Studies of Small Scintillating Cells with Modified Geometries. Alexander Dyshkant for NICADD at NIU. Outline. LC scintillator based calorimeter with incorporated readout electronics: Electronic circuit board with surface mount Multi_Pixel Photon Counters (MPPC). - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/1.jpg)
Studies of Small Scintillating Cells with Modified Geometries
Alexander Dyshkant
for NICADD at NIU
![Page 2: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/2.jpg)
11.18.08 LCWS08 2
Outline• LC scintillator based calorimeter with incorporated readout
electronics:• Electronic circuit board with surface mount Multi_Pixel Photon Counters
(MPPC).• Array of the concave scintillating cells that is directly coupled to the
MPPC on the electronic circuit board.
• Current measurements with MPPC.• Concave cell studies:
– Traditional or flat vs concave cell.– Response MPPC to different diameter of dimples.– Distance impact on a MPPC response.– Response of the adjacent cells and cross talk estimation for the
concave cells with MPPC.
• Summary and Plans.
![Page 3: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/3.jpg)
11.18.08 LCWS08 3
Electronic Circuit Board with Array of Scintillating Cells
Array of scintillating cells directly coupled to the photo detectors that are surface mounted on a printed circuit board. The front-end electronics and all signal and
bias voltage traces for 64 channels are on the same board. Nowadays Fermilab electronics design based on the Minerva FEB 4TriP-t chips.
![Page 4: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/4.jpg)
11.18.08 LCWS08 4
Surface Mounted MPPC
View through the spacer opening. The MPPC active area is about 1 mm² (to the right).
![Page 5: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/5.jpg)
11.18.08 LCWS08 5
Array of Concave Scintillating Cells
Example of 5 x 5 concave scintillating cells that was fabricated as one unit.
![Page 6: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/6.jpg)
11.18.08 LCWS08 6
I_V for CPTA SiPM at room
0.01
0.1
1
10
100
1000
0 5 10 15 20 25 30 35
Bias Voltage, V
Cu
rre
nt,
nA
V-I at LN for SiPM from CPTA
0.01
0.1
1
10
100
1000
0 5 10 15 20 25 30 35
Bias Voltage, V
Cu
rre
nt,
nA
Output Current vs Bias Voltage
0.1
1
10
100
1000
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75
Bias Voltage, V
Curre
nt, n
A
Dark Current With Radioactive Source
V-I for MPPC-33-100C
0.001
0.01
0.1
1
10
100
1000
10000
1.00E-02 1.00E-01 1.00E+00 1.00E+01 1.00E+02
Bias Voltage, V
Cu
rren
t n
A
70.2F
Look at V-I
With low noise photo detector the current measurements are reasonable.
![Page 7: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/7.jpg)
11.18.08 LCWS08 7
Single Cell Preparation• For the cells with an area about 9 cm² and thickness about
200 mils scintillator EJ-200 was used.
• One side of the cell is flat and the other has a concave dimple in the center of the cell.
• The cell edges were painted using reflective coating EJ-510.
• Flat side of the cell was covered with VM2000.
![Page 8: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/8.jpg)
11.18.08 LCWS08 8
Scanner& Scanning• The photo detector was MPPC-10362-11-050C from Hamamatsu and
it was placed under the center of the cell.
• The very top surface of the photo detector was even with the bottom surface of the cell.
• The following scans were performed from one side to the opposite side through the center of the cell.
• A collimated radioactive source Sr90 was used.
• The factory recommended operating voltage was used for MPPC.
• The output current was measured with Kiethley 6485 picoammeter.
• During tests the temperature was measured with an accuracy 0.1ºF.
![Page 9: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/9.jpg)
11.18.08 LCWS08 9
Sr90
Cell
MPPC
Traditional Flat Cell
MPPS is directly coupledat the center of the cell.
xy
Current, nA
![Page 10: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/10.jpg)
11.18.08 LCWS08 10
Concave Cell
Sr90
Cell
MPPC
MPPS is directly coupledat the center of the cell.
