1 Technical Brief on CD4 Technologies (May 2010) Contents 1.0 Aims and objectives 2.0 Target audience 3.0 Introduction 4.0 Background 4.1 Physiological variability in CD4 count 4.2 CD4 as a prognostic marker for HIV progression 4.3 WHO recommendations for the use of CD4 monitoring in the management of adults and adolescents living with HIV 4.3.1 The role of CD4 testing in deciding when to start ART 4.3.2 The role of CD4 testing in deciding when to switch to second line ART 5.0 Description of available technologies for CD4 enumeration 6.0 Considerations when choosing CD4 technology 6.1 Operating characteristics 6.2 Performance characteristics 6.2.1 Accuracy 6.2.1.1 Measures of test accuracy 6.2.2 Precision 7.0 Systematic review of performance of CD4 technologies 7.1 Methods 7.1.1 Inclusion criteria 7.1.2 Data extraction 7.1.3 Assessment of study quality 7.2 Results 7.2.1 Accuracy 7.2.1.1 Bias/mean difference 7.2.1.2 Misclassification probabilities 7.2.2 Precision 7.2.2.1 Withinlaboratory precision 7.2.2.2 Betweenlaboratory precision 7.2.3 Conclusions of systematic review 8.0 Importance of EQA programmes and QC 9.0 Suggested study design for future evaluations of CD4 technology 10.0 What is on the horizon: new CD4 technologies 11.0 References Appendix A: Search strategy and search terms Appendix B: Algorithm for study inclusion

Upload: dangminh

Post on 20-Apr-2018

215 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  1

Technical Brief on CD4 Technologies (May 2010) 

     Contents  1.0 Aims and objectives 2.0 Target audience 3.0 Introduction 4.0 Background 

4.1 Physiological variability in CD4 count 4.2 CD4 as a prognostic marker for HIV progression 4.3 WHO recommendations for the use of CD4 monitoring in the management of adults and        adolescents living with HIV        4.3.1 The role of CD4 testing in deciding when to start ART       4.3.2 The role of CD4 testing in deciding when to switch to second line ART 

5.0 Description of available technologies for CD4 enumeration 6.0 Considerations when choosing CD4 technology 

6.1 Operating characteristics 6.2 Performance characteristics       6.2.1 Accuracy                6.2.1.1 Measures of test accuracy       6.2.2 Precision 

7.0 Systematic review of performance of CD4 technologies 7.1 Methods        7.1.1 Inclusion criteria        7.1.2 Data extraction        7.1.3 Assessment of study quality 7.2 Results        7.2.1 Accuracy                  7.2.1.1 Bias/mean difference                  7.2.1.2 Misclassification probabilities       7.2.2 Precision                 7.2.2.1 Within‐laboratory precision                 7.2.2.2 Between‐laboratory precision       7.2.3 Conclusions of systematic review 

8.0 Importance of EQA programmes and QC 9.0 Suggested study design for future evaluations of CD4 technology 10.0 What is on the horizon: new CD4 technologies 11.0 References 

Appendix A: Search strategy and search terms Appendix B: Algorithm for study inclusion 

Page 2: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  2

Glossary 

Accuracy 

A general  term used  to describe how  similar  the  results obtained  for  individual  specimens with a new CD4 assay are to the results obtained with a reference CD4 assay. Measures of accuracy include bias and misclassification probabilities. 

Bias 

A measure of how much  the CD4  count  that  is obtained using  a new assay differs  from  the CD4 count  that  is  obtained  using  a  reference  assay when  the  same  specimen  is  analysed  using  both assays. (bias need not be constant over the clinically  important range of CD4 counts, and need not be the same for different laboratories using the same reference and new technologies.) 

Coefficient of variation 

The coefficient of variation (CV) is usually defined as the ratio of the standard deviation to the mean, which is usually estimated as the standard deviation divided by the average. This is often multiplied by 100 and expressed as a percentage (CV%). 

Mean 

The average value of a random variable, X. The most common estimate of the mean is the average of all the observations. 

Median 

The value m (of a continuous random variable, X) such that the probability that X  is  less than m  is one half. 

Misclassification probabilities 

The upward and/or downward misclassification probability can be calculated for a particular cut‐off, C. Among specimens with a CD4 count (obtained using the reference technology) which is less than C, the upward misclassification probability is defined as the percentage of specimens for which the CD4 count  (obtained using  the new  technology)  is greater  than or equal  to C  (this  is analogous  to definition of 1 minus  the  sensitivity  in a diagnostic  test with only  two values:  the patient has  the disease and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater than or equal to C, the downward misclassification probability  is defined as the percentage of specimens for which the CD4 count (obtained using the new  technology)  is  less  than  C  (this  is  analogous  to  definition  of  1  minus  the  specificity  in  a diagnostic test with only two values: the patient has the disease and the patient does not have the disease). 

Precision 

Precision  (also  known  as  reproducibility)  is  a  general  term  for how  close  the  results on  replicate specimens are (using a single assay technology). Several sorts of precision can be relevant, including within‐technician,  between‐technician,  within‐laboratory  and  between‐laboratory  precision. Measures of precision include the coefficient of variation (CV%). 

Page 3: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  3

1.0 Aims and objectives  

To provide updated technical guidance to optimize the use of CD4 technologies for HIV care.   

To provide an evaluation of the performance characteristics (accuracy and reproducibility of results) of availableCD4 technologies through a systematic review of published literature.  

To review the operational characteristics of currently available CD4 technologies, and outline new technologies under development. 

To  provide  guidelines  for  conducting  performance  evaluations  of  new  CD4  technology,  using clinically relevant measures of test performance. 

2.0 Target audience 

Ministries of health; national HIV/AIDS control programmes; laboratory managers and experts; manufacturers of CD4 technologies; procurement agencies 

3.0 Introduction  HIV  has  particular  tropism  for  CD4  cells  and  destruction  of  CD4+  cells  results  in  many  of  the deleterious effects of HIV.   Measurement of CD4+ T  lymphocytes  in peripheral blood  is therefore a critical  laboratory parameter  for  the evaluation and monitoring of patients with HIV.    It  is used  to determine  the  immunological  stage  of  HIV  infection,  for  identifying when  to  start  antiretroviral therapy (ART), to identify patients likely to benefit from co‐trimoxazole prophylaxis and to recognise those most at risk of developing immune reconstitution syndrome.  In addition, CD4 may be used in evaluating the response to treatment and for recognizing treatment failure.   

WHO  therefore  encourages  national  programmes  to  increase  access  to  CD4  measurement technologies to improve the quality of HIV prevention, care and treatment programmes. 

A number of  techniques  for determining CD4 exist, however  the  range of cost,  feasibility, ease of use, accuracy and reproducibility vary considerably.    In general the cost and complexity mean that access to CD4 and viral load measurement is still highly restricted in many low income countries. The infrastructure of clinical laboratories located in many developing countries is often inappropriate to support  routine determination of CD4 and viral  load measurements, and high costs of equipment, reagents and problems associated with machine maintenance further restrict access. 

It  is  unclear  what  proportion  of  adults  and  children  living  with  HIV  have  access  to  laboratory monitoring  tests  such  as  CD4  and  viral  load  given  the  cost,  complexity  and  lack  of  capacity  to perform monitoring  tests  in many  resource  limited settings. However significant  investments have been made in developing  laboratory capacity and countries are increasingly gaining access to these technologies. 

