
(Dial/Digital/Vernier)
GTS-WP-29
Based on BS EN ISO
13385-2 (Using slip
gauges and HMS)
Up to 450mm/0.02mm ± 23μm Laboratory
Up to 450mm/0.01mm ±18μm
Height Gauge
(Digital/Dial/Analogu
e)
GTS-WP-25
Based on BS EN ISO
13225/
BS-1643 (Using slip
gauges and HMS)
Up to 600mm ± 14μm Laboratory/
Customer
Premises
(Std. room/
Metrology)
Up to 1000mm ± 17μm
Feeler Gauge GTS-WP-56
Based on BS 957
Up to 1mm 3.1μm
Radius Gauge GTS-WP-81
Based on IS 5273-1969
(using Profile Projector)
Up to 25mm 9.0μm Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 7 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Thread / Screw Pitch
Gauge (Metric/inch)
GTS-WP-126
Based on IS 4211 (using
Profile Projector)
Up to 7mm 5.2μm Laboratory
Up to 42 TPI 170μin
Thread Plug gauges
(Metric / Unified/BSP
(or) G threads )
For simple Pitch
Diameter
GTS-WP-70
Based on Euramet cg-10
(using ULMS & Profile
Projector) Metric Threads
1mm Up to 100mm 1.7μm
>100 Up to 200mm 2.4μm
Inch - Unified / BSP 1/16" to 4" 63μin
Above 4" and including 8" 91μin
Thread Ring gauges
(Metric / Unified/BSP
(or) G threads ) For simple Pitch
diameter
GTS-WP-137
Based on Euramet cg-10
(Using ULMS) Metric
Threads
3mm to 14mm 1.7μm Laboratory >14mm up to and
including 100mm
1.9μm
Inch - Unified / BSP 1/8" to 1/2" 67μin
1/2"< and including 4" 72μin
Thread Plug/Ring GTS-WP-173 1/8" to 1/2" 68μin Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 8 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
gauge – Taper
(NPT / BSPT)
For simple Pitch
Diameter
Based on JIS B 0262 &
Euramet cg-10 (using
ULMS)
1/2"< and including 4" 87μin
Setting / Plain Plug
gauge (Metric/Inch)
GTS-WP-80
Based on ASME B89.1.5
(using ULMS)
1mm to 50mm 0.85μm Laboratory >50mm Up to 100mm 1.1μm
>100mm Up to 400mm 4.2μm
Setting / Plain Ring
gauge (Metric/Inch)
GTS-WP-106
Based on BS EN ISO 4064
(using ULMS and Master
setting ring)
1mm to 14mm 0.68μm 14mm < to 400mm 2.1μm
Height Measuring
System (HMS)/
Digital Height Gauge
L.C: 0.001mm and
greater
GTS-WP-169
Based on BS EN ISO
13225/BS 1643
(Slip Gauges & Long slip
gauges)
Up to 1000mm 0.9+(6.5L/1000)
μm
Where “L” is in mm
Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 9 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Universal Length
Measuring System
(ULMS)
GTS-WP-139
(using slip gauges)
Up to 100mm (absolute)
Up to 600mm
(differential)
0.18 +
(2.5L/1000) μm
(Where L is in
mm)
Laboratory
Profile Projector
GTS-WP-158
Based on JIS 7540
(Using Glass scale and
Cross wire chart)
Up to 200mm 2.8 + (2.2
L/1000) μm
0.19°(11.4’)
(Where L is in
mm)
Laboratory
Thread Measuring
Cylinder
GTS-WP-170
(using ULMS) Based on
BS 3777 & BS 5590
up to 6.35mm 0.67μm Laboratory
Using High Precision
Micrometer
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 10 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Cylindrical Standards
/ Measuring Pins
GTS-WP-170
(using ULMS)
Based on IS-11103
Up to 12mm 0.67μm Laboratory
Using High Precision
Micrometer
Steel Scale GTS-WP-171
Based onOIML R035-1-e&
BS 4372: 1968
(Using Profile Projector,
dig. Micrometer)
Up to 1000mm / 0.5mm 0.30mm Laboratory Up to 1000mm / 1mm 0.60mm
Caliper Checker GTS-WP-164
(using HMS and Ext. Slip
Gauge) Based on
Manufacturer Spec.
Up to 600mm 1+(6.7*L/1000)
μm
Where L is in
mm
Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 11 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Depth Micro Checker
GTS-WP-165 and 166
(using HMS and Ext. Slip
Gauge) Based on
Manufacturer Spec.
Up to 300mm 3.9μm Laboratory
Inside Micro checker GTS-WP-166
(using HMS and Ext. Slip
Gauge) Based on
Manufacturer Spec.
Up to 300mm
3.5μm Laboratory
>300mm Up to 600mm 5.4μm
Dial Calibration
Tester
GTS-WP-167
(using HMS and Slip
Gauge/ULMS)Based on
Manufactured Spec.
