small steam sterilizer - zone medical · 2021. 6. 14. · i hӓtmed small steam sterilizer for...

40
Small Steam Sterilizer User Manual (Internal Water Tank) Q50B/Q70B 2460 HӓTmed (Wuxi) Co., Ltd V01 version

Upload: others

Post on 17-Jul-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

       

Small Steam Sterilizer 

User Manual                (Internal Water Tank) 

  

 

 

 

                                                                                                       Q50B/Q70B 

2460 

 

HӓTmed (Wuxi) Co., Ltd V01 version

Page 2: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

I

  

 

HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms 

to the YY/T 0646‐2015 standard and is a Class B sterilizer.Process evaluation system according to EN13060. 

  Do not  sterilize  any  liquid,Violating  the use of  this product  to 

sterilize liquids will damage the product and the loads. 

 

 

 

   : 

Symbols  Indication  Description 

 Warning 

Identifies  danger  points  or  conditions  and 

the  correct  procedures  that  must  be 

followed to avoid injury. 

 Hot Surfaces  Hot items, take care to avoid burns. 

 Caution 

Identifies  actions  that  could  damage  the 

device/component. 

 Precautions  Draw user's attention to important matters. 

        Symbols on products: 

 

Symbols  Description 

 take care to avoid burns 

 Manufacturer of medical equipment 

 Date of manufacture of this device 

 Serial number 

Symbols description 

Application 

Page 3: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

II

 Name and address of authorized agent of this product in the European Union. 

  Service should be performed by qualified technicians only. Observe all local 

and national directives related to service and safety standards. 

 

The function and life span of this product are directly related to whether the user follows the operating specifications and maintenance requirements in the installation and use manual. 

 

Manufacturer confirms that this product complies with the European Union 

Medical Device Specification. 

 Draw user's attention to important matters. 

 

 

 

 

 

Identifies danger points or  conditions  and  the correct procedures  that must be  followed  to 

avoid injury.  

Liquid sterilization  Do not sterilize any liquid 

Professionals  The operator who use sterilizers and fabric must be got a trained professional. 

Power cord and power plug 

Do not replace the power cable or power plug.Do not use a damaged power cable or 

power plug. 

Do not use this product if the power cable or power plug is damaged.The power cable 

or power plug must be replaced by a professional authorized by the manufacturer of 

this product. 

Do not use force on the power cable,be sure to operate the plug itself when plugging 

and unplugging. 

Make sure the power cable is not knotted。 

Do not place the power cable near a heat source. 

Do not touch the power cable with a sharp object. 

Installation, setup and 

commissioning 

Installation,  setup  and  commissioning  should  be  performed  by  a  professional 

authorized by the manufacturer of this product. 

Safety instruction 

Page 4: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

III

Do not use or operate this product at sites where there is a risk of explosion. 

The electrical connection of  this product and the connection of  the  inlet and outlet 

pipes should be carried out by professionals. 

This product should not be in direct contact with the patient. Place this product outside 

the range of at least 1.5 meters (radius) in the treatment area. 

If an external printer is configured for this product, the printer should not be placed in 

a location where there is a risk of contact with liquid. 

Emergency door 

opening device  Do not use this device to open the door under normal conditions. 

Sewage pretreatment 

and sterilization 

Pretreatment and sterilization of instruments and fabrics should follow the regulations 

of the instrument and fabric manufacturers. 

Strictly  follow  the  relevant  standards  for  disinfection  and  sterilization,e.g.WS506‐

2016,WS310‐2016. 

Use packaging materials or containers that have been confirmed by their manufacturer 

for sterilization. 

Sterilization termination 

This product terminates abnormally at certain stages of the program,high temperature 

steam will emerge from the chamber after opening the door. 

The loaded item may not be sterilized when the sterilization is abnormally terminated. 

Please note  the  status of  the program showed on  the display when  it  terminates.If 

necessary, the loaded items should be repackaged and sterilized. 

Remove the loaded item 

Do not use strong door opening. 

Use the tray handle to take out the tray. Do not touch the sterilized items with bare 

hand,including chamber and  door cover. 

Check if the loaded items is packaged completely after sterilization.If the package is 

damaged,please pack instrument or fabric again. 

Delivery and transportation 

At least 2 people need to move or deliver this product. 

Use proper belts. 

The  shipping  box  provided  by  the  manufacturer  of  this  product  must  be  used  for 

transportation. 

Sterilization load  The load must be sterilized with allowed steam by the manufacturer before it can be 

sterilized. 

maintenance  

Maintenance of this product must be performed by a professional authorized by the 

manufacturer. 

Please maintain the product at the time intervals specified in this manual. 

Must use the accessories supplied by the manufacturer of this product. 

Page 5: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

IV

malfunction 

If the same fault message appears repeatedly in this product, please turn off the power. 

Repair must be performed by a professional authorized by the manufacturer of this 

product. 

 

 

 

 

 

 

Name of sterilization Standard 

134 Standard 

121 Quick B 

Quick S 

134+  Fabric 

Class of sterilization  B  B  B  S  B  B 

Dynamic pressure monitoring  √  √  √  √  √  √ 

Air leak  √  √  √  √  √  √ 

No load  √  √  √  √  √  √ 

Solid load  √  √  √  √  √   

Small amount,porous,permeable mixed load 

√  √      √   

Small amount,porous,permeable  √  √      √  √ 

Porous,permeable full load  √  √      √  √ 

B hollow load  √  √  √  √  √   

A hollow load  √  √  √    √   

Simple packaging  √  √  √  √  √  √ 

Composite packaging  √  √  √    √  √ 

Dry,solid load  √  √  √  √  √   

Dry,porous permeable load  √  √      √  √ 

   

Sterilization Program Introduction

Page 6: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

V

TABLE OF CONTENTS 

1  Device Description ........................................................................................................................................ 1 1.1  Delivery list ........................................................................................................................................... 1 

1.1.1  Standard delivery list ..................................................................................................................... 1 1.2  Product view ......................................................................................................................................... 1 

1.2.1  Front view  ..................................................................................................................................... 2 1.2.2  Front view  ..................................................................................................................................... 2 1.2.3  Side view ....................................................................................................................................... 3 1.2.4  Rear view ....................................................................................................................................... 3 1.2.5  Inside view ..................................................................................................................................... 4 1.2.6  Top view ........................................................................................................................................ 4 

1.3  Technical Parameters ............................................................................................................................ 4 1.4  Loading method .................................................................................................................................... 5 1.5  Intelligent controller and control panel .................................................................................................. 5 