  

 

Page 4: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  4

4.0 Background  

CD4+ T‐lymphocytes, also known as the helper T‐cells, are the coordinators of the immune response which protects  the body against microbial disease and  some  forms of cancer.   The destruction of CD4+ T‐lymphocytes by HIV is the main cause of the progressive weakening of the immune system in HIV infection, and leads ultimately to the acquired immune deficiency syndrome, AIDS.  4.1 Physiological variability in CD4 count 

Diurnal  variation  in  CD4  has  been  observed  in  HIV‐negative  and HIV‐positive  subjects, with  CD4 counts  typically  lower  in  the morning  than  the evening51,52.   Week‐to‐week and seasonal variation may occur, and changes  in CD4 count may also be seen  related  to exercise, smoking,  intercurrent infection, and  some medications other  than ART  (eg  corticosteroids).   Changes  in  total white  cell count and lymphocyte count may contribute more to the fluctuations in absolute CD4 count seen in response  to  these  biological  factors  than  does  the  CD4%51,54.    These  physiological  intrasubject variations  in CD4 count may account  for as much as,  if not more than,  technical variability of CD4 measurement53.    

4.2 CD4 as a prognostic marker for HIV progression 

CD4  count  is  a  strong  predictor  of  HIV  progression  to  AIDS  and  death,  and  is  by  consensus considered  the  best  laboratory  marker  for  deciding  when  to  initiate  antiretroviral  treatment (ART)54,55.   There  is much heterogeneity  in  rate of CD4 decline, with  a minimum period between seroconversion  and AIDS  of  about  2  years,  and  a median  of  11  years54.    In  addition,  the  rate  of decline in CD4 count for an individual patient may be highly variable over time56, making it difficult to predict  at higher CD4  counts  the  likely  time  course of an  individual’s progression  to  advanced disease.   

The use of clinical staging alone  to determine ART  initiation  is  limited by  the uncertain prognostic value of  asymptomatic or mild disease.    In  a Malawian  setting,  56% of  1663  patients with WHO clinical stage 1 or 2 disease had CD4 counts of <350 cells/µl, and 36% had CD4 counts <250 cells/µl57.  The authors also reported high rates of attrition from healthcare among those patients not started on  ART;  it  may  therefore  be  supposed  that,  were  CD4  testing  not  performed,  many  patients unrecognised as eligible for treatment may progress to advanced disease or die before re‐presenting to healthcare. 

Regular monitoring of CD4 counts, in conjunction with clinical criteria, is therefore recommended for monitoring HIV progression, with treatment initiation indicated when CD4 drops to below a defined threshold.        

4.3 WHO recommendations for the use of CD4 monitoring in the management of adults and adolescents living with HIV58  

4.3.1 The role of CD4 testing in deciding when to start ART 

In November 2009,  the WHO  issued a  rapid advice update on  their Antiretroviral  therapy  for HIV infection  in adults and adolescents guideline58.   A clear stimulus  to  this update was  the  increasing evidence  of  improved  survival  and  reduced  HIV‐related  illnesses  with  earlier  initiation  of antiretroviral therapy (ART).  A key recommendation of the updated guideline is therefore that ART 

Page 5: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  5

be  started  in  all  patients  with  HIV  who  have  a  CD4  count  ≤350  cells/µl,  irrespective  of  clinical symptoms.    As  a  result  of  the  uncertain  prognostic  value  of WHO  clinical  stages  1  and  2,  the guidelines recommend that all HIV‐infected individuals with WHO clinical stage 1 or 2 disease should have access to CD4 testing to decide if treatment should be initiated. 

For all patients with WHO clinical stage 3 or 4, ART initiation is recommended irrespective of CD4 count. 

WHO recommendations for when to start ART Start antiretroviral treatment in all patients with HIV who have CD4 count <350 cells/mm3 irrespective of clinical symptoms. (Strong recommendation, moderate quality of evidence) 

CD4 testing is required to identify if patients with HIV and WHO clinical stage 1 or 2 disease need to start antiretroviral treatment. (Strong recommendation, low quality of evidence) 

Start antiretroviral treatment in all patients with HIV and WHO clinical stage 3 or 4 irrespective of CD4 count. (Strong recommendation, low quality of evidence) 

  The criteria for initiating ART for pregnant women are the same as for non‐pregnant adults, with an antenatal CD4 count considered critically important to determine eligibility to ART for the health of the  woman.    Shorter  term  ART  prophylaxis  for  prevention  of  mother  to  child  transmission  is indicated when the mother does not meet the criteria for treatment for her own health.   

4.3.2 The role of CD4 testing in deciding when to switch to second line ART 

Where available,  it  is  recommended  that viral  load be used  for confirmation of  treatment  failure.  However when  viral  load  is  not  available,  the  use  of  immunological  criteria  is  recommended  to confirm clinical failure, having been shown in 2 randomised controlled trials to be superior to clinical monitoring alone59,60.   

Page 6: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  6

5.0 Description of available technologies for CD4 enumeration 

 The gold standard technology for CD4+ T‐cell counting is flow cytometry, a technology that allows a single cell to be measured for a variety of characteristics. Flow cytometers are used in cell counting or cell sorting applications.44  For CD4+ T cell enumeration, cells are stained (fluorescence tagged cell surface  markers)  for  CD4+  T‐cell  counting.  The  standard  methodology  for  determining  the percentage of CD4+ cells is by staining for CD45/CD3/CD4.  By first selecting the lymphocytes (CD45), the CD3/CD4 cytogram can be used to enumerate CD4+ cells. More extensive or simplified staining panels are commercially available.   

While  flow cytometry generally gives the percentage of CD4+ or CD8+ cells, some  flow cytometers precisely record the volume of sample analysed or deliver a fixed volume of sample. This method is called volumetric approach and can directly convert CD4%  to absolute CD4 counts.  In  instruments without this facility, a fixed volume of sample  is spiked with a known number of fluorescent beads (beads‐based  systems).  Since  the  fluorescent beads  can easily be differentiated  from  cells by  the flow cytometer, the counted beads of the sample analysed permits volume determination and hence the concentration of cells. The volumetric approach or systems using  fluorescent beads are called single platform  approaches  and  are  commercially  available  as  TruCount  (Becton Dickinson, USA), FlowCount beads (Beckton Coulter) and Perfect Count (Cytognos) that can be used on practically any flow  cytometer. Besides using  the  traditional  flow  cytometers  (open platforms), dedicated  (often called  closed)  platforms  are  developed  for  CD4+  T‐cell  counts.  The  three  commercially  available dedicated  single  platforms  are:  FACScount  (Becton  Dickinson),  Guava  easy  CD4  (Guava techonologies) and Partec Cyflow Counter (Partec). The dedicated platforms  in comparison to flow cytometers  have  decreased  applicability,  but  allow  CD4+  T‐cell  counting  with  reduced  technical complexity. 

Another  approach used  to  calculate  absolute CD4+  counts  is  to use  a haematology  analyser. The haematology analyser measures the total white cell count. This dual platform approach  introduces variability  in  absolute  CD4+  counts  due  to  combining  results  of  two  platforms  into  a  single calculation. Recent developments in staining panels and data analysis (Panleucogating) offer better results over traditional dual platform methodology, increasing the performance of the dual platform approach up to the single platform level45.  

Besides the  traditional  flow cytometry methodology, manual methods exist. Two manual methods use microscopy  (fluorescence  or  light)  and  haemocytometers.  By  using  beads  (Dynal  T4  Quant, Dynabeads) or spheres (Coulter Manual CD4 count kit, Cytospheres) that become distinctly visible by microscopy when in contact with CD4+ T‐cells, it is possible to count cells.  

Page 7: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  7

6.0 Considerations when choosing CD4 technology 

In deciding upon a technology, both the operating characteristics and the performance characteristics will require careful consideration.   

