sl7000 series universal input data logger 1, 2, 3, 4, … · the sl7000 is a data logger that has a...

8
UNIVERSAL DATA LOGGER SL7000 SERIES 16 Bit Resolution Universal Input Data Logger 1, 2, 3, 4, 5, 6, 7, 8 and 9 Channels LCD Display Option User Replaceable Battery Typical 3 Years battery Life INTRODUCTION The SL7000 is a data logger that has a very flexible combination of inputs from a single temperature input up to 4 universal inputs (RTD, Thermocouple, mA, V, mV, Thermistor, Slidewire) combined with an internal temperature and RH as well as a optional external temperature and RH sensor. All units feature a digital input which can be used as an event trigger or as a counter input enabling counts of up to 65535 per period to be recorded. Powered by its own, user replaceable, internal battery that has a typical life in excess of 3 years, up to 128,000 readings can be recorded at intervals from 1 second to 24 hours. www. isi-be .com The SL7000 features unprecedented accuracy, combining our extensive knowledge of instrumentation precision with data logging technology. High end resolution of 16 Bit for all universal inputs and a modular build ensures that the unit is competitively priced whatever the application. An on-board LCD display (Optional) shows selected channel information and integral LED displays provide immediate scanning and alarm status information. The logger can be set to start immediate, start on an event (either a trigger event or the exercising of an external digital input) or start with the application of an external magnet (Provided). This feature enables loggers to be set up in advance of the mission by a supervisor and to be launched at the start of the journey / log period by an operator. The SL7000 series data loggers are connected to the computer via a USB cable (provided). It is possible to network up to 64 devices. Data are downloaded to a PC computer using the free TempIT-Lite software via its USB interface. TempIT-Lite enables all parameters to be set up and the logger issued, it also provides for data download, data storage and graphing. TempIT-Lite can be upgraded to TempIT-Pro quickly and easily over the telephone to provide additional features such as multi-logger trace overlays, data table display, data export, emailing of graphs, automatic MKT calculation etc. ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC) Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

Upload: dangthuy

Post on 19-Sep-2018

224 views

Category:

Documents


0 download

TRANSCRIPT

UNIVERSAL DATA LOGGERSL7000 SERIES

16 Bit Resolution

Universal Input Data Logger

1, 2, 3, 4, 5, 6, 7, 8 and 9 Channels

LCD Display Option

User Replaceable Battery

Typical 3 Years battery Life

INTRODUCTION

The SL7000 is a data logger that has a very flexible combination ofinputs from a single temperature input up to 4 universal inputs (RTD,Thermocouple, mA, V, mV, Thermistor, Slidewire) combined with aninternal temperature and RH as well as a optional external temperature and RH sensor. All units feature a digital input whichcan be used as an event trigger or as a counter input enabling countsof up to 65535 per period to be recorded.

Powered by its own, user replaceable, internal battery that has atypical life in excess of 3 years, up to 128,000 readings can berecorded at intervals from 1 second to 24 hours.

w w w. i s i - b e . c o m

The SL7000 features unprecedented accuracy, combining our extensive knowledge of instrumentation precision withdata logging technology. High end resolution of 16 Bit for all universal inputs and a modular build ensures that theunit is competitively priced whatever the application.

An on-board LCD display (Optional) shows selected channel information and integral LED displays provide immediate scanning and alarm status information.

The logger can be set to start immediate, start on an event (either a trigger event or the exercising of an externaldigital input) or start with the application of an external magnet (Provided). This feature enables loggers to be setup in advance of the mission by a supervisor and to be launched at the start of the journey / log period by an operator.

The SL7000 series data loggers areconnected to the computer via a USBcable (provided). It is possible to network up to 64 devices.

Data are downloaded to a PC computer using the free TempIT-Litesoftware via its USB interface.TempIT-Lite enables all parameters tobe set up and the logger issued, italso provides for data download, datastorage and graphing.

TempIT-Lite can be upgraded toTempIT-Pro quickly and easily over thetelephone to provide additional features such as multi-logger traceoverlays, data table display, data export, emailing of graphs, automaticMKT calculation etc.

ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC)Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

Every effort has been taken to ensure the accuracy of this specification, however we do not accept responsibility for damage, injury, loss or expense resultingfrom errors and omissions, and we reserve the right of amendment without notice.