1.5.1  Operation buttons ......................................................................................................................... 6 1.6  Emergency door opening device ............................................................................................................ 6 1.7  Equipment workflow ............................................................................................................................. 6 

1.7.1  Sterilization program ..................................................................................................................... 6 1.7.2  Vacuum test ................................................................................................................................... 7 

2  Installation and handling instructions ........................................................................................................... 7 2.1  Unpacking precautions .......................................................................................................................... 7 2.2  Installation site requirements ................................................................................................................ 8 2.3  Device height adjustment ...................................................................................................................... 8 2.4  Connection requirements ...................................................................................................................... 9 

2.4.1  Electrical connections ..................................................................................................................... 9 2.4.2  Distilled water connection ............................................................................................................. 9 2.4.3  Wastewater connection ............................................................................................................... 10 

2.5  Installation service list ......................................................................................................................... 10 2.6  Handling instructions ........................................................................................................................... 10 

3  First time operation .................................................................................................................................... 11 3.1  Startup ................................................................................................................................................ 11 3.2  After startup ....................................................................................................................................... 11 3.3  Open and close the door ...................................................................................................................... 11 3.4  Distilled water injection ....................................................................................................................... 12 3.5  Drainage ............................................................................................................................................. 12 3.6  Sterilization effect verification ............................................................................................................. 13 

4  Sterilization program .................................................................................................................................. 13 4.1  Loading items to be sterilized .............................................................................................................. 13 4.2  Select sterilization program ................................................................................................................. 13 4.3  Auxiliary option ................................................................................................................................... 14 

4.3.1  Warm‐up setting .......................................................................................................................... 14 4.3.2  Automatic printing ....................................................................................................................... 15 4.3.3  Extra drying ................................................................................................................................. 15 

4.4  Running a sterilization program ........................................................................................................... 15 4.4.1  Start sterilization ......................................................................................................................... 15 4.4.2  Program running process ............................................................................................................. 15 

4.5  Manual stop ........................................................................................................................................ 15 4.6  Sterilization process completed ........................................................................................................... 17 4.7  Sterile item storage ............................................................................................................................. 18 

5  Functional testing and other procedures ..................................................................................................... 19 5.1  Vacuum test ........................................................................................................................................ 19 5.2  B&D test.............................................................................................................................................. 20 

Page 7: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

VI

5.3  Helix test ............................................................................................................................................. 20 5.4  Drainage program ................................................................................................................................ 21 

6  System settings ........................................................................................................................................... 21 6.1  Time setting ......................................................................................................................................... 21 6.2  Date setting ......................................................................................................................................... 22 6.3  Language settings ................................................................................................................................ 22 6.4  Warm‐up setting .................................................................................................................................. 23 6.5  Printing settings ................................................................................................................................... 23 6.6  Extra drying settings ............................................................................................................................. 24 6.7  Water quality testing ........................................................................................................................... 24 6.8  System reset ........................................................................................................................................ 25 6.9  System information .............................................................................................................................. 25 

7  Sterilization log ........................................................................................................................................... 25 7.1  Log export ............................................................................................................................................ 26 7.2  Log printing .......................................................................................................................................... 26 7.3  Log interpretation ................................................................................................................................ 26 

8  Small steam sterilizr maintenance ............................................................................................................... 27 8.1  Cleaning of items before sterilization ................................................................................................... 27 8.2  Maintenance ........................................................................................................................................ 27 

8.2.1  Cleaning and maintenance ............................................................................................................ 27 8.2.2  Regular maintenance .................................................................................................................... 29 

9  Failure and common problem ........................................................................................... 错误!未定义书签。 9.1  No information on the display .............................................................................................................. 29 9.2  Large consumption of distilled water .................................................................................................... 30 9.3  Bad drying effect .................................................................................................................................. 30 9.4  No distilled water ................................................................................................................................. 30 9.5  Unable to open the door ...................................................................................................................... 30 9.6  Leakage at the door or dripping underneath ......................................................................................... 31 

 Appendix 1  Safety device description ............................................................................................................. 31 

 Appendix 2  Schematic diagram of piping system  ............................................................................................ 31 

 Appendix 3  Electrical schematic diagram ........................................................................................................ 32 

Page 8: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 1 -

1 Device decription 

1.1 Delivery list 

1.1.1 Standard delivery list 

    The standard accessories sent with sterilizer are listed in the table below. Please check the list when unpacking. 

 

1.2 Product View 

1.2.1 Front View 

1.1.1 Standard delivery list 

Accessory name Accessory Picture 

Q’ty  Description 

Air level  

Air level is placed above the sterilizer during installation. Two foot pads at the front  is higher than the back end. The angle between the bottom of sterilizer and countertop is around 5‐10°. 

Tray  

3 Aluminum tray with holes at the bottom for placing sterilized items. 

Tray handling 

 

1  Tool to take out the tray  

Chamber rack  1  Rack to hold the tray  

Fuse  

2  220V/16A fuse 

Air filter 

 

1 Air filter to ensure that the air entering the sterilization chamber is highly filtered to  reduce bacteria and dust entry. 

Wastewater pipe 

 

2 Transparent silicone tube for draining waste water(Φ9*6 2M). 

Drain connector

(internal tank 

only)   2 

A joint connecting the drain pipe to the sterilizer wastewater interface 

Random Document 

Installation and manual 

 

1  Manual book 

Product Service Manual 

1 User guidance, installation service order, maintenance record 

Operation precautions  

1  common problems and precautions 

Page 9: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 2 -

        

Fig 1.1 front view 

 

1.2.2 Side view 

 

Fig 1.2 side view 

 

Door  handle 

Power switch 

Display 

Page 10: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 3 -

1.2.3 Rear view 

 

 

      Fig 1.4  Rear View 

 

 

 

1.2.4 Inside View 

 

 

Fig 1.5 inside view 

Wastewater  drainage interface A (empty)

Power Switch 

Printer interface 

Test interface 

USB interface 

Emergency door  opening device 

Air filter 

Power cable 

Safety valve 

Wastewater  drainage interface B (half drainage) 

Cooling fan 

Page 11: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 4 -

 

1.2.5 Top View 

 

Fig 1.6 Top view 

  