6.1 Operating characteristics 

Operating characteristics include: 

• Environmental requirements of the technology: instrumentation, power supply, refrigeration of reagents, robustness to heat and humidity 

• Parameters measured – absolute CD4, CD4% • Ease of use, training required, degree of automation • Cost • Throughput • Compatibility with external quality assessment (EQA) programmes and availability of quality 

control (QC) reagents • Compatibility with stabilised specimens 

 The operating characteristics of the currently available CD4 technologies are presented in Tables 1‐4.  Desirable operating characteristics will differ according to the level of the healthcare system, and are likely  to be very different at primary  level health posts  than at national  reference  laboratories.   A tiered approach reflecting the capabilities of, and demands on, each level of the health care system could assist  in  the allocation of  limited  resources.   The desirable operating characteristics  for CD4 testing are described for each level of the health care system in Table 5.    Other  considerations  in  the  context  of  rapid  expansion  of  testing  capacity  include  the  existing instrument base  and  laboratory  infrastructure,  the  expected  changes  in  testing demand over  the subsequent 2‐3  years,  the  availability of human  resources,  and  the need  for CD4%  for paediatric testing50.  Compatibility with EQA programmes and availability of QC reagents are of prime importance and are discussed further in section 8.0.       

 

Page 8: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  8

6.2 Performance characteristics  

In addition  to  considering operating  characteristics when  choosing a  technology,  it  is essential  to consider  performance  characteristics.    Performance  in  this  context  refers  to  the  capability  of  a technology  to give accurate and  reproducible  results, and  therefore  to guide patient management appropriately. 

In evaluating the performance of a test, both accuracy and precision must be considered.   

6.2.1 Accuracy 

Accuracy is a measure of the degree of closeness of the reported value to the true value.  Accuracy of  a  new  technology  is  evaluated  by  comparing  results  generated  by  the  new  test  with  those obtained for the same samples using a reference technology.   

6.2.1.1 Measures of test accuracy  

Correlation and linear regression analyses are frequently used to report accuracy, but this approach is inappropriate in the evaluation of CD4 technology.  In this context, correlation coefficients do not indicate level of agreement, and may be misleading45,46.  Unacceptably high levels of misclassification (a  clinically  relevant measure  of  accuracy) may  be  seen  despite  excellent  correlation.    For  these reasons, assessing correlation alone is insufficient. 

Other measures of accuracy include bias and misclassification. 

Bias or mean difference reflects the average difference between the results of the technology under evaluation and an accepted reference standard. Bias may be reported using Bland‐Altman analysis46, in which the difference between two methods of measurement is plotted against their mean.  If the differences do not vary systematically over the range of CD4 counts studied, then the overall mean difference can be calculated and indicates bias.  If the differences are normally distributed, then 95% of the differences will lie within 2 standard deviations (referred to as ‘limits of agreement’).  If these limits  are  clinically  acceptable,  agreement  between  the  two methods  is  adequate,  and  the  new technology may be used  in place of  the  reference  standard.   However,  in CD4 measurement,  the magnitude and variability of the differences between two methods are nearly always larger at higher counts, making this approach unreliable.  Even within the clinically important range of CD4, bias may not be constant.   

Misclassification probabilities describe the likelihood that a test will incorrectly categorise a result as higher or  lower  than a given cut‐off value.   They have  the advantage of being a clinically  relevant measure of accuracy, as they can be used to assess the  likelihood that a patient will be  incorrectly classified above or below a defined CD4 threshold used in clinical decision making.   

Two types of misclassification can be defined — upward misclassification probability and downward misclassification  probability.    Among  specimens with  a  CD4  count  (obtained  using  the  reference technology) which  is  less  than  a  defined  cut‐off  of  value,  upward misclassification  probability  is defined  as  the  percentage  of  specimens  for  which  the  CD4  count  (obtained  using  the  new technology)  is greater than or equal to that cut‐off.   Among specimens with a CD4 count (obtained using  the  reference  technology)  which  is  greater  than  or  equal  to  the  cut‐off,  the  downward misclassification  probability  is  defined  as  the  percentage  of  specimens  for which  the  CD4  count (obtained using the new technology) is less than the cut‐off45. 

Page 9: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  9

   

Misclassification probabilities  are  therefore  analogous  to 1 minus  the  sensitivity  and 1 minus  the specificity of a test around a determined value.   

Upward misclassification around a treatment threshold may be most clinically important (leading to delay of start of ART or prophylactic treatment in some patients).  Downward misclassification may result in treating large numbers of patients who have CD4 counts above the guideline threshold.  

A  disadvantage  of misclassification  probabilities  as  defined  here  is  that  they  do  not  differentiate between  low‐magnitude  inaccuracy, eg a count of 349 cells/µl being misclassified as >350 cells/µl, which may be reasonably expected from any test, and high‐magnitude inaccuracy.  Misclassification probabilities are therefore heavily influenced by the study population.  A study which includes many CD4 values falling close to the defined cut‐off (as measured by the reference test) will show higher rates of misclassification by the new test than a study where the majority of values lie away from the threshold.    For  this  reason,  reporting  the  range  and  distribution  of  CD4  values  of  the  study population is of great importance. 

A more  important measure might be the probability that a patient with a CD4 test well below the treatment threshold might be incorrectly classified as above that threshold, e.g. the probability that a  patient  with  a  ‘true’  CD4  value  of  <250  cells/µl  might  be  misclassified  as  >350  cells/µl.  Unfortunately these data are currently rarely available in the published literature. 

6.2.2 Precision 

Precision measurements answer the question  ‘how close are the results of replicate testing using a single  technology?’    Precision  is  particularly  important  when  following  a  patient's  serial measurements using the same technology.  Precision can be measured within‐run, between‐run, or between‐laboratory. 

Within‐run variability  is calculated  from  testing multiple  replicates of  the  same  specimen within a single machine run. Between‐run variability is calculated from testing multiple replicates of the same specimen in different runs or at different times. The two measures together describe the amount of variation  that  can  be  expected  among  results  using  the  same  specimen.  Precision  is  also  called reproducibility  and  can  be  thought  of  as  the  degree  to  which  repeated  measurements  under unchanged conditions show the same results. Precision can also be calculated between laboratories to determine how much  variation  is  introduced  into  the  results by having  the  test performed by different technicians in different settings (between‐operator or between‐laboratory variability).  

Data  on  precision  are  often  reported  as  the  coefficient  of  variation  (CV) which  is  a measure  of dispersion  similar  to  standard deviation but  is  calculated with  consideration  for  the data and  can therefore be compared across studies.  A lower value of CV indicates less variation and greater assay reproducibility.   %CV may be badly underestimated if based on too few replicates; a minimum of 8 replicates should be used45.     

Manufacturers  publish  information  on  their  technology’s  accuracy  and  precision;  however, independent, peer‐reviewed evaluations are a more reliable source of  information.   Evaluations of technology  should  include  a  comparison  of  the  test under  evaluation  to  a  reference  standard  of acceptable standing  in the scientific community.   Before deciding to  implement a technology, sites 

Page 10: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  10

need  to  be  satisfied  that  independent  peer‐reviewed  evaluation  data  have  shown  acceptable performance. 