SL7000 TECHNICAL SPECIFICATIONS

Description

Input Variations

Universal Input (maximum 4)Any Combination

Measurement Resolution

Counter Input

Start Options

Channel Indicators

Communication Indicator

Sample Rate

Display

Communication Interface

Universal Input Connector

External Temperature & RH Connector

Sample Methods

Cold Junction Compensation

Detail

Internal Temperature and counterInternal Temperature + RH and counterInternal Temperature, 2 Channel Universal and counterInternal Temperature + RH, 2 Channel Universal and counterInternal Temperature, 4 Channel Universal and counterInternal Temperature, RH, 4 Channel Universal and counter+Additional External Temp and RH option on all versions

Pt 100: 2 or 3 Wire °C or °F (BS EN 60751/JIS1604)Ni100, Ni120: 2 or 3 Wire °C or °FCu100: 2 or 3 Wire °C or °FThermocouples Types J, K, T, R, S, B, N °C or °FThermistor (Various Types) °C or °F(0 to 20) mA Scaled in engineering units±100mVdc Scaled in engineering units(0 to 1)Vdc Scaled in engineering units(0 to 5) Vdc Scaled in engineering units(0 to 10) Vdc Scaled in engineering unitsSlidewire Scaled in engineering units (100 ohm to 10Kohm)

All Universal Inputs - 16 Bit1

Internal & External Temperature - 12 BitInternal & External Relative Humidity - 8 Bit

Used to start logging or as a counter. Maximum Count = 65535per sample period. A current flowing into the counter pin >1 mA<10 mA or a Potential Free Contact (PFC) = logic '1'. Counts canbe actual counts or scaled in engineering units.Maximum frequency = 32768Hz.

Programmable Delay Period (10 minutes to 1 year)Start on external magnet swipeStart when counter input goes high Start on event trigger

8 Green LED's to indicate logging status changing to Red indicatealarm condition.

1 Yellow LED to indicate comms activity

1 Second to 24 hours for all channels

Optional 3.5 Digit LCD with channel indication and low batterywarning. Continuous display with value updated every 4 to 255seconds (programmable). Units displayable are °C, °F, mA, V, %

USB Isolated (mini-USB 1B Male connector) - Cable Supplied

14 Way 2 part plug/socket

6 Pin mini-Din

Point - The sampled measurement is stored directly to memory.Averaging - The average value per period is stored Maximum - The maximum value per period is storedMinimum - The maximum value per period is stored

Uses internal temperature sensor

Note 1 Sample Rate > 5 Minutes = 16bit, <5Minutes > 1 Minute = 15 Bit< 1 Minute> 8 seconds = 14 Bit, <=8 Seconds = 13 Bit

ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC)Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

Every effort has been taken to ensure the accuracy of this specification, however we do not accept responsibility for damage, injury, loss or expense resultingfrom errors and omissions, and we reserve the right of amendment without notice.

SL7000 Accuracy @ 25°C

Description

Internal Temp

External* Temp

Internal/External RH

Pt 100 2 or 3 Wire °C or °F

Ni100/Ni120*/Cu100*

Slidewire

0/20mA

±100mVdc

0/1Vdc

0/5Vdc

0/10Vdc

Thermocouples (J, K, T, R, S, N, B °C or °F)

Thermistor (Various Types, °C or °F)

Memory Size

Memory Utilisation

Battery

Battery Life

Ambient Operating Range

Manifest Text (User)

Manifest Text (Owner)

Alarm Setpoints

Alarm Trigger Time

Event Trigger

Low Battery Warning

Detail

± 0.5 °C (-30°C to +70°C)

± 0.4°C @ 25°C, ±1.75°C (-30°C to +70°C)

± 3.0%RH 20-80%RH (± 2%RH from 10%-90% to special order)± 5.0% RH 0-100%RH (± 4%RH to special order)

± 0.1°C + 0.1% rdg

± 0.2°C ± 0.1% rdg

± 0.02% FSD

± 0.01% FSD

± 0.01% FSD

± 0.01% FSD

± 0.02% FSD

± 0.01% FSD

± 0.1% FSD ± 0.5 °C

Type dependent contact sales office for details

0 or 2 Universal Inputs - 62,000 Readings4 Universal Inputs - 128,000 Readings

Wrap Around (First in, First Out) or Stop when Full (Default).