Up to 25mm 1.5μm Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Dimensional Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and Specification Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 12 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)
Test Sieves GTS-WP-43
Based on ISO 3310-1
(Using Profile Projector)
50μm to 4.3mm
4.0μm Laboratory
GTS-WP-43
Based on ISO 3310-1
(using digital caliper)
5mm to 125mm 32μm Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Temperature Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)
Liquid-in-glass
thermometers
GTS-WP-13 based on BS
1041-2.1
-30 °C up to 150 °C 0.16 °C Laboratory
Direct reading
thermometers
GTS-WP-15 (in house method) -30 °C up to 160 °C 0.16 °C Laboratory
>160 °C up to 500 °C 0.4 °C
>500 °C up to 900 °C 1.3 °C
>900 °C up to 1200 °C 4 °C
Dial thermometers GTS-WP-14 based on EN
13190
-30 °C up to 160 °C 0.16 °C Laboratory
>160 °C up to 400 °C 2 °C
400 °C up to 800 °C 4 °C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Temperature Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 2 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)
Base metal
thermocouples
GTS WP-12 -40°C up to 250°C 0.3 °C Laboratory
250°C up to 600 °C 0.6 °C
600 °C up to 900°C 0.9 °C
900 °C up to 1200 °C 4 °C
Noble metal
thermocouples
GTS-WP-12 0 °C up to 600 °C 0.6 °C Laboratory
600 °C up to 900 °C 0.8 °C
900 °C up to 1200 °C 1.7 °C
Climatic Chamber GTS-WP-09
DKD-R5-7
(9 points)
30 °C up to 180 °C 1.1 °C Laboratory
/ Customer
Premises
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Temperature Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 3 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)
Water bath, incubators GTS-WP-09
DKD-R5-7
(5 points)
5 °C up to 95 °C 1.1 °C Laboratory
/ Customer
Premises
Freezer / Chiller GTS –WP-154
DKD-R5-7
(9 points)
-30 °C up to 95 °C 1.1 °C Laboratory
/ Customer
Premises
Furnace, Oven GTS-WP 09
DKD-R5-7
(9 points, muffle furn.:
1 point)
30 °C up to 180 °C 1.1 °C Laboratory
/ Customer
Premises 180 °C up to 300 °C 1.3 °C
300 °C up to 800 °C 4.0 °C
800 °C up to 1200 °C 9.0 °C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Temperature Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 4 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)
Auto Clave
(Temperature)
GTS-WP-155
DKD-R5-7
(5-9 points)
100 °C up to 140 °C 2 °C Laboratory
/ Customer
Premises
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Sourcing
AC Voltage
Direct Method using Fluke
9100
U= Voltage
1 μV to 10 mV Laboratory
/Customer
Premises
10 Hz - 3 kHz 0.46 x 10-3 U + 0.44 mV
>3 kHz - 10 kHz 0.46 x 10-3 U + 0.59 mV
>10 kHz - 30 kHz 0.69 x 10-3 U + 1.1 mV
>30 kHz - 50 kHz 1.0 x 10-3 U + 2.2 mV
>50 kHz - 100 kHz 2.3 x 10-3 U + 5.9 mV
10 mV to 32 mV
10 Hz - 3 kHz 0.46 x 10-3 U + 0.11 mV
>3 kHz - 10 kHz 0.4 x 10-3 U + 0.15 mV
>10 kHz - 30 kHz 0.70 x 10-3 U + 0.28 mV
>30 kHz - 50 kHz 1.0 x 10-3 U + 0.56 mV
>50 kHz - 100 kHz 2.3 x 10-3 U + 1.5 mV
32 mV to 320 mV
10 Hz - 3 kHz 0.47 x 10-3 U + 22 μV
>3 kHz - 10 kHz 0.47 x 10-3 U + 29 μV
>10 kHz - 30 kHz 0.70 x 10-3 U + 56 μV
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 2 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Voltage
Direct Method using Fluke
9100
U= Voltage
32 mV to 320 mV Laboratory
/Customer
Premises
>30 kHz - 50 kHz 1.1 x 10-3 U + 0.11 mV
>50 kHz - 100 kHz 2.3 x 10-3 U + 0.30 mV
0.32 V to 3.2 V
10 Hz - 3 kHz 0.48 x 10-3 U + 0.22 mV
>3 kHz - 10 kHz 0.47 x 10-3 U + 0.29 mV
>10 kHz - 30 kHz 0.70 x 10-3 U + 0.55 mV
>30 kHz - 50 kHz 1.1 x 10-3 U + 1.1 mV
>50 kHz - 100 kHz 2.3 x 10-3 U + 3.0 mV
3.2 V to 32 V
10 Hz - 3 kHz 0.48 x 10-3 U + 2.2 mV
>3 kHz - 10 kHz 0.71 x 10-3 U + 2.9 mV
>10 kHz - 30 kHz 0.93 x 10-3 U + 5.5 mV
>30 kHz - 50 kHz 1.7 x 10-3 U + 11 mV
>50 kHz - 100 kHz 4.1 x 10-3 U + 37 mV
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 3 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Voltage
Direct Method using Fluke
9100
U= Voltage
32 V to 105 V Laboratory
/Customer
Premises
10 Hz - 3 kHz 0.47 x 10-3 U + 7.3 mV
>3 kHz - 10 kHz 0.70 x 10-3 U + 9.7 mV
>10 kHz - 30 kHz 0.93 x 10-3 U + 18 mV
>30 kHz - 50 kHz 1.7 x 10-3 U + 36 mV
>50 kHz - 100 kHz 4.1 x 10-3 U + 0.12 V
105 V to 320 V
40 Hz to 100 Hz 0.6 x 10-3 U + 22 mV
>100 Hz - 1 kHz 0.6 x 10-3 U + 22 mV
>1 kHz - 3 kHz 0.94 x 10-3 U + 22 mV
>3 kHz - 10 kHz 0.94 x 10-3 U + 37 mV
>20 kHz - 30 kHz 1.7 x 10-3 U + 74 mV
320 V to 800 V
40 Hz to 100 Hz 0.59 x 10-3 U + 73 mV
>100 Hz - 1 kHz 0.59 x 10-3 U + 73 mV
>1 kHz - 3 kHz 0.93 x 10-3 U + 73 mV
>3 kHz - 10 kHz 0.93 x 10-3 U + 0.12 V
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 4 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Voltage
Direct Method using Fluke
9100
U= Voltage
800V to 1050 V Laboratory
/Customer
Premises
40 Hz to 100 Hz 0.59 x 10-3 U + 0.15 V