1.3 Technical parameters 

parameter name  Q50B  Q51B  Q70B  Q71B 

External size  570mm*450mm*470mm  680mm*450mm*470mm 

Packaging size  688mm*576mm*688mm  806mm*576mm*688mm 

Chamber size  Φ253mm*355mm  Φ253mm*475mm 

Capacity of Sterilizer  18 L  23 L 

Pressure vessel weight  4.8 kg  6.2 kg 

Equipment weight  59.6 kg  71.6 kg  64.8 kg  66.8 kg 

Built‐in clean water tank volume 

3.8L 

Built‐in waste water tank volume 

4.5L 

Software version  V00 

Pressure vessel technical indicators 

Working pressure is less than 0.25Mpa 

Operating temperature:115℃‐138℃ 

Working voltage  A.C 220V±22V, 50Hz±1Hz 

Maximum power  2000VA 

Fuse  16A 

Distilled water tank (Internal) 

Page 12: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 5 -

Working environment pressure 

70 Kpa ~ 106 Kpa 

Working temperature  5 oC ~ 40 oC 

Storage and transportation environment temperature 

‐20 oC ~ 55 oC 

environment humidity  ≤ 85% 

Maximum wastewater temperature 

100 oC 

Maximum noise  70 dB(A) 

Electromagnetic compatibility 

Compliance with GB/T18268.1‐2010; Works properly in its electromagnetic environment, but does not constitute an unacceptable electromagnetic 

harassment of anything in the environment。At the same time, it has the ability to resist electromagnetic interference. 

According to GB 4824‐2013,Electromagnetic radiation meets Group 1 Class B requirements. 

Standards compliant  YY/T 0646‐2015; GB4793.1‐2007; GB4793.4‐2001; 

Note 1:The maximum operating voltage range :198V ~ 242V Note 2: Anti‐electromagnetic interference is the ability of the sterilizer to work properly in the presence of electromagnetic interference 

   

 

1.4 Loading method 

The  tray  rack  is  included  in  the delivery of  sterilizer.  The  space  inside  the  tray  rack  is  a  space  that can be 

effectively sterilized, always use the tray rack when loading items to be sterilized. The tray rack can hold 5 trays or 

adapt to thickness of different loads or containers after rotating 90oC. 

 

Fig 1.6 Loading method 

1.5 Intelligent controller and control panel 

Intelligent controller is the core of small steam sterilizer, In addition to completing operations such as rapid 

acquisition,  processing,  fault  detection,  and  control  of  the  system's  real‐time  parameters,  the  controller  also 

intuitively plots the temperature and pressure versus time curve in the chamber. 

The intelligent controller is easy and efficient to operate with LCD screen with 4 buttons. 

Page 13: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 6 -

1.5.1 Operation buttons and Identification 

The use of 4 buttons in different situations is as follows: 

Button position

(Face up from left 

to right) 

1  2  3  4 

       

When the program is selected 

return Previous option 

Next option  Select 

standby mode      Open the door  Startup 

Program runtime        Stop 

When setting        Confirm 

 

There is an icon in the interface to display the door lock status. 

Status Identification  Description 

 Door lock status  The status of the door lock: on or off 

1.6 Emergency door opening device 

Emergency opening is an unconventional way to open the door when sterilizer cannot be powered on due to 

other  reasons or  the  safety door  failure.Please do not  open  the door  in  this way under normal  circumstances. 

Remove the screw on the emergency protection cover on the back of the product (Fig. 1.4),then using your right 

hand  to grasp  the white plastic disc  inside  the protective  cover,gently pull(distance  is  about 1CM).At  the  same 

time,use your left hand to lift the door handle up to open the door 

1.7 Equipment Workflow 

1.7.1 Sterilization program 

Sterilization process consists of 3 stages, pre‐vacuum, sterilization, drying. After the program is started, the 

user can see the stage of the program running on the interface. 

 

Process name  Description 

1.Pre‐vacuum 

During the stage of pre‐vacuum, the process of evacuating and injecting 

steam is proceeded repeatedly, As the mixture of air and steam is 

continuously extracted, the air in the chamber is minimized.The number of 

this process varies according to the selected program, standard sterilization 

procedure repeated 3 times, Helix program and 134+ program repeated 4 

times. 

2.Heating 

After the pre‐vacuum is the heating stage, steam is continuously injected 

into the chamber, the temperature and pressure in the chamber rise 

continuously until the specified value of the selected program is reached. 

Page 14: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 7 -

3.Holding 

After the temperature and pressure reached the set value, holding phase 

begins and maintains stable temperature and pressure for a period of time 

to ensure that the sterilization effect meets the requirements of the 

specification 

4.Depressurizatio

The pressure in the chamber is released after the holding time is over. At the 

same time, the steam in the steam generator is discharged. 

5.Drying 

After sterilization, the sterilized items are dried. Drying begins after pressure 

is released. Under the effect of a high efficiency air filter, the air in the 

chamber is always sterile during the drying process 

6.Intake 

The drying process is also the process of intake, make sure that the pressure 

in the chamber returns to environmental pressure at the end of the 

program. 

 

1.7.2 Vacuum test 

The purpose of the vacuum test  is to test the leak rate of the sterilizer. No sterilization of the items in the 

chamber during the process. Please run the vacuum test in a cold and dry chamber without loads. 

 

Process name  Description 

1.Vacuum The air in the chamber is evacuated till the pressure in the chamber reaches the set value of the vacuum test 

2.Balancing time After waiting for 5 minutes, the vacuum everywhere in the chamber tends to be uniform. 

3. Test holding time 

Test holding time is 10 minutes, The device continuously monitors the pressure in the chamber during the period 

4.End of test After the test is over, the screen displays the test results: vacuum and leak rate. 

  

2 Installation and handling instructions 

2.1 Unpacking precautions 

 

Installation, setup, and commissioning should be performed by a 

professional authorized by the manufacturer of this product. 

 

This product is taken out of the box and transported must be operated by 2 people.  

Total weight:Q50B 59.6 kg;       Q51B 61.6 kg;  

Q70B 64.8 kg;       Q71B 66.8 kg 

Page 15: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 8 -

 

Please  carefully  check  the  box  and  take  out  the  device.  If  the 

package is damaged or the equipment is damaged, please contact 

an authorized dealer or logistics company immediately. 

 

Please check the packing box and the random accessories inside 

the sterilizer. A list of standard accessories can be found in section 

1.1 of this manual. 

 

2.2 Installation site requirements 

This product should be placed in a dust‐free and dry environment。Surrounding humidity is  less than 85%, 

Temperature between 5 oC  and  40 oC. 

 

Do not use or operate this product at a site where there is a risk of 

explosion. 

