Techbox Systems Tel: +44 (0) 1339 755856 Aspen Lodge Email: [email protected] Braemar Road Web: www.techboxsystems.com Ballater AB35 5RQ UK Dynamic Scale Loop The dynamic scale loop is a tube blocking system for the evaluation of scale inhibitors. The basic principal involves mixing two incompatible brines in a narrow bore test coil. Scale begins to form in the coil creating a differential pressure. The time taken for this formation gives an indication of inhibitor performance. Techbox Systems manufacture units to suit all temperature and pressure requirements - from low pressure and temperature to pressures in excess of 6000psi and temperatures over 250°C. Our standard test rigs are the L1 low pressure and H200 high pressure units. We specialise in designing and manufacturing bespoke equipment to meet customer’s specific requirements. Schematic The H200 unit operates at up to 200 Bar and temperatures up to 200°C. The unit uses two preheat coils in addition to the test coil housed within a digitally controlled oven. Techbox H200 DSL

Upload: wjawich

Post on 03-Feb-2016

227 views

Category:

Documents


0 download

DESCRIPTION

Technical Description of the properties of TechBox Dynamic Scale Loop

TRANSCRIPT

Page 1: Techbox Systems DSL

Techbox Systems Tel: +44 (0) 1339 755856 Aspen Lodge Email: [email protected] Braemar Road Web: www.techboxsystems.com Ballater AB35 5RQ UK

Dynamic Scale Loop

The dynamic scale loop is a tube blocking system for the evaluation of scale inhibitors. The basic principal involves mixing two incompatible brines in a narrow bore test coil. Scale begins to form in the coil creating a differential pressure. The time taken for this formation gives an indication of inhibitor performance.

Techbox Systems manufacture units to suit all temperature and pressure requirements - from low pressure and temperature to pressures in excess of 6000psi and temperatures over 250°C.

Our standard test rigs are the L1 low pressure and H200 high pressure units.

We specialise in designing and manufacturing bespoke equipment to meet customer’s specific requirements.

Schematic

The H200 unit operates at up to 200 Bar and temperatures up to 200°C. The unit uses two preheat coils in addition to the test coil housed within a digitally controlled oven.

Techbox H200 DSL

Page 2: Techbox Systems DSL

Techbox Systems Tel: +44 (0) 1339 755856 Aspen Lodge Email: [email protected] Braemar Road Web: www.techboxsystems.com Ballater AB35 5RQ UK

Techbox H200 DSL Software

The control software enables the user to view all data in an easy to understand manner. The graphs display historical pressure and temperature information, whereas the schematic shows all current data in addition to the pumps status.

Model L1 H200

Max Pressure 1 Bar (14.5psig) 200 Bar (2900psig)

Pre-heating Unit S N

Oven O S

Cooling Bath O O

Max Temperature 90°C 200°C

Automatic Testing S S

High Pressure N O

Pump Type Peristaltic HPLC

Exchangeable Test Coil S S

High Accuracy Differential Transducer S S

Piston Flushing N S

Pressures Displayed & Logged S S

Temperature Displayed & Logged S S

Safety Pressure Relief N S

System Pressure Control S S

S-Standard O-Option N-Not Available

Page 3: Techbox Systems DSL

Techbox Systems Tel: +44 (0) 1339 755856 Aspen Lodge Email: [email protected] Braemar Road Web: www.techboxsystems.com Ballater AB35 5RQ UK

Typical result graph

The above graph show typical results. The untreated brine forms scale and starts to block the test coil after about 12 minutes (Blank). The test is then repeated with scale inhibitor present, firstly at 3ppm, automatically reducing to 2ppm after 45 minutes and finally reducing to 1ppm after 90 minutes. At about 110 minutes the test coil starts to block again indicating insufficient inhibitor. At this point the unit automatically cleans itself.

1

2

3

0.00 20.00 40.00 60.00 80.00 100.00

Dif

fere

nti

al P

ress

ure

(p

si)

Test time (Minutes)

Typical Scale Inhibition Perfomance Sulphate Brine at 90°C and 250psi

Untreated Brine

3ppm 1ppm

1psi differential pressure increase

2ppm

Scale Inhibitor

MIC of 2ppm

Page 4: Techbox Systems DSL

Techbox Systems Tel: +44 (0) 1339 755856 Aspen Lodge Email: [email protected] Braemar Road Web: www.techboxsystems.com Ballater AB35 5RQ UK

Specifications

Standard range of systems covering pressures 0 – 200 bar (Higher on request).

High accuracy differential pressure measurement.

Fluid temperature monitoring and recording.

Fully automated enabling a series of tests to be run with decreasing inhibitor concentrations until the minimum ppm is found.

Automated cleaning system.

Replaceable test coil.

Available as 110V or 240V.

Standard test temperatures up to 200°C. (Higher available)

Latest user friendly windows based Techbox software.

Overnight testing possible.

Small sample volumes.

Data saved in Excel compatible format.

Other options available on request including:-

Low temperature option

High pressure option

Manual version

Monel tubing available for heated areas.

Filter

Test Coil