Page 1: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

2402 OxyInstruction manual

Stratos®

BE_2402OXY-CSA.QXD 22.07.2003 15:48 Uhr Seite 1

Page 2: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

KnickElektronische Messgeräte GmbH & Co. KGP.O. Box 37 04 15D-14134 Berlin

Phone: +49 (0)30 - 801 91 - 0Fax: +49 (0)30 - 801 91 - 200Internet: http://[email protected]

WarrantyDefects occurring within 3 years from delivery date shall beremedied free of charge at our plant (carriage and insurancepaid by sender).Sensors, fittings, and accessories: 1 year.

Subject to change without notice

BE_2402OXY-CSA.QXD 22.07.2003 15:48 Uhr Seite 2

3

Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . .7Overview of Stratos e 2402 Oxy . . . . . . . . . . . . . . . . . . . .9Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10

Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10Mounting plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11Pipe mounting, panel mounting . . . . . . . . . . . . . . . . . . . . . . . .12

Installation and connection . . . . . . . . . . . . . . . . . . . . . . .14Information on installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Terminal assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Wiring examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16Protective wiring of relay outputs . . . . . . . . . . . . . . . . . . . . . . .18

User interface and display . . . . . . . . . . . . . . . . . . . . . . . .20Operation: Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23

Sensocheck, Sensoface sensor monitoring . . . . . . . . . . . . . . . .23GainCheck device self test . . . . . . . . . . . . . . . . . . . . . . . . . . . .23Automatic device self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24To activate the Hold mode from outside . . . . . . . . . . . . . . . . . .25

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26Menu structure of configuration . . . . . . . . . . . . . . . . . . . . . . . .27Overview of configuration steps . . . . . . . . . . . . . . . . . . . . . . . .28Output 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30Output 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44Calibration mode, alarm settings . . . . . . . . . . . . . . . . . . . . . . .46Limit function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52Control of rinsing and cleaning systems . . . . . . . . . . . . . . . . . .54

Contents

BE_2402OXY-CSA.QXD 22.07.2003 15:48 Uhr Seite 3

Return of products under warrantyPlease contact our Service Team before returning a defective device. Shipthe cleaned device to the address you have been given. If the device hasbeen in contact with process fluids, it must be decontaminated/disinfectedbefore shipment. In that case, please attach a corresponding certificate, forthe health and safety of our service personnel.

DisposalPlease observe the applicable local or national regulationsconcerning the disposal of “waste electrical and electronicequipment”.

Page 3: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

4 Stratos e 2402 Oxy

Parameter set 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56Default settings of parameter sets . . . . . . . . . . . . . . . . . . . . . .57Parameter set, individual settings . . . . . . . . . . . . . . . . . . . . . . .58

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60Calibration to saturation (SAT) . . . . . . . . . . . . . . . . . . . . . . . . .62Calibration to concentration (Conc) . . . . . . . . . . . . . . . . . . . . .64Zero calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68Adjusting temp probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70

Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . .71Controller functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74

PID controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74Pulse length / pulse frequency controller . . . . . . . . . . . . . . . . . .76

Connecting a rinsing system . . . . . . . . . . . . . . . . . . . . . . .77Operation with automatic cleaning system . . . . . . . . . .77Error messages (error codes) . . . . . . . . . . . . . . . . . . . . . .78

Calibration error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . .80Operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80Sensoface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82

Sensocheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85

Product line and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .85Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86Explosion protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92Warnings and notes to ensure safe operation . . . . . . . . . . . . .93Control Drawing CSA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98

BE_2402OXY-CSA.QXD 22.07.2003 15:48 Uhr Seite 4

5

Safety information

Be sure to read and observe the following instruc-tions!

The device has been designed in accordance with the state ofthe art and complying with the applicable safety regulations.When operating the device, certain conditions may neverthe-less lead to danger for the operator or damage to the device.

Caution!Commissioning may only be carried out by trained experts.Whenever it is likely that protection has been impaired, thedevice shall be made inoperative and secured against un-intended operation.

The protection is likely to be impaired if, for example:• the device shows visible damage• the device fails to perform the intended measurements• after prolonged storage at temperatures above 70 °C• after severe transport stresses

Before recommissioning the device, a professional routine testin accordance with EN 61010-1 must be performed. This testshould be carried out by the manufacturer.

BE_2402OXY-CSA.QXD 22.07.2003 15:48 Uhr Seite 5

Page 4: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

6 Stratos e 2402 Oxy

Intended use

Stratos e 2402 Oxy is used for dissolved oxygen and temperaturemeasurement in biotechnology, pharmaceutical industry, as wellas in the field of environment, food processing, and sewagetreatment.The rugged molded enclosure can be wall or pipe mounted orfixed into a control panel.The protective hood provides additional protection against directweather exposure and mechanical damage.The Stratos has been designed for application with amperometricsensors, e.g. Knick SE 704, SE 705, SE 706, SE 707.

Trademarks

The following names are registered trademarks. For practical rea-sons they are shown without trademark symbol in this manual.Stratos®

Sensocheck®

Sensoface®

Calimatic®

GainCheck®

InPro® and EasyClean® are registered trademarks of Mettler-Toledo.

BE_2402OXY-CSA.QXD 22.07.2003 15:48 Uhr Seite 6

7

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 7

Page 5: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

8 Stratos e 2402 Oxy

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 8

9

Overview of Stratos e 2402 Oxy

1

2

3

4

5

E

D

C

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

DOinput

Hold

input

Control

input

Output 1

Output 2

Alarm

Clean

Power

Do not connect!

Cathode

Guard

Anode

Referenceelectrode

RTD

RTD

shield

HOLD

HOLD/CONTROL

CONTROL

+ Output 1

- Output 1/2

+ Output 2

Relay 1

Relay 1/2

Relay 2

Alarm

Alarm

Clean

Clean

Power

Power

R1

R2

Temp input

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 9

Provided documentation:

• Instruction manual

• Short instructions

• Safety instructions / • EC Declarations of Conformity

• EC-Type Examination Certificate

70084

XXXX (X)Bedienungsanleitung

Stratos®

XXXX X

Safety precautions . . . . . . . . . . . . . . 3Sicherheitshinweise . . . . . . . . . . . . . 5Consignes de sécurité . . . . . . . . . . . 7Norme di Sicurezza . . . . . . . . . . . . . . 9Avisos de seguridad . . . . . . . . . . . . 11Turvallisuusohjeet . . . . . . . . . . . . . . 13Säkerhetsanvisningar . . . . . . . . . . . 15Sikkerhetsinstrukser. . . . . . . . . . . . 17Veiligheidsvoorschriften. . . . . . . . . 19Sikkerhedshenvisninger. . . . . . . . . 21Indicações de segurança . . . . . . . . 23Sigurnosne upute . . . . . . . . . . . . . . 25Saugumo nurodymai. . . . . . . . . . . . 27Wskazówki dotyczącebezpieczeństwa 29Указания по безопасности . . . . . . 31Bezpečnostní pokyny . . . . . . . . . . . 33Biztonsági tudnivalók . . . . . . . . . . . 35

Stratos®

CAL CA. Sol(BUF)

1100 1100

CAL CA. Sol(MAN)

Meß-Modus / Measuring mode / Mode mesure

Kalibrierung / Calibration / Calibrage

AutomatischAutomaticAutomatique

ManuellManualManuel

Cal-InfoKalibrierungProduktkalibrierungVerschiebung Nullpkt.Temp. AbgleichLetzter FehlerKonfigurierungSensormonitorStromgeber 1Stromgeber 2RelaistestRegler manuellParametersatz 1/2Paßzahl-Editor *

FehlermeldungenSteilheit /Asymmetriepotential

Drift der Meßkette

Pufferlösung

Meßbereich pH

Meßbereich mV

Temperatur

Ausgangsstrom 1zu kleinzu großSpanneAusgangstrom 2zu kleinzu großSpanneSensocheckGlaselektrodeBezugselektrodeSpülsondeKommunikationCal-Fehler

Cal / Conf-Daten

Systemausfall

*2-L

eite

r-G

erät

kon

figur

ierb

ar /

2-w

ire d

evic

e co

nfig

urab

le /

app

arei

l 2-f

ils c

onfig

urab

legr

au/g

ray/

gris

: nu

r 4-

Leite

r-G

erät

/ 4

-wire

dev

ice

only

/ a

ppar

eil 4

-fils

seu

lem

ent

Cal InfoCalibrationProduct calibrationZero point adjustm.Temp probe adjustm.Last errorConfigurationSensor monitorCurrent source 1Current source 2Relay testManual controllerParameter set 1/2Passcode editor *

Error messagesSlope /Asymmetry potential

Electrode drift

Buffer solution

pH range

mV range

Temperature

Output current 1too lowtoo highSpanOutput current 2too lowtoo highSpanSensocheckGlass electrodeReference electrodeRinsing probeCommunicationCal error

Cal / Conf data

System failure

Cal-InfoCalibrageCalibrage du produitRéglage du point zéroRéglage sonde temp. Dernière erreurConfigurationContrôle capteurGénérateur courant 1Générateur courant 2Test des relaisRégulateur man.Jeu de paramètres 1/2Editeur code d´accès *

Messages d‘erreurPente / Potentiel asymétrique

Dérive électrode

Solution tampon

Gamme de mesure pH

Gamme de mesure mV

Température

Courant sortie 1trop bastrop hautPlageCourant sortie 2trop bastrop hautPlageSensocheckElectrode de verreEl. de référenceSonde rincageCommunicationErreur cal.