Page 11: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  11

7.0 Systematic review of performance of CD4 technologies 

      We  performed  a  systematic  review  of  studies  evaluating  the  performance  of  CD4  enumeration technologies.    A  search  of  the  Cochrane  Library  and  the  Centre  for  Reviews  and  Dissemination databases  (including DARE, NHS  EDD  and HTA)  found  no  existing  reviews  addressing  the  review objective.    7.1 Methods  We followed standard guidance in performing the review47.  Details of search strategy, search terms and algorithm for study selection are included in appendices A and B.  7.1.1 Inclusion criteria  Inclusion criteria are shown below.   INCLUSION CRITERIA (PICOS) Population Include studies of performance of CD4 enumeration technologies in HIV infected and/or non‐infected adults, in any geographical location 

Interventions (index tests) Include any commercially available technology used for the enumeration of CD4 count, including (but not limited to) manual bead‐based technologies, flow cytometry systems requiring minimal technical input, and flow cytometry systems requiring significant operator input or flow cytometry skill Reference standard/comparators Include studies using any acceptable (currently or at the time of the study) flow cytometric method as the reference standard Outcomes Include studies evaluating accuracy (including e.g. bias and/or misclassification probabilities) and/or precision. Include studies calculating sensitivity, specificity, PPV and NPV for a defined CD4 cut‐off value Study design Include evaluation studies using an acceptable reference technology Other Include English language only, 1990‐present  

Currently, no gold standard technology or internationally recognised reference preparation exists for CD4, and a wide range of flow cytometric methodologies have been used as comparators.   For the purposes of  this  review, we  included  studies using as  reference  technologies any  flow  cytometric methodology considered by the review panel to be acceptable methodology currently or at the time of the study.  

    

Page 12: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  12

    7.1.2 Data extraction  The  following  details  were  extracted  from  studies  accepted  for  inclusion:  location  of  study, laboratory  or  health  care  level,  degree  of  manufacturer  involvement  in  the  study,  index  test, reference  test, population  (HIV positive or HIV positive and negative).   The  following measures of accuracy were extracted:  bias (mean or median difference), including limits of agreement if Bland‐Altman  analyses  were  performed,  and  maximum  observed  difference  when  available; misclassification  probabilities  around  a  defined  CD4  cut‐off  (when  sensitivity  or  specificity  values were given, misclassification probabilities were calculated).   If data were available for subgroups of CD4 values, these were extracted, otherwise where available the range of CD4 counts studied was recorded.    The  following measures  of  precision were  extracted where  available: %coefficient  of variation  for  index and reference tests, no. of replicates per donor, no. of donors, CD4 count(s) of replicate  sample(s).   Where HIV  positive  data  could  be  extracted,  this  alone  has  been  included.  Where this was not possible, combined HIV positive and negative data are included. 

Correlation or linear regression data are inappropriate for evaluation of CD4 test accuracy and were not extracted.  For this reasons, some studies which met the inclusion criteria had no data extracted.   

7.1.3 Assessment of study quality 

Study quality was assessed based on criteria outlined by the STARD (Standards for the Reporting of Diagnostic accuracy studies) initiative48.  Particular focus was given to the following criteria: 

‐Did  the  study  report  the  training  received  by  the  technicians  prior  to performing  the  index  and reference tests? 

‐Did the study report time between blood draw and testing? 

‐Did the study use correlation analysis alone to report accuracy? 

‐Did the study report the distribution/range of CD4 of the sampled population? 

‐Were technicians performing index/reference tests blinded to the results of the other method? 

7.2 Results 

43  studies  were  accepted  for  inclusion  in  the  review1‐43.    At  least  one  published  performance evaluation  study  was  found  for  each  of  the  following  technologies:  Coulter  cytospheres,  Dynal dynabeads, Guava Easy CD4, Partec CyFlow, BD  FACSCount,  singleplatform  flow  cytometry  (SPFC) using TruCount beads (Becton Dickinson), SPFC using Flow‐Count fluorospheres (Beckman Coulter), SPFC  using  Perfect  Count  beads  (Cytognos),  SPFC  using  flow  rate  calibration,  dual  platform panleucogating (PLG), single platform PLG using Beckman Coulter FlowCARE, Sysmex pocH‐100i with Dynal  dynabeads,  i+MED  CD4  Select.    No  published,  peer‐reviewed  performance  evaluations  of PointCareNOW or Guava Auto CD4 were found by our literature search.  The studies took place in a wide  range  of  resource‐rich  and  resource‐poor  countries;  the majority  of  studies  took  place  in research, reference or teaching hospital laboratories. 

Page 13: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  13

7.2.1 Accuracy 

7.2.1.1 Bias/mean difference 

74%  of  studies  reported  bias, mean  difference  or median  difference.    58%  of  studies  performed Bland‐Altman analyses with limits of agreement.  37% of studies reported bias for subgroups of the CD4 count range, e.g. for those samples <500, <350 or <200. 

7.2.1.2 Misclassification probabilities 

30% of studies reported misclassification probabilities or provided data from which misclassification probabilities  could  be  calculated.    Of  these,  the  majority  reported  misclassification  or sensitivity/specificity around a CD4 cut‐off of 200 cells/µl, with only 3 studies providing data around a  cut‐off  of  350  cells/µl.   Misclassification  probability  data  using  a  cut  off  of  350  cells/µl were available  for  the  following  technologies: Coulter  cytospheres, Dynal dynabeads, Guava  Easy  CD4, Partec CyFlow.   

Misclassification data are summarised in figures 1 and 2 below. 

Both misclassification up and down  (particularly down) are  likely  to be underestimates as none of the  studies  are  restricted  to  the most  clinically  relevant  range  and many  included  HIV  negative patients with high CD4 counts.   Lutwana et al5 conducted a  large study of manual  technologies  in Uganda;  they  recruited  only HIV  positive  patients,  the majority  of whom  had  counts within  the clinically  important  range  (range  in  study:  0‐900).    Of  samples  with  ‘true’  counts  <350  cells/µl, Coulter cytospheres misclassified 1% as >350 cells/µl.   However,   of  those with  ‘true’ counts >350 cells/µl, 68% were misclassified as <350 cells/µl    (indicating that 68% of patients not qualifying  for treatment using  flow cytometry would have done so  if cytospheres were used).   Dynal dynabeads had  upward  and  downward misclassification  probabilities  of  6%  and  30%  respectively.    Upward misclassification  by  either  manual  technology  is  however  likely  to  be  an  underestimate  as  the majority of counts were very  low  (less  than 25% of samples had counts >200 cells/µl);  if  the tests were to be used in a population with counts closer to the treatment threshold (as would be likely if used primarily in asymptomatic patients), upwards misclassification would be expected to be higher.   

Karcher  et  al13  compared  Coulter  cytospheres  and  the  Partec  CyFlow  Counter  (both  under  field conditions),  with  dual‐platform  flow  cytometry.    Of  samples  with  counts  <350  cells/µl  by  flow cytometry, 29% were misclassified as having >350  cells/µl by  the CyFlow Counter  (indicating  that 29% of patients eligible  for treatment on the basis of a  flow cytometry CD4 count may have been denied  treatment  if CyFlow‐determined CD4 counts were used as  the only criteria  for  treatment); downward misclassification probability was 7%.   Coulter cytospheres  in this study had upward and downward misclassification probabilities of 16% and 20% respectively.     

Thakar et al9 compared Guava Easy CD4 with flow cytometry, and found 0% upward misclassification and 18% downward misclassification using a  threshold of 350 cells/µl.   The sample size was small however;  calculating  95%  confidence  intervals  gives  confidence  limits  of  0‐6%  for  upward misclassification and 6‐38% for downward misclassification. 

Importantly,  we  cannot  tell  from  these  misclassification  probabilities  the  magnitude  of  the misclassification:  are the samples that are being misclassified restricted to those whose ‘true’ values lie close to the threshold (e.g. within 10 CD4 cells/µl of the treatment cut‐off), or are samples with 

Page 14: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  14

‘true’ values substantially greater or less that the treatment cut off also being misclassified?  A visual impression of this can sometimes be gleaned from observing data plots, but a more useful measure would involve calculating the probability of these high‐magnitude misclassifications.       