Lithium AA (3.6V DC) 2.7 Ah User replacable

Typ 3 Years (2700mAh)

-30°C to + 80°C

256 Characters

256 Characters

Independent High and Low, Latching per channel

Each Setpoint has an associated time. This time dictates theperiod the alarm must be continuously present before the alarmis raised. The options are 0 to 255 sample periods.

A programmable level per channel that will cause all channels tostart recording when the input first crosses the threshold in thespecified direction. An event can also be the exercising of anexternal digit signal or the activation of a magnetic swipe.

Low Battery displayed when battery life has less than one monthto go.

SL7000 General Specification

ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC)Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

SL7000 Order Codes

SL-ACC7000-01SL-ACC7000-02SL-ACC7000-03SL-ACC7000-04SL-ACC7000-05TempIT-LiteTempIT-ProTempIT-HS

External Temp/RH ProbeReplacement Miniature USB connecting leadReplacement Lithium Thionyl Chloride BatteryExternal power supply (cannot be used with ACC7000-01)Spare 14 way input connectorOperating Software (FOC) - requires version 4 or aboveUpgrade to professional versionHigh Security Version

Accessories

SL7000 Mechanical DetailsDimensions in mm

TempIT is freely available from www.signatrol.com

ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC)Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

*Notes:

1. Sample rate affects battery life as does high temperature exposure - see charts overleaf2. “Stop when full” is the only option for sample rates less than 1 minute3. Resolution is programmable as 8 or 11 bit. More data is stored in 8 bit mode and battery life is enhanced4. Application pending5. With ambient temperature less than 50°C

SINGLE CHANNEL TEMPERATURE DATA LOGGERSL51, SL52 & SL53 BUTTON LOGGERS

Miniature

Sample Rates from1 second to 24 hours

Monitor from -40°C to +125°C

Self Contained Temperature Data Logger

Waterproof To 1 Metre

Intrinsically Safe

INTRODUCTION

The SL51T, SL52T and SL53T are self contained single channel temperature data loggers with the ability to record up to 8,000 data points. They operate in a temperature range of -40°C to +125°C depending upon the model chosen. They arevery small (about the same size as a watch battery) which enables them to be inserted into small items and packages.

The SL50R Button Reader™ interfaces to the RS232 or USB port of a PC and is totallycompatible with the TempIT-lite operating software which enables the logger to beset up and the data read and displayed in graphical format. Data can be stored andretrieved at a later time. The software can be up-graded to TempIT-Pro which provides various additional features such as data table view, export data to spreadsheet, email graphs, snap to trace, and automated MKT Calculations etc.

SPECIFICATIONS

DescriptionSensor - Internal Temperature Sensor

Sensor Accuracy

Battery Life

Sample Rates (User Selectable)Memory Size

BatteryResolution

Response TimeSizeCase MaterialWeightSealingCommunicationClock AccuraySampling MethodsAlarmsStart DelayStart on AlarmOwner Manifest DataUser Manifest Data

SL51 SL52 & SL53-40°C to +85°C -40°C to +85°C - SL52

0°C to +125°C - SL531°C from -30°C to +70°C 0.5°C from -10°C to +65°C - SL52

0.5°C from +20°C to +75°C - SL531.3°C Outside this band See Chart

10 Years or 1 million Depends upon sample rate,samples*5 resolution and ambient

temperature*1

1 to 255 Minutes 1 second to 24 hours*1 & 2

2048 Readings 8192 - Low Resolution Mode4096 - High Resolution Mode

3.0V Lithium: Non Replaceable8 Bit (0.5°C) 8 Bit - Low Res Mode (0.5°C)

11 Bit - High Res Mode (0.07°C)Typically 6 seconds in liquid

17mm Diameter x 6mm HeightStainless Steel

4 gramsWater Resistant Waterproof to 1m

RS232 Serial or USB±2 Minutes per Month

Continuous (First in First Out) or Stop When Full*2

High and LowUp to 45 days, 1 Minute Resolution

No YesUp to 32 CharactersUp to 64 Characters

ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC)Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