Données Cal / Conf

Défaillance système

0000 *1100 *1105 *1001 *1015 *0000 *1200 *2222 *5555 *55565557555976541989 *

Err 01:

Err 02:

Err 03:

Err 11:Err 12:Err 13:

Err 21:Err 22:Err 23:

Err 33:Err 34:

Err 41:Err 42:

Err 98:

Err 99:

pHKurzanleitung Short instructions Instructions courtes

TE-194.100-XXX01

Devices only:

Page 6: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

10 Stratos e 2402 Oxy

Package contentsCheck the shipment for transport damage and completeness.The package should contain:• Front unit of Stratos• Lower case• Bag containing small parts• Instruction manual• Specific test report

Assembly

11

10

9

8

7 6 5 4

1

2

3

Fig.: Assembling the enclosure

6 Sealing inserts (1 piece)7 Rubber reducer (1 piece)8 Cable glands (3 pieces)9 Filler plugs (3 pieces)10 Hexagon nuts (5 pieces)11 Sealing plugs (2 pieces),

for sealing in case of wall mounting

1 Jumper (2 pieces)2 Washer (1 piece), for conduit

mounting: Place washer between enclosure and nut

3 Cable ties (3 pieces)4 Hinge pin (1 piece), insertable

from either side5 Enclosure screws (4 pieces)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 10

11

Mounting plan

144

144

15 42

84

8032

21

43

105

27

726,

2

1 2

3

4

1 Cable gland (3 pieces)2 Breakthroughs for cable gland

or conduit 1/2”, dia 21.5 mm (2 breakthroughs)Conduits not included!

3 Breakthroughs for pipe mounting (4 breakthroughs)

4 Breakthroughs for wall mounting (2 breakthroughs)

Fig.: Mounting plan

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 11

Page 7: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

12 Stratos e 2402 Oxy

40 60132

1

2

34

5

Fig.: ZU 0274 pipe-mount kit

1 ZU 0276 protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical or horizontal posts or pipes

5 Self-tapping screws (4 pieces)

1

132165

173

1

Fig.: ZU 0276 protective hood for wall and pipe mounting

Pipe mounting, panel mounting

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 12

13

1

2

3

45

max. 2578 27

1...22

1 Screws (4 pieces) 2 Gasket (1 piece) 3 Control panel 4 Span pieces (4 pieces) 5 Threaded sleeves (4 pieces)

Fig.: ZU 0275 panel-mount kit

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 13

Page 8: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Stratos e 2402 Oxy14

Installation and connection

Information on installationCaution!

• The Stratos may only be installed by trained experts in accordance with this instruction manual and as per applicable local and national codes.

• Be sure to observe the technical specifications and input ratings.• Be sure not to notch the conductor when stripping the

insulation.• Before connecting the device to the power supply, make sure

that its voltage lies within the range 20.5 to 253 V AC/DC.• When commissioning, a complete configuration must be

carried out by the system administrator.

The terminals are suitable for single wires and flexible leads up to 2.5 mm2 (AWG 14).

Warning!Additional safety precautions have to be taken for applications inhazardous locations to CSA (CLI DIV2 GPA,B,C,D T4, Ex nA IIC T4)!(See Pg 93.)

Terminal assignments

Fig.: Terminal assignments Stratos e 2402 Oxy

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 14

15

13

2

4

1 ESD shield covering the sensor inputs(Screw off for assembly)Note: The cable shield must end under the ESD shield.(Cut lines if required)

2 Terminals for temperature probe3 Terminals for sensor4 Connection of power supplyFig.: Information on installation, rear side of Stratos

The connections to the Stratos are incendive andmust be installed in accordance with the NationalElectric Code (ANSI-NFPA 70) Division 2 hazardous(classified) location incendive wiring techniques.

Division 2 wiring

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 15

Page 9: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

16 Stratos e 2402 Oxy

Wiring example 1SE 704/SE 705 sensor (Series InPro6000) and ZU 0231 cable

red

whi

te

blac

k

blac

k

yello

w /

gree

nSE 704 / SE 705 sensor (or InPro6000)

Connection Terminal Cable (e.g. ZU 0231)

1 not connectedcathode 2 white (coax core)guard 3 not connectedref. el. 4 (jumper 4-5)anode 5 red (coax shield)RTD E black RTD D blackshield C yellow / green

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 16

1717

Wiring example 2Sensors with connection via VP cable

red

tran

spar

ent

gree

n

whi

te

yello

w /

gree

n

blue

gray

SE 706 / SE 707 sensor(or InPro6800/6900)

Connection Terminal VP cable (e.g. ZU 0313)

1 not connectedcathode 2 transparent (coax core)guard 3 gray / SE 707 (InPro6900)ref. el. 4 (jumper 4-5), blueanode 5 red (coax shield)RTD E greenRTD D whiteshield C yellow/green

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 17

Page 10: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Stratos e 2402 Oxy18

Protective wiring of switching contactsRelay contacts are subjected to electrical erosion. Especiallywith inductive and capacitive loads, the service life of the con-tacts will be reduced. For suppression of sparks and arcing,components such as RC combinations, nonlinear resistors,series resistors and diodes should be used.

1

23

1 2

3

Typical AC applicationswith inductive load

1 Load2 RC combination, e.g. RIFA PMR 209

Typical RC combinationsfor 230 V AC: Capacitor 0.1µF / 630V,Resistor 100 Ohms / 1 W

3 Contact

Protective wiring of switching outputs

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 18

19

A: DC application with inductive loadB: AC/DC applications with capacitive loadC: Connection of incandescent lamps

A1 Inductive loadA2 Free-wheeling diode, e.g. 1N4007 (Observe polarity)A3 ContactB1 Capacitive loadB2 Resistor, e.g. 8 Ohms/1 W at 24 V / 0.3 AB3 ContactC1 Incandescent lamp, max 60 W / 230 V, 30 W / 115 VC3 Contact

Warning!Make sure that the maximum ratings of the relaycontacts are not exceeded even during switching!

Typical protective wiring measures

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 19

Page 11: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

20 Stratos e 2402 Oxy

User interface and display

1 Display2 Mode indicators (no keys),

from left to right:- Measuring mode- Calibration mode- Alarm- Wash contact- Configuration mode

3 Keypad4 Coding5 Rating plate6 Model designation7 Alarm LED

User interface

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 20

21

1 2 3 4 5 6 7 8 9 10

11

12

13

1617

20

18

19

15 14

1 Mode code entry2 Display of measured variable*3 Temperature4 Current output5 Limit values6 Alarm7 Sensocheck8 Calibration9 Interval/response time

10 Wash contact11 Measurement symbols12 Proceed with enter13 Bar for identifying the device

status, above mode indicators,from left to right:- Measuring mode- Calibration mode- Alarm- Wash contact- Configuration mode

14 Lower display15 Manual temp indicator16 Hold mode active17 Waiting time running18 Sensor data19 Main display20 Sensoface

* Not in use

Display

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 21

Page 12: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

22 Stratos e 2402 Oxy

Operation: Keypad

Start, end calibration

Start, end configuration

Select digit position(selected position flashes)

Edit digit

• Calibration: Continue in program sequence

• Configuration: Confirm entries,next configuration step

• Measuring mode: Display output current

Cal Info, display of zero current and slope

Error Info, display last error message

Start GainCheck device self-test+

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 22

23

Sensoface provides information on the sensor condi-tion.The zero point, slope, and response time during calibra-tion are evaluated. The three Sensoface indicators pro-vide the user with information about wear and requiredmaintenance of the sensor.

GainCheck device self testA display test is carried out, the software version is displayedand the memory and measured value transfer are checked.

Start GainCheck device self-test:

Automatic device self-testThe automatic device self-test checks the memory and meas-ured-value transfer. It runs automatically in the background atfixed intervals.

+

Sensocheck, Sensoface sensor monitoringSensocheck continuously monitors the sensor and leads.Sensocheck can be switched off (Configuration, Pg 47).

Safety functions .

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 23

Page 13: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

24 Stratos e 2402 Oxy

Hold modeDisplay:

The Hold mode is a safety mode during configuration and cali-bration. Output current is frozen (Last) or set to a fixed value(Fix). Alarm and limit contacts are disabled.If the calibration or configuration mode is exited, the Stratosremains in the Hold mode for safety reasons. This preventsundesirable reactions of the connected peripherals due toincorrect configuration or calibration. The measured value and“HOLD” are displayed alternately. The Stratos only returns tomeasuring mode after enter is pressed and a waiting time of20 s has passed.

Safety features

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 24

25

To activate the Hold mode from outside

The Hold mode can be activated from outside by sending asignal to the Hold input (e.g. from the process control system).

Hold active Hold inactive

10 ... 30 V AC/DC 0 ... 2 V AC/DC

6

7

Hold

input

Process control system (PCS)

Power supply12...24 V AC/DC

Stratos e 2402

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 25

Page 14: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Stratos e 2402 Oxy26

ConfigurationIn the Configuration mode you set the device parameters.

Activate with conf

Enter mode code “1200”Edit parameter with and ,confirm/continue with enter.(End with conf enter.)

During configuration the Stratos remainsin the Hold mode for reasons of safety.The output current is frozen (at its lastvalue or at a preset fixed value, depend-ing on the configuration), limit andalarm contacts are inactive. The con-troller is in the configured state,Sensoface is off, mode indicator“Configuration” is on.

Activate

Hold

Input errors The configuration parameters arechecked during the input. In the case ofan incorrect input “Err” is displayed forapprox. 3 sec. The incorrect parameterscannot be stored. Input must be repeat-ed.

End End with conf. The measured value andHold are displayed alternately, “enter”flashes. End Hold mode with enter. The display shows the measured value.The output current remains frozen foranother 20 sec (HOLD icon on, “hour-glass” flashes).

HOLD icon

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 26

27

Menu group Code

Output 1 o1.

Output 2 o2.

Correction CO.

Calibration mode CA.

Alarm settings AL.

Relay / Controller rL.

Rinsing andcleaning probes

Pb.

DisplaySelect menu group

Previousmenu group:

Menu structure of configuration

Menu item 1

Menu item 2

Menu item ...

The configuration steps are assigned to different menu groups. With the arrow keys you can jump between the individual menugroups.Each menu group contains menu items for setting the parameters.Pressing enter opens a menu item. The values are edited using the arrow keys. Pressing enter confirms/stores the settings.Return to measurement: Press conf key.

Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 27

Page 15: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

28 Stratos e 2402 Oxy

Overview of configuration steps

MenuCode Choices

out1 Output 1

o1. Select sensor type Standard (Type A) / Traces (Type B)

Select saturation / concentration % / mg/l, ppm

Select current range 0-20 mA / 4-20 mA

Enter current beginning xxxx

Enter current end xxxx

Time constant of output filter xxxx s

22 mA signal in the case of error ON / OFF

Signal behavior during HOLD Last / Fix

Enter fixed value xxx.x mA

out2 Output 2

o2. Select temperature unit °C / °F

Select temperature probe 22NTC / 30NTC

Select current range 0-20 mA / 4-20 mA

Enter current beginning xxx.x

Enter current end xxx.x

Time constant of output filter xxxx s

22 mA signal in the case of temp error ON / OFF

Signal behavior during HOLD Last / Fix

Enter fixed value xxx.x mA

Corr Correction

Co. Enter polarization voltage 0675 mV / xxxx mV

Select pressure unit bar / kPa / PSi

Select process pressure correction x.xxx bar / 1.013 bar

Enter salt correction xx.xx mg/l

CAL Calibration mode

CA. Select saturation / concentration SAt / Conc

Enter cal timer interval xxxx h

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 28

29

ChoicesCode Auswahl

ALr Alarm settings

AL. Select Sensocheck ON / OFF

Enter alarm delay xxxx s

LED in HOLD mode ON / OFF

rLAY Relay 1/2: Limit values, controller

rL. Select limit function / controller LiMIT / CtROL

Select contact function Lo / HiL1.

Select contact response N/O / N/C

Enter switching point xxxx

Enter hysteresis xxxx

Enter delay xxxx SEC

Select contact function xxxx sL2.

Enter controller setpoint xxxxCt.

Enter neutral zone xxxx

(P) Controller gain KP xxxx %

(I) Reset time TR xxxx SEC

(D) Rate time TD xxxx SEC

Pulse length /Pulse frequency PLC / PFC

PLC: Pulse length xxxx SEC

PFC: Pulse frequency xxxx /min

Select HOLD behavior Y Last / Y Off

PrbE Rinsing and cleaning probes

Pb. Select rinsing/cleaning probe EASYCLN / rinse

Rinsing interval xxx.x hrinse

Rinse duration xxxx SEC

Contact response N/O / N/C

Cleaning interval xxx.x hEASYCLN

Select contact response N/O / N/C

Enter switching point xxxx

Enter hysteresis xxxx

Enter delay xxxx SEC

Lock cleaning interval Off / On

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 29

Page 16: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

* Sensor Type B (traces)

* Sensors Type A (standard applications)

30 Stratos e 2402 Oxy

ConfigurationOutput 1Select sensor type. Measurement procedure.

Select sensor type*

Select meas. procedure

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA in the case of error

Hold mode

End:Press conf, then enter

Code

Output 1 o1.

DisplayMenu group Select menu item

Screw cap Sensor current in air (25 °C)Sensor type

SE 704 / SE 705/ SE 706InPro6100 / InPro6200InPro6800

4-pole4-poleVP

30 to 65 nA50 to 110 nAtyp. 60 nA

0.01 ppm0.01 ppm

0.006 ppm

Screw cap Sensor current in air (25 °C)Sensor type

InPro6900 / SE 707 VP typ. 350 nA 0.001 ppm

Detectionlimit

Detectionlimit

Note: Stratos 2402 Oxy has as device a resolution of 0.01 ppm.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 30

31

Note: Characters represented in gray are flashing and can be edited.

Code Display Action Choices

o1. Select configuration(Press conf key).

Enter mode code “1200”(Select position with key and edit number with key. When the display reads “1200”, press enter to confirm.)

The Stratos is in HOLD mode (HOLDicon is on).

Select sensor Type A / B(see table on left-hand side)Select with keyProceed with enter

Type A(SE 704 /705 / 706,InPro6000...6800)Type B(InPro6900,SE 707)

Select measurement procedure (valid for all following settings):• SAt: Saturation (%)• Conc: Concentration

(mg/l or ppm) Select with keyProceed with enter

%mg/lppm

After correct input a wel-come text (CONF) is dis-played for approx. 3 sec

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 31

Page 17: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

32 Stratos e 2402 Oxy

ConfigurationOutput 1Output current range. Current beginning. Current end.

Select sensor type

Select meas. procedure

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA in the case of error

Hold mode

End:Press conf, then enter

Code

Output 1 o1.

DisplayMenu group Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 32

33

Code

4 - 20 mA(0 - 20 mA)

Display Action Choices

o1. Set output current rangeSelect with keyProceed with enter

Current beginning Enter lower end of scale, dependingon the measurement procedureselected (Saturation orConcentration)Select with key, edit number with

key, proceed with enter

0000 %(mg/l, ppm)

Current endEnter upper end of scale, dependingon the measurement procedureselected (Saturation orConcentration)Select with key, edit number with

key, proceed with enter

0100 %(mg/l, ppm)

Output current

Oxygen saturation

204

100

0

Assignment of measured values:Current beginning and current end

Oxygen saturation

204

70

50[mA]

Example 1: Range 0 to 100 % Example 2: Range 50 to 70%.Advantage: Higher resolution inrange of interest[%]

[mA]

[%]

Output current

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 33

Page 18: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

34 Stratos e 2402 Oxy

ConfigurationOutput 1Time constant of output filter.

Select sensor type

Select meas. procedure

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA in the case of error

Hold mode

End:Press conf, then enter

Code

Output 1 o1.

DisplayMenu group Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 34

35

0 sec(0 - 120 sec)

Code Display Action Choices

o1. Time constant of output filterDefault setting: 0 sec (inactive).To specify a time constant: Select with key, edit number with key, proceed with enter.

Time constant of output filterTo smoothen the current output, a low-pass filter withadjustable filter time constant can be switched on. When thereis a jump at the input (100 %), the output level is 63 % afterthe time constant has been reached. The time constant can be set from 0 to 120 sec.If the time constant is set to 0 sec, the current output followsthe input. Note:The filter only acts on the current output, not on the display, thelimit values, or the controller!

Time constant 0 to 120 sec

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 35

Page 19: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

36 Stratos e 2402 Oxy

ConfigurationOutput 1Output current during Error and HOLD.

Select sensor type

Select meas. procedure

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA in the case of error

Hold mode

End:Press conf, then enter

Code

Output 1 o1.

DisplayMenu group Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 36

37

OFF(ON)

Code Display Action Choices

o1. 22 mA signal for error messageSelect with keyProceed with enter

Output signal during HOLDLAST: During HOLD the last measured valueis maintained at the outputFIX:During HOLD a value (to be entered)is maintained at the outputSelect with keyProceed with enter

LAST(FIX)

Only with FIX selected:Enter current which is to flow at theoutput during HOLDSelect with key, edit number with key, proceed with enter.

21.0 mA(00.0 to21.0 mA

Output signal during HOLD:

Output current[mA]

Output signal HOLDSetting FIX = 21.0 mA

HOLD active

21

4

Output signal HOLDSetting LAST

HOLD active

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 37

Page 20: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

38 Stratos e 2402 Oxy

ConfigurationOutput 2Temperature unit and probe. Output current.

Select °C/°F

Select temp probe

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA for temp error

Hold mode

End:Press conf, then enter

Code DisplayMenu group Select menu item

Output 2 o2.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 38

39

°C(°F)

Code Display Action Choices

o2. Specify temperature unitSelect with keyProceed with enter

Select temperature probeSelect with keyProceed with enter

22NTC(30NTC)

4 - 20 mA(0 - 20 mA)

Set output current rangeSelect with keyProceed with enter

Current beginning: Enter lower end of scale. Select with key, edit number with

key, proceed with enter.

000.0 °C

Current end: Enter upper end of scale. Select with key, edit number with

key, proceed with enter.

100.0 °C

Output current

Process temperature

204

100

0

Process temperature: Current beginning and current end

Process temperature

204

70

50[mA]

Example 1: Range 0 to 100 °C Example 2: Range 50 to 70 °C.Advantage: Higher resolution inrange of interest[°C]

[°C]

[mA]

Output current

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 39

Page 21: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

40 Stratos e 2402 Oxy

ConfigurationOutput 2Time constant of output filter.

Select °C/°F

Select temp probe

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA for temp error

Hold mode

End:Press conf, then enter

Output 2 o2.

Code DisplayMenu group Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 40

41

0 sec(0 - 120 sec)

Code Display Action Choices

o2. Time constant of output filterDefault setting: 0 s (inactive).To specify a time constant: Select with key, edit number with key, proceed with enter.

Time constant of output filterTo smoothen the current output 2, a low-pass filter withadjustable filter time constant can be switched on. When thereis a jump at the input (100 %), the output level is 63 % afterthe time constant has been reached. The time constant can be set from 0 to 120 sec.If the time constant is set to 0 sec (default), the current outputfollows the input. Note:The filter only acts on the current output, not on the display!

Time constant 0 to 120 s

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 41

Page 22: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

42 Stratos e 2402 Oxy

ConfigurationOutput 2Temperature error. Output current during HOLD.

Select °C/°F

Select temp probe

Select 0-20 / 4-20 mA

Enter current beginning

Enter current end

Set output filter

22 mA for temp error

Hold mode

End:Press conf, then enter

Output 2 o2.

Code DisplayMenu group Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 42

43

OFF(ON)

Code Display Action Choices

o2. 22 mA signal for error messageSelect with keyProceed with enter

Output signal during HOLDLAST: During HOLD the last measuredvalue is maintained at the outputFIX:During HOLD a value (to be entered)is maintained at the outputSelect with keyProceed with enter

LAST(FIX)

Only with FIX selected:Enter current which is to flow at theoutput during HOLDSelect with key, edit number with key, proceed with enter.