Page 15: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  15

 

Figure 1a.  Misclassification up, using a CD4 cut‐off of 200 cells/ul Each bar represents results of one study.  Height of bar represents % of samples misclassified as >200 cells/ul by the test under evaluation (i.e. for these samples, the index test gave a result >200, but the reference test gave a result of <200)  

 

Figure 1b. Misclassification down, using a CD4 cut‐off of 200 cells/ul Each bar represents results of one study.  Height of bar represents % of samples misclassified as <200 cells/ul by the test under evaluation (i.e. the index test gave a result <200, but the reference test gave a result of >200) 

Page 16: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  16

 

Figure 2a. Misclassification up, using a CD4 cut‐off of 350 cells/ul Each bar represents results of one study.  Height of bar represents % of samples misclassified as >350 cells/ul by the test under evaluation (i.e. the index test gave a result >350, but the reference test gave a result of <350)   

 

Figure 2b. Misclassification down, using a CD4 cut‐off of 350 cells/ul Each bar represents results of one study.  Height of bar represents % of samples misclassified as <350 cells/ul by the test under evaluation (i.e. the index test gave a result <350, but the reference test gave a result of >350) 

Page 17: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  17

  

7.2.2 Precision 

7.2.2.1 Within‐laboratory precision 

51% of studies reported within‐laboratory precision data.  These are presented in figure 3a below: 

                       Figure 3a.  Within‐laboratory variability.  %CV= coefficient of variation, NC=not clear, CC=Coulter                             cytospheres, DD=Dynal dynabeads 

Gernow et al4 studied  the  reproducibility of Coulter cytospheres and  found poor precision, with a coefficient of variation of 58%.   The study by Landay et al3 however found precision  levels more  in keeping with the other technologies.   The reason for this disparity of results  is not clear.   The CD4 counts of the 19 samples used for replicate analysis by Gernow et al were not stated, however they included 12 HIV negative samples which might be assumed to have high counts.  Given that several papers found cytospheres to have poorer performance at higher counts, this may be in keeping with poor reproducibility  in these replicates.   However, Landay et al found better precision on a sample with a CD4 count of 1200  (%CV 3.5%) than on a sample with a CD4 count of 200  (%CV 10.8%).    It should  be  noted  that  the  manual  methods,  although  employing  simple  technology,  are  labour intensive and require significant user skill.    Inadequate training,  lack of supervision or user  fatigue may lead to poor performance of these techniques. 

 

  0

10 

20 

30 

40 

50 

60 

3  4  14 15 16 16 16 9 17 17 18 18 19 19 20 20 20 11 11 17 17 21 21 22 22 22 22 23 23 24 25 26 26 26 27 28 28 28 28 28 28 29 30 30 30 30 12 12 12 12 12 

%CV 

3

14

16

18

22

26

28

29

12

  

Study   

Index  test   

No. of replicates   

  

  C C ‐ ‐‐ ‐ > DD‐‐ ‐ ‐>   G uava Easy CD4 ‐ ‐‐ > Partec Cy Flow ‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐>FACSCount‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐> TruCount‐‐‐‐‐>FlowCount‐‐‐‐‐‐‐‐‐‐>PLG‐‐‐‐‐‐‐‐‐‐‐‐‐ > Sys mex‐ ‐‐‐‐‐ ‐‐‐‐‐ >   

Coulter cytospheres (CC) 

Dynal dynabeads (DD) 

Guava Easy CD4 

Partec CyFlow 

BD FACSCount 

BD TruCount  

BC FlowCount 

DP panleucogating 

Sysmex POCH‐100i 

  NC NC   2  10   2       4   2     2     3 ‐ 5  3  10 10 10 15 10 10 10 >10 >10 3‐53 10 10   4   4   4   4  NC  NC   5   4    8   8   8    4   8  NC NC   8 NC NC   8   5   5   5   5    10 10 10 10 10        

%CV

 

Page 18: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  18

 

 

7.2.2.2 Between‐laboratory precision 

Six studies reported between‐laboratory precision.  Two were studies evaluating BD FACSCount, two were  evaluating  panleucogating,  and  two  were  studies  evaluating  bead‐based  single  platform technology  (BD  TruCount  beads  and  BC  FlowCount  fluorospheres).    Four  of  the  six  studies  also reported between‐laboratory precision data for the reference test (usually a dual platform method). 

These data are presented in figure 3b: 

 

Figure 3b Between‐laboratory variability. %CV=  coefficient of  variation, NC=not  clear, DPFC=dual platform flow cytometry, DP PLG=dual platform panleucogating, BD=Becton Dickinson, BC=Beckman Coulter. Ref. tests=reference tests 

Single  platform  flow  cytometry  using  TruCount  beads  or  FlowCount  fluorospheres  showed significantly less inter‐laboratory variability than the dual platform comparators in these studies26, 28.  In addition, Denny et al29 demonstrated significantly  improved inter‐laboratory precision using dual platform  panleucogating  compared  with  predicate  technologies  (which  included  dual  or  single platform conventional flow cytometry).   7.2.3 Conclusions of systematic review  This  review  highlighted  the  difficulties  of  answering  clinically  relevant  questions  about  CD4  test performance from the published  literature.   The studies reviewed frequently concluded that a test had  acceptable  performance  based  only  on  correlation  analysis,  or  on  the  demonstration  of  low levels of estimated bias over a wide range of CD4 counts, neither of which are reliable measures of 

 23      23                26     26                28      28     28                 29     29                30     30 Study 

No. of replicates 

Method 

 NC     NC                5        5                 5        5        5                   5       5                   4        4 

%CV

 

Index testsReference tests 

Page 19: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  19

accuracy  in this context.   A minority reported clinically useful measures of accuracy, and few were carried out under field conditions.  The available data do however support the following statements:  

1. Between‐laboratory precision is likely to be superior with single‐platform technologies (using BD  TruCount  beads  or  BC  FlowCount  fluorospheres)  or  panleucogating  than  with conventional dual‐platform technologies 

2. Manual methods, particularly Coulter cytospheres, may be associated with higher  rates of misclassification and worse precision  than  flow‐based methods, and are difficult to quality control 

  It  is difficult  to draw conclusions about  the performance of Guava Easy CD4 or  the Partec CyFlow instruments due to a paucity of appropriate data and,  for the Partec  instruments, a wide  range of instruments,  reagents  and  parameters  used  in  different  studies  making  it  difficult  to  draw generalisable conclusions.  No  published  peer‐reviewed  literature were  found  relating  to  either Guava Auto  CD4/CD%  (now replacing Guava  Easy CD4), or  to  the PointCareNOW  instrument,  and only one, non‐independent study of the CD4 select assay was identified, which lacked precision or misclassification data.  Despite a lack of misclassification data on BD FACSCount, within‐laboratory and between‐laboratory precision were good and these findings are supported by EQA data.  It  can  be  seen  that  whatever  technology  is  chosen,  there  is  variability  associated  with  CD4 measurement.    Performance  characteristics  vary  between  technologies,  and  these  characteristics have  important  implications  both  for  individual  patient  management  and  for  HIV  treatment programmes.    It  is therefore essential to consider performance as well as operating characteristics when  choosing  technology.   However,  these  data  are  not  always  available  in  the  literature  and, currently,  evaluations  are  not  robust  or  comprehensive  enough  to  give  a  clear  idea  of  the comparative merit of different technologies.  What  constitutes  ‘acceptable’  performance  remains  undefined  and  may  vary  between  sites, depending on  local  factors  such as  the distribution of CD4  counts among  asymptomatic patients, how  often  patients  come  back  for  repeat  CD4  testing  and  the  implications  of  potential overtreatment (e.g. cost, promotion of drug resistance).  It is relatively straightforward for a site to decide which technology best  fits their needs based purely on operating characteristics;  it may be harder to decide what performance characteristics are acceptable, and harder still to obtain data on test performance to inform choice.   Given  the  potential  for  error,  participation  in  EQA  programmes  and  access  to  QC  reagents  are essential.    In  the  next  sections we will  describe  the  importance of  quality  control,  and  outline  a suggested study design for future evaluations of CD4 enumeration technology.    