 

Countertop must be able to support the weight of this product : 

 Q50B 59.6 kg;       Q51B 61.6 kg;  

Q70B 64.8 kg;       Q71B 66.8 kg 

Each surface of the device (except the bottom) must have a gap 

of 10 cm outward. Too small a gap will result in heat buildup, 

shorten the life of the equipment, reduce the performance of the 

equipment and extend the running time of the program. 

 

2.3 Device height adjustment 

Take out air level and put it in the horizontal position at the top of the product and adjust the two foot pads at 

the front end to make sure the front is higher than the rear side. The angle between the bottom of sterilizer and 

countertop is around 5‐10°. It is beneficial to realize the drying of the loads and saves distilled water. 

 

              Fig 2.1 Device height adjustment 

 

Rear end Front end 

Page 16: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 9 -

2.4 Connection requirements 

2.4.1 Electrical connections 

 

220V/50Hz power supply. The maximum power supply 

fluctuation range is 198V ~ 242V. 

Use a separate 16A power circuit. Provides 30mA leakage 

protection. 

Make sure the power cable is not knotted or pressed. 

Do not use force to pull the power cable, 

Do not place the power cable near a heat source. 

Do not touch the power cable with a sharp object. 

Do  not  replace  the  power  cable  or  power  plug.  Do  not  use  a 

damaged power cable or power plug 

Do  not  use  this  device  if  the  power  cable  or  power  plug  is 

damaged. The power cable or power plug must be replaced by a 

professional authorized by the manufacturer of this device. 

 

2.4.2 Distilled water injection 

Open up cover of the internal water tank and Inject the distilled water into the internal water tank through the 

water tank deflector, please be noted that the distilled water added should not exceed the height of the water 

deflector, generally stored in a standard 12L distilled water tank.Full tank of distilled water can complete up to 20 

sterilizations. Please pay attention to the amount of distilled water remaining  in the tank to avoid affecting the 

normal operation of this device. 

 

 

Compliance  with  YY/T  0646‐2015  high  quality  distilled  water  as 

specified in appendix B or demineralized water because the degree 

of formation of rust stains depends on whether the water quality of 

the distilled water is up to standard. 

Conductivity ≤20μs/cm; 

If the inlet pressure is >30 Kpa may cause excess distilled water to 

enter the chamber and affect the operation of the device. 

In order to prevent the distilled water from being contaminated, the 

lid of the distilled water tank must be tightened or tight at all times. 

For small steam sterilizer with internal water tank, please drain the 

waste water tank before injecting distilled water. 

Regularly clean the distilled water tank at least once a month. 

Page 17: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 10 -

This  product  has water  quality monitoring device,it will  not work 

properly when unqualified water and will give warning that water 

quality  cannot  reach  the  standard.When  you  see  this  message, 

please replace high quality distilled water in time.  

If  you  need  to  use  the  water  purifier,  please  consult  the 

manufacturer of this device first. 

It is best to tilt and secure all the water pipes connected to the back 

of the device are not bent or knotted. 

 

2.4.3 Wastewater connection 

For small steam sterilizers without internal water tank, take out the drain pipe from the accessory bag and insert the waste water connection to the back of the device (Figure 1.4). 

 

  It  is best  to tilt and secure all  the water pipes connected to  the 

back of the device are not bent or knotted. 

 

2.5 Installation service list 

 

After  installation  please  carefully  fill  out  the  installation  service 

form in the manual which is useful information for maintenance.  

 

2.6 Handling instructions 

This device has to be stored and transported at humidity less than 85% and temperature ‐20oC ~ 55oC. Before transportation, please carry out “drainage program” first and take off all pipes(including distilled 

water pipes and waste water pipes),if printer is connected,remove the printer cable.  

If the tray rack and tray are placed in the chamber during transportation. Before closing the door,place a foam 

pad  between  the  door  and  the  tray  holder  to  prevent  the  tray  frame  from  being  damaged  by  shaking  during 

transportation 

 

For long‐distance transportation, the product must be returned to the package in the order in which it was 

unpacked, and the package must be sealed. 

  Please carry drainage program before handling this device 

 

 

Page 18: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 11 -

 

3 First time operation 

3.1 Startup 

                 Turn on the power switch, the power switch position is in the lower right corner of the device inside view (Figure 1.4).                  

  Make sure to remove the latex plug from the air filter mounting 

hole at  the back of  the device and  install  the air  filter  from  the 

accessories included. 

 

3.2 After Startup 

After startup at the first time, please set up language and carry out pressure test. After the pressure verification 

is successful, the initial setting is completed. The date and time can also be set as required. See Chapter 6 system 

settings in this manual. 

Steps of pressure verification is as follows:  

 

 

 

 

 

 

 

 

 

   

 

 

 

 

3.3 Open/close the door 

 

 

Pressure verification Verification is checking … 

 P: 0.00 Bar 

Pressure verification 

Pressure verification 

successful 

Door  open  to  complete pressure verification 

press button and open up the door  

Wait  for calibration  and make  sure  the door is open. 

press  to select 

1–Sterilization program 2 – Test program 2‐  3 – Sytem setting 4 – File record 

Page 19: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 12 -

Door open up operation:Please press  button at standby mode,you will see door lock icon will change 

its status on the display to indicted the door is opened. At this time, open the door handle on the right side of the 

device upwards by about 90oC and then open the door outward. 

Closing operation:Close the door of sterilizer,The door handle is turned down to the end。 

 

When closing the door, please press the door handle down to the 

end. The strength required after using  for a period of  time may 

increase 

Don’t slam the door at any time. 

Failure  to  comply  with  the  above  requirements may  affect  the 

function and reliability of the door lock device 

 

Open the door only when loading and unloading, Keeping the door 

closed at other times is beneficial to energy saving 

 

3.4 Distilled water injection 

Inject qualified distilled water into the distilled water tank and make sure the water level is between the lowest 

mark (MIN) and the highest mark (MAX). Always check the water  level  in the water tank and replenish distilled 

water in time. 

 

 

Fig 3.1 Inject distilled water 

 

For the sterilizer with internal water tank, please discharge waste 

water tank every time before injecting distilled water.  

 

Page 20: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 13 -

3.5 Drainage 

       For sterilizer with internal water tank. After using it a few times please take out plastic drain connector and pipe from accessory bag when to discharge the waste water tank. Please insert one end of the drain connector without a black rubber seal into the drain pipe and insert the other into drainage interface (Figure 1.4 and Figure 1.5). 

 

For  sterilizer  with  internal  water  tank,  please  empty  the waste 

water tank before injecting the distilled water.  