21.0 mA(00.0 to21.0 mA)

Output signal during HOLD:

Output current[mA]

Output signal HOLDSetting FIX = 21.0 mA

HOLD active

21

4

Output signal HOLDSetting LAST

HOLD active

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 43

Page 23: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

44 Stratos e 2402 Oxy

ConfigurationCorrectionPolarization voltage. Process pressure. Salt correction.

Polarization voltage

Meas. unit (pressure)

Process pressure

Salt correction

End:Press conf, then enter

Correction CO.

Code DisplayMenu group Select menu item

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 44

45

0675 mV

Code Display Action Choices

Co. Enter polarization voltageSelect with key, edit number with

key, proceed with enter.

Select pressure unitSelect with keyProceed with enter

bar(kPa, PSi)

Process pressure correctionEnter process pressure. This value isused to correct oxygen saturation. Ithas no influence on concentrationmeasurement (Conc).Select with key, edit number with key, proceed with enter.

1.013bars

Enter salt correction(salinity)Select with key, edit number with key, proceed with enter.

00.00ppt*

* ppt (parts per thousand) - corresponds to g/kg

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 45

Page 24: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

46 Stratos e 2402 Oxy

ConfigurationCalibration modeAlarm settings

Calibration mode

Cal timer interval

Select Sensocheck

Delay

LED in HOLD mode

Code DisplayMenu group Select menu item

Calibration mode CA.

Alarm settings AL.

End:Press conf, then enter

End:Press conf, then enter

15

16

Alarm Alarm contact

The alarm contact is closed during normal operation(N/C). It opens in the case of alarm or power outage. As a result, a failure message is provided even in thecase of line breakage (fail-safe behavior). For contact ratings, see Specifications.

Error messages can also be signaled by a 22 mA outputcurrent ( see Pg 37,

The operating behavior of the alarm contact is shown onPg 80.

The alarm delay acts on the LED, the 22 mA signaland the alarm contact.

43, 78).

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 46

47

SAt(Conc)

Code Display Action Choices

CA. Specify calibration mode(Calibration to saturation or concentration)Select with keyProceed with enter

Cal timer intervalThe cal timer reminds you to calibrate in time.Select with key, edit number with key,proceed with enter.

0000 h(0 to 9999 h)

ON / OFFAL. Select Sensocheck(continuous monitoring of sensor)Select with keyProceed with enter

Delay for alarmSelect with key, edit number with

key, proceed with enter.

0010 s(xxxx s)

LED in HOLD modeSelect with key, edit number with

key, proceed with enter.

ON / OFF

Parameter setting Alarm HOLD

ON on flashes

OFF flashes off

LED state:

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 47

Page 25: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

48 Stratos e 2402 Oxy

ConfigurationLimit functionRelay 1

Relay / Controller rL.

Code DisplayMenu group Select menu item

End:Press conf, then enter

Contact function

Contact response

Enter switching point

Enter hysteresis

Delay

Relay 2 menu group

Controller menu group

L1.

L2.

Ct.

Use of relays

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 48

49

LiMIT(CtROL)

Code Display Action Choices

rL. Use of relays:• Limit function (LiMIT)• Controller (CtROL)Select with keyProceed with enter

Note: Selecting CtROL leads toController menu group Ct.

Lo (Hi)

Limit 1 function , see Fig. on Pg 51. Select with key, proceed with enter

Limit 1 contact responseN/C: normally closed contactN/O: normally open contactSelect with keyProceed with enter

N/C(N/O)

Limit 1 switching pointSelect with key, edit number with

key, proceed with enter.

0000 %(xxxx %)

Limit 1 hysteresisSelect with key, edit number with

key, proceed with enter.

0001 %(xxxx %)

Limit 1 delayThe contact is activated withdelay (deactivated without delay)Select with key, edit number with key,proceed with enter.

0010 sec(0 - 600 sec)

L1.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 49

Page 26: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

50 Stratos e 2402 Oxy

ConfigurationLimit functionRelay 2

Relay / Controller rL.

Code DisplayMenu group Select menu item

End:Press conf, then enter

Relay 1 menu group

Contact function

Contact response

Enter switching point

Enter hysteresis

Delay

Controller menu group

L1.

L2.

Ct.

Use of relays

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 50

51

Code Display Action Choices

L2. Hi (Lo)

Select Limit 2, see Fig. below. Select with keyProceed with enter

Limit 2 contact responseN/C: normally closed contactN/O: normally open contactSelect with keyProceed with enter

N/C(N/O)

Limit 2 switching pointSelect with key, edit number with

key, proceed with enter.

0500 %(xxxx %)

Limit 2 hysteresisSelect with key, edit number with

key, proceed with enter.

0001 %(xxxx %)

Limit 2 delayThe contact is activated withdelay (deactivated without delay)Select with key, edit number with key,proceed with enter.

0010 s(0 - 600 s)

Limit Lo

Hysteresis +

Switchingpoint

Contact

0

1

SignalLimit Hi

Hysteresis -

Signal

Switchingpoint

Contact

0

1

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 51

Page 27: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

52 Stratos e 2402 Oxy

ConfigurationController (for description see Pg 74)Setpoint. Neutral zone

Relay / Controller rL.

Code DisplayMenu group Select menu item

End:Press conf, then enter

Relay 1 menu group

Relay 2 menu group

Controller setpoint

Enter neutral zone

(P) Controller gain

(I) Reset time TR

(D) Rate time TD

Controller type PLC / PFC

PLC: Pulse length

PFC: Pulse frequency

HOLD behavior

L1.

L2.

Ct.

Use of relays

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 52

53

SetpointSelect with key, edit number with

key, proceed with enter.

Neutral zone (dead band)Select with key, edit number with

key, proceed with enter.

Controller: P-action componentSelect with key, edit number with

key, proceed with enter.

0100 %(xxxx %)

Code Display Action Choices

Ct.

0010 %(xxxx %)

0100 %(xxxx %)

Controller: I-action component (reset time). Select with , editnumber with , proceed withenter.

0000 sec(0–9999 sec)

Controller: D-action component (rate time). Select with , edit number with ,proceed with enter.

0000 sec(0–9999 sec)

Pulse length /Pulse frequencySelect with keyProceed with enter

PLC(PFC)

PLC: Pulse lengthSelect with key, edit number with

key, proceed with enter.

0010 sec(0–600 sec)

PFC: Pulse frequencySelect with key, edit number with

key, proceed with enter.

0060/min(0–180 /min)

Behavior during HOLDSelect with keyProceed with enter

Y Last(Y Off)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 53

Page 28: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Stratos e 2402 Oxy54

ConfigurationControl of rinsing and cleaning systems

Rinsing / Cleaning system

Rinsing interval

Rinse duration

Contact response

Cleaning interval

Cleaningprobes

Pb.

Code DisplayMenu group Select menu item

rinse(EASYCLN)EASYCLN: see oppositepage

Code Display Action (rinse equipment) Selection

Pb. Control of:• Rinse equipment (rinse) • Cleaning system (EasyClean)Select with keyProceed with enter

000.0 h (xxx.x h)

Rinsing intervalSelect with key, edit numberwith key, proceed with enter

Rinse durationSelect with key, edit numberwith key, proceed with enter

0060 s(xxxx s)

Contact typeSelect with keyProceed with enter

N/C(N/O)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 54

55

EASYCLN(rinse)

Code Display Action (cleaning system) Selection

Pb. • Cleaning system (EasyClean)Select with keyProceed with enter

Cleaning interval(EasyClean only)Select with key, edit numberwith key, proceed with enter

000.0 h(xxx.x h)

Off(On)

Lock cleaning interval*

On: The Stratos only starts acleaning interval if the meas-ured value lies within the toler-ated range (Limit Lo/Limit Hi).

*“Lock cleaning interval” function:

Limit Hi

Limit Lo

Locked

t

ppmLocked

The Stratos only starts a cleaning interval if the measuredvalue lies within the toleratedrange (Limit Lo/Limit Hi).(For limit setting, refer to Pages 49, 51).

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 55

Page 29: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

56 Stratos e 2402 Oxy

Parameter set 1/2

Display Action Remark

Select parameter setSelect with keyProceed with enter

Switch between parameter setsPress conf key,enter code 7654Select with key, edit number with key, proceed with enter.

Wrong settings changethe measurementproperties!

If an invalid code isentered, the Stratosreturns to measuringmode.

Welcome text is dis-played for approx.3 sec

Since the complete device config-uration is changed in one step,there is a security prompt(No/Yes). When pressing enter directly, theselection is not stored.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 56

57

Two complete parameter sets are stored in the EEPROM. As delivered, the two sets are identical but can be edited. Note:Fill in your configuration data on the following pages.