Page 20: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  20

  8.0 Importance of EQA programmes and QC  ?DB to help write  9.0 Suggested study design for future evaluations of CD4 technology      10.0 What is on the horizon: new CD4 technologies  Recent  research  and  development  of  CD4  technologies  have  focused  on  creating  accurate  and reliable point‐of‐care (POC) assays.  A POC test for CD4 enumeration could greatly improve access to CD4 testing in low‐income countries with limited laboratory capacity. 

Inverness Medical has  a  POC  CD4  test,  the  Pima CD4, which  is  already  commercially  available  in select markets.  The  test  uses  a  disposable  cartridge  containing  sealed  reagents  and  a  portable analyzer to provide absolute T‐cell counts in whole blood within 20 minutes. The technology is based on  static  image  analysis  and  counting  principles  but  unlike  current  assays  employing  the  same technology, the Pima  is portable and does not require extensive  laboratory equipment or training. Furthermore,  the  assay  requires only 25 µl of  capillary blood  and has  all  test  reagents  contained within  the disposable cartridge. The  reagents are dried and  therefore do not  require  refrigeration and  no  manual  sample  processing  is  required  to  obtain  a  result (http://pimatest.com/en/home.html).  

Although it is commercially available, there have not been any independent evaluations of the Pima comparing the POC test to an acceptable gold standard assay.  

mBio, a division of  the Precision Photonics Corporation, has developed a  system based on planar waveguide technology that  is being applied to CD4 enumeration and the simultaneous diagnosis of HIV and common co‐infections. The system uses an  integrated  fluidic cartridge and a  fluorescence imaging  system  to provide automated  cell  counting. The  technology,  core multiplexed waveguide technology, has received FDA testing; however, there  is no  indication of the future availability of a POC CD4 test (http://www.mbiodx.com/ ). 

Daktari Diagnostics  is a third company  involved  in the manufacture of a POC CD4 test. The Daktari system will  consist of a hand‐held, battery powered  instrument and disposable plastic  cartridges. 

Each test consists of applying a drop of blood from a fingerstick to the cartridge (only 10 μL of blood is required), capping the cartridge, and inserting the cartridge into the instrument. An absolute CD4 cell  count  result  will  be  obtained  within  approximately  six  minutes.    The  system  runs  off  a rechargeable battery, and is designed to be usable by anyone, anywhere in the world.  The system is based on research  in microfluidics and electrical sensing and employs: 1) a process  for performing the assay on a drop of untreated whole blood, eliminating  the  time consuming and complex pre‐processing  steps  required  by  flow  cytometers;  and  2)  a  process  for  counting  cells  using  simple electrical impedance measurements, eliminating the use of costly optical methods. ( http://www.daktaridx.com/). 

Page 21: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  21

The CD4  Initiative was established  in 2005 by an award  from the Bill & Melinda Gates Foundation and is managed through Imperial College of Medicine in London, UK with the objective of developing a new, low‐cost, rapid POC test for measuring CD4+ T‐lymphocytes. The Initiative put forth a call for design proposals  and  current  awards  are held by both  Zyomyx  Inc  and  ImmunoSite  Technologies (http://www1.imperial.ac.uk/medicine/about/divisions/departmentofmedicine/infectiousdiseases/infectious_diseases/cd4_initiative/) .  

Zyomyx  is developing a CD4 gauge similar  to a thermometer  that will quantify CD4 T‐lymphocytes from  whole  blood  obtained  through  a  finger‐prick;  however,  the  prototype  is  still  under development.   

In addition to POC tests for CD4 enumeration, dried blood spots (DBS) have been  investigated as a possible alternative for CD4 testing in settings with limited access to laboratory services. Redd et al49 designed a technique for estimating mature T‐cell numbers by calculating the amount of rearranged T‐cell receptor β genes from dried blood spots of HIV‐infected individuals. This was done by genomic DNA  amplification  with  Real‐Time  PCR.  The  variability  in  the  results  and  the  failure  to  detect lymphocytes  that are not  fully or close‐to‐fully mature, demand more  research before  this can be presented  as  a  viable  alternative  to  current methods.  The  potential  of  combining  HIV  viral  load quantification  and  CD4  enumeration  on  one  single  DBS  could  be  an  easily  implementable  and affordable alternative to the current, expensive methodologies.  

        

Page 22: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  22

11.0 References  1. Balakrishnan, P. et al. An inexpensive, simple, and manual method of CD4 T‐cell quantitation in HIV‐infected individuals for use in developing countries. J. Acquir.Immune Defic. Syndr. 36, 1006–1010 (2004). 

2. Carella, A. V., Moss, M. W., Provost, V. & Quinn, T. C. A manual bead assay for the determination of absolute CD4+ and CD8+ lymphocyte counts in human immunodeficiency virus‐infected individuals. Clin. Diagn. Lab.Immunol. 2, 623–625 (1995). 

3. Landay, A. et al. A rapid manual method for CD4+ T‐cell quantitation for use in developing countries. AIDS 7, 1565–1568 (1993). 

4. Gernow, A., Lisse, I. M., Bottiger, B., Christensen, L. & Brattegaard, K. Determination of CD4+ and CD8+ lymphocytes with the cytosphere assay: a comparative study with flow cytometry and the immunoalkaline phosphatase method. Clin. Immunol. Immunopathol. 76, 135–141 (1995). 

5. Lutwama, F. et al Evaluation of Dynabeads and Cytospheres Compared with flow cytometry to enumerate CD4+ T cells in HIV‐infected Ugandans on antiretroviral therapy. J Acuir Immune Defic Syndr 48(3) July 1, 2008 

6. Idigbe EO et al Comparison of Dynabeads and Capcellia methods with FACScount for the estimation of CD4 T‐lymphocyte levels in HIV/AIDS patients in Lagos, Nigeria. East African Medical Journal, April 2006, vol./is. 83/4(105‐11) 

7. Balakrishnan, P. et al. A reliable and inexpensive EasyCD4 assay for monitoring HIV infected individuals in resource‐limited settings. J. Acquir. Immune Defic. Syndr. 43, 23–26 (2006). 

8. Spacek, L.A . et al. Evaluation of a low‐cost method, the Guava EasyCD4 assay, to enumerate CD4‐ positive lymphocyte counts in HIVinfected patients in the United States and Uganda. J. Acquir. Immune Defic. Syndr. 41, 607–610 (2006). 

9. Thakar, M. R., Kumar, B. K., Mahajan, B. A., Mehendale, S. M. & Paranjape, R. S. Comparison of capillary based microflurometric assay for CD4+ T cell count estimation with dual platform Flow cytometry. AIDS Res. Ther. 3, 26 (2006).  

10. Zijenah, L. S. et al. Affordable flow cytometry for enumeration of absolute CD4+ T lymphocytes to identify subtype C HIV‐1 infected adults requiring antiretroviral therapy (ART) and monitoring response to ART in a resource limited setting. J. Transl. Med. 4, 33(2006).  