>100 Hz - 1 kHz 0.59 x 10-3 U + 0.15 V
>1 kHz - 3 kHz 0.93 x 10-3 U + 0.15 V
>3 kHz - 10 kHz 0.93 x 10-3 U + 0.24 V
AC Voltage
Direct Method using Fluke
5522A
U= Voltage
1.0 mV – 33 mV Laboratory
/Customer
Premises
>10 Hz - 45 Hz 0.93 x 10-3 U + 7.03 μV
>45 Hz -10 kHz 0.18 x 10-3 U + 7.03 μV
>10 kHz - 20 kHz 0.24 x 10-3 U + 7.03 μV
>20 kHz - 50 kHz 1.16 x 10-3 U + 7.03 μV
>50 kHz - 100 kHz 4.05 x 10-3 U + 13.9 μV
>100 kHz - 500 kHz 9.28 x 10-3 U + 57.8 μV
33 mV – 330 mV
10 Hz - 45 Hz 0.35 x 10-3 U + 9.20 μV
>45 Hz -10 kHz 0.18 x 10-3 U + 9.09 μV
>10 kHz - 20 kHz 0.20 x 10-3 U + 9.10 μV
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 5 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Voltage
Direct Method using Fluke
5522A
U = Voltage
33 mV – 330 mV Laboratory
/Customer
Premises
>20 kHz - 50 kHz 0.41 x 10-3 U + 9.20 μV
>50 kHz - 100 kHz 0.93 x 10-3 U + 36.90 μV
>100 kHz - 500 kHz 2.38 x 10-3 U + 79.90 μV
0.33V - 3.3 V
10 Hz - 45 Hz 0.35 x 10-3 U + 57.60 μV
>45 Hz -10 kHz 0.18 x 10-3 U + 68.30 μV
>10 kHz - 20 kHz 0.23 x 10-3 U + 68.60 μV
>20 kHz - 50 kHz 0.35 x 10-3 U + 57.60 μV
>50 kHz - 100 kHz 0.81 x 10-3 U + 0.14 mV
>100 kHz - 500 kHz 2.80 x 10-3 U + 0.69 mV
3.3V – 33 V
10 Hz - 45 Hz 0.35 x 10-3 U + 0.75 mV
>45 Hz -10 kHz 0.18 x 10-3 U + 0.68 mV
>10 kHz - 20 kHz 0.28 x 10-3 U + 0.69 mV
>20 kHz - 50 kHz 0.41 x 10-3 U + 0.69 mV
>50 kHz - 100 kHz 1.05 x 10-3 U + 1.84 mV
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 6 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Voltage
Direct Method using Fluke
5522A
U = Voltage
33 V-330 V Laboratory
/Customer
Premises
45 Hz - 1 kHz 0.23 x 10-3 U + 2.32 mV
>1 kHz - 10 kHz 0.24 x 10-3 U + 6.85 mV
>10 kHz - 20 kHz 0.30 x 10-3 U + 6.87 mV
>20 kHz - 50 kHz 0.36 x 10-3 U + 6.78 mV
>50 kHz - 100 kHz 2.23 x 10-3 U + 57.60 mV
330 V - 1020 V
45 Hz - 1 kHz 0.35 x 10-3 U + 12.10 mV
>1 kHz - 5 kHz 0.30 x 10-3 U + 12.10 mV
>5 kHz - 10 kHz 0.36 x 10-3 U + 11.70 mV
DC Voltage
Direct Method using Fluke
9100
U = Voltage
0.001 mV to 320.000
mV
87 x 10-6 U + 5 μV
Laboratory
/Customer
Premises 0.32001 V to 3.20000
V
89 x 10-6 U + 48 μV
3.2001 V to 32.0000 V 93 x 10-6 U + 0.46 mV
32.001 V to 320.000 V 93 x 10-6 U + 5 mV
320.01 V to 1050.00 V 88 x 10-6 U + 22 mV
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 7 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
DC Voltage
Direct Method using Fluke
5522A
U =Voltage
0 - 329.9999 mV 58 x 10-6 U + 2 μV Laboratory
/Customer
Premises
0.33 V - 3.299999 V 59 x 10-6 U + 1 μV
3.3 V - 32.99999 V 60 x 10-6 U + 7 μV
33 V - 329.9999 V 62 x10-6 U + 0.069 mV
330 - 1020.000 V 61 x 10-6 U + 71 mV
DC Current
Direct Method using Fluke
9100
= Current
0.001 μA to 320.000
μA
0.17 x 10-3 + 0.013 μA Laboratory
/Customer
Premises 0.32001 mA to
3.20000 mA
0.18 x 10-3 + 0.094 μA
3.2001 mA to 32.0000
mA
0.18 x 10-3 + 1.0 μA
32.001 mA to 320.000
mA
0.20 x 10-3 + 11 μA
0.32001 A to 3.20000 A 0.70 x 10-3 + 0.14 mA
3.2001 A to 10.5000 A 0.64 x 10-3 + 1.1 mA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 8 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
DC Current
Direct Method using Fluke
9100
= Current
10-Turn Coil Laboratory
/Customer
Premises
3.2001 A to 32.0000
A
2.4 x 10-3 + 0.51 mA
32.001 A to 105.000
A
2.4 x 10-3 + 3.8 mA
50-Turn Coil
16.001 A to 160.000
A
2.4 x 10-3 + 2.5 mA
160.01 A to 525.00 A 2.4 x 10-3 + 19 mA
525.01 A to 1000.00
A
2.4 x 10-3 + 0.104 A
Direct Method using Fluke
5522A
=Current
0 - 329.999 μA 0.17 x 10-3 + 0.023 μA Laboratory
/Customer
Premises
0.33 mA - 3.29999
mA
0.12 x 10-3 + 0.059 μA
3.3 mA - 32.99999
mA
0.12 x 10-3 + 0.30 μA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 9 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
DC Current Direct Method using Fluke
5522A
=Current
33 mA - 329.999 mA 0.12 x 10-3 + 3.00 μA Laboratory
/Customer
Premises
0.33 A - 1.09999 A 0.23 x 10-3 + 0.047 mA
1.1 A - 2.99999 A 0.44 x 10-3 + 0.046 mA
3 A - 10.9999 A 0.59 x 10-3 + 0.57 mA
11 A - 20.5 A 1.21 x 10-3 + 0.84 mA
50- Turn Coil
0.2 A - 0.33 A 5.84 x 10-3 + 23 mA
0.33 A - 2.9999 A 5.83 x 10-3 + 0.16 A
3 A - 20.5 A 5.88 x 10-3 + 0.58 A
AC Current
Direct Method using Fluke
9100
= Current
1.0 μA to 320.0 μA Laboratory
/Customer
Premises
10 Hz - 3 kHz 0.83 x 10-3 + 0.35 μA
>3 kHz - 10 kHz 1.2 x 10-3 + 0.69 μA
>10 kHz - 20 kHz 2.3 x 10-3 + 2.3 μA
>20 kHz - 30 kHz 2.9 x 10-3 + 3.5 μA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 10 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
9100
= Current
0.32 mA to 3.2 mA Laboratory
/Customer
Premises
10 Hz - 3 kHz 0.85 x 10-3 + 0.34 μA
>3 kHz - 10 kHz 1.2 x 10-3 + 0.68 μA
>10 kHz - 20 kHz 2.4 x 10-3 + 2.3 μA
>20 kHz - 30 kHz 2.9 x 10-3 + 3.5 μA
3.2 mA to 32 mA
10 Hz - 3 kHz 0.85 x 10-3 + 3.6 μA
>3 kHz - 10 kHz 1.2 x 10-3 + 7.3 μA
>10 kHz - 20 kHz 2.4 x 10-3 + 15 μA