Code. Parameter Default setting

o1. Sensor type Ao1. %, mg/l, ppm %o1. 0/4-20 mA 4-20 mAo1. Current beginning 0000 %o1. Current end 0500 %o1. Filter time 0 so1. 22mA signal OFFo1. Hold behavior Lasto1. Fix current 021.0 mAo2. Unit °C / °F °Co2. Temp probe 22 NTCo2. 0/4...20mA 4-20 mAo2. Current beginning 000.0 °Co2. Current end 100.0 °Co2. Filter time 0 so2. 22mA signal OFFo2. Hold behavior Lasto2. Fix current 021.0 mACo. Polariz. voltage 675 mVCo. Pressure unit barCo. Pressure 1,013 barsCo. Salinity 00.00 mg/lCA. Cal mode SatCA. Cal interval 0000 hAL. Sensocheck OFFAL. Alarm delay 0010 secAL. LED Hold off

Code. Parameter Default setting

rL. Relay function LimitL1. Contact function LoL1. Contact response N/CL1. Switching point 0000 %L1. Hysteresis 0001 %L1. Delay 0010 secL2. Contact function HiL2. Contact response N/CL2. Switching point 0500 %L2. Hysteresis 0001 %L2. Delay 0010 secCt. Setpoint 0100 %Ct. Neutral zone 0010 %Ct. P action 0100 %Ct. I action 0000 secCt. D action 0000 secCt. PLC/PFC controller PLCCt. Pulse length 0010 secCt. Pulse frequency 0060 /minCt. Hold behavior LastPb. Probe selection rinsePb. Rinsing interval 000.0 hPb. Rinse duration 0060 secPb. Contact type N/CPb. Cleaning interval 000.0 hPb. Lock interval Off

Default settings of parameter sets

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 57

Page 30: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

58 Stratos e 2402 Oxy

Code. Parameter Setting

o1. Sensor type ____________ ____________o1. %, mg/l, ppm ____________ ____________o1. 0/4–20 mA ____________ ____________o1. Current beginning ____________ ____________o1. Current end ____________ ____________o1. Filter time ____________ ____________o1. 22mA signal ____________ ____________o1. Hold behavior ____________ ____________o1. Fix current ____________ ____________

o2. Unit °C / °F ____________ ____________o2. Temp probe ____________ ____________o2. 0/4–20mA ____________ ____________o2. Current beginning ____________ ____________o2. Current end ____________ ____________o2. Filter time ____________ ____________o2. 22mA signal ____________ ____________o2. Hold behavior ____________ ____________o2. Fix current ____________ ____________

Co. Polarization voltage ____________ ____________Co. Pressure unit ____________ ____________Co. Pressure ____________ ____________Co. Salinity ____________ ____________

CA. Cal mode ____________ ____________CA. Cal interval ____________ ____________

AL. Sensocheck ____________ ____________AL. Alarm delay ____________ ____________AL. LED Hold ____________ ____________

Parameter set – user settings

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 58

59

Code. Parameter Setting

rL. Relay function ____________ ____________

L1. Contact function ____________ ____________L1. Contact response ____________ ____________L1. Switching point ____________ ____________L1. Hysteresis ____________ ____________L1. Delay ____________ ____________

L2. Contact function ____________ ____________L2. Contact response ____________ ____________L2. Switching point ____________ ____________L2. Hysteresis ____________ ____________L2. Delay ____________ ____________

Ct. Setpoint ____________ ____________Ct. Neutral zone ____________ ____________Ct. P action ____________ ____________Ct. I action ____________ ____________Ct. D action ____________ ____________Ct. PLC/PFC controller ____________ ____________Ct. Pulse length ____________ ____________Ct. Pulse frequency ____________ ____________Ct. Hold behavior ____________ ____________

Pb. Probe selection ____________ ____________Pb. Rinsing interval ____________ ____________Pb. Rinse duration ____________ ____________Pb. Contact type ____________ ____________Pb. Cleaning interval ____________ ____________Pb. Lock interval _____________ _____________

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 59

Page 31: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

60 Stratos e 2402 Oxy

CalibrationCalibration adjusts the device to the sensor.

Activate with cal

Enter mode code• Zero point: 1001• Saturation/Concentration: 1100Edit parameter with and ,confirm/continue with enter.(End with cal enter.)

During calibration the Stratos remains inthe Hold mode for reasons of safety.Output current is frozen (last value orpreset fixed value, depending on config-uration), limit and alarm contacts areinactive. The controller is in the config-ured state. Sensoface is off, mode indi-cator “Calibration” is on.

Activate

Hold

Input errors The calibration parameters are checkedduring the input. In the case of an incorrect input ”Err” isdisplayed for approx. 3 sec. The incorrect parameters cannot bestored. Input must be repeated.

End End with cal.The measured value and Hold aredis-played alternately, “enter” flashes.Press enter to end the Hold mode.The measured value is displayed. The output current remains frozen foranother 20 sec (HOLD icon on, “hour-glass” flashes).

HOLD icon

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 60

61

Calibration

It is always recommended to calibrate in air. Compared to water, air is a calibration medium which is easy tohandle, stable, and thus safe. In the most cases, however, thesensor must be dismounted for a calibration in air.When dealing with biotechnological processes which requiresterile conditions, the sensor cannot be removed for calibration.Here, calibration must be performed with aeration directly in theprocess medium (e.g. after sterilisation).In the field of biotechnology, for example, often saturation ismeasured and calibration is performed in the medium for rea-sons of sterility. For other applications where concentration is measured (watercontrol etc.), calibration in air has been proved useful.

Common combination: process variable / calibration mode

Measurement CalibrationSaturation WaterConcentration Air

The calibration procedures for these two common applicationsare described on the following pages. Of course, other combina-tions of process variable and calibration mode are possible.

Note:When a 2-point calibration is required, the zero point calibrationshould be performed prior to saturation or concentration cali-bration, resp.All calibration procedures must be performed by trained person-nel.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 61

Page 32: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

62 Stratos e 2402 Oxy

Display Action Remark

Enter relative humiditySelect with key, edit number with key, proceed with enter.

Enter calibration pressureSelect with key, edit number with key, proceed with enter.

Default for calibrationpressure is theprocess pressure con-figured

Calibration to saturation (SAT)

Activate calibration(Press cal.)Enter mode code 1100Select with key, edit number with key, proceed with enter.

Place sensor in calibration mediumStart with enter

Automatic drift checkDisplay of sensor current (relatedto 25 °C and 1013 mbars normalpressure) and measuring tempera-ture.

The drift check might take sometime.

Drift check can bestopped after > 10 secby pressing cal (accu-racy reduced).

SAT or Conc calibra-tion is selected duringconfiguration.

If an invalid code isentered, the Stratosreturns to measuringmode.

Default for relativehumidity in aqueousmedia:rH = 100 %(in air approx. 50 %)

Stratos is in the Holdmode

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 62

63

Display Action Remark

Enter desired value for saturation

Select with key, edit number with key, proceed with enter.

Display new slope and zero (related to 25°C and 1013 mbars). End calibration with enter.

Default: last valueentered

Place sensor in processThe percent saturation is shown inthe main display alternately with“Hold”; “enter” flashes.End with enter.

After end of calibra-tion, the outputsremain in Hold modefor approx. 20 sec.

New calibration: Press cal key.

Information on saturation calibration (SAT)• The calibration medium must be in equilibrium with air (percent saturation for

water is 100 %).Oxygen exchange between water and air is very slow. To speed up the adjustment processes, make sure that there is a steady medium flow during calibration.

• If the percent saturation is known from a simultaneous measurement, it can be entered manually.

• For 2-point calibration, perform zero point calibration first.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 63

Page 33: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

64 Stratos e 2402 Oxy

Display Action Remark

Enter relative humidity(Select with key, edit number with key, proceed with enter.)

Enter calibration pressure(Select with key, edit number with key, proceed with enter.)

Default for calibrationpressure is normalpressure 1.013 bars.

Calibration to concentration (Conc)

Activate calibration(Press cal.)Enter mode code 1100(Select with key, edit number with key, proceed with enter.)

Place sensor in airStart with enter

Automatic drift checkDisplay of input current (relatedto 25 °C and 1013 mbars) and measuring temperature.

The drift check might take sometime.

Drift check can bestopped after > 10 sec by pressing cal (accu-racy reduced).

SAT or Conc calibra-tion is selected duringconfiguration.

If an invalid code isentered, the Stratosreturns to measuringmode.

Default for relativehumidity in air:rH = 50 %

Stratos is in the Holdmode

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 64

65

Display Action Remark

Enter default for concentration(Select with key, edit number with key, proceed with enter.)

Display of new slope and zero (related to 25 °C and 1013 mbars)

Press enter to end concentrationcalibration.

Default value is calcu-lated from rel. humidi-ty, cal pressure, andcal temperature.

(The unit of measure-ment, ppm or mg/l, ...is preset during config-uration.)

Place sensor in processThe new value is shown in themain display alternately with“Hold”; “enter” flashes.End with enter.

After end of calibra-tion, the outputsremain in Hold modefor approx. 20 sec.

New calibration: Press cal key.

Information on concentration calibration (Conc)Calibration in air. This calibration method is recommended when the sensor canbe removed for calibration. Air has a stable oxygen content. Therefore the adjust-ment processes during calibration run more quickly. • For 2-point calibration, perform zero point calibration first.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 65

Page 34: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

66 Stratos e 2402 Oxy

Zero point calibration

Zero point calibration

The sensor models SE 704, SE 705 and InPro6000 have a verylow zero point current. Therefore, a zero point calibration isonly recommended for measurement of oxygen traces. If a zeropoint calibration is performed, the DO sensor should remain forat least 10 to 30 minutes in the calibration medium in order toobtain stable, non-drifting values. During zero point calibration, a drift check is not performed.Zero point current of a properly functioning sensor is notablyless than 0.5 % of air current. The display (secondary: meas-ured value, main: entered value) does not change until aninput current is entered for the zero point.When measuring in an oxygen-free medium, the displayed cur-rent can be taken directly.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 66

67

Display Action Remark

Main display: Zero point current; store with enteror correct with arrow keysand then store with enter.Lower display: Sensor currentmeasured

Display of slopeDisplay of new zero point cur-rentEnd calibration with enter,place sensor in process

New calibration: Press cal key.

Activate calibration(press cal key)Enter mode code 1001Select with key, edit number with key, proceed with enter.

Zero

Place sensor in oxygen-free medium

The oxygen value is shown inthe main display alternatelywith “Hold”; “enter” flashes.Stop Hold with enter.

After end of calibra-tion, the outputsremain in Hold modefor approx. 20 sec.

Stratos is in the Holdmode.If an invalid code isentered, the Stratosreturns to measuringmode.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 67

Page 35: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

During product calibration the sensor remains in the process. Themeasurement process is only interrupted briefly.

Procedure: During sampling the currently measured value is stored inthe Stratos. The Stratos immediately returns to measuring mode. Thecalibration mode indicator flashes and reminds you that calibration hasnot been terminated. The comparison value is measured on the site,e.g. using a portable DO meter in a bypass. This value is then enteredin the Stratos. The new values for slope and zero are calculated fromthe stored value and the comparison value. From the measured value,the Stratos automatically recognizes whether a new slope or zero mustbe calculated (above approx. 5 % saturation: slope, below: zero). If the sample is invalid, you can take over the measured value storedduring sampling instead of the comparison value. In that case the oldcalibration values remain stored. Afterwards, you can start a new product calibration.The following describes a product calibration with slope correction. Aproduct calibration with zero correction is performed correspondingly.