11. Lynen L; Teav S; Vereecken C; De Munter P; An S; Jacques G; Kestens L Validation of primary CD4 gating as an affordable strategy for absolute CD4 counting in Cambodia Journal of Acquired Immune Deficiency Syndromes: JAIDS, October 2006, vol./is. 43/2(179‐85) 

12. Briggs C et al Measurement of CD4 T cells in point‐of‐care settings with the Sysmex pocH‐100i haematology analyser Int J Lab Haem 2007 

13. Karcher, H., Bohning, D., Downing, R., Mashate, S., Arms, G. Comparison of two alternative methods for CD4+ T‐cell determination (Coulter Manual CD4 Count and Cyflow) Against Standard Dual Platform low Cytometry in Uganda. Cytometry Part B (Clinical Cytometry) 70B:163‐169 (2006) 

14. Diagbouga, S. et al. Successful implementation of a low‐cost method for enumerating CD4+ T lymphocytes in resource‐limited settings: the ANRS 12‐26 study. AIDS 17, 2201–2208 (2003). 

15. Imade, G. E. et al. Comparison of a new, affordable flow cytometric method and the manual magnetic bead technique for CD4 Tlymphocyte counting in a northern Nigerian setting. Clin. Diagn. Lab. Immunol. 12, 224–227 (2005). 

Page 23: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  23

16. Kandathil, A. J. et al. Comparison of microcapillary cytometry technology and flow cytometry for CD4+ and CD8+ T‐cell estimation. Clin. Diagn. Lab. Immunol. 12, 1006–1009 (2005). 

17. Pattanapanyasat K; Phuang‐Ngern Y; Sukapirom K; Lerdwana S; Thepthai C; Tassaneetrithep B Comparison of 5 flow cytometric immunophenotyping systems for absolute CD4+ T‐lymphocyte counts in HIV‐1‐infected patients living in resource‐limited settings. Journal of Acquired Immune Deficiency Syndromes: JAIDS, December 2008, vol./is. 49/4(339‐47) 

18. Fryland, M. et al. The Partec CyFlow Counter could provide an option for CD4+ T‐cell monitoring in the context of scaling‐up antiretroviral treatment at the district level in Malawi. Trans. R. Soc. Trop. Med. Hyg. 100, 980–985 (2006).  

19. Pattanapanyasat, K. et al. Evaluation of a new single‐parameter volumetric flow cytometer (CyFlow(green)) for enumeration of absolute CD4+ T lymphocytes in human immunodeficiency virus type 1‐ infected Thai patients. Clin. Diagn. Lab. Immunol. 12, 1416–1424 (2005). 

20. Manasa J; Musabaike H; Masimirembwa C; Burke E; Luthy R; Mudzori J Evaluation of the Partec flow cytometer against the BD FACSCalibur system for monitoring immune responses of human immunodeficiency virus‐infected patients in Zimbabwe. Clinical & Vaccine Immunology: CVI, March 2007, vol./is. 14/3(293‐8) 

21. Greve B. et al A new no‐lyse, no‐wash flow‐cytometric method for the determination of CD4 T cells in blood samples Transfusion Medicine and Hemotherapy, 2003, vol./is. 30/1(8‐13)  

22. Strauss, K. et al. Performance evaluation of the FACSCount System: a dedicated system for clinical cellular analysis. Cytometry 26, 52– 59 (1996). 

23. Lopez, A. et al. Enumeration of CD4(+) T‐cells in the peripheral blood of HIV‐infected patients: an interlaboratory study of the FACSCount system. Cytometry 38, 231–237 (1999). 

24. Johnson D; Hirschkorn D; Busch MP Evaluation of four alternative methodologies for determination of absolute CD4+ lymphocyte counts. The National Heart, Lung, and Blood Institute Retrovirus Epidemiology Donor Study. Journal of Acquired Immune Deficiency Syndromes & Human Retrovirology, December 1995, vol./is. 10/5(522‐30) 

25. Nicholson J.K.A.; Velleca W.M.; Jubert S.; Green T.A.; Bryan L. Evaluation of alternative CD4 technologies for the enumeration of CD4 lymphocytes Journal of Immunological Methods, 1994, vol./is. 177/1‐2(43‐54) 

26. Schnizlein‐Bick, C. T., Spritzler, J., Wilkening, C. L., Nicholson, J. K. & O'Gorman, M. R. Evaluation of TruCount absolute‐count tubes for determining CD4 and CD8 cell numbers in human immunodeficiency virus‐positive adults. Site Investigators and The NIAID DAIDS New Technologies Evaluation Group. Clin. Diagn. Lab. Immunol. 7, 336–343 (2000). 

27. Nicholson JK; Stein D; Mui T; Mack R; Hubbard M; Denny T Evaluation of a method for counting absolute numbers of cells with a flow cytometer. Clinical & Diagnostic Laboratory Immunology, May 1997, vol./is. 4/3(309‐13) 

28. Reimann, K. A. et al. Multisite comparison of CD4 and CD8 Tlymphocyte counting by singleversus multiple‐platform methodologies: evaluation of Beckman Coulter flow‐count fluorospheres and the tetraONE system.The NIAID DAIDS New Technologies Evaluation Group. Clin. Diagn. Lab. Immunol. 7, 344–351 (2000).  

29. Denny, T. et al. A North American multi‐laboratory study of CD4 counts using Flow Cytometric PanLeukogating (PLG): a NIAIDDAIDS Immunology Quality Assessment Program Study. Cytometry B Clin. Cytom. 74B (Suppl. 1) S52–64 (2008). 

Page 24: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  24

30. Pattanapanyasat, K. et al. A multicenter evaluation of the PanLeucogating method and the use of generic monoclonal antibody reagents for CD4 enumeration in HIV‐infected patients in Thailand. Cytometry B Clin. Cytom. 65B, 29– 36 (2005).  

31. Lyamuya, E. F. et al. Evaluation of the FACScount, TRAx CD4 and Dynabeads methods for CD4 lymphocyte determination. J. Immunol. Methods 195, 103–112 (1996). 

32. Didier, J. M. et al. Comparative assessment of five alternative methods for CD4+ T‐lymphocyte enumeration for implementation in developing countries. J. Acquir. Immune Defic. Syndr. 26, 193–195 (2001). 

33. Pattanapanyasat, K. et al. Evaluation of a single‐platform microcapillary flow cytometer for enumeration of absolute CD4+ T‐lymphocyte counts in HIV‐1 infected Thai patients. Cytometry B Clin. Cytom. 72B, 387–396 (2007).  

34. Dieye, T. N. et al. Absolute CD4 Tcell counting in resource‐poor settings: direct volumetric measurements versus bead‐based clinical flow cytometry instruments. J. Acquir. Immune Defic. Syndr. 39, 32–37 (2005).   

35. Cassens, U. et al. Simplified volumetric flow cytometry allows feasible and accurate determination of CD4 T lymphocytes in immunodeficient patients worldwide. Antiviral Ther. 9, 395–405 (2004).  

36. Young NL et al. Clinical field site evaluation of the FACSCount for absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell count determinations in Thailand. Clinical & Diagnostic Laboratory Immunology, November 1997, vol./is. 4/6(783‐6) 

37. Higgins J. et al Evaluation of a single‐platform technology for lymphocyte immunophenotyping Clinical and Vaccine Immunology, October 2007, vol./is. 14/10(1342‐1348) 

38. Jeganathan S; Bansal M; Smith DE; Gold J Comparison of different methodologies for CD4 estimation in a clinical setting. HIV Medicine, April 2008, vol./is. 9/4(192‐5)  

39. Storie, I. et al. Perfect count: a novel approach for the single platform enumeration of absolute CD4+ T‐lymphocytes. Cytometry B Clin. Cytom. 57, 47–52 (2004).  