>20 kHz - 30 kHz 2.9 x 10-3 + 26 μA
32 mA - 320 mA
10 Hz - 3 kHz 0.9 x 10-3 + 36 μA
>3 kHz - 10 kHz 1.2 x 10-3 + 54 μA
>10 kHz - 20 kHz 2.4 x 10-3 + 74 μA
>20 kHz - 30 kHz 2.9 x 10-3 + 0.11 mA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 11 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
9100
= Current
0.32 A to 3.2 A Laboratory
/Customer
Premises
10 Hz - 3 kHz 1.2 x 10-3 + 0.55 μA
>3 kHz - 10 kHz 2.9 x 10-3 + 3 mA
3.2 A to 10.5 A
10 Hz - 3 kHz 2.3 x 10-3 + 3.5 mA
>3 kHz - 10 kHz 5.8 x 10-3 + 12 mA
10-Turn Coil
3.2001 A to 32.0000 A
10 Hz - 100 Hz 3.3 x 10-3 + 5.2 mA
>100 Hz - 440 Hz 9.3 x 10-3 + 31 mA
32 A to 200 A
10 Hz - 100 Hz 3.3 x 10-3 + 91 mA
>100 Hz - 440 Hz 8.1 x 10-3 + 0.28 A
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 12 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
9100
= Current
50-Turn Coil Laboratory
/Customer
Premises
16.001 A to 160.000
A/10 Hz - 100 Hz
3.3 x 10-3 + 27 mA
160.01 A to 1000.00
A/10 Hz - 100 Hz
3.3 x 10-3 + 0.45 A
Direct Method using Fluke
5522A
= Current
30 - 330 μA
10Hz-20Hz 2.31 x 10-3 + 0.12 μA
>20Hz - 45Hz 1.73 x 10-3 + 0.12 μA
>45Hz - 1kHz 1.45 x 10-3 + 0.12 μA
>1kHz - 5kHz 3.47 x 10-3 + 0.17 μA
>5kHz - 10kH 9.25 x 10-3 + 0.23 μA
>10kHz - 30kHz 18.5 x 10-3 + 0.46 μA
0.33 - 3.3 mA
10Hz - 20Hz 2.31 x 10-3 + 0.17 μA
>20Hz - 45Hz 1.45 x 10-3 + 0.17 μA
>45Hz - 1kHz 1.16 x 10-3 + 0.17 μA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 13 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
5522A
= Current
0.33 - 3.3 mA Laboratory
/Customer
Premises
>1kHz - 5kHz 2.31 x 10-3 + 0.23 μA
>5kHz - 10kHz 5.78 x 10-3 + 0.35 μA
>10kHz - 30kHz 11.6 x 10-3 + 0.69 μA
3.3 - 33 mA
10Hz -20Hz 2.08 x 10-3 + 2.31 μA
>20Hz - 45Hz 1.05 x 10-3 + 2.31 μA
>45Hz - 1kHz 0.49 x 10-3 + 2.26 μA
>1kHz - 5kHz 0.97 x 10-3 + 2.25 μA
>5kHz - 10kHz 2.33 x 10-3 + 2.30 μA
>10kHz - 30kHz 4.64 x 10-3 + 2.31 μA
33 -330 mA
10Hz -20Hz 2.08 x 10-3 + 23.1 μA
>20Hz - 45Hz 1.04 x 10-3 + 23.1 μA
>45Hz - 1kHz 0.47 x 10-3 + 23 μA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 14 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
5522A
= Current
33 -330 mA Laboratory
/Customer
Premises
>1kHz - 5kHz 1.16 x 10-3 + 57.7 μA
>5kHz - 10kHz 2.31 x 10-3 + 0.12 mA
>10kHz - 30kHz 4.63 x 10-3 + 0.23 mA
0.33 - 1.1 A
10Hz - 45Hz 2.08 x 10-3 + 12 mA
>45Hz - 1kHz 0.59 x 10-3 + 11 mA
>1kHz - 5kHz 6.96 x 10-3 + 1.15 mA
>5kHz - 10kHz 28.9 x 10-3 + 5.78 mA
1.11 - 3 A
10Hz - 45Hz 2.08 x 10-3 + 0.12 mA
>45Hz - 1kHz 0.70 x 10-3 + 0.12 mA
>1kHz - 5kHz 6.94 x 10-3 + 1.16 mA
>5kHz - 10kHz 28.9 x 10-3 + 5.78 mA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 15 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
5522A
= Current
3 -11 A Laboratory
/Customer
Premises
45Hz - 100 Hz 0.71 x 10-3 + 2.30 mA
>100Hz - 1kHz 1.16 x 10-3 + 2.30 mA
>1kHz - 5kHz 34.7 x 10-3 + 2.31 mA
11 - 20.5 A
45Hz - 100 Hz 1.4 x 10-3 + 5.74 mA
>100Hz - 1kHz 1.75 x 10-3 + 5.74 mA
>1kHz - 5kHz 34.7 x 10-3 + 5.78 mA
Amp-50 turn
10 -16.5A @ 45Hz-
65Hz
4.08 x 10-3 + 2.83 mA
16.5 – 150 A @
45Hz-65Hz
4.04 x 10-3 + 25.00 mA
150 – 1025 A @
45Hz-65Hz
4.02 x 10-3 + 83.70 mA
10 -16.4999/ 65Hz-
440Hz
9.47 x 10-3 + 3.57 mA
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 16 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Current
Direct Method using Fluke
5522A
= Current
Amp-50 turn Laboratory
/Customer
Premises
16.5 - 149.999/ 65Hz-
440Hz
9.45 x 10-3 + 30.40 mA
150 - 1025/ 65Hz-
440Hz
9.52 x 10-3 + 0.11 A
Resistance
Direct Method using Fluke
9100
R= Resistance
0.0001 Ω to 40.0000 Ω 0.33 x 10-3 R + 12 mΩ Laboratory
/Customer
Premises
40.001 Ω to 400.000 Ω 0.23 x 10-3 R + 23 mΩ
0.40001 kΩ to 4.00000 kΩ
0.17 x 10-3 R + 93 mΩ
4.0001 kΩ to 40.0000 kΩ
0.17 x 10-3 R + 0.93 Ω
40.001 kΩ to 400.000 kΩ
0.21 x 10-3 R + 9.3 Ω
0.40001 MΩ to 4.00000 MΩ
0.23 x 10-3 R + 0.12 kΩ
4.0001 MΩ to 40.0000 MΩ
0.59 x 10-3 R + 2.3 kΩ
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 17 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Resistance
Direct Method using Fluke
9100
R= Resistance
40.001 MΩ to 400.000 MΩ
0.71 x 10-3 R + 46 kΩ Laboratory
/Customer
Premises
Direct Method using Fluke
5522A
R= Resistance
0 - 10.9999 Ω 46 x10-6 R + 12 mΩ
11 Ω - 32.9999 Ω 35 x10-6 R + 17 mΩ
33 Ω - 109.9999 Ω 33 x10-6 R + 17 mΩ
110 Ω - 329.9999 Ω 33 x10-6 R + 23 mΩ
330 Ω - 1.099999 kΩ 33 x10-6 R + 23 mΩ
1.1 kΩ - 3.299999 kΩ 33 x10-6 R + 0.23 Ω
3.3 kΩ -10.99999 kΩ 33 x10-6 R + 0.12 Ω
11 kΩ - 32.99999 kΩ 33 x10-6 R + 1.15 Ω
33 kΩ - 109.9999 kΩ 33 x10-6 R + 1.15 Ω
110 kΩ - 329.99999 kΩ 37 x10-6 R + 12 Ω
330 kΩ - 1.099999 MΩ 38 x10-6 R + 12 Ω
1.1 MΩ - 3.299999 MΩ 70 x10-6 R + 0.17 kΩ
3.3 MΩ - 10.99999 MΩ 0.15 x 10-3 R + 0.29 kΩ
11 MΩ - 32.99999 MΩ 0.30 x 10-3 R + 2.84 kΩ
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 18 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Resistance