3.6 Sterilization effect verification 

Small steam sterilizer installed or used for 12 months or after maintenance shall carry out B&D test without 

load and carry out physical, chemical and biological monitoring with full load.After monitoring is proved successful, 

the sterilizer can be used again. Verification of sterilization results shall comply with the requirements of Appendix 

E of WS506‐2016 

4 Sterilization program 

4.1 Loads to be sterilized 

 

 

Proper loading is a key factor for effective sterilization and drying.You 

must also pay attention to the following when loading 

The  load  must  be  sterilized  by  the  manufacturer  if  steam 

sterilization is permitted. 

Packaging  materials  for  packaged  sterilization  products  shall 

comply with GB/T 19974 standard. 

Sterilized items packed in paper and plastic bags, The paper must 

be faced up and laid flat in the tray. 

For mixed load for sterilization, fabric is on the top and the metal 

is on the lower layer. 

There must be a gap between the sterilization bags. 

Metal bags and dressing bags cannot be attached together. 

The load capacity must not exceed 90% of the chamber capacity. 

 

4.2 Select sterilization program 

This  product  has  7  preset  sterilization  programs.  Please  select  the  appropriate  sterilization  program 

according to the load type and packaging material. The parameter setting of the sterilization program and the typical 

load description are as follows. The maximum load of Q50B/Q51B is 6kg instruments or 2kg fabric.The maximum 

load of Q70/Q71B is 7kg instruments or 2.5kg fabric. 

  

Page 21: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 14 -

      

Table 4‐1 Sterilization program and parameter settings 

Name of 

Sterilization 

(℃) 

Holding 

time 

(Min) 

Drying 

time 

(min) 

Sterilization load 

121 program 

121  >15  1‐99 a  small  amount  of  fabric,Heat‐resistant instrument. 

134  program 

134  >3  1‐99 Mixed  load,long and hollow instruments 

134+ program 

134  >3  1‐99 Mixed  load,long and hollow instruments 

Quick S  134  >3  1‐99 Solid  instruments  and  hollow instruments. 

Quick B  134  >3  1‐99 Wrapped  &  unwrapped  solid instruments 

Fabric  121  >15  1‐99  Fabric 

 

Select the sterilization program as follows: 

 

 

 

 

 

 

 

 

 

 

Fig 4.1 Select sterilization program 

4.3 Auxiliary option 

4.3.1 Warm‐up setting 

The  automatic warm‐up  function  allows  the  sterilizer  to  reach  the  set  temperature before  the program  is 

started to significantly reduce the time required for each sterilization. See section 6.4 for details. 

1 – Sterilization program 2 – Test program 2‐  3 ‐ System settings  4 – File record  Press to select 

134 program 

121 program Quick B Quick S  Press to enter 

Page 22: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 15 -

 

The  chamber  will  automatically  heat  up  to  110  °C  when  the  automatic 

preheat setting is turned on. Please be careful of burns! 

 

4.3.2 Automatic printing 

Turn on the automatic printing function and the sterilizer is properly connected to the printer to automatically print the sterilization record after the program ends. See section 6.5 for details 

 

 

4.3.3 Extra drying 

Additional drying settings based on user preferences, extend the drying time of the preset sterilization program based on the drying time in Table 4‐1 to achieve good drying results for loads, especially cotton fabrics. See section 6.6 for details. 

 

 

4.4 Running a program 

4.4.1 Start sterilization 

After selecting sterilization program,pleaes press  to start up program.Sterilizer will ensure that the door 

is locked and the water quality of the distilled water is monitored after the start of the program. 

 

 

 

 

        

 

Fig 4.2 starting program 

 

4.4.2 Program running process 

Sterilization program includes 3 phases,pre‐vacuum,sterilization and drying. The description of each stage is detailed in section 1.7.1 of this manual. The controller monitors the operation of the sterilizer and the temperature and pressure in the chamber in real time and displays relevant information on the screen. 

 

 

07/25/2017       09:23:58  134 program 

134℃/2.1Bar /5Min 

T:  25.0℃     P: 0.00Bar Door status    

Press to start program 

Page 23: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 16 -

4.5 Manual stop 

A running program can be manually stopped at any stage. If the sterilization process is stopped before 

the start of drying, the loaded items are not sterilized. If the sterilization procedure is manually stopped after the 

start of drying,  the  loaded  items have been effectively sterilized, but not  fully dried.  In  the case of  fully  loaded 

packages, the possibility of wet bags is very high and these items cannot be stored because Because aseptic storage 

places high demands on the dryness of the packaged items. 

 

 After  the  program  is manually  stopped,  there may  be  high  temperature 

steam coming out and causing burns when the door is opened! 

Do not touch the product,  loaded items, chamber or door cover 

with bare hands. 

Be sure to use the tray handle to take the loaded items. 

 

Manually stopping the sterilization process at an earlier stage may cause an 

infection! 

If the sterilization process is stopped before the start of drying, the 

loaded  items  are  not  sterilized,The  use  of  these  items  may 

endanger the health of the operator and the patient. 

If necessary, the loaded items should be repackaged and sterilized. 

If you need to manually stop the current program, please  press  (Fig 4.3). After manual stop, the screen 

displays the program end interface and manually stops the fault information. The intelligent controller is processed 

according to the current situation to achieve a safe pressure inside the chamber and other parts of the device also 

ensure a safe state. After the device is in a safe state, the controller is allowed to operate the door open button, 

the required time and the program running state when manually stopped, usually takes several minutes. 

 

 

 

 

 

 

 

 

 

 

 

 

 

134 program 

P: 0 Bar     T:  25.0℃      Door status            00:00:00 

 

Press to stop the program 

Program stop 

   The  program  stops  in progress…         P: 0.85Bar 

Page 24: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 17 -

 

Fig 4.3 Manual stop 

 

4.6 Sterilization procedure completed 

When the sterilization process ends normally, the program end interface and program running success information are displayed on the screen. If the automatic printing function is set and the printer is connected, the 

product will automatically print the sterilization record. Please press  to open the door.  

 

 

Careful burns! 

Do  not  force  the  door  to  open.  Failure  to  do  so  may  result  in 

damage  to  the  product  and/or  hot  steam  coming  out  of  the 

chamber. 

After the sterilization is finished, wait for a while until the sterilizer 

is fully cooled before opening the door. 

Do not touch any metal surface. 

Do not touch the sterilized loaded items, chamber and door covers 

with bare hands to avoid being burned. 

Be sure to use the tray handle to remove the loaded items. 