40. Storie, I. et al. Flow rate calibration II: a clinical evaluation study using PanLeucoGating as a single‐platform protocol. Cytometry B Clin. Cytom. 55B, 8–13 (2003).  

41. Srithanaviboonchai K et al Novel low cost assay for the monitoring of CD4 counts in HIV infected individuals J Acquir Immune Defic Syndr 2008;47:135–139 

42. Glencross, D., Scott, L. E., Jani, I. V., Barnett, D. & Janossy, G. CD45‐ assisted PanLeucogating for accurate, cost‐effective dual‐platform CD4+ T‐cell enumeration. Cytometry 50, 69–77 (2002). 

43. Sippy‐Chatrani N et al Performance of the Panleucogating protocol for C4 T cell enumeration in an HIV dedicated laboratory facility in Barbados Cytometry Part B (Clinical cytometry) 74B (Suppl 1):S65‐S68 (2008) 

44. Nunez, R., Flow cytometry: principles and instrumentation. Curr Issues Mol Biol, 2001. 3(2): p. 39‐45. 

45. Stevens, W., et al., Evaluating new CD4 enumeration technologies for resource‐constrained countries. Nature Reviews Microbiology, 2008. 6(11): p. S29‐S38. 

46. Bland, J. M. & Altman, D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1, 307–310 (1986). 

47. Systematic Reviews. CRD's guidance for undertaking reviews in health care.Centre for Reviews and Dissemination, University of York, 2008 

Page 25: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  25

48. www.stard‐statement.org 

49. Redd, A.D., et al., T‐cell enumeration from dried blood spots by quantifying rearranged T‐cell receptor‐ β genes, J. Immunol. Methods (2010), doi: 10.1016/j.jim.2010.01.008. 

50. Peter T, Badrichani A, Wu E, Freeman R, Ncube B, Ariki F, Daily J, Shimada Y, Murtagh M. Challenges in implementing CD4 testing in resource‐limited settings. Cytometry B Clin Cytom. 2008;74 Suppl 1:S123‐30. 

51. Malone, Joseph L.; Simms, Thomas E.; Gray, Gregory C.; Wagner, Kenneth F.; Burge, J. Robert; Burke, Donald S. Sources of Variability in Repeated T-Helper Lymphocyte Counts from Human Immunodeficiency Virus Type 1-Infected Patients: Total Lymphocyte Count Fluctuations and Diurnal Cycle Are Important Journal of Acquired Immune Deficiency Syndromes. 3(2):144-151, February 1990

52. Fei DTW, Paxton H, Chen AB. Difficulties in precise quantitation of CD4+ T lymphocytes for clinical trials: a review. Biologicals.1993;21:221-231.

53. Raboud JM, Haley L, Montaner JS, Murphy C, Januszewska M, Schechter MT Quantification of the variation due to laboratory and physiologic sources in CD4 lymphocyte counts of clinically stable HIV-infected individuals. J Acquir Immune Defic Syndr Hum Retrovirol. 1995;10 Suppl 2:S67-73.

54. Giorgi, J. V. Characterization of T lymphocyte subset alterations by flow cytometry in HIV disease. Ann. New York Acad. Sci. 677, 126–137 (1993)

55. Sterling TR, Chaisson RE, Moore RD. HIV-1 RNA, CD4 T-lymphocytes, and clinical response to highly active antiretroviral therapy. AIDS 2001; 15: 2251–2257.

56. Wolbers M, Babiker A, Sabin C, Young J, Dorrucci M, et al. (2010) Pretreatment CD4 Cell Slope and Progression to AIDS or Death in HIV-Infected PatientsInitiating Antiretroviral Therapy—The CASCADE Collaboration: A Collaboration of 23 Cohort Studies. PLoS Med 7(2): e1000239. doi:10.1371/journal.pmed.1000239

57. Tayler-Smith K, Zachariah R, Massaquoi M, Manzi M, Pasulani O, van den Akker T, Bemelmans M, Bauernfeind A, Mwagomba B, Harries AD. Unacceptable attrition among WHO stages 1 and 2 patients in a hospital-based setting in rural Malawi: can we retain such patients within the general health system? Trans R Soc Trop Med Hyg. 2010 May;104(5):313-9. Epub 2010 Feb 6.

58. World Health Organisation Rapid advice: antiretroviral therapy for HIV infection in adults and adolescents -November 2009

59. Coutinho A, Mermin J, Ekwaru J, Were W, R Bunnell, Kaharuza F et al. Utility of Routine Viral Load, CD4 Cell Count, and Clinical Monitoring among HIV-infected Adults in Uganda: A Randomized Trial. 15th Conference on Retroviruses and Opportunistic Infections; 2008; Boston, Abstract 125. [ Available from http://www.retroconference.org/2008/ Abstracts/30881.htm] 60. Mugyenyi P, Walker S Hakim J, Munderi P , Gibb D , Kityo C , et al. Impact of routine laboratory monitoring over 5 years after antiretroviral therapy (ART) initiation on clinical disease progression of HIV-infected African adults: the DART trial final results 2009. 5th IAS Conference on HIV Pathogenesis, Treatment and Prevention; 2009; Cape Town. Abstract TUSS 102 [Available from http://www.ias2009.org/pag/Abstracts.aspx?AID=3807]

Page 26: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  26

APPENDIX A: SEARCH STRATEGY AND SEARCH TERMS 

 Search strategy 

Databases: MEDLINE and EMBASE 

Reference list of Nature supplement: Evaluating diagnostics: the CD4 guide. Nature Review Microbiology November 2008 Vol 6 No 11 

Contacting experts (advisory group) 

Search terms 

Search terms to include: “CD4”, “technolog*”, “methodolog*”, “techn*”, “method*”, “test”, “evaluation”, “validation”, “accuracy”, “comparison”, “efficacy”, “performance”, “reproducibility”, “precision”, “flow AND cytometry”  

Subject headings to include “CD4 antigen”, “Antigens, CD4”, “CD4 lymphocyte count”, “CD4‐positive T‐lymphocytes”, “technology”, “methodology”, “technique”, “evaluation”, “clinical evaluation”, “economic evaluation”, “evaluation research”, “evaluation and follow up”, “instrument validation”, “validation process”, “validation study”, “evaluation studies as topic”, “validation studies”, “flow cytometry” 

Date 

1990‐present 

Language 

English only 

Publication type/status 

Only published work in peer reviewed journals 

 

Page 27: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  27

 APPENDIX B: ALGORITHM FOR STUDY INCLUSION  

Page 28: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 

  28    

433 titles +/‐ abstracts reviewed for relevance  

88 articles identified as potentially relevant (ie evaluations of CD4 enumeration technology) 

345 excluded as not relevant to study question ie not evaluations of CD4 technology  

1 study excluded as did not use flow cytometry as reference standard 

1 study excluded as paediatric only 

1 study excluded as not English language 

14 studies excluded as index test not commercially available at current time 

7 conference abstracts excluded 

40 studies accepted for inclusion 

1 study excluded as evaluating complex flow cytometer (FACSCanto) 

4 studies excluded as EQA studies, not evaluations 

5 studies excluded as not evaluations of new technology (ie old technology or minor modification of existing tech)

13 studies excluded as not evaluations or not evaluations of CD4 technology 

1 study excluded as used outdated FC as reference method 

3 studies  accepted for inclusion 

4 other studies contributed by advisory group

1 excluded as not commercially available 

43 studies accepted for inclusion 

Page 29: Technical Brief on CD4 Technologies (May 2010) and the patient does not have the disease). Among specimens with a CD4 count (obtained using the reference technology) which is greater

 