Direct Method using Fluke
5522A
R= Resistance
33 MΩ - 109.9999 MΩ 0.62 x 10-3 R + 3.30 kΩ Laboratory
/Customer
Premises
110 MΩ - 329.9999 MΩ 3.48 x 10-3 R + 0.12 MΩ
330 MΩ - 1100 MΩ 17.40 x 10-3 R + 0.58 MΩ
Capacitance
Direct Method using Fluke
9100
C= Capacitance
0.5000 nF to 4.0000 nF 25 pF Laboratory
/Customer
Premises
4.0001 nF to 40.000 nF 0.11 nF
40.001 nF to 400.00 nF 0.96 nF
400.01 nF to 4.0000 μF 4.7 x 10-3 C + 1.9 nF
4.0001 μF to 40.000 μF 5.8 x 10-3 C + 19 nF
40.001 μF to 400.00 μF 5.8 x 10-3 C + 0.19 μF
400.01 μF to 4.0000 mF 5.8 x 10-3 C + 1.8 μF
4.0001 mF to 40.000 mF 12 x 10-3 C + 69 μF
Direct Method using Fluke
5522A
C= Capacitance
220 pF - 399.9 pF 9.16 x 10-3 C+ 11 pF
0.4 nF - 1.0999 nF 6.21 x 10-3 C+ 12 pF
1.1 nF - 3.2999 nF 5.93 x 10-3 C + 12 pF
3.3 nF - 10.9999 nF 2.97 x 10-3 C + 12 pF
11 nF - 32.9999 nF 2.95 x 10-3 C + 0.12 nF
33 nF - 109.999 nF 2.96 x 10-3 C + 0.12 nF
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 19 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Capacitance
Direct Method using Fluke
5522A
C= Capacitance
110 nF - 329.999 nF 2.96 x 10-3 C + 0.34 nF Laboratory
/Customer
Premises
0.33 μF - 1.09999 μF 2.97 x 10-3 C + 1.14 nF
1.1 μF- 3.29999 μF 2.96 x 10-3 C + 3.43 nF
3.3 μF - 10.9999 μF 2.96 x 10-3 C + 11.40 nF
11 μF - 32.9999 μF 4.68 x 10-3 C + 34.50 nF
33 μF - 109.999 μF 5.37 x 10-3 C + 0.12 μF
110 μF -329.999 μF 5.37 x 10-3 C + 0.34 μF
0.33 mF - 1.09999 mF 5.37 x 10-3 C + 1.14 μF
1.1 mF - 3.29999 mF 5.37 x 10-3 C + 3.41 μF
3.3 mF - 10.9999 mF 5.37 x 10-3 C + 11.4 μF
11 mF - 32.9999 mF 8.78 x 10-3 C + 34.4 μF
33 mF - 110 mF 12.80 x 10-3 C + 0.12 mF
Frequency
Direct Method using Fluke
9100
f= Frequency
0.5 Hz to 10.0 MHz 29 x 10-6 f Laboratory
/Customer
Premises
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 20 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Frequency
Direct Method using Fluke
5522A
f= Frequency
0.01 Hz - 119.99 Hz 2.87 x 10-6 f + 13 μHz Laboratory
/Customer
Premises
120 Hz - 1199.9 Hz 2.91 x 10-6 f + 26 μHz
1.200 kHz - 11.999 kHz 2.91 x 10-6 f + 0.21 mHz
12.00 kHz - 119.99 kHz 2.91 x 10-6 f + 2.05 mHz
120.00 kHz - 1199.9 kHz 2.91 x 10-6 f + 20.40 mHz
1.200 MHz - 2.000 MHz 2.93 x 10-6 f + 0.29 Hz
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 21 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Measuring
DC Voltage
Direct Method using Fluke
8846A
U= Voltage
100 mv 43 x 10-6 U + 4 μV Laboratory
/Customer
Premises
1 V 31 x 10-6 U + 8 μV
10 V 30 x 10-6 U + 57 μV
100 V 46 x 10-6 U + 0.69 mV
1000 V 49 x 10-6 U + 12 mV
AC Voltage
Direct Method using Fluke
8846A
U= Voltage
100 mV Laboratory
/Customer
Premises
5 Hz to 10 kHz 4.1 x 10-3 U + 46 μV
>10 Hz - 20 kHz 0.72 x 10-3 U + 46 μV
>20 kHz - 50 kHz 1.5 x 10-3 U + 58 μV
>50 kHz - 100 kHz 7 x 10-3 U + 93 μV
1 V
5 Hz to 10 kHz 4.1 x 10-3 U + 0.35 mV
>10 Hz - 20 kHz 0.70 x 10-3 U + 0.35 mV
>20 kHz - 50 kHz 1.4 x 10-3 U + 0.58 mV
>50 kHz - 100 kHz 6.9 x 10-3 U + 0.93 mV
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 22 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Voltage
Direct Method using Fluke
8846A
U= Voltage
10 V Laboratory
/Customer
Premises 5 Hz to 10 kHz 4.1 x 10-3 U + 3.5 mV
>10 Hz - 20 kHz 0.7 x 10-3 U + 3.5 mV
>20 kHz - 50 kHz 1.4 x 10-3 U + 5.8 mV
>50 kHz - 100 kHz 7 x 10-3 U + 9.3 mV
100 V
5 Hz to 10 Hz 4.1 x 10-3 U + 35 mV
>10 Hz - 20 kHz 0.7 x 10-3 U + 35 mV
>20 kHz - 50 kHz 1.4 x 10-3 U + 58 mV
>50 kHz - 100 kHz 7 x 10-3 U + 93 mV
1000 V
5 Hz to 10 Hz 4.1 x 10-3 U + 0.35 V
>10 Hz - 20 kHz 0.72 x 10-3 U + 0.35 V
>20 kHz - 50 kHz 1.8 x 10-3 U + 0.55 V
>50 kHz - 100 kHz 7.03 x 10-3 U + 0.92 V
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 23 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
DC Current
Direct Method using Fluke
8846A
= Current
100 μA 0.59 x 10-3 + 0.03 μA Laboratory
/Customer
Premises 1 mA 0.58 x 10-3 + 0.06 μA
10 mA 0.58 x 10-3 + 2.3 μA
100 mA 0.58 x 10-3 + 5.8 μA
1 A 0.59 x 10-3 + 0.23 mA
10 A 1.8 x 10-3 + 0.92 mA
AC Current
Direct Method using Fluke
8846A
= Current
100 μA/10 Hz to 1 kHz 1.2 x 10-3 + 0.05 μA Laboratory
/Customer
Premises 1 mA/ 10 Hz to 1 kHz 1.2 x 10-3 + 0.46 μA
10 mA /10 Hz to 1 kHz 1.2 x 10-3 + 4.6 μA
100 mA/10 Hz to 1 kHz 1.2 x 10-3 + 46 μA
1 A /10 Hz to 1 kHz 1.2 x 10-3 + 0.46 mA
10 A/ 10 Hz to 1 kHz 1.8 x 10-3 + 6.9 mA
Resistance
Direct Method using Fluke
8846A
R= Resistance
0.0001 Ω to 10.0000 Ω 0.12 x 10-3 R+ 3.5 mΩ Laboratory
/Customer
Premises 10.001 Ω to 100.000 Ω 0.12 x 10-3 R + 4.6 mΩ
0.1001 kΩ to 1.0 kΩ 0.12 x 10-3 R + 12 mΩ
1.0001 kΩ to 10.0000 kΩ
0.12 x 10-3 R + 0.12 Ω
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 24 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Resistance
Direct Method using Fluke
8846A
R= Resistance
10.001 kΩ to 100.000 kΩ
0.12 x 10-3 R + 1.2 Ω Laboratory