Balance when removing  the tray  to prevent  the contents of  the 

tray from slipping. 

 

Damaged packaging devices or items may not be effectively sterilized, 

and exposure to  these  items may endanger  the health of  the operator or 

patient. 

If damaged packaging if found, please repackage the item.  

 

Waiting  for  5  minutes  after  sterilization  before  removing  the 

loaded items, which is good for drying the sterile items. 

The  11th  edition  (2017)  of  the  correct  handling  method  for 

medical  device  cleaning  and  disinfection  issued  by  the 

international authority AKI, Germany. It is recommended that the 

wet  bag  be  judged  as  having  water  drops  remaining  on  the 

package or in the package after 15 minutes of sterilization. 

Physical  parameters  should  be  monitored  for  each  sterilization 

cycle, record and check process variables. When the sterilization 

temperature set in this cycle is 121 ° C, the holding time should be 

no less than 15 minutes. When the sterilization temperature set in 

Page 25: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 18 -

the  cycle  is 134  ° C,  the holding  time  should be not  less  than 3 

minutes. 

Chemical monitoring  should be performed  for  each  sterilization 

cycle,by observing the color change of the chemical  indicator to 

determine  whether  the  sterilization  requirement  is  met 

compliance with  the  information and  requirements provided by 

the manufacturer of the chemical indicator. 

Sterilization  monitoring  shall  comply  with  the  requirements  of 

Annex E of WS506‐2016 

 

The national standards for sterilization,WS310‐2016 stipulate that 

the sterilizer operating cycle parameters should be fully recorded 

and stored for 3 years. 

 

4.7 Sterile Storage 

Sterile package storage time after sterilization varies with packaging material and storage conditions, in full 

compliance with the specification. For example, the condition of WS310‐2016 can be stored for up to 6 months. 

Packaging requirements and storage conditions please follow WS310‐2016. 

 

 

Please follow the packaging manufacturer's recommendations or 

regulations for the storage time of the sterile package. 

Do not store sterile items in contaminated areas. 

Sterile items must be stored in a dust‐free, (In addition to water, 

other  liquids  such  as  alcohol,  disinfectants,  etc.  should  also  be 

avoided) dry temperature suitable environment. 

 

 

 

 

 

 

 

 

 

 

 

 

Page 26: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 19 -

 

 

5 Functional testing and other procedures 

5.1 Vacuum test 

When first  installed, unused for a long time, moved, or periodically tested,vacuum test is recommended to 

detect leakage in the system of this sterilizer: 

    

 

 

 

                    

   

 

Fig 5.1 Vacuum Test 

 

Vacuum test should be carried out when the system is cooled and 

the chamber is dry. 

In  order  to  successfully  complete  the  vacuum  test,please  check 

that the door of sterilizer is closed and the cooling water valve is 

open before starting the program. 

1 –Sterilization program  2 – Test program 3 – System setting 4 – File record 

Vacuum test program  B&D test program HELIX test program 

press to select 

07/25/2017       09:23:58  Vacuum test program  

15Min 

T:  25.0℃    P: 0.00 Bar Door status 

press to enter

press to start up 

Vacuum test program  

 

 

P: 0.00 Bar   T:  25.0℃      Door status              00:00:00 

Page 27: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 20 -

After the vacuum test program is started, when the pressure in the chamber reaches the standard required 

relative vacuum value, the vacuum system stops, and the equilibration time of 5 minutes is turned on. After the 

balancing time elapses, the vacuum test is started for 10 minutes. After the test is completed, the system returns 

to normal pressure. 

After the vacuum test is finished, the system will prompt the result on the program end interface and save the 

record in the memory. If the automatic printing function is set and the printer is connected, the test result will be 

printed at the same time. 

 

 

When the vacuum test is passed, the message is “vacuum test success”. Indicates that the system's leak rate 

meets the design requirements. Print the information as follows: 

‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐  

        PROGRAM PROPERLY EXECUTED!        Vacuum Test Success!     

        Leakage rate: <0.13Kpa/min         Leak rate:<0.13Kpa/min Your system is steady! 

         ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐  When the vacuum test fails, the message is “Vacuum Test Failed”. Print the information as follows: 

         ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐  

PROGRAM ERROR EXECUTED  Vacuum Test Failed!                    Vacuum Test Fail!                 

‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐‐  

 

If the vacuum test fails, please do not continue to use the product 

and contact the manufacturer or authorized dealer immediately. 

  

5.2 B&D Test 

B&D test is used to check the steam penetration of this sterilizer to porous permeable loads, such as cotton 

fabrics. Select the B&D test in the test program interface (Figure 5.1). 

 

 

Test methods and test results are judged in accordance with the 

information  and  requirements  provided  by  the  test  package 

manufacturer. 

WS310‐2016  requires  weekly  B&D  testing  to  monitor  and  save 

records 

 

 

Page 28: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 21 -

5.3 Helix  Test 

The Helix test is used to verify the steam penetration and saturation for A type hollow instruments, such as 

dental handpieces. Select the Helix test in the test program interface (Figure 5.1). 

 

Test methods and test results are judged in accordance with the 

information  and  requirements  provided  by  the  test  package 

manufacturer. 

 

5.4 Drainage program 

This product allows users to perform drainage procedures when needed. This procedure is used to drain water 

remaining in a small steam sterilizer. 

 

When the water in the chamber is obvious after opening the door, 

it is recommended to use the drainage program before loading the 

item for sterilization. Especially when the program is abnormally 

stopped, this situation is easy to occur. Please open the door for 

inspection. 

Please  carry  out  the  drainage  program  before  handling  this 

product. 

6 System setting 

Enter  the  system  settings  interface  (Figure  6.1)  for  time,  date,  language,  warm‐up,  automatic  printing, 

additional drying settings, and system reset. 

     

 Fig 6.1 System setting 

 

6.1 Time setting 

The time setting operation is as follows: 

 

 

 

 

 

   

 

 

1 –Sterilization program 2 – Test program  3 – System setting  4 – File record  

Press to  enter system setting 

Time setting Date setting Language setting Warm‐up setting 

Press to  enter time setting 

press  and   to  set  up 

time.Press to switch between time,minute  and  second.After 

setting  press   to  exit  time setting. 

 Time setting 

 

              10:27:39 

Page 29: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 22 -

 

 

 

  

Fig 6.4  Time Setting  

    

6.2 Date Setting 

Date setting operation is as follows: 

 

 

 

 

 

 

 

 

 

 

 

 

Fig 6.3 Date Setting  

6.3 Language Settings 

Click  the  [Language  Settings]  button  to  enter  the  language  setting  interface  (Figure  6.2),  select  the 

language.Press   for completion. 