x
y
Current, nA
![Page 11: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/11.jpg)
11.18.08 LCWS08 11
Dimensions of the Dimple and Their Meaning for the Concave Cell
Depth
Diameter
![Page 12: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/12.jpg)
11.18.08 LCWS08 12
MPPC-11 Response to Cells
0
15
30
45
60
75
90
105
0 5 10 15 20 25 30 35 40 45
Source Position, mm
No
rma
lize
d R
esp
on
se, %
0
15
30
45
60
75
90
105
19.4
15.6
13.7
15.5
Response of Cell with Different Diameter of Dimple
![Page 13: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/13.jpg)
11.18.08 LCWS08 13
Response of Cell with Different Diameter of Dimple
MPPC-33 Response to Cells
0
15
30
45
60
75
90
105
0 5 10 15 20 25
S ource Position, mm
Nor
mal
ized
Res
pon
se, % flat
larger
smaller
medium size
![Page 14: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/14.jpg)
11.18.08 LCWS08 14
General Considerations• Direct coupling of a photo detector with an effective active area of
about 1 mm² to a concave cell made of plastic scintillator with an about 9 cm² area provides a uniform response across the entire cell without the light yield loss compare to a flat cell (without dimple). This is possible in the case when the top surface of the photo detector is flush with the flat part of the cell concave side. Is such flush position of a photo detector optimal?
• In the case of the surface mount type of photo detector, it will pop out of a printed circuit board into the cell concavity and the cell response can be different.
• So, the vertical position of the photo detector inside the concave cell and its impact on the cell response needs to be clarified. For this test a special stand was designed.
![Page 15: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/15.jpg)
11.18.08 LCWS08 15
Response Versus MPPC Vertical Position for Concave Cell
Cell
MPPC
Positive value
![Page 16: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/16.jpg)
11.18.08 LCWS08 16
25
30
35
40
45
50
-1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
MPPC Vert ical Posit ion, mm
Cur
rent
, nA
Output Current with a Collimated Sr90 on the Top of the Cell Center
Positive value
![Page 17: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/17.jpg)
11.18.08 LCWS08 17
Uniformity Measurement Schematic
MPPC
Sr90
Cell
![Page 18: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/18.jpg)
11.18.08 LCWS08 18
0
15
30
45
-25 -20 -15 -10 -5 0
Source Position, mm
Cur
rent
, nA
0
15
30
45
2 mm
0 mm
Cell Response to Sr90 at Two MPPC Vertical Positions
![Page 19: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/19.jpg)
11.18.08 LCWS08 19
Cross Talk Estimation
• Current response of two concave adjacent cells was measured at the same temperature and the same bias voltage for both MPPC. The MPPCs were install in the center of each concave cell.
• The step of scanning was from 0.2 to 3.2 mm. The scans were performed using precision table.
• The goal was to see a gap between the cell boundaries.
• Measurements with one MPPC help estimate amount of light coming from the neighboring cell through the adjacent boundaries.
![Page 20: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/20.jpg)
11.18.08 LCWS08 20
0.00E+00
1.00E-07
2.00E-07
3.00E-07
4.00E-07
5.00E-07
20 40 60 80 100 120
Source Position, mm
Cu
rre
nt
A+
B, n
ATwo Cells and Two MPPC
A B
![Page 21: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/21.jpg)
11.18.08 LCWS08 21
7.00E-08
1.20E-07
1.70E-07
2.20E-07
2.70E-07
3.20E-07
3.70E-07
65 66 67 68 69 70
Source Position, mm
Cu
rre
nt
A, n
A
Two Cells One Cell
Impact of Adjacent Cell
A
![Page 22: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/22.jpg)
11.18.08 LCWS08 22
Summary
• Concave cell results in uniform response.
• The depth of the sensor within the concavity affects the uniformity of response.
• The uniformity depends on the size of the MPPC.
• The concave scintillating cell indicates a low sensitivity to the diameter of dimple.
• Cross talk between the adjacent concave cells is tolerable.
• An integrated circuit board/scintillator/sensor has been built and under test.
![Page 23: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/23.jpg)
11.18.08 LCWS08 23
Plans
• Test of single concave scintillating cell produced with injection mold process.
• Test of single concave scintillating cells with a thickness less than 5 mm and an area less and more than 9 cm².
• Test of single concave scintillating cell with less diameter of dimple.
![Page 24: Studies of Small Scintillating Cells with Modified Geometries](https://reader035.vdocuments.net/reader035/viewer/2022062217/568149bb550346895db6f0c6/html5/thumbnails/24.jpg)
11.18.08 LCWS08 24
Acknowledgment
• The authors would like to thank NIU Sr. Lab. Mechanic Phillip Stone for preparing the scintillating cells; Marcellinus Demarteau for support.