/Customer
Premises 0.10001 MΩ to 1.00000 MΩ
0.12 x 10-3 R + 11.4 Ω
1.0001 MΩ to 10.0000 MΩ
0.47 x 10-3 R + 0.12 kΩ
10.001 MΩ to 100.000 MΩ
9.3 x 10-3 R + 12 kΩ
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
-200 ⁰C to -80 ⁰C /
RTD-pt385,100 Ω
0.29 x 10-3 t + 0.36 m ⁰C Laboratory
/Customer
Premises -80 ⁰C to 0.003 ⁰C /
RTD-pt385,100 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.03 ⁰C to 100 ⁰C /
RTD-pt385,100 Ω
0.78 x 10-3 t+ 3.45 m ⁰C
100 ⁰C to 300 ⁰C / RTD-
pt385,100 Ω
0.35 x 10-3 t + 0.23 m ⁰C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 25 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
300 ⁰C to 400 ⁰C / RTD-
pt385,100 Ω
0.29 x 10-3 t + 0.12 m ⁰C Laboratory
/Customer
Premises 400 ⁰C to 630 ⁰C / RTD-
pt385,100 Ω
0.22 x 10-3 t + 0.11 m ⁰C
630 ⁰C to 800 ⁰C / RTD-
pt385,100 Ω
0.33 x 10-3 t + 0.051 m ⁰C
-200 ⁰C to -80 ⁰C /
RTD-pt3926,100 Ω
0.29 x 10-3 t + 0.36 m ⁰C
-80 ⁰C to 0.003 ⁰C /
RTD-pt3926,100 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.03 ⁰C to 100 ⁰C /
RTD-pt3926,100 Ω
0.78 x 10-3 t + 3.45 m ⁰C
100 ⁰C to 300 ⁰C / RTD-
pt3926,100 Ω
0.35 x 10-3 t + 0.23 m ⁰C
300 ⁰C to 400 ⁰C / RTD-
pt3926,100 Ω
0.29 x 10-3 t + 0.12 m ⁰C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 26 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
400 ⁰C to 630 ⁰C / RTD-
pt3926,100 Ω
0.22 x 10-3 t + 0.11 m ⁰C Laboratory
/Customer
Premises -200 ⁰C to -190 ⁰C /
RTD-pt3916,100 Ω
1.44 x 10-3 t+ 0.043 m ⁰C
-190 ⁰C to -80 ⁰C /
RTD-pt3916,100 Ω
0.24 x 10-3 t + 0.43 m ⁰C
-80 ⁰C to 0.003 ⁰C /
RTD-pt3916,100 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.03 ⁰C to 100 ⁰C /
RTD-pt3916,100 Ω
0.66 x 10-3 t + 3.45 m ⁰C
100 ⁰C to 260 ⁰C / RTD-
pt3916,100 Ω
0.31 x 10-3 t + 0.27 m ⁰C
260 ⁰C to 300 ⁰C / RTD-
pt3916,100 Ω
0.31 x 10-3 t + 0.14 m ⁰C
300 ⁰C to 400 ⁰C / RTD-
pt3916,100 Ω
0.26 x 10-3 t + 0.14 m ⁰C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 27 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
400 ⁰C to 600 ⁰C / RTD-
pt3916,100 Ω
0.19 x 10-3 t + 0.13 m ⁰C Laboratory
/Customer
Premises 600 ⁰C to 630 ⁰C / RTD-
pt3916,100 Ω
0.42 x 10-3 t + 0.046 m ⁰C
-200 ⁰C to -80 ⁰C /
RTD-pt385,200Ω
0.23 x 10-3 t + 0.45 m ⁰C
-80 ⁰C to 0.003 ⁰C /
RTD-pt385,200 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.03 ⁰C to 100 ⁰C /
RTD-pt385,200 Ω
0.43 x 10-3 t + 3.46 m ⁰C
100 ⁰C to 260 ⁰C / RTD-
pt385,200 Ω
0.22 x 10-3 t + 0.37 m ⁰C
260 ⁰C to 300 ⁰C / RTD-
pt385,200 Ω
0.46 x 10-3 t + 0.093 m ⁰C
300 ⁰C to 400 ⁰C / RTD-
pt385,200 Ω
0.38 x 10-3 t + 0.093 m ⁰C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 28 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
400 ⁰C to 600 ⁰C / RTD-
pt385,200 Ω
0.27 x 10-3 t + 0.093 m ⁰C Laboratory
/Customer
Premises 600 ⁰C to 630 ⁰C / RTD-
pt385,200 Ω
0.29 x 10-3 t + 0.067 m ⁰C
-200 ⁰C to -80 ⁰C /
RTD-pt385,500 Ω
0.23 x 10-3 t + 0.45 m ⁰C
-80 ⁰C to 0.003 ⁰C /
RTD-pt385,500 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.03 ⁰C to 100 ⁰C /
RTD-pt385,500 Ω
0.55 x 10-3 t + 0.3.45 m ⁰C
100 ⁰C to 260 ⁰C / RTD-
pt385,500 Ω
0.27 x 10-3 t + 0.31 m ⁰C
260 ⁰C to 300 ⁰C / RTD-
pt385,500 Ω
0.31 x 10-3 t + 0.14 m ⁰C
300 ⁰C to 400 ⁰C / RTD-
pt385,500 Ω
0.23 x 10-3 t + 0.15 m ⁰C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 29 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
400 ⁰C to 600 ⁰C / RTD-
pt385,500 Ω
0.17 x 10-3 t + 0.14 m ⁰C Laboratory
/Customer
Premises 600 ⁰C to 630 ⁰C / RTD-
pt385,500 Ω
0.20 x 10-3 t + 0.097 m ⁰C
-200 ⁰C to -80 ⁰C /
RTD-pt385,1000 Ω
0.17 x 10-3 t + 0.60 m ⁰C
-80 ⁰C to 0.003 ⁰C /
RTD-pt385,1000 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.03 ⁰C to 100 ⁰C /
RTD-pt385,1000 Ω
0.43 x 10-3 t + 3.46 m ⁰C
100 ⁰C to 260 ⁰C / RTD-
pt385,1000 Ω
0.22 x 10-3 t + 0.37 m ⁰C
260 ⁰C to 300 ⁰C / RTD-
pt385,1000 Ω
0.23 x 10-3 t + 0.19 m ⁰C
300 ⁰C to 400 ⁰C / RTD-
pt385,1000 Ω
0.20 x 10-3 t + 0.17 m ⁰C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 30 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
RTD - simulation Simulation Method using
Fluke 5522A
t= Temperature
400 ⁰C to 600 ⁰C / RTD-
pt385,1000 Ω
0.14 x 10-3 t + 0.19 m ⁰C Laboratory
/Customer
Premises 600 ⁰C to 630 ⁰C / RTD-
pt385,1000 Ω
0.42 x 10-3 t + 0.046 m ⁰C
-80 ⁰C to 0.003 ⁰C /
RTD-ptNi385,120 Ω
1.16 x 10-0 t + 3.47 m ⁰C
0.003 ⁰C to 100 ⁰C /
RTDptNi385,120 Ω
0.89 x 10-3 t + 3.47 m ⁰C
100 ⁰C to 260 ⁰C / RTD-
ptNi385,120 Ω
0.62 x 10-3 t + 0.13 m ⁰C
-100 ⁰C to 260 ⁰C /
RTD-Cu427,10 Ω
0.60 x 10-3 t + 0.19 m ⁰C
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
600 C to 800 C / B-
T/C
0.64 x 10-3 t + 0.31 m C Laboratory
/Customer
Premises 800 C to 1000 C / B-
T/C
0.39 x 10-3 t + 0.38 m C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 31 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
1000 C to 1550 C / B-
T/C
0.22 x 10-3 t + 0.49 m C Laboratory