68 Stratos e 2402 Oxy

Product calibrationCalibration by sampling

Display Action Remark

Product calibration step 1:Activate calibration(press cal key)Enter mode code 1105(Select with key, edit number with key, proceed with enter.)

Store value.Proceed with enter

The type of productcalibration (SAT orConc) is selected dur-ing configuration(measurement proce-dure). If an invalidcode is entered, theStratos returns tomeasuring mode.

Now the sample canbe measured in thelab. The Stratos is inmeasuring mode.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 68

69

Display Action Remark

Measuring mode

Product calibration step 2:When the comparison value hasbeen determined, call up theproduct calibration once more(cal key, mode code 1105).

Display (approx. 3 sec)

Enter comparison value. Confirm with enter.

The new slope iscalculated.

From the flashingCAL mode indicatoryou see that product calibrationhas not been termi-nated.

Display of new slope and zero (related to 25°C and 1013 mbars)End calibration with enter

The measured value is shown inthe main display alternately with“Hold”; “enter” flashes.End with enter.

After end of calibra-tion, the outputsremain in Hold modefor approx. 20 sec.

New calibration: Press cal key.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 69

Page 36: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

70 Stratos e 2402 Oxy

Temperature probe adjustment

Display Remark

In measuring mode the main display shows the configuredprocess variable (%, mg/l, or ppm), the secondary displayshows the temperature. During calibration you can returnto measuring mode by pressing the cal key, during configu-ration by pressing the conf key.(Waiting time for measured value stabilization is approx.20 sec).

Measurement

Display Action Remark

Measure the temperature ofthe process medium using anexternal thermometer. Entermeasured temperature value:Select with key, edit number with key, proceed with enter key.End adjustment with enter.HOLD will be deactivatedafter 20 sec.

Activate calibration(Press cal,enter mode code 1015)Select position with key, edit number with key, proceed with enter key.

Ready for calibration

Wrong settingschange the meas-urement properties!If an invalid code is entered, theStratos returns tomeasuring mode.

Default:Current value ofsecondary display.

Stratos is in Holdmode (Display forapprox. 3 sec)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 70

71

Display of output currentsPress enter while in measuring mode.The current at output 1 is shown in the main display, thecurrent at output 2 in the secondary display.After 5 sec the Stratos returns to measuring mode.

Diagnostics functions

Display Remark

Display of calibration data (Cal Info)Press cal key while in measuring mode and enter code0000. The slope is shown in the main display, the zeropoint current in the secondary display.After 20 sec the Stratos returns to measuring mode (immediate return at pressing cal).

Display of sensor current(Sensor monitor)Press conf key while in measuring mode and enter code2222. The (uncompensated) sensor current is shown in themain display, the measuring temperature in the secondarydisplay. Press enter to return to measurement.

Display of last error message(Error info)Press conf key while in measuring mode and enter code0000. The last error message is displayed forapprox. 20 sec. After that the message will be deleted.(immediate return to measurement at pressing enter).

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 71

Page 37: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Stratos e 2402 Oxy72

Relay test (manual test of contacts)• Press conf key, enter code 5557.The relays are frozen. This state is indicated in the display.The 4 digits in the display correspond to the 4 relays (as onterminal plate):1st digit: R12nd digit: R23rd digit: AL4th digit: CLNFunction test using arrow keys – see left column. When exiting the function (enter), the relays are set corre-sponding to the measured value.

Select a relay

Specify current at output 1• Press conf key, enter code 5555.The current indicated in the main display for output 1 canbe edited.Select with key, edit number with key, proceed with enter.The actually measured current is shown in the secondarydisplay. The Stratos is in Hold mode.Press enter to return to measurement (Hold remainsactive for another 20 sec).

Diagnostics functionsThese functions are used for testing the connected peripherals.

Display Action / Remarks

Specify current at output 2• Press conf key, enter code 5556.The current indicated in the main display for output 2 canbe edited.Select with key, edit number with key, proceed with enter. The actually measured current is shown in the secondarydisplay. The Stratos is in Hold mode. Press enter to return to measurement.

Test 0/1

Return tomeasurement

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 72

73

The arrows indi-cate which relay(valve) is active:

Controller characteristic

+100 %

-100 %

Relay 2 active(Meas. value >setpoint)

Relay 1 active(Meas. value <setpoint)

Setpoint

Controller output

Momentary controller output (adjusted value has not beenstored yet)

Controller test (manual specification of controlleroutput)• Press conf key, enter code 5559.After function activation “Ctrl” is displayed for approx.3 sec. With controller turned off, “OFF” is displayed inaddition, then return to measuring mode.The function is used to start up control loops or check theactuators. For bumpless changeover to automatic operation (exitingthis function), configure an I-action component (resettime).

Specify value:Select with key, edit number with key, proceed with enter.

The Stratos is in Hold mode.Press enter to return to measurement (Hold remains active for another 20 sec).

Controller output -100 to 0 %: Relay 2 active

Controller output 0 to +100 %: Relay 1 active

Display Action / Remarks

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 73

Page 38: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

74 Stratos e 2402 Oxy

P controllerApplication in integrating systems (e.g. closed tank, batch processes).

PI controllerApplication in non-integrating systems(e.g. aeration basin).

PID controller The additional derivative action compensates for measurementpeaks.

Controller characteristic

Controller functionsPID controller

Setpoint

Controller output YP [%]

+100

-100

O2[%]

Neutral zone YP =0

Note:In Hold mode the controller output actsas configured (Y = const. or Y = 0).

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 74

75

Controller equations

Controller output Y = YP + YPdt + TD dt

I action D action

dYP1TR

P-action

100 %

50 %

XwProcess variables: % / ( mg/l, ppm)

KC = 500 % KC = 200 %

KC =100 %

KC =50 %

10 / (1.00) 20 / (2.00) 30 / (3.00) 40 / (4.00)

Deviation

50 / (5.00)

Proportional action YP

Setpoint - Meas. valueConstant

YP= *KC

with:YP Proportional actionTR Reset time [s]TD Rate time [s]KC Controller gain [%]Constant 50 % (for %O2, % Air)

5.00 mg/l (for mg/l)5.00 ppm (for ppm)

Con

trol

ler

outp

ut Y

Proportional action (Gradient KC [%])

Neutral zone (Y=0)Tolerated deviation from setpoint. The setting “010%”, for example, permits a deviation of ± 5 % from the desired value without activating the controller.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 75

Page 39: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

76 Stratos e 2402 Oxy

Controller functionsPulse length / pulse frequency controller

Pulse length controller (PLC)The pulse length controller is used to operate a valve as anactuator. It switches the contact on for a time that depends onthe controller output. The period is constant. A minimum ONtime of 0.5 sec is maintained even if the controller output takescorresponding values.

Pulse frequency controller (PFC)The pulse frequency controller is used to operate a frequency-controlled actuator. It varies the frequency with which the contacts are switched on. The maximum pulse frequency [pulses/min] can be defined. It depends on the actuator.The contact ON time is constant. It is automatically calculatedfrom the user-defined maximum pulse frequency:

Output signal (switching contact) of pulse frequency controller

ON time

Pulse frequency (Y = 20%)

1

0

1

0Pulse frequency (Y = 80%)

Output signal (switching contact) of pulse length controller

ON time (Y = 80 %)

ON time (Y = 20 %)

Pulse length

10

10

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 76

77

Operation with automatic cleaning systemThe Stratos allows connection of an automatic cleaning sys-tem. The cleaning cycle is activated according to the cleaninginterval defined during configuration (Pg 55).

17

18

Clean

e.g. splash cleaning

Power supply

17

18

Clean

7

8

Control -

+

Automaticcleaningsystem

Connecting a rinsing systemThe “Clean” contact can be used to connect a simple splashcleaning system. Rinse time and rinsing interval are definedduring configuration (Pg 54).

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 77

Page 40: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

78 Stratos e 2402 Oxy

Ala

rm c

onta

ct

Red

LED

Out

1 (

22 m

A)

Out

2 (

22 m

A)

Error messages (error codes)

SAT rangeSensor defectiveWrong sensor connectedMeasurement range exceeded

Error DisplayProblem Possible causes

ERR 01 Measuredvalueflashes

xxx

Conc rangeSensor defectiveWrong sensor connectedMeasurement range exceeded

ERR 02 Measuredvalueflashes

xxx

System errorConfiguration or calibration datadefective; completely reconfigure andrecalibrate the deviceMemory error in device program(PROM defective)

ERR 98 “ConF”flashes

xxx x

Factory settingsEEPROM or RAM defectiveThis error message only occurs in thecase of a complete defect. The Stratosmust be repaired and recalibrated atthe factory.