The SL50-ACC01 is a water tight enclosure that allows the SL51T, SL52T and SL53T data loggers to be used in more hostile environments than the standard buttons. Constructed from 316 stainless steel the two part enclosure is sealed on an o-ring and provides greater protection in high pressure (up to 10 Bar), deep water (over 100m) or fluid monitoring as well as aggressive chemical applications

Computer Requirements

Button Interface with RS232 cableSL50R-232

Button Interface with USB cableSL50R-USB

Operating Software (Free from website)TempIT-Lite

ORDER CODE

ACCESSORIES

TempIT-PRO Standard optional operating Software

Every effort has been taken to ensure the accuracy of this specification, however we do not accept responsibility for damage, injury, loss or expense resultingfrom errors and omissions, and we reserve the right of amendment without notice.

Low Cost Data Logging button -40°C to +85°CSL51T

Starter kit containing five SL52T buttons, RS232interface cable and TempIT-Pro software

SL52-KIT-232

IBM PC or 100% compatible Pentium 75 (minimum) with at least one free serial port or USB interface, 16 MB of RAM memory, colour monitor & pointing device and SL50R Reader running Windows® versions 98, 2000, ME, NT & XP.

Software RequirementsTempIT-Lite ( free from Signatrol website: www.signatrol.com ) or TempIT-Pro

APPLICATIONSFood processing verification, pharma-

ceutical storage, laboratories, trans-

portation of temperature sensitive

goods, equipment run time, HVAC sys-

tem testing and balancing, etc.

Error CurveThe SL53 actual error depends upon the ambient temperature according to the following graph:

Data Logging button -40°C to +85°CSL52T

Starter kit containing five SL52T buttons, USBinterface cable and TempIT-Pro software

SL52-KIT-USB

SOFTWARE

Battery LifeBattery life is considerably shortened by fast logging and at temperatures in excess of 45 °C. The following graphs show their effecton the product life of the SL53. Because of this effect , the logger can only be used in 'Stop when Full' mode for sample intervals ofless than 1 Minute. Graphs shown are for high resolution mode, longer life is achieved using low resolution mode (contact sales officefor details.). For extreme worst case applications (1 second log interval @ 125 °C) battery life is restricted to 25 test cycles in high

resolution mode and 76 test cycles in low resolution mode.Increasing the sample time and reducing the temperatureexposure will increase life significantly.

SL53T Battery life, in days when fast logging. Graph shows days against measuring temperature (°C)

SL53T Battery life, in years when normal logging. Graph shows years against measuring temperature (°C).

Plastic Snap-in Fob - Pack of 5SL50-CLIP

Wall Mounting Bracket - Pack of 5SL50-WALL

Starter kit containing five SL51T buttons, RS232 interface cable and TempIT-Pro software

SL51-KIT-232

Starter kit containing five SL51T buttons, USBinterface cable and TempIT-Pro software

SL51-KIT-USB

Data Logging button 0°C to +125°CSL53T

Starter kit containing five SL53T buttons, RS232interface cable and TempIT-Pro software

SL53-KIT-232

Starter kit containing five SL53T buttons, USBinterface cable and TempIT-Pro software

SL53-KIT-USB

ISI sa-nv - Instrumentation for Science and Industry (INTERCONTINENTAL SERVICES INC)Rue du Doyenné 3 Dekenijstraat 1180 Brussels – Belgium - Tel 02/ 343 30 81 Fax 343 12 05 mail : [email protected] web : www.isi-be.com

Telephone: +32 2 343 30 81 - 3 RUE DU DOYENNE -1180 - BRUSSELS.

MINIATURE TEMPERATURE & RELATIVE HUMIDITY DATA LOGGERSL54TH BUTTON LOGGERS

Miniature

Sample Rates from1 second to 24 hours

Monitor from -20°C to +85°C

Temperature & Relative Humidity Data Logger

Self Powered

TempIT Compatible

INTRODUCTION

The SL54TH is our lowest cost temperature and relative humidity data logger. It is aself contained and self powered with the ability to record up to 8,000 data points.The SL54TH can operate in a temperature range of -20°C to +85°C and up to 100%RH.They are very small (about the same size as a watch battery) which enables them tobe inserted into small items and packages.