 

 

 

 

 

 

 

 

Completion

Press to  select  date 

setting.Press to  enter date setting. 

Time setting Date setting Language setting Warm‐up setting 

Press and  to  set  up 

date.Press to  switch  between 

year  month  day.Press  to  exit setting. 

Time setting 

              07/25/2017 

Completion 

Press to  select  language 

setting.  Press to  enter into language setting. 

             Time setting Date setting Language setting 

             Warm‐up setting 

Press and to  select 

language  setting.Press to  exit language setting. 

Language setting    English Chinese    French Back                         Confirm 

Page 30: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 23 -

 

 

 

 

Fig 6.2 Language Setting 

6.4 Warm‐up settings 

Warm up setting operation is as follows: 

 

 

 

 

 

 

 

 

 

 

 

 

Fig 6.5 Warm‐up Setting 

 

The  chamber  will  automatically  heat  up  to  110  °C  when  the  automatic 

preheat setting is turned on. Please be careful about burns! 

6.5 Automatic printing settings 

Automatic printing setting operation is as follows: 

 

 

 

 

 

 

 

 

 

 

 

 

Press to  select 

printer  setting.  Pressto  enter  printer  setting system.  

     Date setting       Language setting       Warm‐up setting 

Printer setting 

Press and  to  setup 

printer.Press to exit printer setting after completion. 

Printer setting       Automatic printing: close 

 Completion 

Completion 

Press to select warm‐up 

setting.Press to  enter into warm‐up setting.  

            Time setting Date setting 

Language setting Warm‐up setting 

Press  and to  adjust. 

Press  to exit warm‐up setting after completion. 

Warm‐up setting  

  Automatic warm‐up:     Close     Back                         Confirm 

Completion 

Page 31: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 24 -

Fig 6.6 Automatic print setting  

6.6 Additional drying setting 

Additional drying setting operation is as follows: 

 

 

 

 

 

   

 

 

 

 

 

 

Fig 6.7 Additional drying setting  

 

6.7 Water quality testing 

Water quality testing operation is as follows: 

 

 

 

 

   

 

 

 

 

 

 

Fig6.8 Water quality test settings 

 

6.8 System reset 

System reset operation is as follows: 

 

Press to  select  additional 

drying  setting.  Press to enter  to  set  up  additional drying. 

       Warm‐up setting        Printer setting Additional drying setting          System reset 

Press and  to  enter  into additional drying setting. 

Press to exit additional drying setting after completion. 

Additional drying setting  

Additional drying:close  

Completion

Press to  select  system 

reset.Press to  enter  into system reset. 

Additional drying setting System reset 

Administrator settings System message 

Press to select water 

supply setting.Press to enter  into  water  supply setting.

Warm‐up setting      Printer setting 

Additional drying setting Water supply setting 

Press and to set up water supply setting. 

Press  to exit water supply setting. 

Water supply setting  

 Water quality testing:     Close     

Back                        Confirm 

Completion 

Page 32: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 25 -

 

 

 

 

 

 

 

 

 

 

 Fig 6.9 System reset interface  

 

6.9 System Information 

System information operation is as follows: 

 

 

 

 

 

 

 

 

 

 

    

Fig 6.9 System information interface  

7 Sterilization Log 

The sterilization log records and stores real‐time data for each phase of the sterilization process. The memory 

of this product can store up to two years of sterilization records. Sterilization records are named after serial numbers, 

such as: 0001.TXT, can be exported through the USB interface or printed by a printer. 

Press and  to  select whether  to  perform  a  system 

reset. Press  to exit after the setting is completed. 

System rest  

             Yes            No 

Completion 

Press to  select  system 

reset.Press to enter into system reset

Press to  exit  System information interface. 

 Completion 

Water supply setting System reset Administrator setting.

System message 

System message MODEL:I‐P‐PRO FW:00.02.19.00 HW:Qx3_V0.2 SN:65252120517050508 Back 

Page 33: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 26 -

7.1 Log export 

For  long‐term  storage  of  sterilization  records,  sterilization  records  can  be  exported  via USB  interface  to  a 

computer or portable storage device. Specific operation: Insert the USB flash drive into the USB interface and export 

the record as follows. After the export is completed, the interface displays the export completion prompt. Please 

remove the USB flash drive. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig 7.1 Record Export 

7.2 Log printing 

When the automatic printing function is turned on and the sterilizer is properly connected to the printer, the 

sterilization record can be automatically printed after the program ends. See 6.5 for details on how to operate. 

 

 

 

 

 

 

File export Export all records? 

              Export 

1 –Sterilization program  2 –Test program  3 –System setting 4 – File record 

按 或 选择

是否导出记录。

按 键进行记

录导出。

File export 

 Exporting file…  Record copy 

File export Export completed     Exit 

Press to exit 

Page 34: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 27 -

 

 

7.3 Log interpretation 

The format of the sterilization record for printer output and internal memory storage is as follows.                                                 

- - - - - - - - - - - - - - - - Quantus Sterilizer - - - - - - - - - - - - - - - - Program : Quick-program-S 134°C unwrapped Data : 2017.07.12 Start Time : 09:11:29 Cycle : 1 SN : 65252120517050513 Preheat Temper : 95.0°C Conductivity :11.0us/cm - - - - - - - - - - - - - - - - Program step press temper time bar°C min Start : -0.0469.6 0:00:00 1.Vacuum one Sub press : -0.8772.00:02:26 Steam press: 0.50109.90:05:32 2.Vacuum two Sub press : -0.5289.20:06:32 Steam press: 2.14134.30:11:51 - - - - - - - - - - - - - - - - Keep : 2.14134.30:11:51 Keep End : 2.17135.0 0:16:51 Realse : 0.10 105.6 0:18:27 Dry : 0.10105.60:18:27 Dry End : -0.9190.3 0:22:27 Ventile : -0.1492.3 0:22:39 Finish : -0.1492.3 0:22:39 - - - - - - - - - - - - - - - - PROGRAM PROPERLY EXECUTED! Temperature : 134.7+0.2/-0.4°C Pressure : 2.14+0.02/-0.03bar Sterilization : 0:22:39 End Time : 09:34:08 Hardware Version: Qx3_V0.2 Software Version: 00.02.16.19 - - - - - - - - - - - - - - - -

Product Number  Program Date Start Time Running times Serial number Preheating temperature Distilled water conductivity     Vacuum twice, recording pressure, Temperature time information    Sterilization holding stage Sterilization holding completion Depressurization Drying start Drying completion Inlet Sterilization completion    Final report, recording sterilization temperature,pressure, program running time and completion time 

Hardware and software version 

Page 35: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 28 -

8 Small Steam Sterilizer Maintenance 

8.1 Cleaning of loaded items prior to sterilization 

  Before sterilization, the load of handpiece must be thoroughly cleared. 