/Customer
Premises 1550 C to 1820 C / B-
T/C
0.21 x 10-3 t + 0.38 m C
0.01 C to 150 C / C-
T/C
2.24 x 10-3 t + 12 m C
150 C to 650 C / C-
T/C
0.46 x 10-3 t + 1.18 m C
650 C to 1000 C / C-
T/C
0.36 x 10-3 t + 0.47 m C
1000 C to 1800 C / C-
T/C
0.32 x 10-3 t + 0.32 m C
1800 C to 2316 C / C-
T/C
0.42 x 10-3 t + 0.16 m C
-250 C to -100 C / E-
T/C
2.31 x 10-3 t + 0.40 m C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 32 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
-100 C to -25 C / E-
T/C
1.84 x 10-3 t + 1.78 m C Laboratory
/Customer
Premises -25 C to 350 C / E-
T/C
0.56 x 10-3 t + 34.70 m C
350 C to 650 C / E-
T/C
0.28 x 10-3 t + 1.03 m C
650 C to 1000 C / E-
T/C
0.24 x 10-3 t + 0.70 m C
-210 C to -100 C / J-
T/C
1.48 x 10-3 t + 0.66 m C
-100 C to -30 C / J-
T/C
1.84 x 10-3 t + 1.55 m C
-30 C to 150 C / J-T/C 1.05 x 10-3 t + 79.1 m C
150 C to 760 C / J-
T/C
0.26 x 10-3 t + 2.01 m C
760 C to 1200 C / J-
T/C
0.22 x 10-3 t + 0.65 m C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 33 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
-200 C to -100 C / K-
T/C
1.91 x 10-3 t + 0.53 m C Laboratory
/Customer
Premises -100 C to -25 C / K-
T/C
2.07 x 10-3 t + 1.59 m C
-25 C to 120 C / K-
T/C
1.29 x 10-3 t + 88 m C
120 C to 1000 C / K-
T/C
0.30 x 10-3 t + 2.02 m C
1000 C to 1372 C /
K-T/C
0.34 x 10-3 t + 0.34 m C
-200 C to -100 C / L-
T/C
2.14 x 10-3 t + 0.47 m C
-100 C to 0.01 C / L-
T/C
1.16 x 10-0 t + 11.6 m C
0.01 C to 800 C / L-
T/C
0.36 x 10-3 t + 11.6 m C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 34 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
800 C to 900 C / L-
T/C
0.22 x 10-3 t + 0.72 m C Laboratory
/Customer
Premises -200 C to -100 C /
N-T/C
2.31 x 10-3 t + 0.43 m C
-100 C to -25 C / N-
T/C
2.53 x 10-3 t + 1.3 m C
-25 C to 120 C / N-
T/C
1.44 x 10-3 t + 96.8 m C
120 C to 410 C / N-
T/C
0.51 x 10-3 t +1.40 m C
410 C to 1300 C / N-
T/C
0.24 x 10-3 t + 0.89 m C
0.01 C to 250 C / R-
T/C
2.59 x 10-3 t +11.50 m C
250 C to 400 C / R-
T/C
1.01 x 10-3 t + 0.43 m C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 35 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
400 C to 1000 C / R-
T/C
0.38 x 10-3 t + 0.61 m C Laboratory
/Customer
Premises 1000 C to 1767 C /
R-T/C
0.26 x 10-3 t + 0.40 m C
0.01 C to 250 C / S-
T/C
2.13 x 10-3 t + 11.5 m C
250 C to 1000 C / S-
T/C
0.42 x 10-3 t + 0.80 m C
1000 C to 1400 C /
S-T/C
0.31 x 10-3 t + 0.38 m C
1400 C to 1767 C /
S-T/C
0.30 x 10-3 t + 0.29 m C
-250 C to -150 C / T-
T/C
2.91 x 10-3 t + 0.25 m C
-150 C to 0.01 C / T-
T/C
1.16 x 10-0 t + 11.6 m C
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 36 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Thermocouple Simulation Method using
Fluke 5522A
t= Temperature
0.01 C to 120 C / T-
T/C
1.45 x 10-3 t + 11.5 m C Laboratory
/Customer
Premises 120 C to 400 C / T-
T/C
0.40 x 10-3 t + 1.76 m C
-200 C to 0.01 C / U-
T/C
1.16 x 10-0 t + 11.6 m C
0.01 C to 600 C / U-
T/C
0.50 x 10-3 t + 11.6 m C
DC Power Direct Method using Fluke
5522A
with PQ Option
P= Power
33 mV / 0.33 mA 0.8 x 10-3 P Laboratory
/Customer
Premises
33 mV / 330 mA 0.7 x 10-3 P 1020 V / 0.33mA 0.6 x 10-3 P 1020 V / 330 mA 0.5 x 10-3 P 33 mV / 0.33 A 0.7 x 10-3 P
33 mV / 3 A 0.1 x 10-3 P 1020 V / 0.33 A 0.5 x 10-3 P
1020 V / 3 A 0.1 x 10-3 P
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 37 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
DC Power Direct Method using Fluke
5522A
with PQ Option
P= Power
33 mV / 3 A 2.6 x 10-3 P Laboratory
/Customer
Premises
33 mV / 20.5 A 1.3 x 10-3 P 1020 V / 3 A 0.7 x 10-3 P
1020 V / 20.5 A 0.9 x 10-3 P
AC Power Direct Method using Fluke
5522A
with PQ Option
P= Power
45 Hz to 65 Hz @ PF=1 Laboratory
/Customer
Premises 33 mV / 3.3 mA 8.6 x 10-3 P
33 mV / 329.99 mA 0.02 x 10-3 P 329.999 mV / 3.3 mA 0.02 x 10-3 P 329.999 mV / 329.99
mA 0.01 x 10-3 P
330 mV / 3.3 mA 2.1 x 10-3 P 330 mV / 329.99 mA 0.01 x 10-3 P
1020 V / 3.3 mA 0.5 x 10-3 P 1020 V / 329.99 mA 0.01 x 10-3 P
33 mV / 0.33 A 0.6 x 10-3 P 33 mV / 20.5 A 0.2 x 10-3 P
329.999 mV / 0.33 A 0.01 x 10-3 P
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 38 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
AC Power Direct Method using Fluke
5522A
with PQ Option
P= Power
329.999 mV / 20.5 A 0.05 x 10-3 P Laboratory
/Customer
Premises
330 mV / 0.33 A 0.4 x 10-3 P 330 mV / 20.5 A 0.08 x 10-3 P 1020 V / 0.33 A 0.21 x 10-3 P 1020 V/ 20.5 A 0.20 x 10-3 P
Resistance Meters
(micro, milli, ohm,
kohm, Meg ohm)
Direct Method using
Decade Resistance Boxes
R= Resistance
Up to 200 μΩ 1.6 x 10-3 R Laboratory
/Customer
Premises
>200 to 400 μΩ 1.5 x 10-3 R >400 μΩ to 600 mΩ 1.5 x 10-3 R
>600 mΩ to 600 Ω 1.9 x 10-3 R >600 Ω to 100 kΩ 2.1 x 10-3 R >100 kΩ to 300 kΩ 1.9 x 10-3 R >300 kΩ to 600 kΩ 1.9 x 10-3 R >600 kΩ to 3 MΩ 14 x 10-3 R