The SL50R Button Reader™ interfaces to the RS232 or USB port of a PC and is totally compatible with the user friendly, TempIT-Lite operating software whichenables the logger to be set up and the data read and displayed in graphical format. Data can be stored and retrieved at a later time. The software can be up-graded to TempIT-Pro which provides various additional features such as datatable view, export data to spreadsheet, email graphs and automated MKTCalculations etc. The TempIT software also automatically calculates dew point fromthe measured temperature and relative humidity readings.

Starter kits are available. The Starter kits contain up to five data logging buttons, thecommunication interface, a single plastic holder and a copy of the TempIT-Pro offering savings if all items were to be purchased separately.

w w w. s i g n a t r o l . c o m

SPECIFICATIONS

Specification SL54THPower Internal SourceBattery Life Dependant upon sample rate and temperature.

Typically 5 years at 10 minute sample rate.

Temperature Range Storage: -40°C to +85°C (-40°F to +185°F)Operating: -20°C to +85°C (-4°F to +185°F)

Temperature Accuracy ±0.5°C from -10°C to +70°CSee graph overleaf outside of his range

Temperature Resolution Low Resolution Mode: 0.5°CHigh Resolution Mode: 0.07°C

Relative Humidity Accuracy 5%RHRelative Humidity Resolution Low Resolution Mode: 0.64%

High Resolution Mode: 0.04%Number of Readings Low Resolution Mode: 8192 (4096 temperature, 4096 RH)

High Resolution Mode: 4096 (2048 temperature, 2048 RH)

Delayed Start Yes, up to 1 year with 1 minute resolution

Sample Rate 1 per second to 1per 24 hours

Software Alarms Yes

Material 305 Stainless Steel

Warranty 1 Year

Dimensions 17mm diameter x 6mm high

Shown with optional key fob holder

Instrumentation for Science and IndustryInstrumentation for Science and IndustryInstrumentation for Science and IndustryInstrumentation for Science and Industry ISI sa-nv Rue du Doyenné 3 - 1180 Brussels Tel +32 (0) 2 343 30 81 Fax 02/ 343 12 05 web : http:// www.isi-be.com mail : [email protected]

Computer Requirements

RS232 Interface cableSL50-INTERFACE-232

USB Interface cableSL50-INTERFACE-USB

Operating Software (Free from website)TempIT-Lite

ORDER CODE

ACCESSORIES

TempIT-PRO Standard Operating Software

Every effort has been taken to ensure the accuracy of this specification, however we do not accept responsibility for damage, injury, loss or expense resultingfrom errors and omissions, and we reserve the right of amendment without notice.

Data Logging button

SL54TH

IBM PC or 100% compatible Pentium 75 (minimum) with at least one free serial port or USB interface, 16 MB of RAM memory, colour monitor & pointing device and SL50R Reader running Windows® versions 98, 2000, ME, NT & XP.

Software RequirementsTempIT-Lite ( free from Signatrol website: www.signatrol.com ) or TempIT-Pro

APPLICATIONSFood processing verification,

pharmaceutical storage, laboratories,

transportation of temperature sensitive

goods, equipment run time, HVAC

system testing and balancing, etc.

Temperature Error CurveThe SL54TH actual error depends upon the ambient temperature according to the following graph:

SOFTWARE

Battery Life

Plastic Snap-in Fob - Pack of 5SL50-CLIP

Wall Mounting Bracket - Pack of 5SL50-WALL

STARTER KITS

SL54TH KIT- - -Interface:USB - USB Interface232 - RS232 Interface

Number of Data Loggers:1,2,3,4 or 5

Example: SL54TH-KIT-USB-5 is a starter kit containing 5SL54TH data loggers, a USB communication interface anda copy of the TempIT-Pro software.

Operating / Storage Temperatures

Instrumentation for Science and IndustryInstrumentation for Science and IndustryInstrumentation for Science and IndustryInstrumentation for Science and Industry ISI sa-nv Rue du Doyenné 3 - 1180 Brussels Tel +32 (0) 2 343 30 81 Fax 02/ 343 12 05 web : http:// www.isi-be.com mail : [email protected]