ERR 99 “FAIL”flashes

xxx x

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 78

79

Ala

rm c

onta

ct

Red

LED

Out

1 (2

2 m

A)

Out

2 (

22 m

A)

Current output 1Current below 0 (3.8) mA

ErrorSymbol(flash-ing)

Problem Possible causes

ERR 11 xxx

Temperature probeOpen or short circuitTemperature range exceeded

ERR 03 xxx x

Current output 1Current above 20.5 mA

ERR 12 xxx

Current output 1Current span too small / too large

ERR 13 xxx

Current output 2Current below 0 (3.8) mA

ERR 21 xxx

Current output 2Current above 20.5 mA

ERR 22 xxx

Current output 2Current span too small / too large

ERR 23 xxx

Rinsing probeCommunication error

ERR 41 xxx

SensocheckSensor: Connecting cable defective

ERR 33 xxx

x

• Zero error, Sensoface active, see Pg 82

• Slope error, Sensoface active, see Pg 82

• Response time exceeded, Sensoface active, s. Pg 82

• Calibration interval expired, Sensoface active, s. Pg 82

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 79

Page 41: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

80 Stratos e 2402 Oxy

Ala

rm c

onta

ct

Cle

anin

gco

ntac

t

LED

Tim

e ou

tOperatingstate

Measurement

Operating states

Rel.1

/2

Con

trol

ler

Rel.1

/2

Lim

it va

lue

Out

2

Out

1

Cal Info(cal) 0000

20sec

Error Info(conf) 0000

20sec

Calibration(cal) 1100

Temp adjustment(cal) 1015

Calibration error messages

Slope out of rangeWrong calibration values specified (relative humidity,pressure, saturation, concentration)

Symbol flashes:Problem Possible causes

Calibration aborted after 12 minutesSensor defective or dirty•No electrolyte in the sensor•Sensor cable insufficiently shielded or defective•Strong electric fields influence the measurement•Temperature fluctuation of calibration solution

In addition “CAL Err” is flashing

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 80

81

Product cal 1(cal) 1105

Product cal 2(cal) 1105

Configuration(conf) 1200

20min

Parameter set 1/2(conf) 7654

20min

Sensor monitor(conf) 2222

20min

Current source 1(conf) 5555

20min

Current source 2(conf) 5556

20min

Relay test(conf) 5557

20min

Manual controller(conf) 5559

20min

Rinsing function

HOLD input

Ala

rm c

onta

ct

Cle

anin

gco

ntac

t

LED

Tim

e ou

tOperatingstate

Rel.1

/2

Con

trol

ler

Rel.1

/2

Lim

it va

lue

Out

2

Out

1

Explanation: active

as configured (Last/Fix or Last/Off)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 81

Page 42: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

82 Stratos e 2402 Oxy

Responsetime Cal timerSlope Zero point

Sensoface(Sensocheck must have been activated during configuration.)

adm.range

The little smiley in the display (Sensoface) alerts for sensorproblems (defective cable, maintenance required). The permitted calibration ranges and the conditions for afriendly, neutral, or sad Sensoface are summarized in the fol-lowing chart. Additional icons refer to the error cause. Replace membrane module or filling solution, if required.

Sensors Type A (SE 704, SE 705, SE 706, InPro6000–6800)

25 to 130 nA -2 to +2 nA max. 720 s

> 35 to < 90 nA > -0.3 to < 0.3 nA ≤ 300 s ≤ 80 %

30 to 35 nAor90 to 110 nA

-0.6 to -0.3 nAor+0.3 to +0.6 nA

300 to 600 s 80 to ≤ 100 %

< 30 nA or >110 nA

< -0.6 nA or >+0.6 nA

>600 s Timer expired

NoteThe worsening of a Sensoface criterion leads to the devaluationof the Sensoface indicator (Smiley becomes “sad”). Animprovement of the Sensoface indicator can only take placeafter calibration or removal of a sensor defect.

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 82

83

Slope Zero point Responsetime Cal timer

Sensor Type B (SE 707, InPro6900)

adm.range

200 to 550 nA -2 to +2 nA max. 720 s

> 250 to < 500 nA > -0.5 to < 0.5 nA < 300 s < 80 %

225 to 250 nAor500 to 525 nA

-1.0 to -0.5 nAor+0.5 to +1.0 nA

300 to 600 s 80 to ≤ 100 %

< 225 nA or >525 nA

< -1.0 nA or >+1.0 nA

> 600 s Timer expired

SensocheckContinuously monitors the sensor and leads for short circuits oropen circuits. Critical values make the Sensoface “sad” and thecorresponding icon flashes:

The Sensocheck message is also output as error message Err33. The alarm contact is active, the red LED is lighted, outputcurrent 1 is set to 22 mA (when configured correspondingly).Sensocheck can be switched off during configuration (thenSensoface is also disabled). Exception: After calibration a Smileyis always displayed for confirmation.

Thermometer and Sensoface:Temperature out of concentration or saturation range

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 83

Page 43: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

84 Stratos e 2402 Oxy

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 84

85

Appendix

Product line and accessories

Devices Order No.

Stratos e 2402 Oxy 2402 Oxy

Mounting accessoriesPipe-mount kit ZU 0274Panel-mount kit ZU 0275Protective hood ZU 0276

For more information concerning our sensors and fittings product line, please refer to our “Sensors, Fittings, Accessories” catalog:Download at http://www.knick.de or request catalog: Phone: +49 (0)30 - 801 91 - 290Fax: +49 (0)30 - 801 91 - 200Mail: [email protected]

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 85

Page 44: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

86 Stratos e 2402 Oxy

DO input Sensor Type A: SE 704 / 705 / 706InPro6000–6800

Sensor Type B: Inpro6900, SE 707

Measuring current -2 to 1800 nA,Resolution 0.05 nA(with Vpol = 800 mV and Vref = 200 mV)

Saturation (-10 to 80 °C) 0 to 500 %Meas. error1,2,3 0.5 % meas.val. + 0.5 %

Concentration (-10 to 80 °C) 0.00 to 50.00 mg/l0.00 to 50.00 ppm

Meas. error1,2,3 0.5 % meas.val. + 0.05 mg/l or 0.05 ppm

Adm. guard current ≤ 20 µAPolarization voltage * 0 to 1000 mV,Process pressure * 0.000 to 9.999 bars

(to 999.9 kPa / to 145.0 psi)Salt correction * 00.00 to 45.00 g/kg

Sensor standardizationOperating modes * DO saturation (automatic)

DO concentration (automatic)Product calibrationZero point calibration

Calibration range Zero point ± 2 nASensor Type A Slope 25 to 130 nA

(at 25 °C, 1013 mbars)Calibration range Zero point ± 2 nASensor Type B Slope 200 to 550 nA

(at 25 °C, 1013 mbars)Calibration timer * 0000 to 9999 hPressure correction * 0.000 to 9.999 bars / 999.9 kPa / 145.0 psi

Sensocheck Monitoring for short circuits / open circuits (can be disabled)

Sensoface Provides information on the sensor conditionEvaluation of zero/slope, response, calibration interval, Sensocheck

Specifications

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 86

87

Temperature NTC 22 kOhms / NTC 30 kOhms*input 2-wire connection, adjustableMeasurement range -20.0 to +150.0 °C / -4 to +302 °FAdjustment range 10 KResolution 0.1 °C / 1 °FMeas. error1,2,3 < 0.5 K (<1 K at >100°C)

HOLD input Galv. separated (OPTO coupler)Function Switches Stratos to HOLD modeSwitching voltage Inactive 0 to 2 V (AC/DC)

Active 10 to 30 V (AC/DC)

CONTROL input Galv. separated (OPTO coupler)Function Control input for automatic cleaning systemSwitching voltage Inactive 0 to 2 V (AC/DC)

Active 10 to 30 V (AC/DC)

Output 1 0/4 to 20 mA, max. 10 V, floating(galv. connected to output 2)

Process variable * DO saturation/DO concentration Overrange * 22 mA in the case of error messagesOutput filter * Low-pass, filter time constant 0 to 120 sMeas. error 1 < 0.3 % current value + 0.05 mA

Start/end of scale Configurable within selected rangeAdm. span 5 to 500 % / 0.5 to 50 mg/l (ppm)

Output 2 0/4 to 20 mA, max. 10 V, floating(galv. connected to output 1)

Process variable TemperatureOverrange * 22 mA in case of temp error messagesOutput filter * Low-pass, filter time constant 0 to 120 sMeas. error 1 < 0.3 % current value + 0.05 mA

Start/end of scale* -20 to +150 °C / -4 to +302 °FAdm. span 20 to 170 K (68 to 338 °F)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 87

Page 45: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

88 Stratos e 2402 Oxy

Alarm contact Relay contact, floatingContact ratings AC < 250 V / < 3 A / < 750 VA

DC < 30 V / < 3 A / < 90 WContact response N/C (fail-safe type)Response delay * 0000 to 0600 s

Limit values Output via relay contacts R1, R2 Contacts R1, R2 floating but inter-connected

Contact ratings AC< 250 V / < 3 A / < 750 VADC< 30 V / < 3 A / < 90 W

Contact response* N/C or N/OResponse delay * 0000 to 0600 sSwitching points* Within selected rangeHysteresis* 000.0 to 050.0 % / 00.00 to 05.00 mg/l (ppm)

PID process Output via relay contacts R1, R2 controller (Relay R1: below setpoint,

Relay R2: above setpoint)Setpoint specification* 0 to 500 % / 0.00 to 50.00 mg/l (ppm)

Neutral zone* 000.0 to 050.0 % / 00.00 to 05.00 mg/l (ppm)P-action component* Controller gain Kc: 0010 to 9999 % I-action component* Reset time Tr: 0000 to 9999 s

(0000 s = no integral action)D-action component* Rate time Td: 0000 to 9999 s

(0000 s = no derivative action)Controller type* Pulse length controller or pulse frequency controllerPulse period* 0001 to 0600 s, min. ON time 0.5 s

(pulse length controller)Max. pulse frequency* 0001 to 0180 min-1

(pulse frequency controller)

Specifications

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 88

89

Cleaning function* Relay contact, floating, for controlling a simple rinsing system or an automatic cleaning system (EasyClean)

Contact ratings AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response* N/C or N/OInterval * 000.0 to 999.9 h

(000.0 h = cleaning function switched off)Rinse duration* 0000 to 1999 s

Display LC display, 7-segment with iconsMain display Character height 17 mm, unit symbols 10 mmSecondary display Character height 10 mm, unit symbols 7 mmSensoface 3 status indicators (friendly, neutral, sad Smiley)

Mode indicators 5 status bars“meas”, “cal”, “alarm”, “cleaning”, “config”18 further icons for configuration and messages

Alarm indication Red alarm LED in case of alarm or HOLD (user defined)

Keypad 5 keys

Service functionsCurrent source Current specifiable for output 1 and 2 (00.00 to 22.00 mA)Manual controller Controller output entered directly (start of control process)Device self-test Automatic memory test (RAM, FLASH, EEPROM)Display test Display of all segmentsLast Error Display of last error occurredSensor monitor Display of direct, uncorrected sensor signalRelay test Manual control of the four switching contacts