Insulation Resistance
Tester
Direct Method using
Decade Meg ohm Box
R= Insulation Resistance
Up to 9.99 MΩ 14 x 10-3 R Laboratory
/Customer
Premises
10 MΩ to 99.9 MΩ 12 x 10-3 R 100 MΩ to 1000 MxΩ 12 x 10-3 R
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Electrical Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 39 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)
Oscilloscope Direct Method using Fluke
5522A with SC1100
Option
Relative Deviation of the
vertical Axis
U= Voltage
6.6 V/ 50 Ω load 2.8 x 10-3 U Laboratory
/Customer
Premises
130 V/ 1 MΩ load 2.8 x 10-3 U
Oscilloscope Band Width
f= Frequency
Up to 1.1 GHz 14 x 10-3 f
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Pressure Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 3 DAC-70-02/F34-B, rev. b (24-11-2016)
Pneumatic Pressure
Calibration of Digital &
Analogue Pressure
Gauges
GTS-WP-02
Based on the requirements of
BS EN 837-1 : 1998 Using
Druck DPI 610, DPI 620
up to 20 bar
0.02%
Laboratory
/ Customer
Premises
Pneumatic Pressure
Transducers,
Transmitters and
Switches
GTS-WP-02
Using Druck DPI 610, DPI 104
and Fluke 8846 multimeter
Up to 100 bar 0.11 %
Hydraulic Pressure
Transducers,
Transmitters and
Switches
GTS-WP-02
Using Budenberg DWT 580HX
Piston Cylinder 030L & Fluke
8846 multimeter
Up to 1200 bar 0.11 % Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Pressure Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 2 of 3 DAC-70-02/F34-B, rev. b (24-11-2016)
Hydraulic pressure
Digital & Analogue
Pressure Gauges &
pressure modules
GTS-WP-02
Using Budenberg DWT 580HX
Piston Cylinder 030L
Up to 1200 bar 0.02 % Laboratory
Vacuum gauge
calibration
GTS-WP-03
Based on the requirements of
BS EN 837-1 : 1998 and
ISO/TS 3567
Using Druck DPI 610, DPI 620
– 1 to 0 bar 1 mbar Laboratory
/ Customer
Premises
Calibration of hydraulic
pressure balance
GTS-WP-143 based on OIML
R110 and EURAMET cg-3
Version 1.0 (03/2011)
Up to 1400 bar 100 ppm
FSD
Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Pressure Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 3 of 3 DAC-70-02/F34-B, rev. b (24-11-2016)
Calibration of pneumatic
pressure balance
GTS-WP-143 based on
OIML R110 and EURAMET
cg-3 Version 1.0 (03/2011)
-1 to 25 bar 100 ppm
FSD Laboratory
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Volume Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 1 DAC-70-02/F34-B, rev. b (24-11-2016)
Pipette
Fixed and Variable
Volume
Gravimetric Method Up to 100 μl 0.57μl Laboratory
100 to 2000 μl 0.57 μl to 1.30
μl
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Force Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 1 DAC-70-02/F34-B, rev. b (24-11-2016)
Force
Verification /Calibration
of Compression testing
machines
GTS-WP-06
based on BS EN ISO 7500-1
3000 kN down to
300 kN
0.24%
of indicating
reading using
force transducer
class 1, ISO 376
Customer
Premises
Force
Verification /Calibration
of Compression testing
machines
GTS-WP-06
based on BS EN ISO 7500-1
3000 kN down to
200 kN
0.45%
of indicating
reading using
force transducer
class 2, ISO 376
Customer
Premises
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Mass Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 1 of 2 DAC-70-02/F34-B, rev. b (24-11-2016)
concrete and asphalt
batching plants (Hopper
Scale)
Hopper Scale calibration of
concrete batching plants &
Euramet CG-18,
ASTM C94/C94M
& NIST Handbook 44
5000 kg 0.05% Customer
Premises
Class M masses Masses according to GTS-WP-
17
Up to 200 g ±0.50 mg Laboratory
1 kg ±2.3 mg
2kg ±9.7mg
5 kg ±12 mg
10 kg ±155 mg
20 kg ±170 mg
DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN
DDEEPPAARRTTMMEENNTT
اعتماد تقييم المطابقة إدارة
SCOPE OF ACCREDITATION
Mass Calibration
General Const. Lab Calibration LLC
Industrial Area # 3, Sharjah- United Arab Emirates
Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL
DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR
THE SCOPE OF ACCREDITATION
Calibration Field/
Measured Quantity
Calibration Method Range and
Specification
Calibration
Measurement
Capability
(CMC)*
Location
Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of
accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC
due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381
Email: [email protected] web site: http://www.dac.gov.ae
Page 2 of 2 DAC-70-02/F34-B, rev. b (24-11-2016)
Top loading direct
reading weighing
balance
GTS-WP-01
Based on the requirements of
ASTM E 898 : 1988
(re approved 2005)
& Euramet CG-18
Calibrated weights
E1 (1 mg to 100g), E2(1 mg to
5 kg),
F1(1 mg to 5 kg),
F2 (10 kg & 20 kg) and
M1 (5 kg to 530 kg)
Up to 100 g 0.0001 g Laboratory/
Customer
Premises Up to 210 g 0.0002 g
Up to 0.5 kg 0.0006 g
Up to 1 kg 0.001 g
Up to 5 kg 0.009 g
Up to 10 kg 0.013 g
Up to 30 kg 0.23 g
Up to 100 kg 0.45 g