Dust and contaminants separated from the load will cause blockage of filters, valves and 

piping, therefore, the sterilization load without cleaning is not allowed to enter the device 

for sterilization. 

Follow all  regulations  for cleaning, disinfection, and sterilization provided by the device 

manufacturer. 

8.2 Small steam sterilizer maintenance 

8.2.1 Cleaning and maintenance 

Regular cleaning and maintenance can effectively extend the life of the sterilizer and ensure a good working 

condition. 

Maintenance item  Maintenance cycle  Special tool  Maintenance operation 

Replace the seal  Sterilize 1000 times    Open the door and check whether the seal is 

deformed or wrinkle, etc. If it is deformed or 

wrinkled, it is recommended to replace it. 

Seal maintenance  2 weeks    Remove  the  seal.Please  rinse  it with water 

and wipe  it  dry with a  soft  cloth.After  that 

replace the lid. 

Note: Pay attention to the installation direction of the seal (the narrow side is close to the chamber cover and the wide side is close to the chamber). 

Door lock maintenance  Sterilize 1000 times  lubricating oil  A small amount of  lubricating oil  is sprayed 

on the left side of the lid through the hole (A 

indicated), and the outer surface is wiped by 

a paper towel. 

 

Cleaning  distilled  water 

tank Sterilize 300 times 

  Wash it with clean water. 

Clean  the  filter  at  both 

ends  of  cooling  water 

inlet pipe. 

Sterilize 100 times 

  Unscrew  the  nut  at  both  ends  of  the  inlet 

pipe and rinse it with clean water. 

A

Page 36: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 29 -

Clean the filter at 

the  bottom  of  the 

sterilization chamber 

Sterilize 200 times 

 Wrench  Unscrew the filter with a wrench and wash it 

with clean water. 

Clean trays  Sterilize 100 times   Rinse it with clean water, if stained, clean it 

with toothpaste 

Sterilization chamber  Every week 

  Wipe the sterilization chamber with a clean 

and wet medical cotton cloth,then wipe the 

chamber with a clean and dry medical cotton 

cloth 

 

 

8.2.2 Regular maintenance 

 

WS310‐2016  and  WS506‐2016  require  a  comprehensive 

maintenance  of  the  small  steam  sterilizer  every  year  to  ensure 

good equipment working condition and sterilization effect 

Maintenance and take care of this device must be performed by a 

professional authorized by the manufacturer. 

  

 

 

 

 

 

 

 

9 Faults and common problem handling 

 

9.1 No display on the screen  

Faulty description: The display does not light up normally when the sterilizer is turned on. 

Solution: 

Check if there is power in the main power outlet 

Check if the power plug of this product is plugged in. 

The screen still does not appear, please contact professional distributor authorized by HӓTmed. 

9.2 Large consumption of distilled water 

Faulty description:Distilled water consumption is much larger than the specified amount 

Page 37: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 30 -

Solution: 

Please check  if the foot pad adjustment  is correct. Under normal circumstances, the front side  is 

slightly higher than the back side. 

Open the door and check if the filter at the bottom of the sterilization chamber is blocked. 

None of the above can help reduce the consumption of distilled water, please contact professional 

distributor authorized by HӓTmed. 

 

9.3 The drying effect is not good 

Faulty description: Drying cannot achieve satisfactory results 

Solution: 

Please check  if the foot pad adjustment  is correct. Under normal circumstances, the front side  is 

slightly higher than the back side. 

Open the door and check if the filter at the bottom of the sterilization chamber is blocked. 

Please check if the load exceeds the maximum loading capacity(especially textiles) and if the  load  

is placed correctly in the chamber(do not touch the wall of the pressure chamber directly) and if the 

standard trays and tray rack are used for loading. 

Activate automatic warm‐up. 

Open the left side cover and check if the filter is clogged 

None of the above can help  improve the effect of drying, please contact professional distributor 

authorized by HӓTmed. 

 

9.4 No distilled water 

Faulty description:No distilled water 

Solution: 

Please check if there is water in distilled water tank.  

Please check if the distilled water pipe is bent or knotted and inserted into place. 

9.5 Unable to open the door 

Faulty description: The door cannot be opened even if the emergency button is activated. 

Solution: 

Please check if the air filter at the rear side of sterilizer is blocked 

Take off side cover,unplug the silicone tube connected to the air filter solenoid valve and discharge 

the negative pressure inside the chamber. 

When you hear the exhaust sound gradually weaken until it disappears, you can open the door. 

Page 38: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 31 -

Return the silicone tube to the original position and clamp. Turn on the vacuum test to check for 

leakage. 

9.6 Leakage at the door or dripping underneath 

Faulty description: Air leakage and water dripping are obvious. 

Solution: 

Please check if the seal is installed correctly(the long side with the word "H" is exposed outside). 

Please check if the seal is damaged. 

 

Appendix 1 Safety Device Description 

Preheating  system  over  temperature  protection  switch:  The  preheating  system  has  an  over‐temperature 

protection switch that automatically disconnects the preheating system power supply circuit when the temperature 

exceeds  the  limitation  of  protection  switch  threshold  temperature  to  realize  the  protection  function  of  the 

preheating system. 

Steam  generator  over  temperature  protection  switch:  The  steam  generator  has  three  temperature 

switches.The  protection  switch  that  automatically  disconnects  steam  generator  power  supply  circuit  when  its 

temperature exceeds threshold temperature. 

Pressure relief valve: When the pressure inside the chamber is 1.3 times greater than the working pressure, 

the pressure relief valve will automatically open in order to keep the pressure  inside the chamber within a safe 

pressure range. 

 

 

 

 

 

 

 

 

 

 

 

 

 

Appendix 2 Schematic diagram of piping system 

 

Page 39: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

- 32 -

 

 

 

Piping system schematic 

 

 

 

 

 

 

 

 

 

 

 

Appendix 3 Electrical schematic diagram 

 

Page 40: Small Steam Sterilizer - Zone Medical · 2021. 6. 14. · I HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms to the YY/T

Small steam sterilizer electrical schematic diagram