Parameter sets Two selectable parameter sets for different applications

Data retention Parameters and calibration data > 10 years (EEPROM)

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 89

Page 46: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Stratos e 2402 Oxy90

Protection against Protective separation of all extra-low-voltage circuits againstelectrical shock mains by double insulation as per EN 61010-1

Power supply 24 (-15%) to 230 V AC/DC (+10%); approx. 5 VA, 2.5 WAC: 45 to 65 Hz; Overvoltage category II, Class II

Nominal operating conditionsAmbient temperature -20 to +55 °CTransport/Storage temp. -20 to +70 °CRelative humidity 10 to 95 % not condensingPower supply 24 (-15%) to 230 V AC/DC (+10%)Frequency for AC 45 to 65 Hz

EMC EN 61326Emitted interference Class B (residential environment)

Class A for mains supply > 60 V DCImmunity to interference Industrial environment

Explosion protectionFM: NI Class I Div 2 Group A, B, C & D, T4 Ta = 55 °C; Type 2

NI Class I Zone 2 Group IIC, T4 Ta = 55°C; Type 2

CSA: Class I Div 2 Groups A, B, C and D, T4Ex nA IIC T4

Enclosure Molded enclosure made of PBT (polybutylene terephtalate)Color Bluish gray RAL 7031Assembly • Wall mounting

• Pipe mounting: dia 40 to 60 mm, 30 to 45 mm• Panel mounting, cutout to DIN 43 700

Sealed against panelDimensions H 144 mm, B 144 mm, T 105 mmIngress protection IP 65 / NEMA 4XCable glands 3 breakthroughs for cable glands M20x1.5

2 breakthroughs for NPT 1/2 “ or Rigid Metallic Conduit

Weight Approx. 1 kg

*) User-defined1) To IEC 746 Part 1, at nominal operating conditions2) ± 1 count3) Plus sensor error

Specifications

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 90

91

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 91

Page 47: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

92 Stratos e 2402 Oxy

Explosion protection

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 92

93

Warnings and notesto ensure safe operation

Warning: Do not disconnect equipment unless power hasbeen switched off.

Warning: Clean only with antistatic moistened cloth.Warning: Substitution of components may impair suitability

for hazardous locations.• The equipment shall be installed and protected from mechanical

impact and ultraviolet (UV) sources.• Clean only with a moistened antistatic cloth as potential

electrostatic hazard may exist. Service equipment only with conductive clothing, footwear and personal grounding devicesto prevent electrostatic accumulation.

• Internal grounding provisions shall be provided for field wiring. Bonding between conduit shall be provided during installation, and all exposed non-current carrying metallic parts shall be bonded and grounded.

• Installation in a Class I, Division 2 or Class I, Zone 2 hazardouslocation shall be in accordance with the Canadian ElectricalCode (CEC Part 1) Section 18 Division 2 wiring methods.

• The equipment shall have a switch or circuit breaker in thebuilding installation (that is in close proximity to the equipment)that is marked as the disconnect switch.

• The enclosure Type 2 is only for indoor use.• The mains supply voltage fluctuations should not exceed

-15/+10 percent of the nominal supply voltage.• Do not use the equipment in a manner not specified in this

documentation.• Caution: Use supply wires suitable for 30 °C above ambient

and rated at least 250 V.• Caution: Use signal wires suitable for at least 250 V.

SPECIAL INSTRUCTIONS FOR FIELD REPRESENTATIVESNone

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 93

Page 48: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

94 Stratos e 2402 Oxy

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 94

95

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 95

Page 49: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

96 Stratos e 2402 Oxy

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 96

97

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 97

Page 50: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

2-point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6122 mA signal for error message . . . . . . . . . . . . . . 37, 43, 78Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Alarm contact . . . . . . . . . . . . . . . . . . . . . . . . . . . .46, 88Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78Operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80

Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Cal Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71Cal timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Calibration error messages . . . . . . . . . . . . . . . . . . . . . . 80Calibration to concentration (Conc) . . . . . . . . . . . . . . . 64Calibration to saturation (SAT) . . . . . . . . . . . . . . . . . . . 62Selection of calibration mode . . . . . . . . . . . . . . . . . . . . 47

Cleaning system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Lock cleaning interval . . . . . . . . . . . . . . . . . . . . . . . . . . 55

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Configuration steps . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Configuration: Correction . . . . . . . . . . . . . . . . . . . . . . . . . 44Polarization voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Salt correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Configuration: Alarm settings . . . . . . . . . . . . . . . . . . . . . . 46Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47LED in HOLD mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Sensocheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Configuration: Calibration mode . . . . . . . . . . . . . . . . . . . 46Cal timer interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

98 Stratos e 2402 Oxy

Index

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 98

Configuration: Controller . . . . . . . . . . . . . . . . . . . . . . . . . 52Configuration: Limit function . . . . . . . . . . . . . . . . . . . . . . 48

Settings for relay 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Settings for relay 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Use of relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Configuration: Output 1 . . . . . . . . . . . . . . . . . . . . . . . . . . 30Measurement procedure . . . . . . . . . . . . . . . . . . . . . . . 30Output current during Error . . . . . . . . . . . . . . . . . . . . . 36Output current range . . . . . . . . . . . . . . . . . . . . . . . . . . 32Output signal during HOLD . . . . . . . . . . . . . . . . . . . . . 37Select sensor type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Time constant of output filter . . . . . . . . . . . . . . . . . . . 34

Configuration: Output 2 . . . . . . . . . . . . . . . . . . . . . . . . . . 38Output current during HOLD . . . . . . . . . . . . . . . . . . . . 42Output current range . . . . . . . . . . . . . . . . . . . . . . . . . . 39Temperature error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Temperature probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Temperature unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Time constant of output filter . . . . . . . . . . . . . . . . . . . 40

Configuration: Rinsing and cleaning systems . . . . . . . . . . 54Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Lock cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Rinsing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Control drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94,96Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Controller equations . . . . . . . . . . . . . . . . . . . . . . . . . . . 75Controller test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73Pulse length / pulse frequency controller . . . . . . . . . . . 76

99

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 99

Page 51: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

CSA Certificate of Compliance . . . . . . . . . . . . . . . . . . . . . 92Current source 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

Controller test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73Display of calibration data . . . . . . . . . . . . . . . . . . . . . . 71Display of last error message . . . . . . . . . . . . . . . . . . . . 71Display of output currents . . . . . . . . . . . . . . . . . . . . . . 71Display of sensor current . . . . . . . . . . . . . . . . . . . . . . . 71Relay test (manual test of contacts) . . . . . . . . . . . . . . . 72Specify output current . . . . . . . . . . . . . . . . . . . . . . . . . 72

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Division 2 wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15EC Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . 7Err . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Error info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

Calibration error messages . . . . . . . . . . . . . . . . . . . . . . 80Display of last error message . . . . . . . . . . . . . . . . . . . . 71

Explosion protection . . . . . . . . . . . . . . . . . . . . . . . . . . 90, 92Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

External activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25LED in HOLD mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Output signal during HOLD . . . . . . . . . . . . . . . . . . 37, 43

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Limit function (LiMIT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Lock cleaning interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Mode codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104Mounting plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

100 Stratos e 2402 Oxy

Index

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 100

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Output current range . . . . . . . . . . . . . . . . . . . . . . . . . 32, 38Output filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35, 41Output signal for HOLD . . . . . . . . . . . . . . . . . . . . . . . . . . 37Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Panel-mount kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Parameter set 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

Default . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Select parameter set . . . . . . . . . . . . . . . . . . . . . . . . . . . 56User settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

PID controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74Pipe mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Pipe-mount kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Polarization voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Product line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Protective hood . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Protective wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Pulse frequency controller . . . . . . . . . . . . . . . . . . . . . . . . . 76Pulse length controller . . . . . . . . . . . . . . . . . . . . . . . . . . . 76Relay 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Relay 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Relay test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Rinsing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Safety functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

CSA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

101

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 101

Page 52: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

Self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Sensocheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23, 83

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Sensoface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23, 82

Permitted calibration ranges . . . . . . . . . . . . . . . . . . . . . 82Sensor monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Sensor type selection . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86Temperature probe adjustment . . . . . . . . . . . . . . . . . . . . . 70Temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Terminal assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Time constant output filter . . . . . . . . . . . . . . . . . . . . . 35, 41Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Typical wirings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20VP cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Wiring examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Zero point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66ZU 0274 pipe-mount kit . . . . . . . . . . . . . . . . . . . . . . . . . . 12ZU 0275 panel-mount kit . . . . . . . . . . . . . . . . . . . . . . . . . 13ZU 0276 protective hood . . . . . . . . . . . . . . . . . . . . . . . . . 12

102 Stratos e 2402 Oxy

BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 102 BE_2402OXY-CSA.QXD 22.07.2003 15:49 Uhr Seite 103

Page 53: Stratos e 2402 Oxy - Sensors | Barben Analyzer … · 2402 Oxy Instruction manual ... pharmaceutical industry, as well as in the field of environment, food processing, and sewage

TA-194.430-KNE03 20070810

Mode codesThe mode codes allow fast access to all functions

Calibration

Key + Code Description

Cal InfoDisplay of zero point and slope

0000

Zero point calibrationAdjusting the zero point (sensor)

1001

Product calibrationAdjusting the zero / slope (product)

1105

1015

Configuration

Error Info Display last error and erase0000

Configuration

Sensor monitorDisplay uncorrected sensor current (nA)

Parameter set 1/2 Selecting parameter set

1200

2222

7654

Key + Code Description

Slope calibrationAdjusting the slope (sensor)

1100

5555

5556

5557

5559

Current source 1 Output current 1 specified

Current source 2 Output current 2 specified

Relay test Manual test of contacts

Manual controllerManual specification of controller output

Temperature probeTemperature probe adjustment

Software version: 1.0